Metrological Guidance for the Calibration of Laboratory Glassware

Open access
Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States), Mackrory, Andrew, USDOE National Nuclear Security Administration (NNSA)·2024·OSTI / U.S. Department of Energy

Abstract

This report provides technical guidance for the calibration of laboratory glassware to help the practitioner achieve traceability to the International System of Units and meet customer quality requirements.The discussion of traceability uses the National Institute of Standards and Technology's seven essential elements of traceability as a framework.The guidance also includes how to determine when calibration is necessary, a discussion of the relationship between tolerances and measurement uncertainty, practical tips, and helpful references.This guidance is intended for United States Department of Energy Contractor Standards Laboratories, but many of the principles and information provided are broadly applicable to other calibration entities and other calibration fields.

Integrity signals

Registry

Crossref · 10.2172/2429959

Cites

0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

12 of 67 cited sources checked · 3 supported · 0 flagged · accuracy not yet measured

  • Premisecitation
    § 2. SEVEN ESSENTIAL ELEMENTS TO ACHIEVE METROLOGICAL TRACEABILITY

    NIST published GMP 13 in Reference [14] for ensuring metrological traceability.

    “NIST published GMP 13 in Reference [14] for ensuring metrological traceability.”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [14] 14. A. Thompson and B. N. Taylor, NIST Special Publication 811: Guide for the Use of the International System of Units (SI), 2nd Printing (November 2008) ed., National Institute of Standards and Technology, U.S. Department of Commerce, 2008. source

      The Abstract identifies this source as NIST’s Guide for the Use of the International System of Units (SI), not as GMP 13. Searches of our copy for “GMP 13” and “metrological traceability” found no passage addressing whether NIST published GMP 13 or its purpose. (LLM, conf=0.75)

      Experimental — accuracy not yet measured
  • Premisecitation
    § 3.7.2. The Relationship Between Tolerances and Measurement Uncertainty

    The uncertainty contributors related to use of the glassware are additive to those reported by the calibration laboratory.

    “The uncertainty contributors related to use of the glassware are additive to those reported by the calibration laboratory [16].”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites [16] 16. G. L. Harris and V. R. Miller, NISTIR 7383 Selected Procedures for Volumetric Calibrations (2019 Ed), National Institute of Standards and Technology, U.S. Department of Commerce, May 2019. source
  • Premisecitation
    § 3.7.2. The Relationship Between Tolerances and Measurement Uncertainty

    ISO/IEC 17025 requires that unless a decision rule is inherent in the standard being used, the decision rule be clearly communicated to and agreed upon with the customer.

    “ISO/IEC 17025 requires that unless a decision rule is inherent in the standard being used, the decision rule be clearly communicated to and agreed upon with the customer [13].”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites [13] 13. G. L. Harris, NISTIR 6969 Selected Laboratory and Measurement Practices and Procedures to Support Basic Mass Calibrations (2019 Ed), National Institute of Standards and Technology, U.S. Department of Commerce, May 2019. source
  • Premisecitation
    § 3.7.2. The Relationship Between Tolerances and Measurement Uncertainty

    ISO/IEC 17025 requires that if a calibration certificate makes a compliance statement (e.g., compliance with a tolerance), the decision rule must also be documented.

    “ISO/IEC 17025 requires that if a calibration certificate makes a compliance statement (e.g., compliance with a tolerance), the decision rule must also be documented.”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [13] 13. G. L. Harris, NISTIR 6969 Selected Laboratory and Measurement Practices and Procedures to Support Basic Mass Calibrations (2019 Ed), National Institute of Standards and Technology, U.S. Department of Commerce, May 2019. source
  • Findingnone given
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    Gravimetric measurements have the lowest achievable measurement uncertainty for the measurement of a substance's quantity.

    “Gravimetric measurements have the lowest achievable measurement uncertainty for the measurement of a substance's quantity.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim. 99
  • May combine two claims. 56
  • Classified as probably not a substantial claim.

    Premisecitation
    § EXECUTIVE SUMMARY

    Metrological traceability requires seven essential elements (from NIST's GMP 13).

    “metrological traceability requires seven essential elements (from NIST's GMP 13: Good Measurement Practice for Ensuring Metrological Traceability) that are discussed in the context of glassware calibration:”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [14] 14. A. Thompson and B. N. Taylor, NIST Special Publication 811: Guide for the Use of the International System of Units (SI), 2nd Printing (November 2008) ed., National Institute of Standards and Technology, U.S. Department of Commerce, 2008. source

      The cited copy is a Guide for the Use of the International System of Units (SI); its §1.2 outline describes guidance on units, not GMP 13. Searches for GMP 13, seven essential elements of metrological traceability, and glassware calibration found no passage addressing the claim. (LLM, conf=0.76)

      Experimental — accuracy not yet measured
  • Classified as probably not a substantial claim.

    Premisecitation
    § EXECUTIVE SUMMARY

    Calibration ensures measurements are metrologically traceable to the International System of Units (SI).

    “Calibration ensures measurements are metrologically traceable to the International System of Units (SI) which is essential to ensuring quality, managing risk, and meeting customer requirements.”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [14] 14. A. Thompson and B. N. Taylor, NIST Special Publication 811: Guide for the Use of the International System of Units (SI), 2nd Printing (November 2008) ed., National Institute of Standards and Technology, U.S. Department of Commerce, 2008. source

      The Introduction (1. Purpose of Guide; page not provided) says the guide assists with using SI when reporting measurement results, but does not say calibration ensures metrological traceability to SI. Searches for “calibration” and “metrological traceability” found no passage addressing that relationship. (LLM, conf=0.76)

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § 1.4. The Need for Critical Thinking

    CCLs with accreditation to ISO/IEC 17025 [13] can be easily approved as calibration suppliers, or nonaccredited CCLs can be audited to confirm compliance in the absence of third-party accreditation.

    “Generally, CCLs with accreditation to ISO/IEC 17025 [13] can be easily approved as calibration suppliers 7 , or nonaccredited CCLs can be audited to confirm compliance in the absence of third-party accreditation.”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites [13] 13. G. L. Harris, NISTIR 6969 Selected Laboratory and Measurement Practices and Procedures to Support Basic Mass Calibrations (2019 Ed), National Institute of Standards and Technology, U.S. Department of Commerce, May 2019. source

      Appendix D (page unavailable) treats accreditation and evaluation of calibration providers as relevant to supplier assessment. But it calls for checking an accredited laboratory’s scope and measurement uncertainty, not easy approval based on ISO/IEC 17025 accreditation alone. The cited passages describe technical audits of traceability evidence, not audits of nonaccredited CCLs to confirm their compliance. (LLM, conf=0.82)

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § 1.4. The Need for Critical Thinking

    It is the responsibility of the calibration lab to provide the supporting evidence of traceability, but it is the responsibility of the customer to assess the validity of the traceability claim.

    “It is the responsibility of the calibration lab to provide the supporting evidence of traceability, but it is the responsibility of the customer to assess the validity of the traceability claim (see Good Measurement Practice (GMP) 13 in Reference [14]).”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites [14] 14. A. Thompson and B. N. Taylor, NIST Special Publication 811: Guide for the Use of the International System of Units (SI), 2nd Printing (November 2008) ed., National Institute of Standards and Technology, U.S. Department of Commerce, 2008. source

      The passages from C.8 Uncertainty in measurement and the References discuss measurement uncertainty, not responsibility for substantiating or assessing traceability. Searches for the calibration laboratory’s supporting evidence and the customer’s assessment of a traceability claim found no passage addressing either responsibility in our copy. (LLM, conf=0.75)

      Experimental — accuracy not yet measured
  • Classified as probably not a substantial claim.

    Premisecitation
    § 2. SEVEN ESSENTIAL ELEMENTS TO ACHIEVE METROLOGICAL TRACEABILITY

    In GMP 13, measurement traceability is characterized by seven essential elements.

    “In GMP 13, measurement traceability is characterized by seven essential elements that are discussed below in the context of glassware volume calibration.”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [14] 14. A. Thompson and B. N. Taylor, NIST Special Publication 811: Guide for the Use of the International System of Units (SI), 2nd Printing (November 2008) ed., National Institute of Standards and Technology, U.S. Department of Commerce, 2008. source

      Our copy is a guide to using SI units. Section 2.1 discusses essential data, and section C.8 discusses measurement uncertainty, but neither describes GMP 13 or seven elements of measurement traceability. Searches for GMP 13 and for seven traceability elements in glassware calibration found no relevant passage. (LLM, conf=0.76)

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Element Five: Documented Measurement Procedure

    Some information not contained in the NIST documents are tolerances for various types of glassware.

    “Some information not contained in the NIST documents are tolerances for various types of glassware.”
    Quote in the paperNot a substantial claim1 of 2 sources checked
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites [14] 14. A. Thompson and B. N. Taylor, NIST Special Publication 811: Guide for the Use of the International System of Units (SI), 2nd Printing (November 2008) ed., National Institute of Standards and Technology, U.S. Department of Commerce, 2008. source

      Section 8.2 (Volume) discusses units of volume, but does not address tolerances for glassware. Searches for glassware tolerances and related calibration terms found no passage establishing whether that information is absent from the NIST documents. (LLM, conf=0.70)

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Element Five: Documented Measurement Procedure

    NISTIR 7383 and NISTIR 6969 contain other necessary SOPs, Good Laboratory Practices, and Good Measurement Practices that are essential for a complete volume calibration capability.

    “Besides SOP 14, NISTIR 7383 [17] and NISTIR 6969 [14] contain other necessary SOPs, Good Laboratory Practices (GLPs), and GMPs that are essential for a complete volume calibration capability.”
    Quote in the paperNot a substantial claim1 of 2 sources checked
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites [14] 14. A. Thompson and B. N. Taylor, NIST Special Publication 811: Guide for the Use of the International System of Units (SI), 2nd Printing (November 2008) ed., National Institute of Standards and Technology, U.S. Department of Commerce, 2008. source

      Our copy is a guide to SI usage: section 2 addresses NIST policy on SI units, and section 1.2 outlines the guide’s contents. Searches for NISTIR 7383, NISTIR 6969, SOP 14, and Good Laboratory and Measurement Practices found no passage addressing what those reports contain or what a complete volume calibration capability requires. (LLM, conf=0.76)

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § 3.7.2. The Relationship Between Tolerances and Measurement Uncertainty

    If a tolerance is assigned to an item, the calibration certificate is required to state both the tolerance assigned and the measurement uncertainty of the calibration.

    “If (as is common) a tolerance is assigned to an item, the calibration certificate is required to state both the tolerance assigned and the measurement uncertainty of the calibration.”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites [13] 13. G. L. Harris, NISTIR 6969 Selected Laboratory and Measurement Practices and Procedures to Support Basic Mass Calibrations (2019 Ed), National Institute of Standards and Technology, U.S. Department of Commerce, May 2019. source

      The Introduction and Criteria Guide Compliance sections require calibration certificates to report measurement uncertainty. The copy discusses applicable tolerances and conformity assessments, but the passages found do not require a certificate to state the tolerance assigned to an item; that part of the claim is unsupported. (LLM, conf=0.84)

      Experimental — accuracy not yet measured
  • May combine two claims.

    Findingnone given
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    Gravimetric measurements can cover the entire range of measured volumes; for all practical purposes, there is no upper limit.

    “Gravimetric measurements can cover the entire range of measured volumes.Data is not shown in the figure beyond 200 cm 3 , but for all practical purposes, there is no upper limit.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingcitation
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    Piston-operated pipettes are typically used over a range from 10% to 100% of their maximum capacity, but the ISO 8655 maximum permissible systematic error specification is constant across this operating range.

    “These are typically used over a range from 10 % to 100 % of their maximum capacity, but the ISO 8655 maximum permissible systematic error specification is constant across this operating range [5].”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [5] 5. ASTM E1272 -02, Standard Specification for Laboratory Glass Graduated Cylinders, Reapproved 2019.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Findingnone given
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    Graduated burets have nearly equivalent measurement uncertainty to volumetric flasks with the advantage of graduations but are limited in the range of volumes that can be measured.

    “Graduated burets have nearly equivalent measurement uncertainty to volumetric flasks with the advantage of graduations but are limited in the range of volumes that can be measured.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingnone given
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    Factors like reading the meniscus in calibration and use, and measurement uncertainty of the balances and reference weights used in the calibration all contribute to uncertainty of the volume measurement.

    “Factors like reading the meniscus in calibration and use, and measurement uncertainty of the balances and reference weights used in the calibration all contribute to uncertainty of the volume measurement.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Premisenone given
    § EXECUTIVE SUMMARY

    The most essential determining factor for whether calibration is required is whether the volume measurement has the potential to impact the quality of a product or service.

    “The most essential determining factor is whether the volume measurement has the potential to impact the quality of a product or service.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationnone given
    § EXECUTIVE SUMMARY

    Gravimetric methods are the preferred alternative for processes requiring accuracy that exceeds the capabilities of volumetric glassware.

    “Gravimetric methods are the preferred alternative for processes requiring accuracy that exceeds the capabilities of volumetric glassware.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S2 (checkable). May combine two claims.

    Premisereasoning
    § EXECUTIVE SUMMARY

    Metrological traceability is essential to ensuring quality, managing risk, and meeting customer requirements.

    “Calibration ensures measurements are metrologically traceable to the International System of Units (SI) which is essential to ensuring quality, managing risk, and meeting customer requirements.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Premisenone given
    § EXECUTIVE SUMMARY

    Measurement uncertainty or tolerances from published standards can be used to estimate achievable accuracy.

    “Measurement uncertainty or tolerances from published standards can be used to estimate achievable accuracy.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationreasoning
    § 1.4. The Need for Critical Thinking

    This introduces a risk that if the user is not familiar with the differences between the two markings, the incorrect graduation line may be used.

    “This introduces a risk that if the user is not familiar with the differences between the two markings, the incorrect graduation line may be used.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Premisenone given
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    The volumetric measurement accuracy achievable using laboratory glassware could be estimated using the tolerances published in various standards.

    “The volumetric measurement accuracy achievable using laboratory glassware could be estimated using the tolerances published in various standards (such as those referenced in Figure 2).”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § 2.4. Element Four: Documented Measurement Uncertainty

    ASTM E542 suggests correction for evaporation becomes important for volumes below 0.1 cm3.

    “ASTM E542 suggests correction for evaporation becomes important for volumes below 0.1 cm 3 [18]; however, this standard does not provide the necessary methods and corrections.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisenone given
    § 2.4. Element Four: Documented Measurement Uncertainty

    The example uncertainty budgets in References [18] and [17] are in accordance with accepted methods.

    “These examples are in accordance with the accepted methods.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S4 (based).

    Premisenone given
    § 2.1.2. Defining the Measurand to Remove Ambiguity or Provide Clarity

    20 °C is the reference temperature for laboratory glassware.

    “20 °C is the reference temperature for laboratory glassware.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S2 (checkable).

    Interpretationreasoning
    § 2.1.2. Defining the Measurand to Remove Ambiguity or Provide Clarity

    If these factors are not included in the definition of the measurand, the definition becomes ambiguous.

    “If these factors are not included in the definition of the measurand, the definition becomes ambiguous.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Interpretationreasoning
    § 1.4. The Need for Critical Thinking

    Plastic labware is generally not desired because of lower accuracy than glassware, but material compatibility may dictate the use of plastic, and this is acceptable provided the impact on measurement uncertainty is appropriately considered.

    “For example, plastic labware is generally not desired because of lower accuracy than glassware, but material compatibility may dictate the use of plastic, and this is acceptable provided the impact on measurement uncertainty is appropriately considered.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § Element Five: Documented Measurement Procedure

    In function, SOP 14 and ASTM E542 are equivalent; however, the NIST documentation tends to have more detailed procedural steps and practical advice.

    “In function, they are equivalent; however, the NIST documentation tends to have more detailed procedural steps and practical advice.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationnone given
    § 2.4. Element Four: Documented Measurement Uncertainty

    If any link in the chain is missing documented quantification of the measurement uncertainty, all dependent links will lack metrological traceability.

    “If any link in the chain is missing documented quantification of the measurement uncertainty, all dependent links will lack metrological traceability.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Interpretationnone given
    § 1.4. The Need for Critical Thinking

    Calibration effort can be reduced with longer calibration intervals, but if glassware is found out of tolerance at a calibration, there is a greater amount of work required to determine the impact on the quality of products and services that are potentially affected.

    “Calibration effort can be reduced with longer calibration intervals, but if glassware is found out of tolerance at a calibration, there is a greater amount of work required to determine the impact on the quality of products and services that are potentially affected.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Interpretationreasoning
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    If the required accuracy is beyond the capability of glassware, then alternative means of measurement are required.

    “Alternatively, if the required accuracy is beyond the capability of glassware, then alternative means of measurement are required.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § 2.2. Element Two: Unbroken Chain of Comparisons

    Each link in the traceability chain is required to have all seven essential elements of traceability.

    “This is often referred to as a traceability chain or hierarchy, and each link in the chain is required to have all seven essential elements of traceability.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim. May combine two claims.

    Premisecitation
    § Element Five: Documented Measurement Procedure

    SOP 14 in NISTIR 7383 and ASTM E542 are accepted and validated practices or procedures for the gravimetric calibration of laboratory glassware.

    “Standard Operating Procedure (SOP) 14 in NISTIR 7383 [17] and ASTM E542 [18] are accepted and validated practices or procedures and are recommended for the gravimetric calibration of laboratory glassware.”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Interpretationnone given
    § 1.4. The Need for Critical Thinking

    Gravimetric measurements are more accurate relative to calibrated glassware, but suitable analytical balances may be higher in cost and require more frequent calibration to meet the needs of a given process.

    “Gravimetric measurements are more accurate relative to calibrated glassware, but suitable analytical balances may be higher in cost and require more frequent calibration to meet the needs of a given process.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisenone given
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    If glassware is the selected method, any class of glassware is acceptable if it meets process requirements.

    “If glassware is the selected method, any class of glassware is acceptable if it meets process requirements.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Premisenone given
    § 1.2. Do You Need to Calibrate?

    If a volume measurement tolerance or uncertainty is specified, calibration is necessary to ensure that requirement is met.

    “If a volume measurement tolerance or uncertainty is specified, calibration is necessary to ensure that requirement is met, but it also must be determined whether glassware can realistically meet this requirement.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationreasoning
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    A more accurate calibration cannot lower the measurement uncertainty in use below some threshold related to the design and materials; for example, ASTM Class B glassware cannot be certified to Class A tolerances simply because of a more accurate calibration process.

    “A more accurate calibration cannot lower the measurement uncertainty in use below some threshold related to the design and materials; for example, ASTM Class B glassware cannot be certified to Class A tolerances simply because of a more accurate calibration process (see Section 3.7: Glassware Accuracy Classes).”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § 2.1.2. Defining the Measurand to Remove Ambiguity or Provide Clarity

    The liquid being used for laboratory glassware is pure air-saturated water.

    “The liquid being used (for laboratory glassware, this will be pure air-saturated water).”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Interpretationnone given
    § 2.4. Element Four: Documented Measurement Uncertainty

    Tolerances are not the same as measurement uncertainty.

    “Note that tolerances are not the same as measurement uncertainty.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § 2.4. Element Four: Documented Measurement Uncertainty

    Evaporation is a more important consideration for smaller volumes, which have a larger surface area to volume ratio, and may be negligible for many volumes.

    “Evaporation is a more important consideration for smaller volumes (with a larger surface area to volume ratio) and may be negligible for many volumes.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Element Five: Documented Measurement Procedure

    The International Organization of Legal Metrology (OIML) is another organization with relevant published standards.

    “The International Organization of Legal Metrology (OIML) 18 is another organization with relevant published standards.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § 2.4. Element Four: Documented Measurement Uncertainty

    ASTM E542-22 Table 6 suggests a range for the thermal coefficient of cubical expansion uncertainty of 5% to 10% of the value.

    “ASTM E542-22 Table 6 [18] suggests a range for the thermal coefficient of cubical expansion (CCE) uncertainty of 5 % to 10 % of the value.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • Classified as probably not a substantial claim.

    Interpretationnone given
    § Element Five: Documented Measurement Procedure

    Not all procedures have equivalent performance due to differences such as the number of measurement repetitions or ability to cancel drift in the process.

    “Not all procedures have equivalent performance due to differences such as the number of measurement repetitions or ability to cancel drift in the process.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationnone given
    § 2.4. Element Four: Documented Measurement Uncertainty

    Every measurement has its own contributing factors, such as repeatability or standard deviation of the calibration process.

    “In addition, every measurement has its own contributing factors (e.g., repeatability or standard deviation of the calibration process).”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationnone given
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    Gravimetric measurement is the most likely solution to a need for higher accuracy.

    “Gravimetric measurement is the most likely solution to a need for higher accuracy, and the process should be redefined if necessary to specify this method.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    ASTM Class B tolerances are generally double Class A tolerances.

    “ASTM Class B tolerances are generally double Class A tolerances.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § 2.4. Element Four: Documented Measurement Uncertainty

    Measurement uncertainty combines several contributing factors; for calibrating glassware these include the accepted standard deviation of the calibration process, operator effects, water temperature measurement, water density, thermal expansion of the glassware, standard masses or calibrated balance, density of standard masses, air density, drain time effects, and evaporation, listed in descending

    “Measurement uncertainty combines the result of several contributing factors, which includes the following for calibrating glassware, listed in descending order of their usual importance [18]:”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • Classified as probably not a substantial claim.

    Premisenone given
    § 2.2. Element Two: Unbroken Chain of Comparisons

    With a measurand defined so it can be traced back to accepted national, international, or intrinsic standards, the second element requires that it is traced back to these standards through a documented and unbroken chain of comparisons.

    “With a measurand defined so it can be traced back to accepted national, international, or intrinsic standards, the second element requires that it is traced back to these standards through a documented and unbroken chain of comparisons (or calibrations).”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Premisenone given
    § 2.1.1. Defining the Measurand in Terms of SI Units

    Mass traceability goes back to national mass standards, and temperature traceability goes back to intrinsic standards, such as fixed-point temperature cells.

    “Mass traceability goes back to national mass standards, and temperature traceability goes back to intrinsic standards, such as fixed-point temperature cells.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Element Five: Documented Measurement Procedure

    Tolerances for glassware are found in the relevant standards specific to the apparatus type, such as the ASTM and ISO standards referenced in Figure 2.

    “This information is found in the relevant standards specific to the apparatus type such as the ASTM and ISO standards referenced in Figure 2.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Premisereasoning
    § 2.1.2. Defining the Measurand to Remove Ambiguity or Provide Clarity

    There are multiple factors that can affect the volume of laboratory glassware, such as temperature, the method by which the meniscus is read, and others.

    “There are multiple factors that can affect the volume of laboratory glassware, such as temperature (because the glassware expands with temperature, changing the volume), the method by which the meniscus is read, and others.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation
    § 2.4. Element Four: Documented Measurement Uncertainty

    Individual staff members can be very repeatable reading a meniscus, but there may be systematic differences between staff members.

    “Individual staff members can be very repeatable reading a meniscus, but there may be systematic differences between staff members.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § 2.4. Element Four: Documented Measurement Uncertainty

    References [18] and [17] give example uncertainty budgets for calibrating glassware.

    “There is no need to start an uncertainty budget from scratch, as References [18] and [17] give example uncertainty budgets for calibrating glassware.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S2 (checkable), S4 (based). May combine two claims.

    Interpretationnone given
    § 1.4. The Need for Critical Thinking

    No audit can detect every possible issue, conditions can change between audits, and mistakes can happen.

    “However, no audit can detect every possible issue, conditions can change between audits, and mistakes can happen.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).

    Interpretationnone given
    § Element Five: Documented Measurement Procedure

    Validating a new procedure can be a substantial effort requiring, for example, a new uncertainty analysis supported by experimental data, and interlaboratory comparisons.

    “Validating a new procedure can be a substantial effort requiring, for example, a new uncertainty analysis supported by experimental data, and interlaboratory comparisons.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § 1.3. Can Volumetric Glassware Meet Process Requirements?

    Calibration is essential in establishing metrological traceability and falls under the scope of verification.

    “Calibration is essential in establishing metrological traceability and falls under the scope of verification 6 .”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationcitation
    § 2.7. Element Seven: Measurement Assurance

    A measurement assurance program ensures the validity of the measurement process and the accuracy of reference standard(s) used at the time the measurements are performed.

    “The final element of metrological traceability is a measurement assurance program 20 that ensures (1) the validity of the measurement process and (2) the accuracy of reference standard(s) used at the time the measurements are performed.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisenone given
    § 3.1. Water as a Standard Reference Material

    The currently accepted expressions to calculate the density of air-saturated water are originally sourced from Tanaka et al. [19] and are duplicated in both ASTM E542 [18] and SOP 14 [17].

    “The currently accepted expressions to calculate the density of air-saturated water are originally sourced from Tanaka et al. [19] and are duplicated in both ASTM E542 [18] and SOP 14 [17].”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites [19] 19. ASTM E438 -92, Standard Specification for Glasses in Laboratory Apparatus, Reapproved 2024.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationreasoning
    § 3.1. Water as a Standard Reference Material

    Distillation, reverse osmosis, ion exchange, or combinations of these processes, followed by filtration, are suitable methods to achieve the required water quality.

    “Distillation, reverse osmosis, ion exchange, or combinations of these processes, followed by filtration, are suitable methods.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § Element Five: Documented Measurement Procedure

    ASTM E542 contains many best practices for use of volumetric glassware, for example, pipetting onto wetted walls; using specified and consistent wait times when emptying; avoiding liquid splashes above the graduation line; and using a burette stopcock fully open until close to the target volume, then slowing the flow.

    “ASTM E542 contains many best practices for use of volumetric glassware, for example, pipetting onto wetted walls; using specified and consistent wait times when emptying; avoiding liquid splashes above the graduation line; using a burette stopcock fully open until close to the target volume target, then slowing the flow; etc.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • May combine two claims.

    Premisecitation
    § 3.1. Water as a Standard Reference Material

    The density-meter equipment is likely to be more costly than controlling water quality and will itself require calibration.

    “However, this equipment is likely to be more costly than controlling the water quality and will itself require calibration [3].”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [3] 3. ASTM E288 -10, Standard Specification for Laboratory Glass Volumetric Flasks, Reapproved 2017.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Interpretationnone given
    § 2.7. Element Seven: Measurement Assurance

    A control chart of check standard measurements can be interpreted to provide assurance that significant changes in systematic error or process variability have not occurred (an internal check).

    “A control chart of check standard measurements can be interpreted to provide assurance that significant changes in systematic error or process variability have not occurred (an internal check).”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationcitation
    § 3.1. Water as a Standard Reference Material

    The density-meter approach may cause contamination of the customer's glassware and is therefore not recommended.

    “This approach may also cause contamination of the customer's glasswarefoot_13 and is therefore not recommended.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim. May combine two claims.

    Interpretationnone given
    § 2.7. Element Seven: Measurement Assurance

    A different number of replicates is a different process, and a standard deviation chart, range chart, and control chart can only be used for data coming from the same process.

    “A different number of replicates is a different process, and a standard deviation chart, range chart, and control chart can only be used for data coming from the same process.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationnone given
    § 2.7. Element Seven: Measurement Assurance

    Check standards used for control charts should ideally have an independently measured reference value that can be compared to the mean value of the control chart data to ensure agreement within measurement uncertainty (an external check).

    “In addition, check standards used for control charts should ideally have an independently measured reference value (i.e., a measurement value from an independent calibration laboratory) that can be compared to the mean value of the control chart data to ensure agreement within measurement uncertainty (an external check).”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § Element Five: Documented Measurement Procedure

    Both GMP 3 and ASTM E542 have two options for reading the meniscus.

    “Both GMP 3 and ASTM E542 have two options for reading the meniscus.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § 3.7.2. The Relationship Between Tolerances and Measurement Uncertainty

    When this approach is used within the DOE, this fraction is 1/4 or 25 %.

    “When this approach is used within the DOE, this fraction is 1/4 or 25 % [2].”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [2] 2. T. Coleman and G. Harris, "Important Technical Guidance on Glassware," August 2005. [Online]. Available: https://www.nist.gov/pml/owm/weights-and-measures-newsletter- archives-field-standards. [Accessed 2 January 2024].

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationnone given
    § Element Five: Documented Measurement Procedure

    The choice of meniscus reading method may depend on how far around the vessel perimeter the graduation lines extend and whether the liquid is opaque or transparent.

    “The choice of method may depend on how far around the vessel perimeter the graduation lines extend and whether the liquid is opaque or transparent.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationnone given
    § Element Five: Documented Measurement Procedure

    Gravimetric calibration has benefits including having fewer procedures and calibration standards to maintain, and a shorter traceability chain with associated lower measurement uncertainty.

    “The benefits include having fewer procedures and calibration standards to maintain, and a shorter traceability chain with associated lower measurement uncertainty.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Element Five: Documented Measurement Procedure

    NISTIR 7383 contains a procedure for calibrating glassware by a volume transfer method.

    “NISTIR 7383 contains a procedure for calibrating glassware by a volume transfer method; however, it is simpler for a glassware calibration lab to calibrate all glassware gravimetrically.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Element Five: Documented Measurement Procedure

    ASTM E542 Section 17.7 notes that when using a calibrated balance under conditions that differ significantly from the reference conditions for conventional mass, additional uncertainty evaluation should be considered.

    “ASTM E542 Section 17.7 notes that when using a calibrated balance under conditions that differ “significantly” from the reference conditions for conventional mass (i.e., 20 °C, 8 g⁄cm 3 reference density, 0.0012 g⁄cm 3 reference air), additional uncertainty evaluation should be considered [18].”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • May combine two claims.

    Interpretationreasoning
    § 3.1. Water as a Standard Reference Material

    Tap water and filtration alone are inadequate because the water purity requirements require removal of both solid and chemical contaminants.

    “Tap water and filtration alone are, however, inadequate, because the water purity requirements are such that both solid and chemical contaminants must be removed.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisenone given
    § 3.7.2. The Relationship Between Tolerances and Measurement Uncertainty

    A widely used practice when calibrating items and determining compliance with a tolerance is to ensure that the uncertainty of the calibration process (expressed at a 95 % level of confidence) is less than or equal to some fraction of the tolerance being tested.

    “A widely used practice when calibrating items and determining compliance with a tolerance is to ensure that the uncertainty of the calibration process (expressed at a 95 % level of confidence) is less than or equal to some fraction of the tolerance being tested.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § 3.1. Water as a Standard Reference Material

    SOP 14 states the requirements for the water thermometer as resolution and uncertainty less than 0.1 °C.

    “SOP 14 states the requirements for the water thermometer as “resolution and uncertainty less than 0.1 °C”, and ASTM E542 requires resolution of 0.01°C and uncertainty (k=2) of 0.1 °C.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S4 (based). May combine two claims.

    Interpretationnone given
    § 3.7.3. Glassware That Does Not Meet Tolerances

    Assigning a correction factor relies on the end user remembering to apply the correction factor and doing it correctly.

    “Assigning a correction factor relies on the end user remembering to apply the correction factor and doing it correctly.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Premisecitation
    § Element Five: Documented Measurement Procedure

    In ASTM E542 Section 17.7, "significantly" is not quantified.

    ““Significantly” is not quantified, so to avoid the additional evaluation, if there is a question as to whether the conditions are sufficiently close to conventional mass reference conditions, mass standards should be used as the reference rather than a calibrated balance.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • May combine two claims.

    Interpretationreasoning
    § 3.1. Water as a Standard Reference Material

    The water serves as a standard reference material, and therefore the purity or quality of the water is important in gravimetric calibrations.

    “The water serves as a standard reference material, and therefore the purity or quality of the water is important in gravimetric calibrations (pure water is not required for volume transfer calibrations, but gravimetric calibrations are recommended over volume transfer methods -see Section 2.5).”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S2 (checkable).

    Interpretationreasoning
    § Element Five: Documented Measurement Procedure

    It is simpler for a glassware calibration lab to calibrate all glassware gravimetrically.

    “NISTIR 7383 contains a procedure for calibrating glassware by a volume transfer method; however, it is simpler for a glassware calibration lab to calibrate all glassware gravimetrically.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § 3.1. Water as a Standard Reference Material

    GLP 10 states that electrical conductivity of ≤ 5 μS/cm (electrical resistivity of ≥ 0.2 MΩ•cm) is recommended as adequate for gravimetric glassware calibrations.

    “GLP 10 in Reference [17] has substantial practical advice for the topic of water purification, and states that while not directly correlated to density, electrical conductivity of ≤ 5 µS/cm (electrical resistivity of ≥ 0.2 MΩ•cm) is recommended as adequate for gravimetric glassware calibrations.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § 3.7.2. The Relationship Between Tolerances and Measurement Uncertainty

    If the ratio of calibration uncertainty to tolerance cannot be met, guardbanding is an option where, provided the tolerance remains greater than 1.5 times the uncertainty, the limits that must be met are tightened when making the compliance decision.

    “If this ratio cannot be met, there are options such as guardbanding, where provided the tolerance remains greater than 1.5 times the uncertainty, the limits that must be met are tightened when making the compliance decision [21].”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [21]

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisenone given
    § 2.6. Element Six: Accredited Technical Competence

    Suitable evidence of technical competence includes appropriate training records, results of proficiency tests or interlaboratory comparisons, and accreditation to ISO/IEC 17025 from a recognized accreditation body.

    “Suitable evidence includes appropriate training records, results of proficiency tests or interlaboratory comparisons, and accreditation to ISO/IEC 17025 (the international standard for competence of testing and calibration laboratories) from a recognized accreditation body.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationreasoning
    § 3.7.2. The Relationship Between Tolerances and Measurement Uncertainty

    The performance achievable in the use of glassware is never the same as the uncertainty in calibration because some of the same contributors (e.g., reading the meniscus) occur in both calibration and use.

    “The performance achievable in the use of glassware is never the same as the uncertainty in calibration because some of the same contributors (e.g., reading the meniscus) occur in both calibration and use.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisenone given
    § 2.6. Element Six: Accredited Technical Competence

    Generally, a recognized accreditation body means an accreditation body that is an ILAC signatory.

    “Generally, this means an accreditation body that is an ILACfoot_11 signatory.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § Element Five: Documented Measurement Procedure

    Examples of specific training for gravimetric calibration include determining an effective density (not average density) when summations of weights are used and understanding that the NIST and ASTM procedures use true mass (mass in a vacuum) in the analysis, not conventional mass.

    “Examples of such specific training are determining an effective density (not average density) when summations of weights are used and understanding that the NIST and ASTM procedures use true mass (mass in a vacuum) in the analysis, not conventional mass.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [3] 3. ASTM E288 -10, Standard Specification for Laboratory Glass Volumetric Flasks, Reapproved 2017.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Classified as probably not a substantial claim.

    Interpretationreasoning
    § Element Five: Documented Measurement Procedure

    A calibration laboratory that adopts the listed procedures will still need its own internal procedure with site-specific requirements such as the specific standards to be used, the number of measurements to be made, record keeping procedures, and safety requirements.

    “While the documents listed above provide almost all the procedure documentation needed for volumetric glassware calibration, a calibration laboratory that adopts these procedures will still need their own internal procedure with site-specific requirements such as the specific standards to be used, the number of measurements to be made, record keeping procedures, safety requirements, etc.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Interpretationnone given
    § 2.7. Element Seven: Measurement Assurance

    Software quality assurance ensures that calculations performed by software are the correct calculations for the intended purpose (validation) and are performed correctly (verification).

    “Software quality assurance to ensure that calculations performed by software are the correct calculations for the intended purpose (validation) and are performed correctly (verification).”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S2 (checkable).

    Interpretationreasoning
    § 3.7.2. The Relationship Between Tolerances and Measurement Uncertainty

    The transparent reporting of the measurement results, their uncertainty (at a stated level of confidence), the assigned tolerance, and the decision rule are necessary for the customer to make their own assessment of the level of false-accept risk.

    “The transparent reporting of the measurement results, their uncertainty (at a stated level of confidence), the assigned tolerance, and the decision rule are necessary for the customer to make their own assessment of the level of false-accept risk.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationreasoning
    § Element Five: Documented Measurement Procedure

    Uncertainty analysis cannot be completed without the process standard deviation from the control charts.

    “Uncertainty analysis (Section 2.4) cannot be completed without the process standard deviation from the control charts.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationnone given
    § 2.6. Element Six: Accredited Technical Competence

    It is not unknown for an accredited laboratory to provide calibration services that fall short of ISO/IEC 17025 requirements.

    “It is not unknown for an accredited laboratory to provide calibration services that fall short of ISO/IEC 17025 requirements, so no calibration certificate should ever be blindly accepted without thorough review.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § 2.7. Element Seven: Measurement Assurance

    SOP 14 calculates the average volume from two runs, but the procedure notes that NIST calibrations generally use five replicates.

    “For the first two items, note that SOP 14 [17] calculates the average volume from two runs, but the procedure notes that NIST calibrations generally use five replicates.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § 2.7. Element Seven: Measurement Assurance

    The appendix of ASTM E542 includes example values from water and air density calculations that could be used for software quality assurance, as does GLP 10 (water density only).

    “The appendix of ASTM E542 [18] includes example values from the water and air density calculations that could be used for this purpose, as does GLP 10 [17] (water density only).”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • May combine two claims.

    Premisecitation
    § 3.1. Water as a Standard Reference Material

    ASTM E542 requires water thermometer resolution of 0.01°C and uncertainty (k=2) of 0.1°C.

    “SOP 14 states the requirements for the water thermometer as “resolution and uncertainty less than 0.1 °C”, and ASTM E542 requires resolution of 0.01°C and uncertainty (k=2) of 0.1 °C.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [18] 18. M. Tanaka, G. Girard, R. Davis, A. Peuto and N. Bignell, "Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports," Metrologia, vol. 38, pp. 301-309, 2001. source
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationnone given
    § 2.6. Element Six: Accredited Technical Competence

    It is the responsibility of the calibration laboratory to provide evidence of traceability and technical competence, but it is the customer's responsibility to assess this evidence and the validity of the traceability claim.

    “As mentioned in Section 1.4, it is the responsibility of the calibration laboratory to provide the evidence of traceability (and technical competence), but it is the customer's responsibility to assess this evidence and the validity of the traceability claim.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Interpretationnone given
    § 3.1. Water as a Standard Reference Material

    The volume of calibrated glassware is calculated as mass divided by density of the water weighed in the process.

    “The volume of calibrated glassware is calculated as mass divided by density of the water weighed in the process.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § 2.7. Element Seven: Measurement Assurance

    Control charts of check standard measurements use a check standard representative of the items being calibrated and measured by the same process.

    “Control charts of check standard measurements, where the check standard is representative of the items being calibrated and is measured by the same process.(See SOP 17 in Reference [17].)”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationnone given
    § Element Five: Documented Measurement Procedure

    Control charts cannot be populated with data without an established procedure.

    “Control charts (Section 2.7) cannot be populated with data without an established procedure.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationnone given
    § 3.7.3. Glassware That Does Not Meet Tolerances

    It is theoretically possible to assign a correction factor to glassware that does not meet the class tolerances that correspond to its markings.

    “While it is theoretically possible to assign a correction factor to glassware that does not meet the class tolerances that correspond to its markings, this is strongly discouraged.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim. May combine two claims.

    Premisecitation
    § 3.11. POVA-Type Pipettes

    ISO 8655 defines the maximum allowable systematic error as the deviation of the mean of a tenfold measurement, and the maximum permissible random error as the standard deviation of a tenfold measurement.

    “For example, ISO 8655 defines the maximum allowable systematic error as the deviation of the mean of a tenfold measurement, and the maximum permissible random error as the standard deviation of a tenfold measurement.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites ISO 8655

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Interpretationnone given
    § Measured Value

    Methods like increasing the number of repeated measurements in the calibration process and using control charts for check standards can manage and minimize the uncertainties and risks.

    “Methods like increasing the number of repeated measurements in the calibration process and using control charts for check standards can manage and minimize the uncertainties and risks.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationreasoning
    § 3.8. Contaminated Glassware (Radioactive or Other Contamination)

    Calibrating contaminated glassware inside the contaminated area transfers the contamination problem to the calibration standards, which also require periodic recalibration and closeout calibration.

    “Calibrating the glassware inside the contaminated area might appear to be a solution, but the calibration standards used are then also subject to potential contamination, and they also require periodic recalibration and closeout calibration, meaning the problem is simply transferred to another item.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationreasoning
    § 3.9. Standardize Before Use

    As the goal of the SBU process is traceability, the same seven essential elements discussed above are applicable, but may have to be modified due to the constraints that prevent conventional calibration.

    “As the goal of the SBU process is traceability, the same seven essential elements discussed above are applicable, but may have to be modified due to the constraints that prevent conventional calibration.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Interpretationnone given
    § Measured Value

    If the solution involves replacing glassware, the required analytical balance will need resolution and measurement uncertainty suitable for glassware calibration, whereas if replacing weight sets, the required balance will need higher resolution and lower uncertainty suitable for weight calibration and will undoubtedly cost more.

    “If the solution involves replacing glassware, the required analytical balance will need resolution and measurement uncertainty suitable for glassware calibration, whereas if replacing weight sets, the required balance will need higher resolution and lower uncertainty suitable for weight calibration and will undoubtedly cost more.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § 3.10. What About Plastic?

    SOP 14 recommends following manufacturer instructions for cleaning and not using cleaning agents that can attack, discolor, or swell the material.

    “SOP 14 recommends following manufacturer instructions for cleaning and not using cleaning agents that can attack, discolor, or swell the material [17].”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [17] 17. ASTM E542 -22, Standard Practice for Gravimetric Calibration of Laboratory Volumetric Instruments.

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § 3.10. What About Plastic?

    Table X1.1 in ASTM E542 provides thermal coefficient of cubical expansion (CCE) values for six different plastics commonly used for laboratory ware.

    “Table X1.1 in ASTM E542 provides thermal coefficient of cubical expansion (CCE) values for six different plastics commonly used for laboratory ware.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites ASTM E542

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § 3.11. POVA-Type Pipettes

    The MPE values in ISO 8655-2:2022 apply to the pipette and tip as a system.

    “The MPE values in ISO 8655-2:2022 apply to the pipette and tip as a system.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [22]

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • May combine two claims.

    Interpretationreasoning
    § Measured Value

    There is a trade-off in costs not only in the number of items that need to be periodically replaced but in the required calibration equipment.

    “There is a trade-off in costs not only in the number of items that need to be periodically replaced but in the required calibration equipment.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § 3.9. Standardize Before Use

    At DOE sites, SBU activities are approved by the CSL for that site.

    “At DOE sites, SBU activities are approved by the CSL for that site.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisenone given
    § 3.10. What About Plastic?

    The cubical coefficient of thermal expansion is a factor in achievable accuracy in normal use and the reason why precision grade (Class A) glassware is made from low-expansion glass.

    “As noted in Section 3.7, the cubical coefficient of thermal expansion is a factor in achievable accuracy in normal use and the reason why precision grade (Class A) glassware is made from lowexpansion glass.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisenone given
    § 3.9. Standardize Before Use

    At Sandia National Laboratories, assets designated as SBU are required to have a written standardization procedure, personnel approved to perform the procedure, PSL-certified standards traceable to the SI, etc.

    “At Sandia National Laboratories, assets designated as SBU are required to have a written standardization procedure, personnel approved to perform the procedure, PSL-certified standards traceable to the SI, etc.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Interpretationnone given
    § 3.10. What About Plastic?

    The key differences in calibrating plastic volumetric apparatus relative to glass are the achievable accuracy, calibration intervals, and cleaning processes.

    “The key differences in calibrating plastic volumetric apparatus relative to glass are the achievable accuracy, calibration intervals, and cleaning processes.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationnone given
    § 3.10. What About Plastic?

    The cleaning methods recommended for glassware may not be compatible with some or all plastics.

    “The cleaning methods recommended for glassware may not be compatible with some or all plastics.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § 3.9. Standardize Before Use

    An uncertainty analysis is also required at a level of rigor commensurate with the application.

    “An uncertainty analysis is also required at a level of rigor commensurate with the application.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisereasoning
    § 3.9. Standardize Before Use

    In lieu of calibration certificates, other records (such as laboratory notebooks) are kept to document standardization activities and results.

    “In lieu of calibration certificates, other records (such as laboratory notebooks) are kept to document standardization activities and results.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisenone given
    § 3.11. POVA-Type Pipettes

    Unlike volumetric glassware, piston-operated pipettes have moving parts and seals that require periodic maintenance, such as cleaning and lubrication, and possibly replacement.

    “Unlike volumetric glassware, piston-operated pipettes have moving parts and seals that require periodic maintenance, such as cleaning and lubrication, and possibly replacement.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationreasoning
    § Measured Value

    If an existing item has tolerance X and is replaced with a like item calibrated to a fraction X/C, the comparison gives a low false accept risk initially, but as the process repeats, the item due for replacement can only be checked to the X tolerance instead of the X/C tolerance.

    “This would give the comparison (or calibration) a low false accept risk when determining whether the existing item is still within tolerance, but as the process repeats, the item due for replacement can only be checked to the 𝑋 tolerance instead of the 𝑋/𝐶 tolerance.”
    Quote in the paperSubstantial claim
    • The authors state this as an assumption: An item freshly calibrated is very unlikely to be out of tolerance.
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Interpretationreasoning
    § 3.10. What About Plastic?

    The meniscus will also exhibit some differences relative to glass due to the hydrophobic nature of a clean and smooth plastic surface.

    “The meniscus will also exhibit some differences relative to glass due to the hydrophobic nature of a clean and smooth plastic surface.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § 3.10. What About Plastic?

    There are other specifications in the standards that could directly affect the achievable accuracy as well, such as thickness of graduation lines and maximum inside diameter at the capacity line.

    “There are other specifications in the standards that could directly affect the achievable accuracy as well, such as thickness of graduation lines and maximum inside diameter at the capacity line.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § 3.11. POVA-Type Pipettes

    The values stated in the ISO standard represent performance that can be realistically achieved, whereas manufacturer specifications may assume specific tips are used or other qualifying assumptions.

    “The values stated in the ISO standard represent performance that can be realistically achieved, whereas manufacturer specifications may assume specific tips are used or other qualifying assumptions [22].”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [22]

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based). May combine two claims.

    Interpretationnone given
    § 3.12. Ethics

    The results of calibrations affect the quality of goods and services provided using the calibrated items, and the quality of these goods and services can potentially impact public safety.

    “The results of calibrations affect the quality of goods and services provided using the calibrated items, and the quality of these goods and services can potentially impact public safety.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § 3.10. What About Plastic?

    The CCE values for plastics given in ASTM E542 range from 5.5 to 45 times higher than the value for Class A glassware and 3.67 to 30 times higher than the value for Class B glassware.

    “The CCE values for plastics given in ASTM E542 range from 5.5 to 45 times higher than the value for Class A glassware and 3.67 to 30 times higher than the value for Class B glassware.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites ASTM E542

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisenone given
    § 3.10. What About Plastic?

    The discussion on calibration intervals and the possibility of calibration intervals as long as 5 years is based on experience with glassware.

    “The discussion on calibration intervals and the possibility of calibration intervals as long as 5 years (see Section 2.3) is based on experience with glassware.”
    Quote in the paperNot a substantial claim
117 of 117
  • The paper relies on this without support.

    The DOE targets a 95% reliability rate at the end of a calibration interval.

    “The DOE targets 95 % reliability rate at the end of a calibration interval.”
  • The paper relies on this without support.

    In practice, the ISO/IEC 17025 requirement that calibration certificates not contain interval recommendations is seldom followed by commercial calibration laboratories.

    “however, in practice this requirement is seldom followed by CCLs.”
  • The authors state this as an assumption.

    After equilibration, the water is air-saturated.

    “After equilibration, it is assumed that the water is air-saturated.”
  • The paper relies on this without support.

    New glassware is not contaminated.

    “which is straightforward since new glassware is not contaminated.”
  • The authors state this as an assumption.

    An item freshly calibrated is very unlikely to be out of tolerance.

    “The whole scheme rests on the assumption that an item freshly calibrated is very unlikely to be out of tolerance, which is not a bad assumption, but still an assumption.”
5 of 5

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

  • The failure-rate citation points to the wrong source

    Minor

    The text attributes the 2005 report of glassware failure rates to reference [3], but bibliography entry 3 is an ASTM flask specification. The Coleman and Harris document dated 2005 is entry 2.

  • The GMP 13 citation identifies an SI guide instead

    Minor

    The cited bibliography entry [14] is a guide to SI units, not the NIST publication containing GMP 13. The NISTIR 6969 publication identified elsewhere in the text as containing GMPs is bibliography entry 13.

2 of 2

Cites

The publisher hasn’t deposited this paper’s reference list.

Research works printed but not linked: 1

  1. [1]1. Joint Committee for Guides in Metrology (JCGM), International vocabulary of metrology -Basic and general concepts and associated terms (VIM), 3rd Edition, 2012.

Authors
Sandia National Laboratories (SNL-NM), A., Mackrory, A., & (NNSA), U. N. N. S. A.
Affiliations
Sandia National Laboratories (US)
Venue
OSTI / U.S. Department of Energy
Published
2024-05-01
OpenAlex
W4402338260
Cited by
0
References
0
Type
report
Language
en
Open access
Locations
OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) (submittedVersion)
Funders
  • USDOE National Nuclear Security Administration (NNSA)
  • DOE (NA0003525)
Topics
  • Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Scientific Measurement and Uncertainty Evaluation
  • Physical Sciences › Engineering › Mechanical Engineering › Advanced Measurement and Metrology Techniques
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
Keywords
traceability, SI units, measurement uncertainty, tolerance, calibration requirements, non-functional requirements

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-11.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-11.

Cited sources are checked by gpt-6-sol.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review