Time in Tight Glucose Range in Type 1 Diabetes: Predictive Factors and Achievable Targets in Real-World Users of the MiniMed 780G System

Open access
Castañeda, Javier, Arrieta, Arcelia, van den Heuvel, Tim, Battelino, Tadej, Cohen, Ohad·2024·Diabetes Care

Abstract

OBJECTIVE: We studied time in tight range (TITR) (70–140 mg/dL) in real-world users of the MiniMed 780G system (MM780G). RESEARCH DESIGN AND METHODS: CareLink Personal data were extracted (August 2020 to December 2022) to examine TITR and its relationship with time in range (TIR; 70–180 mg/dL), factors predicting higher TITR, and which TITR target is a reasonable treatment goal. RESULTS: The 13,461 users (3,762 age ≤15 years and 9,699 age >15 years) showed an average TITR of 48.9% in those age ≤15 years and 48.8% in the older group (vs. TIR 71.2% and 73.9%, respectively). Consistent use of a glucose target (GT) of 100 mg/dL and active insulin time (AIT) of 2 h were the most relevant factors predicting higher TITR (P < 0.0001). In users consistently applying these optimal settings, TITR was 56.7% in those age ≤15 years and 57.0% in the older group, and the relative impact of these settings on TITR was 60% and 86% greater than that on TIR, respectively. TITRs of ∼45% (age ≤15 years 46.3% and older group 45.4%), ∼50% (50.7% and 50.7%) and ∼55% (56.4% and 58.0%) were best associated with glucose management indicators <7.0%, <6.8%, and <6.5%, respectively. TITRs of >45%, >50%, and >55% were achieved in 91%, 74%, and 55% of those age ≤15 years and 93%, 81%, and 57% of older group users, respectively, at optimal settings. CONCLUSIONS: This study demonstrates that 1) mean TIR is high with a high mean TITR in MM780G users (>48%), 2) consistent use of optimal GT/AIT improves TITR (>56%), 3) the impact of these settings on TITR is larger than on TIR, and 4) a TITR target >50% is our suggested treatment goal.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2337/dc23-1581

Cites

8 of 8 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Real-world data also aligned with clinical trial outcomes showing HbA1c reductions of 0.3-0.7% across diverse age groups.

    “Realworld data also aligned with clinical trial outcomes showing HbA₁ c reductions of 0•3-0•7% across diverse age groups.95,101,102”

    From The future of type 1 diabetes therapy (2025) · cites this paper as [102]

    The Results (Part 1; page not provided) report real-world GMI reductions of 0.4% in users aged ≤15 years and 0.3% in older users. The Conclusions say these improvements were consistent with clinical studies, but also state that HbA1c could not be used as an endpoint. This source therefore supports improvement across age groups, not the attributed 0.3–0.7% HbA1c reduction. (LLM, conf=0.88)

    Experimental — accuracy not yet measured
  • The observed results are similar to earlier AID clinical investigations, demonstrating comparable performance in percent time in range to Control-IQ (70.9%), Omnipod 5 (67.8%), and Medtronic 780G (71.2% ≤15 years/73.9% >15 years).

    “Despite the small cohort, our findings are similar to earlier AID clinical investigations, demonstrating comparable performance in %TIR observed in prospective and retrospective observation trials of other systems, such as Control-IQ (70.9%),15 Omnipod 5 (67.8%),16 and Medtronic 780G (71.2% (≤15 years)/73.9% (>15 years)).17”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 13

  1. [3]Montaser E, Fabris C, Kovatchev B. Essential continuous glucose monitoring metrics: the principal dimensions of glycemic control in diabetes. Diabetes Technol Ther 2022; 24: 797– 804 35714355
  2. [7]Cohen O, Basu R, Bock G,. Prediction of postprandial glycemic exposure: utility of fasting and 2-h glucose measurements alone and in combination with assessment of body composition, fitness, and strength. Diabetes Care 2006; 29: 2708– 2713 17130209
  3. [8]Petrovski G, Campbell J, Pasha M,. Simplified meal announcement versus precise carbohydrate counting in adolescents with type 1 diabetes using the MiniMed 780G advanced hybrid closed loop system: a randomized controlled trial comparing glucose control. Diabetes Care 2023; 46: 544– 550 36598841 PMC10148675
  4. [9]Matejko B, Juza A, Kieć-Wilk B,. Transitioning of people with type 1 diabetes from multiple daily injections and self-monitoring of blood glucose directly to MiniMed 780G advanced hybrid closed-loop system: a two-center, randomized, controlled study. Diabetes Care 2022; 45: 2628– 2635 35972259 PMC9862281
  5. [10]Arrieta A, Battelino T, Scaramuzza AE,. Comparison of MiniMed 780G system performance in users aged younger and older than 15 years: evidence from 12 870 real-world users. Diabetes Obes Metab 2022; 24: 1370– 1379 35403792 PMC9545031
  6. [11]Castañeda J, Mathieu C, Aanstoot HJ,. Predictors of time in target glucose range in real-world users of the MiniMed 780G system. Diabetes Obes Metab 2022; 24: 2212– 2221 35791621
  7. [13]Beck RW, Kanapka LG, Breton MD,. A meta-analysis of randomized trial outcomes for the t:slim X2 insulin pump with control-IQ technology in youth and adults from age 2 to 72. Diabetes Technol Ther 2023; 25: 329– 342 37067353 PMC10171957
  8. [16]van den Heuvel T, Kolassa R, Keuthage W,. Advanced hybrid closed loop in adult population with type 1 diabetes: a substudy from the ADAPT randomized controlled trial in users of real-time continuous glucose monitoring. J Diabetes Sci Technol. 22 March 2023 [Epub ahead of print]. DOI: 10.1177/19322968231161320 PMC11418449 36949671 · DOI
  9. [17]Boscari F, Ferretto S, Cavallin F, Bruttomesso cD. Switching from predictive low glucose suspend to advanced hybrid closed loop control: effects on glucose control and patient reported outcomes. Diabetes Res Clin Pract 2022; 185: 109784 35183648
  10. [18]Garcia-Tirado J, Diaz JL, Esquivel-Zuniga R,. Advanced closed-loop control system improves postprandial glycemic control compared with a hybrid closed-loop system following unannounced meal. Diabetes Care 2021; 44: 2379– 2387 34400480
  11. [19]Seget S, Rusak E, Polanska J, Jarosz-Chobot P. Prospective open-label, single-arm, single-center follow-up study of the application of the advanced hybrid closed loop system in well-controlled children and adolescents with type 1 diabetes. Diabetes Technol Ther 2022; 24: 824– 831 35852811 PMC9618370
  12. [20]McVean J, Miller J. MiniMedTM780G insulin pump system with smartphone connectivity for the treatment of type 1 diabetes: overview of its safety and efficacy. Expert Rev Med Devices 2021; 18: 499– 504 34014794
  13. [21]Petersson J, Åkesson K, Sundberg F, Särnblad S. Translating glycated hemoglobin A1c into time spent in glucose target range: a multicenter study. Pediatr Diabetes 2019; 20: 339– 344 30652407

Cited by

Sorted by citations

Authors
Castañeda, J., Arrieta, A., van den Heuvel, T., Battelino, T., & Cohen, O.
Affiliations
  • Medtronic (Switzerland) (CH)
  • 1Medtronic International Trading Sàrl, Tolochenaz, Switzerland
  • University of Ljubljana (SI)
Corresponding author
Cohen, Ohad
Venue
Diabetes Care
Published
2023-12-19
Volume
47
Issue
5
Pages
790-797
ISSN
  • 0149-5992 (ISSN-L)
  • 1935-5548
OpenAlex
W4389942708
Cited by
107
References
8
Citations by year
2026: 31 · 2025: 45 · 2024: 31
FWCI
21.95
Type
journal article
Language
en
Open access
Locations
  • Diabetes Care (publishedVersion)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Funders
Medtronic Europe
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Genetics › Diabetes and associated disorders
Keywords
type 1 diabetes, time in tight range, MiniMed 780G, glycemic targets, time in range, glucose management indicator, predictive factors, treatment goals, adolescents, adults, glycemic control, hypoglycemia, hyperglycemia, continuous glucose monitoring, insulin dose
MeSH terms
Adolescent, Blood Glucose, Diabetes Mellitus, Type 1, Glucose, Humans, Hypoglycemic Agents, Insulin, Blood Glucose Self-Monitoring

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review