The DOI matches its Crossref record.
Application of Chronic Intravascular Blood Glucose Sensor in Dogs
Abstract
An intravenous glucose sensor was implanted in six dogs for 1–15 wk. The glucose sensor is a flexible cylinder, ∼0.2 cm diam and 30 cm long, with a tip containing immobilized glucose oxidase and catalase coupled to a potentiostatic O2 sensor. The sensor and a similar O2 reference sensor were implanted in the superior vena cava near the entrance of the right atrium. The sensor response was conveyed externally either by a telemetry system implanted nearby, surgically accessed leads, or chronically maintained percutaneous leads. Summing over the six implants, there was a total implantation period of 333 days during which glucose sensors were functional on demand. The sensor response showed agreement with conventionally assayed blood samples after accounting for a response lag. Sensor response to glucose showed little change over the implant period. Biocompatibility, enzyme lifetime, O2 availability, O2 sensor stability, and biochemical interference were not limitations. Results demonstrated that this sensor can function effectively as an implant in dogs for a period of months and has the potential for long-term operation.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Electrochemical Glucose Biosensors
Wang, J. (2008) · Chemical Reviews · doi:10.1021/cr068123a · Cited by 3,498
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
Electrochemical Non-enzymatic Glucose Sensors: A Perspective and an Evaluation
Toghill, K. E., & Compton, R. G. (2010) · International Journal of Electrochemical Science · doi:10.1016/s1452-3981(23)15359-4 · Cited by 926
Hydrogels in sensing applications
Buenger, D., Topuz, F., & Groll, J. (2012) · Progress in Polymer Science · doi:10.1016/j.progpolymsci.2012.09.001 · Cited by 762
Glucose Biosensors: 40 Years of Advances and Challenges
Wang, J. (2001) · Electroanalysis · doi:10.1002/1521-4109(200108)13:12<983::aid-elan983>3.0.co;2-# · Cited by 752
Metadata
- Authors
- Armour, J. C., Lucisano, J. Y., McKean, B. D., & Gough, D. A.
- Venue
- Diabetes
- Publisher
- American Diabetes Association
- Published
- 1990-12-01
- Volume
- 39
- Issue
- 12
- Pages
- 1519-1526
- ISSN
- 0012-1797 (ISSN-L)
- 1939-327X
- OpenAlex
- W2023313136
- Cited by
- 175
- References
- 0
- FWCI
- 9.04
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Application of Chronic Intravascular Blood Glucose Sensor in Dogs. https://app.cochat.ai/papers/10.2337/diab.39.12.1519
- The paper
- Armour, J. C., Lucisano, J. Y., McKean, B. D., & Gough, D. A. (1990). Application of Chronic Intravascular Blood Glucose Sensor in Dogs. Diabetes. https://doi.org/10.2337/diab.39.12.1519
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.