Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Lamb, Tyler S, Thakrar, Amar, Ghosh, Mahua, Wilson, Merne P, Wilson, Thomas W · 2010 · Clinical and Investigative Medicine
Cited by 14

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

Objective: To compare blood pressure readings obtained with two commonly used oscillometric monitors: Omron HEM 711 AC (OM) and Welch-Allyn 52000 series NIBP/oximeter (WA) with mercury sphygmomanometers (Merc) in subjects with atrial fibrillation. Methods: We recruited 51 hemodynamically stable subjects with atrial fibrillation. Fifty four subjects in normal sinus rhythm served as controls. Supine blood pressure readings in each arm were recorded simultaneously using one monitor and Merc. The second monitor then replaced the first and readings were repeated. Merc was then switched to the opposite arm, and both monitors retested. Apical heart rates were ascertained with a stethoscope. We used the averaged, same arm Merc readings as “gold standard”. Results: Automated blood pressure readings were obtained in all control subjects and in all but three of those with atrial fibrillation. Both monitors, and operators, noted a difference between apical and radial/brachial pulse rates: apical-recorded: Merc 6.1±15.0; OM 5.5±13.7; WA 10.0±21.2 beats per minute. Both monitors were accurate in controls: over 90% of readings were within 10 mmHg of averaged Merc, and both achieved European Hypertension Society standards. In subjects with atrial fibrillation, about one quarter of all oscillometric readings differed from Merc by more than 10 mmHg. Both falsely high and falsely low readings occurred, some up to 30 mmHg. There was no relation between accuracy and heart rate. Conclusions: Single blood pressure readings, taken with oscillometric monitors in subjects with atrial fibrillation differ, often markedly, from those taken manually. Health care professionals should record multiple readings manually, using validated instruments when making therapeutic decisions.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.25011/cim.v33i1.11838

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 14

Metadata

Authors
Lamb, T. S., Thakrar, A., Ghosh, M., Wilson, M. P., & Wilson, T. W.
Venue
Clinical and Investigative Medicine
Published
2010-02-01
Volume
33
Issue
1
Pages
E54-E62
ISSN
  • 0147-958X (ISSN-L)
  • 1488-2353
OpenAlex
W2124359445
Cited by
14
References
2
FWCI
1.23
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation. https://app.cochat.ai/papers/10.25011/cim.v33i1.11838
The paper
Lamb, T. S., Thakrar, A., Ghosh, M., Wilson, M. P., & Wilson, T. W. (2010). Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation. Clinical and Investigative Medicine. https://doi.org/10.25011/cim.v33i1.11838

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.