Phantom and Clinical Evaluation of the Bayesian Penalized Likelihood Reconstruction Algorithm Q.Clear on an LYSO PET/CT System
Open accessAbstract
UNLABELLED: Q.Clear, a Bayesian penalized-likelihood reconstruction algorithm for PET, was recently introduced by GE Healthcare on their PET scanners to improve clinical image quality and quantification. In this work, we determined the optimum penalization factor (beta) for clinical use of Q.Clear and compared Q.Clear with standard PET reconstructions. METHODS: A National Electrical Manufacturers Association image-quality phantom was scanned on a time-of-flight PET/CT scanner and reconstructed using ordered-subset expectation maximization (OSEM), OSEM with point-spread function (PSF) modeling, and the Q.Clear algorithm (which also includes PSF modeling). Q.Clear was investigated for β (B) values of 100-1,000. Contrast recovery (CR) and background variability (BV) were measured from 3 repeated scans, reconstructed with the different algorithms. Fifteen oncology body (18)F-FDG PET/CT scans were reconstructed using OSEM, OSEM PSF, and Q.Clear using B values of 200, 300, 400, and 500. These were visually analyzed by 2 scorers and scored by rank against a panel of parameters (overall image quality; background liver, mediastinum, and marrow image quality; noise level; and lesion detectability). RESULTS: As β is increased, the CR and BV decreases; Q.Clear generally gives a higher CR and lower BV than OSEM. For the smallest sphere reconstructed with Q.Clear B400, CR is 28.4% and BV 4.2%, with corresponding values for OSEM of 24.7% and 5.0%. For the largest hot sphere, Q.Clear B400 yields a CR of 75.2% and a BV of 3.8%, with corresponding values for OSEM of 64.4% and 4.0%. Scorer 1 and 2 ranked B400 as the preferred reconstruction in 13 of 15 (87%) and 10 of 15 (73%) cases. The least preferred reconstruction was OSEM PSF in all cases. In most cases, lesion detectability was highest ranked for B200, in 9 of 15 (67%) and 10 of 15 (73%), with OSEM PSF ranked lowest. Poor lesion detectability on OSEM PSF was seen in cases of mildly (18)F-FDG-avid mediastinal nodes in lung cancer and small liver metastases due to background noise. Conversely, OSEM PSF was ranked second highest for lesion detectability in most pulmonary nodule evaluation cases. The combined scores confirmed B400 to be the preferred reconstruction. CONCLUSION: Our phantom measurement results demonstrate improved CR and reduced BV when using Q.Clear instead of OSEM. A β value of 400 is recommended for oncology body PET/CT using Q.Clear.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, this difference is directly dependent on the user-defined penalization factor beta during PL reconstruction, and at beta values of 300 to 600, which have been rated as optimal for visual reading of [18F]FDG-PET images by human readers [107,112,113], inter-method SUV differences may no longer be significant [41,108,111].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [107]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“several authors have reported improved SNR and image quality with different radiopharmaceuticals, such as [18F]FDG [107,112,113], [18F]F-PSMA-1007 [143], [68Ga]Ga-PSMA-11 [169,170], [68Ga]Ga-DOTATATE [171] or 89Zr-labelled tracers [172].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [107]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“Using a commercially available PL reconstruction (General Electric [GE] Q.Clear), several phantom studies have shown that PL reconstruction can increase the CR of standardized sphere inserts compared to conventional TOF and PSF reconstruction [41,46,101,107,108].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [107]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Measurement of Observer Agreement for Categorical Data
Landis, J. R., & Koch, G. G. (1977) · Biometrics · doi:10.2307/2529310 · Cited by 81,148
National electrical manufacturers association
Sokole, E. B., & Cradduck, T. D. (1926) · Journal of the A.I.E.E. · doi:10.1109/jaiee.1926.6537344 · Cited by 860
A Systematic Review of the Factors Affecting Accuracy of SUV Measurements
Adams, M. C., Turkington, T. G., Wilson, J. M., & Wong, T. Z. (2010) · American Journal of Roentgenology · doi:10.2214/ajr.10.4923 · Cited by 662
Kinahan, P. E., & Fletcher, J. W. (2010) · Seminars in Ultrasound, CT and MRI · doi:10.1053/j.sult.2010.10.001 · Cited by 631
Statistical methods for tomographic image reconstruction
Geman, S., & McClure, D. E. (1987) · Medical Entomology and Zoology · Cited by 586
Resolution modeling in PET imaging: Theory, practice, benefits, and pitfalls
Rahmim, A., Qi, J., & Sossi, V. (2013) · Medical Physics · doi:10.1118/1.4800806 · Cited by 371
Ahn, S., & Fessler, J. A. (2003) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2003.812251 · Cited by 345
Nuyts, J. L., Beque, D., Dupont, P. J., & Mortelmans, L. (2002) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2002.998681 · Cited by 211
Influence of reconstruction iterations on 18F-FDG PET/CT standardized uptake values.
Jaskowiak, C., Bianco, J. A., Perlman, S. B., & Fine, J. P. (2005) · PubMed · Cited by 162
Teoh, E. J., McGowan, D. R., Bradley, K. M., Belcher, E., Black, E., & Gleeson, F. V. (2016) · European Radiology · doi:10.1007/s00330-015-3832-y · Cited by 109
Self-citation
Flagged first, then by citations
1 of the 22 references couldn’t be matched to a known paper.
Research works printed but not linked: 2
- Ma, 2013, Clinical evaluation of penalized likelihood reconstruction in whole-body PET studies, Eur J Nucl Med Mol Imaging., 40, S109
- Kenny, 2013, Effect of the GE SharpIR PET reconstruction on contrast convergence and measured SUV values in the NEMA image quality phantom and patients, Eur J Nucl Med Mol Imaging., 40, S209
Cited by
Sorted by citations
Items 101–200 of 265
Margail, C., Merlin, C., Billoux, T., Wallaert, M., Otman, H., van Sas, N. C. F., Molnar, I., Guillemin, F., Boyer, L., Guy, L., et al. (2023) · EJNMMI Research · doi:10.1186/s13550-023-00999-y · Cited by 12
Chen, W., Li, Y., Li, Z., Jiang, Y., Cui, Y., Zeng, J., Mo, Y., Tang, S., Li, S., Liu, L., et al. (2024) · Journal of Nuclear Medicine · doi:10.2967/jnumed.123.266948 · Cited by 11
Vallot, D., Ponti, E. D., Morzenti, S., Gramek-Jedwabna, A., Pieczonka, A., Llompart, G. R., Siennicki, J., Deak, P. D., Dutta, C., Uribe, J., et al. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-00294-y · Cited by 12
Xu, L., Li, R.-S., Wu, R., Yang, R., You, Q., Yao, X., Xie, H.-F., Lv, Y., Dong, Y., Wang, F., et al. (2022) · EJNMMI Physics · doi:10.1186/s40658-022-00451-5 · Cited by 12
Liu, Y., Gao, M.-J., Zhou, J., Du, F., Chen, L., Huang, Z., Hu, J., & Lou, C. (2021) · BMC Medical Imaging · doi:10.1186/s12880-021-00664-7 · Cited by 11
Matheoud, R., Al-Maymani, N., Oldani, A., Sacchetti, G. M., Brambilla, M., & Carriero, A. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00380-9 · Cited by 10
Swiha, M. M., Ayati, N., Oprea‐Lager, D. E., Ceci, F., & Emmett, L. (2023) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2023.07.007 · Cited by 12
Murphy, D. J., Royle, L., Chalampalakis, Z., Alves, L., Martins, N., Bassett, P., Breen, R. A. M., Nair, A., Billè, A., Chicklore, S., et al. (2019) · European Journal of Radiology · doi:10.1016/j.ejrad.2019.06.005 · Cited by 11
Post-reconstruction enhancement of [18F]FDG PET images with a convolutional neural network
Ly, J., Minarik, D., Jögi, J., Wollmer, P., & Trägårdh, E. (2021) · EJNMMI Research · doi:10.1186/s13550-021-00788-5 · Cited by 11
Hou, X., Ma, H., Esquinas, P. L., Uribe, C., Tolhurst, S., Bénard, F., Liu, D. M., Rahmim, A., & Celler, A. M. (2020) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aba8b5 · Cited by 11
Q.Clear reconstruction for reducing the scanning time for 68 Ga-DOTA-FAPI-04 PET/MR imaging
Ruan, W., Qin, C., Liu, F., Pi, R., Gai, Y., Liu, Q., & Lan, X. (2023) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-023-06134-2 · Cited by 10
Bourigault, P., Skwarski, M., Macpherson, R. E., Higgins, G. S., & McGowan, D. R. (2022) · Scientific Reports · doi:10.1038/s41598-022-26199-7 · Cited by 10
Self-citationSantoro, M., Gala, G. D., Paolani, G., Zagni, F., Civollani, S., Strolin, S., & Strigari, L. (2023) · Physica Medica · doi:10.1016/j.ejmp.2022.102523 · Cited by 11
Oddstig, J., Brolin, G., Trägårdh, E., & Minarik, D. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00366-7 · Cited by 10
Liu, L., Liu, H., Xu, S., Zhang, S., Yi, T., Mok, G. S. P., & Chen, Y. (2022) · Frontiers in Medicine · doi:10.3389/fmed.2022.845806 · Cited by 10
Mapelli, P., Ghezzo, S., Gajate, A. M. S., Preza, E., Brembilla, G., Cucchiara, V., Bezzi, C., Presotto, L., Bettinardi, V., Savi, A., et al. (2021) · Research Square · doi:10.21203/rs.3.rs-594771/v1 · Cited by 10
Sadeghi, F., Sheikhzadeh, P., Kasraie, N., Farzanehfar, S., Abbasi, M., Salehi, Y., & Ay, M. (2023) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-023-01299-4 · Cited by 9
Neural blind deconvolution for deblurring and supersampling PSMA PET
Sample, C. M., Rahmim, A., Uribe, C., Bénard, F., Wu, J., Fedrigo, R., & Clark, H. (2024) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ad36a9 · Cited by 13
Wagatsuma, K., Sakata, M., Miwa, K., Hamano, Y., Kawakami, H., Kamitaka, Y., Yamao, T., Miyaji, N., Ishibashi, K., Tago, T., et al. (2024) · EJNMMI Physics · doi:10.1186/s40658-024-00641-3 · Cited by 9
Vieira, T. S., Faria, D. B., Moura, C. S., Francisco, E., Barroso, S., & de Oliveira, J. P. (2018) · Medicine · doi:10.1097/md.0000000000011878 · Cited by 9
Ribeiro, D. P., Hallett, W. A., Howes, O. D., McCutcheon, R. A., Nour, M. M., & Tavares, A. (2022) · EJNMMI Research · doi:10.1186/s13550-022-00883-1 · Cited by 9
Bonney, L. M., Kalisvaart, G. M., van Velden, F. H. P., Bradley, K. M., Hassan, A. B., Grootjans, W., & McGowan, D. R. (2025) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-025-07149-7 · Cited by 8
Self-citationMiwa, K., Yamagishi, S., Kamitaki, S., Anraku, K., Sato, S., Yamao, T., Miyaji, N., Wachi, K., Akiya, N., Wagatsuma, K., et al. (2025) · EJNMMI Physics · doi:10.1186/s40658-025-00742-7 · Cited by 8
Miwa, K., Yamao, T., Hashimoto, F., Miyaji, N., Kamitaka, Y., Masubuchi, M., Murata, T., Yoshii, T., Kobayashi, R., Fukuda, S., et al. (2025) · Annals of Nuclear Medicine · doi:10.1007/s12149-025-02088-7 · Cited by 9
Shinohara, H., Hori, K., & Hashimoto, T. (2023) · Annals of Nuclear Medicine · doi:10.1007/s12149-023-01862-9 · Cited by 8
Sadeghi, F., Sheikhzadeh, P., Farzanehfar, S., Ghafarian, P., Moafpurian, Y., & Ay, M. (2023) · EJNMMI Physics · doi:10.1186/s40658-023-00587-y · Cited by 8
Optimization of Q.Clear reconstruction for dynamic 18F PET imaging
Lysvik, E. K., Mikalsen, L. T. G., Revheim, M.-E. R., Emblem, K. E., & Hjørnevik, T. (2023) · EJNMMI Physics · doi:10.1186/s40658-023-00584-1 · Cited by 8
Gabrani-Juma, H., Bimani, Z. A., Zuckier, L. S., & Klein, R. (2020) · Journal of medical imaging · doi:10.1117/1.jmi.7.2.022412 · Cited by 8
Maronnier, Q., Courbon, F., & Caselles, O. (2022) · EJNMMI Physics · doi:10.1186/s40658-022-00496-6 · Cited by 7
Wagatsuma, K., Oda, K., Miwa, K., Inaji, M., Sakata, M., Toyohara, J., Ishiwata, K., Sasaki, M., & Ishii, K. (2017) · Radiological Physics and Technology · doi:10.1007/s12194-017-0414-5 · Cited by 7
Morley, N. C. D., McGowan, D. R., Gleeson, F. V., & Bradley, K. M. (2016) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.201607-1371im · Cited by 7
Self-citationKojima, I., Takanami, K., Ogawa, T., Sakamoto, M., Nagai, H., Miyashita, H., & Iikubo, M. (2020) · Annals of Nuclear Medicine · doi:10.1007/s12149-020-01489-0 · Cited by 7
Lohaus, N., Enderlin, F., Skawran, S., Maurer, A. B., Abukwaik, A. M. A., Franzen, D., Huellner, M. W., & Messerli, M. (2022) · Scientific Reports · doi:10.1038/s41598-022-09904-4 · Cited by 6
Dwivedi, P., Sawant, V., Vajarkar, V., Vatsa, R., Choudhury, S., Jha, A. K., & Rangarajan, V. (2022) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000001631 · Cited by 6
Deep learning-based time-of-flight (ToF) enhancement of non-ToF PET scans for different radiotracers
Mehranian, A., Wollenweber, S. D., Bradley, K. M., Fielding, P. A., Huellner, M. W., Iagaru, A. H., Dedja, M., Colwell, T., Kotasidis, F. A., Johnsen, R., et al. (2025) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-025-07119-z · Cited by 6
Self-citationJung-Soo, K., Park, C., Seok-Hwan, Y., Lee, J.-A., Kim, T., & Yang, H.-J. (2021) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000001368 · Cited by 6
Wu, Z., Guo, B., Huang, B., Zhao, B., Qin, Z., Hao, X., Liang, M., Xie, J., & Li, S. (2021) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.13129 · Cited by 6
Bjöersdorff, M., Puterman, C., Oddstig, J., Amidi, J., Zackrisson, S., Kjölhede, H., Bjartell, A. S., Wollmer, P., & Trägårdh, E. (2022) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.12770 · Cited by 6
Lores, B. F., Lozano, J. A., Fuster-Matanzo, A., Mayorga-Ruiz, I., Moreno-Ruiz, P., Bataller, F. B., Teruel, A. B., Saus, A., Ortiz, A., Villamón-Ribate, E., et al. (2023) · PubMed Central · Cited by 6
Heterogeneous PSMA ligand uptake inside parotid glands
Sample, C. M., Uribe, C., Rahmim, A., Bénard, F., Wu, J., & Clark, H. (2024) · Physica Medica · doi:10.1016/j.ejmp.2024.103366 · Cited by 6
Data-driven gating (DDG)-based motion match for improved CTAC registration
Cook, E. L., Su, K., Higgins, G. S., Johnsen, R., Bouhnik, J.-P., & McGowan, D. R. (2024) · EJNMMI Physics · doi:10.1186/s40658-024-00644-0 · Cited by 6
Self-citationMonaco, L., Ponti, E. D., Crivellaro, C., Morzenti, S., Todde, S., Landoni, C., Elisei, F., Musarra, M., & Guerra, L. (2022) · Current Radiopharmaceuticals · doi:10.2174/1874471015666220107100200 · Cited by 6
Dolci, C., Spadavecchia, C., Crivellaro, C., Ponti, E. D., Todde, S., Morzenti, S., Turolla, E. A., Crespi, A., Guerra, L., & Landoni, C. (2019) · Physica Medica · doi:10.1016/j.ejmp.2018.12.038 · Cited by 6
Initial experience with a PET/computed tomography system using silicon photomultiplier detectors
Park, S. Y., Barrato, L., Hatami, N., Davidzon, G., Gambhir, S. S., & Iagaru, A. H. (2019) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000001088 · Cited by 5
Effect of Bayesian penalty likelihood algorithm on 18F-FDG PET/CT image of lymphoma
Wang, Y., Lin, L., Quan, W., Li, J., & Li, W. (2021) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000001516 · Cited by 5
Anatomically guided reconstruction improves lesion quantitation and detectability in bone SPECT/CT
Vuohijoki, H. E., Constable, C., & Sohlberg, A. O. (2023) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000001675 · Cited by 5
Genc, M., Yıldırım, N., Coskun, N., Özdemir, E., & Türkölmez, Ş. (2023) · Revista Española de Medicina Nuclear e Imagen Molecular (English Edition) · doi:10.1016/j.remnie.2023.07.006 · Cited by 5
Fukuda, S., Wagatsuma, K., Miwa, K., Yakushiji, Y., Kamitaka, Y., Yamao, T., Miyaji, N., & Ishii, K. (2024) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-024-01476-z · Cited by 5
Diagnostic and response assessment FDG PET ‐ CT in neurolymphomatosis
Bruce, D., Eagleton, H., & Subesinghe, M. (2016) · Clinical Case Reports · doi:10.1002/ccr3.734 · Cited by 5
Gamma Ray Emission Imaging in the Medical and Nuclear Safeguards Fields
Dendooven, P. G., & Bubba, T. A. (2022) · Lecture notes in physics · doi:10.1007/978-3-031-10751-1_7 · Cited by 5
Siekkinen, R., Teuho, J., Smith, N. A. S., Fenwick, A., Kirjavainen, A. K., Koskensalo, K., Saraste, A., & Teräs, M. (2020) · Frontiers in Physics · doi:10.3389/fphy.2020.00148 · Cited by 5
Macnab, M.-R. F., Biggans, T. J., McKiddie, F. I., Pether, M. I., Straiton, J., & Staff, R. T. (2020) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000001204 · Cited by 5
Usmani, S., Ahmed, N., Gnanasegaran, G., Rasheed, R., Marafi, F., Alnaaimi, M., Omar, M., Musbah, A., Kandari, F. A., Schepper, S. D., et al. (2021) · British Journal of Radiology · doi:10.1259/bjr.20210043 · Cited by 5
Tatsumi, M., Soeda, F., Kamiya, T., Ueda, J., Katayama, D., Matsunaga, K., Watabe, T., Kato, H., & Tomiyama, N. (2021) · Frontiers in Oncology · doi:10.3389/fonc.2021.707023 · Cited by 4
Bradley, K. M., McGowan, D. R., Gleeson, F. V., Johnson, G. B., Young, J. R., Levin, C. S., Davidzon, G., & Iagaru, A. H. (2018) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.212761 · Cited by 5
Self-citationDoes Q.Clear Processing Change PET Ratios? Quantitative Evidence Using BTXBrain-DAT
Chong, A., Ha, J., & Chung, J. Y. (2025) · Brain Sciences · doi:10.3390/brainsci15101036 · Cited by 4
van de Burgt, A., Dibbets‐Schneider, P., Slump, C. H., Scholte, A. J. H. A., Atsma, D. E., de Geus‐Oei, L., & van Velden, F. H. P. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00393-4 · Cited by 4
Ayati, N., McIntosh, L., Buteau, J. P., Alipour, R., Pudís, M., Daw, N., Jackson, P. A., & Hofman, M. S. (2024) · Cancer Imaging · doi:10.1186/s40644-024-00702-x · Cited by 4
Inukai, J. I., Nogami, M., Tachibana, M., Zeng, F., Nishitani, T., Kubo, K., & Murakami, T. (2023) · Diagnostics · doi:10.3390/diagnostics13111871 · Cited by 4
Development of a bespoke phantom to optimize molecular PET imaging of pituitary tumors
Gillett, D., Marsden, D., Crawford, R., Ballout, S., MacFarlane, J., van der Meulen, M., Gillett, B., Bird, N., Heard, S., Powlson, A., et al. (2023) · EJNMMI Physics · doi:10.1186/s40658-023-00552-9 · Cited by 4
Reconstruction of uptake patterns in PET : The influence of regularizing prior
Bernardi, E. D., Fallanca, F., Gianolli, L., Gilardi, M. C., & Bettinardi, V. (2017) · Medical Physics · doi:10.1002/mp.12205 · Cited by 4
Yamaguchi, N., Odagiri, H., Sato, H., Kodama, H., Tanaka, Y., Takeuchi, T., Takanami, K., Usui, A., & Kaneta, T. (2025) · Journal of Nuclear Medicine Technology · doi:10.2967/jnmt.125.269869 · Cited by 3
Young, J. R., Mugu, V. K., Johnson, G. B., Ehman, E. C., Packard, A. T., Homb, A. C., Nathan, M. A., Thanarajasingam, G., & Kemp, B. J. (2023) · Medicine · doi:10.1097/md.0000000000032665 · Cited by 3
Usefulness of a new anthropomorphic phantom simulating the chest and abdomen regions in PET tests
Sagara, H., Inoue, K., Mano, C., Kajiwara, H., Nagai, Y., Fujii, H., & Inaki, A. (2024) · Annals of Nuclear Medicine · doi:10.1007/s12149-024-02007-2 · Cited by 3
Recent advances in combined Positron Emission Tomography and Magnetic Resonance Imaging
Galve, P., Rodríguez-Vila, B., Herraiz, J. L., García‐Vázquez, V., Malpica, N., Udías, J. M., & Torrado-Carvajal, Á. (2024) · Journal of Instrumentation · doi:10.1088/1748-0221/19/01/c01001 · Cited by 3
Y-90 PET/MR imaging optimization with a Bayesian penalized likelihood reconstruction algorithm
Calatayud-Jordán, J., Vela, N. C., Chimeno-Hernández, J., Carles-Fariña, M., Arroyo, C. O., Bello-Arqués, P., Pérez-Enguix, D., Martí‐Bonmatí, L., & Torres-Espallardo, I. (2024) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-024-01452-7 · Cited by 3
FDG Extravasation in PET / CT Imaging: A Visual Grading Approach Based on Clinical Observations
Bilgiç, S. (2025) · Journal of Medical Imaging and Radiation Oncology · doi:10.1111/1754-9485.13876 · Cited by 2
Fukai, S., Daisaki, H., Yoshida, H., Shimada, N., Motegi, K., Osawa, A., & Terauchi, T. (2025) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-025-01608-z · Cited by 2
Leybourne, N., Prakash, V., Hussein, M., Fenwick, A., Strouhal, P., Evans, P., & Florescu, L. (2025) · EJNMMI Physics · doi:10.1186/s40658-025-00821-9 · Cited by 2
Chism, C. B., RAVIZZINI, G. C., Macapinlac, H. A., & Pan, T. (2017) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000000670 · Cited by 2
Fukai, S., Daisaki, H., Shimada, N., Ishiyama, M., Umeda, T., Yamashita, K., Miyaji, N., Takiguchi, T., Kawakami, H., & Terauchi, T. (2022) · Biomedical engineering and physics express · doi:10.1088/2057-1976/aca90d · Cited by 2
Rovera, G., Grimaldi, S., Dall’Armellina, S., Zotta, M., Finessi, M., Passera, R., & Deandreis, D. (2023) · Diagnostics · doi:10.3390/diagnostics13233535 · Cited by 3
Leybourne, N., Gregory, R., Berry, O., Birch, E., Prakash, V., Scuffham, J. W., Strouhal, P., Meadows, A., Hussein, M., Fenwick, A., et al. (2024) · SPIE Medical Imaging: Physics of Medical Imaging · doi:10.1117/12.3006788 · Cited by 2
Dedja, M., Mehranian, A., Bradley, K. M., Walker, M. D., Fielding, P. A., Wollenweber, S. D., Johnsen, R., & McGowan, D. R. (2024) · EJNMMI Physics · doi:10.1186/s40658-024-00632-4 · Cited by 2
Self-citationMonsef, A., Sheikhzadeh, P., Steiner, J. R., Sadeghi, F., Yazdani, M., & Ghafarian, P. (2024) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/ad3e00 · Cited by 2
Verification of the effect of data-driven brain motion correction on PET imaging
Odagiri, H., Watabe, H., Takanami, K., Akimoto, K., Usui, A., Kawakami, H., Katsuki, A., Uetake, N., Dendo, Y., Tanaka, Y., et al. (2024) · PLoS ONE · doi:10.1371/journal.pone.0301919 · Cited by 2
The role of SiPM-based PET for accurate volume delineation
Leybourne, N., Hussein, M., Fenwick, A., Evans, P., & Florescu, L. (2025) · SPIE Medical Imaging: Physics of Medical Imaging · doi:10.1117/12.3046960 · Cited by 2
Filtering-Based Preconditioner for Accelerated High-Dimensional Cone-Beam CT Image Reconstruction
Wettenhovi, V.-V., Hietanen, A., Niinimäki, K., Vauhkonen, M. J., & Kolehmainen, V. (2025) · Journal of Mathematical Imaging and Vision · doi:10.1007/s10851-025-01276-4 · Cited by 2
Wagatsuma, K., Ikemoto, K., Inaji, M., Kamitaka, Y., Hara, S., Tamura, K., Miwa, K., Tsuzura, K., Tsuruki, T., Miyaji, N., et al. (2024) · Annals of Nuclear Medicine · doi:10.1007/s12149-024-01911-x · Cited by 2
Wu, Z., Qin, Z., Huang, B., Guo, B., Hao, X., Wu, P., Zhao, B., Xie, J., & Li, S. (2020) · doi:10.21203/rs.3.rs-17003/v1 · Cited by 2
Fast PET reconstruction with variance reduction and prior-aware preconditioning
Ehrhardt, M. J., Kereta, Ž., & Schramm, G. (2025) · Frontiers in Nuclear Medicine · doi:10.3389/fnume.2025.1641215 · Cited by 1
Valibeiglou, Z., Islamian, J. P., Soleymani, Y., Farzanehfar, S., Aghahosseini, F., Gilani, N., Rahmim, A., & Sheikhzadeh, P. (2025) · BMC Medical Imaging · doi:10.1186/s12880-025-01774-2 · Cited by 1
Stanzel, S., Nazerani, T., Weitzer, F., Weinke, R., Plhak, E., & Aigner, R. (2025) · Scientific Reports · doi:10.1038/s41598-025-10740-5 · Cited by 1
Minagawa, T., Hori, K., & Hashimoto, T. (2025) · EJNMMI Physics · doi:10.1186/s40658-025-00788-7 · Cited by 1
Naeimi, A., Harsini, S., Aghdam, R. A., Klein, R., & Abbaspour, F. (2025) · Frontiers in Oncology · doi:10.3389/fonc.2025.1597919 · Cited by 1
Maruyama, H., Sato, K., Takanami, K., Watabe, H., Kawakami, H., Taniyama, Y., Takase, K., & Kamei, T. (2025) · Clinical nuclear medicine open. · doi:10.1097/nm9.0000000000000042 · Cited by 1
Clinical Evaluation of BSREM Reconstruction in Pediatric Oncology Using [ 18 F]FDG PET/CT
Shkumat, N. A., Vali, R., & Shammas, A. (2025) · Journal of Nuclear Medicine Technology · doi:10.2967/jnmt.125.269555 · Cited by 1
Yamagata, T., Haramiishi, K., & Fukuchi, K. (2025) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000002097 · Cited by 1
Dambrain, A., Lacombe, M., Dufour, P. A., Lacœuille, F., Morel, O., Testard, A., Bouron, C., Girault, S., & Guillerminet, C. (2026) · EJNMMI Physics · doi:10.1186/s40658-026-00835-x · Cited by 1
Mostert, J. M., Oei, E. H. G., van Beekveld, D., Biswal, S., & van der Heijden, R. A. (2026) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000002164 · Cited by 1
Behr, S. C., Mollard, B. J., Yang, J., Flavell, R. R., Hawkins, R. A., & Seo, Y. (2017) · Molecular Imaging · doi:10.1177/1536012117736703 · Cited by 1
360° CZT gamma cameras for nuclear medicine and molecular imaging
Imbert, L., Bahloul, A., Verger, A., & Marie, P. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822960-6.00087-9 · Cited by 1
Tang, C. Y. L., Lim, G. K. Y., Chua, W. M., Ng, C. W. Q., Koo, S. X., Goh, C. X.-Y., Thang, S. P., Zaheer, S., Lam, W. W.-C., & Huang, H. L. (2023) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000001687 · Cited by 1
[PET] 5. Recent Advances in PET Image Reconstruction Using a Bayesian Penalized Likelihood Algorithm
Miwa, K., Miyaji, N., Yamao, T., Kamitaka, Y., Wagatsuma, K., & Murata, T. (2023) · Japanese Journal of Radiological Technology · doi:10.6009/jjrt.2023-2200 · Cited by 1
Robinson, M., Muirhead, R., McGowan, D. R., Chu, K.-Y., Jacobs, C., & Hawkins, M. (2023) · Clinical Oncology · doi:10.1016/j.clon.2023.06.001 · Cited by 1
Self-citationItagaki, K., Miyake, K. K., Tanoue, M., Oishi, T., Kataoka, M., Kawashima, M., Toi, M., & Nakamoto, Y. (2023) · PubMed Central · doi:10.22038/aojnmb.2023.71598.1501 · Cited by 1
[PET] 6. Overview of Clinical Applications of Artificial Intelligence in PET
Emura, R. (2023) · Japanese Journal of Radiological Technology · doi:10.6009/jjrt.2023-2212 · Cited by 1
Skórkiewicz, K., Łątka, K., Sowa-Staszczak, A., & Hubalewska–Dydejczyk, A. (2023) · Bio-Algorithms and Med-Systems · doi:10.5604/01.3001.0054.1818 · Cited by 1
Tachibana, M., Nogami, M., Inukai, J. I., Zeng, F., Kubo, K., Kurimoto, T., Huellner, M. W., Ueno, Y. R., Tsuboyama, T., Imaoka, I., et al. (2024) · European Journal of Radiology · doi:10.1016/j.ejrad.2024.111678 · Cited by 1
Natsuaki, Y., Leynes, A. P., Wangerin, K. A., Hamdi, M., Rajagopal, A., Kinahan, P. E., Laforest, R. A., Larson, P. E. Z., Hope, T. A., & James, S. S. (2024) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.14507 · Cited by 1
Metadata
- Authors
- Teoh, E. J., McGowan, D. R., Macpherson, R. E., Bradley, K. M., & Gleeson, F. V.
- Affiliations
- Churchill Hospital (GB)
- Venue
- Journal of Nuclear Medicine
- Publisher
- Society of Nuclear Medicine
- Published
- 2015-07-09
- Volume
- 56
- Issue
- 9
- Pages
- 1447-1452
- ISSN
- 0161-5505 (ISSN-L)
- 2159-662X
- PMC
- PMC4558942
- OpenAlex
- W2161249702
- Cited by
- 265
- References
- 22
- Citations by year
- 2026: 14 · 2025: 30 · 2024: 26 · 2023: 32 · 2022: 27 · 2021: 39 · 2020: 28 · 2019: 21 · 2018: 16 · 2017: 19 · 2016: 9 · 2015: 2
- FWCI
- 12.43
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Nuclear Medicine (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- Cancer Research UK (16466)
- National Institute for Health and Care Research (NIHR-HCS-D12-03-13)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Keywords
- Q.Clear, Bayesian penalized likelihood reconstruction, PET image reconstruction, OSEM, point spread function modeling, 18F-FDG PET/CT, image quality, lesion detectability, contrast recovery, background variation, time-of-flight PET, lung cancer, mediastinal lymph nodes, liver metastasis
- MeSH terms
- Machine Learning, Algorithms, Bayes Theorem, Humans, Image Enhancement, Image Interpretation, Computer-Assisted, Pattern Recognition, Automated, Sensitivity and Specificity, Tomography, X-Ray Computed, Reproducibility of Results, Likelihood Functions, Phantoms, Imaging, Positron-Emission Tomography, Multimodal Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review