Accuracy of robot-guided versus freehand fluoroscopy-assisted pedicle screw insertion in thoracolumbar spinal surgery
Open accessThe DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: The quest to improve the safety and accuracy and decrease the invasiveness of pedicle screw placement in spine surgery has led to a markedly increased interest in robotic technology. The SpineAssist from Mazor is one of the most widely distributed robotic systems. The aim of this study was to compare the accuracy of robot-guided and conventional freehand fluoroscopy-guided pedicle screw placement in thoracolumbar surgery. METHODS: This study is a retrospective series of 169 patients (83 women [49%]) who underwent placement of pedicle screw instrumentation from 2007 to 2015 in 2 reference centers. Pathological entities included degenerative disorders, tumors, and traumatic cases. In the robot-assisted cohort (98 patients, 439 screws), pedicle screws were inserted with robotic assistance. In the freehand fluoroscopy-guided cohort (71 patients, 441 screws), screws were inserted using anatomical landmarks and lateral fluoroscopic guidance. Patients treated before 2009 were included in the fluoroscopy cohort, whereas those treated since mid-2009 (when the robot was acquired) were included in the robot cohort. Since then, the decision to operate using robotic assistance or conventional freehand technique has been based on surgeon preference and logistics. The accuracy of screw placement was assessed based on the Gertzbein-Robbins scale by a neuroradiologist blinded to treatment group. The radiological slice with the largest visible deviation from the pedicle was chosen for grading. A pedicle breach of 2 mm or less was deemed acceptable (Grades A and B) while deviations greater than 2 mm (Grades C, D, and E) were classified as misplacements. RESULTS: In the robot-assisted cohort, a perfect trajectory (Grade A) was observed for 366 screws (83.4%). The remaining screws were Grades B (n = 44 [10%]), C (n = 15 [3.4%]), D (n = 8 [1.8%]), and E (n = 6 [1.4%]). In the fluoroscopy-guided group, a completely intrapedicular course graded as A was found in 76% (n = 335). The remaining screws were Grades B (n = 57 [12.9%]), C (n = 29 [6.6%]), D (n = 12 [2.7%]), and E (n = 8 [1.8%]). The proportion of non-misplaced screws (corresponding to Gertzbein-Robbins Grades A and B) was higher in the robot-assisted group (93.4%) than the freehand fluoroscopy group (88.9%) (p = 0.005). CONCLUSIONS: The authors’ retrospective case review found that robot-guided pedicle screw placement is a safe, useful, and potentially more accurate alternative to the conventional freehand technique for the placement of thoracolumbar spinal instrumentation.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Accuracy of Pedicular Screw Placement In Vivo
GERTZBEIN, S. D., & ROBBINS, S. E. (1990) · Spine · doi:10.1097/00007632-199001000-00004 · Cited by 1,381
Devito, D. P., Kaplan, L., Dietl, R., Pfeiffer, M., Horne, D., Silberstein, B., Hardenbrook, M., Kiriyanthan, G., Barzilay, Y., Bruskin, A., et al. (2010) · Spine · doi:10.1097/brs.0b013e3181d323ab · Cited by 407
Accuracy of Robot-Assisted Placement of Lumbar and Sacral Pedicle Screws
Ringel, F. A., Stüer, C., Reinke, A. N., Preuß, A., Behr, M., Auer, F., Stoffel, M. H., & Meyer, B. (2012) · Spine · doi:10.1097/brs.0b013e31824b7767 · Cited by 347
Gautschi, O. P., Schatlo, B., Schaller, K., & Tessitore, E. (2011) · Neurosurgical Focus · doi:10.3171/2011.7.focus11168 · Cited by 268
Self-citationKim, H., Jung, W., Chang, B., Lee, C., Kang, K., & Yeom, J. S. (2017) · The International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.1779 · Cited by 240
Schatlo, B., Molliqaj, G., Cuvinciuc, V., Kotowski, M., Schaller, K., & Tessitore, E. (2014) · Journal of Neurosurgery: Spine · doi:10.3171/2014.3.spine13714 · Cited by 220
Self-citationImage-guided spine surgery: state of the art and future directions
Tjardes, T., Shafizadeh, S., Rixen, D., Paffrath, T., Bouillon, B., Steinhausen, E. S., & Baethis, H. (2010) · European Spine Journal · doi:10.1007/s00586-009-1091-9 · Cited by 206
Unskilled unawareness and the learning curve in robotic spine surgery
Schatlo, B., Martínez, R., Alaid, A. O., von Eckardstein, K., Akhavan‐Sigari, R., Hahn, A., Stockhammer, F., & Rohde, V. (2015) · Acta Neurochirurgica · doi:10.1007/s00701-015-2535-0 · Cited by 131
Self-citation
Cited by
Fatima, N., Massaad, E., Hadzipasic, M., Shankar, G. M., & Shin, J. H. (2021) · The Spine Journal · doi:10.1016/j.spinee.2020.09.007 · Cited by 164
Augmented reality–assisted pedicle screw insertion: a cadaveric proof-of-concept study
Molina, C. A., Theodore, N., Ahmed, A. K., Westbroek, E. M., Mirovsky, Y., Harel, R., Orru’, E., Khan, M., Witham, T., & Sciubba, D. M. (2019) · Journal of Neurosurgery: Spine · doi:10.3171/2018.12.spine181142 · Cited by 160
The Application of Artificial Intelligence in Spine Surgery: A Scoping Review
Shi, L., Wang, H., & Shea, G. K.-H. (2025) · JAAOS: Global Research and Reviews · doi:10.5435/jaaosglobal-d-24-00405 · Cited by 9
Metadata
- Authors
- Molliqaj, G., Schatlo, B., Alaid, A., Solomiichuk, V., Rohde, V., Schaller, K., & Tessitore, E.
- Venue
- Neurosurgical Focus
- Published
- 2017-05
- Volume
- 42
- Issue
- 5
- Pages
- E14
- ISSN
- 1092-0684 (ISSN-L)
- OpenAlex
- W2610991997
- Cited by
- 220
- References
- 8
- FWCI
- 15.67
- Type
- journal article
- Language
- en
- Open access
- diamond
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.