Validating Measures of Electrodermal Activity and Heart Rate Variability Derived From the Empatica E4 Utilized in Research Settings That Involve Interactive Dyadic States
Open accessAbstract
Portable and wireless devices that collect physiological data are becoming more and more sought after in clinical and psychophysiological research as technology swiftly advances. These devices allow for data collection in interactive states, such as dyadic therapy, with reduced restraints compared to traditional laboratory devices. One such portable device is the Empatica E4 wristband (Empatica Srl, Milan, Italy) which allows quantifying cardiac interbeat intervals (IBIs), heart rate variability (HRV), and electro-dermal activity (EDA), as well as several other acceleration and temperature measures. In the current study, we aimed to assess IBI, HRV, and EDA measures, against the same data collected from the well-validated MindWare mobile impedance cardiograph device (MindWare Technology, Gahanna, OH, United States). We assessed the E4 strictly as a research instrument and not as a clinical tool. We were specifically interested in the wristbands' performance during naturalistic interactive face-to-face conversations which inherently involve more hand movements. We collected data from 30 participants, nested in 15 dyads, which were connected to both devices simultaneously, during rest and during a social conversation. After preprocessing and analyses, we found that mean IBIs obtained by the E4 and the MindWare device, were highly similar during rest and during conversation. Medium to high correlations were found between the devices with respect to several HRV measures, with higher correlations during rest compared to conversation. The E4 failed to produce reliable EDA data. We conclude by discussing the strengths and limitations of the E4 during seated conversational states and suggest optimal ways to collect and analyze data with the E4.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“While a complete loss of contact results in a 0-line for the EDA signal, motion artifacts without contact loss result in abrupt level changes or high amplitude fluctuations, and PPG shows considerable deviation from expected signal patterns24,42.”
From Data quality evaluation in wearable monitoring (2022) · cites this paper as [42]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Heart Rate Variability: Standards of Measurement, Physiological Interpretation, and Clinical Use
Electrophysiology, T. F. O. T. E. S. O. C. T. N. A. (1996) · Circulation · doi:10.1161/01.cir.93.5.1043 · Cited by 17,159
Heart rate variability: Standards of measurement, physiological interpretation, and clinical use
Malik, M., Bigger, J. T., Camm, A. J., Kleiger, R. E., Malliani, A., Moss, A. J., & Schwartz, P. J. (1996) · European Heart Journal · doi:10.1093/oxfordjournals.eurheartj.a014868 · Cited by 15,289
Kubios HRV–heart rate variability analysis software.
Tarvainen, M. P., Niskanen, J.-P., Lipponen, J. A., Ranta-aho, P. O., & Karjalainen, P. A. (2014) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2013.07.024 · Cited by 2,720
Shaffer, F., McCraty, R., & Zerr, C. L. (2014) · Frontiers in Psychology · doi:10.3389/fpsyg.2014.01040 · Cited by 1,866 · cited 2×
“As measuring HRV does not require particularly invasive testing, it is extensively acquired in psychological science and is considered a reliable estimate of autonomic nervous system modulation, via the vagus nerve, of sinus node cardiac activity (Blascovich et al., 2011; Kaufmann et al., 2011; Shaffer et al., 2014).”
BackgroundThe polyvagal theory: phylogenetic substrates of a social nervous system.
Porges, S. W. (2001) · International Journal of Psychophysiology · doi:10.1016/s0167-8760(01)00162-3 · Cited by 1,582 · cited 1×
“…conversational states beyond the stress response, as they involve different and perhaps more subtle regulatory capacities of the autonomic nervous system (Porges, 2001), which are of specific interest to many researchers aiming to understand the mechanism underlying changes in HRV and EDA measures.”
BackgroundSome signals and rules for taking speaking turns in conversations.
Duncan, S. (1972) · Journal of Personality and Social Psychology · doi:10.1037/h0033031 · Cited by 1,386 · cited 1×
“A final reason to include a conversation paradigm is the fact that more movement related artifacts occur during speaking, due to gesturing (Duncan, 1972).”
BackgroundMalík, M. (1996) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.1996.tb00275.x · Cited by 1,031 · cited 2×
“SDNN, which is a measure of general variability, HF, which represents cardiovagal activity (Camm et al., 1996) and IBI include both parasympathetic and sympathetic influences (Berntson et al.”
BERNTSON, G. G., CACIOPPO, J. T., & QUIGLEY, K. S. (1993) · Psychophysiology · doi:10.1111/j.1469-8986.1993.tb01731.x · Cited by 998 · cited 1×
“This frequency band in particular has been shown to represent parasympathetic nervous system regulation almost exclusively, and thus is a highly sought after (Berntson et al., 1993).”
BackgroundWearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · cited 8×
“We imported the raw data obtained from the E4 wristband into Kubios, after it was organized to fit Kubios’ specifications with MATLAB 2018a. Kubios HRV Premium software includes an adaptive QRS detection algorithm, pulse wave detectors and tools for automatic and threshold based artifact correction, trend removal and…”
MethodInfluential
Research works printed but not linked: 7
- 2015, Issues with EDA Values.
- 2017, Comparison Procomp Vs Empatica E3 Skin Conductance Signal.
- 2018, Session Start-Time Format and Synchronization.
- 2019, Support Page.
- Kamath, 2012, Heart rate variability (HRV) signal analysis: Clinical applications. Heart Rate Variability (HRV) Signal Analysis: Clinical Applications.
- 2013, R: A Language and Environment for Statistical Computing.
- Tarvainen, 2019, Kubios HRV User’s Guide
Cited by
Sorted by citations
Vos, G., Trinh, K., Sarnyai, Z., & Azghadi, M. R. (2023) · International Journal of Medical Informatics · doi:10.1016/j.ijmedinf.2023.105026 · Cited by 200
“In [77], the authors found the Empatica E4 to be suitable for psychotherapy research focused on Inter-Beat Interval (IBI) and specific HRV measures, but failed to produce reliable EDA data and produced missing IBI data, especially when a subject is being more dynamic.”
BackgroundData quality evaluation in wearable monitoring
Böttcher, S., Vieluf, S., Bruno, E., Joseph, B., Epitashvili, N., Biondi, A., Zabler, N., Glasstetter, M., Dümpelmann, M., Van Laerhoven, K., et al. (2022) · Scientific Reports · doi:10.1038/s41598-022-25949-x · Cited by 91
Magaña, V. C., Scherz, W. D., Seepold, R., Madrid, N. M., Pañeda, X. G., & García, R. (2020) · Sensors · doi:10.3390/s20185274 · Cited by 82
“This device is certified as CE Medical class 2a [66] and has been validated in many works [67,68].”
Methodvan der Mee, D. J., Gevonden, M. J., Westerink, J. H. D. M., & de Geus, E. J. C. (2021) · International Journal of Psychophysiology · doi:10.1016/j.ijpsycho.2021.08.003 · Cited by 70
“14%), although both were modest in keeping with findings for SCL from previous studies (Milstein and Gordon, 2020; Menghini et al., 2019).”
MethodResultXue, T., Ali, A. E., Zhang, T., Ding, G., & Cesar, P. S. (2021) · IEEE Transactions on Multimedia · doi:10.1109/tmm.2021.3124080 · Cited by 77
“(6) We captured participants’ physiological signals through the Empatica E4 wristband6 worn on the non-dominant hand [44], as shown in Fig.”
Background
Metadata
- Authors
- Milstein, N., & Gordon, I.
- Affiliations
- Bar-Ilan University (IL)
- Corresponding author
- Gordon, Ilanit
- Venue
- Frontiers in Behavioral Neuroscience
- Publisher
- Frontiers Media SA
- Published
- 2020-08-18
- Volume
- 14
- Pages
- 148-148
- ISSN
- 1662-5153 (ISSN-L)
- OpenAlex
- W3058974884
- Cited by
- 180
- References
- 43
- Citations by year
- 2026: 28 · 2025: 29 · 2024: 32 · 2023: 44 · 2022: 29 · 2021: 17 · 2020: 1
- FWCI
- 14.08
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Frontiers in Behavioral Neuroscience (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Israel Science Foundation
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Computer Science › Artificial Intelligence › Emotion and Mood Recognition
- Keywords
- Empatica E4, electrodermal activity, heart rate variability, RR interval, device validation, dyadic therapy, face-to-face communication, naturalistic interaction, psychophysiological measurement, measurement reliability
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review