Validity of the Empatica E4 Wristband to Measure Heart Rate Variability (HRV) Parameters: a Comparison to Electrocardiography (ECG)
Open accessAbstract
Wearable monitoring devices are an innovative way to measure heart rate (HR) and heart rate variability (HRV), however, there is still debate about the validity of these wearables. This study aimed to validate the accuracy and predictive value of the Empatica E4 wristband against the VU University Ambulatory Monitoring System (VU-AMS) in a clinical population of traumatized adolescents in residential care. A sample of 345 recordings of both the Empatica E4 wristband and the VU-AMS was derived from a feasibility study that included fifteen participants. They wore both devices during two experimental testing and twelve intervention sessions. We used correlations, cross-correlations, Mann-Whitney tests, difference factors, Bland-Altman plots, and Limits of Agreement to evaluate differences in outcomes between devices. Significant correlations were found between Empatica E4 and VU-AMS recordings for HR, SDNN, RMSSD, and HF recordings. There was a significant difference between the devices for all parameters but HR, although effect sizes were small for SDNN, LF, and HF. For all parameters but RMSSD, testing outcomes of the two devices led to the same conclusions regarding significance. The Empatica E4 wristband provides a new opportunity to measure HRV in an unobtrusive way. Results of this study indicate the potential of the Empatica E4 as a practical and valid tool for research on HR and HRV under non-movement conditions. While more research needs to be conducted, this study could be considered as a first step to support the use of HRV recordings provided by wearables.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The EMPATICA E4 wristband has demonstrated reliability and accuracy for HR and HRV research through comparisons with ECG.
“The reliability and accuracy of EMPATICA E4 for research on HR and HRV have been previously demonstrated through comparisons with electrocardiography (ECG) measurements45.”
From Benefits from one session of deep and slow breathing on vagal tone and anxiety in young and older adults (2021) · cites this paper as [45]
The Conclusions report strong agreement with ECG-based VU-AMS recordings for mean HR and SDNN, supporting the E4’s potential for HR and HRV research while people are at rest. The Discussion reports significant differences for other HRV parameters and poor predictive value for RMSSD, so the unqualified claim of reliability and accuracy for HRV research overstates the findings. (LLM, conf=0.88)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
An overview of heart rate variability metrics and norms
Shaffer, F., & Ginsberg, J. P. (2017) · Frontiers in Public Health · doi:10.3389/fpubh.2017.00258 · Cited by 7,185 · see citing works
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
Kubios HRV–heart rate variability analysis software.
Tarvainen, M. P., Niskanen, J.-P., Lipponen, J. A., Ranta-aho, P. O., & Karjalainen, P. A. (2014) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2013.07.024 · Cited by 2,720 · see citing works
Boucsein, W. (2012) · doi:10.1007/978-1-4614-1126-0 · Cited by 2,526 · see citing works
Mccraty, R., & Shaffer, F. (2015) · Global Advances in Health and Medicine · doi:10.7453/gahmj.2014.073 · Cited by 1,039 · see citing works
Fitbit®: An accurate and reliable device for wireless physical activity tracking
Diaz, K. M., Krupka, D. J., Chang, M. J., Peacock, J., Ma, Y., Goldsmith, J., Schwartz, J. E., & Davidson, K. W. (2015) · International Journal of Cardiology · doi:10.1016/j.ijcard.2015.03.038 · Cited by 405 · see citing works
An introduction to heart rate variability: methodological considerations and clinical applications.
Billman, G. E., Huikuri, H. V., Sacha, J., & Trimmel, K. (2015) · Frontiers in Physiology · doi:10.3389/fphys.2015.00055 · Cited by 269 · see citing works
de Zambotti, M., Baker, F. C., Willoughby, A. R., Godino, J. G., Wing, D., Patrick, K., & Colrain, I. M. (2016) · Physiology & Behavior · doi:10.1016/j.physbeh.2016.03.006 · Cited by 222 · see citing works
Methods of assessing vagus nerve activity and reflexes
Chapleau, M. W., & Sabharwal, R. (2011) · Heart Failure Reviews · doi:10.1007/s10741-010-9174-6 · Cited by 182 · see citing works
Flagged first, then by citations
Research works printed but not linked: 24
- [2]De Looff PC Cornet LJM Embregts PJCM Nijman HLI Didden HCM Associations of sympathetic and parasympathetic activity in job stress and burnout: A systematic review PLoS ONE 2018 13 10 e0205741 10.1371/journal.pone.0205741 30335812 PMC6193670 · DOI
- [4]De Geus EJC Willemsen GHM Klaver CHAM Van Doornen LJP Ambulatory measurement of respiratory sinus arrhytmia and respiration rate Biological Psychology 1995 41 205 227 10.1016/0301-0511(95)05137-6 8608201 · DOI
- [5]Willemsen GHM De Geus EJC Klaver CHAM Van Doornen LJP Carroll D Ambulatory monitoring of the impedance cardiogram Psychophysiology 1996 33 184 193 10.1111/j.1469-8986.1996.tb02122.x 8851246 · DOI
- [6]García-Gonzalez MA Fernández-Chimeno M Guede-Fernández F Ferrer-Mileo V Argelagós-Palau A A methodology to quantify the differences between alternative methods of heart rate variability measurement Physiological Measurement 2016 37 128 144 10.1088/0967-3334/37/1/128 26657196 · DOI
- [7]Garbarino, M., Lai, M., Bender, D., Picard, R. W., & Tognetti, S.. Empatica E3 – A wearable wireless multi-sensor device for real-time computerized biofeedback and data acquisition. In EAI 4thInternational Conference on Wireless Mobile Communication and Healthcare – “Transforming healthcare through innovations in mobile and wireless technologies”, 39-42. (2014)
- [8]McCarthy, C., Pradhan, N., Redpath, C., & Adler, A.. Validation of the Empatica E4 wristband. Proceedings of the 2016 IEEE EMBS International Student Conference (ICS), Ottawa, ON, Canada, 29-31 May 2016, 1-4. (2016)
- [9]Ollander, S., Godin, C., Campagne, A., & Charbonnier, S.. A comparison of wearable and stationary sensors for stress detection. Proceedings of the IEEE International Conference on Systems, Man, and Cybernetics, Budapest, Hungary, 9-12 October 2016, 1-5 (2016).
- [10]Caminal P Sola F Gomis P Guasch E Perera A Soriano N Mont L Validity of the Polar V800 for measuring heart rate variability in mountain running route conditions European Journal of Applied Physiology 2018 118 669 677 10.1007/s00421-018-3808-0 29356949 · DOI
- [11]Giles D Draper N Neil W Validity of the Polar V800 heart rate monitor to measure RR intervals at rest European Journal of Applied Physiology 2016 116 563 571 10.1007/s00421-015-3303-9 26708360 PMC4751190 · DOI
- [14]Kang S Kang JM Ko K Park S Mariani S Validity of a commercial wearable sleep tracker in adult insomnia disorder patients and good sleepers Journal of Psychosomatic Research 2017 97 38 44 10.1016/j.jpsychores.2017.03.009 28606497 · DOI · Paper page
- [15]Jarczok MN Jarczok M Mauss D Koenig J Li J Herr RM Thayer JF Autonomic nervous system activity and workplace stressors—A systematic review Neuroscience & Biobehavioral Reviews 2013 37 8 1810 1823 10.1016/j.neubiorev.2013.07.004 23891906 · DOI
- [16]Kamath, M. V., Watanabe, M., & Upton, A.. Heart Rate Variability (HRV) Signal Analysis: Clinical Applications. CRC Press (2016).
- [17]Trimmel, K., Sacha, J., & Huikuri, H. V.. Heart Rate Variability: Clinical Applications and Interaction between HRV and Heart Rate. Frontiers Media SA (2015). PMC4340167 25762937
- [18]Trotman GP Williams SE Quinton ML Veldhuijzen van Zanten JJCS Challenge and threat states: Examining cardiovascular, cognitieve and affective responses to two distinct laboratory stress tasks International Journal of Psychophysiology 2018 126 42 51 10.1016/j.ijpsycho.2018.02.004 29477547 · DOI
- [19]Van Lier, H. G., Pieterse, M E., Garde, A., Postel, M. G., De Haan, H. A., … & Noordzij, M. L. (2019). A standardized assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor. Behaviour Research Methods. 10.3758/s13428-019-01263-9 PMC7148282 31290128 · DOI
- [20]Zheng Y Poon CCY Wearable devices and their applications in surgical robot control and p-medicine 2016 IEEE 20thInternational Conference on Computer Supported Cooperative Work in Design (CSCWD) 2016 Piscataway IEEE Press 659 663
- [21]Picard RW Fedor S Ayzenberg Y Multiple arousal theory and daily-life electrodermal activity asymmetry Emotion Review 2015 8 62 75 10.1177/1754073914565517 · DOI
- [23]Steckler, A., & McLeroy, K. (2008). The importance of external validity. American Journal of Public Health, 98(1), 9–10. PMC2156062 18048772
- [24]Schuurmans, A. A. T., Nijhof, K. S., Scholte, R., Popma, A., & Otten, R. (2020). A novel approach to improve stress regulation among traumatized youth in residential care: feasibility study testing three game-based meditation interventions. Early Intervention in Psychiatry, 14(4), 476–485. 31502420
- [25]Schuurmans, A. A. T., Nijhof, K. S., Scholte, R., Popma, A., & Otten, R. (2020). Effectiveness of game-based meditation therapy on posttraumatic stress and neurobiological stress reactivity in traumatized adolescents: study protocol for a randomized controlled trial. JMIR Reseearch Protocols.10.2196/preprints.19881. PMC7542410 32965226 · DOI
- [27]Bonett DG Sample size requirements for estimating intraclass correlations with desired precision Statistics in Medicine 2002 21 1331 1335 10.1002/sim.1108 12111881 · DOI · Paper page
- [28]Piferi RL Kline KA Younger J Lawler KA An alternative approach for achieving cardiovascular baseline: viewing an aquatic video International Journal of Psychophysiology 2000 37 207 217 10.1016/S0167-8760(00)00102-1 10832007 · DOI · Paper page
- [36]Zheng Y Ding X Poon CC Zhang H Zhou X Unobtrusive sensing and wearable devices for health informatics IEEE Transactions on Biomedical Engineering 2014 61 1538 1554 10.1109/TBME.2014.2309951 24759283 PMC7176476 · DOI
- [40]Radespiel-Tröger M Rauh R Agreement of two different methods for measurement of heart rate variability Clinical Autonomic Research 2003 13 99 102 10.1007/s10286-003-0085-7 12720094 · DOI
Cited by
Sorted by citations
Magnon, V., Dutheil, F., & Vallet, G. T. (2021) · Scientific Reports · doi:10.1038/s41598-021-98736-9 · Cited by 149 · see citing works
Breaking the Flow: A Study of Interruptions During Software Engineering Activities
Ma, Y., Huang, Y., & Leach, K. (2024) · Proceedings of the IEEE/ACM 46th International Conference on Software Engineering · doi:10.1145/3597503.3639079 · Cited by 12 · see citing works
Real-Time Quality Index to Control Data Loss in Real-Life Cardiac Monitoring Applications
Vila, G., Godin, C., Charbonnier, S., & Campagne, A. (2021) · Sensors · doi:10.3390/s21165357 · Cited by 8 · see citing works
- Authors
- Schuurmans, A. A. T., de Looff, P., Nijhof, K. S., Rosada, C., Scholte, R. H. J., Popma, A., & Otten, R.
- Affiliations
- Radboud University Nijmegen (NL)
- De Bascule (NL)
- Corresponding author
- Schuurmans, Angela A. T.
- ORCID
- Venue
- Journal of Medical Systems
- Published
- 2020-09-23
- Volume
- 44
- Issue
- 11
- Pages
- 190-190
- ISSN
- 0148-5598 (ISSN-L)
- 1573-689X
- PMC
- PMC7511462
- OpenAlex
- W3088371165
- Cited by
- 210
- References
- 10
- Citations by year
- 2026: 37 · 2025: 39 · 2024: 44 · 2023: 35 · 2022: 32 · 2021: 22 · 2020: 1
- FWCI
- 13.92
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- Journal of Medical Systems (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- Research portal (Tilburg University) (publishedVersion, other-oa)
- PubMed (publishedVersion)
- Amsterdam UMC (VU Amsterdam) - Institutional Repository (publishedVersion, other-oa)
- Pure Amsterdam UMC (submittedVersion, cc-by)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- Radboud Repository (Radboud University) (submittedVersion)
- Radboud Repository (Radboud University) (submittedVersion, other-oa)
- Funders
- Dr. Couvee Fonds (NA)
- Innovatiefonds Zorgverzekeraars (NA)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- Empatica E4, wristband, heart rate variability, wearable devices, device validation, electrocardiography comparison, SDNN, RMSSD, adolescents, residential care, trauma, autonomic nervous system
- MeSH terms
- Wearable Electronic Devices, Adolescent, Electrocardiography, Heart Rate, Humans, Electrocardiography, Ambulatory, Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review