Wearable Cardiorespiratory Monitoring Employing a Multimodal Digital Patch Stethoscope: Estimation of ECG, PEP, LVET and Respiration Using a 55 mm Single-Lead ECG and Phonocardiogram

Open access

The DOI matches its Crossref record; 56 of the 57 references it cites were found in a registry or an index.

Abstract

Cardiovascular diseases are the main cause of death worldwide, with sleep disordered breathing being a further aggravating factor. Respiratory illnesses are the third leading cause of death amongst the noncommunicable diseases. The current COVID-19 pandemic, however, also highlights the impact of communicable respiratory syndromes. In the clinical routine, prolonged postanesthetic respiratory instability worsens the patient outcome. Even though early and continuous, long-term cardiorespiratory monitoring has been proposed or even proven to be beneficial in several situations, implementations thereof are sparse. We employed our recently presented, multimodal patch stethoscope to estimate Einthoven electrocardiogram (ECG) Lead I and II from a single 55 mm ECG lead. Using the stethoscope and ECG subsystems, the pre-ejection period (PEP) and left ventricular ejection time (LVET) were estimated. ECG-derived respiration techniques were used in conjunction with a novel, phonocardiogram-derived respiration approach to extract respiratory parameters. Medical-grade references were the SOMNOmedics SOMNO HDTM and Osypka ICON-CoreTM. In a study including 10 healthy subjects, we analyzed the performances in the supine, lateral, and prone position. Einthoven I and II estimations yielded correlations exceeding 0.97. LVET and PEP estimation errors were 10% and 21%, respectively. Respiratory rates were estimated with mean absolute errors below 1.2 bpm, and the respiratory signal yielded a correlation of 0.66. We conclude that the estimation of ECG, PEP, LVET, and respiratory parameters is feasible using a wearable, multimodal acquisition device and encourage further research in multimodal signal fusion for respiratory signal estimation.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3390/s20072033

Cites

56 of 56 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure 1 The wearable multimodal stethoscope patch. (a) Front view. (b) Back view with ECG / IP electrode connectors and 28 mm infant stethoscope membrane.
Figure 1 The wearable multimodal stethoscope patch. (a) Front view. (b) Back view with ECG / IP electrode connectors and 28 mm infant stethoscope membrane.CC BY · PubMed Central
Figure 2 Overview of the processing steps and estimated modalities throughout this work.
Figure 2 Overview of the processing steps and estimated modalities throughout this work.CC BY · PubMed Central
Figure 3 (a) PEP and LVET definitions. (b) Sensor placement throughout the study.
Figure 3 (a) PEP and LVET definitions. (b) Sensor placement throughout the study.CC BY · PubMed Central
Figure 4 Neural network regression model configurations used throughout this work: (a) multi-layer perceptrons. (b) time-delay neural networks.
Figure 4 Neural network regression model configurations used throughout this work: (a) multi-layer perceptrons. (b) time-delay neural networks.CC BY · PubMed Central
Figure 5 Example of the data split in forward-chaining cross-validation with nCV=3 validation rounds.
Figure 5 Example of the data split in forward-chaining cross-validation with nCV=3 validation rounds.CC BY · PubMed Central
Figure 6 ECG and PCG peak detection and classification of S1 and S2 candidates before PEP and LVET estimation.
Figure 6 ECG and PCG peak detection and classification of S1 and S2 candidates before PEP and LVET estimation.CC BY · PubMed Central
Figure 7 Beat-to-beat (B2B) ECG and PCG derived respiration for estimating the respiratory flow F^ and rate RR^ using the estimated ECG and the PCG signals in a model and feature selection framework.
Figure 7 Beat-to-beat (B2B) ECG and PCG derived respiration for estimating the respiratory flow F^ and rate RR^ using the estimated ECG and the PCG signals in a model and feature selection framework.CC BY · PubMed Central
Figure 8 BIC model selection for Einthoven ECG lead estimation, TDNNs only: (a) Selection of models and lines of constant lag length. (b) Minimum BIC values. (c) BIC over the number of delay line lags.
Figure 8 BIC model selection for Einthoven ECG lead estimation, TDNNs only: (a) Selection of models and lines of constant lag length. (b) Minimum BIC values. (c) BIC over the number of delay line lags.CC BY · PubMed Central
Figure 9 Example ECG signals of Subject 1 in the supine position: (a) Patch ECG. (b) Reference and estimated Einthoven I ECGs. (c) Reference and estimated Einthoven II ECGs.
Figure 9 Example ECG signals of Subject 1 in the supine position: (a) Patch ECG. (b) Reference and estimated Einthoven I ECGs. (c) Reference and estimated Einthoven II ECGs.CC BY · PubMed Central
Figure 10 Example ECG signals of Subject 2 in the supine position: (a) Patch ECG. A high intersubject variability of the patch ECG lead can be observed when compared to Figure 9a. (b) Reference and estimated Einthoven I ECGs. (c) Reference and estimated Einthoven II ECGs.
Figure 10 Example ECG signals of Subject 2 in the supine position: (a) Patch ECG. A high intersubject variability of the patch ECG lead can be observed when compared to Figure 9a. (b) Reference and estimated Einthoven I ECGs. (c) Reference and estimated Einthoven II ECGs.CC BY · PubMed Central
Figure 11 ECG and PCG signals with detected fiducial points.
Figure 11 ECG and PCG signals with detected fiducial points.CC BY · PubMed Central
Figure 12 Bland–Altman plots of the STI estimates. (a) PEP. (b) LVET.
Figure 12 Bland–Altman plots of the STI estimates. (a) PEP. (b) LVET.CC BY · PubMed Central
Figure 13 Performance plots of the EDR and PDR based RR estimations over all positions including outliers. (a) Correlation optimization. (b) Respiration rate optimization.
Figure 13 Performance plots of the EDR and PDR based RR estimations over all positions including outliers. (a) Correlation optimization. (b) Respiration rate optimization.CC BY · PubMed Central
Figure 14 Example reference and estimated respiratory signal using ECG- and PCG-derived respiration.
Figure 14 Example reference and estimated respiratory signal using ECG- and PCG-derived respiration.CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 56 references

Research works printed but not linked: 2

  1. Choudhary, T., Bhuyan, M., Bora, K., and Sharma, L. (2020). Design of Breathing-states Detector for m-Health Platform using Seismocardiographic Signal. arXiv.
  2. Klum, 2019, Impedance Plethysmography for Respiratory Flow and Rate Estimation using Multilayer Perceptrons, Int. J. Bioelectromagn., 21, 34

Cited by

Metadata

Authors
Klum, M., Urban, M., Tigges, T., Pielmus, A.-G., Feldheiser, A., Schmitt, T., & Orglmeister, R.
Affiliations
  • Technische Universität Berlin (DE)
  • Department of Electronics and Medical Signal Processing, Technische Universität Berlin, Einsteinufer 17, 10587 Berlin, Germany
  • Kliniken Essen-Mitte (DE)
Corresponding author
Klum, Michael
Venue
Sensors
Publisher
MDPI AG
Published
2020-04-04
Volume
20
Issue
7
Pages
2033
ISSN
1424-8220 (ISSN-L)
OpenAlex
W3014676120
Cited by
75
References
57
Citations by year
2026: 5 · 2025: 13 · 2024: 8 · 2023: 13 · 2022: 15 · 2021: 10 · 2020: 10
FWCI
3.77
Type
journal article
Language
en
Open access
Locations
  • Sensors (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DepositOnce (submittedVersion, cc-by)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • Informationszentrum Lebenswissenschaften (ZB MED) (submittedVersion, cc-by)
  • Technische Universität Berlin – Universitätsbibliothek (cc-by)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
  • Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Phonocardiography and Auscultation Techniques
Keywords
single-lead ECG, ECG-derived respiration, respiratory rate estimation, multimodal signal fusion, sleep-disordered breathing, posture effects
MeSH terms
Wearable Electronic Devices, Electrocardiography, Heart Ventricles, Humans, Phonocardiography, Ventricular Function, Respiratory Rate

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.