Reprocessed Single-Use Semicritical and Critical Medical Devices
Open accessAbstract
What Is the Issue? Reprocessing a medical device includes cleaning, reconditioning, testing, and disinfection to ensure the device can safely be reused. In contrast to reusable medical devices, manufacturers are not required to provide instructions for properly cleaning and sterilizing single-use medical devices (SUMDs). Health Canada regulates third-party device reprocessors and requires they meet the same requirements as new device manufacturers. Health Canada does not provide oversight for hospital onsite reprocessing, deferring to the oversight provided at the provincial and territorial levels. Given the potential economic and environmental benefits of using reprocessed SUMDs, there is a growing interest in determining the clinical safety of reprocessed SUMDs. Current standards for reprocessing medical devices use definitions for sterilization and disinfection based on measurement of bioburden, but not necessarily clinical outcomes such as infection. What Did We Do? To inform decisions about the appropriate use of reprocessed critical and semicritical SUMDs, CADTH sought to identify and summarize literature evaluating the clinical safety of reprocessed SUMDs, defined as infections, mortality, or other adverse events, compared with nonreprocessed (new) SUMDs. Microbiological outcomes, such as bacterial colony counts, were not included. An information specialist searched for peer-reviewed and grey literature sources. This report does not provide a comprehensive list of device reprocessors in Canada or recommend any specific methods of reprocessing medical devices. What Did We Find? We identified 8 studies, including one study based in Canada, that evaluated the use of reprocessed SUMDs compared with new SUMDs; most did not report statistically significant differences in patient outcomes between groups. Most of the included studies were of very low to moderate quality, which limits confidence in the observed outcomes resulting from the reuse of these devices. Half of the included studies were published before the year 2005, which may limit applicability given potential improvements and changes over time in reprocessing standards, surgical approaches, device specifications, and patient care protocols. Most of the studies evaluated a different type of reprocessed single-use medical device for different surgical populations, so there is very limited evidence for the use of a specific device in a specific population or intervention of interest. All included studies evaluated SUMDs classified as critical, and all were conducted in surgical settings; however, it is unclear whether patient risk levels would be different for semicritical devices or in nonsurgical settings. What Does it Mean? Given various devices, clinical applications, and reprocessing methods, it is difficult to draw broad conclusions about the appropriateness of reprocessing SUMDs. While the evidence base in this review was insufficient to conclude whether reprocessed critical SUMDs in surgical settings affect patient outcomes, Canadian standards and other resources exist to help inform decisions around medical device reprocessing based on infection risk. To ensure patient safety, any reprocessing of SUMDs should meet standards for safety, effectiveness, and labelling that follow Health Canada regulations.
AI snapshot
Written from the paper’s full text.
TL;DR
This CADTH rapid review identified 8 studies comparing reprocessed single-use medical devices (SUMDs) with new SUMDs; most found no statistically significant differences in patient outcomes, but the evidence was heterogeneous and of very low to moderate quality. All included studies evaluated critical SUMDs in surgical settings, and none evaluated semicritical devices.
Key findings
- Eight studies met inclusion criteria (1 RCT, 7 nonrandomized); most did not report statistically significant differences in patient outcomes. Only 2 studies evaluated the same device in the same population (balloon catheters for coronary angioplasty) and reached different conclusions.
- For balloon catheters, Plante et al. (1994) found the reprocessed group had more catheters per lesion (2.4 vs 1.2, P < 0.00001), higher initial balloon failure (10.2% vs 3.3%, P < 0.0001), longer procedure times (81 vs 68 minutes, P < 0.0001), more contrast medium (201 vs 165 mL, P < 0.0001), longer hospital stay (5.1 vs 3.4 days, P < 0.0001), and higher clinical failure with adverse events (7.8% vs 3.8%, P < 0.025). Browne et al. (1997) reported no significant difference in fever or WBC count and a 7% device failure rate, which the authors compared with a 10% literature rate for new balloons; P values were not reported.
- For other devices, Jokar et al. (2022) found higher overall port-site infection with reprocessed ports (8.5% vs 1.9%, P = 0.002); Marins et al. (2018) found no significant differences in postoperative infection (12.3% vs 13.4%, P = 0.664), mortality (1.90% vs 2.88%, P = 0.396), or rehospitalization (P = 0.678); Colak et al. (2004) found similar complications (3.2% vs 4.8%, P = 0.50) and infection (1.6% vs 3.2%, P = 0.57); Leung et al. (2019) found no complications and shorter fluoroscopy duration in the remanufactured catheter group (11.8 vs 21.5 minutes, P < 0.0001).
- All included studies evaluated critical SUMDs in surgical settings; none evaluated semicritical devices. Zacharias et al. (2021) reported lower functionality for reprocessed endolaser probes (93.5% vs 99.47%, P < 0.001; OR = 12.99) but zero infectious endophthalmitis in both groups; Gundogdu et al. (1998) reported no wound-site or intra-abdominal infections.
Methods
- Rapid review with an information specialist search of MEDLINE, the Cochrane Database of Systematic Reviews, the International HTA Database, Canadian and major international health technology agency websites, and a focused internet search; search completed September 7, 2023.
- One reviewer screened citations and selected studies using inclusion criteria in Table 1; articles were excluded if they did not report patient-related outcomes or if they evaluated reused but not reprocessed devices. Eight publications were included: 1 RCT and 7 nonrandomized studies.
- Included publications were critically appraised by 1 reviewer using the Downs and Black checklist; no summary scores were calculated, and strengths and limitations were described narratively.
Limitations
- Stated in the paper: only 1 included study was conducted in Canada (Plante et al., 1994); 4 of 8 studies were published before 2005; most evidence was very low to moderate quality; only 1 high-quality RCT was included (2004, Turkey, in-hospital reprocessing); studies were heterogeneous in devices, populations, outcomes, and reprocessing protocols; none evaluated semicritical devices; all were in surgical settings; most used in-house rather than third-party reprocessing; follow-up was often limited and the number of reprocessing cycles was often unclear.
- Evident from the text: many included observational studies lacked randomization, allocation concealment, and blinding; some used historical controls or between-site comparisons with baseline differences (e.g., Plante et al. had higher unstable angina in the reprocessed group); several studies did not report power calculations or loss to follow-up.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 109 cited sources checked · 0 supported · 0 flagged
- Findingown data
In Browne et al., most outcomes were not reported with statistical significance or statistical ranges, including patient safety outcomes.
“Most outcomes were not reported with statistical significance or statistical ranges, including patient safety outcomes.”
Quote in the paperNot a substantial claim - Findingown data
Half of the included studies were published before 2005.
“Half of the included studies were published before the year 2005”
Quote in the paperSubstantial claim - Findingown data
In Browne et al., very limited outcomes were reported for the historical comparison group.
“Very limited outcomes were reported for the historical comparison group.”
Quote in the paperNot a substantial claim - Findingcitation
Zacharias et al. reported on a prospective cohort study comparing reprocessed single-use endolaser probes with new probes in vitrectomy surgeries.
“Zacharias et al. 20 reported on a prospective cohort study comparing reprocessed single-use endolaser probes with new probes in vitrectomy surgeries.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [20]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingown data
Most included studies did not report statistically significant differences in patient outcomes between reprocessed and new single-use medical devices.
“We identified 8 studies, including one study based in Canada, which evaluated the use of reprocessed SUMDs compared with new SUMDs; most did not report statistically significant differences in patient outcomes between groups.”
Quote in the paperSubstantial claim
- Findingown data
Most included studies evaluated a different type of reprocessed single-use medical device for different surgical populations.
“Most of the studies evaluated a different type of reprocessed single-use medical device for different surgical populations, so there is very limited evidence for the use of a specific device in a specific population or intervention of interest.”
Quote in the paperSubstantial claim - Findingcitation
Plante et al. 17 reported hospital stay was significantly greater in the reprocessing group (5.1 vs 3.4 days, P < 0.0001).
“Hospital stay was significantly greater in the reprocessing group (5.1 days versus 3.4 days, P < 0.0001).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Marins et al. 22 reported rehospitalization was required for 28 patients in the reprocessing group and for 33 patients in the nonreprocessed group (P = 0.678).
“Rehospitalization was required for 28 patients in the reprocessing group and for 33 patients in the nonreprocessed group (P = 0.678).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Colak et al. 16 reported median satisfaction rate of the patients was excellent in both groups (P = 0.87).
“The median satisfaction rate of the patients was excellent in both groups (P = 0.87)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingown data
None of the included studies evaluated semicritical use devices.
“None of the included studies evaluated semicritical use devices”
Quote in the paperSubstantial claim - Findingcitation
Leung et al. 2 reported procedure duration was similar in the remanufactured group (178.9 minutes) and the matched new catheter control group (189.5, P = 0.16).
“Leung et al.2 reported that the procedure duration was similar in the remanufactured group (178.9 minutes) and the matched new catheter control group (189.5),P = 0.16).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [2]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingown data
This Rapid Review identified 8 studies evaluating the use of reprocessed SUMDs compared with new SUMDs.
“This Rapid Review identified 8 studies evaluating the use of reprocessed SUMDs compared with new SUMDs”
Quote in the paperSubstantial claim - Findingcitation
Jokar et al. 21 reported incidence of PSI was significantly higher in the reprocessed port group than in the new port group at all evaluated time points up to 2 weeks after surgery.
“Jokar et al.21 reported that the incidence of PSI was significantly higher in the reprocessed port group than in the new port group at all evaluated time points up to 2 weeks after surgery.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [21]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Plante et al. 17 reported higher incidence of initial balloon failure (10.2% vs 3.3%, P < 0.0001).
“and had a higher incidence of initial balloon failure (10.2% versus 3.3%, P < 0.0001).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Plante et al. 17 reported no statistically significant differences in clinical success and failure between the 2 groups.
“No statistically significant differences in clinical success and failure were reported between the 2 groups.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingown data
All of the included studies took place in surgical settings.
“and all of the included studies took place in surgical settings.”
Quote in the paperNot a substantial claim - Findingcitation
Marins et al. 22 reported differences in postoperative infection rate between the reprocessed and nonreprocessed group were not statistically significant.
“Marins et al. 22 reported that the differences in postoperative infection rate between the reprocessed and nonreprocessed group were not statistically significant.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Marins et al. 22 reported no statistically significant difference in mortality (P = 0.396).
“There was no statistically significant difference in mortality (P = 0.396).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingnone given
The majority of studies did not report statistically significant impacts on patient outcomes.
“The majority of studies did not report statistically significant impacts on patient outcomes.”
Quote in the paperSubstantial claim - Findingreasoning
The primary outcome of interest in Gundogdu et al. 18 seemed to be culturing bacteria on surgical equipment and at the patient's surgical site rather than actual patient outcomes such as infection.
“The primary outcome of interest seemed to be on culturing bacteria on surgical equipment and at the patient's surgical site rather than on actual patient outcomes such as infection.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [18]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
The frequency of infectious endophthalmitis was zero in both groups in Zacharias et al. 20.
“The frequency of infectious endophthalmitis was zero in both groups.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [20]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Plante et al. 17 reported the reprocessing group exhibited longer procedure times (81 vs 68 minutes, P < 0.0001).
“The reprocessing group exhibited longer procedure times (81 minutes versus 68 minutes, P < 0.0001)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Plante et al. 17 reported fever in 3 patients in the reprocessing group and in 1 patient in the single-use centre.
“Plante et al. 17 reported that fever was noted in 3 patients in the reprocessing group (in association with urinary tract infection, pneumonia and the postoperative period after urgent bypass surgery) and in 1 patient in the single-use centre (flulike illness)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Other causes explained the elevated WBC in the reprocessing group in 9 patients (MI in 4, hematoma in 3, steroids in 1, and urinary tract infection in 1).
“Other causes explained the elevated WBC in the reprocessing group in 9 patients (MI in 4, hematoma in 3, steroids in 1, and urinary tract infection in 1).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Marins et al. 22 reported length of hospital stay was approximately 10 days in both groups (P = 0.881).
“and length of hospital stay was approximately 10 days in both groups (P = 0.881).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Leung et al. 2 reported statistically significant higher overall fluoroscopy duration in the new catheter group (21.5 minutes) versus the remanufactured group (11.8 minutes, P < 0.0001).
“There was a statistically significant higher overall fluoroscopy duration in the new catheter group (21.5 minutes) versus the remanufactured group (11.8 minutes), P < 0.0001).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [2]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Plante et al. 17 reported rate of clinical failure with adverse clinical events was significantly higher in the reuse group (7.8% vs 3.8%, P < 0.025).
“Still, the rate of clinical failure with adverse clinical events was significantly higher in the reuse group (7.8% versus 3.8%.P < 0.025).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Zacharias et al. 20 reported statistically significant functionality rates between groups.
“and reported statistically significant functionality rates between groups.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [20]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
The second phase of Zacharias et al. 20 included 590 endolaser probes, of which 375 were original and 215 were reprocessed.
“The second phase included 590 endolaser probes, of which 375 were original and 215 were reprocessed”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [20]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Plante et al. 17 reported angiographic success rate in both the reprocessed and single-use groups was identical at 88%.
“Plante et al. 17 reported that the angiographic success rate in both the reprocessed and single-use groups was identical at 88%.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Leung et al. 2 reported 1 mapping catheter failure in the manufactured group versus no failures in the new use group.
“There was 1 mapping catheter failure in the manufactured group versus no failures in the new use group.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [2]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
None of the patients in Gundogdu et al. 18 had infection at the wound site or intra-abdominally.
“None of the patients had infection at the wound site or intra-abdominally.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [18]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Leung et al. 2 reported no minor or major complications in either group.
“There were no minor or major complications in either group.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [2]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Marins et al. 22 reported no statistically significant differences in antibiotic consumption.
“There were also no statistically significant differences in antibiotic consumption.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
In Plante et al. 17, in all cases fever did not appear to be related to the catheterization procedure.
“and in all cases, fever did not appear to be related to the catheterization procedure.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Plante et al. 17 reported the reprocessing site utilized more balloon catheters per lesion (mean 2.4 vs 1.2, P < 0.00001).
“The reprocessing site utilized more balloon catheters per lesion (mean 2.4 catheters in the reprocessing group versus 1.2 in the single-use group, P < 0.00001)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Plante et al. 17 reported the reprocessing group utilized an increased volume of contrast medium (201 vs 165 mL, P < 0.0001).
“and utilized an increased volume of contrast medium.(201 mL versus 165 mL, P < 0.0001)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingcitation
Plante et al. 17 reported increased adverse events, particularly in patients with unstable angina (7.8% vs 3.8%, P < 0.25).
“and increased adverse events, particularly in patients with unstable angina (7.8% versus 3.8%, P < 0.25).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingnone given
The majority of studies did not report statistically significant differences in patient outcomes with the use of reprocessed SUMDs.
“The majority of studies did not report statistically significant differences in patient outcomes with the use of reprocessed SUMDs.”
Quote in the paperSubstantial claim - Findingcitation
Marins et al. 22 reported a wide range of differences in postoperative infection rate depending on the severity of the condition and the type of surgery.
“There was a wide range of differences in postoperative infection rate depending on the severity of the condition and the type of surgery.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning
In Leung et al., there are bias and confounding concerns due to the lack of randomization and blinding, typical of observational studies.
“most importantly, there are bias and confounding concerns due to the lack of randomization and blinding, which is typical of observational studies.”
Quote in the paperNot a substantial claim - Premisecitation
Provincial policies generally refer to using third-party commercial processors in accordance with Health Canada requirements.
“Provincial policies generally refer to using third-party commercial processors in accordance with Health Canada requirements.7,8”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [7]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [8]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning
The evidence base in this review was insufficient to conclude whether reprocessed critical single-use medical devices in surgical settings affect patient outcomes.
“While the evidence base in this review was insufficient to conclude whether reprocessed critical SUMDs in surgical settings affect patient outcomes, Canadian standards and other resources exist to help inform decisions around medical device reprocessing based on infection risk.”
Quote in the paperSubstantial claim - Interpretationreasoning
The pre-2005 publication of half the included studies may limit applicability given potential improvements and changes over time in reprocessing standards, surgical approaches, device specifications, and patient care protocols.
“which may limit applicability given potential improvements CADTH Health Technology Review Key Messagesand changes over time in reprocessing standards, surgical approaches, device specifications, and patient care protocols.”
Quote in the paperSubstantial claim - Interpretationreasoning
In Plante et al., the use of a control group at a second centre has the potential for differences in patient groups, practice patterns, and other factors that could confound observed results.
“the use of a control group at a second centre has the potential for differences in patient groups, practice patterns and other factors that could confound the observed results.”
Quote in the paperSubstantial claim - Interpretationreasoning
In Plante et al., the observed outcomes may have been at least partly due to differences in the groups at baseline.
“Therefore, the observed outcomes (such as longer procedure duration, adverse events, etc.) may have been at least partly due to differences in the groups at baseline.”
Quote in the paperSubstantial claim - Interpretationnone given
In Jokar et al., follow-up may not have been sufficiently long-term to assess all adverse events.
“Follow-up may not have been sufficiently long-term to assess all adverse events.”
Quote in the paperNot a substantial claim - Interpretationreasoning
Given various devices, clinical applications, and reprocessing methods, it is difficult to draw broad conclusions about the appropriateness of reprocessing single-use medical devices.
“Given various devices, clinical applications, and reprocessing methods, it is difficult to draw broad conclusions about the appropriateness of reprocessing SUMDs.”
Quote in the paperSubstantial claim - Interpretationreasoning
There is very limited evidence for the use of a specific reprocessed single-use medical device in a specific population or intervention of interest.
“Most of the studies evaluated a different type of reprocessed single-use medical device for different surgical populations, so there is very limited evidence for the use of a specific device in a specific population or intervention of interest.”
Quote in the paperSubstantial claim - Interpretationreasoning
Due to heterogeneity across studies, including reported outcomes, it is difficult to make comparisons or draw conclusions about the safety of reusing SUMDs.
“Due to the heterogeneity across studies, including reported outcomes, it is difficult to make comparisons or draw conclusions about the safety of reusing SUMDs.”
Quote in the paperSubstantial claim - Study design and conductreasoning
The Gundogdu et al. 18 study had very little focus on patient outcomes.
“There was very little focus on patient outcomes.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [18]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning
The lack of observed differences between groups in Colak et al. 16 may have been affected by the absence of a power calculation.
“which may have affected the lack of observed differences between groups.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning
One can have confidence in the results of the Colak et al. 16 study.
“and that 1 can have confidence in the results of the study.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation
Colak et al. 16 conducted an RCT evaluating the reuse of disposable laparoscopic instruments compared with new use.
“16 conducted an RCT evaluating the reuse of disposable laparoscopic instruments compared with new use.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation
Gundogdu et al. 18 conducted a prospective cohort comparing new versus reprocessed single-use plastic trocars.
“Gundogdu et al. 18 conducted a prospective cohort comparing new versus reprocessed single-use plastic trocars.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [18]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Study design and conductnone given
In the included studies, it was not always clear how often a SUMD was reprocessed, which is an important consideration in any reprocessing policy.
“In the included studies, it was not always clear how often a SUMD was reprocessed, which is an important consideration in any reprocessing policy.”
Quote in the paperSubstantial claim - Interpretationreasoning
The Colak et al. 16 study results are generalizable to similar populations.
“meaning that the study results are generalizable to similar populations”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Study design and conductcitation
There is just 1 high-quality RCT (Colak et al. 16) that evaluated the use of reprocessed SUMDs.
“and there is just 1 high-quality RCT 16 that evaluated the use of reprocessed SUMDs.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Study design and conductreasoning
The body of evidence was appraised to be of very low to moderate quality.
“The body of evidence was appraised to be of very low to moderate quality”
Quote in the paperSubstantial claim - Interpretationreasoning
It is difficult to compare outcomes or draw conclusions on the safety of reprocessed SUMDs based on the existing body of evidence since the population under study, the intervention, the outcomes reported, and the method of reprocessing varies.
“It is difficult to compare outcomes or draw conclusions on the safety of reprocessed SUMDs based on the existing body of evidence since the population under study, the intervention, the outcomes reported, and the method of reprocessing varies.”
Quote in the paperSubstantial claim - Premisecitation
In community health care settings, reprocessing of critical and semicritical medical equipment/devices, when not performed according to current standards, has been linked with health care-associated infections and outbreaks.
“In community health care settings (defined as any location outside of an acute care hospital where health care is provided), reprocessing of critical and semicritical medical equipment/devices, when not performed according to current standards, has been linked with health care-associated infections and outbreaks.23”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning
In Marins et al., without a power calculation, it is unclear whether nonsignificant differences between groups were real or related to underpowering of the study to detect a difference between groups.
“Without a power calculation, it is unclear whether nonsignificant differences between groups were real or related to the underpowering of the study to detect a difference between groups.”
Quote in the paperNot a substantial claim - Study design and conductnone given
The patients in both groups in Browne et al. (1997) were recruited from the same population.
“The patients in both groups were recruited from the same population”
Quote in the paperNot a substantial claim - Study design and conductnone given
No attempt was made to blind patients or outcome assessors in Plante et al. (1994).
“No attempt to blind patients or outcome assessors was”
Quote in the paperNot a substantial claim - Study design and conductnone given
The patients in the different intervention groups in Plante et al. (1994) were not recruited from the same population.
“The patients in the different intervention groups were not recruited from the same population”
Quote in the paperNot a substantial claim - Study design and conductnone given
There was adequate adjustment for confounding factors between treatment groups at baseline in Jokar et al. (2022).
“There was adequate adjustment for confounding factors between treatment groups at baseline”
Quote in the paperNot a substantial claim - Study design and conductnone given
There was adequate adjustment for confounding factors between treatment groups at baseline in Colak et al. (2004).
“There was adequate adjustment for confounding factors between treatment groups at baseline”
Quote in the paperNot a substantial claim
- May combine two claims. 83
- Classified as probably not a substantial claim. 157
- The quote only partly states this claim. 3
May combine two claims.
Findingown dataMost included studies were of very low to moderate quality, which limits confidence in the observed outcomes resulting from reuse of these devices.
“Most of the included studies were of very low to moderate quality, which limits confidence in the observed outcomes resulting from the reuse of these devices.”
Quote in the paperSubstantial claimMay combine two claims.
FindingcitationThe Plante et al. study was a prospective cohort study evaluating balloon catheters for coronary angioplasty at 2 sites in the same province in Canada, with 1 site using reprocessed single-use catheters and 1 site using new catheters.
“The Plante et al. 17 study was a prospective cohort study evaluating balloon catheters for coronary angioplasty at 2 sites in the same province in Canada, with 1 site using reprocessed single-use catheters and 1 site using new catheters.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataIn Leung et al., loss to follow-up was not reported and a power calculation was not described.
“Loss to followup was not reported, a power calculation was not described”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn Browne et al., follow-up for fever, chills, and elevated WBC was reported only up to 24 hours postprocedure, and patients were followed only until hospital discharge, with no long-term follow-up.
“Follow-up for fever, chills and elevated WBC was reported only up to 24 hours postprocedure, and therein patients were followed only until hospital discharge.There was no long-term follow-up.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataAll included studies evaluated single-use medical devices classified as critical, and all were conducted in surgical settings.
“All included studies evaluated SUMDs classified as critical, and all were conducted in surgical settings; however, it is unclear whether patient risk levels would be different for semicritical devices or in nonsurgical settings.”
Quote in the paperSubstantial claimMay combine two claims.
FindingreasoningIn Plante et al., at baseline there were statistically significant differences in the incidence of unstable angina (and related IV heparin and nitroglycerin use before the procedure) between the 2 centres, which may suggest that more high-risk lesions were treated in the reprocessed device group.
“At baseline, there were statistically significant differences in the incidence of unstable angina (and related IV heparin and nitroglycerin use before the procedure) between the 2 centres, which may suggest that more high-risk lesions were treated in the reprocessed device group.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingnone givenIn Browne et al., the study was powered to detect a 5% difference in angiographic failure rates of new and reused balloons, yet the failure rate for the comparison group did not appear to be provided.
“The study was powered to detect a 5% difference in the angiographic failure rates of new and reused balloons, yet the failure rate for the comparison group did not appear to be provided.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataIn Jokar et al., there was no description of blinding of patients or outcome assessors or allocation concealment.
“There was no description of blinding of patients or outcome assessors or allocation concealment.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingnone givenIn Zacharias et al., there was no description of the patient groups or any indication of how patients were selected or assigned to treatment.
“There was no description of the patient groups or any indication of how the patients were selected or assigned to treatment.”
Quote in the paperSubstantial claimMay combine two claims.
Findingnone givenIn Browne et al., the retroactive comparison group was not well-described and a list of potential confounders was not provided.
“The retroactive comparison group was not well-described, and a list of potential confounders was not provided.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe literature search identified 8 publications meeting inclusion criteria, comprising 1 randomized controlled trial and 7 nonrandomized studies.
“Of these potentially relevant articles, 25 publications were excluded for various reasons, and 8 publications met the inclusion criteria and were included in this report.These comprised 1 randomized controlled trial (RCT), and 7 nonrandomized studies.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
FindingreasoningIn Jokar et al., there appears to be a small loss to follow-up of 3 patients that was not reported or addressed by the authors, though this small number is unlikely to have impacted observed outcomes.
“There appears to be a small loss to follow-up of 3 patients that was not reported or addressed by the authors, though this small number is unlikely to have impacted the observed outcomes.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
FindingcitationBrowne et al. 15 reported that it is unknown how many cycles of restoration a single balloon angioplasty catheter can tolerate, and that data are required to evaluate the degradation of these products with each subsequent cycle.
“Browne et al. 15 reported that it is unknown how many cycles of restoration a single balloon angioplasty catheter can tolerate, and that data are required to evaluate the degradation of these products with each subsequent cycle.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Classified as probably not a substantial claim. May combine two claims.
FindingcitationPlante et al. 17 emphasized that before a policy of balloon catheter reuse is initiated, it is imperative to establish clear and rigorous guidelines for cleaning and sterilization, with adequate monitoring and quality control in place at all stages.
“Plante et al. 17 have emphasized that before a policy of balloon catheter reuse is initiated, it is imperative to establish clear and rigorous guidelines for cleaning and sterilization, and with adequate monitoring and quality control in place at all stages.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
FindingcitationBrowne et al. 15 reported incidence of device failure was 7%, which the authors claim compares with a 10% rate reported in the literature for new balloon catheters.
“The incidence of device failure was 7%, which the authors claim compares with a 10% rate reported in the literature for new balloon catheters.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationMarins et al. 22 reported no reoperations in either group.
“There were no reoperations in either group.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Findingown dataIn Marins et al., length of follow-up and loss to follow-up were not reported, and a power calculation was not performed.
“Length of follow-up and loss to follow-up were not reported, and a power calculation was not performed.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationMarins et al. 22 reported duration of the surgical intervention was on average 2 hour and 23 minute (P = 0.161).
“The duration of the surgical intervention was on average 2 hour and 23 minute (P = 0.161)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingcitationColak et al. 16 reported total incidence of complications (3.2% vs 4.8%, P = 0.50) and infection rates (1.6% vs 3.2%, P = 0.57) were similar.
“The total incidence of complications (3.2% vs 4.8%, P = 0.50) and infection rates (1.6% vs 3.2%, P = 0.57) were also similar.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Classified as probably not a substantial claim. May combine two claims.
FindingcitationBrowne et al. 15 emphasized important differences between in vitro evaluation and in vivo evaluation and stated that methods of restoration must be highly controlled and reproducible.
“Browne et al. 15 emphasized the important differences between in vitro evaluation and in vivo evaluation and stated that methods of restoration must be highly controlled and reproducible.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Findingnone givenMost of the included studies had limited follow-up, so long-term outcomes for many patients were not available.
“Most of the included studies had limited follow-up, so long-term outcomes for many patients were not available.”
Quote in the paperSubstantial claimMay combine two claims.
FindingcitationThree studies evaluated the same study population, patients undergoing laparoscopic cholecystectomy, but each evaluated a different intervention: disposable laparoscopic instruments, plastic trocars, and surgical ports.
“Three studies evaluated the same study population, patients undergoing laparoscopic cholecystectomy, but in each case evaluated a different intervention: disposable laparoscopic instruments, 16 plastic trocars, 18 and surgical ports.21”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [18]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [21]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingcitationAmong reprocessed probes in Zacharias et al. 20, 201 (93.5%) were functioning and 14 (6.5%) were nonfunctioning (P < 0.001) for an odds ratio OR = 12.99 (P < 0.001).
“Among the reprocessed ones, 201 (93.5%) were functioning and 14 (6.5%) were nonfunctioning (P < 0.001) for an odds ratio OR = 12.99 (P < 0.001).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [20]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingcitationThe 2 studies that reported on the same intervention (balloon catheters) in the same population (coronary angiography) came to different conclusions, with Plante et al. 17 reporting higher rates of adverse events and other procedure-related issues in the reprocessed SUMD group, while Browne et al. 15 described much more favourable results.
“The 2 studies that reported on the same intervention (balloon catheters) in the same population (coronary angiography) came to different conclusions, with Plante et al 17 reporting higher rates of adverse events and other procedure-related issues in the reprocessed SUMD group, while Browne et al 15 described much more favourable results.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Findingown dataMost of the 8 studies employed a different reprocessing protocol and evaluated different reprocessed SUMDs in patients undergoing different surgical interventions.
“most of which employed a different reprocessing protocol, and evaluated different reprocessed SUMDs in patients undergoing different surgical interventions.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingnone givenGundogdu et al. 18 focused mainly on nonpatient outcomes in a prospective cohort study.
“Gundogdu et al. 18 focused mainly on nonpatient outcomes in a prospective cohort study.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [18]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingcitationPlante et al. 17 reported overall clinical success rate and rate of clinical failure without adverse clinical events were comparable in the 2 groups.
“Overall clinical success rate and the rate of clinical failure without adverse clinical events were comparable in the 2 groups.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingcitationColak et al. 16 reported no statistically significant difference between the new and the reprocessed disposable laparoscopic instrument groups in mean operating time, LAS, duration and amount of analgesic administration, or hospital stay.
“Colak et al. 16 reported that there was no statistically significant difference between the new and the reprocessed disposable laparoscopic instrument groups in mean operating time, LAS, duration and amount of analgesic administration, or hospital stay.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationDevice failure was not reported for the comparison group in the retrospective review in Browne et al. 15.
“Device failure was not reported for the comparison group in the retrospective review.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationThe only patient outcome reported in Gundogdu et al. 18 was infection, evaluated using Cruise's criteria.
“The only patient outcome reported was infection which was evaluated using Cruise's criteria.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [18]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingcitationIn Jokar et al. 21, all PSIs were considered superficial, and no symptoms of deep infection were reported.
“All PSIs were considered superficial, and no symptoms of deep infection were reported.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [21]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingcitationFor laparoscopic cholecystectomy patients, use of reprocessed surgical ports led to a significant increase in PSI, while use of reprocessed plastic trocars and other disposable laparoscopic instruments reportedly did not increase infection rates.
“For laparoscopic cholecystectomy patients, the use of reprocessed surgical ports led to a significant increase in PSI, 21 while the use of reprocessed plastic trocars and other disposable laparoscopic instruments reportedly did not increase infection rates.16,18”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [21]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [18]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationBrowne et al. 15 reported mean procedure time was 67 minutes in the reprocessing group compared with 83 minutes in the comparison group.
“Browne et al. 15 reported that the mean procedure time was 67 minutes in the reprocessing group compared with 83 minutes in the comparison group.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationBrowne et al. 15 reported mean fluoroscopy time was 13 minutes in the reprocessing group compared with 18 minutes in the comparison group.
“The mean fluoroscopy time for the procedures was 13 minutes in the reprocessing group compared with 18 minutes in the comparison group.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Findingown dataOf the 8 included studies, only 2 studies evaluated the same device in the same population (balloon catheters for coronary angiography 15,17) while the rest evaluated a variety of devices across different populations.
“Of the 8 included studies, only 2 studies evaluated the same device in the same population (balloon catheters for coronary angiography 15,17 ) while the rest of the included studies evaluated a variety of devices across different populations.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingcitationAmong original probes in Zacharias et al. 20, 373 (99.47%) were functioning and 2 (0.53%) were nonfunctioning.
“Among the original probes, 373 (99.47%) were functioning and 2 (0.53%) were nonfunctioning.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [20]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Classified as probably not a substantial claim.
FindingreasoningOutcomes were reported in which the impact on patient outcomes was not made clear.
“In addition, outcomes were reported in which the impact on patient outcomes was not made clear.”
Quote in the paperSubstantial claimMay combine two claims.
FindingcitationMarins et al. 22 reported hospital mortality in the reprocessing group was 1.90% (6 deaths), and 2.88% (12 deaths) in the nonreprocessing group.
“Hospital mortality in the reprocessing group was 1.90%, resulting in 6 deaths, and it was 2.88% (12 deaths) in the nonreprocessing group.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
FindingcitationMarins et al. 22 point out that stringent criteria allow the detection of any type of anomaly, which is not the case in original medical devices whose quality control is done by sampling.
“They point out that these stringent criteria allow the detection of any type of anomaly, which is not the case in original medical devices whose quality control is done by sampling.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Findingown dataThree studies (2, 15, 19) used third-party reprocessing or remanufacturing of SUMDs, while the rest described in-hospital reprocessing protocols that varied by institution.
“Three studies 2,15,19 used third-party reprocessing or remanufacturing of SUMDs, while the rest described in-hospital reprocessing protocols that varied by institution.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [2]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [19]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Findingnone givenStudies evaluating semicritical devices were not identified, and all included studies evaluated critical category devices in surgical settings.
“Studies evaluating semicritical devices were not identified, and all included studies evaluated critical category devices in surgical settings.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationBrowne et al. 15 reported mean dye volume was 275 mL compared with 307 mL in the comparison group.
“The mean dye volume was 275 mL compared with 307 mL in the comparison group.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Classified as probably not a substantial claim.
FindingcitationBrowne et al. 15 reported that using simple methods of cleaning and decontamination followed by radiation sterilization does not offer enough of a guarantee of safety to permit reuse.
“Browne et al. 15 also reported that using simple methods of cleaning and decontamination followed by radiation sterilization does not offer enough of a guarantee of safety to permit reuse.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingcitationNo patient in the control group had chills, while 1 person in the reprocessing group had chills.
“No patient in the control group had chills, while 1 person in the reprocessing group had chills.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationMarins et al. 22 emphasize that a certified reprocessing process requires rigorous inspection, with demanding criteria and guidelines specific to each device.
“Marins et al 22 emphasize that a certified reprocessing process requires rigorous inspection, with demanding criteria and guidelines specific to each device.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingreasoningAlthough the Marins et al. study scored well in external validity and generalizability and in many aspects related to confounding and bias, it had flaws typical of observational studies, including lack of randomization of patients to each intervention arm with assignment concealment, and no blinding of patients and outcome assessors described.
“Although the study scored well in terms of external validity and therefore generalizability, and also in many aspects related to confounding and bias, it had a few flaws typical of observational studies, including lack of randomization of patients to each intervention arm with assignment concealment, and there was no blinding of patients and outcome assessors described.”
Quote in the paperNot a substantial claimMay combine two claims.
FindingcitationPlante et al. 17 used an in-house reprocessing protocol, whereas Browne et al. 15 used a third-party reprocessor.
“However, Plante et al 17 used an in-house reprocessing protocol, whereas Browne et al 15 used a third-party reprocessor”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
FindingcitationIn Jokar et al. 21, individuals in the disposable port group had a maximum of 2 symptoms of superficial infection, but there were usually more than 2 symptoms in the reprocessed port group.
“Individuals in the disposable port group had a maximum of 2 symptoms of superficial infection, but there were usually more than 2 symptoms in the reprocessed port group.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [21]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims. The quote only partly states this claim.
Findingown dataIn Plante et al. (1994), the volume of contrast medium was 201 mL +/-86 in the reprocessed group versus 165 +/-61 mL in the new device group, P<0.025.
“Volume of contrast medium: 201 mL +/-86”
- Also stated in: § Additional information · plain
“165 +/-61 mL medium: Volume of contrast”
Quote in the paperNot a substantial claim- Also stated in: § Additional information · plain
May combine two claims.
FindingcitationIn Plante et al. (1994), the clinical failure rate with adverse clinical events was 7.8% (n=25) in the reprocessed group versus 3.8% (n=14) in the new device group, P<0.025.
“Clinical failure with adverse clinical events: n = 25 (7.8%)”
- Also stated in: § Additional information · plain
“Clinical failure with adverse clinical events: N = 14 (3.8%) P < 0.025”
Quote in the paperSubstantial claim- Also stated in: § Additional information · plain
May combine two claims. The quote only partly states this claim.
FindingcitationIn Plante et al. (1994), the initial balloon failure rate was 10.2% in the reprocessed group versus 3.3% in the new device group, P<0.0001.
“Initial balloon failure: 10.2%”
- Also stated in: § Additional information · plain
“Initial balloon failure: 3.3%; P < 0.0001”
Quote in the paperNot a substantial claim- Also stated in: § Additional information · plain
The quote only partly states this claim.
FindingcitationIn Plante et al. (1994), procedure times were 81 +/-41 minutes in the reprocessed group versus 68 +/-32 minutes in the new device group, P<0.0001.
“Procedure times: 81 +/-41 minutes”
- Also stated in: § Additional information · plain
“Procedure times: 68 +/-32 minutes, P < 0.0001”
Quote in the paperNot a substantial claim- Also stated in: § Additional information · plain
May combine two claims.
FindingcitationIn Plante et al. (1994), mean balloon catheter use per lesion was 2.4 +/-1.5 in the reprocessed group versus 1.2 +/-0.5 in the new device group, P<0.00001.
“Mean 2.4 +/-1.5 catheters”
- Also stated in: § Additional information · plain
“Mean 1.2 +/-0.5 catheters”
Quote in the paperSubstantial claim- Also stated in: § Additional information · plain
May combine two claims.
FindingcitationIn Plante et al. (1994), no statistically significant differences in clinical success and clinical failure without adverse events were reported between the 2 groups.
“No statistically significant differences in clinical success and clinical failure without adverse events were reported between the 2 groups.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningIn Zacharias et al., because of flaws in study design, there is considerable likelihood of bias, confounding, and lack of generalizability, limiting confidence in the reported results.
“Because of the flaws in the study design there is considerable likelihood of bias, confounding, and lack of generalizability, limiting confidence in the reported results.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Interpretationnone givenLeung et al. conducted a well-reported prospective cohort study with historical controls.
“Leung et al. 2 conducted a well-reported prospective cohort study with historical controls.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [2]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningIn Browne et al., it is not described whether participants were consecutive or whether potential participants might have been excluded, limiting generalizability and validity.
“The included participants were selected from the same hospital, but it is not described whether they are consecutive, or whether any potential participants might have been excluded, limiting the generalizability and validity of the study results.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningIn Jokar et al., adverse events were largely determined through patient self-report, which may not be as objective as if assessed by a health care professional using a validated outcome measure.
“Though adverse events were reported, they were largely determined through patient self-report (which may not be as objective as if assessed by a health care professional using a validated outcome measure).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataIncluded publications were critically appraised by 1 reviewer using the Downs and Black checklist for randomized and nonrandomized studies as a guide.
“The included publications were critically appraised by 1 reviewer using the Downs and Black checklist 13 for randomized and nonrandomized studies as a guide.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [13]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
InterpretationreasoningThe Browne et al. prospective cohort study with historical controls demonstrated serious reporting issues and external and internal validity issues.
“The Browne et al. 15 prospective cohort study with historical controls demonstrated serious reporting issues, external and internal validity issues.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
InterpretationcitationJokar et al. conducted a well-reported prospective cohort study evaluating the incidence of port-site infections in patients undergoing laparoscopic cholecystectomy.
“Jokar et al.21 conducted a well-reported prospective cohort study evaluating the incidence of PSIs in patients undergoing laparoscopic cholecystectomy.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [21]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Classified as probably not a substantial claim. May combine two claims.
InterpretationreasoningIt is unclear whether patient risk levels would be different for semicritical devices or in nonsurgical settings.
“All included studies evaluated SUMDs classified as critical, and all were conducted in surgical settings; however, it is unclear whether patient risk levels would be different for semicritical devices or in nonsurgical settings.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationcitationMarins et al. conducted a well-reported prospective cohort study evaluating surgical scissors and suture machines.
“Marins et al. 22 conducted a well-reported prospective cohort study evaluating surgical scissors and suture machines.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
PremisecitationHealth Canada regulates third-party device reprocessors and requires they meet the same requirements as new device manufacturers.
“In 2016, Health Canada updated its policy on the purchase and use of reprocessed SUMDs, concluding that it has authority to require that commercially reprocessed devices meet appropriate standards for safety, effectiveness, and labelling, with third-party reprocessing companies held to the same requirements as manufacturers of new devices.6”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [6]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Classified as probably not a substantial claim. May combine two claims.
InterpretationreasoningOverall, the Plante et al. study was well-reported but demonstrates issues with external and internal validity, particularly confounding.
“Overall, the Plante et al. 17 study was well-reported, but demonstrates issues with external validity and internal validity, particularly confounding.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
InterpretationreasoningThe Zacharias et al. study demonstrated issues with reporting, external validity, and internal validity.
“The study demonstrated issues with reporting, external validity, and internal validity.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenSummary scores were not calculated for included studies; strengths and limitations of each included publication were described narratively.
“Summary scores were not calculated for the included studies; rather, the strengths and limitations of each included publication were described narratively.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningIn Jokar et al., quasi-randomized allocation to treatment groups (odd/even patient ID) may have limited bias to some extent.
“They did report quasi-randomized allocation to treatment groups (if the last digit of the ID patient was odd, the patient was selected for the disposable port group, and if it was even, the patient was selected for the reprocessed port group), which may have limited bias to some extent.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningIn Leung et al., because historical controls were used, treatment groups were not recruited over the same period, so differences between patients, providers, or other factors may have impacted observed outcomes.
“Also, as the study used historical controls, the treatment groups were not recruited over the same period, which means there could be differences between patients, providers, or other factors that may have impacted the observed outcomes.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThere are potential economic and environmental benefits of using reprocessed single-use medical devices.
“Given the potential economic and environmental benefits of using reprocessed SUMDs, [9][10][11][12] yet limited guidance for hospitals to pursue onsite SUMD reprocessing, there is a growing interest in determining the clinical safety of reprocessed SUMDs.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [9]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [10]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [11]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationA 2019 position statement for community health care settings published by Infection Prevention and Control Canada states that critical and semicritical medical equipment/devices labelled as single-use must not be reprocessed and reused unless a licensed reprocessor does the reprocessing.
“Therefore, a 2019 position statement for community health care settings published by Infection Prevention and Control Canada, states that critical and semicritical medical equipment/devices labelled as single-use must not be reprocessed and reused unless a licensed reprocessor does the reprocessing.23”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [23]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe reuse of SUMDs also has important potential economic and environmental benefits, which have been evaluated elsewhere.
“The reuse of SUMDs also has important potential economic and environmental benefits, which have been evaluated elsewhere.[9][10][11][12]”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [9]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [10]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [11]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe subjects in Browne et al. (1997) appeared representative of the population from which they were recruited.
“The subjects asked to participate in the study appeared to be representative of the entire population from which they were recruited”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataOf the 8 included studies, 4 [15][16][17][18] were published before 2005, which potentially limits the applicability of at least half of the studies reported.
“In fact, of the 8 included studies,4 [15][16][17][18] were published before the year 2005, which potentially limits the applicability of at least half of the studies reported.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
Study design and conductnone givenThe objective of the study by Browne et al. (1997) is clearly described.
“The objective of the study is clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenBrowne et al. (1997) was powered to detect balloon failure rates.
“The study was powered to detect balloon failure rates”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoningNo power calculation was described in Colak et al. 16.
“No power calculation was described”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Study design and conductreasoningThe Gundogdu et al. 18 study demonstrated serious issues with reporting, external validity, and internal validity (bias and confounding).
“The study demonstrated serious issues with reporting and external validity, and internal validity, with regards to bias and confounding.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [18]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationEstimates of random variability for the main outcomes and important adverse events are provided in Browne et al. (1997).
“Estimates of the random variability in the data for the main outcomes are provided, and important adverse events have been reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenLoss to follow-up was not reported in Browne et al. (1997).
“Loss to follow-up was not reported”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationnone givenThe study conducted in Canada was published in 1994, and it is unknown if the same surgical methods, device specifications, patient care protocols, and reprocessing standards would be relevant today.
“In addition, the study conducted in Canada was published in 1994, and it is unknown if the same surgical methods, device specifications, patient care protocols, and reprocessing standards would be relevant today.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductcitationIn Colak et al. 16, it is not known if blinding of outcome assessors occurred, though patients and surgeons were blinded.
“it is not known if blinding of outcome assessors occurred, though patients and surgeons were blinded.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoningThe characteristics of the patient group in Browne et al. (1997) are not clearly described.
“The characteristics of the patient group are not clearly described.While we know the source, no inclusion/exclusion criteria are reported.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataThere is just 1 study conducted in Canada (Plante et al. 17), and results of studies conducted in other countries may not be generalizable to Canadian populations and settings.
“There is just 1 study conducted in Canada, 17, and the results of the studies conducted in the other countries may not be generalizable to Canadian populations and settings.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Study design and conductreasoningThe Colak et al. 16 RCT was well-reported and scored very well on external validity and internal validity (bias and confounding).
“The RCT was well-reported, and scored very well with regard to external validity, and internal validity, both bias and confounding”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Interpretationown dataWhile 8 studies met the inclusion criteria, they mostly looked at different devices for different interventions, which limits the amount of evidence available to evaluate patient outcomes for a specific device/intervention.
“While 8 studies met the inclusion criteria in this Rapid Review, they mostly looked at different devices for different interventions, which limits the amount of evidence available to evaluate the patient outcomes for a specific device/intervention.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenThe staff, places, and facilities in Browne et al. (1997) appeared representative of the treatment the majority of patients receive.
“It appears that the staff, places, and facilities where the patients were treated were representative of the treatment the majority of patients receive”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationThe Canadian Standards Association has published a 2023 standard regarding Canadian medical device reprocessing for all health care settings (CSA Z314:23), superseding an earlier version published in 2018, but excludes reprocessing of SUMDs.
“The Canadian Standards Association has published a 2023 standard regarding Canadian medical device reprocessing for all health care settings (CSA Z314:23), superseding an earlier version published in 2018, but excludes reprocessing of SUMDs.24”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [24]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationP values were not reported for any outcomes in Browne et al. 15.
“P values were not reported for any outcomes.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Classified as probably not a substantial claim. May combine two claims.
InterpretationreasoningReported outcomes such as procedure duration, fluoroscopy duration, and device failure would likely have some impact on patient outcomes, but it was not always clear what these impacts might be.
“For example, reported outcomes such as procedure duration, fluoroscopy duration, and device failure would likely have some impact on patient outcomes, but it was not always clear what these impacts might be.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationnone givenThe majority of studies employed an in-house reprocessing protocol, and not the presumably more rigorous third-party reprocessing protocol, which has the potential to impact study results.
“Finally, the majority of the studies employed an in-house reprocessing protocol, and not the presumably more rigorous third-party reprocessing protocol, which has the potential to impact study results.”
Quote in the paperNot a substantial claim- Rests on an assumption: Third-party reprocessing protocols are more rigorous than in-house reprocessing protocols.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe intervention and comparator in Browne et al. (1997) are described.
“The intervention and comparator are described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenFollow-up was the same for all patients in Browne et al. (1997).
“Follow-up was the same for all patients”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationIn 2018, the FDA reviewed and officially reported on the remanufacturing industry over the previous 2 decades and found no evidence that using remanufactured devices increased risk to health.
“In 2018, the FDA (FDA) reviewed and officially reported on the remanufacturing industry over the previous 2 decades and found no evidence that using remanufactured devices increased risk to health. 2”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [2]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenP values were not reported in Browne et al. (1997).
“P values were not reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenOutcomes in Browne et al. (1997) are not presented clearly, and many outcomes were not detailed in the introduction or methods.
“Outcomes are not presented clearly, and many outcomes are presented which were not detailed in the introduction or methods”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe Medicines and Health care Products Regulatory Agency also permits remanufacturing single-use devices within strict guidelines.
“The Medicines and Health care Products Regulatory Agency also permits remanufacturing single-use devices within strict guidelines.25”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [25]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenOnly some of the outcomes are described in the introduction/methods of Browne et al. (1997).
“Only some of the outcomes are described in the introduction/methods”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductreasoningThere is no description of the characteristics of the retrospective control group in Browne et al. (1997), so the distribution of potential confounders is unknown.
“There is no description of the characteristics of the retrospective control group, though it is described as being case-matched, and therefore the distribution of potential confounders is unknown”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenThe proportion of study participants that agreed to participate is not stated in Browne et al. (1997), nor is it known for the comparator group.
“The proportion of study participants that agreed to participate is not stated, nor is this known for the comparator group”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe outcome measures in Browne et al. (1997) appeared valid and reliable.
“Outcome measures appeared to be valid and reliable”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningIt is likely that device specifications, reprocessing protocols, and surgical approaches and patient care protocols have changed since some of these studies were published.
“It is likely that device specifications, reprocessing protocols, and surgical approaches and patient care protocols have changed since some of these studies were published.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataMost of the published evidence was of very low to moderate quality, published in countries outside Canada, with at least half published before 2005, which raises concerns regarding generalizability and applicability.
“In addition, most of the published evidence was of very low to moderate quality, and published in countries outside of Canada, with at least half of the evidence published before 2005, which raises concerns regarding generalizability and applicability.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningThe Colak et al. 16 RCT was published in 2004, conducted in Turkey, and evaluates SUMDs reprocessed using an in-hospital protocol, which could limit its generalizability to the Canadian population and Canadian health care settings, as well as settings where third-party reprocessing is utilized.
“However, the RCT was published in 2004, was conducted in Turkey, and evaluates SUMDs that were reprocessed using an in-hospital protocol, which could limit its generalizability to the Canadian population and Canadian health care settings, as well as setting where third-party reprocessing is utilized.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationIn Colak et al. 16, the proportion of patients that agreed to participate was not stated.
“The only issues noted were that the proportion of patients that agreed to participate in the study was not stated”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [16]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Study design and conductnone givenThe statistical tests in Browne et al. (1997) seemed appropriate.
“The statistical tests seemed appropriate”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenStudy objectives of Marins et al. (2018) were clearly described.
“Study objectives were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenIt is not stated whether any adjustment was made for confounding in Zacharias et al. (2021), and potential confounders are not described.
“It is not stated whether any adjustment was made for confounding, and potential confounders at not described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenFollow-up was the same between groups in Colak et al. (2004).
“Follow-up was the same between groups”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenIt is not known if blinding of outcome assessors occurred in Colak et al. (2004).
“It is not known if blinding of outcome assessors occurred”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S1 (asserted), S3 (load-bearing), S4 (based).
Study design and conductnone givenCharacteristics of the patients included in Colak et al. (2004) were clearly described.
“Characteristics of the patients included in the study were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenNo power calculation was described in Plante et al. (1994).
“No power calculation was described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThere were confounders between treatment groups in Plante et al. (1994) that may not have been considered in the analysis, such as higher rates of unstable angina in the reprocessed group at baseline.
“There were confounders between treatment groups that may not have been considered in the analysis (i.e., higher rates of unstable angina in the reprocessed group at baseline)”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenStudy objectives of Jokar et al. (2022) were clearly described.
“Study objectives were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenNo unplanned analyses occurred in Jokar et al. (2022).
“No unplanned analyses occurred”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenBoth groups in Jokar et al. (2022) were recruited over the same period of time.
“Both groups were recruited over the same period of time”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenLoss to follow-up was not reported in Leung et al. (2019).
“Loss to follow-up was not reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenNo unplanned analyses occurred in Colak et al. (2004).
“No unplanned analyses occurred”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenAll important adverse events appear to have been reported in Plante et al. (1994).
“It appears that all-important adverse events have been reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenBoth groups in Leung et al. (2019) were not recruited over the same period of time.
“Both groups were not recruited over the same period of time”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenMain outcome measures used in Colak et al. (2004) were valid and appropriate.
“Main outcome measures used were valid and appropriate”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoningThere does not appear to be any adjustment for confounding in Browne et al. (1997) since potential confounders are not described.
“There does not appear to be any adjustment for confounding since potential confounders are not described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenBlinding of outcome assessors was not described in Leung et al. (2019).
“Blinding of outcome assessors was not described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenThe statistical tests used to assess the main outcomes in Jokar et al. (2022) appeared appropriate.
“The statistical tests used to assess the main outcomes appeared to be appropriate”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe interventions of interest in Jokar et al. (2022) were clearly described.
“The interventions of interest were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenFollow-up was the same between groups in Leung et al. (2019).
“Follow-up was the same between groups”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenThe interventions of interest in Colak et al. (2004) were clearly described.
“The interventions of interest were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe distributions of principal confounders in each group of Jokar et al. (2022) were clearly described.
“The distributions of principal confounders in each group of subjects to be compared were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenOnly 3 patients appeared to be lost to follow-up in Jokar et al. (2022).
“Only 3 patients appeared to be lost to follow-up”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Study design and conductnone givenThe main findings of Colak et al. (2004) were clearly described.
“The main findings of the study were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoningThe subjects in Jokar et al. (2022) seemed representative of the population from which they were recruited, given any patient requiring LC was eligible.
“The subjects asked to participate in the study seemed to be representative of the entire population from which they were recruited, given any patient requiring LC was eligible (excepting detailed inclusion and exclusion criteria)”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe outcome measures in Jokar et al. (2022) were clearly described.
“The outcome measures were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenBlinding of patients to treatment allocation occurred in Colak et al. (2004).
“Blinding of patients to treatment allocation occurred”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenStudy objectives of Leung et al. (2019) were clearly described.
“Study objectives were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataNo power calculation was described in Zacharias et al. (2021).
“No power calculation was described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenAll important adverse events that may be a consequence of the intervention were reported in Colak et al. (2004).
“All important adverse events that may be a consequence of the intervention were reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenOutcomes of Leung et al. (2019) were clearly described.
“Outcomes were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenIt is not known if patients and outcome assessors were blinded to treatment assignment in Browne et al. (1997).
“It is not known if patients and outcome assessors were blinded to treatment assignment”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenStatistical analysis in Leung et al. (2019) was appropriate.
“Statistical analysis was appropriate”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenA very simple power calculation was performed in Jokar et al. (2022).
“A very simple power calculation was performed”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe proportion of those asked who agreed to participate in Leung et al. (2019) was not stated.
“The proportion of those asked who agreed to participate in the study was not stated.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenActual P values were reported in Plante et al. (1994).
“Actual P values were reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenThe staff, places, and facilities in Colak et al. (2004) seemed representative of the treatment the majority of similar patients would receive.
“The staff, places, and facilities where the patients were treated seemed to be representative of the treatment the majority of similar patients would receive”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
Study design and conductnone givenThe main findings of Jokar et al. (2022) were clearly described.
“The main findings of the study were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenNo power calculation was described in Colak et al. (2004).
“No power calculation was described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Study design and conductnone givenThe statistical analyses in Plante et al. (1994) seemed appropriate.
“Statistical analyses seemed to be appropriate”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductown dataAdverse events in Jokar et al. (2022) were largely determined through patient self-report, and follow-up may not have been sufficiently long-term to assess all adverse events.
“Though adverse events were reported, they were largely determined through patient self-report which is not ideal, and follow-up may not have been sufficiently long-term to assess all adverse events”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductnone givenThe randomized intervention assignment in Jokar et al. (2022) was not described as being concealed from patients and would not likely be concealed for health care staff.
“The randomized intervention assignment was not described as being concealed from patients and would not likely be concealed for health care staff”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenNo power calculation was described in Leung et al. (2019).
“No power calculation was described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S1 (asserted), S3 (load-bearing), S4 (based).
Study design and conductnone givenThe distributions of principal confounders in each group of Colak et al. (2004) were clearly described.
“The distributions of principal confounders in each group of subjects to be compared were clearly described”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductnone givenPatients in Leung et al. (2019) were not randomized to treatment groups and allocation concealment was not implemented.
“Patients were not randomized to treatment groups and allocation concealment was not implemented”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenBoth groups in Marins et al. (2018) were recruited over the same period of time.
“Both groups were recruited over the same period of time”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Study design and conductnone givenThe staff, places, and facilities in Plante et al. (1994) seemed representative of the treatment the majority of patients receive.
“It appears that the staff, places, and facilities where the patients were treated were representative of the treatment the majority of patients receive”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S1 (asserted), S3 (load-bearing), S4 (based).
Study design and conductnone givenThe distributions of principal confounders in each group of Leung et al. (2019) were clearly described.
“The distributions of principal confounders in each group of subjects to be compared were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenJokar et al. (2022) does not provide estimates of the random variability in the data for the outcome.
“The study does not provide estimates of the random variability in the data for the outcome”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataQuasi-randomization was performed to assign patients to each group in Jokar et al. (2022).
“Quasi-randomization was performed to assign patients to each group”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenThe objectives, main outcomes, interventions, and characteristics of included patients of Plante et al. (1994) were well-described.
“The objectives of the study, the main outcomes, description of the interventions, and the characteristics of the included patients were well-described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenActual P values and estimates of random variability were reported in Leung et al. (2019).
“Actual P values were reported and estimates of random variability were reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThere were no confounding factors between treatment groups at baseline in Leung et al. (2019).
“There were no confounding factors between treatment groups at baseline”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenBoth groups in Jokar et al. (2022) were recruited from the same population.
“Both groups were recruited from the same population”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataNo power calculation was reported in Marins et al. (2018).
“No power calculation was reported.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenLength of follow-up was the same in both groups in Jokar et al. (2022).
“Length of follow-up was the same in both groups”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenPatients in the different treatment groups in Plante et al. (1994) were recruited over the same period of time.
“Patients in the different treatment groups were recruited over the same period of time”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenOutcomes of Colak et al. (2004) were clearly described.
“Outcomes were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Study design and conductnone givenThe main findings of Leung et al. (2019) were clearly described.
“The main findings of the study were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenThe proportion lost to follow-up in Colak et al. (2004) was too small to affect the main findings.
“The proportion lost to follow-up was too small to affect the main findings”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenBoth groups in Marins et al. (2018) were recruited from the same population.
“Both groups were recruited from the same population”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenCharacteristics of the patients included in Jokar et al. (2022) were clearly described.
“Characteristics of the patients included in the study were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenMain outcome measures used in Leung et al. (2019) were valid and appropriate.
“Main outcome measures used were valid and appropriate”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenNo adjustment appeared to be made for different lengths of follow-up in Marins et al. (2018).
“No adjustment appeared to be made for different lengths of follow-up”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenStatistical analysis in Colak et al. (2004) was appropriate.
“Statistical analysis was appropriate”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenAll important adverse events that may be a consequence of the intervention were reported in Leung et al. (2019).
“All important adverse events that may be a consequence of the intervention were reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenBoth groups in Colak et al. (2004) were recruited over the same period of time.
“Both groups were recruited over the same period of time”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenIt is unknown if loss to follow-up occurred and was addressed in Zacharias et al. (2021).
“It is unknown if loss to follow-up occurred and was addressed if so”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenActual P values were reported in Jokar et al. (2022).
“Actual P values were reported”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductnone givenPatients in Plante et al. (1994) were not randomized to intervention groups and randomization concealment did not occur.
“Patients were not randomized to intervention groups and randomization concealment did not occur”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataPatients in Colak et al. (2004) were randomized to treatment groups and allocation concealment was implemented.
“Patients were randomized to treatment groups and allocation concealment was implemented (only the nurse knew the allocation)”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe proportion of those asked who agreed to participate was not provided in Plante et al. (1994).
“The proportion of those asked who agreed to participate was not provided”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenActual P values and estimates of random variability were reported in Colak et al. (2004).
“Actual P values were reported and estimates of random variability were reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Study design and conductnone givenThe interventions of interest in Zacharias et al. (2021) are clearly described.
“The interventions of interest are clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenBoth groups in Leung et al. (2019) were recruited from the same population.
“Both groups were recruited from the same population”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenBoth groups in Colak et al. (2004) were recruited from the same population.
“Both groups were recruited from the same population”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe interventions of interest in Leung et al. (2019) were clearly described.
“The interventions of interest were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenRandomization and allocation concealment was not reported in Marins et al. (2018).
“Randomization and allocation concealment was not reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe study objectives of Zacharias et al. (2021) were clearly described.
“The study objectives were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Study design and conductnone givenThe main outcome measures in Plante et al. (1994) seemed reliable.
“The main outcomes measures used seemed to be reliable”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenLoss to follow-up was not reported in Plante et al. (1994).
“Loss to follow-up was not reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenThe staff, places, and facilities in Jokar et al. (2022) seemed representative of the treatment the majority of similar patients would receive.
“The staff, places, and facilities where the patients were treated seemed to be representative of the treatment the majority of similar patients would receive”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenNo unplanned analyses occurred in Leung et al. (2019).
“No unplanned analyses occurred”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe subjects in Leung et al. (2019) seemed representative of the population from which they were recruited.
“The subjects asked to participate in the study seemed to be representative of the entire population from which they were recruited”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenIt is not stated whether any adjustment was made for confounding in Marins et al. (2018).
“It is not stated whether any adjustment was made for confounding”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe distributions of principal confounders in each group of Plante et al. (1994) were clearly described.
“The distributions of principal confounders in each group of subjects to be compared clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe patients in each group in Browne et al. (1997) were recruited over different time periods.
“The patients in each group were recruited over different time periods”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe subjects in Plante et al. (1994) seemed representative of the population from which they were recruited.
“The subjects asked to participate in the study seemed to be representative of the entire population from which they were recruited”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenThe staff, places, and facilities in Leung et al. (2019) seemed representative of the treatment the majority of similar patients would receive.
“The staff, places, and facilities where the patients were treated seemed to be representative of the treatment the majority of similar patients would receive”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenNo unplanned analyses occurred and follow-up appeared the same between both groups in Plante et al. (1994).
“No unplanned analyses occurred and follow-up appeared to be the same between both groups”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe proportion of patients that agreed to participate was not stated in Colak et al. (2004).
“The proportion of patients that agreed to participate was not stated”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenStudy objectives of Colak et al. (2004) were clearly described.
“Study objectives were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenIt is unknown whether the outcome assessor was blinded to the intervention in Marins et al. (2018).
“It is unknown whether the outcome assessor was blinded to the intervention”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenUnplanned outcomes and analyses were reported in Browne et al. (1997).
“Unplanned outcomes and analyses were reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenBlinding of patients to treatment allocation did not occur in Leung et al. (2019).
“Blinding of patients to treatment allocation and did not occur”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductreasoningThe patients in Browne et al. (1997) were not randomized and no randomization concealment occurred.
“The patients were not randomized and therefore no randomization concealment”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe subjects in Colak et al. (2004) seemed representative of the population from which they were recruited.
“The subjects asked to participate in the study seemed to be representative of the entire population from which they were recruited”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe very limited length of follow-up in Plante et al. (1994) may not be adequate to assess outcomes.
“Very limited length of follow-up may not be adequate to assess outcomes”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductown dataThe percentage of subjects asked to participate in Jokar et al. (2022) was not reported, so it is not known if participants are representative of the entire population from which they were recruited.
“The percentage of subjects asked to participate in the study was not reported, and therefore it is not known if those that participated are representative of the entire population from which they were recruited”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenCharacteristics of the patients included in Leung et al. (2019) were clearly described.
“Characteristics of the patients included in the study were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
Study design and conductnone givenThe main findings of Plante et al. (1994) are clearly described.
“The main findings are clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenLoss to follow-up was reported in Colak et al. (2004).
“Loss to follow-up was reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenOutcomes of Jokar et al. (2022) were clearly described.
“Outcomes were clearly described”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe proportion of those asked who agreed to participate was not stated in Marins et al. (2018).
“The proportion of those asked who agreed to participate was not stated”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe main outcomes of Zacharias et al. (2021) were described in the introduction.
“The main outcomes were described in the introduction”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenLoss to follow-up was not reported in Marins et al. (2018).
“Loss to follow-up was not reported”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataNo attempt was described to blind study subjects or outcome assessors in Jokar et al. (2022).
“No attempt was described to blind study subjects or outcome assessors to the intervention they received”
Quote in the paperNot a substantial claim
The paper relies on this without support.
Manufacturers of single-use medical devices are not obliged to provide instructions for proper cleaning and sterilization, unlike manufacturers of reusable devices.
“Manufacturers of SUMDs are not obliged to provide instructions for proper cleaning and sterilization, as they are for reusable devices.”
The paper relies on this without support.
Health Canada does not provide oversight for hospital onsite reprocessing or in community health care settings, deferring to provincial and territorial oversight.
“However, Health Canada does not provide oversight for hospital onsite reprocessing or in community health care settings, deferring to the oversight provided at the provincial and territorial level.”
The paper relies on this without support.
Current standards for reprocessing medical devices use definitions for sterilization and disinfection based on measurement of bioburden, but not necessarily clinical outcomes such as infection.
“Current standards for reprocessing medical devices use definitions for sterilization and disinfection based on measurement of bioburden, but not necessarily clinical outcomes such as infection.”
The authors state this as an assumption.
Reprocessing protocols vary by third-party processor.
“Protocols varied by institution, by device, and it is assumed, also by third-party processor.”
The authors state this as an assumption.
There would not be greater risks to patient safety when considering use of reprocessed semicritical SUMDs than reprocessed critical SUMDs.
“However, it may be reasonable to assume that there would not be greater risks to patient safety when considering use of reprocessed semicritical SUMDs than reprocessed critical SUMDs.”
The authors state this as an assumption.
All adverse events have been reported in Gundogdu et al. (1998) 18.
“It is assumed that all adverse events have been reported”
The paper relies on this without support.
Third-party reprocessing protocols are more rigorous than in-house reprocessing protocols.
“not the presumably more rigorous third-party reprocessing protocol”
The authors state this as an assumption.
All important adverse events that may be a consequence of the intervention were reported in Jokar et al. (2022).
“It is assumed that all-important adverse events that may be a consequence of the intervention were reported”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Marins publication year differs from the reference
MinorThe study is dated 2018 in the appraisal, but its cited reference lists publication in 2019.
“Marins et al. (2018) 22”
“22. de Sousa Marins B, Queiroz e Melo J, Logarinho Monteiro JL, Rente G, Teixeira Bastos P. Reprocessing of single-use medical devices: Clinical and financial results. Port J Public Health. 2019;36(3):150-156.”
Table 4 reprocessed-group rehospitalization percentage does not match its count
MinorTable 4 places 33 rehospitalizations in the reprocessed group, which had 316 patients. That is 10.44%, not the reported 7.91%; 7.91% instead corresponds to 33 of the 417 nonreprocessed patients.
Reported 7.91 · recomputed 10.443038
“Rehospitalization n = 33 (7.91%)Mortality n = 6 (1.90%)”
Browne study has conflicting design labels
MinorThe text describes Browne et al. as a prospective cohort with historical controls, whereas Table 4 labels the same study a retrospective case-control study.
“The Browne et al. 15 prospective cohort study with historical controls demonstrated serious reporting issues, external and internal validity issues.”
“| Browne et al. | PTCA Balloon | Fever: n = 11 | Fever: n = 12 | P values were not reported | | (1997) 15 | Catheter | Elevated WBC: n = 12 | Elevated WBC: n = 14 | Angiographic failure rate | | Retrospective case-control”
Rehospitalization counts are assigned to opposite groups
MinorThe narrative reports 28 rehospitalizations in the reprocessing group and 33 in the nonreprocessed group, but Table 4 assigns 33 to the reprocessed group and 28 to the new-device group.
“Rehospitalization was required for 28 patients in the reprocessing group and for 33 patients in the nonreprocessed group (P = 0.678).”
“Rehospitalization n = 33 (7.91%)Mortality n = 6 (1.90%) Reoperations n = 0 Postoperative Infection Rate 13.4% P = 0.664 Consumption of antibiotics (defined daily dose) 5.61 ± 15.21 P = 0.319 Length of Hospital Stay 10.36 ± 12.00 days P = 0.881 Duration of Surgery 2.45 ± 1.52 hours P = 0.161 Rehospitalization n = 28 (8.86%)P = 0.678”
Table 4 new-device rehospitalization percentage does not match its count
MinorTable 4 places 28 rehospitalizations in the nonreprocessed group, which had 417 patients. That is 6.71%, not the reported 8.86%; 8.86% instead corresponds to 28 of the 316 reprocessed patients.
Reported 8.86 · recomputed 6.714628
“Rehospitalization n = 28 (8.86%)P = 0.678”
Cites
Liberati, A., Altman, D. G., Tetzlaff, J., Mulrow, C., Gøtzsche, P. C., Ioannidis, J. P. A., Clarke, M., Devereaux, P. J., Kleijnen, J., & Moher, D. (2009) · Annals of Internal Medicine · doi:10.7326/0003-4819-151-4-200908180-00136 · Cited by 10,349 · see citing works
Downs, S. H., & Black, N. (1998) · Journal of Epidemiology and Community Health · doi:10.1136/jech.52.6.377 · Cited by 8,252 · see citing works
Liberati, A., Altman, D. G., Tetzlaff, J., Mulrow, C., Gøtzsche, P. C., Ioannidis, J. P. A., Clarke, M., Devereaux, P. J., Kleijnen, J., & Moher, D. (2009) · Italian Journal of Public Health · doi:10.2427/5768 · Cited by 3,374 · see citing works
Keil, M., Viere, T., Helms, K., & Rogowski, W. H. (2022) · European Journal of Public Health · doi:10.1093/eurpub/ckac174 · Cited by 139 · see citing works
Bolten, A., Kringos, D. S., Spijkerman, I. J. B., & Weiland, N. H. S. (2022) · Journal of Hospital Infection · doi:10.1016/j.jhin.2022.07.011 · Cited by 62 · see citing works
Reuse of balloon catheters for coronary angioplasty: A potential cost-saving strategy?
Planté, S., Strauss, B. H., Goulet, G., Watson, R. K., & Chisholm, R. J. (1994) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(94)90142-2 · Cited by 55 · see citing works
Hailey, D. M., Jacobs, P., Ries, N. M., & Polisena, J. (2008) · International Journal of Technology Assessment in Health Care · doi:10.1017/s0266462308080562 · Cited by 47 · see citing works
Economic Analysis of Reprocessing Single-Use Medical Devices: A Systematic Literature Review
Jacobs, P., Polisena, J., Hailey, D. M., & Lafferty, S. (2008) · Infection Control and Hospital Epidemiology · doi:10.1086/529587 · Cited by 40 · see citing works
Initial Experience With Reuse of Coronary Angioplasty Catheters in the United States
Browne, K. F., Maldonado, R., Telatnik, M., Vlietstra, R. E., & Brenner, A. S. (1997) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(97)00362-8 · Cited by 37 · see citing works
Interventions for sustainable surgery: a systematic review
Lam, K., Gadi, N. R., Acharya, A., Beatty, J. W., Darzi, A. W., & Purkayastha, S. (2023) · International Journal of Surgery · doi:10.1097/js9.0000000000000359 · Cited by 33 · see citing works
Çolak, T., Ersöz, G., Akça, T., Kanık, A., & Aydın, S. (2004) · Surgical Endoscopy · doi:10.1007/s00464-004-8112-x · Cited by 28 · see citing works
Remanufactured circular mapping catheters: safety, effectiveness and cost
Leung, L. W. M., Evranos, B., Grimster, A., Li, A. C., Norman, M. W., Bajpai, A., Zuberi, Z., Sohal, M., & Gallagher, M. M. (2018) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-018-0497-x · Cited by 18 · see citing works
Reprocessing of Single-Use Medical Devices: Clinical and Financial Results
Martins, B. D. S., Melo, J., Monteiro, J. L., Rente, G., & Bastos, P. T. (2018) · Portuguese Journal of Public Health · doi:10.1159/000496299 · Cited by 14 · see citing works
Safety and cost-effectiveness of single-use endolaser probe reprocessing in vitreoretinal surgery
Zacharias, L. C., Conci, L. D. S., Megnis, B. P., Falabretti, J. G., Neto, T. D. S. R., Neto, E. D. D. S., Preti, R. C., Cunha, L. P., & Monteiro, M. L. R. (2021) · International Journal of Retina and Vitreous · doi:10.1186/s40942-021-00292-x · Cited by 7 · see citing works
Jokar, M., Larti, N., Zarei, M., Melali, H., Zabihirad, J., Maraki, F., Dorcheh, S. J., & Kelidari, B. (2022) · Surgical Laparoscopy Endoscopy & Percutaneous Techniques · doi:10.1097/sle.0000000000001104 · Cited by 5 · see citing works
Zayyan, K. S., & Sellu, D. P. (1999) · Journal of Laparoendoscopic & Advanced Surgical Techniques · doi:10.1089/lap.1999.9.117 · Cited by 3 · see citing works
Flagged first, then by citations
Research works printed but not linked: 2
- Best Practices for Cleaning, Disinfection and Sterilization of Medical Equipment/Devices In All Health Care Settings, 3rd edition Toronto (ON): Public Health Ontario; 2013: https:// www .publichealthontario .ca/ -/ media/ Documents/ B/ 2013/ bp -cleaning -disinfection -sterilization -hcs .pdf ?rev = 7a 36d8e52664 4eb794a9ef 066d97a120 & sc _lang = en. Accessed 2023 Sep 25.
- de Sousa Marins B, Queiroz e Melo J, Logarinho Monteiro JL, Rente G, Teixeira Bastos P. Reprocessing of single-use medical devices: Clinical and financial results. Port J Public Health. 2019;36(3):150-156.
Cited by
Sorted by citations
Patel, S. S., Bains, A., Ali, N., & Chawla, P. (2025) · Circular Economy and Sustainability · doi:10.1007/s43615-025-00611-6 · Cited by 2 · see citing works
Assessment of applications for reprocessing single-use devices: experiences from Taiwan
Lee, C., Saffari, M., Aljaberi, M. A. A., Shih, C., Potenza, M. N., & Lin, C. (2025) · Journal of Hospital Infection · doi:10.1016/j.jhin.2025.08.001 · Cited by 2 · see citing works
Tartari, E., Meşe, E. A., Bulabula, A. N. H., Dancer, S. J., Giacobbe, D. R., Krone, M., Kilpatrick, C., Marek, A., Petrosillo, N., Presterl, E., et al. (2026) · Clinical Microbiology and Infection · doi:10.1016/j.cmi.2026.04.017 · Cited by 2 · see citing works
Planetary Health in Otolaryngology: A Call to Depollute Care
Liu, C., MacNeill, A. J., Strychowsky, J. E., Chadha, N. K., & Westerberg, B. D. (2026) · Journal of Otolaryngology - Head and Neck Surgery · doi:10.1177/19160216261463216
The hidden reuse of single-use surgical consumables: An underrecognized patient safety concern
Bagheri, M. (2026) · Perioperative Care and Operating Room Management · doi:10.1016/j.pcorm.2026.100745
- Authors
- Kaulback, K., & Horton, J.
- ORCID
- Horton, Jennifer
- Venue
- Canadian Journal of Health Technologies
- Published
- 2024-03-11
- Volume
- 4
- Issue
- 3
- ISSN
- 2563-6596 (ISSN-L)
- OpenAlex
- W4392706467
- Cited by
- 5
- References
- 16
- Citations by year
- 2026: 3 · 2025: 2
- FWCI
- 1.30
- Type
- journal article
- Language
- en
- Open access
- diamond
- doi.org
- Locations
- Canadian Journal of Health Technologies (publishedVersion)
- Topics
- Life Sciences › Immunology and Microbiology › Microbiology › Medical Device Sterilization and Disinfection
- Health Sciences › Medicine › Infectious Diseases › Infection Control in Healthcare
- Health Sciences › Health Professions › Health Information Management › Healthcare Regulation
- Keywords
- single-use medical devices, medical device reprocessing, clinical safety, infection outcomes, patient mortality, adverse events, microbial load, surgical setting, medical device reuse, Canada, Ontario, Quebec, British Columbia, Alberta
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-11-06.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-11-06.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review