The Variability of Results Between Point-of-Care Testing Glucose Meters and the Central Laboratory Analyzer

Khan, Adil I., Vasquez, Yolanda, Gray, Jacquelyn, Wians Jr, Frank H., Kroll, Martin H. · 2006 · Archives of Pathology & Laboratory Medicine
Cited by 133

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

Context. —Point-of-care testing glucose meters are strongly recommended in the management of diabetes and are increasingly being used for making therapeutically important decisions. Thus, it is essential that their results correlate well with those of laboratory analyzers. Objectives. —To test the reliability of point-of-care testing glucose meters. Design. —Two studies were performed: (1), an in-house study comparing accuracy of point-of-care testing glucose meters with a reference analyzer using fresh whole blood specimens (2), a real-time comparison of (a) 2 successive glucose meter readings and (b) glucose meter reading to central laboratory analyzer reading. Setting. —(1), Seven glucose meters from 4 manufacturers were compared with the Yellow Springs YSI 2300 blood glucose analyzer using whole blood without preservative. (2), (a) Whole blood samples were read within 5 minutes of each other using Accu-Chek meters and (b) between a glucose meter and a Hitachi laboratory analyzer. Results. —(1) Within the Accu-Chek group of glucose meters, fresh, preservative-free whole blood samples showed the lowest bias. (2) At the hypoglycemic level, successive glucose meter readings agreed well, but there was considerable disagreement between glucose meter and central laboratory values. Because laboratory analyzers are of proven accuracy, they are used as the reference. In the glucose meter–central laboratory analyzer correlation, for both hypoglycemic and hyperglycemic values, readings in which the differences were greater than 10% occurred more than 61% of the time. In the hypoglycemic range, differences greater than 20% occurred 57% of the time. Conclusions. —One should scrutinize point-of-care testing glucose meter readings at the hypoglycemic and hyperglycemic levels and whenever possible to corroborate these clinical results with central laboratory analyzers.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.5858/2006-130-1527-tvorbp

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 133

Metadata

Authors
Khan, A. I., Vasquez, Y., Gray, J., Wians Jr, F. H., & Kroll, M. H.
Venue
Archives of Pathology & Laboratory Medicine
Published
2006-10
Volume
130
Issue
10
Pages
1527-1532
ISSN
  • 0003-9985 (ISSN-L)
  • 1543-2165
OpenAlex
W1959236711
Cited by
133
References
2
FWCI
6.86
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: The Variability of Results Between Point-of-Care Testing Glucose Meters and the Central Laboratory Analyzer. https://app.cochat.ai/papers/10.5858/2006-130-1527-tvorbp
The paper
Khan, A. I., Vasquez, Y., Gray, J., Wians Jr, F. H., & Kroll, M. H. (2006). The Variability of Results Between Point-of-Care Testing Glucose Meters and the Central Laboratory Analyzer. Archives of Pathology & Laboratory Medicine. https://doi.org/10.5858/2006-130-1527-tvorbp

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.