Effects of Different Hematocrit Levels on Glucose Measurements With Handheld Meters for Point-of-Care Testing
The DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
Objectives. —To determine the effects of low, normal, and high hematocrit levels on glucose meter measurements and to assess the clinical risks of hematocrit errors. Design. —Changes in glucose measurements between low and high hematocrit levels were calculated to determine hematocrit effects. The differences between glucose measured with meters and with a plasma glucose method (YSI 2300) also were compared. Setting. —Six handheld glucose meters were assessed in vitro at low (19.1%), normal (38.5%), and high (58.3%) hematocrit levels, and at 6 glucose concentrations ranging from 2.06 mmol/L (37.1 mg/dL) to 30.24 mmol/L (544.7 mg/dL). Results. —Most systems, regardless of the reference to which they were calibrated, demonstrated positive bias at lower hematocrit levels and negative bias at higher hematocrit levels. Low, normal, and high hematocrit levels progressively lowered Precision G and Precision QID glucose measurements. Hematocrit effects on the other systems were more dependent on the glucose concentration. Overall, Accu-Chek Comfort Curve showed the least sensitivity to hematocrit changes, except at the lowest glucose concentration. Conclusions. —We strongly recommend that clinical professionals choose glucose systems carefully and interpret glucose measurements with extreme caution when the patient's hematocrit value changes, particularly if there is a simultaneous change in glucose level.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“In strip-based glucose assays, high hematocrit values, due to increased blood viscosity, can reduce the diffusion of plasma and determine underestimated results, whereas low hematocrit levels, like in anemia, are associated with overestimated readings 101,102.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [102]
In the citing paper
“Glucose extraction is approximately in linear relation with the density and duration of iontophoretic current (<0.5 mA/cm2) 102.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [102]
In the citing paper
“Important variables that may influence the results of bedside glucose monitoring include changes in hematocrit (114), altitude, environmental temperature or humidity, hypotension, hypoxia and high triglyceride concentrations (115), and various drugs.”
From Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus (2011) · cites this paper as [114]
In the citing paper
“Quite often the relevance of factors such as HCT is not considered by the treating physician and, subsequently, by the PwD [28].”
From Self-Monitoring of Blood Glucose as an Integral Part in the Management of People with Type 2 Diabetes Mellitus (2022) · cites this paper as [28]
Citations
Cites
A Text-Book of Physical Chemistry
(1936) · Nature · doi:10.1038/138864b0 · Cited by 1,146
A textbook of physical chemistry
Adamson, A. W. (1973) · Cited by 343
Tang, Z., Du, X., Louie, R. F., & Kost, G. J. (2000) · American Journal of Clinical Pathology · doi:10.1309/qaw1-x5xw-bvrq-5lkq · Cited by 207
Self-citationFingerstick Glucose Determination in Shock
Atkin, S. H., Dasmahapatra, A., Jaker, M. A., Chorost, M. I., & Reddy, S. (1991) · Annals of Internal Medicine · doi:10.7326/0003-4819-114-12-1020 · Cited by 189
Louie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0257-pocgt · Cited by 180
Self-citationBurrin, J. M., & Price, C. P. (1985) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456328502200401 · Cited by 150
Kost, G. J., Vu, H.-T., Lee, J. H., Bourgeois, P., Kiechle, F. L., Martin, C., Miller, S. S., Okorodudu, A. O., Podczasy, J. J., Webster, R., et al. (1998) · Critical Care Medicine · doi:10.1097/00003246-199803000-00036 · Cited by 120
Self-citationCommercial biosensors : applications to clinical, bioprocess, and environmental samples
Ramsay, G. (1998) · Wiley eBooks · Cited by 109
Accuracy of fingerstick glucose values in shock patients
Sylvain, H., Pokorny, M., English, S., Benson, N., Whitley, T., Ferenczy, C., & Harrison, J. (1995) · American Journal of Critical Care · doi:10.4037/ajcc1995.4.1.44 · Cited by 89
Neonatal Hyperviscosity: I. Incidence
Wirth, F. H., Goldberg, K. E., & Lubchenco, L. O. (1979) · PEDIATRICS · doi:10.1542/peds.63.6.833 · Cited by 79
Research works printed but not linked: 4
- Hills, L., G.Azurin, and X.Wang. et al. Glutathione as an interferent in near patient whole blood glucose devices.In: Proceedings of the 17th International Symposium of the International Federation of Clinical Chemistry. Madison, Wis: Omnipress. 1998. 207–219.
- Ramsay, G. Commercial Biosensors: Applications to Clinical, Bioprocess, and Environmental Samples. New York, NY: John Wiley & Sons Inc. 1998. 23.
- Adamson, A. W. A Textbook of Physical Chemistry. New York, NY: Academic Press. 1973. 339–437.
- Hills, L., G.Azurin, and X.Wang. et al. Glutathione as an interferent in near patient whole blood glucose devices.In: Proceedings of the 17th International Symposium of the International Federation of Clinical Chemistry. Madison, Wis: Omnipress. 1998. 207–219.
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
Management of diabetes and hyperglycemia in hospitals.
Clement, S., Braithwaite, S. S., Magee, M. F., Ahmann, A., Smith, E. P., Schafer, R. G., Hirsch, I. B., & on behalf of the Diabetes in Hospitals Writing Committee (2004) · Diabetes Care · doi:10.2337/diacare.27.2.553 · Cited by 1,266
Glucose Biosensors: An Overview of Use in Clinical Practice
Yoo, E.-H., & Lee, S.-Y. (2010) · Sensors · doi:10.3390/s100504558 · Cited by 1,137
Sacks, D. B., Bruns, D. E., Goldstein, D. E., Maclaren, N. K., McDonald, J. M., & Parrott, M. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.3.436 · Cited by 1,130
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Diabetes Care · doi:10.2337/dc11-9998 · Cited by 718
Metadata
- Authors
- Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J.
- Affiliations
- University of California, Davis (US)
- ORCID
- Kost, Gerald J.
- Venue
- Archives of Pathology & Laboratory Medicine
- Published
- 2000-08
- Volume
- 124
- Issue
- 8
- Pages
- 1135-1140
- ISSN
- 0003-9985 (ISSN-L)
- 1543-2165
- OpenAlex
- W1883317613
- Cited by
- 293
- References
- 23
- Citations by year
- 2026: 10 · 2025: 6 · 2024: 1 · 2023: 5 · 2022: 10 · 2021: 8 · 2020: 9 · 2019: 10 · 2018: 13 · 2017: 10 · 2016: 10 · 2015: 14 · 2014: 14 · 2013: 21 · 2012: 11
- FWCI
- 3.06
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Archives of Pathology & Laboratory Medicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- hematocrit, glucose measurement, point-of-care testing, measurement bias, plasma glucose, meter calibration
- MeSH terms
- Blood Glucose, California, Diagnostic Techniques, Endocrine, Evaluation Studies as Topic, Hematocrit, Humans, Predictive Value of Tests, Reproducibility of Results, Biosensing Techniques, Point-of-Care Systems, Diagnostic Equipment
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Effects of Different Hematocrit Levels on Glucose Measurements With Handheld Meters for Point-of-Care Testing. https://app.cochat.ai/papers/10.5858/2000-124-1135-eodhlo
- The paper
- Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000). Effects of Different Hematocrit Levels on Glucose Measurements With Handheld Meters for Point-of-Care Testing. Archives of Pathology & Laboratory Medicine. https://doi.org/10.5858/2000-124-1135-eodhlo
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.