Accuracy of Self Monitoring Blood Glucose Systems in a Clinical Setting: Application of New Planned ISO- Standards
The DOI matches its Crossref record.
Abstract
BACKGROUND: The accuracy of glucose meters (GMs) for self-monitoring of blood glucose (BG) is essential for preventing insulin dosage errors in insulin injecting diabetic patients. We evaluated the accuracy of 19 commercially available GMs in a clinical setting applying the new planned ISO criteria. METHODS: Capillary blood was taken at 10-minute intervals in insulin dependent patients during induction of a BG increase and decrease by administration of a standardized carbohydrate rich meal and/or injection of an appropriate dosage of insulin under medical supervision. BG was measured simultaneously by GMs and a laboratory method (Hitado Super GL). The GM results were compared with a laboratory reference method according to the current ISO criteria (ISO 15197:2003) and the proposed (more stringent) criteria. RESULTS: Evaluating GM results according to the current ISO criteria, all GMs reached the minimum acceptable system accuracy criteria. Analysing GM results according to the more stringent planned ISO standard criteria, only 8 of 19 GMs reached the minimum acceptable accuracy criteria, i.e. > or = 95% of the individual glucose results fell within +/- 15 mg/dL of the results of the reference value at glucose concentrations < 100 mg/dL and within +/- 15% at glucose concentrations > or = 100 mg/dL. When results obtained in BG ranges < and > or = 100 mg/dL were separately analysed, readings of 10 and 8 GMs, respectively, fell in the area of acceptable accuracy. CONCLUSIONS: The study shows that more than half of the tested GMs would fail the proposed tighter ISO criteria. Most GMs reveal similar accuracy in BG measurement when compared at lower and higher BG ranges.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Freckmann, G., Schmid, C., Baumstark, A., Rutschmann, M., Haug, C., & Heinemann, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815580160 · Cited by 70
A review of blood glucose monitor accuracy
King, F., Ahn, D., Hsiao, V., Porco, T., & Klonoff, D. C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0232 · Cited by 37
Boettcher, C., Dost, A., Wudy, S. A., Flechtner-Mors, M., Borkenstein, M., Schiel, R., Weitzel, D., Bechtold-Dalla Pozza, S., Wolf, J., Holl, R. W., et al. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0262 · Cited by 35
Hasslacher, C., Kulozik, F., & Platten, I. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814522804 · Cited by 22
Self-citationYu-Fei, W., Wei-Ping, J., Ming-Hsun, W., Miao-O, C., Ming-Chang, H., Chi-Pin, W., & Ming-Shih, L. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817705143 · Cited by 15
Freckmann, G., Baumstark, A., Jendrike, N., Mende, J., Schauer, S., Link, M., Pleus, S., & Haug, C. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820948873 · Cited by 12
Metadata
- Authors
- Hasslacher, C., Kulozik, F., & Platten, I.
- Venue
- Clinical Laboratory
- Publisher
- Clinical Laboratory Publications
- Published
- 2013
- Volume
- 59
- Issue
- 07+08/2013
- Pages
- 727-33
- ISSN
- 1433-6510 (ISSN-L)
- OpenAlex
- W2074902079
- Cited by
- 26
- References
- 0
- FWCI
- 2.53
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.