Correction algorithm for online continuous flow δ 13 C and δ 18 O carbonate and cellulose stable isotope analyses
Open accessAbstract
We describe an algorithm to correct for scale compression, runtime drift, and amplitude effects in carbonate and cellulose oxygen and carbon isotopic analyses made on two online continuous flow isotope ratio mass spectrometry (CF-IRMS) systems using gas chromatographic (GC) separation. We validate the algorithm by correcting measurements of samples of known isotopic composition which are not used to estimate the corrections. For carbonate δ13C (δ18O) data, median precision of validation estimates for two reference materials and two calibrated working standards is 0.05‰ (0.07‰); median bias is 0.04‰ (0.02‰) over a range of 49.2‰ (24.3‰). For α-cellulose δ13C (δ18O) data, median precision of validation estimates for one reference material and five working standards is 0.11‰ (0.27‰); median bias is 0.13‰ (−0.10‰) over a range of 16.1‰ (19.1‰). These results are within the 5th–95th percentile range of subsequent routine runtime validation exercises in which one working standard is used to calibrate the other. Analysis of the relative importance of correction steps suggests that drift and scale-compression corrections are most reliable and valuable. If validation precisions are not already small, routine cross-validated precision estimates are improved by up to 50% (80%). The results suggest that correction for systematic error may enable these particular CF-IRMS systems to produce δ13C and δ18O carbonate and cellulose isotopic analyses with higher validated precision, accuracy, and throughput than is typically reported for these systems. The correction scheme may be used in support of replication-intensive research projects in paleoclimatology and other data-intensive applications within the geosciences.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 18 cited sources checked · 0 supported · 0 flagged
- Findingown data
For the LSVEC reference material, the precision achieved is better than reported in Stichler [1995], which was 0.15 permil for d13C and 0.25 permil for d18O.
“For the LSVEC reference material, the precision achieved is better than reported in Stichler [1995], which was 0.15 and 0.25& for d C and d O , respectively.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Stichler, 1995
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Findingown data
For carbonate d18O data, median precision of validation estimates for two reference materials and two calibrated working standards is 0.07 permil.
“For carbonate d 13 C (d 18 O) data, median precision of validation estimates for two reference materials and two calibrated working standards is 0.05& (0.07&);”
- Also stated in: § 5. Discussion · plain
“For the carbonate materials we analyzed, we found that the median validation precisions across the carbonates analyzed are 0.05 and 0.07& for d C and d O , respectively (Table 1).”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Also stated in: § 5. Discussion · plain
- Findingown data
These biases are small relative to the calibrated ranges in δ13C and δ18O across all materials analyzed (16.1 and 19.1‰, respectively).
“but these biases are again small relative to the calibrated ranges in d C and d O across all materials analyzed (16.1 and 19.1&, respectively; Table 2 and Figures 1c and 1d)”
Quote in the paperSubstantial claim - Findingown data
These validation results are within the 5th-95th percentile range of subsequent routine runtime validation exercises in which one working standard is used to calibrate the other.
“These results are within the 5th-95th percentile range of subsequent routine runtime validation exercises in which one working standard is used to calibrate the other.”
Quote in the paperSubstantial claim - Findingown data
For carbonate d13C data, median bias is 0.04 permil over a range of 49.2 permil.
“median bias is 0.04& (0.02&) over a range of 49.2& (24.3&).”
- Also stated in: § 5. Discussion · plain
“However, the median biases (0.04 and 0.02& for d C and d O , respectively; Table 1) are small relative to the known range across in d C and d O across working standards (49.2 and 24.3&, respectively; Table 1 and Figures 1a and 1b).”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Also stated in: § 5. Discussion · plain
- Findingown data
For carbonate d18O data, median bias is 0.02 permil over a range of 24.3 permil.
“median bias is 0.04& (0.02&) over a range of 49.2& (24.3&).”
- Also stated in: § 5. Discussion · plain
“However, the median biases (0.04 and 0.02& for d C and d O , respectively; Table 1) are small relative to the known range across in d C and d O across working standards (49.2 and 24.3&, respectively; Table 1 and Figures 1a and 1b).”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Also stated in: § 5. Discussion · plain
- Findingown data
For carbonate d13C data, median precision of validation estimates for two reference materials and two calibrated working standards is 0.05 permil.
“For carbonate d 13 C (d 18 O) data, median precision of validation estimates for two reference materials and two calibrated working standards is 0.05& (0.07&);”
- Also stated in: § 5. Discussion · plain
“For the carbonate materials we analyzed, we found that the median validation precisions across the carbonates analyzed are 0.05 and 0.07& for d C and d O , respectively (Table 1).”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Also stated in: § 5. Discussion · plain
- Findingown data
Of 12 comparisons, 5 are statistically distinguishable from zero bias.
“include 5 of 12 comparisons that are statistically distinguishable from zero bias”
Quote in the paperNot a substantial claim - Findingown data
For a-cellulose d18O data, median bias is 20.10 permil over a range of 19.1 permil.
“median bias is 0.13& (20.10&) over a range of 16.1& (19.1&).”
Quote in the paperNot a substantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Findingown data
In practice over additional routine tests, the amplitude correction is rarely significant at P < 0.10 for both correction standards.
“In practice over additional routine tests (last two columns of Tables 3 and 4), the amplitude correction is rarely significant at P < 0.10 for both correction standards.”
Quote in the paperSubstantial claim - Findingown data
If validation precisions are not already small, routine cross-validated precision estimates are improved by up to 50% for d13C and 80% for d18O.
“If validation precisions are not already small, routine cross-validated precision estimates are improved by up to 50% (80%).”
Quote in the paperSubstantial claim - Findingown data
Differential reaction and equilibration times for batch acid versus sequential acid injection cause a systematic drift in d18O measurements.
“analyses of batch acid versus sequential acid injection and analyses (results not shown) suggest that differential reaction and equilibration times for the former procedure cause a systematic drift in the d 18 O measurements.”
Quote in the paperSubstantial claim - Findingown data
In the single validation studies presented, some spline and amplitude corrected validation precisions are higher than raw precisions.
“In the results of the single validation studies presented (Tables 1 and 2), some spline and amplitude corrected validation precisions are higher than raw precisions”
Quote in the paperNot a substantial claim - Findingreasoning
Because precisions are small, the biases in some of the validation data sets (JTB1, EACARB) are statistically different from zero.
“Because precisions are small, the biases in some of the validation data sets (JTB1, EACARB) are statistically different from zero.”
Quote in the paperNot a substantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Findingown data
For a-cellulose d18O data, median precision of validation estimates for one reference material and five working standards is 0.27 permil.
“For a-cellulose d 13 C (d 18 O) data, median precision of validation estimates for one reference material and five working standards is 0.11& (0.27&);”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Findingown data
For a-cellulose d13C data, median precision of validation estimates for one reference material and five working standards is 0.11 permil.
“For a-cellulose d 13 C (d 18 O) data, median precision of validation estimates for one reference material and five working standards is 0.11& (0.27&);”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Findingown data
For a-cellulose d13C data, median bias is 0.13 permil over a range of 16.1 permil.
“median bias is 0.13& (20.10&) over a range of 16.1& (19.1&).”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Findingown data
Validation study precisions and biases for both carbonate and cellulose experiments fall within the 5th-95th percentile ranges of the routine runtime precision and bias estimates.
“Validation study precisions and biases for both carbonate and cellulose experiments (Tables 1 and 2) fall within the 5th-95th percentile ranges of the routine runtime precision and bias estimates (Figures 2 and 3).”
Quote in the paperSubstantial claim - Findingreasoning
Two-point precision is approximately that achieved after the drift and amplitude correction in the validation study and in routine cross-validation of many batch runs.
“Two-point precision is approximately that achieved after the drift and amplitude correction in the validation study and in routine cross-validation of many batch runs.”
Quote in the paperSubstantial claim - Findingown data
Reversing the order of drift and amplitude corrections for both the carbonate and cellulose validation experiments gives similar validation precision and bias estimates reported in the abstract and as shown in Tables 1 and 2.
“Reversing the order of drift and amplitude corrections for both the carbonate and cellulose validation experiments gives similar validation precision and bias estimates reported in the abstract and as shown in Tables 1 and 2.”
Quote in the paperNot a substantial claim - Findingnone given
The algorithm described here detects and corrects for three potential sources of systematic uncertainty in CF-IRMS data streams.
“Although the algorithm described here detects and corrects for three potential sources of systematic uncertainty in CF-IRMS data streams”
Quote in the paperSubstantial claim - Premisecitation
Measurements in cellulose pyrolysis are commonly subject to runtime isotopic drift, arising from changes in the state of the thermal conversion reactor as consumables are used and elemental sample carbon is added to the system.
“Measurements are also commonly subject to runtime isotopic drift, arising from changes in the state of the thermal conversion reactor as consumables are used and elemental sample carbon is added to the system [e.g., Young et al., 2011].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Young et al., 2011 source
- Study design and conductown data
Instrument linearity for the carbonate system was confirmed to be less than 0.03 permil/nA over a range of approximately 2-14 nA.
“Prior to batch analysis, instrument linearity (dependence of d 18 O and d 13 C on signal at the collectors) was confirmed to be <0.03&/nAover a range of approximately 2-14 nA.”
Quote in the paperNot a substantial claim - Interpretationnone given
If systematic errors in isotopic analyses can be detected, they can be modeled and corrected for, resulting in more precise and accurate results.
“In principle, if systematic errors in isotopic analyses can be detected, we can model and correct for them, resulting in more precise and accurate results.”
Quote in the paperSubstantial claim - Premisereasoning
Dual C/O measurements on cellulose samples in a low-temperature pyrolysis system at 1080 degrees C may be subject to scale compression and bias in both isotopes.
“Similarly, dual C/O measurements on cellulose samples in a low-temperature pyrolysis system (10808C) [Werner et al., 1996] may be subject to scale compression and bias in both isotopes.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Werner et al., 1996 source
- Interpretationown data
The spline and two-point corrections are most valuable when raw precisions are poor.
“Results as a function of correction step (Tables 3 and 4) suggest that the spline and two-point corrections are most valuable when raw precisions are poor (Tables 3 and 4, last two columns).”
Quote in the paperSubstantial claim - Interpretationreasoning
Uncertainty in the correction scheme presented here arises from both empirical and structural uncertainties.
“Uncertainty in the correction scheme presented here arises from both empirical and structural uncertainties.”
Quote in the paperSubstantial claim - Study design and conductown data
Separation of CO and N2 peaks was verified by analysis of the chromatogram.
“Separation of CO and N 2 peaks (a small amount of N, with characteristically low 30/28 ratio, may be present in industrial or extracted a-cellulose [Anchukaitis et al., 2008]) was verified by analysis of the chromatogram.”
Quote in the paperNot a substantial claim - Interpretationown data
There might be artificial skill in these corrections if raw precision is already low.
“suggesting that there might be artificial skill in these corrections if raw precision is already low (first four columns of Tables 3 and 4).”
Quote in the paperSubstantial claim - Interpretationown data
Drift and scale-compression corrections are the most reliable and valuable correction steps.
“Analysis of the relative importance of correction steps suggests that drift and scale-compression corrections are most reliable and valuable.”
Quote in the paperSubstantial claim - Interpretationown data
Correction for systematic error may enable these particular CF-IRMS systems to produce d13C and d18O carbonate and cellulose isotopic analyses with higher validated precision, accuracy, and throughput than is typically reported for these systems.
“may enable these particular CF-IRMS systems to produce d 13 C and d 18 O carbonate and cellulose isotopic analyses with higher validated precision, accuracy, and throughput than is typically reported for these systems.”
Quote in the paperSubstantial claim - Interpretationown data
The algorithm corrects for scale compression, runtime drift, and amplitude effects in carbonate and cellulose oxygen and carbon isotopic analyses made on two online CF-IRMS systems using gas chromatographic separation.
“We describe an algorithm to correct for scale compression, runtime drift, and amplitude effects in carbonate and cellulose oxygen and carbon isotopic analyses made on two online continuous flow isotope ratio mass spectrometry (CF-IRMS) systems using gas chromatographic (GC) separation.”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
- Interpretationown data
The long-term systematic bias in corrected data is not significantly different from zero.
“and that the long-term systematic bias in corrected data is not significantly different from zero.”
Quote in the paperSubstantial claim - Study design and conductown data
No measurement blank correction was applied to carbonate analyses because routine analyses on empty, He-flushed and acidified sample vials give signals of 0.05-0.1 nA.
“No measurement blank correction has been applied, as routine analyses on empty, He-flushed and acidified sample vials give signals of 0.05-0.1 nA.”
Quote in the paperNot a substantial claim - Study design and conductcitation
Using working standards of known isotopic composition and similar chemical composition as the basis for corrections, and analyzing them in batches together with unknowns, automatically removes differential working-standard-versus-unknown-sample biases (the IT principle).
“Because we use working standards of known isotopic composition and similar chemical composition as the basis for these corrections, and because these working standards are analyzed in batches together with the unknowns, differential (working standard versus unknown sample) biases are automatically removed, an approach known as the principle of Identical Treatment (IT principle) [Werner and Brand,”
Quote in the paperSubstantial claim0 of 2 sources checked- Rests on an assumption: The working standards used for correction have chemical composition similar to the unknown samples.
- Interpretationcitation
The α-cellulose validation precisions are comparable to those achieved routinely in high-temperature pyrolysis systems with near-quantitative conversion of α-cellulose into analyte CO above 1400°C.
“These precisions are comparable to those achieved routinely in high-temperature pyrolysis systems with near-quantitative conversion of a-cellulose into analyte CO above 14008C [Leuenberger and Filot, 2007].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Leuenberger and Filot, 2007 source
- Interpretationreasoning
These biases may be negligible in practical measurement and application.
“and may be negligible in practical measurement and application”
Quote in the paperSubstantial claim - Premisecitation
Runtime isotopic drift in oxygen isotopic composition may arise in continuous flow preparation systems from systematic differences in the time between acidification and analysis of evolved CO2 analyte gas.
“Depending on analytical method, runtime isotopic drift in oxygen isotopic composition may arise in continuous flow preparation systems from systematic differences in the time between acidification and analysis of evolved CO 2 analyte gas [Spotl, 2011];”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Spotl, 2011 source
- Premisecitation
Amplitude or integrated peak area biases in measured isotopic composition may arise in the analysis of carbonate isotopic composition by continuous flow mass spectrometry.
“For example, it has been observed that amplitude or integrated peak area biases in measured isotopic composition may arise in the analysis of carbonate isotopic composition by continuous flow mass spectrometry [Spotl and Vennemann, 2003].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Spotl and Vennemann, 2003 source
- Study design and conductown data
Instrument linearity for the cellulose system was tested daily and confirmed to be less than 0.03 permil/nA over the 1-10 nA range.
“Instrument linearity (dependence of d 18 O and d 13 C on signal amplitude at the collectors) was tested daily and confirmed to be <0.03&/nAover the 1-10 nA range.”
Quote in the paperNot a substantial claim - Study design and conductown data
The algorithm was validated by correcting measurements of samples of known isotopic composition that were not used to estimate the corrections.
“We validate the algorithm by correcting measurements of samples of known isotopic composition which are not used to estimate the corrections.”
- Also stated in: § PUBLICATIONS · plain
“We validate the algorithm in two online CF-IRMS contexts-measurement on carbonate-derived analyte CO 2 and measurement on a-cellulose-derived analyte CO-by applying corrections to samples of known isotopic composition but which were not used to estimate the corrections themselves.”
Quote in the paperSubstantial claim - Also stated in: § PUBLICATIONS · plain
- Interpretationcitation
Results are similar to those achieved by Young et al. [2011].
“results are similar to those achieved by Young et al. [2011].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Young et al. [2011] source
- Interpretationnone given
Empirical sources of uncertainty include the statistical estimation of the spline, amplitude, and two-point corrections.
“Empirical sources of uncertainty include the statistical estimation of the spline, amplitude, and two-point corrections.”
Quote in the paperSubstantial claim - Interpretationnone given
CF-IRMS analyses using GC separation may be subject to runtime drift, signal amplitude and scale compression biases arising from changes in preparation system and mass analyzer conditions.
“CF-IRMS analyses using GC separation may be subject to runtime drift, signal amplitude and scale compression biases arising from changes in preparation system and mass analyzer conditions.”
Quote in the paperNot a substantial claim - Interpretationreasoning
Detection and correction of these biases may be used to produce corrected data sets with improved and validated precision and accuracy.
“Detection and correction of these biases may be used to produce corrected data sets with improved and validated precision and accuracy.”
Quote in the paperSubstantial claim - Interpretationown data
There is no structural reason why the amplitude correction might not be performed prior to the drift correction.
“However, there is no structural reason why the amplitude correction might not be performed prior to the drift correction.”
Quote in the paperNot a substantial claim - Interpretationown data
The data suggest that results are not sensitive to order of operations.
“the data suggest that results are not sensitive to order of operations.”
Quote in the paperNot a substantial claim - Interpretationreasoning
Together with objective and automated data correction algorithms such as those presented here, high-throughput CF-IRMS systems enable rapid development of validated, precise, and accurate data streams in support of a wide range of data-intensive geoscientific applications.
“Together with objective and automated data correction algorithms such as those presented here, high-throughput CF-IRMS systems enable rapid development of validated, precise, and accurate data streams in support of a wide range of data-intensive geoscientific applications.”
Quote in the paperNot a substantial claim - Interpretationown data
The small effect of amplitude scaling reflects scaling factors routinely determined to be within 10% of unity in both systems studied here.
“This reflects the small effect of amplitude scaling by scaling factors routinely determined to be within 10% of unity in both systems studied here.”
Quote in the paperNot a substantial claim
- May combine two claims. 10
- Classified as probably not a substantial claim. 11
May combine two claims.
Findingown dataFor the α-cellulose materials analyzed, median validation precisions across five α-cellulose working standards and one reference material are 0.11‰ for δ13C and 0.27‰ for δ18O.
“For the a-cellulose materials we analyzed, we found that the median validation precisions across five acellulose working standards and one reference material are 0.11 and 0.27& for d C and d O , respectively (Table 2).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataMedian bias for α-cellulose validation is 0.13‰ for δ13C and -0.10‰ for δ18O.
“Median bias (0.13 and 20.10& for d C and d O , respectively; Table 2)”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor the carbonate materials analyzed, the median validation precisions across the carbonates analyzed are 0.05 permil for d13C and 0.07 permil for d18O.
“For the carbonate materials we analyzed, we found that the median validation precisions across the carbonates analyzed are 0.05 and 0.07& for d C and d O , respectively (Table 1).”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
May combine two claims.
Findingown dataThe median biases (0.04 permil for d13C and 0.02 permil for d18O) are small relative to the known range across working standards (49.2 permil for d13C and 24.3 permil for d18O).
“However, the median biases (0.04 and 0.02& for d C and d O , respectively; Table 1) are small relative to the known range across in d C and d O across working standards (49.2 and 24.3&, respectively; Table 1 and Figures 1a and 1b).”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
InterpretationreasoningThe correction scheme may be used in support of replication-intensive research projects in paleoclimatology and other data-intensive applications within the geosciences.
“The correction scheme may be used in support of replication-intensive research projects in paleoclimatology and other data-intensive applications within the geosciences.”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductown dataNo measurement blank correction was applied to cellulose analyses because routine analyses on empty silver sample capsules give signals of less than 0.1 nA.
“No measurement blank correction has been applied, as routine analyses on empty silver sample capsules give signals of less than 0.1 nA.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductreasoningBecause materials of known and substantially different isotopic composition are few, the algorithm is applied in validation to samples of materials that were also used to correct the reported data (IAEA-CO-1, LSVEC, and AKC).
“Note that because materials of known and substantially different isotopic composition are few, the algorithm is applied in validation to samples of materials that were also used to correct the reported data (i.e., IAEA-CO-1, LSVEC, and AKC).”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationScale compression and bias in cellulose pyrolysis measurements are due to the presence of catalytic glassy carbon and deposition of excess sample carbon in the pyrolysis reactor, and thermal conversion into both CO and CO2 at this temperature.
“This is because of the presence of catalytic glassy carbon and deposition of excess sample carbon in the pyrolysis reactor [Young et al., 2011], and thermal conversion into both CO and CO 2 at this temperature [Leuenberger and Filot, 2007].”
Quote in the paperSubstantial claim0 of 2 sources checkedClassified as probably not a substantial claim.
Study design and conductreasoningAnalysis of samples of known isotopic composition which were not themselves used to estimate the corrections is required for assessment of the results.
“Analysis of samples of known isotopic composition which were not themselves used to estimate the corrections is required for assessment of the results.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
PremisereasoningBiases associated with measurement conditions common to both working standards and unknown samples, if present, must be detected and removed.
“However, biases associated with measurement conditions common to both working standards and unknown samples, if present, must be detected and removed.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataMedian biases are small relative to the distribution of routine runtime cross-validation precisions, and hence may be negligibly small in practical measurement and application.
“Median biases are also small relative to the distribution of routine runtime cross-validation precisions (Figure 2), and hence may be negligibly small in practical measurement and application.”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
Classified as probably not a substantial claim. May combine two claims.
Study design and conductreasoningIn each experiment, the sample data were not used to develop the corrections; they are out-of-sample tests of the validity of the corrections.
“In each experiment, the ''samples'' were of materials of a priori known isotopic composition, but the ''sample'' data were not used to develop the corrections; in other words, these are ''out-of-sample'' tests of the validity of the corrections.”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
Probably not a substantial claim: may not pass S2 (checkable).
PremisecitationMean and variance-calibrated, high-throughput stable isotope analyses are increasingly required in paleoclimatology.
“Mean and variance-calibrated, high-throughput stable isotope analyses are increasingly required in paleoclimatology.”
Quote in the paperNot a substantial claim0 of 3 sources checkedClassified as probably not a substantial claim. May combine two claims.
PremisecitationCF-IRMS is a rapid, economical, but relatively low-precision approach to high-throughput stable isotopic analyses of carbonate, cellulose, and similar materials.
“Continuous flow isotope ratio mass spectrometry (CF-IRMS) is a rapid, economical, but relatively low-precision approach to high-throughput stable isotopic analyses of these materials [Brenna et al., 1997;Werner et al., 1996;Spotl and Vennemann, 2003;Spotl, 2011].”
Quote in the paperSubstantial claim0 of 3 sources checkedProbably not a substantial claim: may not pass S2 (checkable).
Study design and conductreasoningThe more complex the suite of corrections, the more important that their efficacy is evaluated using validation tests, because of the potential for artificial skill from overfitting of the calibration data.
“However, the more complex the suite of corrections, the more important that their efficacy is evaluated using validation tests, because of the potential for artificial skill from overfitting of the calibration data.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
PremisecitationIsotopic analyses should be corrected for scale differences arising from using intermediate working standards to correct to international reference scales.
“Isotopic analyses should also be corrected for scale differences arising from using intermediate working standards to correct to international reference scales [Coplen et al., 2006].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Coplen et al., 2006 source
Classified as probably not a substantial claim.
Study design and conductnone givenBy construction, the means of the final corrected calibration data are the known isotopic values of the working standards on the international reference scales.
“By construction, the means of the final corrected calibration data are the known isotopic values of the working standards on the international reference scales.”
Quote in the paperSubstantial claim- Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
InterpretationreasoningThe algorithm merely contributes to the comprehensive set of quality controls and standard operating protocols that produce long-term stability, precision, accuracy, and traceability of corrected data.
“it merely contributes to the comprehensive set of quality controls and standard operating protocols that produce long-term stability, precision, accuracy, and traceability of corrected data.”
Quote in the paperNot a substantial claim
The paper relies on this without support.
The precision on validation data should be approximately equal to the precision of the corrected calibration data.
“The precision on validation data should be approximately equal to the precision of the corrected calibration data.”
The paper relies on this without support.
The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
“with known d 13 C and d 18 O values for SAC and AKC with respect to the V-PDB and V-SMOW reference standards previously determined and as listed in Table 2.”
The paper relies on this without support.
The working standards used for correction have chemical composition similar to the unknown samples.
“Because we use working standards of known isotopic composition and similar chemical composition as the basis for these corrections,”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Cellulose significance count disagrees with Table 2
MinorThe discussion says five of twelve bias comparisons are statistically distinguishable from zero, but Table 2 marks six comparisons with asterisks.
“include 5 of 12 comparisons that are statistically distinguishable from zero bias”
“| AWS | 222.22 (0.07) | 20.08* | 21.42 (0.22) | 20.12 | | H1 | 229.91 (0.26) | 0.17* | 20.13 (0.27) | 20.23* | | L6 | 229.23 (0.28) | 20.07 | 39.24 (0.75) | 0.56* | | Switsur | 224.31 (0.09) | 0.19* | 27.38 (0.32) | 0.08 | | IAEA-CH3 | 224.96 (0.13) | 0.24* | 32.61 (0.27) | 20.09 |”
Cites
McCarroll, D., & Loader, N. J. (2004) · Quaternary Science Reviews · doi:10.1016/j.quascirev.2003.06.017 · Cited by 1,817 · see citing works
Referencing strategies and techniques in stable isotope ratio analysis
Werner, R. A., & Brand, W. A. (2001) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.258 · Cited by 1,033 · see citing works
El Niño/Southern Oscillation and tropical Pacific climate during the last millennium
Cobb, K. M., Charles, C. D., Cheng, H., & Edwards, R. L. (2003) · Nature · doi:10.1038/nature01779 · Cited by 924 · see citing works
New Guidelines for <i>δ</i> <sup>13</sup> C Measurements
Coplen, T. B., Brand, W. A., Gehre, M., Gröning, M., Meijer, H. A. J., Toman, B., & Verkouteren, R. M. (2006) · Analytical Chemistry · doi:10.1021/ac052027c · Cited by 906 · see citing works
Assessment of international reference materials for isotope-ratio analysis (IUPAC Technical Report)
Brand, W. A., Coplen, T. B., Vogl, J., Rosner, M., & Prohaska, T. (2014) · Pure and Applied Chemistry · doi:10.1515/pac-2013-1023 · Cited by 704 · see citing works
Continuous‐flow isotope ratio mass spectrometric analysis of carbonate minerals
Spötl, C., & Vennemann, T. W. (2003) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.1010 · Cited by 679 · see citing works
Highly Variable El Niño–Southern Oscillation Throughout the Holocene
Cobb, K. M., Westphal, N., Sayani, H. R., Watson, J. T., Lorenzo, E. D., Cheng, H., Edwards, R. L., & Charles, C. D. (2013) · Science · doi:10.1126/science.1228246 · Cited by 486 · see citing works
Koutavas, A., deMenocal, P. B., Olive, G. C., & Lynch-Stieglitz, J. (2006) · Geology · doi:10.1130/g22810a.1 · Cited by 307 · see citing works
Boettger, T., Haupt, M., Knöller, K., et al. (2007) · Analytical Chemistry · doi:10.1021/ac0700023 · Cited by 222 · see citing works
Tracking the extent of the South Pacific Convergence Zone since the early 1600s
Linsley, B. K., Kaplan, A., Gouriou, Y., Salinger, J., deMenocal, P. B., Wellington, G. M., & Howe, S. S. (2006) · Geochemistry Geophysics Geosystems · doi:10.1029/2005gc001115 · Cited by 152 · see citing works
A weak El Niño/Southern Oscillation with delayed seasonal growth around 4,300 years ago
McGregor, H. V., Fischer, M. J., Gagan, M. K., Fink, D. W., Phipps, S. J., Wong, H. K. Y., & Woodroffe, C. D. (2013) · Nature Geoscience · doi:10.1038/ngeo1936 · Cited by 125 · see citing works
Linsley, B. K., Zhang, P., Kaplan, A., Howe, S. S., & Wellington, G. M. (2008) · Paleoceanography · doi:10.1029/2007pa001539 · Cited by 110 · see citing works
On-line determination of δ18O values of organic substances
Werner, R. A., Kornexl, B. E., Roβmann, A., & Schmidt, H.-L. (1996) · Analytica Chimica Acta · doi:10.1016/0003-2670(95)00468-8 · Cited by 108 · see citing works
Tropical cloud forest climate variability and the demise of the Monteverde golden toad
Anchukaitis, K. J., & Evans, M. N. (2010) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0908572107 · Cited by 101 · see citing works
Self-citationSpötl, C. (2011) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.5037 · Cited by 96 · see citing works
Consequences of a Rapid Cellulose Extraction Technique for Oxygen Isotope and Radiocarbon Analyses
Anchukaitis, K. J., Evans, M. N., Lange, T. E., Smith, D. R. R., Leavitt, S. W., & Schrag, D. P. (2008) · Analytical Chemistry · doi:10.1021/ac7020272 · Cited by 72 · see citing works
Self-citationAdaptation of the unterzaucher procedure for determination of Oxygen-18 in organic substances
Santrock, J., & Hayes, J. M. (1987) · Analytical Chemistry · doi:10.1021/ac00128a025 · Cited by 57 · see citing works
Leuenberger, M. C., & Filot, M. S. (2007) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.2998 · Cited by 37 · see citing works
Young, G. H. F., Loader, N. J., & McCarroll, D. (2011) · Palaeogeography, Palaeoclimatology, Palaeoecology · doi:10.1016/j.palaeo.2010.11.018 · Cited by 24 · see citing works
Evans, M. N. (2008) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.3603 · Cited by 20 · see citing works
Self-citationBrenna (1997) · doi:10.1002/(sici)1098-2787(1997)16:5
Flagged first, then by citations
1 of the 22 references couldn’t be matched to a known paper.
Research works printed but not linked: 4
- 2008, Handbook of Stable Isotope Analytical Techniques, 2, 1398
- IsoPrime Ltd., 2009, IsoPrime User Guide 2009A
- Stichler, 1995, Reference and Intercomparison Materials for Stable Isotopes of Light Elements, 67
- Thermo Electron Corporation, 2005, DELTA V Plus Operating Manual, Revision A - 118 4270
Cited by
Sorted by citations
Cui, H., Kaufman, A. J., Xiao, S., Zhou, C., & Liu, X. (2016) · Chemical Geology · doi:10.1016/j.chemgeo.2016.12.010 · Cited by 161 · see citing works
“…scripting on the VPDB and LSVEC scale using periodic in-run measurement of international reference carbonate materials and in-house standard carbonates, from which empirical corrections for signal amplitude, sequential drift, and one or two-point mean corrections were applied (Evans et al., 2016).”
MethodCui, H., Xiao, S., Cai, Y., Peek, S., Plummer, R. E., & Kaufman, A. J. (2019) · Geological Magazine · doi:10.1017/s0016756819000293 · Cited by 80 · see citing works
“…flow isotope ratio mass spectrometry in the University of Maryland Palaeoclimate Laboratory using a refined method for the analysis and correction of carbon ( δ 13 C carb ) and oxygen Y a n g t z e ( δ 18 O carb ) isotopic compositions of 100 μ g carbonate samples (Spötl, 2011; Evans et al. 2016).”
BackgroundMethodInfluentialSelf-citationBelmecheri, S., Wright, W. E., Szejner, P., et al. (2018) · Plant Cell & Environment · doi:10.1111/pce.13401 · Cited by 74 · see citing works
Tree-ring isotopes capture interannual vegetation productivity dynamics at the biome scale
Lévesque, M., Andreu‐Hayles, L., Smith, W. K., Williams, P., Hobi, M. L., Allred, B., & Pederson, N. (2019) · Nature Communications · doi:10.1038/s41467-019-08634-y · Cited by 60 · see citing works
Andreu-Hayles, L., Lévesque, M., Martin‐Benito, D., Huang, W., Harris, R. A., Oelkers, R., Leland, C., Fernández, J. M., Anchukaitis, K., & Helle, G. (2019) · Chemical Geology · doi:10.1016/j.chemgeo.2018.09.007 · Cited by 45 · see citing works
“At LDEO, data from the IRMS are corrected using a two-point calibration (e.g., Evans et al., 2016) with two international well-defined reference materials (IAEA) that fully cover the expected δ-range of the samples (Fig.”
BackgroundMethodInfluentialLévesque, M., Andreu‐Hayles, L., & Pederson, N. (2017) · Scientific Reports · doi:10.1038/srep46158 · Cited by 49 · see citing works
“For each sample, 200 μ g of cellulose were weighted and put in silver capsules. δ 13 C and δ 18 O were measured simultaneously using high-temperature pyrolysis in a Costech elemental analyzer interfaced with an Elementar Isoprime mass spectrometer at the Department of Geology at the University of Maryland, USA 51 .”
Szejner, P., Clute, T., Anderson, E., et al. (2020) · New Phytologist · doi:10.1111/nph.16484 · Cited by 32 · see citing works
Self-citationLorrey, A. M., Brookman, T. H., Evans, M. N., Fauchereau, N., Macinnis‐Ng, C., Barbour, M. M., Criscitiello, A. S., Eischeid, G., Fowler, A. M., Horton, T. W., et al. (2016) · Dendrochronologia · doi:10.1016/j.dendro.2016.03.012 · Cited by 20 · see citing works
Self-citationComparative Foliar Atmospheric Mercury Accumulation across Functional Types in Temperate Trees
Zhang, X., Kang, H., Liu, X., et al. (2025) · Environmental Science & Technology · doi:10.1021/acs.est.4c09462 · Cited by 16 · see citing works
Zeng, X., Ni, P., Li, Y., et al. (2021) · Dendrochronologia · doi:10.1016/j.dendro.2021.125895 · Cited by 12 · see citing works
Xu, G., Liu, X., Hu, J., Dorado‐Liñán, I., Gagen, M. H., Szejner, P., Chen, T., & Trouet, V. M. (2022) · Tree Physiology · doi:10.1093/treephys/tpac076 · Cited by 12 · see citing works
Kulenguski, J. T., Gilleaudeau, G. J., Kaufman, A. J., Kipp, M. A., Tissot, F. L. H., Goepfert, T. J., Pitts, A. D., Pierantoni, P. P., Evans, M. N., & Elrick, M. (2023) · Palaeogeography Palaeoclimatology Palaeoecology · doi:10.1016/j.palaeo.2023.111756 · Cited by 10 · see citing works
Self-citationGea‐Izquierdo, G., Férriz, M., Conde, M., Evans, M. N., Querejeta, J., & Martin‐Benito, D. (2025) · Agricultural and Forest Meteorology · doi:10.1016/j.agrformet.2025.110430 · Cited by 10 · see citing works
Self-citationAssessing the efficacy of isotopic provenancing of human remains in Colombia
Eck, C. J., DiGangi, E. A., & Bethard, J. D. (2019) · Forensic Science International · doi:10.1016/j.forsciint.2019.109919 · Cited by 9 · see citing works
Mil, T. D., Meko, M. D., Belmecheri, S., et al. (2021) · Dendrochronologia · doi:10.1016/j.dendro.2021.125879 · Cited by 8 · see citing works
Li, Y., Zeng, X., Liu, X., et al. (2023) · Journal of Geophysical Research Biogeosciences · doi:10.1029/2023jg007455 · Cited by 7 · see citing works
Self-citationZeng, X., Ni, P., Liu, X., et al. (2024) · Journal of Forestry Research · doi:10.1007/s11676-024-01716-8 · Cited by 7 · see citing works
“The two-point calibration method was employed for stable isotope measurement (Evans et al 2016) using two standard cellulose materials (IAEA-CH 3 and Sigma-Aldrich-cellulose).”
The Inskie Springs: New Insights into Low-Radon Waters
Novikov, D. A., Kopylova, Y. G., Сухорукова, А. Ф., Вакуленко, Л. Г., Pyryaev, A. N., Maksimova, A. A., Деркачев, А. С., Faguet, A., Khvashchevskaya, A. A., Федорович, Д. Ф., et al. (2022) · Russian Geology and Geophysics · doi:10.2113/rgg20214337 · Cited by 6 · see citing works
Okubo, J., Kaufman, A. J., Warren, L. V., Evans, M. N., Marroquín, S., Varni, M. A., Misi, A., Bahniuk, A. M., & Xiao, S. (2022) · Journal of the Geological Society · doi:10.1144/jgs2021-091 · Cited by 5 · see citing works
“The uncertainty (1σ) of measurements based on multiple standard analyses run in the same analytical session was 0.05‰ for carbon and 0.1‰ for oxygen (Evans et al. 2016). δ34SCAS and δ34Spyrite analyses Sulfur isotopic compositions of carbonate-associated sulfate (CAS) from the Januária and Sete…”
BackgroundInfluentialSelf-citationNapier, T. J., Hendy, I. L., Hinnov, L. A., Brown, E. T., & Shevenell, A. (2018) · Quaternary Science Reviews · doi:10.1016/j.quascirev.2018.05.003 · Cited by 4 · see citing works
Xu, G., Liu, X., Sun, W., et al. (2018) · Journal of Arid Land · doi:10.1007/s40333-018-0070-6 · Cited by 4 · see citing works
Ge, W., Liu, X., Li, X., et al. (2022) · Forests · doi:10.3390/f13070986 · Cited by 3 · see citing works
Nurhati, I. S., Evans, M. N., Cahyarini, S. Y., D'Arrigo, R. D., Yoshimura, K., & Herho, S. (2024) · Paleoceanography and Paleoclimatology · doi:10.1029/2023pa004758 · Cited by 3 · see citing works
“We corrected the data using the algorithm and software described in Evans et al. (2016).”
MethodInfluentialSelf-citationRogers, J. A., Jackson, K., Pluckhahn, T. J., Garland, C. J., Thompson, V. D., & Herbert, G. S. (2026) · The Journal of Island and Coastal Archaeology · doi:10.1080/15564894.2025.2574637 · Cited by 2 · see citing works
“Measurement followed established procedures (Burman et al. 2005; Duhr and Hilkert 2004; Evans et al. 2016; R � ev � esz and Landwehr 2002; Sp € otl 2011; Sp € otl and Vennemann 2003).”
Novikov, D. A., Pyrayev, A. N., Maksimova, A. A., Федорович, Д. Ф., & Chernykh, A. V. (2023) · Georesursy · doi:10.18599/grs.2023.4.16 · Cited by 1 · see citing works
ISOTOPE COMPOSITION OF NATURAL WATERS IN THE VERKH-TULA SETTLEMENT (THE NOVOSIBIRSK REGION)
Pyrayev, A. N., Novikov, D. A., Федорович, Д. Ф., et al. (2021) · Interexpo GEO-Siberia · doi:10.33764/2618-981x-2021-2-1-207-217 · Cited by 1 · see citing works
Ge, W., Liu, X., Xiaoqin, L. I., et al. (2022) · SSRN Electronic Journal · doi:10.2139/ssrn.4056853
ISTOPIC COMPOSITION (H, O AND C) OF NATURAL WATERS OF THE NOVOSIBIRSK CITY AGLOMERATION
Novikov, D. A., Pyrayev, A. N., Федорович, Д. Ф., et al. (2021) · Interexpo GEO-Siberia · doi:10.33764/2618-981x-2021-2-1-149-159
CHARACTERISTICS OF STABLE ISOTOPES IN NATURAL WATERS OF THE VITIM RIVER BASIN
Pyrayev, A. N., Novikov, D. A., & Maksimova, A. A. (2021) · Interexpo GEO-Siberia · doi:10.33764/2618-981x-2021-2-1-218-226
Evans, M. N., Herho, S. H. S., Wollney, J. C., Carter, A. N., Gbenro, M. A., Tripp, G. R., D'Arrigo, R. D., & Cahyarini, S. Y. (2026) · Geophysical Research Letters · doi:10.1029/2025gl117991
“Procedures for developing the δ 18 O c analyses are described previously (Anchukaitis et al., 2008; Evans et al., 2016; Nurhati et al., 2024) (see also Supporting Information S1, Section 1).”
Self-citationSun, P., Huo, Y., Hu, B., Fan, C., Gao, J., & Li, Y. (2026) · SSRN Electronic Journal · doi:10.2139/ssrn.7571332
- Authors
- Evans, M. N., Selmer, K. J., Breeden, B. T., Lopatka, A. S., & Plummer, R. E.
- Affiliations
- Earth System Science Interdisciplinary Center (US)
- Agricultural Research Service (US)
- University of Utah (US)
- University of Maryland, College Park (US)
- Corresponding author
- Evans, M. N.
- Venue
- Geochemistry Geophysics Geosystems
- Publisher
- American Geophysical Union (AGU)
- Published
- 2016-09
- Volume
- 17
- Issue
- 9
- Pages
- 3580-3588
- ISSN
- 1525-2027 (ISSN-L)
- OpenAlex
- W2795974910
- Cited by
- 31
- References
- 22
- Citations by year
- 2026: 3 · 2025: 2 · 2024: 2 · 2023: 3 · 2022: 5 · 2021: 5 · 2020: 1 · 2019: 3 · 2018: 4 · 2017: 1 · 2016: 2
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- Locations
- Geochemistry Geophysics Geosystems (publishedVersion)
- Funders
- the editorial direction of J. Blichert-Toft (NSF/EAR0929983 and NSF/ATM0902794)
- Topics
- Physical Sciences › Earth and Planetary Sciences › Paleontology › Archaeology and ancient environmental studies
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Scientific Measurement and Uncertainty Evaluation
- Physical Sciences › Earth and Planetary Sciences › Atmospheric Science › Geology and Paleoclimatology Research
- Keywords
- carbonate stable isotopes, carbon-13, oxygen-18, amplitude correction, bias correction, gas chromatographic separation, paleoclimatology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-11.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-11.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review