Correction algorithm for online continuous flow δ 13 C and δ 18 O carbonate and cellulose stable isotope analyses

Open access
Evans, M. N., Selmer, K. J., Breeden, B. T., Lopatka, A. S., Plummer, R. E.·2016·Geochemistry Geophysics Geosystems

Abstract

We describe an algorithm to correct for scale compression, runtime drift, and amplitude effects in carbonate and cellulose oxygen and carbon isotopic analyses made on two online continuous flow isotope ratio mass spectrometry (CF-IRMS) systems using gas chromatographic (GC) separation. We validate the algorithm by correcting measurements of samples of known isotopic composition which are not used to estimate the corrections. For carbonate δ13C (δ18O) data, median precision of validation estimates for two reference materials and two calibrated working standards is 0.05‰ (0.07‰); median bias is 0.04‰ (0.02‰) over a range of 49.2‰ (24.3‰). For α-cellulose δ13C (δ18O) data, median precision of validation estimates for one reference material and five working standards is 0.11‰ (0.27‰); median bias is 0.13‰ (−0.10‰) over a range of 16.1‰ (19.1‰). These results are within the 5th–95th percentile range of subsequent routine runtime validation exercises in which one working standard is used to calibrate the other. Analysis of the relative importance of correction steps suggests that drift and scale-compression corrections are most reliable and valuable. If validation precisions are not already small, routine cross-validated precision estimates are improved by up to 50% (80%). The results suggest that correction for systematic error may enable these particular CF-IRMS systems to produce δ13C and δ18O carbonate and cellulose isotopic analyses with higher validated precision, accuracy, and throughput than is typically reported for these systems. The correction scheme may be used in support of replication-intensive research projects in paleoclimatology and other data-intensive applications within the geosciences.

Integrity signals

Registry

Crossref · 10.1002/2016gc006469

Cites

21 of 21 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

0 of 18 cited sources checked · 0 supported · 0 flagged

  • Findingown data

    For the LSVEC reference material, the precision achieved is better than reported in Stichler [1995], which was 0.15 permil for d13C and 0.25 permil for d18O.

    “For the LSVEC reference material, the precision achieved is better than reported in Stichler [1995], which was 0.15 and 0.25& for d C and d O , respectively.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Stichler, 1995

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
  • Findingown data

    For carbonate d18O data, median precision of validation estimates for two reference materials and two calibrated working standards is 0.07 permil.

    “For carbonate d 13 C (d 18 O) data, median precision of validation estimates for two reference materials and two calibrated working standards is 0.05& (0.07&);”
    • Also stated in: § 5. Discussion · plain
      “For the carbonate materials we analyzed, we found that the median validation precisions across the carbonates analyzed are 0.05 and 0.07& for d C and d O , respectively (Table 1).”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
  • Findingown data

    These biases are small relative to the calibrated ranges in δ13C and δ18O across all materials analyzed (16.1 and 19.1‰, respectively).

    “but these biases are again small relative to the calibrated ranges in d C and d O across all materials analyzed (16.1 and 19.1&, respectively; Table 2 and Figures 1c and 1d)”
    Quote in the paperSubstantial claim
  • Findingown data

    These validation results are within the 5th-95th percentile range of subsequent routine runtime validation exercises in which one working standard is used to calibrate the other.

    “These results are within the 5th-95th percentile range of subsequent routine runtime validation exercises in which one working standard is used to calibrate the other.”
    Quote in the paperSubstantial claim
  • Findingown data

    For carbonate d13C data, median bias is 0.04 permil over a range of 49.2 permil.

    “median bias is 0.04& (0.02&) over a range of 49.2& (24.3&).”
    • Also stated in: § 5. Discussion · plain
      “However, the median biases (0.04 and 0.02& for d C and d O , respectively; Table 1) are small relative to the known range across in d C and d O across working standards (49.2 and 24.3&, respectively; Table 1 and Figures 1a and 1b).”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
  • May combine two claims. 10
  • Classified as probably not a substantial claim. 11
  • May combine two claims.

    Findingown data

    For the α-cellulose materials analyzed, median validation precisions across five α-cellulose working standards and one reference material are 0.11‰ for δ13C and 0.27‰ for δ18O.

    “For the a-cellulose materials we analyzed, we found that the median validation precisions across five acellulose working standards and one reference material are 0.11 and 0.27& for d C and d O , respectively (Table 2).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Median bias for α-cellulose validation is 0.13‰ for δ13C and -0.10‰ for δ18O.

    “Median bias (0.13 and 20.10& for d C and d O , respectively; Table 2)”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    For the carbonate materials analyzed, the median validation precisions across the carbonates analyzed are 0.05 permil for d13C and 0.07 permil for d18O.

    “For the carbonate materials we analyzed, we found that the median validation precisions across the carbonates analyzed are 0.05 and 0.07& for d C and d O , respectively (Table 1).”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
  • May combine two claims.

    Findingown data

    The median biases (0.04 permil for d13C and 0.02 permil for d18O) are small relative to the known range across working standards (49.2 permil for d13C and 24.3 permil for d18O).

    “However, the median biases (0.04 and 0.02& for d C and d O , respectively; Table 1) are small relative to the known range across in d C and d O across working standards (49.2 and 24.3&, respectively; Table 1 and Figures 1a and 1b).”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
  • Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).

    Interpretationreasoning

    The correction scheme may be used in support of replication-intensive research projects in paleoclimatology and other data-intensive applications within the geosciences.

    “The correction scheme may be used in support of replication-intensive research projects in paleoclimatology and other data-intensive applications within the geosciences.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Study design and conductown data

    No measurement blank correction was applied to cellulose analyses because routine analyses on empty silver sample capsules give signals of less than 0.1 nA.

    “No measurement blank correction has been applied, as routine analyses on empty silver sample capsules give signals of less than 0.1 nA.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductreasoning

    Because materials of known and substantially different isotopic composition are few, the algorithm is applied in validation to samples of materials that were also used to correct the reported data (IAEA-CO-1, LSVEC, and AKC).

    “Note that because materials of known and substantially different isotopic composition are few, the algorithm is applied in validation to samples of materials that were also used to correct the reported data (i.e., IAEA-CO-1, LSVEC, and AKC).”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    Scale compression and bias in cellulose pyrolysis measurements are due to the presence of catalytic glassy carbon and deposition of excess sample carbon in the pyrolysis reactor, and thermal conversion into both CO and CO2 at this temperature.

    “This is because of the presence of catalytic glassy carbon and deposition of excess sample carbon in the pyrolysis reactor [Young et al., 2011], and thermal conversion into both CO and CO 2 at this temperature [Leuenberger and Filot, 2007].”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites Young et al., 2011 source
    • cites Leuenberger and Filot, 2007 source
  • Classified as probably not a substantial claim.

    Study design and conductreasoning

    Analysis of samples of known isotopic composition which were not themselves used to estimate the corrections is required for assessment of the results.

    “Analysis of samples of known isotopic composition which were not themselves used to estimate the corrections is required for assessment of the results.”
    Quote in the paperSubstantial claim
  • Classified as probably not a substantial claim.

    Premisereasoning

    Biases associated with measurement conditions common to both working standards and unknown samples, if present, must be detected and removed.

    “However, biases associated with measurement conditions common to both working standards and unknown samples, if present, must be detected and removed.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationown data

    Median biases are small relative to the distribution of routine runtime cross-validation precisions, and hence may be negligibly small in practical measurement and application.

    “Median biases are also small relative to the distribution of routine runtime cross-validation precisions (Figure 2), and hence may be negligibly small in practical measurement and application.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
  • Classified as probably not a substantial claim. May combine two claims.

    Study design and conductreasoning

    In each experiment, the sample data were not used to develop the corrections; they are out-of-sample tests of the validity of the corrections.

    “In each experiment, the ''samples'' were of materials of a priori known isotopic composition, but the ''sample'' data were not used to develop the corrections; in other words, these are ''out-of-sample'' tests of the validity of the corrections.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
  • Probably not a substantial claim: may not pass S2 (checkable).

    Premisecitation

    Mean and variance-calibrated, high-throughput stable isotope analyses are increasingly required in paleoclimatology.

    “Mean and variance-calibrated, high-throughput stable isotope analyses are increasingly required in paleoclimatology.”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Cobb et al., 2003 source
    • cites McCarroll and Loader, 2004 source
    • cites Koutavas et al., 2006 source
  • Classified as probably not a substantial claim. May combine two claims.

    Premisecitation

    CF-IRMS is a rapid, economical, but relatively low-precision approach to high-throughput stable isotopic analyses of carbonate, cellulose, and similar materials.

    “Continuous flow isotope ratio mass spectrometry (CF-IRMS) is a rapid, economical, but relatively low-precision approach to high-throughput stable isotopic analyses of these materials [Brenna et al., 1997;Werner et al., 1996;Spotl and Vennemann, 2003;Spotl, 2011].”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites Brenna et al., 1997 source
    • cites Werner et al., 1996 source
    • cites Spotl and Vennemann, 2003 source
  • Probably not a substantial claim: may not pass S2 (checkable).

    Study design and conductreasoning

    The more complex the suite of corrections, the more important that their efficacy is evaluated using validation tests, because of the potential for artificial skill from overfitting of the calibration data.

    “However, the more complex the suite of corrections, the more important that their efficacy is evaluated using validation tests, because of the potential for artificial skill from overfitting of the calibration data.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Premisecitation

    Isotopic analyses should be corrected for scale differences arising from using intermediate working standards to correct to international reference scales.

    “Isotopic analyses should also be corrected for scale differences arising from using intermediate working standards to correct to international reference scales [Coplen et al., 2006].”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Coplen et al., 2006 source
  • Classified as probably not a substantial claim.

    Study design and conductnone given

    By construction, the means of the final corrected calibration data are the known isotopic values of the working standards on the international reference scales.

    “By construction, the means of the final corrected calibration data are the known isotopic values of the working standards on the international reference scales.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.
  • Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).

    Interpretationreasoning

    The algorithm merely contributes to the comprehensive set of quality controls and standard operating protocols that produce long-term stability, precision, accuracy, and traceability of corrected data.

    “it merely contributes to the comprehensive set of quality controls and standard operating protocols that produce long-term stability, precision, accuracy, and traceability of corrected data.”
    Quote in the paperNot a substantial claim
18 of 18
  • The paper relies on this without support.

    The precision on validation data should be approximately equal to the precision of the corrected calibration data.

    “The precision on validation data should be approximately equal to the precision of the corrected calibration data.”
  • The paper relies on this without support.

    The known d13C and d18O values for the working standard a-celluloses SAC and AKC with respect to V-PDB and V-SMOW are accurate.

    “with known d 13 C and d 18 O values for SAC and AKC with respect to the V-PDB and V-SMOW reference standards previously determined and as listed in Table 2.”
  • The paper relies on this without support.

    The working standards used for correction have chemical composition similar to the unknown samples.

    “Because we use working standards of known isotopic composition and similar chemical composition as the basis for these corrections,”
3 of 3

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

  • Cellulose significance count disagrees with Table 2

    Minor

    The discussion says five of twelve bias comparisons are statistically distinguishable from zero, but Table 2 marks six comparisons with asterisks.

1 of 1

Cites

Flagged first, then by citations

1 of the 22 references couldn’t be matched to a known paper.

Research works printed but not linked: 4

  1. 2008, Handbook of Stable Isotope Analytical Techniques, 2, 1398
  2. IsoPrime Ltd., 2009, IsoPrime User Guide 2009A
  3. Stichler, 1995, Reference and Intercomparison Materials for Stable Isotopes of Light Elements, 67
  4. Thermo Electron Corporation, 2005, DELTA V Plus Operating Manual, Revision A - 118 4270

Cited by

Sorted by citations

Authors
Evans, M. N., Selmer, K. J., Breeden, B. T., Lopatka, A. S., & Plummer, R. E.
Affiliations
  • Earth System Science Interdisciplinary Center (US)
  • Agricultural Research Service (US)
  • University of Utah (US)
  • University of Maryland, College Park (US)
Corresponding author
Evans, M. N.
Venue
Geochemistry Geophysics Geosystems
Published
2016-09
Volume
17
Issue
9
Pages
3580-3588
ISSN
1525-2027 (ISSN-L)
OpenAlex
W2795974910
Cited by
31
References
22
Citations by year
2026: 3 · 2025: 2 · 2024: 2 · 2023: 3 · 2022: 5 · 2021: 5 · 2020: 1 · 2019: 3 · 2018: 4 · 2017: 1 · 2016: 2
Type
journal article
Language
en
Open access
Locations
Geochemistry Geophysics Geosystems (publishedVersion)
Funders
the editorial direction of J. Blichert-Toft (NSF/EAR0929983 and NSF/ATM0902794)
Topics
  • Physical Sciences › Earth and Planetary Sciences › Paleontology › Archaeology and ancient environmental studies
  • Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Scientific Measurement and Uncertainty Evaluation
  • Physical Sciences › Earth and Planetary Sciences › Atmospheric Science › Geology and Paleoclimatology Research
Keywords
carbonate stable isotopes, carbon-13, oxygen-18, amplitude correction, bias correction, gas chromatographic separation, paleoclimatology

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-11.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-11.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review