Wireless, Skin-Interfaced Devices for Pediatric Critical Care: Application to Continuous, Noninvasive Blood Pressure Monitoring
Open accessAbstract
Indwelling arterial lines, the clinical gold standard for continuous blood pressure (BP) monitoring in the pediatric intensive care unit (PICU), have significant drawbacks due to their invasive nature, ischemic risk, and impediment to natural body movement. A noninvasive, wireless, and accurate alternative would greatly improve the quality of patient care. Recently introduced classes of wireless, skin-interfaced devices offer capabilities in continuous, precise monitoring of physiologic waveforms and vital signs in pediatric and neonatal patients, but have not yet been employed for continuous tracking of systolic and diastolic BP-critical for guiding clinical decision-making in the PICU. The results presented here focus on materials and mechanics that optimize the system-level properties of these devices to enhance their reliable use in this context, achieving full compatibility with the range of body sizes, skin types, and sterilization schemes typically encountered in the PICU. Systematic analysis of the data from these devices on 23 pediatric patients, yields derived, noninvasive BP values that can be quantitatively validated against direct recordings from arterial lines. The results from this diverse cohort, including those under pharmacological protocols, suggest that wireless, skin-interfaced devices can, in certain circumstances of practical utility, accurately and continuously monitor BP in the PICU patient population.
Overview
CoChat Summary
TL;DR
A pair of wireless, skin-interfaced devices—chest ECG and limb PPG—was optimized for pediatric critical care and used to estimate continuous systolic and diastolic blood pressure from pulse arrival time and heart rate, calibrated against indwelling arterial lines in 23 PICU patients over 82 hours. The multivariate PAT/HR model showed the strongest agreement with arterial-line BP, meeting adult FDA criteria for DBP and nearly meeting them for SBP.
Key findings
- Across 23 PICU patients and 290,108 paired SBP/DBP comparisons, the multivariate PAT/HR model had the strongest correlation with arterial-line BP: r = 0.85 for SBP and r = 0.77 for DBP. Bland-Altman mean ± SD differences were 1.7 ± 8.7 mmHg for SBP and 1.8 ± 6.2 mmHg for DBP.
- DBP estimates met adult FDA noninvasive BP criteria (mean difference ≤5 mmHg and SD ≤8 mmHg) across the four models. For SBP, mean differences were within 5 mmHg, but the smallest SD was 8.7 mmHg, slightly exceeding the 8 mmHg criterion.
- Optimized limb-device encapsulation reduced strain magnitude by ~33% and shear strain by ~96% at x = 18 mm versus the unoptimized design; effective modulus decreased by ~83% versus the unoptimized design with the same materials and by ~96% versus the unoptimized design with A-20 as both shell and filler. Natural frequency and damping ratio decreased by ~49% and ~38%, respectively.
- The devices captured hemodynamic changes after lorazepam, methadone, hydromorphone, and dexamethasone, with trends consistent with arterial-line measurements; 82 cumulative hours of recordings occurred without adverse events.
Methods
- Prospective, single-center, observational pilot study at Ann & Robert H. Lurie Children’s Hospital of Chicago. Eligible patients were under 3 years of age, admitted to the PICU, and had an indwelling arterial line for clinical BP monitoring; n = 23 (13 male, 10 female; mean age 53 ± 47 weeks; mean weight 8.2 ± 4.1 kg), with 82 cumulative hours of recordings.
- Time-synchronized chest ECG and limb PPG devices derived PAT (ECG R-peak to PPG onset) and HR (successive R-peak intervals). For each patient, an initial 10-minute window of PAT and/or HR was calibrated to arterial-line SBP and DBP using four models: multivariate linear PAT/HR, univariate linear PAT, univariate linear HR, and inverse-square PAT; no recalibration was performed. Agreement was assessed with Bland-Altman analysis, Pearson correlation, and paired t-tests.
- Device mechanics and materials were characterized using digital image correlation, high-speed tracking, finite element analysis, shelf aging, autoclave sterilization, UVC exposure with FTIR, water contact angle, and peel-force measurements.
Limitations
- The paper states that the approach required an initial calibration period using the indwelling arterial line, which is invasive; it suggests future use of a noninvasive cuff reference. It also states that larger cohorts and different ICU populations—such as neonatal, neurocritical, and cardiac critical care—are needed, and that interpatient variability requires further investigation.
- The paper notes that the SBP SD of 8.7 mmHg with the multivariate PAT/HR model slightly exceeds the FDA 8 mmHg criterion, and that arterial-line reference measurements can be subject to over- or underestimation due to improper waveform dampening. It also notes that possible UVC-induced near-surface chemical changes in polyorganosiloxanes require more surface-sensitive techniques and that penetration depth remains debated.
- Evident from the text: this was a single-center observational pilot with n = 23, mostly patients with respiratory failure, restricted to children under 3 years old, with no external validation, randomization, or recalibration after the initial 10-minute period.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 Valid · 1 Not valid · 14 Not checked · 35 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Here, the obtained model equations are: SBP (mmHg) = [-0.02 ms -1 (PAT (ms)) + 0.20 bpm -1 (HR (bpm)) + 71.79 mmHg], and DBP (mmHg) = [-0.04 ms -1 (PAT (ms)) + 0.24 bpm -1 (HR (bpm)) + 38.42 mmHg].”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Performance of wireless, skin-interfaced devices, calibrated using [a(PAT)+b(HR)+c], in capture of transient blood pressure and other hemodynamic changes after pharmacologic interventions.”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“These devices continuously track short- and long-term BP trends, including during periods of pharmacological interventions.”
- Also stated in: § WILEY-VCH · p. 45 · plain
“Performance of wireless, skin-interfaced devices, calibrated using [a(PAT)+b(HR)+c], in capture of transient blood pressure and other hemodynamic changes after pharmacologic interventions.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
- Also stated in: § WILEY-VCH · p. 45 · plain
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“As FEA simulations indicate (Figure 2K, Figure S9), soft materials are important factors in determining the effective mechanical properties of the overall system.”
hedgedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“In particular, this configuration reduces the strain magnitude (Figure 2C-D; Supplementary Video 1) and shear strain (Figure 2E-F; Supplementary Video 2) distributions upon uniaxial stretching deformations of the serpentine wiring that connects the LEDs to the photodetector on the main body of the device.”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
21 more claims
Create a free account to see all 50 claims and to verify the 4 references still open.
- It may state two separate claims as one. 16
- Probably not a substantial claim. 11
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated major: it would touch a result — a value a result rests on, reported and recomputed p-values on opposite sides of the significance level, two passages that conflict with no sign which is right, or a cited work retracted before this paper was published.
In the paper
§ WILEY-VCH · p. 16
“The Bland-Altman results in our study (Figure 5A-D, Figure 6A-D) indicate that the devices and analysis approaches (Equation 2-5) meet these adult FDA specifications for measurements of DBP.”
§ 2.3. Evaluation of BP Prediction Model Performance · p. 15
“For DBP prediction (Figure 6A-D), the mean and SD of the difference in BP is 1.8 ± 6.2 mmHg, 0.7 ± 9.0 mmHg, -1.1 ± 9.0 mmHg, and 0.5 ± 9.7 mmHg, respectively.”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ WILEY-VCH · p. 63
“| 9 | -1.4 | 6.1 |”
§ WILEY-VCH · p. 63
“| 9 | 5.6 | 6.7 |”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ WILEY-VCH · p. 59
“Statistical annotation key: **** p ≤ 0.0001, *** p ≤ 0.001, ** p ≤ 0.01, * p ≤ 0.05, ns p ≤ 0.05.”
§ WILEY-VCH · p. 26
“Significance was applied to p <0.05.”
Citations
Cites
Surviving Sepsis Campaign: International guidelines for management of sepsis and septic shock: 2016.
Rhodes, A., Evans, L. E., Alhazzani, W., Levy, M. M., Antonelli, M., Ferrer, R., Kumar, A., Sevransky, J. E., Sprung, C. L., Nunnally, M. E., et al. (2017) · Intensive Care Medicine · doi:10.1007/s00134-017-4683-6 · Cited by 6,621
Ncorr: Open-Source 2D Digital Image Correlation Matlab Software
Blaber, J., Adair, B., & Antoniou, A. (2015) · Experimental Mechanics · doi:10.1007/s11340-015-0009-1 · Cited by 2,260
Intensive care management of coronavirus disease 2019 (COVID-19): challenges and recommendations
Phua, J., Weng, L., Ling, L., Egi, M., Lim, C.-M., Divatia, J. V., Shrestha, B. R., Arabi, Y. M., Ng, J., Gomersall, C. D., et al. (2020) · The Lancet Respiratory Medicine · doi:10.1016/s2213-2600(20)30161-2 · Cited by 1,509
Mechanical characterization of bulk Sylgard 184 for microfluidics and microengineering
Johnston, I. D., McCluskey, D. K., Tan, C. K. L., & Tracey, M. C. (2014) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/24/3/035017 · Cited by 1,495
Pandharipande, P. P., Pun, B. T., Herr, D. L., Maze, M., Girard, T. D., Miller, R. R., Shintani, A. K., Thompson, J. L., Jackson, J. C., Deppen, S. A., et al. (2007) · JAMA · doi:10.1001/jama.298.22.2644 · Cited by 1,451
Characterization of Polydimethylsiloxane (PDMS) Properties for Biomedical Micro/Nanosystems
Mata, A., Fleischman, A. J., & Roy, S. (2005) · Biomedical Microdevices · doi:10.1007/s10544-005-6070-2 · Cited by 1,318
The influence of heart rate on augmentation index and central arterial pressure in humans
Wilkinson, I. B., MacCallum, H., Flint, L., Cockcroft, J. R., Newby, D. E., & Webb, D. J. (2000) · The Journal of Physiology · doi:10.1111/j.1469-7793.2000.t01-1-00263.x · Cited by 1,091
Hydrogel Adhesion: A Supramolecular Synergy of Chemistry, Topology, and Mechanics
Yang, J., Bai, R., Chen, B., & Suo, Z. (2020) · Advanced Functional Materials · doi:10.1002/adfm.201901693 · Cited by 1,033
Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., et al. (2019) · Science · doi:10.1126/science.aau0780 · Cited by 807
Self-citationEfimenko, K., Wallace, W. E., & Genzer, J. (2002) · Journal of Colloid and Interface Science · doi:10.1006/jcis.2002.8594 · Cited by 757
Research works printed but not linked: 4
- Cattivelli F. S., 2009, IEEE 2009 Sixth Int. Workshop on BSN, 114
- Wang R., 2014, IEEE 2014 12th Int. Conf. on Signal Processing (ICSP), 115
- Alinejad S., 2015, EXCLI J., 14, 577
- Khemani R. G., 2009, Cochrane Database Syst. Rev., 3, CD001000
Cited by
Thin, soft, wearable system for continuous wireless monitoring of artery blood pressure
Li, J., Jia, H., Zhou, J., Huang, X., Xu, L., Jia, S., Gao, Z., Yao, K., Li, D., Zhang, B., et al. (2023) · Nature Communications · doi:10.1038/s41467-023-40763-3 · Cited by 293
“14).”
Self-citationUltra‐Thin Flexible Encapsulating Materials for Soft Bio‐Integrated Electronics
Sang, M., Kim, K., Shin, J., & Yu, K. J. (2022) · Advanced Science · doi:10.1002/advs.202202980 · Cited by 175
Sang, M., Kang, K., Zhang, Y., Zhang, H., Kim, K., Cho, M., Shin, J., Hong, J., Kim, T., Lee, S. K., et al. (2021) · Advanced Materials · doi:10.1002/adma.202105865 · Cited by 153
Self-citationSoft shape-programmable surfaces by fast electromagnetic actuation of liquid metal networks
Ni, X., Luan, H., Kim, J.-T., Rogge, S. I., Bai, Y., Kwak, J. W., Liu, S., Yang, D. S., Li, S., Li, S., et al. (2022) · Nature Communications · doi:10.1038/s41467-022-31092-y · Cited by 110
Self-citationMin, S., An, J., Lee, J. H., Kim, J. H., Joe, D. J., Eom, S. H., Yoo, C. D., Ahn, H., Hwang, J.-Y., Xu, S. H., et al. (2025) · Nature Reviews Cardiology · doi:10.1038/s41569-025-01127-0 · Cited by 157
“Depending on the measurement location, the time differences of pulse waves are classified into two categories: pulse arrival time 176–178 and pulse transit time 178–182 .”
Self-citation
Metadata
- Authors
- Liu, C., Kim, J., Kwak, S. S., Hourlier‐Fargette, A., Avila, R., Vogl, J., Tzavelis, A., Chung, H. U., Lee, J. Y., Kim, D. H., Ryu, D., Fields, K. B., Ciatti, J. L., Li, S., Irie, M., Bradley, A., Shukla, A., Chavez, J., Dunne, E. C., Kim, S. S., Kim, J., Park, J. B., Jo, H. H., Kim, J., Johnson, M. C., Kwak, J. W., Madhvapathy, S. R., Xu, S., Rand, C. M., Marsillio, L. E., Hong, S. J., Huang, Y., Weese‐Mayer, D. E., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- Querrey Simpson Institute for Bioelectronics Northwestern University Evanston IL 60208 USA
- Lurie Children's Hospital (US)
- Sibel (United States) (US)
- Corresponding author
- Huang, Yonggang
- Weese‐Mayer, Debra E.
- Rogers, John A.
- ORCID
- Liu, Claire
- Kim, Jin‐Tae
- Kwak, Sung Soo
- Avila, Raudel
- Vogl, Jamie
- Tzavelis, Andreas
- Chung, Ha Uk
- Lee, Jong Yoon
- Kim, Dong Hyun
- Ciatti, Joanna L.
- Li, Shupeng
- Chavez, Jairo
- Dunne, Emma C.
- Kim, Jungwoo
- Park, Jun Bin
- Jo, Han Heul
- Kim, Joohee
- Johnson, Michael C.
- Kwak, Jean Won
- Madhvapathy, Surabhi R.
- Xu, Shuai
- Rand, Casey M.
- Marsillio, Lauren E.
- Hong, Sue J.
- Huang, Yonggang
- Weese‐Mayer, Debra E.
- Rogers, John A.
- Venue
- Advanced Healthcare Materials
- Publisher
- Wiley
- Published
- 2021-05-03
- Volume
- 10
- Issue
- 17
- Pages
- e2100383-e2100383
- ISSN
- 2192-2640 (ISSN-L)
- 2192-2659
- OpenAlex
- W3158794625
- Cited by
- 57
- References
- 75
- Citations by year
- 2026: 7 · 2025: 8 · 2024: 10 · 2023: 14 · 2022: 14 · 2021: 4
- FWCI
- 3.19
- Type
- journal article
- Language
- en
- Open access
- green
- hal.science
- Licence
- Locations
- Advanced Healthcare Materials (publishedVersion)
- PubMed (publishedVersion)
- HAL (submittedVersion, other-oa)
- Funders
- Gerber Foundation
- National Science Foundation (DGE‐1842165)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- non-invasive blood pressure monitoring, pediatric intensive care, neonatal patients, continuous blood pressure measurement, pharmacological interventions
- MeSH terms
- Blood Pressure, Child, Critical Care, Humans, Infant, Newborn, Monitoring, Physiologic, Skin, Vital Signs
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-29, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-29 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Wireless, Skin-Interfaced Devices for Pediatric Critical Care: Application to Continuous, Noninvasive Blood Pressure Monitoring. https://app.cochat.ai/papers/10.1002/adhm.202100383
- The paper
- Liu, C., Kim, J., Kwak, S. S., Hourlier‐Fargette, A., Avila, R., Vogl, J., Tzavelis, A., Chung, H. U., Lee, J. Y., Kim, D. H., Ryu, D., Fields, K. B., Ciatti, J. L., Li, S., Irie, M., Bradley, A., Shukla, A., Chavez, J., Dunne, E. C., . . . Rogers, J. A. (2021). Wireless, Skin-Interfaced Devices for Pediatric Critical Care: Application to Continuous, Noninvasive Blood Pressure Monitoring. Advanced Healthcare Materials. https://doi.org/10.1002/adhm.202100383
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.