Wireless, Skin-Interfaced Devices for Pediatric Critical Care: Application to Continuous, Noninvasive Blood Pressure Monitoring
Open accessThe DOI matches its Crossref record; 72 of the 75 references it cites were found in a registry or an index; none of the 32 sources its claims cite is checked for support yet.
Abstract
Indwelling arterial lines, the clinical gold standard for continuous blood pressure (BP) monitoring in the pediatric intensive care unit (PICU), have significant drawbacks due to their invasive nature, ischemic risk, and impediment to natural body movement. A noninvasive, wireless, and accurate alternative would greatly improve the quality of patient care. Recently introduced classes of wireless, skin-interfaced devices offer capabilities in continuous, precise monitoring of physiologic waveforms and vital signs in pediatric and neonatal patients, but have not yet been employed for continuous tracking of systolic and diastolic BP-critical for guiding clinical decision-making in the PICU. The results presented here focus on materials and mechanics that optimize the system-level properties of these devices to enhance their reliable use in this context, achieving full compatibility with the range of body sizes, skin types, and sterilization schemes typically encountered in the PICU. Systematic analysis of the data from these devices on 23 pediatric patients, yields derived, noninvasive BP values that can be quantitatively validated against direct recordings from arterial lines. The results from this diverse cohort, including those under pharmacological protocols, suggest that wireless, skin-interfaced devices can, in certain circumstances of practical utility, accurately and continuously monitor BP in the PICU patient population.
Overview
CoChat Summary
TL;DR
A pair of wireless, skin-interfaced devices—chest ECG and limb PPG—was optimized for pediatric critical care and used to estimate continuous systolic and diastolic blood pressure from pulse arrival time and heart rate, calibrated against indwelling arterial lines in 23 PICU patients over 82 hours. The multivariate PAT/HR model showed the strongest agreement with arterial-line BP, meeting adult FDA criteria for DBP and nearly meeting them for SBP.
Key findings
- Across 23 PICU patients and 290,108 paired SBP/DBP comparisons, the multivariate PAT/HR model had the strongest correlation with arterial-line BP: r = 0.85 for SBP and r = 0.77 for DBP. Bland-Altman mean ± SD differences were 1.7 ± 8.7 mmHg for SBP and 1.8 ± 6.2 mmHg for DBP.
- DBP estimates met adult FDA noninvasive BP criteria (mean difference ≤5 mmHg and SD ≤8 mmHg) across the four models. For SBP, mean differences were within 5 mmHg, but the smallest SD was 8.7 mmHg, slightly exceeding the 8 mmHg criterion.
- Optimized limb-device encapsulation reduced strain magnitude by ~33% and shear strain by ~96% at x = 18 mm versus the unoptimized design; effective modulus decreased by ~83% versus the unoptimized design with the same materials and by ~96% versus the unoptimized design with A-20 as both shell and filler. Natural frequency and damping ratio decreased by ~49% and ~38%, respectively.
- The devices captured hemodynamic changes after lorazepam, methadone, hydromorphone, and dexamethasone, with trends consistent with arterial-line measurements; 82 cumulative hours of recordings occurred without adverse events.
Methods
- Prospective, single-center, observational pilot study at Ann & Robert H. Lurie Children’s Hospital of Chicago. Eligible patients were under 3 years of age, admitted to the PICU, and had an indwelling arterial line for clinical BP monitoring; n = 23 (13 male, 10 female; mean age 53 ± 47 weeks; mean weight 8.2 ± 4.1 kg), with 82 cumulative hours of recordings.
- Time-synchronized chest ECG and limb PPG devices derived PAT (ECG R-peak to PPG onset) and HR (successive R-peak intervals). For each patient, an initial 10-minute window of PAT and/or HR was calibrated to arterial-line SBP and DBP using four models: multivariate linear PAT/HR, univariate linear PAT, univariate linear HR, and inverse-square PAT; no recalibration was performed. Agreement was assessed with Bland-Altman analysis, Pearson correlation, and paired t-tests.
- Device mechanics and materials were characterized using digital image correlation, high-speed tracking, finite element analysis, shelf aging, autoclave sterilization, UVC exposure with FTIR, water contact angle, and peel-force measurements.
Limitations
- The paper states that the approach required an initial calibration period using the indwelling arterial line, which is invasive; it suggests future use of a noninvasive cuff reference. It also states that larger cohorts and different ICU populations—such as neonatal, neurocritical, and cardiac critical care—are needed, and that interpatient variability requires further investigation.
- The paper notes that the SBP SD of 8.7 mmHg with the multivariate PAT/HR model slightly exceeds the FDA 8 mmHg criterion, and that arterial-line reference measurements can be subject to over- or underestimation due to improper waveform dampening. It also notes that possible UVC-induced near-surface chemical changes in polyorganosiloxanes require more surface-sensitive techniques and that penetration depth remains debated.
- Evident from the text: this was a single-center observational pilot with n = 23, mostly patients with respiratory failure, restricted to children under 3 years old, with no external validation, randomization, or recalibration after the initial 10-minute period.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 Valid · 1 Not valid · 14 Not checked · 35 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Here, the obtained model equations are: SBP (mmHg) = [-0.02 ms -1 (PAT (ms)) + 0.20 bpm -1 (HR (bpm)) + 71.79 mmHg], and DBP (mmHg) = [-0.04 ms -1 (PAT (ms)) + 0.24 bpm -1 (HR (bpm)) + 38.42 mmHg].”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Performance of wireless, skin-interfaced devices, calibrated using [a(PAT)+b(HR)+c], in capture of transient blood pressure and other hemodynamic changes after pharmacologic interventions.”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“These devices continuously track short- and long-term BP trends, including during periods of pharmacological interventions.”
- Also stated in: § WILEY-VCH · p. 45 · plain
“Performance of wireless, skin-interfaced devices, calibrated using [a(PAT)+b(HR)+c], in capture of transient blood pressure and other hemodynamic changes after pharmacologic interventions.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
- Also stated in: § WILEY-VCH · p. 45 · plain
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“As FEA simulations indicate (Figure 2K, Figure S9), soft materials are important factors in determining the effective mechanical properties of the overall system.”
hedgedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“In particular, this configuration reduces the strain magnitude (Figure 2C-D; Supplementary Video 1) and shear strain (Figure 2E-F; Supplementary Video 2) distributions upon uniaxial stretching deformations of the serpentine wiring that connects the LEDs to the photodetector on the main body of the device.”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
21 more claims
Create a free account to see all 50 claims and to verify the 4 references still open.
- It may state two separate claims as one. 16
- Probably not a substantial claim. 11
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated major: it would touch a result — a value a result rests on, reported and recomputed p-values on opposite sides of the significance level, two passages that conflict with no sign which is right, or a cited work retracted before this paper was published.
In the paper
§ WILEY-VCH · p. 16
“The Bland-Altman results in our study (Figure 5A-D, Figure 6A-D) indicate that the devices and analysis approaches (Equation 2-5) meet these adult FDA specifications for measurements of DBP.”
§ 2.3. Evaluation of BP Prediction Model Performance · p. 15
“For DBP prediction (Figure 6A-D), the mean and SD of the difference in BP is 1.8 ± 6.2 mmHg, 0.7 ± 9.0 mmHg, -1.1 ± 9.0 mmHg, and 0.5 ± 9.7 mmHg, respectively.”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ WILEY-VCH · p. 63
“| 9 | -1.4 | 6.1 |”
§ WILEY-VCH · p. 63
“| 9 | 5.6 | 6.7 |”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ WILEY-VCH · p. 59
“Statistical annotation key: **** p ≤ 0.0001, *** p ≤ 0.001, ** p ≤ 0.01, * p ≤ 0.05, ns p ≤ 0.05.”
§ WILEY-VCH · p. 26
“Significance was applied to p <0.05.”
Citations
Cites
Surviving Sepsis Campaign: International guidelines for management of sepsis and septic shock: 2016.
Rhodes, A., Evans, L. E., Alhazzani, W., Levy, M. M., Antonelli, M., Ferrer, R., Kumar, A., Sevransky, J. E., Sprung, C. L., Nunnally, M. E., et al. (2017) · Intensive Care Medicine · doi:10.1007/s00134-017-4683-6 · Cited by 6,621
Ncorr: Open-Source 2D Digital Image Correlation Matlab Software
Blaber, J., Adair, B., & Antoniou, A. (2015) · Experimental Mechanics · doi:10.1007/s11340-015-0009-1 · Cited by 2,260
Intensive care management of coronavirus disease 2019 (COVID-19): challenges and recommendations
Phua, J., Weng, L., Ling, L., Egi, M., Lim, C.-M., Divatia, J. V., Shrestha, B. R., Arabi, Y. M., Ng, J., Gomersall, C. D., et al. (2020) · The Lancet Respiratory Medicine · doi:10.1016/s2213-2600(20)30161-2 · Cited by 1,509
Mechanical characterization of bulk Sylgard 184 for microfluidics and microengineering
Johnston, I. D., McCluskey, D. K., Tan, C. K. L., & Tracey, M. C. (2014) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/24/3/035017 · Cited by 1,495
Pandharipande, P. P., Pun, B. T., Herr, D. L., Maze, M., Girard, T. D., Miller, R. R., Shintani, A. K., Thompson, J. L., Jackson, J. C., Deppen, S. A., et al. (2007) · JAMA · doi:10.1001/jama.298.22.2644 · Cited by 1,451
Characterization of Polydimethylsiloxane (PDMS) Properties for Biomedical Micro/Nanosystems
Mata, A., Fleischman, A. J., & Roy, S. (2005) · Biomedical Microdevices · doi:10.1007/s10544-005-6070-2 · Cited by 1,318
The influence of heart rate on augmentation index and central arterial pressure in humans
Wilkinson, I. B., MacCallum, H., Flint, L., Cockcroft, J. R., Newby, D. E., & Webb, D. J. (2000) · The Journal of Physiology · doi:10.1111/j.1469-7793.2000.t01-1-00263.x · Cited by 1,091
Hydrogel Adhesion: A Supramolecular Synergy of Chemistry, Topology, and Mechanics
Yang, J., Bai, R., Chen, B., & Suo, Z. (2020) · Advanced Functional Materials · doi:10.1002/adfm.201901693 · Cited by 1,033
Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., et al. (2019) · Science · doi:10.1126/science.aau0780 · Cited by 807
Self-citationEfimenko, K., Wallace, W. E., & Genzer, J. (2002) · Journal of Colloid and Interface Science · doi:10.1006/jcis.2002.8594 · Cited by 757
Far-UVC light (222 nm) efficiently and safely inactivates airborne human coronaviruses
Buonanno, M., Welch, D., Shuryak, I., & Brenner, D. J. (2020) · Scientific Reports · doi:10.1038/s41598-020-67211-2 · Cited by 686
Chung, H. U., Rwei, A. Y., Hourlier-Fargette, A., Xu, S., Lee, K., Dunne, E. C., Xie, Z., Liu, C., Carlini, A., Kim, D. H., et al. (2020) · Nature Medicine · doi:10.1038/s41591-020-0792-9 · Cited by 482
Self-citationSchrader, M. E. (1995) · Langmuir · doi:10.1021/la00009a049 · Cited by 465
Raeiszadeh, M., & Adeli, B. (2020) · ACS Photonics · doi:10.1021/acsphotonics.0c01245 · Cited by 464
Zimon, A. D. (1969) · Springer eBooks · doi:10.1007/978-1-4899-6600-1 · Cited by 427
White, W. B., Berson, A. S., Robbins, C., Jamieson, M. J., Prisant, L. M., Roccella, E., & Sheps, S. G. (1993) · Hypertension · doi:10.1161/01.hyp.21.4.504 · Cited by 385
Decreasing PICU Catheter-Associated Bloodstream Infections: NACHRI's Quality Transformation Efforts
Miller, M. R., Griswold, M., Harris, J. M., Yenokyan, G., Huskins, W. C., Moss, M., Rice, T. B., Ridling, D., Campbell, D., Margolis, P., et al. (2010) · Pediatrics · doi:10.1542/peds.2009-1382 · Cited by 348
Medical device related pressure ulcers in hospitalized patients
Black, J. M., Cuddigan, J. E., Walko, M. A., Didier, L. A., Lander, M. J., & Kelpe, M. R. (2010) · International Wound Journal · doi:10.1111/j.1742-481x.2010.00699.x · Cited by 329
Less invasive hemodynamic monitoring in critically ill patients
Teboul, J.-L., Saugel, B., Cecconi, M., De Backer, D., Hofer, C. K., Monnet, X., Perel, A., Pinsky, M. R., Reuter, D. A., Rhodes, A., et al. (2016) · Intensive Care Medicine · doi:10.1007/s00134-016-4375-7 · Cited by 328
UV-C irradiation is highly effective in inactivating SARS-CoV-2 replication
Biasin, M., Bianco, A., Pareschi, G., Cavalleri, A., Cavatorta, C., Fenizia, C., Galli, P., Lessio, L., Lualdi, M., Tombetti, E., et al. (2021) · Scientific Reports · doi:10.1038/s41598-021-85425-w · Cited by 319
UV/ozone modification of poly(dimethylsiloxane) microfluidic channels
Berdichevsky, Y., Khandurina, J., Guttman, A., & Lo, Y.-H. (2004) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2003.09.022 · Cited by 318
Monitoring approaches for health-care workers during the COVID-19 pandemic
Bielicki, J. A., Duval, X., Gobat, N., Goossens, H., Koopmans, M., Tacconelli, E., & van der Werf, S. (2020) · The Lancet Infectious Diseases · doi:10.1016/s1473-3099(20)30458-8 · Cited by 315
Relation between blood pressure and pulse wave velocity for human arteries
Ma, Y., Choi, J., Hourlier-Fargette, A., Xue, Y., Chung, H. U., Lee, J. Y., Wang, X., Xie, Z., Kang, D., Wang, H., et al. (2018) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1814392115 · Cited by 309
Self-citationConversion of Some Siloxane Polymers to Silicon Oxide by UV/Ozone Photochemical Processes
Ouyang, M., Yuan, C., Muisener, R. J., Boulares, A., & Koberstein, J. T. (2000) · Chemistry of Materials · doi:10.1021/cm990770d · Cited by 287
Martin, W. R. (1973) · Archives of General Psychiatry · doi:10.1001/archpsyc.1973.01750320112017 · Cited by 264
Cuff-Less and Continuous Blood Pressure Monitoring: A Methodological Review
Sharma, M., Barbosa, K., Ho, V., Griggs, D., Ghirmai, T., Krishnan, S., Hsiai, T., Chiao, J.-C., & Cao, H. (2017) · Technologies · doi:10.3390/technologies5020021 · Cited by 226
Barr, J., Zomorodi, K., Bertaccini, E. J., Shafer, S. L., & Geller, E. (2001) · Anesthesiology · doi:10.1097/00000542-200108000-00007 · Cited by 209
Zimon, A. D. (1982) · Springer eBooks · doi:10.1007/978-1-4615-8576-3 · Cited by 207
Pressure Ulcers in Neonates and Children: An NPUAP White Paper
Baharestani, M. M., & Ratliff, C. R. (2007) · Advances in Skin & Wound Care · doi:10.1097/01.asw.0000266646.43159.99 · Cited by 175
The role of diffusion phenomena in polymer-to-polymer adhesion
Voyutskii, S. S., & Vakula, V. L. (1963) · Journal of Applied Polymer Science · doi:10.1002/app.1963.070070207 · Cited by 161
Infectious risk associated with arterial catheters compared with central venous catheters*
Lucet, J., Bouadma, L., Zahar, J., Schwebel, C., Geffroy, A., Pease, S., Hérault, M.-C., Haouache, H., Adrie, C., Thuong, M., et al. (2010) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181d4502e · Cited by 147
Cunningham, S., Symon, A., Elton, R. A., et al. (1999) · Early Human Development · doi:10.1016/s0378-3782(99)00038-9 · Cited by 144
Comparative hemodynamic effects of inotropic and vasodilator drugs in severe heart failure.
Mikulic, E., Cohn, J. N., & Franciosa, J. A. (1977) · Circulation · doi:10.1161/01.cir.56.4.528 · Cited by 131
Regulatory, safety, and privacy concerns of home monitoring technologies during COVID-19
Gerke, S., Shachar, C., Chai, P. R., & Cohen, I. G. (2020) · Nature Medicine · doi:10.1038/s41591-020-0994-1 · Cited by 126
Hopf, R., Bernardi, L., Menze, J., Zündel, M., Mazza, E., & Ehret, A. E. (2016) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2016.02.022 · Cited by 125
Variation of Arterial and Central Venous Catheter Use in United States Intensive Care Units
Gershengorn, H. B., Garland, A., Kramer, A., Scales, D. C., Rubenfeld, G., & Wunsch, H. (2014) · Anesthesiology · doi:10.1097/aln.0000000000000008 · Cited by 121
Cuff-free blood pressure estimation using pulse transit time and heart rate
Wang, R., Jia, W., Mao, Z., Sclabassi, R. J., & Sun, M. (2014) · International Conference on Signal Processing · doi:10.1109/icosp.2014.7014980 · Cited by 120
Femoral Vascular Catheterization in Critically Ill Infants and Children
Venkataraman, S. T., Thompson, A. E., & Orr, R. A. (1997) · Clinical Pediatrics · doi:10.1177/000992289703600601 · Cited by 116
A systematic review of the cardiotoxicity of methadone.
Alinejad, S., Kazemi, T., Zamani, N., Hoffman, R. S., & Mehrpour, O. (2015) · EXCLI journal · doi:10.17179/excli2015-553 · Cited by 113
Thermal ageing studies on room-temperature vulcanised polysiloxane rubbers
Patel, M., & Skinner, A. R. (2001) · Polymer Degradation and Stability · doi:10.1016/s0141-3910(01)00118-5 · Cited by 110
Ohlsson, A., Calvert, S., Hosking, M., et al. (1992) · Acta Paediatrica · doi:10.1111/j.1651-2227.1992.tb12096.x · Cited by 110
Autonomic control of heart rate after brain injury in children
Goldstein, B., Kempski, M. H., Deking, D. E., Cox, C., DeLong, D. J., Kelly, M. M., & Woolf, P. D. (1996) · Critical Care Medicine · doi:10.1097/00003246-199602000-00009 · Cited by 88
Clinical monitoring of systemic hemodynamics in critically ill newborns
de Boode, W.-P. (2010) · Early Human Development · doi:10.1016/j.earlhumdev.2010.01.031 · Cited by 87
Gao, M., Olivier, N. B., & Mukkamala, R. (2016) · Physiological Reports · doi:10.14814/phy2.12768 · Cited by 86
van Paridon, B. M., Sheppard, C., Garcia Guerra, G., Joffe, A. R., & for the Alberta Sepsis Network (2015) · Critical Care · doi:10.1186/s13054-015-1010-x · Cited by 84
Association Between Arterial Catheter Use and Hospital Mortality in Intensive Care Units
Gershengorn, H. B., Wunsch, H., Scales, D. C., Zarychanski, R., Rubenfeld, G., & Garland, A. (2014) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2014.3297 · Cited by 82
Placet, V., & Delobelle, P. (2015) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/25/3/035009 · Cited by 73
Kain, M. P., Bolker, B. M., & McCoy, M. W. (2015) · PeerJ · doi:10.7717/peerj.1226 · Cited by 71
Tian, K., Bae, J., Suo, Z., & Vlassak, J. J. (2018) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b16445 · Cited by 70
Farhadinejad, Z., Ehsani, M., Ahmadi‐Joneidi, I., et al. (2012) · IEEE Transactions on Dielectrics and Electrical Insulation · doi:10.1109/tdei.2012.6311523 · Cited by 55
Kim, Y. C., Kim, S., Kim, D., et al. (2010) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2010.01.017 · Cited by 52
Self-citationUltraviolet light degrades the mechanical and structural properties of human stratum corneum
Lipsky, Z. W., & German, G. K. (2019) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2019.103391 · Cited by 51
Anast, N., Olejniczak, M. J., Ingrande, J., & Brock‐Utne, J. G. (2015) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/s12630-015-0509-6 · Cited by 50
Stress relaxation studies of model silicone RTV networks
Stein, J. A., & Prutzman, L. C. (1988) · Journal of Applied Polymer Science · doi:10.1002/app.1988.070360305 · Cited by 44
Gowardman, J. R., Lipman, J., & Rickard, C. M. (2010) · Journal of Hospital Infection · doi:10.1016/j.jhin.2010.01.005 · Cited by 41
Ryan, R. M., Wilding, G. E., Wynn, R. J., Welliver, R. C., Holm, B. A., & Leach, C. L. (2011) · Journal of Perinatology · doi:10.1038/jp.2011.16 · Cited by 41
Patel, M., Skinner, A. R., & Maxwell, R. S. (2005) · Polymer Testing · doi:10.1016/j.polymertesting.2005.03.013 · Cited by 39
Femoral artery catheterization in neonates and infants
DuMond, A. A., da Cruz, E. M., Almodovar, M. C., et al. (2011) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0b013e3182192c7b · Cited by 38
Direct fabrication of PDMS waveguides via low-cost DUV irradiation for optical sensing
Valouch, S., Sieber, H., Kettlitz, S. W., et al. (2012) · Optics Express · doi:10.1364/oe.20.028855 · Cited by 31
Zhang, L., Gowardman, J. R., Morrison, M., Krause, L., Playford, E. G., & Rickard, C. M. (2014) · European Journal of Clinical Microbiology & Infectious Diseases · doi:10.1007/s10096-014-2058-2 · Cited by 30
Filler-induced softening effect in thermally aged polydimethylsiloxane elastomers
Yang, A. C. ‐. (1994) · Polymer · doi:10.1016/0032-3861(94)90123-6 · Cited by 28
Solav, D., Rubin, M. B., Cereatti, A., Camomilla, V., & Wolf, A. (2015) · Annals of Biomedical Engineering · doi:10.1007/s10439-015-1384-6 · Cited by 25
Cyclical variation of blood pressure and heart rate in neonates.
Cunningham, S., Deere, S., & McIntosh, N. (1993) · Archives of Disease in Childhood · doi:10.1136/adc.69.1_spec_no.64 · Cited by 24
Skin Care Management Practices for Premature Infants
Baker, S. F., Smith, B. J., Donohue, P., et al. (1999) · Journal of Perinatology · doi:10.1038/sj.jp.7200246 · Cited by 24
On the dynamics of air bubbles in Rayleigh–Bénard convection
Kim, J., Nam, J., Shen, S., Lee, C., & Chamorro, L. P. (2020) · Journal of Fluid Mechanics · doi:10.1017/jfm.2020.148 · Cited by 19
Self-citationThe importance of a centralized line service during the COVID-19 pandemic
Ilonzo, N., Rao, A., Soundararajan, K., Vouyouka, A. G., Han, D., Tadros, R. O., Kim, S. Y., Love, B. A., Ting, W., Marin, M. L., et al. (2020) · Journal of Vascular Surgery · doi:10.1016/j.jvs.2020.05.016 · Cited by 17
Self-citationArterial waveforms: Monitoring changes in configuration
Campbell, B. (1997) · Heart & Lung · doi:10.1016/s0147-9563(97)90057-1 · Cited by 12
Design of Clinical Trials to Validate Cuffless Blood Pressure Monitors
Verberk, W. J. (2019) · Springer eBooks · doi:10.1007/978-3-030-24701-0_12 · Cited by 9
Does the Radial Arterial Line Degrade the Performance of a Pulse Oximeter?
Morris, R. W., Nairn, M., & Beaudoin, M. (1990) · Anaesthesia and Intensive Care · doi:10.1177/0310057x9001800117 · Cited by 6
Jansaithong, J. (2001) · Journal for Specialists in Pediatric Nursing · doi:10.1111/j.1744-6155.2001.tb00242.x · Cited by 5
Dhillon, M. S., & Banet, M. J. (2019) · Springer eBooks · doi:10.1007/978-3-030-24701-0_5 · Cited by 5
Cyclical variation of blood pressure and heart rate in neonates.
Dunster, K. R., Colditz, P. B., & Joy, G. J. (1994) · Archives of Disease in Childhood Fetal & Neonatal · doi:10.1136/fn.70.1.f77 · Cited by 4
Research works printed but not linked: 4
- Cattivelli F. S., 2009, IEEE 2009 Sixth Int. Workshop on BSN, 114
- Wang R., 2014, IEEE 2014 12th Int. Conf. on Signal Processing (ICSP), 115
- Alinejad S., 2015, EXCLI J., 14, 577
- Khemani R. G., 2009, Cochrane Database Syst. Rev., 3, CD001000
Cited by
Thin, soft, wearable system for continuous wireless monitoring of artery blood pressure
Li, J., Jia, H., Zhou, J., Huang, X., Xu, L., Jia, S., Gao, Z., Yao, K., Li, D., Zhang, B., et al. (2023) · Nature Communications · doi:10.1038/s41467-023-40763-3 · Cited by 293
“14).”
Self-citationUltra‐Thin Flexible Encapsulating Materials for Soft Bio‐Integrated Electronics
Sang, M., Kim, K., Shin, J., & Yu, K. J. (2022) · Advanced Science · doi:10.1002/advs.202202980 · Cited by 161
Sang, M., Kang, K., Zhang, Y., Zhang, H., Kim, K., Cho, M., Shin, J., Hong, J., Kim, T., Lee, S. K., et al. (2021) · Advanced Materials · doi:10.1002/adma.202105865 · Cited by 155
Self-citationSoft shape-programmable surfaces by fast electromagnetic actuation of liquid metal networks
Ni, X., Luan, H., Kim, J.-T., Rogge, S. I., Bai, Y., Kwak, J. W., Liu, S., Yang, D. S., Li, S., Li, S., et al. (2022) · Nature Communications · doi:10.1038/s41467-022-31092-y · Cited by 110
Self-citationMin, S., An, J., Lee, J. H., Kim, J. H., Joe, D. J., Eom, S. H., Yoo, C. D., Ahn, H., Hwang, J.-Y., Xu, S. H., et al. (2025) · Nature Reviews Cardiology · doi:10.1038/s41569-025-01127-0 · Cited by 175
“Depending on the measurement location, the time differences of pulse waves are classified into two categories: pulse arrival time 176–178 and pulse transit time 178–182 .”
Self-citation
Metadata
- Authors
- Liu, C., Kim, J., Kwak, S. S., Hourlier‐Fargette, A., Avila, R., Vogl, J., Tzavelis, A., Chung, H. U., Lee, J. Y., Kim, D. H., Ryu, D., Fields, K. B., Ciatti, J. L., Li, S., Irie, M., Bradley, A., Shukla, A., Chavez, J., Dunne, E. C., Kim, S. S., Kim, J., Park, J. B., Jo, H. H., Kim, J., Johnson, M. C., Kwak, J. W., Madhvapathy, S. R., Xu, S., Rand, C. M., Marsillio, L. E., Hong, S. J., Huang, Y., Weese‐Mayer, D. E., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- Querrey Simpson Institute for Bioelectronics Northwestern University Evanston IL 60208 USA
- Lurie Children's Hospital (US)
- Sibel (United States) (US)
- Corresponding author
- Huang, Yonggang
- Weese‐Mayer, Debra E.
- Rogers, John A.
- ORCID
- Liu, Claire
- Kim, Jin‐Tae
- Kwak, Sung Soo
- Avila, Raudel
- Vogl, Jamie
- Tzavelis, Andreas
- Chung, Ha Uk
- Lee, Jong Yoon
- Kim, Dong Hyun
- Ciatti, Joanna L.
- Li, Shupeng
- Chavez, Jairo
- Dunne, Emma C.
- Kim, Jungwoo
- Park, Jun Bin
- Jo, Han Heul
- Kim, Joohee
- Johnson, Michael C.
- Kwak, Jean Won
- Madhvapathy, Surabhi R.
- Xu, Shuai
- Rand, Casey M.
- Marsillio, Lauren E.
- Hong, Sue J.
- Huang, Yonggang
- Weese‐Mayer, Debra E.
- Rogers, John A.
- Venue
- Advanced Healthcare Materials
- Publisher
- Wiley
- Published
- 2021-05-03
- Volume
- 10
- Issue
- 17
- Pages
- e2100383-e2100383
- ISSN
- 2192-2640 (ISSN-L)
- 2192-2659
- OpenAlex
- W3158794625
- Cited by
- 57
- References
- 75
- Citations by year
- 2026: 7 · 2025: 8 · 2024: 10 · 2023: 14 · 2022: 14 · 2021: 4
- FWCI
- 3.19
- Type
- journal article
- Language
- en
- Open access
- green
- hal.science
- Licence
- Locations
- Advanced Healthcare Materials (publishedVersion)
- PubMed (publishedVersion)
- HAL (submittedVersion, other-oa)
- Funders
- Gerber Foundation
- National Science Foundation (DGE‐1842165)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- non-invasive blood pressure monitoring, pediatric intensive care, neonatal patients, continuous blood pressure measurement, pharmacological interventions
- MeSH terms
- Blood Pressure, Child, Critical Care, Humans, Infant, Newborn, Monitoring, Physiologic, Skin, Vital Signs
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-29, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-29 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.