Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units

Open access

Abstract

Standard clinical care in neonatal and pediatric intensive-care units (NICUs and PICUs, respectively) involves continuous monitoring of vital signs with hard-wired devices that adhere to the skin and, in certain instances, can involve catheter-based pressure sensors inserted into the arteries. These systems entail risks of causing iatrogenic skin injuries, complicating clinical care and impeding skin-to-skin contact between parent and child. Here we present a wireless, non-invasive technology that not only offers measurement equivalency to existing clinical standards for heart rate, respiration rate, temperature and blood oxygenation, but also provides a range of important additional features, as supported by data from pilot clinical studies in both the NICU and PICU. These new modalities include tracking movements and body orientation, quantifying the physiological benefits of skin-to-skin care, capturing acoustic signatures of cardiac activity, recording vocal biomarkers associated with tonality and temporal characteristics of crying and monitoring a reliable surrogate for systolic blood pressure. These platforms have the potential to substantially enhance the quality of neonatal and pediatric critical care. Soft electronic patches worn on the skin of infants or children in intensive-care units have a wide range of capabilities in aiding critical care, including monitoring of hemodynamic parameters, cardiac activity, movement and crying.

Overview

CoChat Summary

TL;DR

A pair of soft, skin-interfaced wireless biosensors was developed for neonatal and pediatric intensive care, with pilot studies showing clinical-quality vital signs plus added modalities such as movement/orientation, kangaroo-care tracking, seismocardiography, cry analysis, and a systolic blood pressure surrogate. The system was tested in 50 NICU/PICU patients across a wide age range, with comparisons to standard clinical instruments.

Key findings

  • In NICU/PICU pilot studies, the wireless system measured HR, SpO2, temperature, and RR with close agreement to standard monitoring: mean differences were −0.02 bpm for HR, 0.11% for SpO2, and 0.21 °C for temperature (n = 3); RR mean difference was 0.11 breaths per minute with SD 1.95 (n = 6, 41 data points).
  • SpO2 accuracy root mean square was 2.99%, and a larger comparison included over 0.4 million aggregated HR/SpO2 data points from n = 20 patients, within FDA regulatory guidelines for HR and reflectance-mode SpO2.
  • Time-synchronized chest and limb units enabled PAT- and PTT-derived systolic blood pressure estimates versus arterial-line measurements, with mean differences of 1.60 and −0.04 mmHg and SDs of 7.99 and 7.86 mmHg, respectively (n = 5), within ANSI/AAMI SP10 limits.
  • Additional modalities included kangaroo-care orientation/activity tracking, SCG, and cry analysis: KC events showed phi/theta of 2.85 ± 0.10 rad and −0.29 ± 0.28 rad over 2.4 h; activity was 0.07 ± 0.02 g/s during rest/KC versus 0.24 ± 0.05 g/s during hands-on care; 11 cry events in n = 3 infants showed duration difference −3.9 ± 13.9 s and fundamental frequency 410.7 ± 47.9 Hz.

Methods

  • Design: two mechanically soft, wireless devices—a chest unit for ECG, accelerometry, temperature, and acoustic/vibration signals, and a limb unit for reflectance PPG and temperature—using Bluetooth Low Energy, with modular/embedded/battery-free power options.
  • Data/sample: pilot clinical studies in NICU and PICU patients, including 50 patients for safety/skin tolerability across 23–40 weeks gestational age and 1 week–4 years chronological age; vital-sign comparisons used smaller subsets (e.g., n = 3 for HR/SpO2/temperature, n = 6 for RR, n = 20 for aggregated HR/SpO2, n = 5 for BP).
  • Analysis: Bland-Altman comparisons against standard clinical monitors; time-synchronized PAT/PTT calibrated against arterial-line SBP; accelerometer filtering for KC orientation/activity and cry detection; MANOVA for position/activity comparisons.

Limitations

  • The paper states that in instances of low SpO2 (<90%), mean disagreement was higher than the overall comparison, likely reflecting different SpO2 de-noising algorithms rather than a true measurement difference.
  • The paper notes that comparisons reflect real-world vital-sign calculations during clinical care, feeding, imaging, and other procedures, so motion artifact and noise are inherent and processed differently between the experimental system and clinical comparator.
  • Evident from the text: key validation cohorts were small (e.g., n = 3 for HR/SpO2/temperature, n = 5 for BP, n = 3 for KC/cry), thermal safety testing used a subset of 3 patients over 24 h, and BP estimation required initial calibration against 5 min of arterial-line SBP data.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1038/s41591-020-0792-9

Cites

54 of 54 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Fig. 1. Designs, mechanical characterization results and photographs of a soft, wireless chest unit for physiological monitoring of neonatal and pediatric patients. a, Schematic diagram and expanded view illustration of a device with a modular primary battery. The main body consists of buckled serpentine interconnects between islands of electrical components contained within a soft, elastomeric enclosure. The battery interfaces to the system via reversible magnetic coupling. Thin silicone pads establish electrical connections between measurement electrodes and a hydrogel interface to the skin, to yield a completely sealed, waterproof device. LMS stands for low modulus silicone. b, Illustration of a detachable wireless power harvesting system. c, Illustration of a powering option that involves an integrated, wirelessly rechargeable battery. This option uses a different top layer encapsulation, without the magnets. d, Photograph of the chest unit with modular battery on a realistic model of a neonate. e, Computed stresses (right: normal; left: shear) at the interface between the skin and a device during uniaxial stretching to a strain of 20%, with a thickness of 200 μm. f-h, Photograph of a representative device during stretching (f), twisting (g), and bending (g). i-j, Photographs of a device with integrated battery (i) and with a wireless energy harvester (j), both mounted on a model.
Fig. 1. Designs, mechanical characterization results and photographs of a soft, wireless chest unit for physiological monitoring of neonatal and pediatric patients. a, Schematic diagram and expanded view illustration of a device with a modular primary battery. The main body consists of buckled serpentine interconnects between islands of electrical components contained within a soft, elastomeric enclosure. The battery interfaces to the system via reversible magnetic coupling. Thin silicone pads establish electrical connections between measurement electrodes and a hydrogel interface to the skin, to yield a completely sealed, waterproof device. LMS stands for low modulus silicone. b, Illustration of a detachable wireless power harvesting system. c, Illustration of a powering option that involves an integrated, wirelessly rechargeable battery. This option uses a different top layer encapsulation, without the magnets. d, Photograph of the chest unit with modular battery on a realistic model of a neonate. e, Computed stresses (right: normal; left: shear) at the interface between the skin and a device during uniaxial stretching to a strain of 20%, with a thickness of 200 μm. f-h, Photograph of a representative device during stretching (f), twisting (g), and bending (g). i-j, Photographs of a device with integrated battery (i) and with a wireless energy harvester (j), both mounted on a model.CC BY · PubMed Central
Fig. 2. Designs and photographs of a wireless limb unit for physiological monitoring of neonatal and pediatric patients and block diagram of the system operation. a, Schematic diagram and expanded view illustration of the limb unit, designed to measure PPG, SpO2 and peripheral skin temperature. b, Photograph of a device while bent and twisted. c-e, Placement of a device on (c) a model of a neonate at the ankle-to-base of the foot, (d) a model of a pediatric patient (2-year-old CA) at the wrist-to-hand, and (e) at the foot-to-toe. f, Photograph of the chest and limb unit on a model of a neonate in a NICU isolette, with a tablet computer displaying representative data through a graphical user interface. g, Block diagram showing the operational scheme of two time-synchronized devices, with an analog-front-end for ECG processing, 3-axis accelerometer, thermometer IC, and BLE SoC for the chest unit and a pulse oximeter IC, thermometer IC, and BLE SoC for the limb unit. Three different options for power supply appear at the bottom of g.
Fig. 2. Designs and photographs of a wireless limb unit for physiological monitoring of neonatal and pediatric patients and block diagram of the system operation. a, Schematic diagram and expanded view illustration of the limb unit, designed to measure PPG, SpO2 and peripheral skin temperature. b, Photograph of a device while bent and twisted. c-e, Placement of a device on (c) a model of a neonate at the ankle-to-base of the foot, (d) a model of a pediatric patient (2-year-old CA) at the wrist-to-hand, and (e) at the foot-to-toe. f, Photograph of the chest and limb unit on a model of a neonate in a NICU isolette, with a tablet computer displaying representative data through a graphical user interface. g, Block diagram showing the operational scheme of two time-synchronized devices, with an analog-front-end for ECG processing, 3-axis accelerometer, thermometer IC, and BLE SoC for the chest unit and a pulse oximeter IC, thermometer IC, and BLE SoC for the limb unit. Three different options for power supply appear at the bottom of g.CC BY · PubMed Central
Fig. 3. Photographs of wireless wearable devices on neonatal and pediatric patients in the NICU and PICU, respectively, and of parental hands-on care with healthy neonates. a-c, Photographs of the chest unit on two different ~2-year-old children (a) and (b) and on a neonatal preterm infant in (c) (27 w GA, 6 w CA) with apnea of prematurity and respiratory failure. Here, the device rests on the back. d, Photograph of the limb unit mounted on the ankle-to-foot interface of an infant (27 w GA, 6 w CA) and e, another infant (25 w GA, 44 w CA). f, Photograph of a similar unit on the foot-to-toe interface of the pediatric patient in (b). g, Photograph of a device on the wrist-to-hand interface of a pediatric patient (35 w GA, 6w CA). h-j, Photograph of a pair of devices on a 40 w GA healthy neonate during (h) KC, (i) feeding, and (j) hands-on care (diaper change).
Fig. 3. Photographs of wireless wearable devices on neonatal and pediatric patients in the NICU and PICU, respectively, and of parental hands-on care with healthy neonates. a-c, Photographs of the chest unit on two different ~2-year-old children (a) and (b) and on a neonatal preterm infant in (c) (27 w GA, 6 w CA) with apnea of prematurity and respiratory failure. Here, the device rests on the back. d, Photograph of the limb unit mounted on the ankle-to-foot interface of an infant (27 w GA, 6 w CA) and e, another infant (25 w GA, 44 w CA). f, Photograph of a similar unit on the foot-to-toe interface of the pediatric patient in (b). g, Photograph of a device on the wrist-to-hand interface of a pediatric patient (35 w GA, 6w CA). h-j, Photograph of a pair of devices on a 40 w GA healthy neonate during (h) KC, (i) feeding, and (j) hands-on care (diaper change).CC BY · PubMed Central
Fig. 4. Representative data collected in the NICU and PICU. a, Representative ECG, PPG, SCG and respiration waveforms collected from a neonate (29 w GA). b, Comparison of HR, SpO2, RR, temperature, and temperature gradient between the chest and the foot to standard clinical measurements. c-f, Corresponding Bland-Altman plots for (c) HR, (d) SpO2, (e) RR, and (f) temperature.
Fig. 4. Representative data collected in the NICU and PICU. a, Representative ECG, PPG, SCG and respiration waveforms collected from a neonate (29 w GA). b, Comparison of HR, SpO2, RR, temperature, and temperature gradient between the chest and the foot to standard clinical measurements. c-f, Corresponding Bland-Altman plots for (c) HR, (d) SpO2, (e) RR, and (f) temperature.CC BY · PubMed Central
Fig. 5. Time-synchronized operation of chest and limb units for measurements of systolic blood pressure, with comparison to arterial line data collected in the PICU on pediatric patients. a, Block diagram of the scheme for time-synchronization. b, Definition of pulse arrival time (PAT) and pulse transit time (PTT), as derived from ECG, SCG, and PPG waveforms. PEP and AO stands for pre-ejection period and aorta opening, respectively. c, Representative PAT and PTT data from a pediatric patient in the PICU during a study over 5 h. d, Results of SPB determined with PAT and PTT and with an arterial line (A-line) for an infant (34 w GA, 40 w CA) and e, another infant (40 w GA, 50 w CA). f,g, Bland-Altman plot for PAT-derived SBP (n = 5) and Bland-Altman plot for PTT-derived SBP (n = 5), respectively. Data points in the red circle indicates the comparison result at the incidents of motion artifact in 5e.
Fig. 5. Time-synchronized operation of chest and limb units for measurements of systolic blood pressure, with comparison to arterial line data collected in the PICU on pediatric patients. a, Block diagram of the scheme for time-synchronization. b, Definition of pulse arrival time (PAT) and pulse transit time (PTT), as derived from ECG, SCG, and PPG waveforms. PEP and AO stands for pre-ejection period and aorta opening, respectively. c, Representative PAT and PTT data from a pediatric patient in the PICU during a study over 5 h. d, Results of SPB determined with PAT and PTT and with an arterial line (A-line) for an infant (34 w GA, 40 w CA) and e, another infant (40 w GA, 50 w CA). f,g, Bland-Altman plot for PAT-derived SBP (n = 5) and Bland-Altman plot for PTT-derived SBP (n = 5), respectively. Data points in the red circle indicates the comparison result at the incidents of motion artifact in 5e.CC BY · PubMed Central
Fig. 6. Advanced monitoring modalities based on measurements of orientation, activity and vibratory motions. a, Definition of device axes and rotation angle between the device and reference frames. b, Orientation data extracted from low bandpass filtering of accelerometry data collected from the chest unit and derived rotation angles for various scenarios: resting in supine and right lateral positions, non-KC and KC holding. c, Filtered accelerometry data and d, rotational angles for neonates in various body positions in the NICU (n = 3). e, Representative HR, SpO2, chest and peripheral limb temperature, orientation and activity data defined as the root-mean-square of the 3-axes acceleration values between 1 and 10 Hz before, during, and after KC with a premature neonate (31 w GA). f, Raw accelerometry signal (top) and spectrogram of time-frequency signal (bottom) during crying and non-crying events from a neonate (37 w GA, large-for-gestational-age) with feeding difficulties. g, Comparison of cry duration determined with the device and by manual recording from neonates (n = 3), with a total of 11 cry events. The resolution of cry duration from our device was 0.2s, whereas manual count ranged from one second to one minute, dependent on the recorder. Error bars were defined as half of the resolution of the measurement.
Fig. 6. Advanced monitoring modalities based on measurements of orientation, activity and vibratory motions. a, Definition of device axes and rotation angle between the device and reference frames. b, Orientation data extracted from low bandpass filtering of accelerometry data collected from the chest unit and derived rotation angles for various scenarios: resting in supine and right lateral positions, non-KC and KC holding. c, Filtered accelerometry data and d, rotational angles for neonates in various body positions in the NICU (n = 3). e, Representative HR, SpO2, chest and peripheral limb temperature, orientation and activity data defined as the root-mean-square of the 3-axes acceleration values between 1 and 10 Hz before, during, and after KC with a premature neonate (31 w GA). f, Raw accelerometry signal (top) and spectrogram of time-frequency signal (bottom) during crying and non-crying events from a neonate (37 w GA, large-for-gestational-age) with feeding difficulties. g, Comparison of cry duration determined with the device and by manual recording from neonates (n = 3), with a total of 11 cry events. The resolution of cry duration from our device was 0.2s, whereas manual count ranged from one second to one minute, dependent on the recorder. Error bars were defined as half of the resolution of the measurement.CC BY · PubMed Central

Verification

Claims the paper makes

9 Valid · 9 Not valid · 13 Not checked · 63 Own results and interpretations

45 more claims

Create a free account to see all 94 claims and to verify the 1 reference still open.

See all claims — free account
  • It may state two separate claims as one. 39
  • Probably not a substantial claim. 10

Create a free account to see these claims.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

2 more claims about this paper

Create a free account to see every claim other papers make about this one.

See all claims — free account

Citations

Cites

Show all 55 references

Research works printed but not linked: 2

  1. Wheeler, D. S., Wong, H. R. & Shanley, T. P. Pediatric Critical Care Medicine: Basic Science and Clinical Evidence (Springer, 2007).
  2. Pandia, K., Inan, O.T. & Kovacs, G.T.A. in 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) 6881–6884 (2013).

Cited by

See all 482

Metadata

Authors
Chung, H. U., Rwei, A. Y., Hourlier-Fargette, A., Xu, S., Lee, K., Dunne, E. C., Xie, Z., Liu, C., Carlini, A., Kim, D. H., Ryu, D., Kulikova, E., Cao, J., Odland, I. C., Fields, K. B., Hopkins, B., Banks, A., Ogle, C., Grande, D., Park, J. B., Kim, J., Irie, M., Jang, H., Lee, J., Park, Y., Kim, J., Jo, H. H., Hahm, H., Avila, R., Xu, Y., Namkoong, M., Kwak, J. W., Suen, E., Paulus, M. A., Kim, R. J., Parsons, B. V., Human, K. A., Kim, S. S., Patel, M., Reuther, W., Kim, H. S., Lee, S. H., Leedle, J. D., Yun, Y., Rigali, S., Son, T., Jung, I., Arafa, H., Soundararajan, V. R., Ollech, A., Shukla, A., Bradley, A., Schau, M., Rand, C. M., Marsillio, L. E., Harris, Z. L., Huang, Y., Hamvas, A., Paller, A. S., Weese-Mayer, D. E., Lee, J. Y., & Rogers, J. A.
Affiliations
  • Northwestern University (US)
  • Lurie Children's Hospital (US)
  • Dalian University of Technology (CN)
  • Sibel (United States) (US)
  • University of Illinois Urbana-Champaign (US)
  • North Carolina State University (US)
  • Kyung Hee University (KR)
Corresponding author
  • Lee, Jong Yoon
  • Rogers, John A.
Venue
Nature Medicine
Published
2020-03
Volume
26
Issue
3
Pages
418-429
ISSN
  • 1078-8956 (ISSN-L)
  • 1546-170X
OpenAlex
W3012163988
Cited by
482
References
55
Citations by year
2026: 49 · 2025: 77 · 2024: 76 · 2023: 85 · 2022: 77 · 2021: 89 · 2020: 29
FWCI
21.94
Type
journal article
Language
en
Open access
Licence
Locations
  • Nature Medicine (publishedVersion)
  • PubMed (publishedVersion)
  • PubMed Central (acceptedVersion, cc-by)
  • eScholarship (California Digital Library) (submittedVersion, cc-by)
Funders
  • Bill and Melinda Gates Foundation (OPP1182909, OPP1193311)
  • Gerber Foundation
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
  • Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
Keywords
wireless physiological monitoring, neonatal intensive care, pediatric intensive care, non-invasive vital sign monitoring, skin-to-skin care, vocal biomarkers, crying
MeSH terms
Child, Child, Preschool, Electrocardiography, Equipment Design, Humans, Infant, Newborn, Intensive Care Units, Neonatal, Monitoring, Physiologic, Skin, Time Factors, Intensive Care Units, Pediatric, Biosensing Techniques, Photoplethysmography, Blood Pressure Monitoring, Ambulatory, Wireless Technology

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units. https://app.cochat.ai/papers/10.1038/s41591-020-0792-9
The paper
Chung, H. U., Rwei, A. Y., Hourlier-Fargette, A., Xu, S., Lee, K., Dunne, E. C., Xie, Z., Liu, C., Carlini, A., Kim, D. H., Ryu, D., Kulikova, E., Cao, J., Odland, I. C., Fields, K. B., Hopkins, B., Banks, A., Ogle, C., Grande, D., . . . Rogers, J. A. (2020). Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units. Nature Medicine. https://doi.org/10.1038/s41591-020-0792-9

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.