Diagnostic yield of automatic and patient‐triggered ambulatory cardiac event recording in the evaluation of patients with palpitations, dizziness, or syncope
Open accessThe DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index; none of the 20 sources its claims cite is checked for support yet.
Abstract
BACKGROUND: Recent studies have shown that patient-triggered cardiac event recorders (CER) have an increased diagnostic yield and are more cost effective than conventional 24-h-Holter electrocardiograms (ECGs) for the evaluation of sporadic, potentially arrhythmia-related symptoms. HYPOTHESIS: The aim of this study was to determine the diagnostic yield of a patient-triggered CER combined with continuous automatic arrhythmia detection in the evaluation of sporadic dizziness/syncope or palpitations and its clinical relevance in assessing the further management. METHODS: We investigated 101 consecutive outpatients (54 +/- 20 years, 40 women), referred for evaluation of sporadic dizziness and syncope (36%) or palpitations (64%) of suspected rhythmogenic origin. All were monitored by patient-triggered CER with continuous automatic arrhythmia detection. RESULTS: After a mean monitoring period of 103 +/- 38 h, 83 patients registered symptoms and 57 patients had diagnostic or therapeutic relevant arrhythmias (relA). A total of 196 episodes of relA were recorded; 31 (16%) episodes were patient-triggered and 165 (84%) automatically recorded. Diagnostic relevant episodes (relA and/or typical symptoms) occurred in 94 patients, in 54% after the first 24 h of monitoring. According to the results of the CER, 80 patients needed no further diagnostic evaluation; 20 had additional diagnostic tests. CONCLUSIONS: Cardiac event recorders with a continuous automatic arrhythmia detection function are a well-tolerated device for sporadic, potentially arrhythmia-related symptoms. The patient-triggered mode alone is not sufficiently reliable; the automatic continuous arrhythmia detection function has additional diagnostic and therapeutic consequences. In 54% of all patients, the first diagnostic event would not have been recorded with a single conventional 24-h-Holter ECG.
Overview
CoChat Summary
TL;DR
In 101 outpatients with sporadic palpitations, dizziness, or syncope, combined automatic and patient-triggered cardiac event recording found relevant arrhythmias in 57 patients and diagnostic-relevant episodes in 94. Automatic detection captured 84% of relevant arrhythmia episodes, and 54% of first diagnostic events occurred after the first 24 h, suggesting patient-triggered recording alone is insufficient.
Key findings
- After mean monitoring of 103 ± 38 h, 83 patients registered symptoms and 57 had relevant arrhythmias; 196 relevant arrhythmia episodes were recorded (1.94 per patient): 8 ventricular tachycardia, 14 ventricular bigeminus, 33 atrial fibrillation/flutter, 92 supraventricular tachycardia, 42 bradycardias, and 7 pauses.
- Of 196 relevant arrhythmia episodes, 31 (16%) were patient-triggered and 165 (84%) were automatically recorded. Among 18 asymptomatic patients, 11 (61%) had relevant arrhythmias detected only by automatic arrhythmia detection.
- Diagnostic-relevant episodes (relevant arrhythmia and/or typical symptoms) occurred in 94 patients; in 54% the first event occurred after the first 24 h and would have been missed by a single 24-h Holter ECG. CER results led to no further evaluation in 80 patients, additional diagnostic tests in 20, therapy in 23, and reassurance in 57.
- Diagnostic yield was higher with structural heart disease: 72% (38/53) had relevant arrhythmias versus 40% (17/48) without structural heart disease, while symptom registration was similar (81% vs 83%).
Methods
- Prospective study of 101 consecutive outpatients referred for sporadic (<1 episode/day) dizziness/syncope (36%) or palpitations (64%) of suspected rhythmogenic origin; mean age 54 ± 20 years, 40 women. Patients with daily symptoms or a cause identified by basic workup were excluded.
- All patients were monitored with a commercially available cardiac event recorder (R-Test Evolution, Novacor, France) combining patient-triggered recordings with continuous automatic arrhythmia detection for up to 7 days; pre- and post-triggering delays were 20 and 10 s.
- ECG strips from all patient-triggered and automatically recorded events were analyzed independently by two cardiologists blinded to the study. Relevant arrhythmias were predefined as ventricular tachycardia (>100 beats/min and ≥3 beats), ventricular bigeminus, supraventricular tachycardia (>100 beats/min and ≥3 beats), atrial fibrillation or flutter, bradycardia (<40 beats/min), and pauses (RR interval ≥3 s).
Limitations
- Stated: the study had a relatively small sample of patients studied in a tertiary referral center.
- Stated: there was no recording during hours when the CER was not in place (e.g., during bathing), with the possibility of missing relevant events; also, 52% of patients had known cardiovascular pathology and were likely to have an arrhythmia within the week of monitoring, so benefit may be greatest in patients with a higher likelihood of an event and may not apply to those without structural heart disease who experience arrhythmic events more rarely.
- Evident: this was a single-center observational study without a concurrent control group; comparisons with 24-h Holter ECG and other cardiac event recorder studies are indirect.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 Valid · 1 Not valid · 10 Not checked · 33 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Of those who did not trigger the recording of any relevant arrhythmia during their symptoms, 63% nevertheless showed relevant arrhythmias in the automatically registered ECG strips (Fig. 1).”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Of the 18 patients who were asymptomatic during the monitoring period, 11 (61%) had relevant arrhythmias registered by the automatic arrhythmia detection function only (18 episodes of supraventricular tachycardia, 6 episodes of atrial fibrillation or flutter, 6 episodes of ventricular bigeminus, and 1 pause).”
plainn = 18- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Diagnostically relevant episodes, defined as relevant arrhythmias and/or typical symptoms, occurred in 94 patients.”
plainn = 94- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Of the 80 patients who had no need for further investigations, 41 (51%) had their first diagnostic event after the first 24 h of monitoring.”
plainn = 80- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The number of patients registering symptoms during the monitoring period was not significantly different between these two groups: 81% (43/53) compared with 83% (40/48) (Table III).”
plainn = 48- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
17 more claims
Create a free account to see all 44 claims and every source verdict.
- Probably not a substantial claim. 14
- It may state two separate claims as one. 11
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Discussion · p. 4
“Of those who did not trigger the recording of any relevant arrhythmia during their symptoms, 63% nevertheless showed relevant arrhythmias in the automatically registered ECG strips (Fig. 1).”
§ Discussion · p. 3
“| Patient-triggered relA | 1 | 11 | 5 | 0 | 17 | | Automatically triggered relA | 6 | 7 | 16 | 0 | 29 | | No relA | 8 | 3 | 25 | 1 | 37 |”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
“Of the 289 registered patient-triggered ECG strips, 225 (78%) showed normal sinus rhythm, 28 (9%) sinus tachycardia, 31 (11%) relevant arrhythmias”
Reported 9 · recomputed 9.688581
Citations
Cites
Evaluation and outcome of emergency room patients with transient loss of consciousness
Day, S. C., Cook, E. F., Funkenstein, H., & Goldman, L. (1982) · The American Journal of Medicine · doi:10.1016/0002-9343(82)90913-5 · Cited by 587
The Prevalence of Symptoms in Medical Outpatients and the Adequacy of Therapy
Arrington, M. E. (1990) · Archives of Internal Medicine · doi:10.1001/archinte.1990.00040031685016 · Cited by 367
Kinlay, S., Leitch, J. W., Neil, A., Chapman, B. L., Hardy, D. B., & Fletcher, P. J. (1996) · Annals of Internal Medicine · doi:10.7326/0003-4819-124-1_part_1-199601010-00003 · Cited by 245
Incremental diagnostic yield of loop electrocardiographic recorders in unexplained syncope
Linzer, M., Pritchett, E. L. C., Pontinen, M., McCarthy, E., & Divine, G. W. (1990) · The American Journal of Cardiology · doi:10.1016/0002-9149(90)90591-n · Cited by 239
Zimetbaum, P. J., Kim, K. Y., Josephson, M. E., Goldberger, A. L., & Cohen, D. J. (1998) · Annals of Internal Medicine · doi:10.7326/0003-4819-128-11-199806010-00002 · Cited by 133
Utility and cost of event recorders in the diagnosis of palpitations, presyncope, and syncope
Fogel, R. I., Evans, J. J., & Prystowsky, E. N. (1997) · The American Journal of Cardiology · doi:10.1016/s0002-9149(96)00717-5 · Cited by 83 · cited 1×
“Age (years) 54 ± 20 Men/women (n) 61/40 Indications for monitoring Palpitations (%) 65 (64) Dizziness/syncope (%) 36 (36) Known cardiovascular disease a (%) 53 (52) Coronary artery disease (%) 29 (29) Hypertension (%) 9 (9) Valvular heart disease (%) 19 (19) Cardiomyopathy (%) 4 (4) Others (%) 6 (6)”
Utility of Patient-Activated Cardiac Event Recorders in General Clinical Practice
Zimetbaum, P., Kim, K. Y., Ho, K. K. L., Zebede, J., Josephson, M. E., & Goldberger, A. L. (1997) · The American Journal of Cardiology · doi:10.1016/s0002-9149(96)00766-7 · Cited by 77
Roche, F., Gaspoz, J.-M., Pichot, V., Costes, F., Isaaz, K., Ferron, C., Roche, C., Geyssant, A., Lacour, J.-R., & Barthélémy, J.-C. (1997) · The American Journal of Cardiology · doi:10.1016/s0002-9149(97)00614-0 · Cited by 26
Research works printed but not linked: 1
- Simonetti G, 1998, Prospektive Vergleichsuntersuchung eines Aktivierungs‐Holters mit einem konventionellen Holter‐EKG bei Patienten mit Synkopen, Präsynkopen and Palpitationen, Schweiz Med Wochenschr, 128, 29S
Cited by
ROTHMAN, S. A., LAUGHLIN, J. C., SELTZER, J., WALIA, J. S., BAMAN, R. I., SIOUFFI, S. Y., SANGRIGOLI, R. M., & KOWEY, P. R. (2007) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2006.00729.x · Cited by 194
Joshi, A. K., Kowey, P. R., Prystowsky, E. N., Benditt, D. G., Cannom, D. S., Pratt, C. M., McNamara, A., & Sangrigoli, R. M. (2005) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2004.12.015 · Cited by 110
Long-term continuous external electrocardiographic recording: a review
Enseleit, F., & Duru, F. (2006) · EP Europace · doi:10.1093/europace/euj054 · Cited by 62
Cardiovascular assessment of falls in older people
Tan, M. P., & Kenny, R. A. (2006) · Clinical Interventions in Aging · doi:10.2147/ciia.2006.1.1.57 · Cited by 52
“Cardiac event recorders with a continuous automatic arrhythmia detection function (Balmelli et al 2003) may be more appropriate for older people with unexplained falls, but symptom correlation may not be possible.”
BackgroundEfficacy of diagnostic tools for detecting cardiac arrhythmias: systematic literature search
Hoefman, E., Bindels, P., & van Weert, H. C. P. M. (2010) · Netherlands Heart Journal · doi:10.1007/s12471-010-0831-0 · Cited by 50
Metadata
- Authors
- Balmelli, N., Naegeli, B., & Bertel, O.
- Affiliations
- Triemli Hospital (CH)
- Corresponding author
- Naegeli, Barbara
- Venue
- Clinical Cardiology
- Publisher
- Wiley
- Published
- 2003-04
- Volume
- 26
- Issue
- 4
- Pages
- 173-176
- ISSN
- 0160-9289 (ISSN-L)
- 1932-8737
- PMC
- PMC6654562
- OpenAlex
- W1968170366
- Semantic Scholar
- 19583da41d362264af0c16e6551f55f2326b2666
- Cited by
- 35
- References
- 8
- Citations by year
- 2026: 2 · 2023: 1 · 2022: 1 · 2021: 2 · 2018: 3 · 2017: 2 · 2016: 3 · 2014: 1 · 2013: 1 · 2012: 2
- FWCI
- 1.60
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Clinical Cardiology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Syncope and Autonomic Disorders
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac electrophysiology and arrhythmias
- Keywords
- automatic arrhythmia detection, ambulatory cardiac monitoring, palpitations, syncope, dizziness, arrhythmia diagnosis, diagnostic yield
- MeSH terms
- Arrhythmias, Cardiac, Dizziness, Electrocardiography, Female, Humans, Male, Middle Aged, Syncope, Electrocardiography, Ambulatory, Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-12, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-12 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Diagnostic yield of automatic and patient‐triggered ambulatory cardiac event recording in the evaluation of patients with palpitations, dizziness, or syncope. https://app.cochat.ai/papers/10.1002/clc.4960260405
- The paper
- Balmelli, N., Naegeli, B., & Bertel, O. (2003). Diagnostic yield of automatic and patient‐triggered ambulatory cardiac event recording in the evaluation of patients with palpitations, dizziness, or syncope. Clinical Cardiology. https://doi.org/10.1002/clc.4960260405
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.