Feasibility and Diagnostic Performance of the Fibroscan XL Probe for Liver Stiffness Measurement in Overweight and Obese Patients

Myers, Robert P., Pomier–Layrargues, Gilles, Kirsch, Richard, Pollett, Aaron, Duarte–Rojo, Andres, Wong, David, Beaton, Melanie, Levstik, Mark, Crotty, Pam, Elkashab, Magdy·2012·Hepatology

Abstract

Failure of liver stiffness measurement (LSM) by transient elastography (TE, FibroScan) and unreliable results occur in ≈5% and 15% of patients, respectively, mainly due to obesity. In this multicenter study, we evaluated the feasibility and performance of the novel FibroScan XL probe in 276 patients with chronic liver disease (42% viral hepatitis, 46% nonalcoholic fatty liver disease [NAFLD]) and a body mass index (BMI) ≥28 kg/m2. Patients underwent liver biopsy and TE with the standard M and XL probes. TE failure was defined as no valid LSMs and unreliable examinations as <10 valid LSMs or an interquartile range (IQR)/LSM >30% or success rate <60%. Probe performance for diagnosing ≥F2 fibrosis and cirrhosis (F4) versus biopsy were examined using areas under receiver operating characteristic curves (AUROC). FibroScan failure was less frequent with the XL probe than the M probe (1.1% versus 16%) and the XL probe was more often reliable (73% versus 50%; both P < 0.00005). Reliable results with the XL probe were obtained in 61% of patients in whom the M probe was unreliable. Among 178 patients with ≥10 valid LSMs using both probes, liver stiffness was highly correlated between probes (ρ = 0.86; P < 0.0005); however, median liver stiffness was lower using the XL probe (6.8 versus 7.8 kPa; P < 0.00005). The AUROC of the XL and M probes were similar for ≥F2 fibrosis (0.83 versus 0.86; P = 0.19) and cirrhosis (0.94 versus 0.91; P = 0.28). Conclusion : Compared with the M probe, the FibroScan XL probe reduces TE failure and facilitates reliable LSM in obese patients. Although the probes have comparable accuracy, lower liver stiffness cutoffs will be necessary when the XL probe is used to noninvasively assess liver fibrosis. (Hepatology 2012)

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1002/hep.24624

Cites

27 of 27 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 27 references

Cited by

See all 503

Metadata

Authors
Myers, R. P., Pomier–Layrargues, G., Kirsch, R., Pollett, A., Duarte–Rojo, A., Wong, D., Beaton, M., Levstik, M., Crotty, P., & Elkashab, M.
Venue
Hepatology
Published
2012-01
Volume
55
Issue
1
Pages
199-208
ISSN
0270-9139 (ISSN-L)
OpenAlex
W2154739921
Cited by
503
References
27
FWCI
28.31
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Feasibility and Diagnostic Performance of the Fibroscan XL Probe for Liver Stiffness Measurement in Overweight and Obese Patients. https://app.cochat.ai/papers/10.1002/hep.24624
The paper
Myers, R. P., Pomier–Layrargues, G., Kirsch, R., Pollett, A., Duarte–Rojo, A., Wong, D., Beaton, M., Levstik, M., Crotty, P., & Elkashab, M. (2012). Feasibility and Diagnostic Performance of the Fibroscan XL Probe for Liver Stiffness Measurement in Overweight and Obese Patients. Hepatology. https://doi.org/10.1002/hep.24624

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.