Performance of a 200‐MHz cryogenic RF probe designed for MRI and MRS of the murine brain
Abstract
A cryogenic 200‐MHz RF transmit/receive probe (CryoProbe) made of copper and designed for murine brain studies is described. The probe operates at 30K and its performance was compared to a home‐made surface coil of equal dimensions and a mouse head volume resonator, both operating at room temperature (RT). Since for small volumes of tissue interrogated (<5 mL) sample noise and thermal noise of the receiver are of comparable magnitude, considerable noise reduction can be achieved by cooling of the receiver system, e.g., using cold helium gas. The effect of temperature lowering on the signal‐to‐noise ratio (SNR) was assessed in phantom and in vivo murine brain studies using conventional MRI sequences (spin and gradient echo) and single voxel MR spectroscopy. Consistent sensitivity increases were observed for the CryoProbe. SNR increases by a factor ≥2 for the phantom experiments and by factors of 1.8 to 1.9 for in vivo experiments when compared to the RT surface coil for both MRI and MRS experiments. These results are in good agreement with estimated SNR gains for a circular surface coil probe cooled to 30K. The use of a cryogenic probe represents an economically attractive option to enhance the sensitivity in small animal MRI/MRS. Magn Reson Med, 2008. © 2008 Wiley‐Liss, Inc.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Cites
RARE imaging: A fast imaging method for clinical MR
Hennig, J., Nauerth, A., & Friedburg, H. (1986) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910030602 · Cited by 2,088 · see citing works
Thermal Agitation of Electricity in Conductors
Johnson, J. B. (1928) · Physical Review · doi:10.1103/physrev.32.97 · Cited by 1,811 · see citing works
Spatial Localization in NMR Spectroscopy in Vivo
Bottomley, P. A. (1987) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1987.tb32915.x · Cited by 1,382 · see citing works
Biological magnetic resonance imaging using laser-polarized 129Xe
Albert, M. S., Cates, G. D., Driehuys, B., et al. (1994) · Nature · doi:10.1038/370199a0 · Cited by 1,059 · see citing works
In vivo1H NMR spectroscopy of rat brain at 1 ms echo time
Tk�?, I., uk, Z. S., Choi, I., & Rolf Gruetter, A. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199904)41:4 · Cited by 994 · see citing works
The Return of the Frequency Sweep: Designing Adiabatic Pulses for Contemporary NMR
Garwood, M., & DelaBarre, L. J. (2001) · Journal of Magnetic Resonance · doi:10.1006/jmre.2001.2340 · Cited by 954 · see citing works
The sensitivity of the zeugmatographic experiment involving human samples
Hoult, D. I., & Lauterbur, P. C. (1979) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(79)90019-2 · Cited by 778 · see citing works
The intrinsic signal-to-noise ratio in NMR imaging
Edelstein, W. A., Glover, G. H., Hardy, C. J., & Redington, R. W. (1986) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910030413 · Cited by 758 · see citing works
MR Imaging with Hyperpolarized 3He Gas
Middleton, H., Black, R. D., Saam, B. T., et al. (1995) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910330219 · Cited by 359 · see citing works
Imaging of the active B1 field in vivo
Stollberger, R., & Wach, P. (1996) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910350217 · Cited by 357 · see citing works
Radio frequency coil technology for small‐animal MRI
Doty, F. D., Entzminger, G., Kulkarni, J., et al. (2007) · NMR in Biomedicine · doi:10.1002/nbm.1149 · Cited by 210 · see citing works
A High-Temperature Superconducting Receiver for Nuclear Magnetic Resonance Microscopy
Black, R. D., Early, T. A., Roemer, P. B., et al. (1993) · Science · doi:10.1126/science.8430331 · Cited by 207 · see citing works
Perspectives with cryogenic RF probes in biomedical MRI
Darrasse, L. (2003) · Biochimie · doi:10.1016/j.biochi.2003.09.016 · Cited by 201 · see citing works
Assaf, Y., & Cohen, Y. (2000) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(200002)43:2 · Cited by 184 · see citing works
A high-resolution NMR probe in which the coil and preamplifier are cooled with liquid helium
Styles, P., Soffe, N., Scott, C. A., et al. (1984) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(84)90050-7 · Cited by 139 · see citing works
Diffusion tensor microscopy of the intervertebral disc anulus fibrosus
Hsu, E. W., & Setton, L. A. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199905)41:5 · Cited by 127 · see citing works
MR properties of rat sciatic nerve following trauma
Stanisz, G. J., Midha, R., Munro, C. A., & Henkelman, R. M. (2001) · Magnetic Resonance in Medicine · doi:10.1002/1522-2594(200103)45:3 · Cited by 95 · see citing works
Thin film high temperature superconducting RF coils for low field MRI
van Heteren, J. G., James, T. W., & Bourne, L. C. (1994) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910320315 · Cited by 61 · see citing works
Use of high temperature superconductor in a receiver coil for magnetic resonance imaging
Hall, A. S., Alford, N. M., Button, T. W., et al. (1991) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910200218 · Cited by 59 · see citing works
Performance of a High-Temperature Superconducting Resonator for High-Field Imaging
Black, R. D., Early, T. A., & Johnson, G. A. (1995) · Journal of Magnetic Resonance Series A · doi:10.1006/jmra.1995.1058 · Cited by 43 · see citing works
A superconducting probehead applicable for nuclear magnetic resonance microscopy at 7 T
Odoj, F., Rommel, E., von Kienlin, M., et al. (1998) · Review of Scientific Instruments · doi:10.1063/1.1149003 · Cited by 32 · see citing works
Potential benefits of a cryogenically cooled NMR probe for room-temperature samples
Jerosch‐Herold, M., & Kirschman, R. (1989) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(89)90327-2 · Cited by 29 · see citing works
Progress in high field MRI at the University of Florida
Beck, B. L. (2002) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1016/s1352-8661(01)00140-5 · Cited by 29 · see citing works
Performance of Large-Size Superconducting Coil in 0.21T MRI System
Lee, K. H., Cheng, M.-S., Chan, K. C., et al. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.831539 · Cited by 22 · see citing works
Progress in high field MRI at the University of Florida
Beck, B. L., Plant, D. P., Grant, S. C., et al. (2001) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/bf02678590 · Cited by 15 · see citing works
Buckley (1999) · doi:10.1002/(sici)1522-2594(199901)41:1
Flagged first, then by citations
Research works printed but not linked: 1
- VesterM SteinmeyerF RoasB ThummesG KlundtK.High temperature superconducting surface coils with liquid nitrogen or pulse tube refrigeration. In: Proc 5th Annual Meeting ISMRM Vancouver Canada;1997:1528.
Cited by
Sorted by citations
Micro MRI of the mouse brain using a novel 400 MHz cryogenic quadrature RF probe
Baltes, C., Radzwill, N., Bosshard, S., Marek, D., & Rudin, M. (2009) · NMR in Biomedicine · doi:10.1002/nbm.1396 · Cited by 149 · see citing works
Self-citationMüller, H.-P., Roselli, F., Rasche, V., & Kassubek, J. (2020) · Frontiers in Neuroscience · doi:10.3389/fnins.2020.00734 · Cited by 12 · see citing works
- Authors
- Ratering, D., Baltes, C., Nordmeyer‐Massner, J., Marek, D., & Rudin, M.
- Affiliations
- ETH Zurich (CH)
- Bruker (Switzerland) (CH)
- University of Zurich (CH)
- Corresponding author
- Rudin, Markus
- ORCID
- Rudin, Markus
- Venue
- Magnetic Resonance in Medicine
- Publisher
- Wiley
- Published
- 2008-04-17
- Volume
- 59
- Issue
- 6
- Pages
- 1440-1447
- ISSN
- 0740-3194 (ISSN-L)
- 1522-2594
- OpenAlex
- W2043164692
- Semantic Scholar
- 077bcfe8e17bd309e7af292350f220cc7ae0cb20
- Cited by
- 86
- References
- 26
- Citations by year
- 2025: 3 · 2024: 2 · 2023: 9 · 2022: 4 · 2021: 5 · 2020: 5 · 2019: 4 · 2018: 2 · 2017: 4 · 2016: 6 · 2015: 4 · 2014: 6 · 2013: 3 · 2012: 7
- FWCI
- 4.79
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Magnetic Resonance in Medicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Physical Sciences › Chemistry › Spectroscopy › Advanced NMR Techniques and Applications
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Keywords
- MR spectroscopy, small animal MRI, signal-to-noise ratio, thermal noise reduction, surface coil
- MeSH terms
- Animals, Body Temperature, Brain Mapping, Cold Temperature, Copper, Equipment Design, Image Enhancement, Magnetic Resonance Imaging, Mice, Inbred C57BL, Magnetic Resonance Spectroscopy, Radio Waves, Phantoms, Imaging, Mice
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review