Continuous flow 2 H/ 1 H and 18 O/ 16 O analysis of water samples with dual inlet precision

Gehre, M., Geilmann, H., Richter, J., Werner, R. A., Brand, W. A.·2004·Rapid Communications in Mass Spectrometry

Abstract

A method for isotope ratio analysis of water samples is described comprising an on‐line high‐temperature reduction technique in a helium carrier gas. Using a gas‐tight syringe, injection of 0.5 to 1 μL sample is made through a heated septum into a glassy carbon reactor at temperatures in excess of 1300°C. More than 150 injections can be made per day and both isotope ratios of interest, δ 2 H and δ 18 O, can be measured with the same setup. The technique has the capability to transfer high‐precision stable isotope ratio analysis of water samples from a specialized to a routine laboratory task compatible with other common techniques (automated injection for GC, LC, etc.). Experiments with an emphasis on the reactor design were made in two different laboratories using two different commercially available high‐temperature elemental analyser (EA) systems. In the Jena TC/EA unit, sample‐to‐sample memory (single injection) has been reduced to ∼1% and high precision of about 0.1‰ for δ 18 O and < 1‰ for δ 2 H has been achieved by a redesign of the glassy carbon reactor and by redirecting the gas flow of the commercial TC/EA unit. With the modified reactor, the contact of water vapour with surfaces other than glassy carbon is avoided completely. The carrier gas is introduced at the bottom of the reactor thereby flushing the outer tube compartment of the tube‐in‐tube assembly before entering the active heart of the reactor. With the Leipzig high‐temperature reactor (HTP) similar precision was obtained with a minor modification (electropolishing) of the injector metal sleeve. With this system, the temperature dependence of the reaction has been studied between 1100 and 1450°C. Complete yield and constant isotope ratio information has been observed only for temperatures above 1325°C. For temperatures above 1300°C the reactor produces an increasing amount of CO background from reaction of glass carbon with the ceramic tube. This limits the usable temperature to a maximum of 1450°C. Relevant gas permeation through the Al 2 O 3 walls has not been detected up to 1600°C. Copyright © 2004 John Wiley & Sons, Ltd.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1002/rcm.1672

Cites

15 of 15 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 15 references

Cited by

See all 258

Metadata

Authors
Gehre, M., Geilmann, H., Richter, J., Werner, R. A., & Brand, W. A.
Venue
Rapid Communications in Mass Spectrometry
Publisher
Wiley
Published
2004-10-12
Volume
18
Issue
22
Pages
2650-2660
ISSN
  • 0951-4198 (ISSN-L)
  • 1097-0231
OpenAlex
W1986789727
Cited by
258
References
15
FWCI
10.71
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Continuous flow 2 H/ 1 H and 18 O/ 16 O analysis of water samples with dual inlet precision. https://app.cochat.ai/papers/10.1002/rcm.1672
The paper
Gehre, M., Geilmann, H., Richter, J., Werner, R. A., & Brand, W. A. (2004). Continuous flow 2 H/ 1 H and 18 O/ 16 O analysis of water samples with dual inlet precision. Rapid Communications in Mass Spectrometry. https://doi.org/10.1002/rcm.1672

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.