CoChat

Development, validation, and clinical assessment of a liquid chromatography-tandem mass spectrometry serum assay for per- and polyfluoroalkyl substances (PFAS) recommended by the National Academies of Science, Engineering, and Medicine (NASEM)

Open access
Dui, Wen, Smith, Michael P., Bartock, Sarah H. · 2024 · Analytical and Bioanalytical Chemistry

The DOI matches its Crossref record; 57 of the 57 references it cites were found in a registry or an index.

Abstract

Per- and polyfluoroalkyl substances (PFAS) are widely used in industry, residential, and consumer products. Studies have shown associations between high PFAS exposure and adverse health effects. In 2022, the National Academies of Science, Engineering, and Medicine (NASEM) published Guidance on PFAS Exposure, Testing, and Clinical Follow-up providing laboratory and clinical direction. The Guidance suggests nine PFAS should be measured in serum or plasma specimens and summed to provide a total PFAS concentration using a NASEM-recommended method. Follow-up clinical recommendations are based on the calculated PFAS NASEM summation. We developed and validated a liquid chromatography-tandem mass spectrometry (LC–MS/MS) method in accordance with NASEM recommendations but distinguished by the ability to separate closely related structural isomers. As part of our validation, PFAS prevalence was evaluated in a population survey comprised of clinical donor and remnant specimens ( n = 1023 in total). In this study, 82.2% of the specimens had PFAS NASEM summations of 2 to < 20 ng/mL and 2.5% had a summation ≥ 20 ng/mL. The median PFAS NASEM summation was 4.65 ng/mL in this study, lower than the 7.74 ng/mL median observed in the 2017–2020 Centers for Disease Control and Prevention, National Health and Nutrition Examination Survey ( n = 3072). This lower median PFAS NASEM summation may reflect a decline in PFAS population levels over time or sample population exposure differences. Graphical Abstract

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1007/s00216-024-05519-y

Cites

57 of 57 references verified0 citation errors · 0 not checkable

Statistics

Figures

Fig. 1 Chromatographical view of ten PFAS analytes in LC–MS/MS method, scheduled multiple reaction monitoring (sMRM) scan mode with a detection time window of 60 s (PFUnDA 40 s). (a) Extracted ion chromatogram (XIC) view of all analytes at 1.25 ng/mL. (b) Individual chromatograms of quantifier at 0.1 ng/mL, analyte name on the top followed by the retention time (min)
Fig. 1 Chromatographical view of ten PFAS analytes in LC–MS/MS method, scheduled multiple reaction monitoring (sMRM) scan mode with a detection time window of 60 s (PFUnDA 40 s). (a) Extracted ion chromatogram (XIC) view of all analytes at 1.25 ng/mL. (b) Individual chromatograms of quantifier at 0.1 ng/mL, analyte name on the top followed by the retention time (min)CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 57 references

Research works printed but not linked: 3

  1. Approved American National Standard/Academy Standard Board (ANSI/ASB) Standard 113. Standard for Identification Criteria in Forensic Toxicology. First Edition 2023 by the AAFS Standard Board, LLC.
  2. Approved American National Standard/Academy Standard Board (ANSI/ASB) Standard 098. Standard for Mass Spectral Analysis in Forensic Toxicology. First Edition 2023 by the AAFS Standard Board, LLC.
  3. Kissa E. Fluorinated surfactants and repellents, 2nd edit. New York: Marcel Dekker; 2005.

Cited by

See all 9

Metadata

Authors
Dui, W., Smith, M. P., & Bartock, S. H.
Affiliations
Quest Diagnostics (United States) (US)
Corresponding author
Bartock, Sarah H.
Venue
Analytical and Bioanalytical Chemistry
Published
2024-09-13
Volume
416
Issue
28
Pages
6333-6344
ISSN
  • 1618-2642 (ISSN-L)
  • 1618-2650
OpenAlex
W4402529063
Cited by
9
References
56
Citations by year
2026: 6 · 2025: 3
FWCI
1.18
Type
journal article
Language
en
Open access
Licence
Locations
  • Analytical and Bioanalytical Chemistry (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
Topics
  • Physical Sciences › Environmental Science › Environmental Chemistry › Per- and polyfluoroalkyl substances research
  • Physical Sciences › Chemistry › Organic Chemistry › Fluorine in Organic Chemistry
  • Physical Sciences › Environmental Science › Health, Toxicology and Mutagenesis › Toxic Organic Pollutants Impact
Keywords
per- and polyfluoroalkyl substances, structural isomer separation, LC-MS/MS method development, method validation, clinical follow-up
MeSH terms
Chromatography, Liquid, Environmental Exposure, Environmental Pollutants, Fluorocarbons, Humans, United States, Reproducibility of Results, National Academies of Science, Engineering, and Medicine, U.S., Health and Medicine Division, Tandem Mass Spectrometry, Limit of Detection

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Development, validation, and clinical assessment of a liquid chromatography-tandem mass spectrometry serum assay for per- and polyfluoroalkyl substances (PFAS) recommended by the National Academies of Science, Engineering, and Medicine (NASEM). https://app.cochat.ai/papers/10.1007/s00216-024-05519-y
The paper
Dui, W., Smith, M. P., & Bartock, S. H. (2024). Development, validation, and clinical assessment of a liquid chromatography-tandem mass spectrometry serum assay for per- and polyfluoroalkyl substances (PFAS) recommended by the National Academies of Science, Engineering, and Medicine (NASEM). Analytical and Bioanalytical Chemistry. https://doi.org/10.1007/s00216-024-05519-y

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.