Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques
Open accessAbstract
INTRODUCTION: Positron emission tomography (PET) is a fully quantitative technology for imaging metabolic pathways and dynamic processes in vivo. Attenuation correction of raw PET data is a prerequisite for quantification and is typically based on separate transmission measurements. In PET/CT attenuation correction, however, is performed routinely based on the available CT transmission data. OBJECTIVE: Recently, combined PET/magnetic resonance (MR) has been proposed as a viable alternative to PET/CT. Current concepts of PET/MRI do not include CT-like transmission sources and, therefore, alternative methods of PET attenuation correction must be found. This article reviews existing approaches to MR-based attenuation correction (MR-AC). Most groups have proposed MR-AC algorithms for brain PET studies and more recently also for torso PET/MR imaging. Most MR-AC strategies require the use of complementary MR and transmission images, or morphology templates generated from transmission images. We review and discuss these algorithms and point out challenges for using MR-AC in clinical routine. DISCUSSION: MR-AC is work-in-progress with potentially promising results from a template-based approach applicable to both brain and torso imaging. While efforts are ongoing in making clinically viable MR-AC fully automatic, further studies are required to realize the potential benefits of MR-based motion compensation and partial volume correction of the PET data.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Spatial registration and normalization of images
Friston, K. J., Ashburner, J., Frith, C. D., Poline, J. ‐., Heather, J. D., & Frackowiak, R. S. J. (1995) · Human Brain Mapping · doi:10.1002/hbm.460030303 · Cited by 3,744
Simultaneous PET-MRI: a new approach for functional and morphological imaging
Judenhofer, M. S., Wehrl, H. F., Newport, D. F., Catana, C., Siegel, S. B., Becker, M., Thielscher, A., Kneilling, M., Lichy, M. P., Eichner, M., et al. (2008) · Nature Medicine · doi:10.1038/nm1700 · Cited by 981
Self-citationAttenuation correction for a combined 3D PET/CT scanner
Kinahan, P. E., Townsend, D. W., Beyer, T., & Sashin, D. (1998) · Medical Physics · doi:10.1118/1.598392 · Cited by 946 · cited 1×
“511 keV [3, 4].”
Self-citationSUV: standard uptake or silly useless value?
Keyes, J. W. (1995) · PubMed · Cited by 803
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners
Kinahan, P. E., Hasegawa, B. H., & Beyer, T. (2003) · Seminars in Nuclear Medicine · doi:10.1053/snuc.2003.127307 · Cited by 515
Self-citation2000 IEEE Nuclear Science Symposium Conference Record
(2000) · 2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149) · doi:10.1109/nssmic.2000.949858 · Cited by 496
Hofmann, M., Steinke, F., Scheel, V., Charpiat, G., Farquhar, J., Aschoff, P., Brady, M., Schölkopf, B., & Pichler, B. J. (2008) · Journal of Nuclear Medicine · doi:10.2967/jnumed.107.049353 · Cited by 495
Self-citationBurger, C., Goerres, G., Schoenes, S., Buck, A., Lonn, A., & von Schulthess, G. (2002) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-002-0796-3 · Cited by 428
Simultaneous PET and MR imaging
Shao, Y., Cherry, S. R., Farahani, K., Meadors, K., Siegel, S., Silverman, R. W., & Marsden, P. K. (1997) · Physics in Medicine and Biology · doi:10.1088/0031-9155/42/10/010 · Cited by 374
Improvements in cancer staging with PET/CT: literature-based evidence as of September 2006.
Czernin, J. G., Allen-Auerbach, M., & Schelbert, H. R. (2007) · PubMed · Cited by 355
Cited by
MR-based synthetic CT generation using a deep convolutional neural network method
Han, X. (2017) · Medical Physics · doi:10.1002/mp.12155 · Cited by 711
Drzezga, A., Souvatzoglou, M., Eiber, M., Beer, A. J., Fürst, S., Martinez-Möller, A., Nekolla, S. G., Ziegler, S., Ganter, C., Rummeny, E. J., et al. (2012) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.098608 · Cited by 508
Advances in multimodal data fusion in neuroimaging: Overview, challenges, and novel orientation
Zhang, Y.-D., Dong, Z., Wang, S.-H., Yu, X., Yao, X., Zhou, Q., Hu, H., Li, M., Jiménez-Mesa, C., Ramirez, J., et al. (2020) · Information Fusion · doi:10.1016/j.inffus.2020.07.006 · Cited by 453
PET/MRI: Paving the Way for the Next Generation of Clinical Multimodality Imaging Applications
Pichler, B. J., Kolb, A., Nägele, T., & Schlemmer, H.-P. (2010) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.061853 · Cited by 418
Self-citationDeep Learning MR Imaging–based Attenuation Correction for PET/MR Imaging
Liu, F., Jang, H., Kijowski, R., Bradshaw, T., & McMillan, A. B. (2018) · Radiology · doi:10.1148/radiol.2017170700 · Cited by 412
Design and performance evaluation of a whole-body Ingenuity TF PET–MRI system
Zaidi, H., Ojha, N., Morich, M., Griesmer, J., Hu, Z., Maniawski, P., Ratib, O., Izquierdo-Garcia, D., Fayad, Z. A., & Shao, L. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/10/013 · Cited by 402
A review of substitute CT generation for MRI-only radiation therapy
Edmund, J. M., & Nyholm, T. (2017) · Radiation Oncology · doi:10.1186/s13014-016-0747-y · Cited by 378
EANM procedure guidelines for brain PET imaging using [18F]FDG, version 3
Guedj, E., Varrone, A., Boellaard, R., Albert, N. L., Barthel, H., van Berckel, B., Brendel, M., Cecchin, D., Ekmekcioglu, O., Garibotto, V., et al. (2022) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-021-05603-w · Cited by 300
Automatic, three-segment, MR-based attenuation correction for whole-body PET/MR data
Schulz, V., Torres-Espallardo, I., Renisch, S., Hu, Z., Ojha, N., Börnert, P., Perkuhn, M., Niendorf, T., Schäfer, W. M., Brockmann, H., et al. (2011) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1603-1 · Cited by 292
Hofmann, M., Bezrukov, I., Mantlik, F., Aschoff, P., Steinke, F., Beyer, T., Pichler, B. J., & Schölkopf, B. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.110.078949 · Cited by 288
Self-citationSamarin, A., Burger, C., Wollenweber, S. D., Crook, D. W., Burger, I. A., Schmid, D. T., von Schulthess, G. K., & Kuhn, F. P. (2012) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-012-2113-0 · Cited by 251
Mehranian, A., Arabi, H., & Zaidi, H. (2016) · Medical Physics · doi:10.1118/1.4941014 · Cited by 152
Heusch, P., Buchbender, C., Beiderwellen, K., Nensa, F., Hartung-Knemeyer, V., Lauenstein, T. C., Bockisch, A., Forsting, M., Antoch, G., & Heusner, T. A. (2013) · European Journal of Radiology · doi:10.1016/j.ejrad.2013.01.008 · Cited by 96
Boellaard, R., Rausch, I., Beyer, T., Delso, G., Yaqub, M., Quick, H. H., & Sattler, B. (2015) · Medical Physics · doi:10.1118/1.4930962 · Cited by 62
Self-citation
Metadata
- Authors
- Hofmann, M., Pichler, B., Schölkopf, B., & Beyer, T.
- Venue
- European Journal of Nuclear Medicine and Molecular Imaging
- Published
- 2008-12-23
- Volume
- 36
- Issue
- S1
- Pages
- 93-104
- ISSN
- 1619-7070 (ISSN-L)
- 1619-7089
- OpenAlex
- W2002129123
- Cited by
- 343
- References
- 28
- FWCI
- 10.75
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.