EANM/EARL FDG-PET/CT accreditation - summary results from the first 200 accredited imaging systems
Open accessAbstract
PURPOSE: From 2010 until July 2016, the EANM Research Ltd. (EARL) FDG-PET/CT accreditation program has collected over 2500 phantom datasets from approximately 200 systems and 150 imaging sites worldwide. The objective of this study is to report the findings and impact of the accreditation program on the participating PET/CT systems. METHODS: To obtain and maintain EARL accredited status, sites were required to complete and submit two phantom scans - calibration quality control (CalQC), using a uniform cylindrical phantom and image quality control (IQQC), using a NEMA NU2-2007 body phantom. Average volumetric SUV bias and SUV recovery coefficients (RC) were calculated and the data evaluated on the basis of quality control (QC) type, approval status, PET/CT system manufacturer and submission order. RESULTS: SUV bias in 5% (n = 96) of all CalQC submissions (n = 1816) exceeded 10%. After corrective actions following EARL feedback, sites achieved 100% compliance within EARL specifications. 30% (n = 1381) of SUVmean and 23% (n = 1095) of SUVmax sphere recoveries from IQQC submissions failed to meet EARL accreditation criteria while after accreditation, failure rate decreased to 12% (n = 360) and 9% (n = 254), respectively. Most systems demonstrated longitudinal SUV bias reproducibility within ±5%, while RC values remained stable and generally within ±10% for the four largest and ±20% for the two smallest spheres. CONCLUSIONS: Regardless of manufacturer or model, all investigated systems are able to comply with the EARL specifications. Within the EARL accreditation program, gross PET/CT calibration errors are successfully identified and longitudinal variability in PET/CT performances reduced. The program demonstrates that a harmonising accreditation procedure is feasible and achievable.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Figures






Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Although all these errors are potentially severe, it should be noted that they can easily be avoided by following existing guidelines published by the EANM [24] or the EANM Research Ltd. (EARL) initiative [67].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [67]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0
Boellaard, R., Delgado-Bolton, R., Oyen, W. J. G., Giammarile, F., Tatsch, K., Eschner, W., Verzijlbergen, F. J., Barrington, S. F., Pike, L. C., Weber, W. A., et al. (2015) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-014-2961-x · Cited by 3,303
Self-citationRecommendations on the Use of 18 F-FDG PET in Oncology
Fletcher, J. W., Djulbegović, B., Soares, H. P., Siegel, B. A., Lowe, V. J., Lyman, G. H., Coleman, R. E., Wahl, R. L., Paschold, J. C., Avril, N. E., et al. (2008) · Journal of Nuclear Medicine · doi:10.2967/jnumed.107.047787 · Cited by 1,113
Standards for PET Image Acquisition and Quantitative Data Analysis
Boellaard, R. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.057182 · Cited by 906
Self-citationClinical biomarkers in drug discovery and development
Frank, R., & Hargreaves, R. J. (2003) · Nature Reviews Drug Discovery · doi:10.1038/nrd1130 · Cited by 807
Procedure guideline for tumor imaging with 18F-FDG PET/CT 1.0.
Delbeke, D., Coleman, R. E., Guiberteau, M. J., Brown, M. L., Royal, H. D., Siegel, B. A., Townsend, D. W., Berland, L. L., Parker, J. A., Hubner, K., et al. (2006) · PubMed · Cited by 689
Shankar, L. K., Hoffman, J. M., Bacharach, S. L., Graham, M. M., Karp, J. S., Lammertsma, A. A., Larson, S. M., Mankoff, D. A., Siegel, B. A., den Abbeele, A. V., et al. (2006) · PubMed · Cited by 677
A Systematic Review of the Factors Affecting Accuracy of SUV Measurements
Adams, M. C., Turkington, T. G., Wilson, J. M., & Wong, T. Z. (2010) · American Journal of Roentgenology · doi:10.2214/ajr.10.4923 · Cited by 662
Kinahan, P. E., & Fletcher, J. W. (2010) · Seminars in Ultrasound, CT and MRI · doi:10.1053/j.sult.2010.10.001 · Cited by 631
Procedure guideline for tumor imaging with F-18 FDG PET/CT
Delbeke, D., Coleman, R. E., Guiberteau, M. J., Brown, M. L., Royal, H. D., Siegel, B. A., Townsend, D. W., Berland, L. L., Parker, J. A., Hübner, K. F., et al. (2006) · Cited by 620
Understanding the standardized uptake value, its methods, and implications for usage.
Thie, J. A. (2004) · PubMed · Cited by 546
Research works printed but not linked: 1
- Kinahan P, Wahl R, Shao L, Frank R, Perlman E. The QIBA profile for quantitative FDG-PET/CT oncology imaging. J Nucl Med. 2014;55:1520.
Cited by
Sorted by citations
Swart, L. E., Gomes, A., Scholtens, A. M., Sinha, B. N. M., Tanis, W., Lam, M. G. E. H., van der Vlugt, M. J., Streukens, S. A. F., Aarntzen, E. H. J. G., Bucerius, J. A., et al. (2018) · Circulation · doi:10.1161/circulationaha.118.035032 · Cited by 196
EANM practice guideline for quantitative SPECT-CT
Dickson, J. C., Armstrong, I. S., Gabiña, P. M., Denis-Bacelar, A. M., Krizsan, A. K., Gear, J. I., WYNGAERT, T. V. D., de Geus‐Oei, L., & Herrmann, K. (2022) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-022-06028-9 · Cited by 125
Recent Developments in SPECT/CT
Ritt, P. (2022) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2022.01.004 · Cited by 108
Schöder, H. M., Polley, M. C., Knopp, M. V., Hall, N. C., Kostakoglu, L., Zhang, J., Higley, H. R., Kelloff, G. J., Liu, H., Zelenetz, A. D., et al. (2020) · Blood · doi:10.1182/blood.2019003277 · Cited by 95
Nuclear medicine and molecular imaging advances in the 21st century
Vaz, S. C., Oliveira, F., Herrmann, K., & Veit-Haibach, P. (2020) · The British Journal of Radiology · doi:10.1259/bjr.20200095 · Cited by 86
Metadata
- Authors
- Kaalep, A., Sera, T., Oyen, W., Krause, B. J., Chiti, A., Liu, Y., & Boellaard, R.
- Affiliations
- North Estonia Medical Centre (EE)
- Institute of Cancer Research (GB)
- Universitätsmedizin Rostock (DE)
- Humanitas University (IT)
- European Organisation for Research and Treatment of Cancer (NL)
- University Medical Center Groningen (NL)
- Corresponding author
- Kaalep, Andres
- Venue
- European Journal of Nuclear Medicine and Molecular Imaging
- Published
- 2017-12-01
- Volume
- 45
- Issue
- 3
- Pages
- 412-422
- ISSN
- 1619-7070 (ISSN-L)
- 1619-7089
- PMC
- PMC5787222
- OpenAlex
- W2773306697
- Cited by
- 174
- References
- 44
- Citations by year
- 2026: 18 · 2025: 30 · 2024: 20 · 2023: 22 · 2022: 22 · 2021: 20 · 2020: 14 · 2019: 13 · 2018: 15
- FWCI
- 8.82
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- European Journal of Nuclear Medicine and Molecular Imaging (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Pure Amsterdam UMC (publishedVersion, cc-by)
- SZTE Publicatio Repozitórium (University of Szeged) (acceptedVersion, other-oa)
- Archivio Istituzionale della Ricerca - Università Vita-Salute San Raffaele (submittedVersion, cc-by)
- Amsterdam UMC (VU Amsterdam) - Institutional Repository (submittedVersion, other-oa)
- University of Groningen research database (University of Groningen / Centre for Information Technology) (submittedVersion, cc-by)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Physical Sciences › Physics and Astronomy › Radiation › Advanced Radiotherapy Techniques
- Keywords
- quality control, image quality control, standardized uptake value
- MeSH terms
- Accreditation, Positron Emission Tomography Computed Tomography, Calibration, Nuclear Medicine, Societies, Medical, Phantoms, Imaging, Fluorodeoxyglucose F18
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review