A systematic review of the utility of 1·5 versus 3 Tesla magnetic resonance brain imaging in clinical practice and research
The DOI matches its Crossref record; 61 of the 62 references it cites were found in a registry or an index, and 1 are not checked yet.
Abstract
Objective MRI at 3 T is said to be more accurate than 1.5 T MR, but costs and other practical differences mean that it is unclear which to use.Methods We systematically reviewed studies comparing diagnostic accuracy at 3 T with 1.5 T. We searched MEDLINE, EMBASE and other sources from 1 January 2000 to 22 October 2010 for studies comparing diagnostic accuracy at 1.5 and 3 T in human neuroimaging. We extracted data on methodology, quality criteria, technical factors, subjects, signal-to-noise, diagnostic accuracy and errors according to QUADAS and STARD criteria.Results Amongst 150 studies (4,500 subjects), most were tiny, compared old 1.5 T with new 3 T technology, and only 22 (15 %) described diagnostic accuracy. The 3 T images were often described as "crisper", but we found little evidence of improved diagnosis. Improvements were limited to research applications [functional MRI (fMRI), spectroscopy, automated lesion detection]. Theoretical doubling of the signal-to-noise ratio was not confirmed, mostly being 25 %. Artefacts were worse and acquisitions took slightly longer at 3 T.Conclusion Objective evidence to guide MRI purchasing decisions and routine diagnostic use is lacking. Rigorous evaluation accuracy and practicalities of diagnostic imaging technologies should be the routine, as for pharmacological interventions, to improve effectiveness of healthcare.Key points • Higher field strength MRI may improve image quality and diagnostic accuracy. • There are few direct comparisons of 1.5 and 3 T MRI. • Theoretical doubling of the signal-to-noise ratio in practice was only 25 %. • Objective evidence of improved routine clinical diagnosis is lacking. • Other aspects of technology improved images more than field strength.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“A field strength of 3·0 T might be preferred, but images from modern 1·5 T MRI scanners are often of a similarly high definition,80 and are therefore acceptable”
From Neuroimaging standards for research into small vessel disease and its contribution to ageing and neurodegeneration (2013) · cites this paper as [80]
Citations
Cites
Towards complete and accurate reporting of studies of diagnostic accuracy: The STARD Initiative.
Bossuyt, P. M., Reitsma, J. B., Bruns, D. E., Gatsonis, C. A., Glasziou, P. P., Irwig, L. M., Lijmer, J. G., Moher, D., Rennie, D., de Vet, H. C. W., et al. (2003) · Annals of Internal Medicine · doi:10.7326/0003-4819-138-1-200301070-00010 · Cited by 3,219
Towards complete and accurate reporting of studies of diagnostic accuracy: the STARD initiative
Bossuyt, P. M., Reitsma, J. B., Bruns, D. E., Gatsonis, C. A., Glasziou, P. P., Irwig, L. M., Lijmer, J. G., Moher, D., Rennie, D., & de Vet, H. C. W. (2003) · BMJ · doi:10.1136/bmj.326.7379.41 · Cited by 3,215
Towards Complete and Accurate Reporting of Studies of Diagnostic Accuracy: The STARD Initiative
Bossuyt, P. M., Reitsma, J. B., Bruns, D. E., Gatsonis, C. A., Glasziou, P. P., Irwig, L. M., Lijmer, J. G., Moher, D., Rennie, D., & de Vet, H. C. W. (2003) · Radiology · doi:10.1148/radiol.2261021292 · Cited by 3,215
Exposure to Low-Dose Ionizing Radiation from Medical Imaging Procedures
Fazel, R., Krumholz, H. M., Wang, Y., Ross, J. S., Chen, J., Ting, H. H., Shah, N. D., Nasir, K., Einstein, A. J., & Nallamothu, B. K. (2009) · New England Journal of Medicine · doi:10.1056/nejmoa0901249 · Cited by 1,338
Evaluation of QUADAS, a tool for the quality assessment of diagnostic accuracy studies
Whiting, P. F., Weswood, M. E., Rutjes, A. W., Reitsma, J. B., Bossuyt, P. N., & Kleijnen, J. (2006) · BMC Medical Research Methodology · doi:10.1186/1471-2288-6-9 · Cited by 769
Characterizing iron deposition in multiple sclerosis lesions using susceptibility weighted imaging
Haacke, E. M., Makki, M., Ge, Y., Maheshwari, M., Sehgal, V., Hu, J., Selvan, M., Wu, Z., Latif, Z., Xuan, Y., et al. (2009) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21676 · Cited by 329 · cited 1×
“Of the 22 studies (15 %) that did provide some sensitivity and specificity information, 7 calculated accuracy on a per patient basis [17, 26–31], 9 calculated accuracy on a per lesion basis [32–40], 4 calculated both [16, 41–43] and in two studies, sensitivity and specificity were irrelevant [44, 45], precluding a for…”
BackgroundfMRI at 1.5, 3 and 7 T: Characterising BOLD signal changes
van der Zwaag, W., Francis, S., Head, K., Peters, A., Gowland, P., Morris, P., & Bowtell, R. (2009) · NeuroImage · doi:10.1016/j.neuroimage.2009.05.015 · Cited by 281 · cited 1×
“Of the 22 studies (15 %) that did provide some sensitivity and specificity information, 7 calculated accuracy on a per patient basis [17, 26–31], 9 calculated accuracy on a per lesion basis [32–40], 4 calculated both [16, 41–43] and in two studies, sensitivity and specificity were irrelevant [44, 45], precluding a for…”
BackgroundComparison of 18F-FET PET and 5-ALA fluorescence in cerebral gliomas
Floeth, F. W., Sabel, M., Ewelt, C., Stummer, W., Felsberg, J., Reifenberger, G., Steiger, H. J., Stoffels, G., Coenen, H. H., & Langen, K.-J. (2011) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1690-z · Cited by 281
3T phased array MRI improves the presurgical evaluation in focal epilepsies: A prospective study
Knake, S., Triantafyllou, C., Wald, L. L., Wiggins, G., Kirk, G. P., Larsson, P. G., Stufflebeam, S. M., Foley, M. T., Shiraishi, H., Dale, A. M., et al. (2005) · Neurology · doi:10.1212/01.wnl.0000179355.04481.3c · Cited by 225
Magnetic Resonance Imaging at 3.0 Tesla: Challenges and Advantages in Clinical Neurological Imaging
Frayne, R., Goodyear, B. G., Dickhoff, P., Lauzon, M. L., & Sevick, R. J. (2003) · Investigative Radiology · doi:10.1097/01.rli.0000073442.88269.c9 · Cited by 212 · cited 1×
“However, the potential price is more artefacts [3], greater magnetic field heterogeneity, increased vigilance required for patient screening, increased biological effects [4, 5] and increased cost of equipment purchase, installation, maintenance and operation.”
Background
Research works printed but not linked: 1
- Wellcome Trust (2011) Human functional brain imaging 1990–2009: portfolio review. Wellcome Trust, London
Cited by
Wardlaw, J. M., Smith, E. E., Biessels, G. J., Cordonnier, C., Fazekas, F., Frayne, R., Lindley, R. I., O'Brien, J. T., Barkhof, F., Benavente, O. R., et al. (2013) · The Lancet Neurology · doi:10.1016/s1474-4422(13)70124-8 · Cited by 5,544
Self-citationWhat are White Matter Hyperintensities Made of?: Relevance to Vascular Cognitive Impairment
Wardlaw, J. M., Valdés Hernández, M. C., & Muñoz‐Maniega, S. (2015) · Journal of the American Heart Association · doi:10.1161/jaha.114.001140 · Cited by 902
Self-citationSmith, E. E., Saposnik, G., Biessels, G. J., Doubal, F. N., Fornage, M., Gorelick, P. B., Greenberg, S. M., Higashida, R. T., Kasner, S. E., & Seshadri, S. (2017) · Stroke · doi:10.1161/str.0000000000000116 · Cited by 404
Clancy, Ú., Gilmartin, D., Jochems, A. C. C., Knox, L., Doubal, F., & Wardlaw, J. M. (2021) · The Lancet Psychiatry · doi:10.1016/s2215-0366(20)30431-4 · Cited by 182
Self-citationClinical Impact of Deep Learning Reconstruction in MRI
Kiryu, S., Akai, H., Yasaka, K., Tajima, T., Kunimatsu, A., Yoshioka, N., Akahane, M., Abe, O., & Ohtomo, K. (2023) · Radiographics · doi:10.1148/rg.220133 · Cited by 183
Patterns of functional connectivity in an aging population: The Rotterdam Study
Zonneveld, H. I., Pruim, R. H., Bos, D., Vrooman, H. A., Muetzel, R. L., Hofman, A., Rombouts, S. A., van der Lugt, A., Niessen, W. J., Ikram, M. A., et al. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2019.01.041 · Cited by 162
A naturalistic neuroimaging database for understanding the brain using ecological stimuli
Aliko, S., Huang, J., Gheorghiu, F., Meliß, S., & Skipper, J. I. (2020) · Scientific Data · doi:10.1038/s41597-020-00680-2 · Cited by 142
Perivascular Spaces, Glymphatic System and MR
Yu, L., Hu, X., Li, H., & Zhao, Y. (2022) · Frontiers in Neurology · doi:10.3389/fneur.2022.844938 · Cited by 126
Hirjak, D., Thomann, P. A., Kubera, K. M., Wolf, N. D., Sambataro, F., & Wolf, R. C. (2015) · Schizophrenia Research · doi:10.1016/j.schres.2015.10.022 · Cited by 120
“…have also shown the occurrence of NSS in healthy individuals (Hirjak et al., 2015b; Hirjak et al., 2014a; Thomann et al., 2015) suggesting that NSS might reflect a neurodevelopmental signature of motor system dysfunction, probably as a continuum between health and disease (Whitty et al., 2009).”
Zhu, X., Helpman, L., Papini, S., Schneier, F. R., Markowitz, J. C., Meter, P. E. V., Lindquist, M. A., Wager, T. D., & Neria, Y. (2016) · Depression and Anxiety · doi:10.1002/da.22594 · Cited by 100
“Although a 1.5 T MRI is still completely adequate for most fMRI scans (Wardlaw et al., 2012), our data warrant replication in higher resolution scanners with eye-tracking systems, larger samples, and longer scan time.”
BackgroundInfluentialWardlaw, J. M., Brazzelli, M. G., Miranda, H., Chappell, F. M., McNamee, P., Scotland, G., Quayyum, Z., Martin, D., Shuler, K., Sandercock, P. A., et al. (2014) · Health Technology Assessment · doi:10.3310/hta18270 · Cited by 99
Self-citationDe Guio, F., Jouvent, E., Biessels, G. J., Black, S. E., Brayne, C., Chen, C., Cordonnier, C., De Leeuw, F.-E., Dichgans, M., Doubal, F., et al. (2016) · Journal of Cerebral Blood Flow & Metabolism · doi:10.1177/0271678x16647396 · Cited by 92
“Indeed there are relatively few direct comparisons of different scanner field strengths in any situation.(41) In 10 patients with SVD, Theysohn et al.”
BackgroundSelf-citationBlair, G. W., Doubal, F., Thrippleton, M. J., Marshall, I., & Wardlaw, J. M. (2016) · Journal of Cerebral Blood Flow & Metabolism · doi:10.1177/0271678x16631756 · Cited by 93
“5 and 3T.(24) Therefore, more data are required to determine the true differences in BOLD signal and artefacts due to field strength.”
BackgroundSelf-citationManagement of Multi Organ Dysfunction in Neonatal Encephalopathy
O’Dea, M. I., Sweetman, D., Bonifacio, S. L., El‐Dib, M., Austin, T., & Molloy, E. J. (2020) · Frontiers in Pediatrics · doi:10.3389/fped.2020.00239 · Cited by 82
Krishnamurthy, U., Neelavalli, J., Mody, S. S., Yeo, L., Jella, P. K., Saleem, S., Korzeniewski, S. J., Cabrera, M. D., Ehterami, S., Bahado-Singh, R. O., et al. (2014) · Journal of Perinatal Medicine · doi:10.1515/jpm-2014-0268 · Cited by 71
“0T [83–86].”
Diffusion Tensor Imaging (DTI) Findings Following Pediatric Non-Penetrating TBI: A Meta-Analysis
Roberts, R., Mathias, J. L., & Rose, S. (2014) · Developmental Neuropsychology · doi:10.1080/87565641.2014.973958 · Cited by 67
Fullana, M. A., Zhu, X., Alonso, P., Cardoner, N., Real, E., López-Solà, C., Segalàs, C., Subirà, M., Galfalvy, H., Menchón, J. M., et al. (2017) · Journal of Psychiatry and Neuroscience · doi:10.1503/jpn.160215 · Cited by 63
White Matter Hyperintensities Are Under Strong Genetic Influence
Sachdev, P. S., Thalamuthu, A., Mather, K. A., Ames, D. J., Wright, M. J., Wen, W. X., Bowden, J. L., Lee, T., Brodaty, H., Crawford, J., et al. (2016) · Stroke · doi:10.1161/strokeaha.116.012532 · Cited by 60
Rua, C., Clarke, W. T., Driver, I. D., Mougin, O. E., Morgan, A., Clare, S., Francis, S., Muir, K. W., Wise, R. G., Carpenter, T. A., et al. (2020) · NeuroImage · doi:10.1016/j.neuroimage.2020.117358 · Cited by 57
Meijer, F. J. A., van Rumund, A., Fasen, B. A. C. M., Titulaer, I., Aerts, M., Esselink, R. A. J., Bloem, B. R., Verbeek, M. M., & Goraj, B. M. (2014) · American Journal of Neuroradiology · doi:10.3174/ajnr.a4140 · Cited by 57
“5T scanner.(36,37) On the basis of their 3T MR imaging SWI study, Haller et al(34) found increased susceptibility in the thalamus and left substantia nigra in PD, and they reported good discrimination between PD and AP by using a support-vector analysis.”
BackgroundReimer, C., Deike‐Hofmann, K., Graf, M., Reimer, P., Wiestler, B., Floca, R. O., Kickingereder, P., Schlemmer, H.-P., Wick, W., Bendszus, M., et al. (2017) · PLoS ONE · doi:10.1371/journal.pone.0174620 · Cited by 53
Impact of deep learning reconstruction on intracranial 1.5 T magnetic resonance angiography
Yasaka, K., Akai, H., Sugawara, H., Tajima, T., Akahane, M., Yoshioka, N., Kabasawa, H., Miyo, R., Ohtomo, K., Abe, O., et al. (2021) · Japanese Journal of Radiology · doi:10.1007/s11604-021-01225-2 · Cited by 50
Assessment of reduced field of view in diffusion tensor imaging of the lumbar nerve roots at 3 T
Budzik, J., Verclytte, S., Lefebvre, G., Monnet, A., Forzy, G., & Cotten, A. (2012) · European Radiology · doi:10.1007/s00330-012-2710-0 · Cited by 43
“5 T versus 3 T in the field of neuroradiology [17] reports that no convincing evidence of improved diagnostic accuracy or reduced total examination times was found.”
Background3 T: the good, the bad and the ugly
Graves, M. J. (2021) · British Journal of Radiology · doi:10.1259/bjr.20210708 · Cited by 41
Buchanan, C. R., Maniega, S. M., Hernández, M. V., Ballerini, L., Barclay, G., Taylor, A. M., Russ, T. C., Tucker–Drob, E. M., Wardlaw, J. M., Deary, I. J., et al. (2021) · Human Brain Mapping · doi:10.1002/hbm.25473 · Cited by 37
“5T technology in research settings (Wardlaw et al., 2012).”
MethodSelf-citationHasford, F., Mumuni, A. N., Trauernicht, C. J., Ige, T. A., Inkoom, S., Okeji, M. C., Nathaniel, E.-E. U., Khelassi-Toutaoui, N., Geraldo, M., Konfé, A., et al. (2022) · Physica Medica · doi:10.1016/j.ejmp.2022.09.016 · Cited by 32
“It has been reported elsewhere [49] that both 1.”
BackgroundSelf-citationBrain Development in Fetuses of Mothers with Diabetes: A Case-Control MR Imaging Study
Denison, F. C., Macnaught, G., Semple, S. I. K., Terris, G., Walker, J., Anblagan, D., Serag, A., Reynolds, R. M., & Boardman, J. P. (2017) · American Journal of Neuroradiology · doi:10.3174/ajnr.a5118 · Cited by 26
Hierarchical complexity of the adult human structural connectome
Smith, K., Bastin, M. E., Cox, S. R., Hernández, M. V., Wiseman, S. J., Escudero, J., & Sudlow, C. L. M. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2019.02.028 · Cited by 26
Self-citationYap, K. H., Manan, H. A., Yahya, N. A., Azmin, S., Mukari, S. A. M., & Ibrahim, N. M. (2022) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.859651 · Cited by 25
Cortical morphometry in anorexia nervosa: An out‐of‐sample replication study
Leppänen, J., Sedgewick, F., Cardi, V., Treasure, J. L., & Tchanturia, K. (2019) · European Eating Disorders Review · doi:10.1002/erv.2686 · Cited by 25
Predictors of Anesthetic Exposure in Pediatric MRI
Machado‐Rivas, F., Leitman, E. M., Jaimes, C. E., Conklin, J., Caruso, P. A., Liu, C. A., & Gee, M. S. (2020) · American Journal of Roentgenology · doi:10.2214/ajr.20.23601 · Cited by 23
High Resolution Three-Dimensional MR Imaging of the Skull Base
Blitz, A. M., Aygün, N., Herzka, D. A., Ishii, M., & Gallia, G. L. (2016) · Radiologic Clinics of North America · doi:10.1016/j.rcl.2016.08.011 · Cited by 23
The technological future of 7 T MRI hardware
WEBB, A. G., & de Moortele, P. V. (2015) · NMR in Biomedicine · doi:10.1002/nbm.3315 · Cited by 22
Life cycle costing as a decision making tool for technology acquisition in radio-diagnosis
Chakravarty, A., & Debnath, J. (2014) · Medical Journal Armed Forces India · doi:10.1016/j.mjafi.2014.10.004 · Cited by 21
“Another important issue that needs to be kept in mind is that doubling the magnetic field strength does not automatically translate into increasing advantages, though requiring much higher inputs in financial terms.(9,10) There are strengths and weaknesses of alternate decision models that need to be looked into befor…”
BackgroundSquarzoni, P., Duran, F., Busatto, G. F., & Alves, T. C. T. F. (2018) · Journal of Alzheimer s Disease · doi:10.3233/jad-161036 · Cited by 20
“However,itshouldbenotedthatimprovedperformance of3Teslascannershasbeenmoreclearlydemon-strated for functional rather than for morphometric MRI brain acquisitions to date [95].”
Basal ganglia volume and shape in anorexia nervosa
Leppänen, J., Cardi, V., Sedgewick, F., Treasure, J. L., & Tchanturia, K. (2019) · Appetite · doi:10.1016/j.appet.2019.104480 · Cited by 20
Caeyenberghs, K., Imms, P. E., Irimia, A., Monti, M. M., Esopenko, C., de Souza, N. L., D, J. F. D., Newsome, M. R., Dobryakova, E., Cwiek, A. P., et al. (2024) · NeuroImage Clinical · doi:10.1016/j.nicl.2024.103585 · Cited by 19
“Even though high-field MRI scanners (3T or more) increase the risk of susceptibility artifacts, they are preferred for rsfMRI studies as they provide better signal-to-noise ratio (SNR) and a higher spatial resolution (Wardlaw et al., 2012).”
Radbruch, A., Paech, D., Gassenmaier, S., Luetkens, J. A., Isaak, A., Herrmann, J., Othman, A. E., Schäfer, J. F., & Nikolaou, K. (2021) · Investigative Radiology · doi:10.1097/rli.0000000000000818 · Cited by 20
“5 T were smaller than 5mmor located in the periventricular zone.(30,31) Similar results were reported in patients with acute optic neuritis, where MRI at 3.”
ResultCortés, M. S., Bargalló, N., Arranz, Á., Simoes, R. V., Figueras, F., & Gratacós, E. (2016) · Fetal Diagnosis and Therapy · doi:10.1159/000445947 · Cited by 19
Clinical application of brain MRI in parkinsonism. From basic to advanced imaging
Meijer, F. J. A. (2015) · Radboud Repository (Radboud University) · Cited by 19
Vessel distance mapping: A novel methodology for assessing vascular-induced cognitive resilience
Garcia‐Garcia, B., Mattern, H., Vockert, N., Yakupov, R., Schreiber, F., Spallazzi, M., Perosa, V., Haghikia, A., Speck, O., Düzel, E., et al. (2023) · NeuroImage · doi:10.1016/j.neuroimage.2023.120094 · Cited by 18
Klenk, C., Gawande, R. S., Tran, V. T., Leung, J. T., Chi, K. W., Owen, D. R., Luna‐Fineman, S., Sakamoto, K. M., McMillan, A., Quon, A., et al. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.161646 · Cited by 17
“5-T systems differ in diagnostic accuracy for pediatric tumors (37).”
Backgroundvan Overbeek, E. C., Staals, J., & van Oostenbrugge, R. J. (2013) · PLoS ONE · doi:10.1371/journal.pone.0078100 · Cited by 17
“5T yielded largely unchanged conclusions (results not shown) [26].”
Radojewski, P., Piredda, G. F., Bonanno, G., Lövblad, K., Vargas, M. I., Sutter, R., Nanz, D., Karrer, T., Salanti, G., & Wiest, R. (2024) · European Journal of Neurology · doi:10.1111/ene.16557 · Cited by 15
“T MRI in 2012, Wardlaw et al. stated that “ob-jective evidence to guide MRI purchasing decisions and routine diagnostic use ( of 3T MRI ) is lacking” [ 12].”
Treatment of glioblastoma using MRIdian® A3i BrainTx™: Imaging and treatment workflow demonstration
Rosa, A. L., Mittauer, K. E., Rzepczynski, A. E., Chuong, M. D., Kütük, T., Bassiri, N., McAllister, N. C., Hall, M. D., McCulloch, J., Alvarez, D., et al. (2023) · Medical dosimetry · doi:10.1016/j.meddos.2023.02.001 · Cited by 14
Ippoliti, M., Adams, L. C., Brenner, W., Hamm, B., Spincemaille, P., Wang, Y., & Makowski, M. R. (2018) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26045 · Cited by 14
“5T is adequate for most examinations and still accounts for the majority of scanners that are currently in clinical use.(14,15) Therefore, QSM might need to be optimized for clinical use on 1.”
BackgroundFluschnik, N., Tahir, E., Erley, J., Müllerleile, K., Metzner, A., Wenzel, J., Guerreiro, H., Adam, G., Blankenberg, S., Kirchhof, P. F., et al. (2022) · EP Europace · doi:10.1093/europace/euac213 · Cited by 14
“MRI is increasingly performed at 3T to improve image quality.(10) Data on the safety of MRI imaging at 3T in patients with CIEDs are scarce.”
Chi, S., Wen, X., Yu, Y., Wang, G., Zhang, J., Xue, C., Zhang, X., Wang, Z., Gesang, M., Chen, J., et al. (2022) · Parkinsonism & Related Disorders · doi:10.1016/j.parkreldis.2022.11.010 · Cited by 13
Skouras, S., Gray, M. A., Critchley, H., & Koelsch, S. (2014) · NeuroImage · doi:10.1016/j.neuroimage.2014.07.007 · Cited by 13
Salice, S., Esposito, R., Ciavardelli, D., Pizzi, S. D., di Bastiano, R., & Tartaro, A. (2016) · PLoS ONE · doi:10.1371/journal.pone.0159047 · Cited by 12
Doniselli, F. M., Ramos, J. N., Hilario, A., Opančina, V., van der Hoorn, A., Castellano, A., van der Molen, A. J., Quattrocchi, C. C., Yousry, T. A., Rovira, À., et al. (2025) · European Radiology · doi:10.1007/s00330-025-11646-6 · Cited by 11
“Using 3 T instead of 1.5 T increases the signal-to-noise ratio by approximately 25% [95].”
InfluentialLi, L., Liu, W., Cai, Q., Liu, Y., Hu, W., Zuo, Z., Ma, Q., He, S., & Jin, K. (2023) · Frontiers in Immunology · doi:10.3389/fimmu.2023.1152235 · Cited by 11
“0-T MRI scanners for the routine clinical diagnosis of diseases, except for research applications, including functional MRI, spectroscopy, and automated lesion detection (35).”
Guarnaschelli, J. N., Vagal, A. S., McKenzie, J. T., McPherson, C. M., Warnick, R. E., Batra, V., Breneman, J. C., & Lamba, M. A. S. (2014) · Practical Radiation Oncology · doi:10.1016/j.prro.2013.11.003 · Cited by 11
Sinko, K., Czerny, C., Jagsch, R., Baumann, A., & Kulinna-Cosentini, C. (2015) · Dentomaxillofacial Radiology · doi:10.1259/dmfr.20150028 · Cited by 9
“While a higher field strength is expected to improve image quality, the theoretical doubling of the signal-to-noise ratio was no higher than 25%.(15) In agreement with our results, Debatin and Patak(16) conclude that 3.”
ResultWhat is the role of 3 T MRI in sports medicine? Revisiting the marriage after the honeymoon
Roemer, F. W., & Guermazi, A. (2015) · British Journal of Sports Medicine · doi:10.1136/bjsports-2015-095139 · Cited by 9
Stegmann, T., Joundi, R. A., Srivastava, A., Sinnecker, T., Aeschbacher, S., Eikelboom, J. W., Wachter, R., Smith, E. E., Rodondi, N., Bosch, J., et al. (2025) · European Heart Journal · doi:10.1093/eurheartj/ehaf550 · Cited by 8
Brain atrophy pattern in patients with mild cognitive impairment: MRI study
Calandrelli, R., Panfili, M., Onofrj, V., Tran, H. E., Piludu, F., Guglielmi, V., Colosimo, C., & Pilato, F. (2022) · Translational Neuroscience · doi:10.1515/tnsci-2022-0248 · Cited by 8
“5T and 3T scans do not significantly differ in their power to detect differences in quantitative values [56].”
BackgroundBing, F., Berger, I., Fabry, A., Moroni, A.-L., Casile, C., Morel, N., M’Biene, S., Guellerin, J., Pignal-Jacquard, C., Vadot, W., et al. (2021) · Revue Neurologique · doi:10.1016/j.neurol.2021.08.007 · Cited by 7
Loos, C., van Oostenbrugge, R. J., & Staals, J. (2012) · Cerebrovascular Diseases · doi:10.1159/000344003 · Cited by 7
Wattjes, M. P., & Barkhof, F. (2012) · European Radiology · doi:10.1007/s00330-012-2552-9 · Cited by 7
Automated Feature Extraction for Predicting Multiple Sclerosis Patient Disability using Brain MRI
Muslim, A. M., Mashohor, S., Mahmud, R., Gawwam, G. A., & Hanafi, M. B. (2022) · International Journal of Advanced Computer Science and Applications · doi:10.14569/ijacsa.2022.0130353 · Cited by 6
“First, different MRI Tesla provides different image characteristics in terms of clarity, details, and noise reduction and provides different amounts of signal received from the human body during an MRI scan [24].”
BackgroundMotiei, M., Mansouri, S. S., Tamimi, A., Farokhi, S., Fakouri, A., Rassam, K., Sedighi-Pirsaraei, N., & Rad, A. H. (2025) · International Journal of Surgery · doi:10.1097/js9.0000000000002588 · Cited by 6
“As it has been noted, the sensitivity and specificity of MRI at different field strengths are affected by the anatomical location of the tumor and technical considerations [ 39 ] , including imaging sequences, coil configurations, and the application of contrast agents [ 40 ] .”
Arrigoni, F., Calloni, S. F., Huisman, T. A. G. M., & Chiapparini, L. (2018) · Handbook of clinical neurology · doi:10.1016/b978-0-444-63956-1.00013-8 · Cited by 7
Neuroimaging Techniques for Neuroplasticity Quantification in Stroke Patients
Metropolitana, U. A., Rodríguez-García, M. E., Marín-Arriaga, N., Ibarra”, I. N. D. R. “. G. I., Macías-Arriaga, S. G., Ibarra”, I. N. D. R. “. G. I., Salazar-Cárdenas, B., Ibarra”, I. N. D. R. “. G. I., Ramírez-Rodríguez, T., Ibarra”, I. N. D. R. “. G. I., et al. (2023) · Revista Mexicana de Ingeniería Biomédica · doi:10.17488/rmib.44.2.5 · Cited by 6
Ghaderi, A., Wang, H., & Protzner, A. B. (2026) · Neural Computation · doi:10.1162/neco.a.1488 · Cited by 5
Advanced Magnetic Resonance Techniques: 3 T
Bhatti, L., Hoang, J. K., Dale, B. M., & Bashir, M. R. (2015) · Radiologic Clinics of North America · doi:10.1016/j.rcl.2014.12.001 · Cited by 5
Task-based modulation of functional connectivity of dorsal attention network in adult-ADHD
Metin, B., Kayaalp, S. D., Farhad, S., Çiftçi, E., Er, B. G., & Tarhan, N. (2024) · Neuroscience Letters · doi:10.1016/j.neulet.2024.137998 · Cited by 4
A molecular/genetic approach to cerebral small-vessel disease: Beyond aging and hypertension
Martini, S. R., Williams, S., Moretti, P. M., Woo, D., & Worrall, B. B. (2015) · Brain Circulation · doi:10.4103/2394-8108.166376 · Cited by 4
Lai, H. Y., Musrah, A. T. S., Yap, K. H., Chow, Y., Salim, N., & Voon, N. S. (2025) · Behavioural Brain Research · doi:10.1016/j.bbr.2025.115770 · Cited by 3
A High-Performance Conformal Metamaterial for 1.5 T Wrist Magnetic Resonance Imaging
Qi, Y., Chi, Z., Li, X., Wang, Y., Hu, W., Gao, J., Zheng, Z., Peng, R., Li, B., & Zhao, Q. C. (2026) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.6c09276 · Cited by 3
Moreno-Ayure, M., Páez, C., López-Arias, M. A., Mendez-Betancurt, J. L., Ordóñez-Rubiano, E. G., Rudas, J., Pulido, C., Gómez, F. A., Martínez, D., Enciso-Olivera, C. O., et al. (2020) · Medicine · doi:10.1097/md.0000000000021125 · Cited by 3
Utilization of Portable Brain Magnetic Resonance Imaging in an Acute Care Setting
Wang, A., Siddiqi, I., Marino, M. A., Maniakhina, L., Li, J. J., Ku, A., Ko, K., & Miulli, D. E. (2022) · Cureus · doi:10.7759/cureus.33067 · Cited by 3
Filippi, M., & Agosta, F. (2021) · Springer eBooks · doi:10.1007/978-3-030-66773-3_2 · Cited by 3
Wang, X. (2014) · ERA · Cited by 2
Dynamic MR Spectroscopy of brain metabolism using a non-conventional spectral averaging scheme
Mumuni, A. N., & McLean, J. (2016) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2016.12.011 · Cited by 1
Self-citationFilippi, M., & Agosta, F. (2021) · Springer eBooks · doi:10.1007/978-3-030-66773-3_1 · Cited by 1
3.0 T MRI Diagnostic Features: Comparison with Lower Magnetic Fields
Scarabino, T., Giannatempo, G. M., Pollice, S., Nasuto, M., Balzano, R. F., & Popolizio, T. (2017) · Springer eBooks · doi:10.1007/978-3-319-44174-0_3 · Cited by 1
Vessel distance mapping: a novel methodology for assessing vascular‐induced cognitive resilience
Garcia‐Garcia, B., Mattern, H., Vockert, N., Yakupov, R., Schreiber, F., Spallazzi, M., Perosa, V., Speck, O., Düzel, E., Maaß, A., et al. (2022) · Alzheimer s & Dementia · doi:10.1002/alz.063391 · Cited by 1
Developing a standardized tool for interpretation of radiology diagnostic accuracy trials
Schwarz, B. A. (2016) · Middlesex University Research Repository (Middlesex University Of London) · Cited by 1
Squarzoni, P. (2017) · Digital Library of Theses and Dissertations (Universidade de São Paulo) · doi:10.11606/t.5.2017.tde-07032017-162343 · Cited by 1
Wardlaw, J. M., Brazzelli, M. G., Miranda, H., Chappell, F. M., McNamee, P., Scotland, G., Quayyum, Z., Martin, D., Shuler, K., Sandercock, P. A., et al. (2015) · Health Technology Assessment · doi:10.3310/hta18270-c201509 · Cited by 1
Self-citationSingle voxel proton magnetic resonance spectroscopy of childhood brain tumours
Gill, S. K. (2014) · University of Birmingham Institutional Research Archive (University of Birmingham) · Cited by 1
Hypnosis versus Anesthesia: a study with children undergoing Magnetic Resonance Imaging procedures
Malafronte, M. (2021) · Rovira & Virgili repository · Cited by 1
Liu, Y. (2023) · White Rose eTheses Online (University of Leeds, The University of Sheffield, University of York) · Cited by 1
Autoimmune Diseases in Childhood
Karmazsin, L., Péter, F., & Szombathy, G. (2024) · Frontiers research topics · doi:10.3389/978-2-8325-5731-0
Setting the Report and Support of the Functional Findings
Tartaro, A., & Salice, S. (2017) · Springer eBooks · doi:10.1007/978-3-319-44174-0_13
Avancini, T. M. C. (2017) · Digital Library of Theses and Dissertations (Universidade de São Paulo) · doi:10.11606/t.5.2017.tde-06062017-103414
Kovačević-Dragosavljević, V., Katana, D., & Grujić-Vujmilović, D. (2017) · Биомедицинска истраживања · doi:10.7251/bii1701001v
Thrippleton, M. J., Parikh, J., Semple, S. I. K., Harris, B. A., Andrews, P., Wardlaw, J. M., & Marshall, I. (2017) · PLoS ONE · doi:10.1371/journal.pone.0189872
Self-citationKim, J. S., Kim, M., Kang, S. H., Oh, K., Suh, S., & Seo, W. (2018) · Precision and Future Medicine · doi:10.23838/pfm.2018.00114
“Lacunar infarction was also excluded, defined as following; size 3 to 15 mm, T2 high signal, FLAIR low signal with gliotic rim, and T1 low signal [9].”
Backgroundvan Overbeek (2018) · Research Publications (Maastricht University) · doi:10.26481/dis.20181102eo
Arana, E., & Arribas, L. (2020) · Neuro-Oncology · doi:10.1093/neuonc/noaa176
Rodrigues, L. (2020) · Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT)
Kaufmann, T. J., Smits, M., Boxerman, J. L., Huang, R. Y., Barboriak, D. P., Weller, M., Chung, C. D., Tsien, C. I., Brown, P. D., Shankar, L. K., et al. (2020) · Neuro-Oncology · doi:10.1093/neuonc/noaa202
El diagnóstico en la epilepsia
Navarro, J. O., Prior, M. J. A.-A., Singer, P. A., García, M. R., & Tejeror, E. D. (2023) · Medicine - Programa de Formación Médica Continuada Acreditado · doi:10.1016/j.med.2023.02.011
Assessing the Volume of the Head of the Mandibular Condyle Using 3T-MRI—A Preliminary Trial
Balma, A. M., Cavagnetto, D., Pavone, L., & Mussano, F. D. (2024) · Dentistry Journal · doi:10.3390/dj12070220
“…it is primarily evident that all the MRI replicates In this case, performing a 3D reconstruction from 3T-MRI sequences allowed a better definition of the bone edges thanks to the better resolution in soft tissue representation [30]. replicates show an underestimation of both the values of S and V.”
BackgroundEissa, L., ElBannan, S. M., Idriss, H. F., & Aziz, A. M. A. (2025) · The Egyptian Journal of Radiology and Nuclear Medicine · doi:10.1186/s43055-025-01517-x
“However, there are also many drawbacks as the potential price is more artifacts in the higher-field MRI scanners, greater magnetic field heterogeneity, increased biological effects, and increased cost of equipment purchase, installation, maintenance and operation [63].”
Durgut, B. D., Kamaşak, T., Kul, S., Arslan, E. A., Şahin, S., Dilber, B., & Cansu, A. (2025) · Turkish Journal of Pediatric Disease · doi:10.12956/tjpd.2025.1106
\nImaging and fluid biomarkers in uncertain cases of parkinsonism
van Rumund, A. (2023) · Radboud Repository (Radboud University)
Metadata
- Authors
- The SINAPSE Collaborative Group, Wardlaw, J. M., Brindle, W., Casado, A. M., Shuler, K., Henderson, M., Thomas, B., Macfarlane, J., Muñoz Maniega, S., Lymer, K., Morris, Z., Pernet, C., Nailon, W., Ahearn, T., Mumuni, A. N., Mugruza, C., McLean, J., Chakirova, G., Tao, Y., Simpson, J., Stanfield, A. C., Johnston, H., Parikh, J., Royle, N. A., De Wilde, J., Bastin, M. E., Weir, N., Farrall, A., & Valdes Hernandez, M. C.
- Affiliations
- Western General Hospital (GB)
- University of Edinburgh (GB)
- University of Dundee (GB)
- University of Aberdeen (GB)
- University of Glasgow (GB)
- University of Stirling (GB)
- University of St Andrews (GB)
- Higher Education Academy (GB)
- Edinburgh Royal Infirmary (GB)
- Corresponding author
- Wardlaw, Joanna M.
- ORCID
- Venue
- European Radiology
- Published
- 2012-06-09
- Volume
- 22
- Issue
- 11
- Pages
- 2295-2303
- ISSN
- 0938-7994 (ISSN-L)
- 1432-1084
- OpenAlex
- W1964031016
- Cited by
- 94
- References
- 61
- Citations by year
- 2026: 8 · 2025: 6 · 2024: 4 · 2023: 6 · 2022: 8 · 2021: 8 · 2020: 7 · 2019: 4 · 2018: 5 · 2017: 9 · 2016: 9 · 2015: 8 · 2014: 7 · 2013: 2 · 2012: 3
- FWCI
- 0.86
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- European Radiology (publishedVersion)
- PubMed (publishedVersion)
- Discovery Research Portal (University of Dundee) (publishedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiology practices and education
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Keywords
- 3T MRI, 1.5T MRI, brain MRI, diagnostic accuracy, signal-to-noise ratio, image artifacts, functional MRI, MRI spectroscopy, automated lesion detection, MRI field strength
- MeSH terms
- Brain, Brain Mapping, Diagnostic Imaging, Humans, Magnetic Resonance Imaging, Multiple Sclerosis, Neoplasms, Predictive Value of Tests, Research Design, Reproducibility of Results, Biomedical Research, Signal-To-Noise Ratio, Neuroimaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.