Validity, reliability and stability of the portable Cortex Metamax 3B gas analysis system
Open accessThe DOI matches its Crossref record; 24 of the 24 references it cites were found in a registry or an index.
Abstract
This study investigated the performance of the portable Cortex Metamax 3B (MM3B) automated gas analysis system during both simulated and human exercise using adolescents. Repeated measures using a Gas Exchange System Validator (GESV) across a range of simulated metabolic rates, showed the MM3B to be adequately reliable (both percentage errors, and percentage technical error of measurements <2%) for measuring expired ventilation (V (E)), oxygen consumption (VO(2)), and carbon dioxide production (VCO(2)). Over a 3 h period, the MM3B was shown to be acceptably stable in measuring gas fractions, as well as V (E), VO(2), and VCO(2) generated by the GESV, especially at moderate and high metabolic rates (drifts <2% and of minor physiological significance). Using eight healthy adolescents during rest, moderate, and vigorous cycle ergometry, the validity of the MM3B was tested against the primary criterion Douglas bag method (DBM) and a secondary criterion machine known to be accurate, the Jaeger Oxycon Pro system. No significant errors in V (E) were noted, yet the MM3B significantly overestimated both VO(2) and VCO(2) by approximately 10-17% at moderate and vigorous exercise as compared to the DBM and at all exercise levels compared to the Oxycon Pro. No significant differences were seen in any metabolic variable between the two criterion systems (DBM and Oxycon Pro). It is concluded the MM3B produces acceptably stable and reliable results, but is not adequately valid during moderate and vigorous exercise without some further correction of VO(2) and VCO(2).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Qualitative Methods in Public Health: A Field Guide for Applied Research
Ulin, P. R., Robinson, E. T., & Tolley, E. E. (2005) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000172593.20181.14 · Cited by 1,700
Wasserman, K. (2004) · Cited by 861
Durnin, J. V. G. A., & PASSMORE, R. (1967) · Heinemann Educational Books · Cited by 597
Complete Counterbalancing of Immediate Sequential Effects in a Latin Square Design
Bradley, J. V. (1958) · Journal of the American Statistical Association · doi:10.1080/01621459.1958.10501456 · Cited by 343 · cited 1×
“The DBM collections of expired gas were used as the primary criterion and the order of measurement system followed a counterbalanced Latin-Square process so as to avoid any order effect (Bradley 1958).”
MethodAutomated Metabolic Gas Analysis Systems: A Review
Macfarlane, D. J. (2001) · Sports Medicine · doi:10.2165/00007256-200131120-00002 · Cited by 198
Self-citationValidity and reliability of the Cortex MetaMax3B portable metabolic system
Vogler, A. J., Rice, A. J., & Gore, C. J. (2010) · Journal of Sports Sciences · doi:10.1080/02640410903582776 · Cited by 134 · cited 13×
“The relatively large percentage errors reported in our study for VO2 and VE by the MM3B are generally higher than those reported in the above validation studies, and also exceed the 4–10% guidelines recommended by some (Brehm et al. 2004; Laurent et al. 2008; Vogler et al. 2010), although these limits are not universa…”
BackgroundMethodResultInfluentialValidity and Stability of a Computerized Metabolic System with Mixing Chamber
Foss, Ø., & Hallén, J. (2005) · International Journal of Sports Medicine · doi:10.1055/s-2004-821317 · Cited by 127 · cited 2×
“The Oxycon Pro system acted as a ‘‘secondary criterion’’ as it has been shown to be a valid metabolic system in its mixing-chamber mode (Foss and Hallen 2005).”
BackgroundResultA respiratory gas exchange simulator for routine calibration in metabolic studies
Huszczuk, A., Whipp, B., & Wasserman, K. (1990) · European Respiratory Journal · doi:10.1183/09031936.93.03040465 · Cited by 119 · cited 1×
“The GESV (MedGraphics, MA, USA; a similar, but updated, GESV is now sold by Vacumed, CA, USA) was a mechanical gas exchange system simulating human breathing and has a reported accuracy in producing VO2 and VCO2 of ±2% (Huszczuk et al. 1990).”
BackgroundExamination of the Metamax I and II oxygen analysers during exercise studies in the laboratory
Medbø, J. I., Mamen, A., Welde, B., von Heimburg, E. D., & Stokke, R. (2002) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/003655102764654321 · Cited by 113 · cited 1×
“tem (Cortex, Leipzig, Germany), that is also marketed as the VmaxST in many countries, which supersedes the earlier Metamax I and II models that were shown to be valid and reliable (Medbo et al. 2002).”
BackgroundRenal Function and Plasma Volume Following Ultramarathon Cycling
Neumayr, G., Pfister, R., Hoertnagl, H., Mitterbauer, G., Prokop, W., & Joannidis, M. (2005) · International Journal of Sports Medicine · doi:10.1055/s-2004-815717 · Cited by 98
Validation of the portable VmaxST system for oxygen-uptake measurement.
Brehm, M., Harlaar, J., & Groepenhof, H. (2004) · Gait & Posture · doi:10.1016/s0966-6362(03)00097-3 · Cited by 94 · cited 7×
“The relatively large percentage errors reported in our study for VO2 and VE by the MM3B are generally higher than those reported in the above validation studies, and also exceed the 4–10% guidelines recommended by some (Brehm et al. 2004; Laurent et al. 2008; Vogler et al. 2010), although these limits are not universa…”
MethodResultInfluentialValidity and reliability of selected commercially available metabolic analyzer systems
Hodges, L. D., Brodie, D. A., & Bromley, P. D. (2005) · Scandinavian Journal of Medicine and Science in Sports · doi:10.1111/j.1600-0838.2005.00477.x · Cited by 93 · cited 2×
“…studies on the MM3B have been performed against the DBM (Brehm et al. 2004; Prieur et al. 2003; Vogler et al. 2010), and despite its limitations the DBM is often considered to be the criterion (Hodges et al. 2005), especially when precise metabolic calibrators are not available (Gore et al. 1997).”
BackgroundAmbulatory Gas Exchange Measurements - Current Status and Future Options
Meyer, T., Davison, R., & Kindermann, W. (2004) · International Journal of Sports Medicine · doi:10.1055/s-2004-830507 · Cited by 82 · cited 3×
“Meyer et al. (2005) reported that despite the need for establishing the reliability of portable gas analysis devices, few studies have addressed this issue; however, three studies appear to have reported data on the reliability of the Metamax 3B.”
BackgroundMethodAutomated ??VO2max calibrator for open-circuit indirect calorimetry systems
GORE, C. J., CATCHESIDE, P. G., FRENCH, S. N., BENNETT, J. M., & LAFORGIA, J. (1997) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-199708000-00016 · Cited by 62 · cited 3×
“…studies on the MM3B have been performed against the DBM (Brehm et al. 2004; Prieur et al. 2003; Vogler et al. 2010), and despite its limitations the DBM is often considered to be the criterion (Hodges et al. 2005), especially when precise metabolic calibrators are not available (Gore et al. 1997).”
BackgroundMethodPerformance Characteristics of Gas Analysis Systems: What We Know and What We Need to Know
Atkinson, G., Davison, R. C. R., & Nevill, A. M. (2005) · International Journal of Sports Medicine · doi:10.1055/s-2004-830505 · Cited by 59
Ein tragbarer Apparat zur Bestimmung des Grasstoffwechsels
Kofrányi, E., & Michaelis, H. (1940) · European Journal of Applied Physiology · doi:10.1007/bf02015045 · Cited by 53
Portable respiratory gas collection equipment.
Daniels, J. (1971) · Journal of Applied Physiology · doi:10.1152/jappl.1971.31.1.164 · Cited by 41 · cited 1×
“The DBM can be used in field trials (Daniels 1971), but it remains very limited due to its bulk, added air-resistance, and inability to easily undertake sequential measurements (Durnin and Passmore 1967).”
MethodA Methodology to Assess the Accuracy of a Portable Metabolic System (VmaxSTTM)
Prieur, F., Castells, J., & Denis, C. (2003) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000065003.82941.b0 · Cited by 37 · cited 14×
“over a 20 min period (Prieur et al. 2003), but for some studies, the calibration of a portable system in the field may be problematic due to the need for carrying bulky calibration equipment, including large pressurized bottles of alpha-standard calibration gases.”
BackgroundMethodResultInfluentialValidity of the Oxycon Mobile metabolic system under field measuring conditions
Eriksson, J. S., Rosdahl, H. G., & Schantz, P. (2011) · European Journal of Applied Physiology · doi:10.1007/s00421-011-1985-1 · Cited by 35 · cited 1×
“Using eight healthy adolescents during rest, moderate, and vigorous cycle ergometry, the validity of the MM3B was tested against the primary criterion Douglas bag method (DBM) and a secondary criterion machine known to be accurate, the Jaeger Oxycon Pro system.”
MethodA versatile system for measuring oxygen consumption in man.
Johnson, R. E., Robbins, F., Schilke, R., Mole, P. A., Harris, J., & Wakat, D. (1967) · Journal of Applied Physiology · doi:10.1152/jappl.1967.22.2.377 · Cited by 26 · cited 2×
“the Second World War and often referred to as the Kofranyi–Michaelis respirometer (Johnson et al. 1967) after the authors of an early publication (Kofranyi and Michaelis”
BackgroundMethodValidation of the Sensormedics (S2900Z) Metabolic Cart for Pediatric Exercise Testing
Unnithan, V. B., Wilson, J., Buchanan, D., et al. (1994) · Canadian Journal of Applied Physiology · doi:10.1139/h94-039 · Cited by 14 · cited 2×
“Furthermore, it is important to match the type of validation as to how the testing system will be used (Unnithan et al. 1994), and although many automated metabolic systems are used on pediatric/adolescent populations, few systems have been actually validated using these groups (Unnithan et al.…”
BackgroundValidity of the VmaxST portable metabolic measurement system
Laurent, C. M., Meyers, M. C., Robinson, C. A., Strong, L. R., Chase, C. M., & Goodwin, B. G. (2008) · Journal of Sports Sciences · doi:10.1080/02640410701758685 · Cited by 13 · cited 2×
“One of the earliest portable systems was the fully mechanical Max-Planck respirometer, developed during the Second World War and often referred to as the Kofranyi–Michaelis respirometer (Johnson et al. 1967) after the authors of an early publication (Kofranyi and Michaelis 1949).”
BackgroundMethodReliability and Validity of the VmaxST Portable Metabolic Analyzer
Perkins, C. D., Pivarnik, J., & Green, M. R. (2004) · Journal of Physical Activity and Health · doi:10.1123/jpah.1.4.413 · Cited by 12 · cited 7×
“The study by Perkins et al. (2004) examined reliability of the MM3B but used repeated measurements on human participants, which inflates the variability as it combines the relatively large biological error of the participants and the smaller technical error of the machine [their contributions to…”
BackgroundMethodResultInfluential
Cited by
Wahl, Y., Düking, P., Droszez, A., Wahl, P., & Mester, J. (2017) · Frontiers in Physiology · doi:10.3389/fphys.2017.00725 · Cited by 133
Energy system contribution during competitive cross-country skiing
Losnegard, T. (2019) · European Journal of Applied Physiology · doi:10.1007/s00421-019-04158-x · Cited by 99
Validity and Reliability of the VO2 Master Pro for Oxygen Consumption and Ventilation Assessment
(2020) · International Journal of Exercise Science · doi:10.70252/thjt1177 · Cited by 39
Beltrami, F. G., Froyd, C., Mamen, A., & Noakes, T. D. (2014) · European Journal of Applied Physiology · doi:10.1007/s00421-014-2825-x · Cited by 5
Metadata
- Authors
- Macfarlane, D. J., & Wong, P.
- Venue
- European Journal of Applied Physiology
- Published
- 2011-11-11
- Volume
- 112
- Issue
- 7
- Pages
- 2539-2547
- ISSN
- 0020-9376 (ISSN-L)
- 1439-6319
- 1439-6327
- OpenAlex
- W2076406139
- Cited by
- 167
- References
- 24
- FWCI
- 10.76
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.