Photoplethysmography and its application in clinical physiological measurement

Open access
Allen, John·2007·Physiological Measurement

Abstract

Photoplethysmography (PPG) is a simple and low-cost optical technique that can be used to detect blood volume changes in the microvascular bed of tissue. It is often used non-invasively to make measurements at the skin surface. The PPG waveform comprises a pulsatile ('AC') physiological waveform attributed to cardiac synchronous changes in the blood volume with each heart beat, and is superimposed on a slowly varying ('DC') baseline with various lower frequency components attributed to respiration, sympathetic nervous system activity and thermoregulation. Although the origins of the components of the PPG signal are not fully understood, it is generally accepted that they can provide valuable information about the cardiovascular system. There has been a resurgence of interest in the technique in recent years, driven by the demand for low cost, simple and portable technology for the primary care and community based clinical settings, the wide availability of low cost and small semiconductor components, and the advancement of computer-based pulse wave analysis techniques. The PPG technology has been used in a wide range of commercially available medical devices for measuring oxygen saturation, blood pressure and cardiac output, assessing autonomic function and also detecting peripheral vascular disease. The introductory sections of the topical review describe the basic principle of operation and interaction of light with tissue, early and recent history of PPG, instrumentation, measurement protocol, and pulse wave analysis. The review then focuses on the applications of PPG in clinical physiological measurements, including clinical physiological monitoring, vascular assessment and autonomic function.

Integrity signals

Registry

Crossref · 10.1088/0967-3334/28/3/r01

Cites

277 of 277 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • PPG sensors used to measure HR are liable to poor accuracy during activities with increased physical exertion or activities involving repetitive contractions of forearm skeletal muscles.

    “Prior research has suggested that PPG sensors used to measure the HR are liable to poor accuracy during activities with increased physical exertion or activities involving repetitive contractions of forearm skeletal muscles [36-38].”
  • During activities involving sustained muscle contractions or higher intensity exercises, the contact between the device's PPG sensor and skin is decreased, leading to a disruption in the signal quality and causing poor quality data.

    “It has been suggested that during activities involving sustained muscle contractions or higher intensity exercises, the contact between the device’s PPG sensor and skin is decreased, leading to a disruption in the signal quality and causing poor quality data [36,37].”
  • ECG monitoring via wearable sensors poses challenges for athletics: surface EMG, increased electrode impedance, respiration induced baseline drift, and electrode contact movement can cause noise and motion artifacts.

    “Firstly, surface EMG, increased electrode impedance, respiration induced baseline drift, and electrode contact movement can cause noise and motion artifacts.88”
  • Intrinsic PPG disturbances originating from physiological processes such as respiration, vasoconstriction due to temperature changes or stress have not been considered.

    “However, the intrinsic PPG disturbances originating from physiological processes, such as respiration, vasoconstriction due to temperature changes or stress [57], have not been considered.”
  • The photoplethysmogram (PPG) signal is an optical measure of arterial blood volume that fluctuates with each heartbeat and is strongly influenced by the heart and blood vessels.

    “The photoplethysmogram (PPG) signal is a measure of arterial blood volume, which fluctuates with each heartbeat and is used by many wearables to monitor heart rate (HR).”

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Allen, J.
Affiliations
Freeman Hospital (GB)
Corresponding author
Allen, John
Venue
Physiological Measurement
Published
2007-02-20
Volume
28
Issue
3
Pages
R1-R39
ISSN
  • 0967-3334 (ISSN-L)
  • 1361-6579
OpenAlex
W2083872334
Cited by
3,967
References
275
Citations by year
2026: 308 · 2025: 292 · 2024: 310 · 2023: 317 · 2022: 323 · 2021: 314 · 2020: 291 · 2019: 304 · 2018: 299 · 2017: 227 · 2016: 185 · 2015: 216 · 2014: 159 · 2013: 132 · 2012: 86
FWCI
3.44
Type
journal article
Language
en
Open access
Locations
  • Physiological Measurement (publishedVersion)
  • PubMed (publishedVersion)
  • Newcastle University ePrints (Newcastle Univesity)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
Keywords
photoplethysmography, blood volume changes, pulse wave analysis, non-invasive physiological monitoring, oxygen saturation, blood pressure measurement, cardiac output, autonomic function, peripheral vascular disease, vascular assessment, light-tissue interaction
MeSH terms
Autonomic Nervous System, Blood Vessels, Humans, Monitoring, Physiologic, Vascular Diseases, Photoplethysmography

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review