The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis

Open access

Abstract

Not much is known about how accurate and reproducible different thermometers are at diagnosing patients with suspected fever. The study aims at evaluating which peripheral thermometers are more accurate and reproducible. We searched Medline, Embase, Scopus, WOS, CENTRAL, and Cinahl to perform: (1) diagnostic accuracy meta-analysis (MA) using rectal mercury-in-glass or digital thermometry as reference, and bivariate models for pooling; (2) network MA to estimate differences in mean temperature between devices; (3) Bland-Altman method to estimate 95% coefficient of reproducibility. PROSPERO registration: CRD42020174996. We included 46 studies enrolling more than 12,000 patients. Using 38 °C (100.4 ℉) as cut-off temperature, temporal infrared thermometry had a sensitivity of 0.76 (95% confidence interval, 0.65, 0.84; low certainty) and specificity of 0.96 (0.92, 0.98; moderate certainty); tympanic infrared thermometry had a sensitivity of 0.77 (0.60, 0.88; low certainty) and specificity of 0.98 (0.95, 0.99; moderate certainty). For all the other index devices, it was not possible to pool the estimates. Compared to the rectal mercury-in-glass thermometer, mean temperature differences were not statistically different from zero for temporal or tympanic infrared thermometry; the median coefficient of reproducibility ranged between 0.53 °C [0.95 ℉] for infrared temporal and 1.2 °C [2.16 ℉] for axillary digital thermometry. Several peripheral thermometers proved specific, but not sensitive for diagnosing fever with rectal thermometry as a reference standard, meaning that finding a temperature below 38 °C does not rule out fever. Fixed differences between temperatures together with random error means facing differences between measurements in the order of 2 °C [4.5 ℉]. This study informs practitioners of the limitations associated with different thermometers; peripheral ones are specific but not sensitive.

Overview

CoChat Summary

## TL;DR

This systematic review and network meta-analysis found that peripheral thermometers are generally specific but not sensitive for detecting fever when compared with rectal thermometry at a 38 °C threshold. Temporal artery infrared and tympanic infrared thermometers performed best overall, but substantial random error remained across devices.

## Key findings

  • The review included 46 studies with 12,602 patients; 43 studies contributed to quantitative analyses.
  • For fever detection at 38 °C, temporal artery infrared thermometry had pooled sensitivity 0.76 (95% CI 0.65–0.84) and specificity 0.96 (0.92–0.98) in 9 studies (2,533 participants; 885 with fever).
  • For fever detection at 38 °C, tympanic infrared thermometry had pooled sensitivity 0.77 (0.60–0.88) and specificity 0.98 (0.95–0.99) in 9 studies (3,862 participants; 1,279 with fever).
  • Compared with rectal mercury-in-glass thermometry, axillary digital temperature was significantly lower by −0.67 °C (−0.98, −0.37), and axillary mercury-in-glass by −0.55 °C (−0.87, −0.23); the median 95% coefficient of reproducibility ranged from 0.53 °C for infrared temporal to 1.2 °C for axillary digital thermometry.
  • The authors report that the mean 95% coefficient of reproducibility across 73 direct comparisons was 1.06 °C, and that fixed plus random error could lead to differences of about 2 °C in the worst case.

## Methods

  • Systematic review and meta-analysis conducted according to Cochrane DTA guidance and reported per PRISMA-DTA; registered on PROSPERO CRD42020174996.
  • Searched Medline, Embase, Web of Science, Scopus, CENTRAL, and CINAHL up to March 2020 for studies of adults or children in emergency, hospital inpatient, or neonatal settings comparing thermometer types.
  • Used bivariate meta-analysis for diagnostic accuracy, and network meta-analysis for mean differences between thermometer types; analyses were centered on a 38 °C fever threshold and rectal mercury-in-glass or digital thermometry as reference standard.

## Limitations

  • The paper states that there were too few studies to estimate accuracy for some devices/sites and settings, including a specific setting such as the ED.
  • The authors state that overall certainty was moderate because of limitations in study design and considerable heterogeneity across studies, including different body sites, mixed child/adult populations, threshold effects, and operator variability.
  • Evident from the text: many studies had unclear enrollment, limited blinding, and some thermometers/sites had only one to three studies, which restricts precision and generalizability.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1007/s11739-020-02556-0

Cites

67 of 67 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Fig. 1 PRISMA flow diagram
Fig. 1 PRISMA flow diagramCC BY · PubMed Central
Fig. 2 Forest plots of the accuracy of infrared thermometer at the temporal and the tympanic sites (index devices) versus mercury-in-glass or digital thermometer at the rectal site (reference standard device) among studies using 38 °C as cut-off values of temperature for index and reference standard devices. TP true positive, FP false positive, FN false negative, TN true negative, CI confidence interval
Fig. 2 Forest plots of the accuracy of infrared thermometer at the temporal and the tympanic sites (index devices) versus mercury-in-glass or digital thermometer at the rectal site (reference standard device) among studies using 38 °C as cut-off values of temperature for index and reference standard devices. TP true positive, FP false positive, FN false negative, TN true negative, CI confidence intervalCC BY · PubMed Central
Fig. 3 Forest plot of mean differences among thermometers at different anatomical sites from network meta-analysis
Fig. 3 Forest plot of mean differences among thermometers at different anatomical sites from network meta-analysisCC BY · PubMed Central
Fig. 4 Forest plot of 95% coefficient of reproducibility (95% CR) of thermometers at different anatomical sites from the meta-analysis
Fig. 4 Forest plot of 95% coefficient of reproducibility (95% CR) of thermometers at different anatomical sites from the meta-analysisCC BY · PubMed Central

Verification

Claims the paper makes

4 Valid · 0 Not valid · 7 Not checked · 41 Own results and interpretations

13 more claims

Create a free account to see all 52 claims and to verify the 2 references still open.

See all claims — free account
  • Probably not a substantial claim. 17
  • It may state two separate claims as one. 23

Create a free account to see these claims.

Citations

Cites

Show all 69 references

Cited by

See all 49

Metadata

Authors
Pecoraro, V., Petri, D., Costantino, G., Squizzato, A., Moja, L., Virgili, G., & Lucenteforte, E.
Affiliations
  • University of Pisa (IT)
  • University of Milan (IT)
  • University of Insubria (IT)
  • Azienda Ospedaliero-Universitaria Careggi (IT)
Corresponding author
Lucenteforte, Ersilia
Venue
Internal and Emergency Medicine
Published
2020-11-25
Volume
16
Issue
4
Pages
1071-1083
ISSN
  • 1828-0447 (ISSN-L)
  • 1970-9366
OpenAlex
W3107110570
Cited by
49
References
69
Citations by year
2026: 9 · 2025: 10 · 2024: 9 · 2023: 6 · 2022: 12 · 2021: 3
FWCI
0.77
Type
journal article
Language
en
Open access
Licence
Locations
  • Internal and Emergency Medicine (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • Florence Research (University of Florence) (submittedVersion, cc-by)
  • Archivio Istituzionale della Ricerca (Universita Degli Studi Di Milano) (submittedVersion, other-oa)
  • CINECA IRIS Institutial research information system (University of Pisa) (submittedVersion)
  • IrInSubria (University of Insubria) (submittedVersion)
Funders
  • Fondazione Cassa si Risparmio di Firenze (24383)
  • Università di Pisa
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
  • Health Sciences › Medicine › Critical Care and Intensive Care Medicine › Thermal Regulation in Medicine
  • Health Sciences › Medicine › Physiology (medical) › Thermoregulation and physiological responses
Keywords
body temperature measurement, fever diagnosis, infrared thermometry, tympanic thermometer, digital thermometer, mercury-in-glass thermometer, rectal thermometry, diagnostic accuracy, reproducibility, sensitivity and specificity
MeSH terms
Body Temperature, Diagnostic Tests, Routine, Equipment Design, Humans, Sensitivity and Specificity, Thermometers, Reproducibility of Results

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis. https://app.cochat.ai/papers/10.1007/s11739-020-02556-0
The paper
Pecoraro, V., Petri, D., Costantino, G., Squizzato, A., Moja, L., Virgili, G., & Lucenteforte, E. (2021). The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis. Internal and Emergency Medicine. https://doi.org/10.1007/s11739-020-02556-0

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.