Digital PET/CT allows for shorter acquisition protocols or reduced radiopharmaceutical dose in [18F]-FDG PET/CT
Open accessThe DOI matches its Crossref record; 27 of the 27 references it cites were found in a registry or an index.
Abstract
Purpose: To establish the feasibility of shorter acquisition times (and by analogy, applied activity) on tumour detection and lesion contrast in digital PET/CT. Methods: Twenty-one randomly selected patients who underwent oncological [18 F]-FDG PET/CT on a digital PET/CT were retrospectively evaluated. Scan data were anonymously obtained and reconstructed in list-mode acquisition for a standard 2 min/bed position (bp), 1 min/bp and 30 s/bp (100%, 50% and 25% time or applied activity, respectively). Scans were randomized and read by two nuclear medicine physicians in a consensus read. Readers were blind to clinical details. Scans were evaluated for the number of pathological lesions detected. Measured uptake for lesions was evaluated by maximum and mean standardized uptake value (SUVmax and SUVmean, respectively) and tumour-to-backround ratio (TBR) were compared. Agreement between the three acquisitions was compared by Krippendorf’s alpha. Results: Overall n = 100 lesions were identified in the 2 min and 1 min/bp acquisitions and n = 98 lesions in the 30 s/bp acquisitions. Agreement between the three acquisitions with respect to lesion number and tumour-to-background ratio showed almost perfect agreement (K’s α = 0.999). SUVmax, SUVmean and TBR likewise showed > 98% agreement, with longer acquisitions being associated with slightly higher mean TBR (2 min/bp 7.94 ± 4.41 versus 30 s/bp 7.84 ± 4.22, p < 0.05). Conclusion: Shorter acquisition times have traditionally been associated with reduced lesion detectability or the requirement for larger amounts of radiotracer activity. These data confirm that this is not the case for new-generation digital PET scanners, where the known higher sensitivity results in clinically adequate images for shorter acquisitions. Only a small variation in the semi-quantitative parameters SUVmax, SUVmean and TBR was seen, confirming that either reduction of acquisition time or (by analogy) applied activity can be reduced as much as 75% in digital PET/CT without apparent clinical detriment.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures
![Fig. 1 Tumour-to-background ratio for the three acquisitions [30 s/bed position (bp); 1 min/bp; 2 min/bp]. The bars represent the standard error in the mean. No clear relationship between list-mode acquisition length and TBR was observed](https://figures.cochat.ai/paper-figures/b21fd7e003914da185ec6f288bbd9fe5/digital-pet-ct-allows-for-shorter-acquisition-protocols-or-reduced-figure-1.jpg)



Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“This is especially true for SUVmean and, to a lesser degree, for SUVpeak, which are both relatively stable even after reduction of acquisition time to 50% or less [76,77,78,79,80].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [79]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, standardized reading criteria do not negate the dissimilarities in the images obtained using different types of PET hardware and methods of image reconstruction, and both SiPM technology [79] and PSF reconstruction [225] have been found to result in systematically higher lesion conspicuity despite standardized reporting criteria (PSMA-RADS) [216].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [79]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“However, systematic SUVmax increases of >5–10% have not typically been observed at acquisition times per bed position of >30–60s, at least during investigations of SiPM-equipped systems and/or patients with BMI <25 kg/m2 [79,81,82].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [79]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
From RECIST to PERCIST: Evolving Considerations for PET Response Criteria in Solid Tumors
Wahl, R. L., Jacene, H., Kasamon, Y., & Lodge, M. A. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.057307 · Cited by 3,777
FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0
Boellaard, R., Delgado-Bolton, R., Oyen, W. J. G., Giammarile, F., Tatsch, K., Eschner, W., Verzijlbergen, F. J., Barrington, S. F., Pike, L. C., Weber, W. A., et al. (2015) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-014-2961-x · Cited by 3,303
Performance Characteristics of the Digital Biograph Vision PET/CT System
van Sluis, J., de Jong, J., Schaar, J., Noordzij, W., van Snick, P., Dierckx, R., Borra, R., Willemsen, A., & Boellaard, R. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215418 · Cited by 484
Noise Considerations for PET Quantification Using Maximum and Peak Standardized Uptake Value
Lodge, M. A., Chaudhry, M. A., & Wahl, R. L. (2012) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.101733 · Cited by 226
Marcus, C., Ciarallo, A., Tahari, A. K., Mena, E., Koch, W., Wahl, R. L., Kiess, A. P., Kang, H., & Subramaniam, R. M. (2014) · Journal of Nuclear Medicine · doi:10.2967/jnumed.113.136796 · Cited by 191
Performance Evaluation of the Vereos PET/CT System According to the NEMA NU2-2012 Standard
Rausch, I., Ruiz, A., Valverde-Pascual, I., Cal-González, J., Beyer, T., & Carrio, I. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215541 · Cited by 168
Nguyen, N. C., Vercher-Conejero, J. L., Sattar, A., Miller, M. A., Maniawski, P. J., Jordan, D. W., Muzic, R. F., Su, K.-H., O’Donnell, J. K., & Faulhaber, P. F. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.114.148338 · Cited by 154
Variations in Clinical PET/CT Operations: Results of an International Survey of Active PET/CT Users
Beyer, T., Czernin, J., & Freudenberg, L. S. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.110.079624 · Cited by 136
Comparison of image quality and lesion detection between digital and analog PET/CT
López-Mora, D. A., Flotats, A., Fuentes-Ocampo, F., Camacho, V., Fernández, A., Ruiz, A., Duch, J., Sizova, M., Domènech, A., Estorch, M., et al. (2019) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-019-4260-z · Cited by 94
Reduction of 18 F-FDG Dose in Clinical PET/MR Imaging by Using Silicon Photomultiplier Detectors
Sekine, T., Delso, G., Zeimpekis, K. G., de Galiza Barbosa, F., ter Voert, E. E. G. W., Huellner, M., & Veit-Haibach, P. (2018) · Radiology · doi:10.1148/radiol.2017162305 · Cited by 79
Research works printed but not linked: 1
- van Sluis J Boellaard R Dierckx RA Stormezand G Glaudemans A Noordzij W Image quality and activity optimization in oncological (18)F-FDG PET using the digital Biograph Vision PET/CT J Nucl Med 2019 10.2967/jnumed.119.234351 31628214 · DOI · Paper page
Cited by
Alberts, I., Hünermund, J.-N., Prenosil, G., Mingels, C., Bohn, K. P., Viscione, M., Sari, H., Vollnberg, B., Shi, K., Afshar-Oromieh, A., et al. (2021) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-021-05282-7 · Cited by 286
Self-citationXue, S., Guo, R., Bohn, K. P., Matzke, J., Viscione, M., Alberts, I., Meng, H., Sun, C., Zhang, M., Zhang, M., et al. (2021) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-021-05644-1 · Cited by 113
Self-citationAlberts, I., Sari, H., Mingels, C., Afshar‐Oromieh, A., Pyka, T., Shi, K., & Rominger, A. (2023) · Cancer Imaging · doi:10.1186/s40644-023-00540-3 · Cited by 98
Self-citationInfluences on PET Quantification and Interpretation
Rogasch, J. M. M., Hofheinz, F., van Heek, L., Voltin, C.-A., Boellaard, R., & Kobe, C. (2022) · Diagnostics · doi:10.3390/diagnostics12020451 · Cited by 57
“However, standardized reading criteria do not negate the dissimilarities in the images obtained using different types of PET hardware and methods of image reconstruction, and both SiPM technology [79] and PSF reconstruction [225] have been found to result in systematically higher lesion conspicuity despite standardize…”
BackgroundA review of harmonization strategies for quantitative PET
Akamatsu, G., Tsutsui, Y., Daisaki, H., Mitsumoto, K., Baba, S., & Sasaki, M. (2023) · Annals of Nuclear Medicine · doi:10.1007/s12149-022-01820-x · Cited by 57
Digital PET vs Analog PET: Clinical Implications?
López-Mora, D. A., Carrió, I., & Flotats, A. (2021) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2021.10.004 · Cited by 54
Abbreviated scan protocols to capture 18F-FDG kinetics for long axial FOV PET scanners
Viswanath, V., Sari, H., Pantel, A. R., Conti, M., Daube-Witherspoon, M. E., Mingels, C., Alberts, I., Eriksson, L., Shi, K., Rominger, A., et al. (2022) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-022-05747-3 · Cited by 53
Self-citationAlberts, I., Schepers, R., Zeimpekis, K., Sari, H., Rominger, A., & Afshar‐Oromieh, A. (2022) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-022-05961-z · Cited by 49
Self-citationSingh, M. K. (2023) · Radiography · doi:10.1016/j.radi.2023.10.004 · Cited by 48
Sari, H., Teimoorisichani, M., Mingels, C., Alberts, I., Panin, V., Bharkhada, D., Xue, S., Prenosil, G., Shi, K., Conti, M., et al. (2022) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-022-05909-3 · Cited by 40
Self-citationLong axial field-of-view PET/CT devices: are we ready for the technological revolution?
Filippi, L., Dimitrakopoulou‐Strauss, A., Evangelista, L., & Schillaci, O. (2022) · Expert Review of Medical Devices · doi:10.1080/17434440.2022.2141111 · Cited by 38
Yamagishi, S., Miwa, K., Kamitaki, S., Anraku, K., Sato, S., Yamao, T., Kubo, H., Miyaji, N., & Oguchi, K. (2023) · Journal of Nuclear Medicine · doi:10.2967/jnumed.123.266140 · Cited by 36
Imaging Procedure and Clinical Studies of [18F]FP-CIT PET
Sung, C., Oh, S. J., & Kim, J. S. (2024) · Nuclear Medicine and Molecular Imaging · doi:10.1007/s13139-024-00840-x · Cited by 31
Dose Reduction in Pediatric Oncology Patients with Delayed Total-Body [18F]FDG PET/CT
Mingels, C., Spencer, B. A., Nalbant, H., Omidvari, N., Rokni, M., Rominger, A., Sen, F., Cherry, S. R., Badawi, R. D., Abdelhafez, Y. G., et al. (2024) · Journal of Nuclear Medicine · doi:10.2967/jnumed.124.267521 · Cited by 30
Self-citationPET-CT in Clinical Adult Oncology: I. Hematologic Malignancies
Salem, A. E., Shah, H., Covington, M. F., Koppula, B. R., Fine, G., Wiggins, R. H., Hoffman, J. M., & Morton, K. A. (2022) · Cancers · doi:10.3390/cancers14235941 · Cited by 26
The impact of long axial field of view (LAFOV) PET on oncologic imaging
Cook, G., Alberts, I., Wagner, T., Fischer, B. M., Nazir, M. S., & Lilburn, D. M. L. (2024) · European Journal of Radiology · doi:10.1016/j.ejrad.2024.111873 · Cited by 23
Self-citationCosta, P. F., Jentzen, W., Brahmer, A., Mavroeidi, I. A., Zarrad, F., Umutlu, L., Fendler, W. P., Rischpler, C., Herrmann, K., Conti, M., et al. (2022) · BMC Cancer · doi:10.1186/s12885-022-09993-4 · Cited by 22
Filippi, L., & Schillaci, O. (2022) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-022-05791-z · Cited by 22
Diagnostic accuracy in NSCLC lymph node staging with Total-Body and conventional PET/CT
Mingels, C., Madani, M. H., Sen, F., Nalbant, H., Riess, J. W., Abdelhafez, Y. G., Ghasemiesfe, A., Rominger, A., Guindani, M., Badawi, R. D., et al. (2025) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-025-07177-3 · Cited by 20
Self-citationAlves, V. D. P. V., Brady, S. L., Ata, N. K. A., Li, Y., MacLean, J. R., Zhang, B., Sharp, S. E., & Trout, A. T. (2022) · American Journal of Roentgenology · doi:10.2214/ajr.22.27894 · Cited by 20
Mingels, C., Sari, H., Gözlügöl, N., Bregenzer, C., Knappe, L. M., Krieger, K., Afshar‐Oromieh, A., Pyka, T., Nardo, L., Gräni, C., et al. (2023) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-023-06435-6 · Cited by 18
Self-citationLong Axial Field-of-View PET/CT Could Answer Unmet Needs in Gynecological Cancers
Triumbari, E. K. A., Rufini, V., Mingels, C., Rominger, A., Alavi, A. A., Fanfani, F., Badawi, R. D., & Nardo, L. (2023) · Cancers · doi:10.3390/cancers15092407 · Cited by 15
Self-citationDjekidel, M., Alsadi, R., Akl, M. A., Bouhali, O., & O’Doherty, J. (2023) · Frontiers in Nuclear Medicine · doi:10.3389/fnume.2023.1183471 · Cited by 14
Medical Radiation Exposure Reduction in PET via Super-Resolution Deep Learning Model
Yoshimura, T., Hasegawa, A., Kogame, S., Magota, K., Kimura, R., Watanabe, S., Hirata, K., & Sugimori, H. (2022) · Diagnostics · doi:10.3390/diagnostics12040872 · Cited by 14
Peng, L., Liao, Y., Zhou, R., Zhong, Y., Jiang, H., Wang, J., Fu, Y., Xue, L., Zhang, X., Sun, M., et al. (2023) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-023-06273-6 · Cited by 13
Long-axial field-of-view PET/CT improves radiomics feature reliability
Alberts, I., Xue, S., Sari, H., Cavinato, L., Prenosil, G., Afshar‐Oromieh, A., Mingels, C., Shi, K., Caobelli, F., Rahmim, A., et al. (2024) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-024-06921-5 · Cited by 13
Self-citationLong Axial Field-of-View (LAFOV) PET/CT in Prostate Cancer
Sachpekidis, C., & Dimitrakopoulou‐Strauss, A. (2024) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2024.05.004 · Cited by 13
Self-citationSeo, M., Kim, Y., Ye, B. D., Park, S. H., Kim, S.-Y., Jung, J. H., Hwang, S. W., Chae, S. Y., Lee, D. Y., Lee, S. J., et al. (2022) · Journal of Nuclear Medicine · doi:10.2967/jnumed.121.263289 · Cited by 12
Long-Axial Field-of-View PET Imaging in Patients with Lymphoma
Mingels, C., Nalbant, H., Sari, H., Godinez, F., Şen, F., Spencer, B. A., Esteghamat, N., Tuscano, J. M., & Nardo, L. (2024) · PET Clinics · doi:10.1016/j.cpet.2024.05.005 · Cited by 11
Self-citationMei, R., Pyka, T., Sari, H., Fanti, S., Afshar‐Oromieh, A., Giger, R., Caobelli, F., Rominger, A., & Alberts, I. (2023) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-023-06516-6 · Cited by 11
Self-citationBagni, O., Danieli, R., Bianconi, F., Palumbo, B., & Filippi, L. (2025) · Biomedicines · doi:10.3390/biomedicines13081887 · Cited by 9
Literature review: Imaging in prostate cancer
Mingels, C., Loebelenz, L. I., Huber, A. T., Alberts, I., Rominger, A., Afshar‐Oromieh, A., & Obmann, V. C. (2023) · Current Problems in Cancer · doi:10.1016/j.currproblcancer.2023.100968 · Cited by 8
Self-citationQuarter-Century Transformation of Oncology
Ulaner, G. A., Vaz, S. C., & Groheux, D. (2024) · PET Clinics · doi:10.1016/j.cpet.2023.12.002 · Cited by 7
Technical NEMA NU2–2018 Performance Assessment of Time-of-Flight-Integrated Digital PET-CT System
Singh, M. K., Agarwal, K. K., Vishwakarma, M., & Patel, H. (2024) · World Journal of Nuclear Medicine · doi:10.1055/s-0044-1778709 · Cited by 7
Lymphoma Therapy Response Assessment with Low-Dose [ 18 F]FDG Total-Body PET/CT
Mingels, C., Chung, K. J., Nalbant, H., Abdelhafez, Y. G., Esteghamat, N., Rokni, M., Zoghi, S., Tuscano, J. M., Rominger, A., Badawi, R. D., et al. (2025) · Journal of Nuclear Medicine · doi:10.2967/jnumed.124.268841 · Cited by 6
Self-citationRovera, G., Urso, L., Stracuzzi, F., Laudicella, R., Frantellizzi, V., Cottignoli, C., Gazzilli, M., Guglielmo, P., Panareo, S., Evangelista, L., et al. (2024) · Clinical and Translational Imaging · doi:10.1007/s40336-024-00650-3 · Cited by 5
Filippi, L., Bianconi, F., Ferrari, C., Linguanti, F., Battisti, C., Urbano, N., Minestrini, M., Messina, S. G., Buci, L., Baldoncini, A., et al. (2025) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-025-07508-4 · Cited by 5
Rovera, G., Urso, L., Stracuzzi, F., Laudicella, R., Frantellizzi, V., Cottignoli, C., Gazzilli, M., Guglielmo, P., Panareo, S., Evangelista, L., et al. (2024) · Clinical and Translational Imaging · doi:10.1007/s40336-024-00653-0 · Cited by 4
Time to shift to Digital PET/CT?
Kwon, S. J., & Park, H. L. (2024) · Nuclear Medicine and Molecular Imaging · doi:10.1007/s13139-024-00900-2 · Cited by 4
Velo, P., Ismail, M. I., Mohandas, K. K., & Kasilingam, L. (2022) · Journal of medical imaging and radiation sciences · doi:10.1016/j.jmir.2022.09.010 · Cited by 4
Weissensee, A., Bregenzer, C., Viscione, M., Mingels, C., Sari, H., Caobelli, F., Seifert, R., Rominger, A., & Pyka, T. (2025) · Clinical Nuclear Medicine · doi:10.1097/rlu.0000000000006278 · Cited by 3
Self-citationFulcheri, C. P. L., Sguazzini, O., Parenti, V. D., Scarlattei, M., Baldari, G., Cervati, V., Graziani, T., Ruffini, L., & Ghetti, C. (2026) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncag003 · Cited by 3
Cumulative radiation doses due to nuclear medicine examinations: a systematic review
Brambilla, M., Kuchcińska, A., Matheoud, R., & Muni, A. (2021) · British Journal of Radiology · doi:10.1259/bjr.20210444 · Cited by 3
Eom, Y., Park, Y.-J., Lee, S., Lee, S. J., An, Y., Park, B.-N., & Yoon, J. (2024) · Tomography · doi:10.3390/tomography10120151 · Cited by 3
Faist, D., Gnesin, S., Medici, S., Khan, A., Lalonde, M. N., Schaefer, N. G., Depeursinge, A., Conti, M., Schaefferkoetter, J. D., Prior, J. O., et al. (2025) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-025-07259-2 · Cited by 2
18F-FDG PET Adults Whole-Body Scan Injected Dose Optimization: A Mini Review
(2023) · Journal of Population Therapeutics and Clinical Pharmacology · doi:10.47750/jptcp.2023.1113 · Cited by 2
Impact of Long–Axial-Field-of-View PET/CT Workflow on the Patient Experience
Chen, J., Rae, W. I. D., Kench, P. L., Bailey, D. L., Bailey, E. A., ROACH, P. J., & Meikle, S. R. (2025) · Journal of Nuclear Medicine Technology · doi:10.2967/jnmt.125.270843 · Cited by 2
Fukai, S., Daisaki, H., Shimada, N., Ishiyama, M., Umeda, T., Yamashita, K., Miyaji, N., Takiguchi, T., Kawakami, H., & Terauchi, T. (2022) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/aca90d · Cited by 2
Inomata, T., Sato, K., Ibaraki, M., Kominami, M., Kinoshita, F., Shinohara, Y., Kato, M., & Kinoshita, T. (2023) · Japanese Journal of Radiological Technology · doi:10.6009/jjrt.2023-1365 · Cited by 2
Leybourne, N., Prakash, V., Hussein, M., Fenwick, A., Strouhal, P., Evans, P., & Florescu, L. (2025) · EJNMMI Physics · doi:10.1186/s40658-025-00821-9 · Cited by 2
Digitale SPECT und PET: Klinische Konsequenzen
Rogasch, J. M. M. (2022) · Angewandte Nuklearmedizin · doi:10.1055/a-1715-5305 · Cited by 1
Inomata, T., Sato, K., Ibaraki, M., Kominami, M., Shinohara, Y., Kinoshita, F., Yamamoto, H., Kato, M., Kinoshita, T., & Chida, K. (2025) · Applied Sciences · doi:10.3390/app152212292 · Cited by 1
Digitale Positronenemissionstomografie – Entwicklung, Detektortechnologie und Perspektiven
Kazek, S. L., Jentzen, W., Seifert, R., Herrmann, K., & Kersting, D. (2022) · Angewandte Nuklearmedizin · doi:10.1055/a-1715-5184 · Cited by 1
Alberts, I., Schepers, R., Zeimpekis, K., Sari, H., Rominger, A., & Afshar‐Oromieh, A. (2022) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-022-06071-6 · Cited by 1
Self-citationDwivedi, P., Jha, A. K., Mithun, S., Sawant, V., Vajarkar, V., Chauhan, M. K. H., Choudhury, S., & Rangarajan, V. (2024) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncae179 · Cited by 1
Estimation of Effective Dose from 18F-FDG PET/CT in Pediatric Oncology Patients
Volzhencev, A. A., Shermakova, A. А., Yadgarov, M. Y., Glotova, I. V., Kiselev, K. D., Ryzhov, S. A., & Likar, Y. N. (2026) · Meditsinskaya Fizika · doi:10.52775/1810-200x-2026-110-2-76-88 · Cited by 1
Maffione, A. M., Ferretti, A., Bartoletti, P., Pavan, L., Massaro, A., Chondrogiannis, S., Osele, A., Gava, M. M., & Marzola, M. C. (2026) · Clinical and Translational Imaging · doi:10.1007/s40336-026-00791-7
Silva, L. C., Constantino, C. S., Teixeira, R. J. V., Castanheira, J. C., Oliveira, F. P. M., & Costa, D. C. (2025) · Journal of Imaging Informatics in Medicine · doi:10.1007/s10278-025-01638-9
The Role of Artificial Intelligence in PET Imaging
Veziroglu, E. M., Dayma, K., Hasani, N., Kulkarni, R., Farhadi, F., & Klyuzhin, I. S. (2026) · Radiologic Clinics of North America · doi:10.1016/j.rcl.2026.07.010
Fragoso Costa, P., Jentzen, W., Brahmer, A., Mavroeidi, I.-A., Zarrad, F., Umutlu, L., Fendler, W. P., Rischpler, C., Herrmann, K., Conti, M., et al. (2022) · Figshare · doi:10.6084/m9.figshare.20506869
Digitale SPECT und PET: Klinische Konsequenzen
Rogasch, J. M. M. (2023) · Radiopraxis · doi:10.1055/a-1993-7420
Petryakova, A. V., Чипига, Л. А., Andreev, G. А., Gromova, E. A., Катаева, Г. В., Kovalev, K. V., Onischenko, S. S., & Bezumova, E. V. (2024) · Radiacionnaâ gigiena · doi:10.21514/1998-426x-2024-17-1-44-53
Ogasawara, R., Irikawa, A., Watanabe, Y., Harada, T., Hosokawa, S., Koyama, K., Tsuda, K., Kimura, T., Okuda, K., & Takahashi, Y. (2025) · Radiation · doi:10.3390/radiation5020018
Legodi, K., Qebetu, M., Banda, K., Goorhoo, P., López-Montes, A., Shi, K., Sathekge, M. M., & Ramonaheng, K. (2026) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/ae7577
Self-citationYang, R., Xie, Y., Cai, D., Lin, Y., Yu, H., Zheng, Z., He, Y., & Shi, H. (2025) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000002073
Leike, T., Lima, T. V. M., Saltybaeva, N., Grünig, H., Stenger-Weisser, A., Bhure, U., Leimgruber, A., & Strobel, K. (2026) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000002135
Li, B., Geng, M., Hong, R., Mao, B., Li, P., Yuan, X., Fang, L., Zhang, B., Liu, Y., Liu, Q., et al. (2026) · Journal of Nuclear Medicine and Molecular Imaging · doi:10.65457/jnmmi-2026-0017
Azahar, Q., Hashim, H., Hamid, K. A., Xin, L., & Mansor, S. (2026) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-026-01776-6
Review of Methods of [18F]FDG Dose Optimization for PET/CT
Averyanova-Onishchenko, S. S., & Рыжкин, С. А. (2025) · Meditsinskaya Fizika · doi:10.52775/1810-200x-2025-107-3-85-94
Ryzhov, S. A., Volzhencev, A. A., Shermakova, A. А., Yadgarov, M. Y., Glotova, I. V., Kiselev, K. D., & Likar, Y. N. (2026) · Meditsinskaya Fizika · doi:10.52775/1810-200x-2026-111-3-94-108
Baek, S., Kim, J. Y., Joo, J. Y., Kim, H., Kwon, O., Lee, M., Lee, M.-G., Park, C., & Chae, S. Y. (2026) · Diagnostics · doi:10.3390/diagnostics16172825
Bedair, K. M., Hammoury, S. I., Moustafa, M. H., Welaya, E. M., & Mohamed, E. I. (2026) · Radiography · doi:10.1016/j.radi.2026.103374
Digitalni PET/CT: prednosti in slabosti
Leskovar, N. (2023) · Repository of the University of Ljubljana (University of Ljubljana)
Metadata
- Authors
- Alberts, I., Sachpekidis, C., Prenosil, G., Viscione, M., Bohn, K. P., Mingels, C., Shi, K., Ashar-Oromieh, A., & Rominger, A.
- Affiliations
- University of Bern (CH)
- Corresponding author
- Alberts, Ian
- Venue
- Annals of Nuclear Medicine
- Published
- 2021-02-07
- Volume
- 35
- Issue
- 4
- Pages
- 485-492
- ISSN
- 0914-7187 (ISSN-L)
- 1864-6433
- PMC
- PMC7981298
- OpenAlex
- W3127712352
- Cited by
- 70
- References
- 27
- Citations by year
- 2026: 10 · 2025: 12 · 2024: 14 · 2023: 13 · 2022: 17 · 2021: 4
- FWCI
- 6.33
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- Annals of Nuclear Medicine (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- BORIS (University Library Bern) (submittedVersion, cc-by)
- eScholarship (California Digital Library) (submittedVersion, cc-by)
- Funders
- Universität Bern
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Keywords
- digital PET/CT, 18F-FDG, acquisition time reduction, list-mode reconstruction, cancer detection, lesion contrast, tumor-to-background ratio, standardized uptake value, scanner sensitivity
- MeSH terms
- Positron Emission Tomography Computed Tomography, Body Weight, Humans, Image Processing, Computer-Assisted, Neoplasms, Retrospective Studies, Sensitivity and Specificity, Radiopharmaceuticals, Fluorodeoxyglucose F18, Signal-To-Noise Ratio
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.