Image Quality and Activity Optimization in Oncologic 18F-FDG PET Using the Digital Biograph Vision PET/CT System
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
The first Biograph Vision PET/CT system (Siemens Healthineers) was installed at the University Medical Center Groningen. Improved performance of this system could allow for a reduction in activity administration or scan duration. This study evaluated the effects of reduced scan duration in oncologic 18F-FDG PET imaging on quantitative and subjective imaging parameters and its influence on clinical image interpretation. Methods: Patients referred for a clinical PET/CT scan were enrolled in this study, received a weight-based 18F-FDG injected activity, and underwent list-mode PET acquisition at 180 s per bed position (s/bp). Acquired PET data were reconstructed using the vendor-recommended clinical reconstruction protocol (hereafter referred to as “clinical”), using the clinical protocol with additional 2-mm gaussian filtering (hereafter referred to as “clinical+G2”), and—in conformance with European Association of Nuclear Medicine Research Ltd. (EARL) specifications—using different scan durations per bed position (180, 120, 60, 30, and 10 s). Reconstructed images were quantitatively assessed for comparison of SUVs and noise. In addition, clinically reconstructed images were qualitatively evaluated by 3 nuclear medicine physicians. Results: In total, 30 oncologic patients (22 men, 8 women; age: 48–88 y [range], 67 ± 9.6 y [mean ± SD]) received a single weight-based (3 MBq/kg) 18F-FDG injected activity (weight: 45–123 kg [range], 81 ± 15 kg [mean ± SD]; activity: 135–380 MBq [range], 241 ± 47.3 MBq [mean ± SD]). Significant differences in lesion SUVmax were found between the 180-s/bp images and the 30- and 10-s/bp images reconstructed using the clinical protocols, whereas no differences were found in lesion SUVpeak. EARL-compliant images did not show differences in lesion SUVmax or SUVpeak between scan durations. Quantitative parameters showed minimal deviation (∼5%) in the 60-s/bp images. Therefore, further subjective image quality assessment was conducted using the 60-s/bp images. Qualitative assessment revealed the influence of personal preference on physicians’ willingness to adopt the 60-s/bp images in clinical practice. Although quantitative PET parameters differed minimally, an increase in noise was observed. Conclusion: With the Biograph Vision PET/CT system for oncologic 18F-FDG imaging, scan duration or activity administration could be reduced by a factor of 3 or more with the use of the clinical+G2 or the EARL-compliant reconstruction protocol.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Performance Characteristics of the Digital Biograph Vision PET/CT System
van Sluis, J., de Jong, J., Schaar, J., Noordzij, W., van Snick, P., Dierckx, R., Borra, R., Willemsen, A., & Boellaard, R. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215418 · Cited by 484 · cited 1×
“When EARL-compliant reconstructions are applied, even further reductions are potentially achievable (6).”
BackgroundSelf-citationStudies of a Next-Generation Silicon-Photomultiplier–Based Time-of-Flight PET/CT System
Hsu, D. F. C., Ilan, E., Peterson, W. T., Uribe, J., Lubberink, M., & Levin, C. S. (2017) · Journal of Nuclear Medicine · doi:10.2967/jnumed.117.189514 · Cited by 241
Performance Evaluation of the Vereos PET/CT System According to the NEMA NU2-2012 Standard
Rausch, I., Ruiz, A., Valverde-Pascual, I., Cal-González, J., Beyer, T., & Carrio, I. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215541 · Cited by 168
Initial experience with a SiPM-based PET/CT scanner: influence of acquisition time on image quality
Sonni, I., Baratto, L., Park, S., Hatami, N., Srinivas, S., Davidzon, G., Gambhir, S. S., & Iagaru, A. (2018) · EJNMMI Physics · doi:10.1186/s40658-018-0207-x · Cited by 74 · cited 1×
“With regard to other commercially available digital PET/CT systems, a comparable study stated that a reduction in activity or scan time to 90 s/bp (which equals a factor of 2) is feasible when the Discovery MI PET/CT system (GE Healthcare) is used (16).”
Method
Cited by
Prenosil, G. A., Sari, H., Fürstner, M., Afshar-Oromieh, A., Shi, K., Rominger, A., & Hentschel, M. (2022) · Journal of Nuclear Medicine · doi:10.2967/jnumed.121.261972 · Cited by 338
Alberts, I., Hünermund, J.-N., Prenosil, G., Mingels, C., Bohn, K. P., Viscione, M., Sari, H., Vollnberg, B., Shi, K., Afshar-Oromieh, A., et al. (2021) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-021-05282-7 · Cited by 286
Zhang, Y.-Q., Hu, P.-C., Wu, R.-Z., Gu, Y.-S., Chen, S.-G., Yu, H.-J., Wang, X.-Q., Song, J., & Shi, H.-C. (2020) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-020-04823-w · Cited by 123
Alberts, I., Sachpekidis, C., Prenosil, G., Viscione, M., Bohn, K. P., Mingels, C., Shi, K., Ashar-Oromieh, A., & Rominger, A. (2021) · Annals of Nuclear Medicine · doi:10.1007/s12149-021-01588-6 · Cited by 70
Pilz, J., Hehenwarter, L., Zimmermann, G., Rendl, G., Schweighofer-Zwink, G., Beheshti, M., & Pirich, C. (2021) · EJNMMI Research · doi:10.1186/s13550-021-00784-9 · Cited by 6
“Further benefits can be experienced in the paediatric field where PET/CT studies with short acquisition times might lead to a decrease in anaesthesia or sedation time [15, 31].”
Background
Metadata
- Authors
- van Sluis, J., Boellaard, R., Dierckx, R. A. J. O., Stormezand, G. N., Glaudemans, A. W. J. M., & Noordzij, W.
- Venue
- Journal of Nuclear Medicine
- Publisher
- Society of Nuclear Medicine
- Published
- 2019-10-18
- Volume
- 61
- Issue
- 5
- Pages
- 764-771
- ISSN
- 0161-5505 (ISSN-L)
- 2159-662X
- OpenAlex
- W2980576527
- Cited by
- 65
- References
- 4
- FWCI
- 4.95
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Image Quality and Activity Optimization in Oncologic 18F-FDG PET Using the Digital Biograph Vision PET/CT System. https://app.cochat.ai/papers/10.2967/jnumed.119.234351
- The paper
- van Sluis, J., Boellaard, R., Dierckx, R. A. J. O., Stormezand, G. N., Glaudemans, A. W. J. M., & Noordzij, W. (2020). Image Quality and Activity Optimization in Oncologic 18F-FDG PET Using the Digital Biograph Vision PET/CT System. Journal of Nuclear Medicine. https://doi.org/10.2967/jnumed.119.234351
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.