Comparative accuracy testing of non-contact infrared thermometers and temporal artery thermometers in an adult hospital setting

Open access
Khan, Shahrukh, Saultry, Bridey, Adams, Scott, Kouzani, Abbas Z., Decker, Kelly, Digby, Robin, Bucknall, Tracey·2021·American Journal of Infection Control

Abstract

Background NCIT are non-invasive devices for fever screening in children. However, evidence of their accuracy for fever screening in adults is lacking. This study aimed to compare the accuracy of non-contact infrared thermometers (NCIT) with temporal artery thermometers (TAT) in an adult hospital.Methods A prospective observational study was conducted on a convenience sample of non-infectious inpatients in 2 Australian hospitals. NCIT and TAT devices were used to collect body temperature recordings. Participant characteristics included age, gender, skin color, highest temperature, and antipyretic medications recorded in last 24-hour.Results In 265 patients, a mean difference of ± 0.26°C was recorded between the NCIT (36.64°C) and the reference TAT (36.90°C) temperature devices. Bland-Altman analysis showed that NCIT and TAT temperatures were closely aligned at temperatures <37.5°C, but not at temperatures >37.5°C. NCIT had low sensitivity (16.13%) at temperatures ≥37.5°C. An AUROC score of 0.67 (SD 0.05) demonstrated poor accuracy of the NCIT device at temperatures ≥37.5°C.Conclusion This is the first study to compare accuracy of NCIT thermometers to TAT in adult patients. Although mass fever screening is currently underway using NCIT, these results indicate that the NCIT may not be the most accurate device for fever mass screening during a pandemic.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.ajic.2020.09.012

Cites

15 of 15 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Previous studies using clinical data from hospitals and transit centers have been inconclusive regarding the use and effectiveness of NCITs as a screening method during SARS and influenza outbreaks.

    “Overall, previous studies using clinical data from hospitals and transit centers have been inconclusive regarding the use and effectiveness of NCITs as a screening method during SARS and influenza outbreaks.”

    From Clinical evaluation of non-contact infrared thermometers (2021) · cites this paper as [8]

    The Introduction (page unavailable) says previous studies of NCIT accuracy in children reported conflicting results, and the Discussion describes differing accuracy findings across studies. The paper does not establish that those studies used clinical data from transit centers or assessed screening during SARS and influenza outbreaks. (LLM, conf=0.82)

    Experimental — accuracy not yet measured
  • Clinical studies have reported mixed performance in terms of NCIT accuracies.

    “Although NCITs may be the primary tools for temperature screening during a pandemic, clinical studies have reported mixed performance in terms of their accuracies5–8.”

    From Clinical evaluation of non-contact infrared thermometers (2021) · cites this paper as [8]

  • Khan et al. quantified the mean difference between NCIT and temporal artery thermometer in adults to be ±0.26 °C.

    “Khan et al.8 quantified the mean difference between NCIT and temporal artery thermometer in adults to be ± 0.26 °C.”

    From Clinical evaluation of non-contact infrared thermometers (2021) · cites this paper as [8]

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Khan, S., Saultry, B., Adams, S., Kouzani, A. Z., Decker, K., Digby, R., & Bucknall, T.
Affiliations
  • Deakin University (AU)
  • The Alfred Hospital (AU)
Corresponding author
Bucknall, Tracey
Venue
American Journal of Infection Control
Publisher
Elsevier BV
Published
2021-05
Volume
49
Issue
5
Pages
597-602
ISSN
0196-6553 (ISSN-L)
OpenAlex
W3090527447
Cited by
49
References
15
Citations by year
2026: 3 · 2025: 6 · 2024: 7 · 2023: 10 · 2022: 10 · 2021: 12 · 2020: 1
FWCI
4.02
Type
journal article
Language
en
Open access
Locations
  • American Journal of Infection Control (publishedVersion)
  • PubMed (publishedVersion)
  • Deakin Research Online (Deakin University) (submittedVersion)
  • Figshare (submittedVersion, cc-by)
  • Minerva Access (University of Melbourne) (submittedVersion)
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
  • Health Sciences › Medicine › Critical Care and Intensive Care Medicine › Thermal Regulation in Medicine
  • Health Sciences › Medicine › Physiology › Thermoregulation and physiological responses
Keywords
non-contact infrared thermometer, fever screening, hospitalized patients, antipyretic use, skin color, COVID-19 pandemic, hospital infection control
MeSH terms
Adult, Australia, Body Temperature, Child, Hospitals, Humans, Prospective Studies, Sensitivity and Specificity, Temporal Arteries, Thermometers

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review