Clinical evaluation of non-contact infrared thermometers
Open accessPublished version of a preprint
The DOI matches its Crossref record; 9 of the 10 references it cites were found in a registry or an index; 9 of the 25 sources its claims cite were checked for support.
Abstract
Non-contact infrared thermometers (NCITs) are being widely used during the COVID-19 pandemic as a temperature-measurement tool for screening and isolating patients in healthcare settings, travelers at ports of entry, and the general public. To understand the accuracy of NCITs, a clinical study was conducted with 1113 adult subjects using six different commercially available NCIT models. A total of 60 NCITs were tested with 10 units for each model. The NCIT-measured temperature was compared with the oral temperature obtained using a reference oral thermometer. The mean difference between the reference thermometer and NCIT measurement (clinical bias) was different for each NCIT model. The clinical bias ranged from just under - 0.9 °C (under-reporting) to just over 0.2 °C (over-reporting). The individual differences ranged from - 3 to + 2 °C in extreme cases, with the majority of the differences between - 2 and + 1 °C. Depending upon the NCIT model, 48% to 88% of the individual temperature measurements were outside the labeled accuracy stated by the manufacturers. The sensitivity of the NCIT models for detecting subject's temperature above 38 °C ranged from 0 to 0.69. Overall, our results indicate that some NCIT devices may not be consistently accurate enough to determine if subject's temperature exceeds a specific threshold of 38 °C. Model-to-model variability and individual model accuracy in the displayed temperature were found to be outside of acceptable limits. Accuracy and credibility of the NCITs should be thoroughly evaluated before using them as an effective screening tool.
Overview
CoChat Summary
## TL;DR
This clinical study evaluated six commercially available non-contact infrared thermometers (NCITs) against oral reference temperatures in more than 1,000 adults and found substantial model-to-model variability. Several devices showed clinically important bias, poor sensitivity for detecting 38 °C, and many readings outside manufacturers’ stated accuracy.
## Key findings
- Among 1022 subjects in the analyzed results, oral reference temperatures ranged from 36.1 to 40.3 °C, with an average of 36.9 °C in both trials.
- Clinical bias across the six NCIT models ranged from under-reporting by −0.87 °C to over-reporting by 0.21 °C; Model E had the largest bias (−0.89 °C) and Model C the smallest (0.14 °C).
- Depending on model, 48% to 88% of individual measurements were outside the manufacturers’ labeled accuracy claims.
- Sensitivity for detecting a temperature of 38 °C ranged from 0 to 0.69, and specificity ranged from 0.97 to 1.
- The difference between reference temperature and NCIT temperature was statistically significant for all six models; only Models C and F showed intra-model consistency, and only some model pairs were not significantly different from each other.
## Methods
- Controlled clinical study of adult subjects at the University Health Center, University of Maryland, College Park; 1113 subjects were enrolled, and the analysis reported 1022 subjects after exclusions.
- Six commercially available NCIT models were tested, with 10 units per model; each subject had oral temperature measured with a Welch Allyn SureTemp Plus 690 reference thermometer and then NCIT measurements at the forehead, repeated in two trials.
- Analyses included clinical bias calculations, ANOVA with Tukey post-hoc testing, correlation of measurement error with oral temperature, and sensitivity/specificity calculations across threshold temperatures from 37.0 to 38.0 °C.
## Limitations
- The paper states that subjects under 18 were not included.
- The number of subjects with temperatures ≥38 °C was about 5% of the sample; this was stated by the authors as a limitation, although they said there were still sufficient subjects for analysis.
- The study focused on six NCIT models from different manufacturers that targeted the center of the forehead; the authors note that other potentially important clinical factors such as sex, age, skin tone, and weight were not considered and should be evaluated in future studies.
- Evident from the design: the study was conducted under controlled conditions, so the findings may not fully reflect real-world screening settings with greater user and environmental variability.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures




Verification
Claims the paper makes
5 Valid · 1 Not valid · 8 Not checked · 70 Own results and interpretations
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“Although NCITs may be the primary tools for temperature screening during a pandemic, clinical studies have reported mixed performance in terms of their accuracies5–8.”
§ Clinical evaluation of non-contact infrared thermometers > Introduction
What the sources say
- Not checked — no full text of the sourcecites [5] Dante, A., Franconi, I., Marucci, A. R., Alfes, C. M. & Lancia, L. Evaluating the interchangeability of forehead, tympanic, and axillary thermometers in italian paediatric clinical settings: Results of a multicentre observational study. J. Pediatr. Nurs. 52, e21–e25. https://doi.org/10.1016/j.pedn.2019.11.014 (2020). source
- Not checked — no full text of the sourcecites [6] Franconi, I., La Cerra, C., Marucci, A. R., Petrucci, C. & Lancia, L. Digital axillary and non-contact infrared thermometers for children. Clin. Nurs. Res. 27, 180–190. https://doi.org/10.1177/1054773816676538 (2018). source
- Supportedcites [7] Hayward, G. et al. Non-contact infrared versus axillary and tympanic thermometers in children attending primary care: A mixed-methods study of accuracy and acceptability. Br. J. Gen. Pract. 70, e236–e244. https://doi.org/10.3399/bjgp20X708845 (2020). source
The Introduction reports variable agreement between NCITs and conventional thermometers and says studies found differing sensitivity for fever detection. The Comparison with existing literature section likewise describes studies finding both low and much higher sensitivity. No page numbers are provided in our copy. (LLM, conf=0.94)
- Supportedcites [8] Khan, S. et al. Comparative accuracy testing of non-contact infrared thermometers and temporal artery thermometers in an adult hospital setting. Am. J. Infect. Control https://doi.org/10.1016/j.ajic.2020.09.012 (2020). source
The Introduction says studies of NCIT accuracy and reliability have reported conflicting results. The Discussion contrasts this study’s poor fever-detection accuracy with higher accuracy reported in other clinical studies. Page numbers are not supplied in the provided passages. (LLM, conf=0.93)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Overall, previous studies using clinical data from hospitals and transit centers have been inconclusive regarding the use and effectiveness of NCITs as a screening method during SARS and influenza outbreaks.”
§ Clinical evaluation of non-contact infrared thermometers > Introduction
What the sources say
- Not checked — no full text of the sourcecites [5] Dante, A., Franconi, I., Marucci, A. R., Alfes, C. M. & Lancia, L. Evaluating the interchangeability of forehead, tympanic, and axillary thermometers in italian paediatric clinical settings: Results of a multicentre observational study. J. Pediatr. Nurs. 52, e21–e25. https://doi.org/10.1016/j.pedn.2019.11.014 (2020). source
- Not checked — no full text of the sourcecites [6] Franconi, I., La Cerra, C., Marucci, A. R., Petrucci, C. & Lancia, L. Digital axillary and non-contact infrared thermometers for children. Clin. Nurs. Res. 27, 180–190. https://doi.org/10.1177/1054773816676538 (2018). source
- Partly supportedcites [7] Hayward, G. et al. Non-contact infrared versus axillary and tympanic thermometers in children attending primary care: A mixed-methods study of accuracy and acceptability. Br. J. Gen. Pract. 70, e236–e244. https://doi.org/10.3399/bjgp20X708845 (2020). sourceExperimental — accuracy not yet measured
The Introduction says prior NCIT studies produced variable results and lacked clear conclusions; it also says most were conducted in paediatric inpatient or other hospital settings. It does not discuss transit-centre screening or NCIT use during SARS or influenza outbreaks, so those parts of the claim are unsupported. (LLM, conf=0.86)
- Partly supportedcites [8] Khan, S. et al. Comparative accuracy testing of non-contact infrared thermometers and temporal artery thermometers in an adult hospital setting. Am. J. Infect. Control https://doi.org/10.1016/j.ajic.2020.09.012 (2020). sourceExperimental — accuracy not yet measured
The Introduction (page unavailable) says previous studies of NCIT accuracy in children reported conflicting results, and the Discussion describes differing accuracy findings across studies. The paper does not establish that those studies used clinical data from transit centers or assessed screening during SARS and influenza outbreaks. (LLM, conf=0.82)
- Not matchedcites [9] Non-contact Thermometers for Detecting Fever: A Review of Clinical Effectiveness. Report No. PMID: 25520984 (Canadian Agency for Drugs and Technologies in Health).
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Could not verifycites [10] Chen, G. et al. Validity of wrist and forehead temperature in temperature screening in the general population during the outbreak of 2019 novel coronavirus: A prospective real-world study. medRxiv https://doi.org/10.1101/2020.03.02.20030148 (2020). sourceExperimental — accuracy not yet measured
Our copy of Chen et al. (2020) contains no reference to SARS or influenza outbreaks, to transit-center screening, or to prior studies being inconclusive; it is a single prospective COVID-19-era study at Ningbo First Hospital (Introduction; Study population, Results). Its Discussion instead states 'this study was the first to explore the reliability and validity of wrist and forehead temperature measurements in mass screening,' and notes only that forehead NCIT measurements 'can be affected by a person's physiological and environmental conditions (8, 16).' Searched: SARS, influenza, transit centers/airports, previous studies inconclusive, limited sensitivity — no passage addresses the claim's point, so neither support nor contradiction is available from our copy. (LLM, conf=0.72)
- Partly supportedcites [11] Bitar, D., Goubar, A. & Desenclos, J. International travels and fever screening during epidemics: a literature review on the effectiveness and potential use of non-contact infrared thermometers. Eurosurveillance 14(6), 19115 (2009). sourceExperimental — accuracy not yet measured
The Abstract (page unavailable) reports widely varying results from studies mainly in hospitals, while Materials and methods (page unavailable) says airport SARS screening reports lacked enough data to assess NCIT performance. This supports uncertainty about effectiveness, but the paper assesses influenza screening under a hypothetical pandemic; it does not report clinical screening studies during an influenza outbreak. (LLM, conf=0.86)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The clinical bias ranged from just under − 0.9 °C (under-reporting) to just over 0.2 °C (over-reporting).”
§ Clinical evaluation of non-contact infrared thermometers > Abstract
plainn = 1,113- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The mean difference between the reference thermometer and NCIT measurement (clinical bias) was different for each NCIT model.”
§ Clinical evaluation of non-contact infrared thermometers > Abstract
plainstates an associationn = 1,113- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The sensitivity of the NCIT models for detecting subject’s temperature above 38 °C ranged from 0 to 0.69.”
§ Clinical evaluation of non-contact infrared thermometers > Abstract
plainn = 1,113- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
41 more claims
Create a free account to see all 84 claims and to verify the 2 references still open.
- Probably not a substantial claim. 27
- It may state two separate claims as one. 11
Citations
Cites
Bitar, D., Goubar, A., & Desenclos, J. C. (2009) · Eurosurveillance · doi:10.2807/ese.14.06.19115-en · Cited by 150
Antibody responses to endemic coronaviruses modulate COVID-19 convalescent plasma functionality
Morgenlander, W. R., Henson, S. N., Monaco, D. R., Chen, A., Littlefield, K. E., Bloch, E. M., Fujimura, E., Ruczinski, I., Crowley, A. R., Natarajan, H., et al. (2021) · Journal of Clinical Investigation · doi:10.1172/jci146927 · Cited by 87
Khan, S., Saultry, B., Adams, S., Kouzani, A. Z., Decker, K., Digby, R., & Bucknall, T. (2021) · American Journal of Infection Control · doi:10.1016/j.ajic.2020.09.012 · Cited by 49
Chen, G., Xie, J., Dai, G., Zheng, P., Hu, X., Lu, H., Xu, L., Chen, X., & Chen, X. (2020) · Iranian Journal of Public Health · doi:10.18502/ijph.v49is1.3670 · Cited by 42
Chen, G., Xie, J., Dai, G., Zheng, P., Hu, X., Lu, H., Xu, L., Chen, X., & Chen, X. (2020) · doi:10.1101/2020.03.02.20030148 · Cited by 40
Hayward, G., Verbakel, J. Y., Ismail, F. A., Edwards, G., Wang, K., Fleming, S., Holtman, G. A., Glogowska, M., Morris, E., Curtis, K., et al. (2020) · British Journal of General Practice · doi:10.3399/bjgp20x708845 · Cited by 26
Adams, S. D., Bucknall, T. K., & Kouzani, A. Z. (2021) · Scientific Reports · doi:10.1038/s41598-021-91361-6 · Cited by 14
Digital Axillary and Non-Contact Infrared Thermometers for Children
Franconi, I., La Cerra, C., Marucci, A. R., Petrucci, C., & Lancia, L. (2018) · Clinical Nursing Research · doi:10.1177/1054773816676538 · Cited by 13
Dante, A., Franconi, I., Marucci, A. R., Alfes, C. M., & Lancia, L. (2020) · Journal of Pediatric Nursing · doi:10.1016/j.pedn.2019.11.014 · Cited by 13
Research works printed but not linked: 1
- Non-contact Thermometers for Detecting Fever: A Review of Clinical Effectiveness. Report No. PMID: 25520984 (Canadian Agency for Drugs and Technologies in Health).
Cited by
Zhao, Y., & Bergmann, J. (2023) · Sensors · doi:10.3390/s23177439 · Cited by 103
“Sullivan [20] hospital NCIT1, brand unknown”
Wang, Q., Zhou, Y., Ghassemi, P., McBride, D., Casamento, J. P., & Pfefer, T. J. (2021) · Sensors · doi:10.3390/s22010215 · Cited by 48
“Data acquired for each subject include thermal images, NCIT readings (analyzed in [45]) and reference sublingual temperature (Tre f ).”
MethodResultInfluentialSelf-citationSolomon, D., Arumugam, V., Sakthirajan, R., et al. (2024) · Kidney International Reports · doi:10.1016/j.ekir.2024.02.014 · Cited by 37
Ajčević, M., Stella, A. B., Furlanis, G., Caruso, P., Naccarato, M., Accardo, A., & Manganotti, P. (2022) · Sensors · doi:10.3390/s22134760 · Cited by 40
Flexible Fabric Temperature Sensor Based on Vo2/Pedot:Pss with High Performance
Wang, Y., Hong, Y., Hu, X., Ye, Y., Wang, P., Luo, J., Yin, A., Ren, Z., Liu, H., Qi, X., et al. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202300898 · Cited by 30
Metadata
- Authors
- Sullivan, S. J. L., Rinaldi, J. E., Hariharan, P., Casamento, J. P., Baek, S., Seay, N., Vesnovsky, O., & Topoleski, L. D. T.
- Affiliations
- United States Food and Drug Administration (US)
- University of Maryland, Baltimore County (US)
- Corresponding author
- Hariharan, Prasanna
- Venue
- Scientific Reports
- Published
- 2021-11-11
- Volume
- 11
- Issue
- 1
- Pages
- 22079-22079
- ISSN
- 2045-2322 (ISSN-L)
- PMC
- PMC8586154
- OpenAlex
- W3193296718
- Semantic Scholar
- 5bd3c18799b8aa8cbdf27f7c613dd899a0bffc32
- Cited by
- 44
- References
- 10
- Citations by year
- 2026: 9 · 2025: 4 · 2024: 11 · 2023: 10 · 2022: 8 · 2021: 2
- FWCI
- 4.08
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- Scientific Reports (publishedVersion, cc-by)
- Research Square (acceptedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Maryland Shared Open Access Repository (USMAI Consortium) (submittedVersion, cc-by)
- Digital Repository at the University of Maryland (University of Maryland College Park) (cc-by)
- Funders
- National Institute of General Medical Sciences (R25GM055036)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
- Physical Sciences › Engineering › Aerospace Engineering › Calibration and Measurement Techniques
- Health Sciences › Medicine › Physiology (medical) › Thermoregulation and physiological responses
- Keywords
- non-contact infrared thermometer, body temperature measurement, measurement accuracy, clinical bias, COVID-19 screening
- MeSH terms
- COVID-19, Adult, Body Temperature, Female, Fever, Humans, Infrared Rays, Male, Sensitivity and Specificity, Thermometers, Young Adult, Pandemics
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol and qwen3.8-omni-flash).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Clinical evaluation of non-contact infrared thermometers. https://app.cochat.ai/papers/10.1038/s41598-021-99300-1
- The paper
- Sullivan, S. J. L., Rinaldi, J. E., Hariharan, P., Casamento, J. P., Baek, S., Seay, N., Vesnovsky, O., & Topoleski, L. D. T. (2021). Clinical evaluation of non-contact infrared thermometers. Scientific Reports. https://doi.org/10.1038/s41598-021-99300-1
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.