Clinical evaluation of non-contact infrared thermometers

Open access

Published version of a preprint

The DOI matches its Crossref record; 9 of the 10 references it cites were found in a registry or an index; 9 of the 25 sources its claims cite were checked for support.

Abstract

Non-contact infrared thermometers (NCITs) are being widely used during the COVID-19 pandemic as a temperature-measurement tool for screening and isolating patients in healthcare settings, travelers at ports of entry, and the general public. To understand the accuracy of NCITs, a clinical study was conducted with 1113 adult subjects using six different commercially available NCIT models. A total of 60 NCITs were tested with 10 units for each model. The NCIT-measured temperature was compared with the oral temperature obtained using a reference oral thermometer. The mean difference between the reference thermometer and NCIT measurement (clinical bias) was different for each NCIT model. The clinical bias ranged from just under - 0.9 °C (under-reporting) to just over 0.2 °C (over-reporting). The individual differences ranged from - 3 to + 2 °C in extreme cases, with the majority of the differences between - 2 and + 1 °C. Depending upon the NCIT model, 48% to 88% of the individual temperature measurements were outside the labeled accuracy stated by the manufacturers. The sensitivity of the NCIT models for detecting subject's temperature above 38 °C ranged from 0 to 0.69. Overall, our results indicate that some NCIT devices may not be consistently accurate enough to determine if subject's temperature exceeds a specific threshold of 38 °C. Model-to-model variability and individual model accuracy in the displayed temperature were found to be outside of acceptable limits. Accuracy and credibility of the NCITs should be thoroughly evaluated before using them as an effective screening tool.

Overview

CoChat Summary

## TL;DR

This clinical study evaluated six commercially available non-contact infrared thermometers (NCITs) against oral reference temperatures in more than 1,000 adults and found substantial model-to-model variability. Several devices showed clinically important bias, poor sensitivity for detecting 38 °C, and many readings outside manufacturers’ stated accuracy.

## Key findings

  • Among 1022 subjects in the analyzed results, oral reference temperatures ranged from 36.1 to 40.3 °C, with an average of 36.9 °C in both trials.
  • Clinical bias across the six NCIT models ranged from under-reporting by −0.87 °C to over-reporting by 0.21 °C; Model E had the largest bias (−0.89 °C) and Model C the smallest (0.14 °C).
  • Depending on model, 48% to 88% of individual measurements were outside the manufacturers’ labeled accuracy claims.
  • Sensitivity for detecting a temperature of 38 °C ranged from 0 to 0.69, and specificity ranged from 0.97 to 1.
  • The difference between reference temperature and NCIT temperature was statistically significant for all six models; only Models C and F showed intra-model consistency, and only some model pairs were not significantly different from each other.

## Methods

  • Controlled clinical study of adult subjects at the University Health Center, University of Maryland, College Park; 1113 subjects were enrolled, and the analysis reported 1022 subjects after exclusions.
  • Six commercially available NCIT models were tested, with 10 units per model; each subject had oral temperature measured with a Welch Allyn SureTemp Plus 690 reference thermometer and then NCIT measurements at the forehead, repeated in two trials.
  • Analyses included clinical bias calculations, ANOVA with Tukey post-hoc testing, correlation of measurement error with oral temperature, and sensitivity/specificity calculations across threshold temperatures from 37.0 to 38.0 °C.

## Limitations

  • The paper states that subjects under 18 were not included.
  • The number of subjects with temperatures ≥38 °C was about 5% of the sample; this was stated by the authors as a limitation, although they said there were still sufficient subjects for analysis.
  • The study focused on six NCIT models from different manufacturers that targeted the center of the forehead; the authors note that other potentially important clinical factors such as sex, age, skin tone, and weight were not considered and should be evaluated in future studies.
  • Evident from the design: the study was conducted under controlled conditions, so the findings may not fully reflect real-world screening settings with greater user and environmental variability.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1038/s41598-021-99300-1

Cites

9 of 9 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Figure 1 Accuracy performance statistics for each NCIT model for Trial #1. The midline indicates the median, the box top captures 25% of the data above the median and the box bottom captures 25% of the data below the median. The whiskers (error bars) represent that maximum and the minimum ΔT. The circles represent outlier data.
Figure 1 Accuracy performance statistics for each NCIT model for Trial #1. The midline indicates the median, the box top captures 25% of the data above the median and the box bottom captures 25% of the data below the median. The whiskers (error bars) represent that maximum and the minimum ΔT. The circles represent outlier data.CC BY · PubMed Central
Figure 2 Total counts per error value, per NCIT model for Trial #1 and #2. Green area indicates ± 0.3 °C laboratory accuracy zone; dashed black line indicates the zero error line.
Figure 2 Total counts per error value, per NCIT model for Trial #1 and #2. Green area indicates ± 0.3 °C laboratory accuracy zone; dashed black line indicates the zero error line.CC BY · PubMed Central
Figure 3 ΔT (°C) as a function of Tref (°C).
Figure 3 ΔT (°C) as a function of Tref (°C).CC BY · PubMed Central
Figure 4 Sensitivity and specificity of all NCIT models obtained from Trial #1 dataset
Figure 4 Sensitivity and specificity of all NCIT models obtained from Trial #1 datasetCC BY · PubMed Central

Verification

Claims the paper makes

5 Valid · 1 Not valid · 8 Not checked · 70 Own results and interpretations

41 more claims

Create a free account to see all 84 claims and to verify the 2 references still open.

See all claims — free account
  • Probably not a substantial claim. 27
  • It may state two separate claims as one. 11

Create a free account to see these claims.

Citations

Cites

Show all 9 references

Research works printed but not linked: 1

  1. Non-contact Thermometers for Detecting Fever: A Review of Clinical Effectiveness. Report No. PMID: 25520984 (Canadian Agency for Drugs and Technologies in Health).

Cited by

See all 44

Metadata

Authors
Sullivan, S. J. L., Rinaldi, J. E., Hariharan, P., Casamento, J. P., Baek, S., Seay, N., Vesnovsky, O., & Topoleski, L. D. T.
Affiliations
  • United States Food and Drug Administration (US)
  • University of Maryland, Baltimore County (US)
Corresponding author
Hariharan, Prasanna
Venue
Scientific Reports
Published
2021-11-11
Volume
11
Issue
1
Pages
22079-22079
ISSN
2045-2322 (ISSN-L)
OpenAlex
W3193296718
Cited by
44
References
10
Citations by year
2026: 9 · 2025: 4 · 2024: 11 · 2023: 10 · 2022: 8 · 2021: 2
FWCI
4.08
Type
journal article
Language
en
Open access
Licence
Locations
  • Scientific Reports (publishedVersion, cc-by)
  • Research Square (acceptedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • Maryland Shared Open Access Repository (USMAI Consortium) (submittedVersion, cc-by)
  • Digital Repository at the University of Maryland (University of Maryland College Park) (cc-by)
Funders
National Institute of General Medical Sciences (R25GM055036)
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
  • Physical Sciences › Engineering › Aerospace Engineering › Calibration and Measurement Techniques
  • Health Sciences › Medicine › Physiology (medical) › Thermoregulation and physiological responses
Keywords
non-contact infrared thermometer, body temperature measurement, measurement accuracy, clinical bias, COVID-19 screening
MeSH terms
COVID-19, Adult, Body Temperature, Female, Fever, Humans, Infrared Rays, Male, Sensitivity and Specificity, Thermometers, Young Adult, Pandemics

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol and qwen3.8-omni-flash).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Clinical evaluation of non-contact infrared thermometers. https://app.cochat.ai/papers/10.1038/s41598-021-99300-1
The paper
Sullivan, S. J. L., Rinaldi, J. E., Hariharan, P., Casamento, J. P., Baek, S., Seay, N., Vesnovsky, O., & Topoleski, L. D. T. (2021). Clinical evaluation of non-contact infrared thermometers. Scientific Reports. https://doi.org/10.1038/s41598-021-99300-1

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.