Beam hardening artifacts in micro-computed tomography scanning can be reduced by X-ray beam filtration and the resulting images can be used to accurately measure BMD
Open accessThe DOI matches its Crossref record; 53 of the 53 references it cites were found in a registry or an index.
Abstract
Bone mineral density (BMD) measurements are critical in many research studies investigating skeletal integrity. For preclinical research, micro-computed tomography (μCT) has become an essential tool in these studies. However, the ability to measure the BMD, directly from μCT images can be biased by artifacts such as beam hardening, in the image. This three-part study was designed to understand how the image acquisition process can affect the resulting BMD measurements and to verify that the BMD measurements are accurate. In the first part of this study, the effect of beam hardening-induced cupping artifacts on BMD measurements was examined. In the second part of this study, the number of bones in the X-ray path and the sampling process during scanning was examined. In the third part of this study, μCT-based BMD measurements were compared with ash weights to verify the accuracy of the measurements. The results indicate that beam hardening artifacts of up to 32.6% can occur in sample sizes of interest in studies investigating mineralized tissue and affect mineral density measurements. Beam filtration can be used to minimize these artifacts. The results also indicate that, for murine femora, the scan setup can impact densitometry measurements for both cortical and trabecular bone and morphologic measurements of trabecular bone. Last, when a scan setup that minimized all of these artifacts was used, the μCT based measurements correlated used well with ash weight measurements (R2 = 0.983 when air was excluded), indicating that μCT can be an accurate tool for murine bone densitometry.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Applied Multiple Regression/Correlation Analysis for the Behavioral Sciences
Cohen, P., Cohen, P., West, S. G., & Aiken, L. S. (2014) · Psychology Press eBooks · doi:10.4324/9781410606266 · Cited by 20,737
Applied Multiple Regression/Correlation Analysis for the Behavioral Sciences.
Henry, N. W., Cohen, J., & Cohen, P. (1977) · Contemporary Sociology · doi:10.2307/2064799 · Cited by 7,089
Feldkamp, L. A., Davis, L. C., & Kress, J. W. (1984) · Journal of the Optical Society of America A · doi:10.1364/josaa.1.000612 · Cited by 6,319 · cited 1×
“These projections were corrected using low-end and highend outlier replacement in conjunction with a sinogram based longterm trend correction and reconstructed using a filtered cone-beam backprojection algorithm with a Ram-Lak filter to generate images with an isotropic voxel size of 18 μm [43,44].”
Applied Multiple Regression/Correlation Analysis for the Behavioral Sciences.
Tufte, E. R., Cohen, J., & Cohen, P. (1979) · Journal of the American Statistical Association · doi:10.2307/2286442 · Cited by 5,180 · cited 1×
“In order to compare the ability of the BMCfull and BMCexclude to predict the ash weight, a statistical test comparing the correlation coefficients between these two methods was used [49].”
Energy-selective reconstructions in X-ray computerised tomography
Alvarez, R. E., & Macovski, A. (1976) · Physics in Medicine & Biology · doi:10.1088/0031-9155/21/5/002 · Cited by 2,013 · cited 1×
“This corrects for beam hardening and can be used to minimize cupping, streak, and flare artifacts [10,12,13].”
Artifacts in CT: Recognition and Avoidance
Barrett, J. F., & Keat, N. (2004) · RadioGraphics · doi:10.1148/rg.246045065 · Cited by 1,688 · cited 1×
“This relative density is difficult to verify because the CT image acquisition process is subject to artifacts from partial-voluming, photon starvation, photon scatter, under-sampling, and beam hardening [9].”
(2020) · Elsevier eBooks · doi:10.1016/c2015-1-01622-2 · Cited by 1,683
Ramachandran, G. N., & Lakshminarayanan, A. V. (1971) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.68.9.2236 · Cited by 984 · cited 1×
“These projections were corrected using low-end and highend outlier replacement in conjunction with a sinogram based longterm trend correction and reconstructed using a filtered cone-beam backprojection algorithm with a Ram-Lak filter to generate images with an isotropic voxel size of 18 μm [43,44].”
Beam hardening in X-ray reconstructive tomography
Brooks, R. A., & Chiro, G. D. (1976) · Physics in Medicine and Biology · doi:10.1088/0031-9155/21/3/004 · Cited by 721 · cited 2×
“The acrylic is similar to water and provided the same path length effect that water bags provided in early CT scanners [35], so it was a reasonable first choice as a material.”
Statistical image reconstruction for polyenergetic X-ray computed tomography
Elbakri, I. A., & Fessler, J. A. (2002) · IEEE Transactions on Medical Imaging · doi:10.1109/42.993128 · Cited by 691 · cited 1×
“Last, iterative reconstruction approaches based on Poisson distributions have also been proposed, both with and without the need for segmentation [17–19].”
Cited by
Guidelines for assessment of bone microstructure in rodents using micro–computed tomography
Bouxsein, M. L., Boyd, S. K., Christiansen, B. A., Guldberg, R. E., Jepsen, K. J., & Müller, R. (2010) · Journal of Bone and Mineral Research · doi:10.1002/jbmr.141 · Cited by 4,433
Maire, E., & Withers, P. J. (2014) · International Materials Reviews · doi:10.1179/1743280413y.0000000023 · Cited by 1,374
High-resolution Computed Tomography for Clinical Imaging of Bone Microarchitecture
Burghardt, A. J., Link, T. M., & Majumdar, S. (2011) · Clinical Orthopaedics & Related Research · doi:10.1007/s11999-010-1766-x · Cited by 296
Metadata
- Authors
- Meganck, J. A., Kozloff, K. M., Thornton, M. M., Broski, S. M., & Goldstein, S. A.
- Venue
- Bone
- Publisher
- Elsevier BV
- Published
- 2009-12
- Volume
- 45
- Issue
- 6
- Pages
- 1104-1116
- ISSN
- 1873-2763 (ISSN-L)
- 8756-3282
- OpenAlex
- W2126609373
- Cited by
- 185
- References
- 53
- FWCI
- 8.96
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Beam hardening artifacts in micro-computed tomography scanning can be reduced by X-ray beam filtration and the resulting images can be used to accurately measure BMD. https://app.cochat.ai/papers/10.1016/j.bone.2009.07.078
- The paper
- Meganck, J. A., Kozloff, K. M., Thornton, M. M., Broski, S. M., & Goldstein, S. A. (2009). Beam hardening artifacts in micro-computed tomography scanning can be reduced by X-ray beam filtration and the resulting images can be used to accurately measure BMD. Bone. https://doi.org/10.1016/j.bone.2009.07.078
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.