Coaxial electrospinning multicomponent functional controlled-release vascular graft: Optimization of graft properties
Abstract
Small diameter vascular grafts possessing desirable biocompatibility and suitable mechanical properties have become an urgent clinic demand. Herein, heparin loaded fibrous grafts of collagen/chitosan/poly(l-lactic acid-co-ε-caprolactone) (PLCL) were successfully fabricated via coaxial electrospinning. By controlling the concentration of heparin and the ratio of collagen/chitosan/PLCL, most grafts had the heparin encapsulation efficiency higher than 70%, and the heparin presented sustained release for more than 45 days. Particularly, such multicomponent grafts had relative low initial burst release, and after heparin releasing for 3 weeks, the grafts still showed good anti-platelet adhesion ability. In addition, along with the excellent cell biocompatibility, the fabricated grafts possessed suitable mechanical properties including good tensile strength, suture retention strength, burst pressure and compliance which could well match the native blood vessels. Thus, the optimized graft properties could be properly addressed for vascular tissue application via coaxial electrospinning.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Adding heparin to PLCL vascular stents enhances anticoagulant performance.
“most studies choose to add heparin to PLCL vascular stents to enhance the anticoagulant performance of the stents [156–158].”
From Application of PLCL as a biodegradable polymer in biomedical engineering (2026) · cites this paper as [158]
Cites
1 retracted
Zhai, W., Qiu, L., Mo, X., et al. (2013) · Journal of Biomedical Materials Research Part B Applied Biomaterials · doi:10.1002/jbm.b.32972 · Cited by 31 · see citing works
Self-citationRetractedLutolf, M. P., & Hubbell, J. A. (2005) · Nature Biotechnology · doi:10.1038/nbt1055 · Cited by 4,551 · see citing works
Structure and process relationship of electrospun bioabsorbable nanofiber membranes
Zong, X., Kim, K., Fang, D., Ran, S., Hsiao, B. S., & Chu, B. (2002) · Polymer · doi:10.1016/s0032-3861(02)00275-6 · Cited by 1,931 · see citing works
Human tissue-engineered blood vessels for adult arterial revascularization
L’Heureux, N., Dusserre, N., König, G., Victor, B. E., Keire, P. A., Wight, T. N., Chronos, N. A. F., Kyles, A. E., Gregory, C. R., Hoyt, G. E., et al. (2006) · Nature Medicine · doi:10.1038/nm1364 · Cited by 1,083 · see citing works
König, G., McAllister, T. N., Dusserre, N., Garrido, S. A., Iyican, C., Marini, A., Fiorillo, A., Avila, H., Wystrychowski, W., Zagalski, K., et al. (2008) · Biomaterials · doi:10.1016/j.biomaterials.2008.11.011 · Cited by 582 · see citing works
Saphenous Vein Versus PTFE for Above-Knee Femoropopliteal Bypass. A Review of the Literature
Klinkert, P., Post, P. N., Breslau, P. J., & van Bockel, J. H. (2004) · European Journal of Vascular and Endovascular Surgery · doi:10.1016/j.ejvs.2003.12.027 · Cited by 518 · see citing works
Compressive and shear properties of alginate gel: Effects of sodium ions and alginate concentration
LeRoux, M. A., Guilak, F., & Setton, L. A. (1999) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(199910)47:1 · Cited by 449 · see citing works
Chlupáč, J., Filová, E., & Bačáková, L. (2009) · Physiological Research · doi:10.33549/physiolres.931918 · Cited by 409 · see citing works
Xu, C., Inai, R., Kotaki, M., & Ramakrishna, S. (2004) · Tissue Engineering · doi:10.1089/ten.2004.10.1160 · Cited by 399 · see citing works
Compliance properties of conduits used in vascular reconstruction
Tai, N. R. M., Salacinski, H. J., Edwards, A., et al. (2000) · British journal of surgery · doi:10.1046/j.1365-2168.2000.01566.x · Cited by 269 · see citing works
Su, Y., Su, Q., Liu, W., Lim, M., Venugopal, J. R., Mo, X., Ramakrishna, S., Al‐Deyab, S. S., & El-Newehy, M. H. (2011) · Acta Biomaterialia · doi:10.1016/j.actbio.2011.11.002 · Cited by 261 · see citing works
Self-citationHeparin release from thermosensitive hydrogels
Gutowska, A., Bae, Y. H., Feijén, J., & Kim, S. W. (1992) · Journal of Controlled Release · doi:10.1016/0168-3659(92)90194-v · Cited by 200 · see citing works
Mechanisms of Stiffening and Strengthening in Media-Equivalents Fabricated Using Glycation
Girton, T. S., Oegema, T. R., Grassl, E. D., Isenberg, B. C., & Tranquillo, R. T. (2000) · Journal of Biomechanical Engineering · doi:10.1115/1.429652 · Cited by 164 · see citing works
Fabrication of silk fibroin blended P(LLA‐CL) nanofibrous scaffolds for tissue engineering
Zhang, K., Wang, H., Huang, C., Su, Y., Mo, X., & Ikada, Y. (2009) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32504 · Cited by 151 · see citing works
Self-citationFunctionalized scaffolds to enhance tissue regeneration
Guo, B., Lei, B., Li, P., & Ma, P. X. (2015) · Regenerative Biomaterials · doi:10.1093/rb/rbu016 · Cited by 148 · see citing works
Li, Y., Neoh, K. G., & Kang, E. T. (2005) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.30286 · Cited by 138 · see citing works
Rogers, C., Karnovsky, M. J., & Edelman, E. R. (1993) · Circulation · doi:10.1161/01.cir.88.3.1215 · Cited by 135 · see citing works
Thiol‐ene/acrylate substrates for softening intracortical electrodes
Ware, T. H., Simon, D., Liu, H. C., Musa, T., Vasudevan, S., Sloan, A. M., Keefer, E. W., Rennaker, R. L., & Voit, W. (2013) · Journal of Biomedical Materials Research Part B Applied Biomaterials · doi:10.1002/jbmb.32946 · Cited by 126 · see citing works
Yin, A., Zhang, K., McClure, M. J., Huang, C., Wu, J., Fang, J., Mo, X., Bowlin, G. L., Al‐Deyab, S. S., & El-Newehy, M. H. (2012) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.34434 · Cited by 117 · see citing works
Self-citationHealing characteristics of electrospun polyurethane grafts with various porosities
Bergmeister, H., Schreiber, C., Grasl, C., Walter, I., Plasenzotti, R., Stoiber, M., Bernhard, D., & Schima, H. (2012) · Acta Biomaterialia · doi:10.1016/j.actbio.2012.12.009 · Cited by 117 · see citing works
Perivascular graft heparin delivery using biodegradable polymer wraps
Edelman, E. R., Nathan, A. T., Katada, M., Gates, J. D., & Karnovsky, M. J. (2000) · Biomaterials · doi:10.1016/s0142-9612(00)00154-x · Cited by 84 · see citing works
Preparation of core‐shell biodegradable microfibers for long‐term drug delivery
Huang, H., He, C., Wang, H., & Mo, X. (2009) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32543 · Cited by 77 · see citing works
Self-citationHuang, C., Wang, S., Qiu, L., Ke, Q., Zhai, W., & Mo, X. (2013) · ACS Applied Materials & Interfaces · doi:10.1021/am400099p · Cited by 71 · see citing works
Self-citationFabrication and properties of core‐shell structure P(LLA‐CL) nanofibers by coaxial electrospinning
Li, X., Su, Y., Chen, R., et al. (2008) · Journal of Applied Polymer Science · doi:10.1002/app.29056 · Cited by 54 · see citing works
Su, Y., Li, X., Liu, Y., et al. (2010) · Journal of Biomaterials Science Polymer Edition · doi:10.1163/092050609x12583785588757 · Cited by 41 · see citing works
Punet, X., Mauchauffé, R., Rodríguez‐Cabello, J. C., et al. (2015) · Regenerative Biomaterials · doi:10.1093/rb/rbv014 · Cited by 36 · see citing works
Yin, A., Li, J., Bowlin, G. L., et al. (2014) · Colloids and Surfaces B Biointerfaces · doi:10.1016/j.colsurfb.2014.04.011 · Cited by 33 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 1
- Cavender, 1990, The effects of heparin bonded tantalum stents on thrombosis and neointimal proliferation, Circulation, 82, 541
Cited by
Sorted by citations
Application of PLCL as a biodegradable polymer in biomedical engineering
Zhao, Y., & Liu, H. (2026) · Journal of Materials Science Materials in Medicine · doi:10.1007/s10856-026-07016-3 · Cited by 7 · see citing works
“In order to solve this problem, most studies choose to add heparin to PLCL vascular stents to enhance the anticoagulant performance of the stents [156 – 158].”
- Authors
- Yin, A., Luo, R., Li, J., Mo, X., Wang, Y., & Zhang, X.
- Affiliations
- Sichuan University (CN)
- Donghua University (CN)
- Corresponding author
- Yin, Anlin
- Mo, Xiumei
- Wang, Yunbing
- Venue
- Colloids and Surfaces B: Biointerfaces
- Publisher
- Elsevier BV
- Published
- 2017-04
- Volume
- 152
- Pages
- 432-439
- ISSN
- 0927-7765 (ISSN-L)
- OpenAlex
- W2580556381
- Cited by
- 45
- References
- 27
- Citations by year
- 2026: 2 · 2025: 4 · 2024: 6 · 2023: 5 · 2022: 9 · 2021: 4 · 2020: 7 · 2019: 5 · 2018: 3
- FWCI
- 1.48
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Colloids and Surfaces B Biointerfaces (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Natural Science Foundation of China (31500784)
- China Postdoctoral Science Foundation Funded Project (2015M580790)
- Post-doctoral Special Fund of Sichuan University
- Program of Introducing Talents of Discipline to Universities (B16033)
- Topics
- Physical Sciences › Materials Science › Biomaterials › Electrospun Nanofibers in Biomedical Applications
- Health Sciences › Medicine › Surgery › Tissue Engineering and Regenerative Medicine
- Physical Sciences › Materials Science › Biomaterials › Silk-based biomaterials and applications
- Keywords
- coaxial electrospinning, vascular grafts, heparin, controlled release, collagen, chitosan, PLCL, biocompatibility, mechanical properties, suture retention strength, burst pressure, compliance, tensile strength
- MeSH terms
- Animals, Biocompatible Materials, Blood Vessel Prosthesis, Collagen, Delayed-Action Preparations, Humans, Materials Testing, Polyesters, Tensile Strength, Tissue Engineering, Chitosan
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review