Gating Approaches in Cardiac PET Imaging
Open accessAbstract
Cardiac PET provides high sensitivity and high negative predictive value in the diagnosis of coronary artery disease and cardiomyopathies. Cardiac, respiratory as well as bulk patient motion have detrimental effects on thoracic PET imaging, in particular on cardiovascular PET imaging where the motion can affect the PET images quantitatively as well as qualitatively. Gating can ameliorate the unfavorable impact of motion additionally enabling evaluation of left ventricular systolic function. In this article, the authors review the recent advances in gating approaches and highlight the advances in data-driven approaches, which hold promise in motion detection without the need for complex hardware setup.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In cardiac PET imaging, contraction of the myocardium contributes to quantitative inaccuracy of uncorrected static PET protocols.
“In cardiac PET imaging, contraction of the myocardium further contributes to the quantitative inaccuracy of uncorrected static PET protocols [90,91].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [90]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Joshi, N. V., Vesey, A. T., Williams, M. C., Shah, A. S. V., Calvert, P. A., Craighead, F., Yeoh, S. E., Wallace, W. A., Salter, D. M., Fletcher, A. M., et al. (2013) · The Lancet · doi:10.1016/s0140-6736(13)61754-7 · Cited by 953 · see citing works
Martínez-Möller, A., Souvatzoglou, M., Delso, G., Bundschuh, R. A., Chefd’hotel, C., Ziegler, S. I., Navab, N., Schwaiger, M., & Nekolla, S. G. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.054726 · Cited by 712 · see citing works
Blankstein, R., Osborne, M. T., Naya, M., Waller, A. H., Kim, C. K., Murthy, V. L., Kazemian, P., Kwong, R. Y. K., Tokuda, M., Skali, H., et al. (2013) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2013.09.022 · Cited by 709 · see citing works
Dilsizian, V., Bacharach, S. L., Beanlands, R. S., Bergmann, S. R., Delbeke, D., Dorbala, S., Gropler, R. J., Knuuti, J. M., Schelbert, H. R., & Travin, M. I. (2016) · Journal of Nuclear Cardiology · doi:10.1007/s12350-016-0522-3 · Cited by 616 · see citing works
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners
Kinahan, P. E., Hasegawa, B. H., & Beyer, T. (2003) · Seminars in Nuclear Medicine · doi:10.1053/snuc.2003.127307 · Cited by 515 · see citing works
Coronary Arterial 18F-Sodium Fluoride Uptake
Dweck, M. R., Chow, M. W. L., Joshi, N. V., Williams, M. C., Jones, C., Fletcher, A. M., Richardson, H., White, A. C., McKillop, G., van Beek, E. J. R., et al. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2011.12.037 · Cited by 485 · see citing works
Treatment of axial data in three-dimensional PET.
Daube-Witherspoon, M. E., & Muehllehner, G. (1987) · PubMed · Cited by 357 · see citing works
Focus on time-of-flight PET: the benefits of improved time resolution
Conti, M. (2011) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1711-y · Cited by 335 · see citing works
Respiratory Motion in Positron Emission Tomography/Computed Tomography: A Review
Nehmeh, S. A., & Erdi, Y. E. (2008) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2008.01.002 · Cited by 318 · see citing works
Correlation of lung tumor motion with external surrogate indicators of respiration
Hoisak, J. D. P., Sixel, K. E., Tirona, R., Cheung, P. C. F., & Pignol, J. (2004) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2004.07.681 · Cited by 318 · see citing works
Imaging cardiac amyloidosis: a pilot study using 18F-florbetapir positron emission tomography
Dorbala, S., Vangala, D., Semer, J., Strader, C., Bruyere, J. R., Carli, M. F. D., Moore, S. C., & Falk, R. H. (2014) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-014-2787-6 · Cited by 316 · see citing works
Gould, K. L., Pan, T., Loghin, C., Johnson, N. P., Guha, A., & Sdringola, S. M. (2007) · Journal of Nuclear Medicine · doi:10.2967/jnumed.107.039792 · Cited by 292 · see citing works
Clinical Myocardial Perfusion PET/CT
Carli, M. F. D., Dorbala, S., Meserve, J. D., Fakhri, G. E., Sitek, A., & Moore, S. C. (2007) · Journal of Nuclear Medicine · doi:10.2967/jnumed.106.032789 · Cited by 289 · see citing works
SARRAZIN, J., Philippon, F., Tessier, M., Guimond, J. V., Molin, F., Champagne, J., Nault, I., Blier, L. E., Nadeau, M., Charbonneau, L., et al. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2011.11.059 · Cited by 284 · see citing works
Cardiac Positron Emission Tomography
Bengel, F. M., Higuchi, T., Javadi, M. S., & Läutamaki, R. (2009) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.02.065 · Cited by 274 · see citing works
Dorbala, S., Hachamovitch, R., Curillova, Z., Thomas, D., Vangala, D., Kwong, R. Y. K., & Carli, M. F. D. (2009) · JACC. Cardiovascular imaging · doi:10.1016/j.jcmg.2009.04.009 · Cited by 271 · see citing works
Cardiac sarcoidosis: state-of-the-art review
Kouranos, V., & Sharma, R. (2021) · Heart · doi:10.1136/heartjnl-2019-316442 · Cited by 257 · see citing works
Dorbala, S., Vangala, D., Sampson, U. K. A., Limaye, A. M., Kwong, R. K., & Carli, M. F. D. (2007) · PubMed · Cited by 240 · see citing works
The impact of respiratory motion on tumor quantification and delineation in static PET/CT imaging
Liu, C., Pierce, L. A., Alessio, A. M., & Kinahan, P. E. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/24/007 · Cited by 233 · see citing works
Dawood, M., Büther, F., Lang, N., Schober, O., & Schäfers, K. P. (2007) · Medical Physics · doi:10.1118/1.2748104 · Cited by 211 · see citing works
List Mode–Driven Cardiac and Respiratory Gating in PET
Büther, F., Dawood, M., Stegger, L., Wübbeling, F., Schäfers, M. A., Schober, O., & Schäfers, K. P. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.059204 · Cited by 205 · see citing works
Cardiac sarcoidosis-state of the art review.
Hulten, E. A., Aslam, S. N., Osborne, M. T., Abbasi, S. A., Bittencourt, M. S., & Blankstein, R. (2016) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas) · doi:10.3978/j.issn.2223-3652.2015.12.13 · Cited by 199 · see citing works
Sureshbabu, W., & Mawlawi, O. R. (2005) · PubMed · Cited by 196 · see citing works
Martínez-Möller, A., Souvatzoglou, M., Navab, N., Schwaiger, M., & Nekolla, S. G. (2007) · PubMed · Cited by 166 · see citing works
Displacement and velocity of the coronary arteries: cardiac and respiratory motion
Shechter, G., Resar, J. R., & McVeigh, E. R. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2005.862752 · Cited by 164 · see citing works
Lertsburapa, K., Ahlberg, A. W., Bateman, T. M., Katten, D. M., Volker, L. L., Cullom, S. J., & Heller, G. V. (2008) · Journal of Nuclear Cardiology · doi:10.1016/j.nuclcard.2008.06.168 · Cited by 150 · see citing works
Phase and amplitude binning for 4D-CT imaging
Abdelnour, A. F., Nehmeh, S. A., Pan, T., Humm, J. L., Vernon, P., Schöder, H. M., Rosenzweig, K., Mageras, G. S., Yorke, E. D., Larson, S. M., et al. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/12/012 · Cited by 150 · see citing works
Strategies for Motion Tracking and Correction in PET
Rahmim, A., ROUSSET, O. G., & Zaidi, H. (2007) · PET Clinics · doi:10.1016/j.cpet.2007.08.002 · Cited by 143 · see citing works
Motion Correction in Dual Gated Cardiac PET Using Mass-Preserving Image Registration
Gigengack, F., Ruthotto, L., Burger, M., Wolters, C. H., Jiang, X., & Schäfers, K. P. (2011) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2011.2175402 · Cited by 138 · see citing works
18F-FDG PET/CT for the Assessment of Myocardial Sarcoidosis
Skali, H., Schulman, A. R., & Dorbala, S. (2013) · Current Cardiology Reports · doi:10.1007/s11886-013-0370-6 · Cited by 137 · see citing works
Dual cardiac–respiratory gated PET: implementation and results from a feasibility study
Martínez-Möller, A., Zikic, D., Botnar, R. M., Bundschuh, R. A., Howe, W., Ziegler, S. I., Navab, N., Schwaiger, M., & Nekolla, S. G. (2007) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-007-0374-9 · Cited by 131 · see citing works
Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI
Keller, S. H., Holm, S., Hansen, A. E., Sattler, B., Andersen, F. L., Klausen, T. L., Højgaard, L., Kjær, A., & Beyer, T. (2012) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-012-0345-4 · Cited by 125 · see citing works
Wolthaus, J., van Herk, M., Muller, S. H., Belderbos, J. S. A., Lebesque, J. V., de Bois, J. A., Rossi, M. M. G., & Damen, E. M. F. (2005) · Physics in Medicine and Biology · doi:10.1088/0031-9155/50/7/017 · Cited by 124 · see citing works
Self-gated MRI motion modeling for respiratory motion compensation in integrated PET/MRI
Grimm, R., Fürst, S., Souvatzoglou, M., Forman, C., Hutter, J. M., Dregely, I., Ziegler, S. I., Kiefer, B., Joachim Hornegger, A., Block, K. T., et al. (2014) · Medical Image Analysis · doi:10.1016/j.media.2014.08.003 · Cited by 114 · see citing works
on behalf of the European Society of Cardiovascular Radiology (ESCR), Nensa, F., Bamberg, F., Rischpler, C., Menezes, L., Poeppel, T. D., la Fougère, C. J., Beitzke, D., Rasul, S., Loewe, C., et al. (2018) · European Radiology · doi:10.1007/s00330-017-5008-4 · Cited by 113 · see citing works
Self-citationAttenuation correction of PET cardiac data with low‐dose average CT in PET/CT
Pan, T., Mawlawi, O. R., Luo, D., Liu, H. H., Chi, P. M., Mar, M. V., Gladish, G. W., Truong, M. T., Erasmus, J., Liao, Z. X., et al. (2006) · Medical Physics · doi:10.1118/1.2349843 · Cited by 111 · see citing works
Respiratory Gating for Radiotherapy: Main Technical Aspects and Clinical Benefits
Giraud, P., & Houle, A. (2013) · ISRN Pulmonology · doi:10.1155/2013/519602 · Cited by 106 · see citing works
18F-FDG PET/CT for the Assessment of Myocardial Sarcoidosis
Skali, H., Schulman, A. R., & Dorbala, S. (2013) · Current Cardiology Reports · doi:10.1007/s11886-013-0352-8 · Cited by 104 · see citing works
Quiescent period respiratory gating for PET/CT
Liu, C., Alessio, A. M., Pierce, L. A., Thielemans, K., Wollenweber, S. D., Ganin, A., & Kinahan, P. E. (2010) · Medical Physics · doi:10.1118/1.3480508 · Cited by 104 · see citing works
Influence of Cardiac and Respiratory Motion on Tomographic Reconstructions of the Heart
Ter‐Pogossian, M. M., Bergmann, S. R., & Sobel, B. E. (1982) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198212000-00016 · Cited by 99 · see citing works
FDG PET/CT for Early Detection and Localization of Left Ventricular Assist Device Infection
Kim, J., Feller, E. D., Chen, W., Liang, Y., & Dilsizian, V. (2018) · JACC. Cardiovascular imaging · doi:10.1016/j.jcmg.2018.01.024 · Cited by 95 · see citing works
Adenosine stimulates respiration in man.
Watt, A., & Routledge, P. (1985) · British Journal of Clinical Pharmacology · doi:10.1111/j.1365-2125.1985.tb05108.x · Cited by 95 · see citing works
Bundschuh, R. A., Martínez-Möller, A., Essler, M., Martínez, M.-J., Nekolla, S. G., Ziegler, S. I., & Schwaiger, M. (2007) · Journal of Nuclear Medicine · doi:10.2967/jnumed.106.035279 · Cited by 92 · see citing works
Kesner, A., & Kuntner, C. (2010) · Medical Physics · doi:10.1118/1.3483784 · Cited by 90 · see citing works
Advances in SPECT and PET Hardware
Slomka, P. J., Pan, T., Berman, D. S., & Germano, G. (2015) · Progress in Cardiovascular Diseases · doi:10.1016/j.pcad.2015.02.002 · Cited by 88 · see citing works
Self-citationBateman, T. M., Dilsizian, V., Beanlands, R. S., DePuey, E. G., Heller, G. V., & Wolinsky, D. A. (2016) · Journal of Nuclear Medicine · doi:10.2967/jnumed.116.180448 · Cited by 85 · see citing works
Motion Correction of 18F-NaF PET for Imaging Coronary Atherosclerotic Plaques
Rubeaux, M., Joshi, N. V., Dweck, M. R., Fletcher, A. M., Motwani, M., Thomson, L., Germano, G., Dey, D., Li, D., Berman, D. S., et al. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.162990 · Cited by 83 · see citing works
Self-citationOptimal number of respiratory gates in positron emission tomography: A cardiac patient study
Dawood, M., Büther, F., Stegger, L., Jiang, X., Schober, O., Schäfers, M. A., & Schäfers, K. P. (2009) · Medical Physics · doi:10.1118/1.3112422 · Cited by 79 · see citing works
Kesner, A., Schleyer, P. J., Büther, F., Walter, M. A., Schäfers, K. P., & Koo, P. J. (2014) · EJNMMI Physics · doi:10.1186/2197-7364-1-8 · Cited by 78 · see citing works
Büther, F., Ernst, I., Dawood, M., Kraxner, P., Schäfers, M. A., Schober, O., & Schäfers, K. P. (2010) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1533-y · Cited by 78 · see citing works
Dual-gated cardiac PET–Clinical feasibility study
Teräs, M., Kokki, T., Durand-Schaefer, N., Noponen, T., Pietilä, M., Kišš, J., Hoppela, E., Sipilä, H. T., & Knuuti, J. M. (2009) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-009-1252-4 · Cited by 69 · see citing works
Beyer, T., Lassen, M. L., Boellaard, R., Delso, G., Yaqub, M., Sattler, B., & Quick, H. H. (2016) · Magnetic Resonance Materials in Physics, Biology and Medicine · doi:10.1007/s10334-015-0505-4 · Cited by 66 · see citing works
Self-citationRole of PET to evaluate coronary microvascular dysfunction in non-ischemic cardiomyopathies
Bravo, P. E., Carli, M. F. D., & Dorbala, S. (2017) · Heart Failure Reviews · doi:10.1007/s10741-017-9628-1 · Cited by 66 · see citing works
An investigation of sensitivity limits in PET scanners
Eriksson, L., Townsend, D. W., Conti, M., Eriksson, M., Rothfuß, H., Schmand, M. J., Casey, M. E., & Bendriem, B. (2007) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2007.06.112 · Cited by 65 · see citing works
Evaluation of respiratory and cardiac motion correction schemes in dual gated PET/CT cardiac imaging
Lamare, F., Maitre, A. L., Dawood, M., Schäfers, K. P., Fernandez, P., Rimoldi, O. E., & Visvikis, D. (2014) · Medical Physics · doi:10.1118/1.4881099 · Cited by 58 · see citing works
Armstrong, I. S., Tonge, C. M., & Arumugam, P. (2014) · Journal of Nuclear Cardiology · doi:10.1007/s12350-014-9858-8 · Cited by 55 · see citing works
A Novel Method for Respiratory Motion Gated With Geometric Sensitivity of the Scanner in 3D PET
He, J., O'Keefe, G. J., Gong, S. J., Jones, G. R., Saunder, T., Scott, A. M., & Geso, M. (2008) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2008.2001187 · Cited by 55 · see citing works
Dual-Gated Motion-Frozen Cardiac PET with Flurpiridaz F 18
Slomka, P. J., Rubeaux, M., Meunier, L. L., Dey, D., Lazewatsky, J. L., Pan, T., Dweck, M. R., Newby, D. E., Germano, G., & Berman, D. S. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.164285 · Cited by 53 · see citing works
Self-citationHybrid Imaging: Instrumentation and Data Processing
Cal-González, J., Rausch, I., Sundar, L. K. S., Lassen, M. L., Muzik, O., Moser, E., Papp, L., & Beyer, T. (2018) · Frontiers in Physics · doi:10.3389/fphy.2018.00047 · Cited by 52 · see citing works
Self-citationMuñoz, C., Neji, R., da Cruz, G. J. L., Mallia, A., Jeljeli, S., Reader, A. J., Botnar, R. M., & Prieto, C. (2017) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26690 · Cited by 51 · see citing works
Contrôle de la respiration en radiothérapie: principaux aspects techniques et intérêts cliniques
Giraud, P., & Garcia, R. (2010) · Bulletin du Cancer · doi:10.1684/bdc.2010.1143 · Cited by 50 · see citing works
Motion frozen 18F-FDG cardiac PET
Meunier, L. L., Slomka, P. J., Dey, D., Ramesh, A., Thomson, L. E. J., Hayes, S. W., Friedman, J. D., Cheng, V., Germano, G., & Berman, D. S. (2010) · Journal of Nuclear Cardiology · doi:10.1007/s12350-010-9322-3 · Cited by 49 · see citing works
Self-citationLassen, M. L., Kwieciński, J., Cadet, S. J., Dey, D., Wang, C., Dweck, M. R., Berman, D. S., Germano, G., Newby, D. E., & Slomka, P. J. (2018) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.217877 · Cited by 49 · see citing works
Self-citationEnhancing Cardiac PET by Motion Correction Techniques
Rubeaux, M., Doris, M. K., Alessio, A., & Slomka, P. J. (2017) · Current Cardiology Reports · doi:10.1007/s11886-017-0825-2 · Cited by 48 · see citing works
Self-citationChander, A., Brenner, M., Läutamaki, R., Voicu, C., Merrill, J., & Bengel, F. M. (2008) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.053819 · Cited by 48 · see citing works
Assessment of attenuation correction for myocardial PET imaging using combined PET/MRI
Lassen, M. L., Rasul, S., Beitzke, D., Stelzmueller, M.-E., Cal-González, J., Hacker, M., & Beyer, T. (2017) · Journal of Nuclear Cardiology · doi:10.1007/s12350-017-1118-2 · Cited by 47 · see citing works
Self-citationBrown, T. L. Y., Merrill, J., Volokh, L., & Bengel, F. M. (2007) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-007-0603-2 · Cited by 47 · see citing works
AlJaroudi, W., Alraies, M. C., Hachamovitch, R., Jaber, W. A., Brunken, R. C., Cerqueira, M. D., & Marwick, T. H. (2012) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-012-2171-3 · Cited by 43 · see citing works
[ 18 F]-NaF PET/CT Identifies Active Calcification in Carotid Plaque
Cocker, M. S., Spence, J. D., Hammond, R. R., Wells, G., deKemp, R. A., Lum, C., Adeeko, A., Yaffe, M. J., Leung, E., Hill, A. C., et al. (2016) · JACC. Cardiovascular imaging · doi:10.1016/j.jcmg.2016.03.005 · Cited by 43 · see citing works
Kwieciński, J., Adamson, P. D., Lassen, M. L., Doris, M. K., Moss, A. J., Cadet, S., Jansen, M. A., Dey, D., Lee, S., Yun, M., et al. (2018) · Circulation Cardiovascular Imaging · doi:10.1161/circimaging.118.008325 · Cited by 42 · see citing works
Self-citationMR-Based Cardiac and Respiratory Motion-Compensation Techniques for PET-MR Imaging
Muñoz, C., Kolbitsch, C., Reader, A. J., Marsden, P. K., Schaeffter, T., & Prieto, C. (2016) · PET Clinics · doi:10.1016/j.cpet.2015.09.004 · Cited by 42 · see citing works
Muñoz, C., Kunze, K., Neji, R., Vitadello, T., Rischpler, C., Botnar, R. M., Nekolla, S. G., & Prieto, C. (2018) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-018-4047-7 · Cited by 37 · see citing works
Left ventricular dyssynchrony assessment by phase analysis from gated PET-FDG scans
Pazhenkottil, A. P., Buechel, R. R., Nkoulou, R. N., Ghadri, J. R., Herzog, B. A., Husmann, L., Wolfrum, M., Kuest, S. M., Fiechter, M., Gaemperli, O., et al. (2011) · Journal of Nuclear Cardiology · doi:10.1007/s12350-011-9411-y · Cited by 32 · see citing works
Joint PET-MR respiratory motion models for clinical PET motion correction
Manber, R., Thielemans, K., Hutton, B. F., Wan, S., McClelland, J. R., Barnes, A., Arridge, S., Sebastien Ourselin, A., & Atkinson, D. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/17/6515 · Cited by 30 · see citing works
Memmott, M. J., Tonge, C. M., Saint, K. J., & Arumugam, P. (2017) · Journal of Nuclear Cardiology · doi:10.1007/s12350-016-0767-x · Cited by 29 · see citing works
Kong, E., Lee, S., & Cho, I.-H. (2013) · Nuclear Medicine and Molecular Imaging · doi:10.1007/s13139-013-0201-0 · Cited by 24 · see citing works
Motion Correction and Its Impact on Absolute Myocardial Blood Flow Measures with PET
Piccinelli, M., Votaw, J. R., & Garcia, E. V. (2018) · Current Cardiology Reports · doi:10.1007/s11886-018-0977-8 · Cited by 22 · see citing works
Hacker, M. (2011) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-011-2027-2 · Cited by 13 · see citing works
Demons versus level-set motion registration for coronary 18F-sodium fluoride PET
Rubeaux, M., Joshi, N. V., Dweck, M. R., Fletcher, A. M., Motwani, M., Thomson, L., Germano, G., Dey, D., Berman, D. S., Newby, D. E., et al. (2016) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2217179 · Cited by 12 · see citing works
Self-citationDasari, P. K. R., Jones, J. P., Casey, M., Liang, Y., Dilsizian, V., & Smith, M. F. (2018) · Journal of Nuclear Cardiology · doi:10.1007/s12350-018-1311-y · Cited by 9 · see citing works
Data-driven dual-gating for cardiac PET
Thielemans, K., Schleyer, P., Marsden, P. K., Teuho, J., Teräs, M., Bettinardi, V., Menezes, L., Manjeshwar, R. M., & Stearns, C. W. (2014) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2014.7430902 · Cited by 8 · see citing works
Respiratory gating in cardiac PET: Effects of adenosine and dipyridamole
Lassen, M. L., Rasmussen, T., Christensen, T. E., Kjær, A., & Hasbak, P. (2016) · Journal of Nuclear Cardiology · doi:10.1007/s12350-016-0631-z · Cited by 8 · see citing works
Self-citationClinical Impact of Cardiac-Gated PET Imaging
Nekolla, S. G., Dinges, J., & Rischpler, C. (2012) · PET Clinics · doi:10.1016/j.cpet.2012.10.002 · Cited by 7 · see citing works
Technical consideration for dual ECG/respiratory-gated cardiac PET imaging
Hyun, M. C., Gerlach, J., Rubeaux, M., & Slomka, P. J. (2016) · Journal of Nuclear Cardiology · doi:10.1007/s12350-016-0741-7 · Cited by 5 · see citing works
Self-citationGould, K. L. (2016) · Journal of Nuclear Cardiology · doi:10.1007/s12350-016-0666-1 · Cited by 1 · see citing works
Flagged first, then by citations
2 of the 87 references couldn’t be matched to a known paper.
Research works printed but not linked: 3
- [62]Munoz C, Neji R, Cruz G, Motion-corrected simultaneous cardiac positron emission tomography and coronary MR angiography with high acquisition efficiency. Magn Reson Med. 2018; 79(1): 339– 350. doi: 10.1002/mrm.26690 28426162 PMC5763353 · DOI
- [69]Chesler DA. 3-Dimensional activity distribution from multiple positron scintigraphs. In: Journal of Nuclear Medicine. Vol 12; 1971: 347.
- [74]Martinez-Möller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG. Artifacts from Misaligned CT in Cardiac Perfusion Solutions. J Nucl Med. 2007; 48(2): 188– 194. 17268013
Cited by
Sorted by citations
Assessment of myocardial fibrosis in patients with systemic sclerosis using [68Ga]Ga-FAPI-04-PET-CT
Treutlein, C., Distler, J. H. W., Taşçılar, K., Fakhouri, S. C., Györfi, A., Atzinger, A., Matei, A., Dees, C., Büttner-Herold, M. J., Kuwert, T., et al. (2022) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-022-06081-4 · Cited by 58 · see citing works
Lassen, M. L., Kwieciński, J., Dey, D., Cadet, S. J., Germano, G., Berman, D. S., Adamson, P. D., Moss, A. J., Dweck, M. R., Newby, D. E., et al. (2019) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-019-04437-x · Cited by 55 · see citing works
Self-citationInfluences on PET Quantification and Interpretation
Rogasch, J. M. M., Hofheinz, F., van Heek, L., Voltin, C.-A., Boellaard, R., & Kobe, C. (2022) · Diagnostics · doi:10.3390/diagnostics12020451 · Cited by 50 · see citing works
“In cardiac PET imaging, contraction of the myocardium further contributes to the quantitative inaccuracy of uncorrected static PET protocols [90,91].”
BackgroundNon-invasive photoacoustic computed tomography of rat heart anatomy and function
Lin, L., Tong, X., Cavallero, S., Zhang, Y., Na, S., Cao, R., Hsiai, T. K., & Wang, L. V. (2023) · Light Science & Applications · doi:10.1038/s41377-022-01053-7 · Cited by 42 · see citing works
Andrews, J. P. M., Macnaught, G., Moss, A. J., Doris, M. K., Pawade, T., Adamson, P. D., van Beek, E. J. R., Lucatelli, C., Lassen, M. L., Robson, P. M., et al. (2019) · Journal of Nuclear Cardiology · doi:10.1007/s12350-019-01962-y · Cited by 29 · see citing works
Self-citationMotion-correction strategies for enhancing whole-body PET imaging
Wang, J., Bermudez, D. H., Chen, W., Durgavarjhula, D., Randell, C., Uyanik, M., & McMillan, A. B. (2024) · Frontiers in Nuclear Medicine · doi:10.3389/fnume.2024.1257880 · Cited by 22 · see citing works
Advances in the Assessment of Coronary Artery Disease Activity with PET/CT and CTA
Kwieciński, J., Wolny, R., Chwała, A., & Slomka, P. J. (2023) · Tomography · doi:10.3390/tomography9010026 · Cited by 16 · see citing works
Self-citationAdvances in Quantitative Analysis of 18 F-Sodium Fluoride Coronary Imaging
Kwieciński, J., Lassen, M. L., & Slomka, P. J. (2021) · Molecular Imaging · doi:10.1155/2021/8849429 · Cited by 12 · see citing works
Self-citationSeparated Respiratory Phases for In Vivo Ultrasonic Thermal Strain Imaging
Yin, C., Wang, G., Xie, Y., Tu, J., Sun, W., Kong, X., Guo, X., & Zhang, D. (2022) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2022.3149287 · Cited by 11 · see citing works
Long Axial Field of View PET/CT: Technical Aspects in Cardiovascular Diseases
Martínez-Lucio, T. S., Mendoza-Ibáñez, O. I., Liu, W., Mostafapour, S., Li, Z., Providência, L., de Souza, G. S., Mohr, P., Dobrolińska, M. M., van Leer, B., et al. (2024) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2024.10.009 · Cited by 9 · see citing works
Lehtonen, E., Teuho, J., Koskinen, J., Tadi, M. J., Klén, R., Siekkinen, R., Gambin, J. R., Vasankari, T., & Saraste, A. (2021) · Sensors · doi:10.3390/s21123983 · Cited by 8 · see citing works
Value of PET ECG gating in a cross-validation study of cardiac function assessment by PET/MR imaging
Asiares, A. V., Vitadello, T., Bogdanović, B., Solari, E. L., McIntosh, L., Schachoff, S., Ibrahim, T., & Nekolla, S. G. (2022) · Journal of Nuclear Cardiology · doi:10.1007/s12350-022-03105-2 · Cited by 8 · see citing works
Motion in nuclear cardiology imaging: types, artifacts, detection and correction techniques
Mohammadi, I., Castro, I. F., Rahmim, A., & Veloso, J. F. C. A. (2021) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ac3dc7 · Cited by 7 · see citing works
Gupta, A., Park, J. Y., Choi, H., Choi, T. H., Chung, Y., Kim, D., & Lee, Y. (2024) · Molecular Pharmaceutics · doi:10.1021/acs.molpharmaceut.4c00412 · Cited by 6 · see citing works
Lassen, M. L., Tzolos, E., Pan, T., Kwieciński, J., Cadet, S. J., Dey, D., Berman, D. S., & Slomka, P. J. (2022) · Medical Physics · doi:10.1002/mp.15834 · Cited by 6 · see citing works
Self-citationUtility of cardiac imaging in patients with ventricular tachycardia
Hawson, J. J., Joshi, S. B., Al‐Kaisey, A., Das, S. K., Anderson, R. D., Morton, J. B., Kumar, S., Kistler, P. M., Kalman, J. M., & Lee, G. (2023) · Indian Pacing and Electrophysiology Journal · doi:10.1016/j.ipej.2023.03.003 · Cited by 6 · see citing works
Veneziano, F. A., Gioia, F. A., & Gentile, F. (2025) · Journal of Cardiovascular Development and Disease · doi:10.3390/jcdd12090338 · Cited by 5 · see citing works
Hashimoto, H., Kuronuma, K., Hyun, M. C., Han, D., Builoff, V. M., Cadet, S., Dey, D., Berman, D. S., Kwieciński, J., & Slomka, P. J. (2024) · Journal of Nuclear Cardiology · doi:10.1016/j.nuclcard.2024.102045 · Cited by 5 · see citing works
Self-citationEstimation of optimal number of gates in dual gated 18F-FDG cardiac PET
Klén, R., Teuho, J., Noponen, T., Thielemans, K., Hoppela, E., Lehtonen, E., Sipilä, H. T., Teräs, M., & Knuuti, J. M. (2020) · Scientific Reports · doi:10.1038/s41598-020-75613-5 · Cited by 5 · see citing works
Notohamiprodjo, S., Scheidhauer, K., Eiber, M. J., Yakushev, I. Y., Asiares, A. V., Kraus-Deuringer, J., Krebs-Fleischmann, H. W. A., Kleiner, S., Eggerstedt, R., Eglseder, B., et al. (2024) · Journal of Nuclear Cardiology · doi:10.1016/j.nuclcard.2024.102010 · Cited by 5 · see citing works
Fanta, O., Patil, S., Werner, T. J., Torigian, D. A., & Alavi, A. A. (2025) · Current Cardiology Reports · doi:10.1007/s11886-025-02240-9 · Cited by 4 · see citing works
Błach, A., & Kwieciński, J. (2023) · Diagnostics · doi:10.3390/diagnostics13101791 · Cited by 4 · see citing works
Deep generative denoising networks enhance quality and accuracy of gated cardiac PET data
Jafaritadi, M., Teuho, J., Lehtonen, E., Klén, R., Saraste, A., & Levin, C. S. (2024) · Annals of Nuclear Medicine · doi:10.1007/s12149-024-01945-1 · Cited by 4 · see citing works
Advanced Cardiac Imaging in the Assessment of Aortic Stenosis
Grodecki, K., Warniełło, M., Śpiewak, M., & Kwieciński, J. (2023) · Journal of Cardiovascular Development and Disease · doi:10.3390/jcdd10050216 · Cited by 3 · see citing works
DeTransUnet: attenuation correction of gated cardiac images without structural information
Wang, B., Lu, L., & Liu, H. (2022) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ac840e · Cited by 3 · see citing works
Novel PET Applications and Radiotracers for Imaging Cardiovascular Pathophysiology
Kwieciński, J. (2023) · Cardiology Clinics · doi:10.1016/j.ccl.2023.01.002 · Cited by 2 · see citing works
Guo, L., Liu, C., & Soultanidis, G. M. (2025) · PET Clinics · doi:10.1016/j.cpet.2025.07.004 · Cited by 2 · see citing works
Gözlügöl, N., Zangeneh, F. A., Sari, H., Bosbach, W. A., Zeimpekis, K., Rominger, A., Gräni, C., & Caobelli, F. (2026) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-026-07892-5 · Cited by 2 · see citing works
Luo, Y., & Wang, L. V. (2025) · Current Treatment Options in Cardiovascular Medicine · doi:10.1007/s11936-025-01092-4 · Cited by 2 · see citing works
Novel PET Applications and Radiotracers for Imaging Cardiovascular Pathophysiology
Kwieciński, J. (2025) · Heart Failure Clinics · doi:10.1016/j.hfc.2025.01.017 · Cited by 2 · see citing works
Device-Less Data-Driven Cardiac and Respiratory Gating Using LAFOV PET Histo Images
Overbeck, N., Andersen, T. L., Rodell, A. B., Cabello, J., Birge, N., Schleyer, P., Conti, M., Korsholm, K., Fischer, B. M., Andersen, F. L., et al. (2024) · Diagnostics · doi:10.3390/diagnostics14182055 · Cited by 2 · see citing works
Coronary artery disease imaging with total-body PET
Clark, L. E., Balmforth, C., Gil, M., Kwieciński, J., Slomka, P. J., Newby, D. E., Dweck, M. R., & Williams, M. C. (2026) · British Journal of Radiology · doi:10.1093/bjr/tqag093 · Cited by 1 · see citing works
Self-citationA Cardiac-Gating Method for Human Cerebral Hemodynamic Signal Extraction
Chen, L., Wang, M., Li, M., Yang, F., & Xu, S. (2025) · IEEE MTT-S International Conference on Numerical Electromagnetic and Multiphysics Modeling and Optimization · doi:10.1109/nemo62710.2025.11215541
Deng, J., Sun, H., Wang, H., Hong, X., Xu, W., Wang, F., Ma, J., Lian, C., & Lu, L. (2026) · EJNMMI Physics · doi:10.1186/s40658-026-00839-7
Bajaj, R., Wan, M. Y. S., Thielemans, K., Bakır, E. O., Maung, S., Sivananthan, A., Crake, T., Mathur, A., Torii, R., Baumbach, A., et al. (2026) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2026.1767077
Kim, H., Tang, R., Shrestha, U. M., Chae, H., Levi, J., Flavell, R. R., & Seo, Y. (2026) · Medical Physics · doi:10.1002/mp.70606
Investigation of CardioFreeze Motion Correction for Cardiac PET/CT
Wang, A. R., Guo, L., Fazzone‐Chettiar, R., Weyman, C. A., Hayden, C., Spottiswoode, B. S., Thorn, S., Liu, Y.-H., Liu, C., & Miller, E. J. (2026) · Journal of Nuclear Cardiology · doi:10.1016/j.nuclcard.2026.103000
- Authors
- Lassen, M. L., Kwiecinski, J., & Slomka, P. J.
- Affiliations
- Cedars-Sinai Medical Center (US)
- Corresponding author
- Slomka, Piotr J.
- Venue
- PET Clinics
- Publisher
- Elsevier BV
- Published
- 2019-04
- Volume
- 14
- Issue
- 2
- Pages
- 271-279
- ISSN
- 1556-8598 (ISSN-L)
- PMC
- PMC6793993
- OpenAlex
- W2913989203
- Cited by
- 38
- References
- 87
- Citations by year
- 2026: 6 · 2025: 6 · 2024: 8 · 2023: 6 · 2022: 6 · 2021: 3 · 2020: 1 · 2019: 2
- FWCI
- 0.45
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- PET Clinics (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Heart, Lung, and Blood Institute (R01HL135557)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Keywords
- cardiac PET, gating, respiratory motion, cardiac motion, motion correction, left ventricular systolic function, coronary artery disease, cardiomyopathy
- MeSH terms
- Electrocardiography, Humans, Cardiomyopathies, Sensitivity and Specificity, Magnetic Resonance Angiography, Positron-Emission Tomography, Myocardial Perfusion Imaging, Cardiac-Gated Imaging Techniques, Respiratory-Gated Imaging Techniques, Multimodal Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review