Investigating the state-of-the-art in whole-body MR-based attenuation correction: an intra-individual, inter-system, inventory study on three clinical PET/MR systems
Open accessAbstract
OBJECTIVE: We assess inter- and intra-subject variability of magnetic resonance (MR)-based attenuation maps (MRμMaps) of human subjects for state-of-the-art positron emission tomography (PET)/MR imaging systems. MATERIALS AND METHODS: Four healthy male subjects underwent repeated MR imaging with a Siemens Biograph mMR, Philips Ingenuity TF and GE SIGNA PET/MR system using product-specific MR sequences and image processing algorithms for generating MRμMaps. Total lung volumes and mean attenuation values in nine thoracic reference regions were calculated. Linear regression was used for comparing lung volumes on MRμMaps. Intra- and inter-system variability was investigated using a mixed effects model. RESULTS: Intra-system variability was seen for the lung volume of some subjects, (p = 0.29). Mean attenuation values across subjects were significantly different (p < 0.001) due to different segmentations of the trachea. Differences in the attenuation values caused noticeable intra-individual and inter-system differences that translated into a subsequent bias of the corrected PET activity values, as verified by independent simulations. CONCLUSION: Significant differences of MRμMaps generated for the same subjects but different PET/MR systems resulted in differences in attenuation correction factors, particularly in the thorax. These differences currently limit the quantitative use of PET/MR in multi-center imaging studies.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Several sources of error exist in MR-AC, including extracorporeal components, incorrect attenuation correction factors for bone and lung tissue, incorrect segmentation, and MR image truncation.
“However, several sources of error exist, including extracorporeal components such as headphones, dense hair or MR coils [59], incorrect attenuation correction factors for bone and lung tissue [60], incorrect segmentation of such compartments, and MR image truncation [61].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [61]
Inter-vendor differences in MR-AC maps can lead to SUV differences in phantom spheres of 10-20%.
“In such phantoms, inter-vendor differences in MR-AC maps [61] can lead to SUV differences in phantom spheres of 10–20% [58].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [61]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0
Boellaard, R., O’Doherty, M. J., Weber, W., Mottaghy, F. M., Lonsdale, M. N., Stroobants, S. G., Oyen, W. J. G., Kotzerke, J., Hoekstra, O. S., Pruim, J., et al. (2009) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-009-1297-4 · Cited by 1,329 · see citing works
Self-citationPerformance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner
Delso, G., Fürst, S., Jakoby, B., Ladebeck, R., Ganter, C., Nekolla, S. G., Schwaiger, M., & Ziegler, S. I. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.092726 · Cited by 870 · see citing works
Self-citationMartínez-Möller, A., Souvatzoglou, M., Delso, G., Bundschuh, R. A., Chefd’hotel, C., Ziegler, S. I., Navab, N., Schwaiger, M., & Nekolla, S. G. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.054726 · Cited by 712 · see citing works
Self-citationHofmann, M., Steinke, F., Scheel, V., Charpiat, G., Farquhar, J. D. R., Aschoff, P., Brady, M., Schölkopf, B., & Pichler, B. J. (2008) · Journal of Nuclear Medicine · doi:10.2967/jnumed.107.049353 · Cited by 495 · see citing works
Design and performance evaluation of a whole-body Ingenuity TF PET–MRI system
Zaidi, H., Ojha, N., Morich, M. A., Griesmer, J. J., Hu, Z., Maniawski, P. J., Ratib, O. M., Izquierdo‐Garcia, D., Fayad, Z. A., & Shao, L. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/10/013 · Cited by 400 · see citing works
CT substitute derived from MRI sequences with ultrashort echo time
Johansson, A., Karlsson, M. C. F., & Nyholm, T. (2011) · Medical Physics · doi:10.1118/1.3578928 · Cited by 323 · see citing works
Hofmann, M., Bezrukov, I., Mantlik, F., Aschoff, P., Steinke, F., Beyer, T., Pichler, B. J., & Schölkopf, B. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.110.078949 · Cited by 287 · see citing works
Self-citationSamarin, A., Burger, C., Wollenweber, S. D., Crook, D. W., Burger, I. A., Schmid, D. T., von Schulthess, G. K., & Kuhn, F. P. (2012) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-012-2113-0 · Cited by 251 · see citing works
Paulus, D. H., Quick, H. H., Geppert, C., Fenchel, M., Zhan, Y., Hermosillo, G., Faul, D. D., Boada, F. E., Friedman, K. P., & Koesters, T. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.156000 · Cited by 175 · see citing works
Self-citationSimultaneous Reconstruction of Activity and Attenuation for PET/MR
Salomon, A., Goedicke, A., Schweizer, B., Aach, T., & Schulz, V. (2010) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2010.2095464 · Cited by 174 · see citing works
Why is TOF PET reconstruction a more robust method in the presence of inconsistent data?
Conti, M. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/1/010 · Cited by 164 · see citing works
MR-Based PET Attenuation Correction for PET/MR Imaging
Bezrukov, I., Mantlik, F., Schmidt, H., Schölkopf, B., & Pichler, B. J. (2012) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2012.08.002 · Cited by 154 · see citing works
Completion of a Truncated Attenuation Image From the Attenuated PET Emission Data
Nuyts, J. L., Bal, G., Kehren, F., Fenchel, M., Michel, C. J., & Watson, C. C. (2012) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2012.2220376 · Cited by 140 · see citing works
Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI
Keller, S. H., Holm, S., Hansen, A. E., Sattler, B., Andersen, F. L., Klausen, T. L., Højgaard, L., Kjær, A., & Beyer, T. (2012) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-012-0345-4 · Cited by 125 · see citing works
Self-citationAbdominal fat assessed by computed tomography: Sex difference in distribution
Dixon, A. K. (1983) · Clinical Radiology · doi:10.1016/s0009-9260(83)80303-1 · Cited by 124 · see citing works
Routine quality control recommendations for nuclear medicine instrumentation
Sokole, E. B., With contribution from the EANM Working Group on Nuclear Medicine Instrumentation Quality Control:, Płachcińska, A., Britten, A. J., Georgosopoulou, M. L., Tindale, W., & Klett, R. (2010) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-009-1347-y · Cited by 124 · see citing works
Navalpakkam, B. K., Braun, H., Kuwert, T., & Quick, H. H. (2013) · Investigative Radiology · doi:10.1097/rli.0b013e318283292f · Cited by 123 · see citing works
Self-citationSimultaneous PET/MR imaging: MR-based attenuation correction of local radiofrequency surface coils
Paulus, D. H., Braun, H., Aklan, B., & Quick, H. H. (2012) · Medical Physics · doi:10.1118/1.4729716 · Cited by 114 · see citing works
Self-citationAccurate PET/MR Quantification Using Time of Flight MLAA Image Reconstruction
Boellaard, R., Hofman, M. B. M., Hoekstra, O. S., & Lammertsma, A. A. (2014) · Molecular Imaging and Biology · doi:10.1007/s11307-013-0716-x · Cited by 87 · see citing works
Self-citationTellmann, L., Quick, H. H., Bockisch, A., Herzog, H. R., & Beyer, T. (2011) · Medical Physics · doi:10.1118/1.3582699 · Cited by 82 · see citing works
Self-citationMR‐based field‐of‐view extension in MR/PET: B0 homogenization using gradient enhancement (HUGE)
Blumhagen, J. O., Ladebeck, R., Fenchel, M., & Scheffler, K. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24555 · Cited by 74 · see citing works
Integrated PET/MR imaging: Automatic attenuation correction of flexible RF coils
Kartmann, R., Paulus, D. H., Braun, H., Aklan, B., Ziegler, S., Navalpakkam, B. K., Lentschig, M. G., & Quick, H. H. (2013) · Medical Physics · doi:10.1118/1.4812685 · Cited by 71 · see citing works
Self-citationAznar, M. C., Sersar, R., Saabye, J., Ladefoged, C. N., Andersen, F. L., Rasmussen, J. H., Löfgren, J., & Beyer, T. (2014) · European Journal of Radiology · doi:10.1016/j.ejrad.2014.03.022 · Cited by 69 · see citing works
Self-citationSchramm, G., Maus, J., Hofheinz, F., Petr, J., Beuthien‐Baumann, B., Platzek, I., Steinbach, J., Kotzerke, J., & van den Hoff, J. (2012) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-012-0328-5 · Cited by 66 · see citing works
Impact of incorrect tissue classification in Dixon-based MR-AC: fat-water tissue inversion
Ladefoged, C. N., Hansen, A. E., Keller, S. H., Holm, S., Law, I. G., Beyer, T., Højgaard, L., Kjær, A., & Andersen, F. L. (2014) · EJNMMI Physics · doi:10.1186/s40658-014-0101-0 · Cited by 51 · see citing works
Self-citationSystematic Evaluation of Phantom Fluids for Simultaneous PET/MR Hybrid Imaging
Ziegler, S., Braun, H., Ritt, P., Hocke, C., Kuwert, T., & Quick, H. H. (2013) · Journal of Nuclear Medicine · doi:10.2967/jnumed.112.116376 · Cited by 50 · see citing works
Self-citationBlumhagen, J. O., Braun, H., Ladebeck, R., Fenchel, M., Faul, D. D., Scheffler, K., & Quick, H. H. (2014) · Medical Physics · doi:10.1118/1.4861097 · Cited by 48 · see citing works
Self-citationBailey, D. L., Barthel, H., Beyer, T., Boellaard, R., Gückel, B., Hellwig, D., Herzog, H. R., Pichler, B. J., Quick, H. H., Sabri, O., et al. (2013) · Molecular Imaging and Biology · doi:10.1007/s11307-013-0623-1 · Cited by 41 · see citing works
Self-citationInfluence and Compensation of Truncation Artifacts in MR-Based Attenuation Correction in PET/MR
Schramm, G., Langner, J., Hofheinz, F., Petr, J., Lougovski, A., Beuthien‐Baumann, B., Platzek, I., & van den Hoff, J. (2013) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2013.2272660 · Cited by 39 · see citing works
Bailey, D. L., Barthel, H., Beuthien‐Baumann, B., Beyer, T., Bisdas, S., Boellaard, R., Czernin, J. G., Drzezga, A. E., Ernemann, U., Franzius, C., et al. (2014) · Molecular Imaging and Biology · doi:10.1007/s11307-014-0725-4 · Cited by 39 · see citing works
Self-citationWollenweber, S. D., Delso, G., Deller, T. W., Goldhaber, D. M., Hüllner, M. W., & Veit‐Haibach, P. (2013) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-013-0388-1 · Cited by 30 · see citing works
Self-citationCompletion of a truncated attenuation image from the attenuated PET emission data
Nuyts, J. L., Michel, C. J., Fenchel, M., Bal, G., & Watson, C. (2010) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2010.5874154 · Cited by 18 · see citing works
Magnetic Resonace–Based Attenuation Correction for Micro–Single-Photon Emission Computed Tomography
Keereman, V., Fierens, Y., Vanhove, C., Lahoutte, T., & Vandenberghe, S. (2012) · Molecular Imaging · doi:10.2310/7290.2011.00036 · Cited by 9 · see citing works
Flagged first, then by citations
1 of the 34 references couldn’t be matched to a known paper.
Research works printed but not linked: 1
- Deller T, Delso G, Grant A, et al (2014) PET NEMA Perfor- mance Measurements for a SiPM-Based Time-of-Flight PET/ MR System" IEEE Medical Imaging Conference, Seattle #1618
Cited by
Sorted by citations
Deep Learning MR Imaging–based Attenuation Correction for PET/MR Imaging
Liu, F., Jang, H., Kijowski, R., Bradshaw, T., & McMillan, A. B. (2018) · Radiology · doi:10.1148/radiol.2017170700 · Cited by 412 · see citing works
Aide, N., Lasnon, C., Veit-Haibach, P., Sera, T., Sattler, B., & Boellaard, R. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3740-2 · Cited by 344 · see citing works
Self-citationDixon-VIBE Deep Learning (DIVIDE) Pseudo-CT Synthesis for Pelvis PET/MR Attenuation Correction
Torrado-Carvajal, Á., Vera-Olmos, J., Izquierdo‐Garcia, D., Catalano, O. A., Morales, M. A., Margolin, J., Soricelli, A., Salvatore, M., Malpica, N., & Catana, C. (2018) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.209288 · Cited by 132 · see citing works
Mannheim, J. G., Schmid, A. M., Schwenck, J., Katiyar, P., Herfert, K., Pichler, B. J., & Disselhorst, J. A. (2018) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2018.02.011 · Cited by 70 · see citing works
International EANM-SNMMI-ISMRM consensus recommendation for PET/MRI in oncology
Veit‐Haibach, P., Ahlström, H. K., Boellaard, R., Bolton, R. C. D., Hesse, S., Hope, T. A., Huellner, M. W., Iagaru, A. H., Johnson, G. B., Kjær, A., et al. (2023) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-023-06406-x · Cited by 51 · see citing works
Self-citationInfluences on PET Quantification and Interpretation
Rogasch, J. M. M., Hofheinz, F., van Heek, L., Voltin, C.-A., Boellaard, R., & Kobe, C. (2022) · Diagnostics · doi:10.3390/diagnostics12020451 · Cited by 50 · see citing works
“However, several sources of error exist, including extracorporeal components such as headphones, dense hair or MR coils [59], incorrect attenuation correction factors for bone and lung tissue [60], incorrect segmentation of such compartments, and MR image truncation [61].”
BackgroundSelf-citationPET/MRI in Infection and Inflammation
Sollini, M., Berchiolli, R. N., Kirienko, M., Rossi, A., Glaudemans, A. W. J. M., Slart, R. H. J. A., & Erba, P. A. (2018) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2018.02.003 · Cited by 50 · see citing works
Technical and instrumentational foundations of PET/MRI
Rausch, I., Quick, H. H., Cal-González, J., Sattler, B., Boellaard, R., & Beyer, T. (2017) · European Journal of Radiology · doi:10.1016/j.ejrad.2017.04.004 · Cited by 47 · see citing works
Self-citationAssessment of attenuation correction for myocardial PET imaging using combined PET/MRI
Lassen, M. L., Rasul, S., Beitzke, D., Stelzmueller, M.-E., Cal-González, J., Hacker, M., & Beyer, T. (2017) · Journal of Nuclear Cardiology · doi:10.1007/s12350-017-1118-2 · Cited by 47 · see citing works
Self-citationAttenuation correction for human PET/MRI studies
Catana, C. (2020) · Physics in Medicine and Biology · doi:10.1088/1361-6560/abb0f8 · Cited by 47 · see citing works
Lindemann, M. E., Oehmigen, M., Blumhagen, J. O., Gratz, M., & Quick, H. H. (2017) · Medical Physics · doi:10.1002/mp.12449 · Cited by 41 · see citing works
Self-citationOehmigen, M., Lindemann, M. E., Gratz, M., Kirchner, J., Ruhlmann, V., Umutlu, L., Blumhagen, J. O., Fenchel, M., & Quick, H. H. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3864-4 · Cited by 39 · see citing works
Self-citationGating Approaches in Cardiac PET Imaging
Lassen, M. L., Kwiecinski, J., & Slomka, P. J. (2019) · PET Clinics · doi:10.1016/j.cpet.2018.12.008 · Cited by 38 · see citing works
Self-citationA review of PET attenuation correction methods for PET-MR
Krokos, G., Mackewn, J. E., Dunn, J., & Marsden, P. K. (2023) · EJNMMI Physics · doi:10.1186/s40658-023-00569-0 · Cited by 36 · see citing works
MLAA-based attenuation correction of flexible hardware components in hybrid PET/MR imaging
Heußer, T., Rank, C. M., Berker, Y., Freitag, M. T., & Kachelrieß, M. (2017) · EJNMMI Physics · doi:10.1186/s40658-017-0177-4 · Cited by 31 · see citing works
Aortic valve stenosis—multimodality assessment with PET/CT and PET/MRI
Tzolos, E., Andrews, J. P., & Dweck, M. R. (2019) · British Journal of Radiology · doi:10.1259/bjr.20190688 · Cited by 31 · see citing works
Oehmigen, M., Lindemann, M. E., Lanz, T. E., Kinner, S., & Quick, H. H. (2016) · Medical Physics · doi:10.1118/1.4959546 · Cited by 29 · see citing works
Self-citationWhole-Body [18F]-FDG-PET/MRI for Oncology: A Consensus Recommendation
Umutlu, L., Beyer, T., Grueneisen, J. S., Rischpler, C., Quick, H. H., Veit‐Haibach, P., Eiber, M., Purz, S., Antoch, G., Gatidis, S., et al. (2019) · Nuklearmedizin - NuclearMedicine · doi:10.1055/a-0830-4453 · Cited by 29 · see citing works
Self-citationAndrews, J. P. M., Macnaught, G., Moss, A. J., Doris, M. K., Pawade, T., Adamson, P. D., van Beek, E. J. R., Lucatelli, C., Lassen, M. L., Robson, P. M., et al. (2019) · Journal of Nuclear Cardiology · doi:10.1007/s12350-019-01962-y · Cited by 29 · see citing works
Self-citationPET/MRI attenuation estimation in the lung: A review of past, present, and potential techniques
Lillington, J., Brusaferri, L., Kläser, K., Shmueli, K., Neji, R., Hutton, B. F., Fraioli, F., Arridge, S., Cardoso, M. J., Sebastien Ourselin, A., et al. (2019) · Medical Physics · doi:10.1002/mp.13943 · Cited by 28 · see citing works
EFOMP’s protocol quality controls in PET/CT and PET/MR
Matheoud, R., Boellaard, R., Pike, L. C., Ptáček, J., Reynés‐Llompart, G., Soret, M., Vandenberghe, S., Zorz, A., Julyan, P. J., Rausch, I., et al. (2022) · Physica Medica · doi:10.1016/j.ejmp.2022.11.010 · Cited by 22 · see citing works
Self-citationWhole-Body [18F]-FDG-PET/MRI for Oncology: A Consensus Recommendation
Umutlu, L., Beyer, T., Grueneisen, J. S., Rischpler, C., Quick, H. H., Veit‐Haibach, P., Eiber, M., Purz, S., Antoch, G., Gatidis, S., et al. (2019) · RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren · doi:10.1055/a-0828-8654 · Cited by 22 · see citing works
Self-citationRischpler, C., Nekolla, S. G., Heusch, G., Umutlu, L., Rassaf, T., Heusch, P., Herrmann, K., & Nensa, F. (2019) · European Journal of Hybrid Imaging · doi:10.1186/s41824-018-0050-2 · Cited by 20 · see citing works
Wang, C., Wu, P.-H., Lin, G., Huang, Y.-L., Lin, Y., Chang, Y.-P. E., & Weng, J. (2022) · Journal of Personalized Medicine · doi:10.3390/jpm12030361 · Cited by 19 · see citing works
PET/MR Imaging in Cardiovascular Imaging
Rischpler, C., & Woodard, P. K. (2019) · PET Clinics · doi:10.1016/j.cpet.2018.12.005 · Cited by 18 · see citing works
PET attenuation correction for flexible MRI surface coils in hybrid PET/MRI using a 3D depth camera
Frohwein, L. J., Heß, M., Schlicher, D., Bolwin, K., Büther, F., Jiang, X., & Schäfers, K. P. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa9e2f · Cited by 18 · see citing works
Olin, A. B., Ladefoged, C. N., Langer, N. H., Keller, S. H., Löfgren, J., Hansen, A. E., Kjær, A., Langer, S. W., Fischer, B. M., & Andersen, F. L. (2017) · Journal of Nuclear Medicine · doi:10.2967/jnumed.117.198853 · Cited by 17 · see citing works
Recent Advances in Positron Emission Tomography/Magnetic Resonance Imaging Technology
Arabi, H., & Zaidi, H. (2023) · Magnetic Resonance Imaging Clinics of North America · doi:10.1016/j.mric.2023.06.002 · Cited by 16 · see citing works
Impact of improved attenuation correction on 18F-FDG PET/MR hybrid imaging of the heart
Lindemann, M. E., Nensa, F., & Quick, H. H. (2019) · PLoS ONE · doi:10.1371/journal.pone.0214095 · Cited by 15 · see citing works
Self-citationCatana, C., Quick, H. H., & Zaidi, H. (2018) · Medical Physics · doi:10.1002/mp.12963 · Cited by 15 · see citing works
Self-citationRausch, I., Valladares, A., Sundar, L. K. S., Beyer, T., Hacker, M., Meyerspeer, M., & Unger, E. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00364-9 · Cited by 11 · see citing works
Self-citationGrafe, H., Lindemann, M. E., Ruhlmann, V., Oehmigen, M., Hirmas, N., Umutlu, L., Herrmann, K., & Quick, H. H. (2020) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-020-04738-6 · Cited by 10 · see citing works
Self-citationSousa, J. M., Appel, L., Mérida, I., Heckemann, R. A., Costes, N., Engström, M., Papadimitriou, S., Nyholm, D., Ahlström, H. K., Hammers, A., et al. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-00347-2 · Cited by 10 · see citing works
Practical Considerations for Clinical PET/MR Imaging
Galgano, S. J., Viets, Z., Fowler, K. J., Gore, L., Thomas, J. V., McNamara, M. M., & McConathy, J. (2017) · PET Clinics · doi:10.1016/j.cpet.2017.09.002 · Cited by 10 · see citing works
Magnetic Resonance Imaging: Historical Overview, Technical Developments, and Clinical Applications
Jahng, G., Park, S., Ryu, C., & Cho, Z. (2020) · Progress in Medical Physics · doi:10.14316/pmp.2020.31.3.35 · Cited by 9 · see citing works
Oehmigen, M., Lindemann, M. E., Gratz, M., Neji, R., Hammers, A., Sauer, M., Lanz, T. E., & Quick, H. H. (2018) · Medical Physics · doi:10.1002/mp.13171 · Cited by 9 · see citing works
Self-citationKuttner, S., Lassen, M. L., Øen, S. K., Sundset, R., Beyer, T., & Eikenes, L. (2019) · Acta Radiologica · doi:10.1177/0284185119848118 · Cited by 8 · see citing works
Self-citationPET/CT and PET/MR Imaging of the Post-treatment Head and Neck
Guzmán, G., & Ivanidze, J. (2021) · Neuroimaging Clinics of North America · doi:10.1016/j.nic.2021.09.003 · Cited by 7 · see citing works
Practical Considerations for Clinical PET/MR Imaging
Galgano, S. J., Viets, Z., Fowler, K. J., Gore, L., Thomas, J. V., McNamara, M. M., & McConathy, J. (2017) · Magnetic Resonance Imaging Clinics of North America · doi:10.1016/j.mric.2016.12.006 · Cited by 7 · see citing works
Bruckmann, N. M., Lindemann, M. E., Grueneisen, J. S., Grafe, H., Li, Y., Sawicki, L. M., Rischpler, C., Herrmann, K., Umutlu, L., Quick, H. H., et al. (2021) · European Journal of Radiology · doi:10.1016/j.ejrad.2021.109691 · Cited by 7 · see citing works
Self-citationLindemann, M. E., Gratz, M., Grafe, H., Jannusch, K., Umutlu, L., & Quick, H. H. (2023) · Medical Physics · doi:10.1002/mp.16863 · Cited by 6 · see citing works
Self-citationAttenuation correction and truncation completion for breast PET/MR imaging using deep learning
Li, X., Johnson, J. M., Strigel, R. M., Bancroft, L. C. H., Hurley, S. A., Estakhraji, S. I. Z., Kumar, M., Fowler, A. M., & McMillan, A. B. (2024) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ad2126 · Cited by 6 · see citing works
PET/MRI: Attenuation Correction
Delso, G., & Nuyts, J. L. (2018) · Springer eBooks · doi:10.1007/978-3-319-68517-5_4 · Cited by 6 · see citing works
Self-citationImportance of attenuation correction in PET/MR image quantification: Methods and applications
Torrado-Carvajal, Á. (2020) · Revista Española de Medicina Nuclear e Imagen Molecular (English Edition) · doi:10.1016/j.remnie.2020.03.002 · Cited by 6 · see citing works
Allen, T. J., Bancroft, L. C. H., Kumar, M., Bradshaw, T., Strigel, R. M., McMillan, A. B., & Fowler, A. M. (2022) · Medical Physics · doi:10.1002/mp.15781 · Cited by 5 · see citing works
Essentials of Quantitative Imaging with PET
Lammertsma, A. A. (2019) · Springer eBooks · doi:10.1007/978-3-319-95564-3_10 · Cited by 4 · see citing works
Lindemann, M. E., Gratz, M., Blumhagen, J. O., Jakoby, B., & Quick, H. H. (2022) · Medical Physics · doi:10.1002/mp.15446 · Cited by 4 · see citing works
Self-citationTorrado-Carvajal, Á. (2020) · Revista Española de Medicina Nuclear e Imagen Molecular · doi:10.1016/j.remn.2020.03.004 · Cited by 3 · see citing works
Grafe, H., Lindemann, M. E., Weber, M., Kirchner, J., Binse, I., Umutlu, L., Herrmann, K., & Quick, H. H. (2022) · Cancers · doi:10.3390/cancers14133040 · Cited by 3 · see citing works
Self-citationLindemann, M. E., Oehmigen, M., Lanz, T. E., Grafe, H., Bruckmann, N. M., Umutlu, L., & Quick, H. H. (2022) · Medical Physics · doi:10.1002/mp.15535 · Cited by 3 · see citing works
Self-citationPET/CT and PET/MR Tomographs: Image Acquisition and Processing
Belcari, N., Boellaard, R., & Morrocchi, M. (2019) · Springer eBooks · doi:10.1007/978-3-319-95564-3_9 · Cited by 3 · see citing works
Self-citationQuick, H. H. (2017) · Zeitschrift für Medizinische Physik · doi:10.1016/j.zemedi.2017.09.002 · Cited by 3 · see citing works
Self-citationPET/MRI: technical and methodological aspects
Torrado-Carvajal, Á., & Catana, C. (2022) · Elsevier eBooks · doi:10.1016/b978-0-323-88537-9.00003-9 · Cited by 2 · see citing works
Systems, Physics, and Instrumentation of PET/MRI for Cardiovascular Studies
Quick, H. H. (2017) · Current Cardiovascular Imaging Reports · doi:10.1007/s12410-017-9414-4 · Cited by 2 · see citing works
Self-citationThe Future of PET-MRI Beyond “PET Plus MRI”
Saboury, B., Morris, M. A., Borja, A. J., & Alavi, A. A. (2020) · Advances in Clinical Radiology · doi:10.1016/j.yacr.2020.04.008 · Cited by 2 · see citing works
Future Trends of PET/MR and Utility of AI in Multi-Modal Imaging
Hung, S., Liu, M., Yap, P., Shen, D. G., Lin, W., & Castillo, M. (2021) · Springer eBooks · doi:10.1007/978-3-030-82367-2_9 · Cited by 1 · see citing works
Peters, H. A., Scheuer, M., Weiß, D., Boschheidgen, M., Ivan, V. L., Dietzel, F., Mohrmann, S., Ruckhäberle, E., Herrmann, K., Quick, H. H., et al. (2026) · Diagnostics · doi:10.3390/diagnostics16111712 · Cited by 1 · see citing works
Self-citationCardiovascular PET/MRI: Technical Considerations and Outlook
Nekolla, S. G., van Marwick, S., Schachoff, S., Kunze, K., & Rischpler, C. (2017) · Current Cardiovascular Imaging Reports · doi:10.1007/s12410-017-9435-z · Cited by 1 · see citing works
Lassen, M. L., & Slomka, P. J. (2020) · Journal of Nuclear Cardiology · doi:10.1007/s12350-019-01994-4
Self-citationLassen, M. L., & Slomka, P. J. (2020) · Journal of Nuclear Cardiology · doi:10.1007/s12350-020-02057-9
Self-citationPET/MR: Perfusion and Viability
Kessler, L., & Rischpler, C. (2021) · Springer eBooks · doi:10.1007/978-3-030-83167-7_12
Quick, H. H. (2017) · Springer eBooks · doi:10.1007/978-3-319-69641-6_2
Self-citationRadiomics and artificial intelligence
Ghezzo, S., Bezzi, C., Neri, I., Mapelli, P., Presotto, L., Gajate, A. M. S., Bettinardi, V., Garibotto, V., COBELLI, F. D., Scifo, P., et al. (2022) · Elsevier eBooks · doi:10.1016/b978-0-323-88537-9.00002-7
Rausch, I., & Beyer, T. (2024) · Springer eBooks · doi:10.1007/978-3-662-67192-4_3
Self-citationMilošević, A., Chodyla, M.-K., Bruckmann, N. M., Lindemann, M. E., Grueneisen, J. S., Haubold, J., Fendler, W. P., Umutlu, L., Quick, H. H., & Schaarschmidt, B. M. (2024) · Clinical and Translational Imaging · doi:10.1007/s40336-024-00628-1
Self-citation
- Authors
- Beyer, T., Lassen, M. L., Boellaard, R., Delso, G., Yaqub, M., Sattler, B., & Quick, H. H.
- Affiliations
- Vienna General Hospital (AT)
- University Medical Center Groningen (NL)
- University Hospital Zurich (CH)
- Gemeente Amsterdam (NL)
- University Hospital Leipzig (DE)
- Friedrich-Alexander-Universität Erlangen-Nürnberg (DE)
- Corresponding author
- Beyer, Thomas
- ORCID
- Venue
- Magnetic Resonance Materials in Physics, Biology and Medicine
- Published
- 2016-01-06
- Volume
- 29
- Issue
- 1
- Pages
- 75-87
- ISSN
- 0968-5243 (ISSN-L)
- 1352-8661
- OpenAlex
- W2229540381
- Cited by
- 66
- References
- 34
- Citations by year
- 2026: 1 · 2024: 4 · 2023: 4 · 2022: 9 · 2021: 5 · 2020: 9 · 2019: 12 · 2018: 6 · 2017: 15 · 2016: 1
- FWCI
- 9.85
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- other-oa
- www.springer.com/tdm (tdm)
- Locations
- Magnetic Resonance Materials in Physics Biology and Medicine (publishedVersion, other-oa)
- PubMed (publishedVersion)
- Amsterdam UMC (VU Amsterdam) - Institutional Repository (publishedVersion)
- Pure Amsterdam UMC (submittedVersion, other-oa)
- University of Groningen research database (University of Groningen / Centre for Information Technology) (submittedVersion, other-oa)
- Data Archiving and Networked Services (DANS) (submittedVersion)
- Funders
- Vanderbilt University Medical Center
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
- Keywords
- MR-based attenuation correction, PET/MR, within-person variability, lung volume, attenuation map
- MeSH terms
- Adult, Algorithms, Computer Simulation, Humans, Image Processing, Computer-Assisted, Lung, Magnetic Resonance Imaging, Male, Middle Aged, Reproducibility of Results, Positron-Emission Tomography, Whole Body Imaging, Multimodal Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review