Lead integrity alert algorithm decreases inappropriate shocks in patients who have Sprint Fidelis pace-sense conductor fractures

Kallinen, Linda M., Hauser, Robert G., Tang, Chuen, Melby, Daniel P., Almquist, Adrian K., Katsiyiannis, William T., Gornick, Charles C.·2010·Heart Rhythm

Abstract

Background The Medtronic Sprint Fidelis high-voltage implantable cardioverter-defibrillator (ICD) lead is prone to fracture. The majority of fractures involve the pace-sense (P/S) conductors and may result in multiple inappropriate shocks. The Medtronic lead integrity alert (LIA) algorithm was designed to improve early detection of transient P/S conductor fractures and to decrease the incidence and number of inappropriate shocks.Objective The purpose of this prospective single-center study was to assess the effectiveness of the LIA algorithm for warning patients of an impending Sprint Fidelis P/S conductor fracture and for decreasing the incidence and number of inappropriate shocks.Methods The study population included all patients who had Sprint Fidelis leads and Medtronic ICD pulse generators that were implanted and followed at the Minneapolis Heart Institute. Patients were evaluated in the clinic every 3 to 4 months or by remote monitoring using the Medtronic CareLink system. When the LIA algorithm was released in August 2008, the RAMware was downloaded to the pulse generator of all patients with the Sprint Fidelis lead. Patients and family members received educational materials and were given a demonstration of the audible alerts.Results Between October 2004 and January 2010, 52 (11.3%) of 461 Sprint Fidelis leads failed in the study population. Inappropriate shocks were the first sign of lead failure in 18 (69%) of the 26 patients who did not have the LIA compared to 4 (17%) of 23 patients who had the LIA (P = .0004). Patients who experienced inappropriate shocks without the LIA received an average of 13.2 +/- 13.6 inappropriate shocks (range 2-54) versus 3.0 +/- 2.0 inappropriate shocks (range 2-6) in patients who had the LIA (P = .017). The audible alert was effective in 70% (16/23) and 35% (6/17) of patients with and without the LIA, respectively, whose alerts were programmed ON (P = .053). Overall, 8 (32%) of 25 patients whose audible alerts were triggered did not immediately hear or recognize the tone.Conclusion The LIA appears to be an effective method for detecting most Sprint Fidelis lead fractures and for decreasing the incidence and number of inappropriate shocks. However, a better method for alerting patients and caregivers is needed.

Integrity signals

Registry

Crossref · 10.1016/j.hrthm.2010.05.015

Cites

16 of 16 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Prospective although smaller studies have also demonstrated that lead integrity alerts decrease shocks.

    “Prospective although smaller studies have also demonstrated that lead integrity alerts decrease shocks.75”

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Kallinen, L. M., Hauser, R. G., Tang, C., Melby, D. P., Almquist, A. K., Katsiyiannis, W. T., & Gornick, C. C.
Affiliations
Minneapolis Heart Institute Foundation (US)
Corresponding author
Hauser, Robert G.
Venue
Heart Rhythm
Publisher
Elsevier BV
Published
2010-08
Volume
7
Issue
8
Pages
1048-1055
ISSN
1547-5271 (ISSN-L)
OpenAlex
W2011174551
Cited by
62
References
16
Citations by year
2025: 2 · 2024: 1 · 2023: 1 · 2022: 3 · 2021: 1 · 2020: 2 · 2018: 2 · 2017: 2 · 2016: 4 · 2015: 3 · 2014: 5 · 2013: 8 · 2012: 12
FWCI
7.68
Type
journal article
Language
en
Open access
closed
Locations
  • Heart Rhythm (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
  • Health Sciences › Medicine › Epidemiology › Infective Endocarditis Diagnosis and Management
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac electrophysiology and arrhythmias
Keywords
inappropriate shocks, implantable cardioverter-defibrillator, remote monitoring, patient education
MeSH terms
Adult, Aged, Algorithms, Cardiac Pacing, Artificial, Electrodes, Implanted, Equipment Failure, Female, Humans, Male, Middle Aged, Defibrillators, Implantable

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review