Guidelines for Managing Pacemaker and Implantable Defibrillator Advisories
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Since 1990, more than 60 device or lead advisories have been announced, affecting more than 1 million device patients.
“Since 1990, >60 device or lead advisories have been announced, affecting >1 million device patients.66”
From Managing with pacemakers and implantable cardioverter defibrillators (2013) · cites this paper as [66]
In the ICD Sports Safety Registry, freedom from lead malfunction was 97% at 5 years from implant and 90% at 10 years, similar to previously reported malfunction rates in unselected populations.
“Freedom from lead malfunction was 97% at 5 years (from implant) and 90% at 10 years, similar to previously-reported malfunction rates in unselected populations.66”
From Managing with pacemakers and implantable cardioverter defibrillators (2013) · cites this paper as [66]
Cites
The AVID Investigators (1997) · New England Journal of Medicine · doi:10.1056/nejm199711273372202 · Cited by 3,221 · see citing works
Mirowski, M., Reid, P. R., Mower, M. M., Watkins, L., Gott, V. L., Schauble, J. F., Langer, A. A., Heilman, M. S., Kolenik, S. A., Fischell, R. E., et al. (1980) · New England Journal of Medicine · doi:10.1056/nejm198008073030607 · Cited by 1,356 · see citing works
Wilkoff, B. L., Love, C. J., BYRD, C. L., Bongiorni, M. G., Carrillo, R. G., Crossley, G. H., Epstein, L. M., Friedman, R. A., Kennergren, C. E. H., Mitkowski, P. P., et al. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.05.020 · Cited by 1,024 · see citing works
Catastrophic outcomes of noncardiac surgery soon after coronary stenting
Kałuża, G. L., Joseph, J., Lee, J. R., Raizner, M. E., & Raizner, A. E. (2000) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(00)00521-0 · Cited by 720 · see citing works
Risk Factors Related to Infections of Implanted Pacemakers and Cardioverter-Defibrillators
Klug, D., Balde, M., Pavin, D., Hidden‐Lucet, F., Clémenty, J., Sadoul, N., Rey, J. L., Lande, G., Lazarus, A., Victor, J., et al. (2007) · Circulation · doi:10.1161/circulationaha.106.678664 · Cited by 697 · see citing works
Kleemann, T., Becker, T., Doenges, K., Vater, M., Senges, J., Schneider, S., Saggau, W., Weisse, U., & Seidl, K. (2007) · Circulation · doi:10.1161/circulationaha.106.663807 · Cited by 633 · see citing works
Lead Extraction in the Contemporary Setting: The LExICon Study
Wazni, O. M., Epstein, L. M., Carrillo, R. G., Love, C. J., Adler, S. W., Riggio, D. W., Karim, S. S., Bashir, J. G., Greenspon, A. J., Dimarco, J., et al. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.08.070 · Cited by 577 · see citing works
Wilkoff, B. L., Auricchio, A., Brugada, J., Cowie, M., Ellenbogen, K. A., Gillis, A. M., Hayes, D. L., Howlett, J. G., Kautzner, J., Love, C. J., et al. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.04.013 · Cited by 364 · see citing works
Pacemaker and ICD Generator Malfunctions
Maisel, W. H., Moynahan, M., Zuckerman, B. D., Gross, T. P., Tovar, O. H., Tillman, D.-B., & Schultz, D. B. (2006) · JAMA · doi:10.1001/jama.295.16.1901 · Cited by 344 · see citing works
Self-citationClinical Study of the Laser Sheath for Lead Extraction: The Total Experience in the United States
BYRD, C. L., Wilkoff, B. L., Love, C. J., Sellers, T. D., & Reiser, C. (2002) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2002.t01-1-00804.x · Cited by 308 · see citing works
Examining the psychosocial impact of implantable cardioverter defibrillators: A literature review
Camm, A. J., Sears, S. F., Todaro, J. F., Lewis, T. S., Sotile, W. M., & CONTI, J. B. (1999) · Clinical Cardiology · doi:10.1002/clc.4960220709 · Cited by 292 · see citing works
Gould, P. A. (2006) · JAMA · doi:10.1001/jama.295.16.1907 · Cited by 284 · see citing works
Five‐Years Experience with Intravascular Lead Extraction
SMITH, H. J., Fearnot, N. E., BYRD, C. L., Wilkoff, B. L., Love, C. J., Sellers, T. D., & DATABASE, F. T. U. S. L. E. (1994) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1994.tb03792.x · Cited by 278 · see citing works
Hauser, R. G., Katsiyiannis, W. T., Gornick, C. C., Almquist, A. K., & Kallinen, L. M. (2010) · Europace · doi:10.1093/europace/eup375 · Cited by 264 · see citing works
Complications Related to Permanent Pacemaker Therapy
KIVINIEMI, M. S., PIRNES, M. A., ERÄNEN, H. J. K., Kettunen, R., & Hartikainen, J. E. K. (1999) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1999.tb00534.x · Cited by 259 · see citing works
Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator lead
Hauser, R. G., Kallinen, L. M., Almquist, A. K., Gornick, C. C., & Katsiyiannis, W. T. (2007) · Heart Rhythm · doi:10.1016/j.hrthm.2007.03.041 · Cited by 246 · see citing works
Adverse Events With Transvenous Implantable Cardioverter-Defibrillators
Rosenqvist, M., Beyer, T., Block, M. S., den Dulk, K., Minten, J., & Lindemans, F. W. (1998) · Circulation · doi:10.1161/01.cir.98.7.663 · Cited by 231 · see citing works
Recalls and Safety Alerts Involving Pacemakers and Implantable Cardioverter-Defibrillator Generators
Maisel, W. H. (2001) · JAMA · doi:10.1001/jama.286.7.793 · Cited by 210 · see citing works
Self-citationJones, S. O., Eckart, R. E., Albert, C. M., & Epstein, L. M. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.01.009 · Cited by 208 · see citing works
A single-centre experience of over one thousand lead extractions
Kennergren, C. E. H., Bjurman, C., Wiklund, R., & Gäbel, J. (2009) · EP Europace · doi:10.1093/europace/eup054 · Cited by 206 · see citing works
Medical Device Regulation: An Introduction for the Practicing Physician
Maisel, W. H. (2004) · Annals of Internal Medicine · doi:10.7326/0003-4819-140-4-200402170-00012 · Cited by 194 · see citing works
Self-citationAl‐Khatib, S. M., Lucas, F. L., Jollis, J. G., Malenka, D. J., & Wennberg, D. E. (2005) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2005.04.063 · Cited by 193 · see citing works
Curtis, J. P. (2009) · JAMA · doi:10.1001/jama.2009.547 · Cited by 181 · see citing works
Carlson, M. D., Wilkoff, B. L., Maisel, W. H., Carlson, M. D., Ellenbogen, K. A., Saxon, L. A., Prystowsky, E. N., Alpert, J. S., Cain, M. E., Ching, E. A., et al. (2006) · Heart Rhythm · doi:10.1016/j.hrthm.2006.08.029 · Cited by 177 · see citing works
Self-citationPacemaker and ICD Generator Reliability
Maisel, W. H. (2006) · JAMA · doi:10.1001/jama.295.16.1929 · Cited by 170 · see citing works
Self-citationDetection and management of animplantable cardioverter defibrillator lead failure
Ellenbogen, K. A., Wood, M. A., Shepard, R. K., Clemo, H. F., Vaughn, T., Holloman, K., Dow, M., Leffler, J., Abeyratne, A., & Verness, D. (2002) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(02)02620-7 · Cited by 164 · see citing works
TOMPKINS, C. M., Cheng, A., Dalal, D., Brinker, J., Leng, C. T., Marine, J. E., Nazarian, S., SPRAGG, D. D., Sinha, S. K., Halperin, H. R., et al. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.12.056 · Cited by 158 · see citing works
Necessity for Surgical Revision of Defibrillator Leads Implanted Long-Term
Eckstein, J., Koller, M., Zabel, M., Kalusche, D., Schaer, B. A., Osswald, S., & Sticherling, C. (2008) · Circulation · doi:10.1161/circulationaha.107.740670 · Cited by 150 · see citing works
Kron, J., Herre, J. M., Renfroe, E. G., Rizo-Patrón, C., Raitt, M. H., Halperin, B. D., Gold, M. R., Goldner, B. G., Wathen, M. S., Wilkoff, B. L., et al. (2001) · American Heart Journal · doi:10.1067/mhj.2001.111261 · Cited by 149 · see citing works
Swerdlow, C. D., Gunderson, B. D., Ousdigian, K. T., Abeyratne, A., Stadler, R. W., Gillberg, J. M., Patel, A. S., & Ellenbogen, K. A. (2008) · Circulation · doi:10.1161/circulationaha.108.796136 · Cited by 141 · see citing works
Transvenous Defibrillation Leads: High Incidence of Failure During Long‐Term Follow‐Up
Dorwarth, U., Frey, B., Dugas, M., Matis, T., Fiek, M., Schmoeckel, M., REMP, T., DURCHLAUB, I., GERTH, A., Steinbeck, G., et al. (2003) · Journal of Cardiovascular Electrophysiology · doi:10.1046/j.1540-8167.2003.02305.x · Cited by 135 · see citing works
Risks of Spontaneous Injury and Extraction of an Active Fixation Pacemaker Lead
Kay, G. N., Brinker, J., Kawanishi, D. T., Love, C. J., Lloyd, M. A., Reeves, R. C., PIOGER, G., Fee, J., Overland, M., Ensign, L. G., et al. (1999) · Circulation · doi:10.1161/01.cir.100.23.2344 · Cited by 117 · see citing works
Recommendations from the Heart Rhythm Society Task Force on Lead Performance Policies and Guidelines
Maisel, W. H., Hauser, R. G., Hammill, S. C., Hauser, R. G., Ellenbogen, K. A., Epstein, A. E., Hayes, D. L., Alpert, J. S., Berger, R. D., Curtis, A. B., et al. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.04.024 · Cited by 114 · see citing works
Self-citationRisk of Failure of Transvenous Implantable Cardioverter-Defibrillator Leads
Borleffs, C. J. W., van Erven, L., van Bommel, R. J., van der Velde, E. T., van der Wall, E. E., Bax, J. J., Rosendaal, F. R., & Schalij, M. J. (2009) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.108.834093 · Cited by 107 · see citing works
Safety Issues Involving Medical Devices
Maisel, W. H. (2005) · JAMA · doi:10.1001/jama.294.8.955 · Cited by 104 · see citing works
Self-citationGold, M. R., Peters, R. W., Johnson, J. W., & Shorofsky, S. R. (1996) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(96)00314-2 · Cited by 102 · see citing works
Semper Fidelis — Consumer Protection for Patients with Implanted Medical Devices
Maisel, W. H. (2008) · New England Journal of Medicine · doi:10.1056/nejmp0800495 · Cited by 98 · see citing works
Self-citationImplantable Cardioverter-Defibrillator Lead Performance
Maisel, W. H., & Kramer, D. B. (2008) · Circulation · doi:10.1161/circulationaha.108.776807 · Cited by 96 · see citing works
Self-citationEffective Management of ICD Patient Psychosocial Issues and Patient Critical Events
Sears, S. F., Matchett, M., & CONTI, J. B. (2009) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2009.01526.x · Cited by 92 · see citing works
Kallinen, L. M., Hauser, R. G., Lee, K. W., Almquist, A. K., Katsiyiannis, W. T., Tang, C. Y., Melby, D. P., & Gornick, C. C. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.02.039 · Cited by 92 · see citing works
Hammill, S. C., Kremers, M. S., Kadish, A. H., Stevenson, L. W., Heidenreich, P. A., Lindsay, B. D., Mirro, M. J., Radford, M. J., McKay, C., Wang, Y., et al. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.07.015 · Cited by 91 · see citing works
Maytin, M., Love, C. J., Fischer, A., Carrillo, R. G., Garisto, J. D., Bongiorni, M. G., Segreti, L., John, R. M., Michaud, G. F., Albert, C. M., et al. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.03.058 · Cited by 89 · see citing works
Jamula, E., Douketis, J. D., & Schulman, S. (2008) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2008.03080.x · Cited by 86 · see citing works
Transvenous Implantable Cardioverter-Defibrillator Leads
Maisel, W. H. (2007) · Circulation · doi:10.1161/circulationaha.107.698597 · Cited by 78 · see citing works
Self-citationDo abandoned leads pose risk to implantable cardioverter-defibrillator patients?
Glikson, M., Suleiman, M. A., Luria, D. M., Martin, M. L., Hodge, D. O., Shen, W.-K., Bradley, D. J., Munger, T. M., Rea, R. F., Hayes, D. L., et al. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.10.012 · Cited by 75 · see citing works
Outcome of advisory implantable cardioverter-defibrillator replacement: One-year follow-up
Gould, P. A., Gula, L. J., Champagne, J., Healey, J. S., Cameron, D., Simpson, C., Thibault, B., Pintér, A., Tung, S., Sterns, L. D., et al. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.09.020 · Cited by 70 · see citing works
Kirkpatrick, J. N., Ghani, S. N., Burke, M. C., & Knight, B. P. (2007) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2007.00773.x · Cited by 68 · see citing works
Outcome of Young Patients with Abandoned, Nonfunctional Endocardial Leads
Silvetti, M. S., & Drago, F. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01017.x · Cited by 63 · see citing works
Kallinen, L. M., Hauser, R. G., Tang, C., Melby, D. P., Almquist, A. K., Katsiyiannis, W. T., & Gornick, C. C. (2010) · Heart Rhythm · doi:10.1016/j.hrthm.2010.05.015 · Cited by 62 · see citing works
Management of Recalled Pacemakers and Implantable Cardioverter-Defibrillators
Amin, M. S., Matchar, D. B., Wood, M. A., & Ellenbogen, K. A. (2006) · JAMA · doi:10.1001/jama.296.4.412 · Cited by 62 · see citing works
Thal, S. G., Moukabary, T., Boyella, R., Shanmugasundaram, M., Pierce, M., Thai, H. M., & Goldman, S. (2010) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02674.x · Cited by 61 · see citing works
Long‐Term Structural Failure of Coaxial Polyurethane Implantable Cardioverter Defibrillator Leads
Hauser, R. G., Cannom, D. S., Hayes, D. L., Parsonnet, V., HAYES, J., Ratliff, N. B., Tyers, G. F. O., Epstein, A. E., Vlay, S. C., Furman, S., et al. (2002) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2002.t01-1-00879.x · Cited by 61 · see citing works
Learning Curve Characteristics of Pacing Lead Extraction with a Laser Sheath
Bracke, F. A. L. E., Meijer, A., & van Gelder, B. (1998) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1998.tb01172.x · Cited by 59 · see citing works
Predictors and mode of detection of transvenous lead malfunction in implantable defibrillators
Luria, D. M., Glikson, M., Brady, P. A., LEXVOLD, N. Y., Rasmussen, M. J., Hodge, D. O., Chugh, S. S., Rea, R. F., Hayes, D. L., Hammill, S. C., et al. (2001) · The American Journal of Cardiology · doi:10.1016/s0002-9149(00)01536-8 · Cited by 57 · see citing works
Recalls and Safety Alerts Affecting Automated External Defibrillators
Shah, J. S., & Maisel, W. H. (2006) · JAMA · doi:10.1001/jama.296.6.655 · Cited by 57 · see citing works
Self-citationComplications Associated with Generator Replacement in Response to Device Advisories
Costea, A., Rardon, D. P., Padanilam, B. J., Fogel, R. I., & Prystowsky, E. N. (2007) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2007.01047.x · Cited by 54 · see citing works
Sears, S. F., & CONTI, J. B. (2006) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2006.02.066 · Cited by 54 · see citing works
Laser Lead Extraction: Is There a Learning Curve?
Ghosh, N., Yee, R. K., Klein, G. J., Quantz, M., Novick, R. J., Skanes, A. C., & Krahn, A. D. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.09368.x · Cited by 50 · see citing works
Undavia, M. B., Goldstein, N. E., Cohen, P., Sinthawanarong, K., SINGSON, M., Bhutani, D., Munson, T., Gomes, J. A., Fischer, A., & Mehta, D. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01204.x · Cited by 41 · see citing works
Physician Management of Pacemaker and Implantable Cardioverter Defibrillator Advisories
Maisel, W. H. (2004) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2004.00460.x · Cited by 38 · see citing works
Self-citationChanging Trends in Pacemaker and Implantable Cardioverter Defibrillator Generator Advisories
Maisel, W. H., Stevenson, W. G., & Epstein, L. M. (2002) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2002.01670.x · Cited by 35 · see citing works
Self-citationAlert Tones Are Frequently Inaudible among Patients with Implantable Cardioverter‐Defibrillators
SIMONS, E. C., FEIGENBLUM, D. Y., Nemirovsky, D., & Simons, G. R. (2009) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02480.x · Cited by 30 · see citing works
Psychological Distress in Patients with ICD Recall
Cuculi, F., Herzig, W., Kobza, R., & Erné, P. (2006) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2006.00523.x · Cited by 30 · see citing works
Performance of implantable Cardiac Rhythm Management Devices
SONG, S. L. (1994) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1994.tb02410.x · Cited by 30 · see citing works
Gibson, D. P., Kuntz, K. K., Levenson, J. L., & Ellenbogen, K. A. (2008) · EP Europace · doi:10.1093/europace/eun082 · Cited by 28 · see citing works
Birnie, D. H., Sears, S. F., Green, M. S., Lemery, R., Gollob, M. H., & AMYOTTE, B. (2008) · EP Europace · doi:10.1093/europace/eun317 · Cited by 26 · see citing works
Kitamura, S., Satomi, K., Kurita, T., Shimizu, W., Suyama, K., Aihara, N., Niwaya, K., Kobayashi, J., & Kamakura, S. (2006) · Circulation Journal · doi:10.1253/circj.70.273 · Cited by 24 · see citing works
Kawanishi, D. T., SONG, S. L., Furman, S., Parsonnet, V., PIOGER, G., PETITOT, J. C., GODIN, J., & for the BILITCH REGISTRY and STIMAREC (1996) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1996.tb03232.x · Cited by 23 · see citing works
Estimates of Implantable Cardioverter-Defibrillator Complications
Anderson, K. P. (2009) · Circulation · doi:10.1161/circulationaha.108.841452 · Cited by 20 · see citing works
Complications of replacing implantable devices in response to advisories: A single center experience
Moore, J. W., Barrington, W., Bazaz, R. R., Jain, S. K., Němec, J., Ngwu, O., Schwartzman, D. A., Shalaby, A. A., & Saba, S. (2008) · International Journal of Cardiology · doi:10.1016/j.ijcard.2007.12.070 · Cited by 19 · see citing works
Patient Evaluation of ICD Recall Communication Strategies:A Vignette Study
Stutts, L. A., CONTI, J. B., Aranda, J. M., Miles, W. M., Burkart, T. A., & Sears, S. F. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00820.x · Cited by 18 · see citing works
Corbisiero, R., & Armbruster, R. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01076.x · Cited by 17 · see citing works
To Replace or Not to Replace: A Systematic Approach to Respond to Device Advisories
Priori, S. G., Auricchio, A., Nisam, S., & YONG, P. (2008) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2008.01291.x · Cited by 15 · see citing works
Luria, D. M., Bar-Lev, D. S., Gurevitz, O. T., Granit, H., Rotstein, Z., Eldar, M., & Glikson, M. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.00204.x · Cited by 15 · see citing works
Automatic Impedance Monitoring and Patient Alert Feature in Implantable Cardioverter Defibrillators:
Duru, F., Luechinger, R., Scharf, C., & Brunckhorst, C. (2005) · Journal of Cardiovascular Electrophysiology · doi:10.1046/j.1540-8167.2005.40592.x · Cited by 14 · see citing works
Fisher, J. D., Koulogiannis, K. P., Lewallen, L., KATZ, D., Kim, S. G., Ferrick, K. J., GROSS, J. N., KRUMERMAN, A. K., Johnston, D. R., & MERCALDI, B. C. (2009) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02433.x · Cited by 14 · see citing works
Proceedings of the ICD Lead Performance Conference
Maisel, W. H., & Hauser, R. G. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.05.034 · Cited by 13 · see citing works
Self-citationKoplan, B. A., Weiner, S., GILLIGAN, D., Thackeray, L., Merkel, S., Berg, K. C., & for the REFLEx Study Investigators (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00924.x · Cited by 13 · see citing works
FDA Recalls: How Do Pacemaker Manufacturers Compare?
Tyers, G. F. O. (1989) · The Annals of Thoracic Surgery · doi:10.1016/s0003-4975(10)62864-0 · Cited by 13 · see citing works
High Quality Performance of Pacemakers and Implantable Defibrillators
LEVINE, P. A., Stanton, M. S., & SIMS, J. J. (2002) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2002.01667.x · Cited by 11 · see citing works
Mahajan, T., Dubin, A. M., Atkins, D. L., BRADLEY, D. J., SHANNON, K. M., Erickson, C. C., Franklin, W. H., Cecchin, F., Berul, C. I., & members of the Pediatric and Congenital Electrophysiology Society (2008) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2008.01259.x · Cited by 10 · see citing works
Clinical Judgment Versus Decision Analysis for Managing Device Advisories
Amin, M. S., Wood, M. A., Shepard, R. K., Kalahasty, G., & Ellenbogen, K. A. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01171.x · Cited by 9 · see citing works
Maisel, W. H. (2005) · American Journal of Therapeutics · doi:10.1097/01.mjt.0000155117.55919.43 · Cited by 8 · see citing works
Self-citationKüpper, B., Yee, R. K., O’Hara, G. L., Simpson, C. S., Bella, P. E. D., Workel, M., HULL, M. L., & Heuer, H. (2007) · EP Europace · doi:10.1093/europace/eum123 · Cited by 8 · see citing works
Implantable cardioverter-defibrillator lead complication: When is an outbreak out of bounds?
Maisel, W. H. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.10.014 · Cited by 8 · see citing works
Self-citationKrahn, A. D., Simpson, C. S., Parkash, R., Yee, R., Champagne, J., Healey, J. S., Cameron, D., Thibault, B., Mangat, I., Tung, S., et al. (2009) · Canadian Journal of Cardiology · doi:10.1016/s0828-282x(09)70502-1 · Cited by 7 · see citing works
Short and Medium Time Experience with a Tined, Multilumen Steroid Eluting Defibrillation Lead
Aass, H., & Ilvento, J. P. (2002) · Journal of Interventional Cardiac Electrophysiology · doi:10.1023/a:1014184425250 · Cited by 6 · see citing works
Amin, M. S., & Ellenbogen, K. A. (2009) · Current Opinion in Cardiology · doi:10.1097/hco.0b013e328333d375 · Cited by 4 · see citing works
Basile, E. M., & Lorell, B. H. (2006) · PubMed · Cited by 4 · see citing works
Product-experience reporting on endocardial defibrillation leads: a 4-year national perspective
Padeletti, L., Pappone, C., Curnis, A., Zanotto, G., Calò, L., Ricciardi, G., Pieragnoli, P., Dondina, C., Raciti, G., & Michelucci, A. (2009) · Expert Review of Medical Devices · doi:10.1586/erd.09.22 · Cited by 1 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Lampert, R., Olshansky, B., Heidbuchel, H., Lawless, C., Saarel, E., Ackerman, M., Calkins, H., Estes, N. A. M., Link, M. S., Maron, B. J., et al. (2013) · Circulation · doi:10.1161/circulationaha.112.000447 · Cited by 286 · see citing works
Safety of Sports for Athletes With Implantable Cardioverter-Defibrillators
Lampert, R. J., Olshansky, B., Heidbüchel, H., Lawless, C. E., Saarel, E. V., Ackerman, M. J., Calkins, H. G., Estes, N. A. M., Link, M. S., Maron, B. J., et al. (2017) · Circulation · doi:10.1161/circulationaha.117.027828 · Cited by 156 · see citing works
Managing with pacemakers and implantable cardioverter defibrillators
Lampert, R. (2013) · Circulation · doi:10.1161/circulationaha.113.001555 · Cited by 34 · see citing works
Evaluation and Management of Arrhythmia in the Athletic Patient
Lampert, R. J. (2012) · Progress in Cardiovascular Diseases · doi:10.1016/j.pcad.2012.01.002 · Cited by 29 · see citing works
Sport Participation in Patients with Implantable Cardioverter-Defibrillators
Lampert, R. J. (2019) · Current Treatment Options in Cardiovascular Medicine · doi:10.1007/s11936-019-0772-x · Cited by 12 · see citing works
Update on Sports Participation for Athletes with Implantable Cardioverter Defibrillators
Lampert, R. J. (2019) · International Journal of Cardiovascular Sciences · doi:10.5935/2359-4802.20190059 · Cited by 3 · see citing works
Changing Views: Safety and Efficacy of Implantable Cardioverter-Defibrillator Therapy in Athletes
Pavlů, L., Hutyra, M., & Táborský, M. (2015) · Central European Journal of Public Health · doi:10.21101/cejph.a4228 · Cited by 2 · see citing works
Salvador, J. J. G. (2015) · Archivos de medicina del deporte: revista de la Federación Española de Medicina del Deporte y de la Confederación Iberoamericana de Medicina del Deporte
- Authors
- Kramer, D. B., & Maisel, W. H.
- ORCID
- Kramer, Daniel B.
- Venue
- Clinical Cardiac Pacing, Defibrillation and Resynchronization Therapy
- Publisher
- Elsevier
- Published
- 2011
- Pages
- 1028-1039
- OpenAlex
- W2503463318
- Cited by
- 8
- References
- 90
- Citations by year
- 2019: 2 · 2017: 1 · 2015: 2 · 2013: 2 · 2012: 1
- FWCI
- 6.59
- Type
- book chapter
- Language
- en
- Open access
- closed
- Locations
- Elsevier eBooks (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Health Sciences › Health Professions › Health Information Management › Healthcare Regulation
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac electrophysiology and arrhythmias
- Keywords
- cardiac pacing, cardiac resynchronization therapy, implantable cardioverter-defibrillator
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review