Cited by 3Congenital Heart Disease StudiesHemodynamic Monitoring and TherapyNon-Invasive Vital Sign Monitoring

Non-invasive cardiac output monitoring (NICOM) in adult congenital heart disease patients with Fontan palliation

Open access
Quain, Angela, Hoyer, Mark, Ephrem, Georges, Kay, W. Aaron·2021·International Journal of Cardiology Congenital Heart Disease

Abstract

Fontan palliation for single ventricle malformations is an increasingly common reason for heart failure in the adult population. Cardiac output (CO) measurement in Fontan physiology is achieved by invasive cardiac catheterization (RHC). Noninvasive CO monitors using thoracic bioreactance (NICOM) have been validated in non-congenital patients but have not been studied in adult Fontan patients. To compare RHC obtained values of CO using the Fick equation with those measured simultaneously by NICOM in a cohort of adults with Fontan palliation. In nineteen patients undergoing routine outpatient RHC, we compared CO values as determined by Fick with those generated by the Starling SV NICOM device. Bland-Altman plots and intraclass correlation coefficients (ICCs) revealed internal consistency within NICOM measurements, however the agreement between RHC and NICOM for CO was poor (ICCs ​∼ ​0.40). We performed sub-analyses using two-sample T-tests and ICCs to determine if clinical cyanosis, acute desaturation, or Fontan pressure affected the difference observed between RHC and NICOM. Neither chronic hypoxia, acute desaturation, nor Fontan pressure measures were found to be associated with the observed difference between the RHC and NICOM measured CO. Our study did not find a correlation between RHC and NICOM derived measures of CO in a cohort of Fontan patients, even in sub-analyses of confounders of Fontan physiology. We observed internal consistency within the device, which may open a role for monitoring of trends rather than absolute values in Fontan patients. Our study was limited due to small sample size.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.ijcchd.2021.100287

Cites

11 of 11 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • ICG has high error compared to reference standard invasive measures in broader populations with cardiovascular diseases and in patients with higher BMI.

    “In healthy subjects with low body mass index (BMI), this technique can provide accurate measures of STIs; however, in broader populations with cardiovascular diseases and in patients with higher BMI, ICG has been shown to have high error compared to reference standard invasive measures [14,15]”

    From A Chest‐Conformable, Wireless Electro‐Mechanical E‐Tattoo for Measuring Multiple Cardiac Time Intervals (2023) · cites this paper as [14]

    The Results report poor agreement between noninvasive NICOM cardiac-output measurements and invasive right-heart catheterization in adults with Fontan palliation (ICC approximately 0.40). The paper studies thoracic bioreactance, not ICG, and this finding does not establish high error across broader cardiovascular populations or in patients with higher BMI. No page numbers were available in our copy. (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • ICG provides accurate measures of STIs in healthy subjects with low body mass index (BMI).

    “In healthy subjects with low body mass index (BMI), this technique can provide accurate measures of STIs; however, in broader populations with cardiovascular diseases and in patients with higher BMI, ICG has been shown to have high error compared to reference standard invasive measures [14,15]”

    From A Chest‐Conformable, Wireless Electro‐Mechanical E‐Tattoo for Measuring Multiple Cardiac Time Intervals (2023) · cites this paper as [14]

    The Methods and Results compare NICOM-derived cardiac output with invasive measurements in adults with Fontan palliation; they do not assess ICG measures of systolic time intervals in healthy, low-BMI subjects. Searches for healthy subjects, low BMI, ICG, and STIs found no passage addressing the attributed claim. Page numbers are unavailable in our copy. (LLM, conf=0.75)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Quain, A., Hoyer, M., Ephrem, G., & Kay, W. A.
Affiliations
  • University of Michigan (US)
  • Riley Hospital for Children (US)
  • Indiana University Indianapolis (US)
Corresponding author
Kay, W. Aaron
Venue
International Journal of Cardiology Congenital Heart Disease
Publisher
Elsevier BV
Published
2021-12
Volume
6
Pages
100287
ISSN
2666-6685 (ISSN-L)
OpenAlex
W3212047978
Cited by
3
References
11
Citations by year
2024: 1 · 2023: 2
FWCI
0.54
Type
journal article
Language
en
Open access
Locations
  • International Journal of Cardiology Congenital Heart Disease (publishedVersion, cc-by)
  • Purdue University Indianapolis (Indiana University) (submittedVersion, cc-by)
  • IUScholarWorks (Indiana University) (submittedVersion, cc-by)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
Methodist Health Foundation
Topics
  • Health Sciences › Medicine › Epidemiology › Congenital Heart Disease Studies
  • Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
Keywords
noninvasive cardiac output monitoring, Nicom, Fontan circulation, adult congenital heart disease, right heart catheterization, Fick equation, device validation, chronic hypoxia, cyanosis, Fontan pressure

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review