Single-Beat Noninvasive Imaging of Cardiac Electrophysiology of Ventricular Pre-Excitation
Open accessAbstract
Objectives The aim of this study was to determine whether noninvasive imaging of cardiac electrophysiology (NICE) is feasible in patients with Wolff-Parkinson-White (WPW) syndrome in the clinical setting of a catheter laboratory and to test the accuracy of the noninvasively obtained ventricular activation sequences as compared with that of standard invasive electroanatomic mapping.Background NICE of ventricular activation could serve as a useful tool in the treatment of cardiac arrhythmias and might help improve our understanding of arrhythmia mechanisms.Methods NICE works by fusing the data from high-resolution electrocardiographic mapping and a model of the patient's cardiac anatomy obtained by magnetic resonance imaging. The ventricular activation sequence was computed with a bidomain theory-based heart model to solve this inverse problem. Noninvasive imaging of cardiac electrophysiology was performed in 7 patients with WPW syndrome undergoing catheter ablation of the accessory pathway. The position error of NICE was defined as the distance between the site of earliest activation computed by NICE and the successful ablation site identified by electroanatomic mapping (CARTO; Biosense Webster, Diamond Bar, California) for normal atrioventricular (AV) conduction as well as for adenosine-induced AV block.Results The error introduced by geometric coupling of the CARTO data and the NICE model was 5 +/- 3 mm (model discretization 10 mm). All ventricular accessory pathway insertion sites were identified with an accuracy of 18.7 +/- 5.8 mm (baseline) and 18.7 +/- 6.4 mm (adenosine).Conclusions The individual cardiac anatomy model obtained for each patient enables accurate noninvasive electrocardiographic imaging of ventricular pre-excitation in patients with WPW syndrome. Noninvasive imaging of cardiac electrophysiology might be used as a complementary noninvasive approach to localize the origin and help identify and understand the underlying mechanisms of cardiac arrhythmias.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
BSPM has nowadays got a real clinical breakthrough as a method for non-invasive 3D EAI, with the competitive concept of ECG imaging (ECGi), a technology that has rapidly evolved in clinically available devices, increasingly used for non-invasive pre-interventional assessment of patients undergoing catheter ablation procedures.
“BSPM has nowadays got a real clinical breakthrough as a method for non-invasive 3D EAI, with the competitive concept of “ECG imaging” (ECGi) (239–241), a technology that has rapidly evolved in clinically available devices (242), increasingly used for non-invasive pre-interventional assessment of patients undergoing catheter ablation procedures (243, 244).”
From Clinical magnetocardiography: the unshielded bet—past, present, and future (2023) · cites this paper as [240]
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Noninvasive electrocardiographic imaging for cardiac electrophysiology and arrhythmia
Ramanathan, C., Ghanem, R. N., Jia, P., Ryu, K., & Rudy, Y. (2004) · Nature Medicine · doi:10.1038/nm1011 · Cited by 646
Nonfluoroscopic, in vivo navigation and mapping technology
Ben‐Haim, S. A., Osadchy, D., Scnuster, I., Gepstein, L., Hayam, G., & Josephson, M. E. (1996) · Nature Medicine · doi:10.1038/nm1296-1393 · Cited by 370
Dong, J., Calkins, H. G., Solomon, S. B., Lai, S., Dalal, D., Lardo, A., Brem, E., Preiss, A., Berger, R. D., Halperin, H. R., et al. (2006) · Circulation · doi:10.1161/circulationaha.105.565200 · Cited by 234
A new method for myocardial activation imaging
Huiskamp, G. J. M., & Greensite, F. S. (1997) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.581930 · Cited by 182
A Noninvasive Imaging Modality for Cardiac Arrhythmias
Burnes, J. E., TACCARDI, B., & Rudy, Y. (2000) · Circulation · doi:10.1161/01.cir.102.17.2152 · Cited by 125
Model-based imaging of cardiac electrical excitation in humans
TILG, B., Fischer, G., Modre, R., Hanser, F., Messnarz, B., Schocke, M. F. H., Kremser, C., Berger, T., Hintringer, F., & Roithinger, F. X. (2002) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2002.804438 · Cited by 106
Self-citationModre, R., TILG, B., Fischer, G., & Wach, P. (2002) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2002.803519 · Cited by 88
Self-citationCheng, L. K., Bodley, J. M., & Pullan, A. J. (2003) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2002.807326 · Cited by 88
Noninvasive three-dimensional electrocardiographic imaging of ventricular activation sequence
Zhang, X., RAMACHANDRA, I., Liu, Z., Muneer, B., Pogwizd, S. M., & He, B. (2005) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00639.2005 · Cited by 82
Atrial Noninvasive Activation Mapping of Paced Rhythm Data
Modre, R., TILG, B., Fischer, G., Hanser, F., Messnarz, B., Seger, M., Schocke, M. F. H., Berger, T., Hintringer, F., & Roithinger, F. X. (2003) · Journal of Cardiovascular Electrophysiology · doi:10.1046/j.1540-8167.2003.02558.x · Cited by 57
Self-citationOn modeling the Wilson terminal in the boundary and finite element method
Fischer, G., TILG, B., Modre, R., Hanser, F., Messnarz, B., & Wach, P. (2002) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.983455 · Cited by 48
Self-citationComparison of Gold Versus Platinum Electrodes on Myocardial Lesion Size Using Radiofrequency Energy
Simmons, W. N., Mackey, S., He, D. S., & Marcus, G. M. (1996) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1996.tb06509.x · Cited by 28
Which patient should be referred to an electrophysiologist: supraventricular tachycardia
SCHILLING, R. J. (2002) · Heart · doi:10.1136/heart.87.3.299 · Cited by 27
Computationally Efficient Noninvasive Cardiac Activation Time Imaging
Pfeifer, B. E., Seger, M., Hintermüller, C., Hanser, F., Modre, R., TILG, B., Trieb, T., Kremser, C., Roithinger, F. X., Hintringer, F., et al. (2005) · Methods of Information in Medicine · doi:10.1055/s-0038-1634024 · Cited by 20
Self-citationA Signal Processing Pipeline for Noninvasive Imaging of Ventricular Preexcitation
Hanser, F., Pfeifer, B. E., Seger, M., Hintermüller, C., Modre, R., TILG, B., Trieb, T., Berger, T., Roithinger, F. X., Hintringer, F., et al. (2005) · Methods of Information in Medicine · doi:10.1055/s-0038-1634001 · Cited by 19
Self-citationAtrial and Ventricular Myocardium Extraction Using Model-based Techniques
Fischer, G., Hanser, F., Seger, M., Hintermüller, C., Modre‐Osprian, R., Trieb, T., TILG, B., & Pfeifer, B. E. (2006) · Methods of Information in Medicine · doi:10.1055/s-0038-1634032 · Cited by 14
Self-citationGreensite, F. S. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.250576 · Cited by 10
Flagged first, then by citations
Cited by
Sorted by citations
Noninvasive Electrocardiographic Imaging of Arrhythmogenic Substrates in Humans
Rudy, Y. (2013) · Circulation Research · doi:10.1161/circresaha.112.279315 · Cited by 185
Non-Invasive Imaging of Cardiac Activation and Recovery
van Dam, P. M., Oostendorp, T. F., Linnenbank, A. C., & van Oosterom, A. (2009) · Annals of Biomedical Engineering · doi:10.1007/s10439-009-9747-5 · Cited by 175
Cluitmans, M. J. M., Brooks, D. H., MacLeod, R., Dössel, O., Guillem, M. S., van Dam, P. M., Švehlíková, J., He, B., Sapp, J. L., Wang, L., et al. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.01305 · Cited by 144
The Inverse Problem of Electrocardiography
Pullan, A. J., Cheng, L. K., Nash, M. P., Ghodrati, A., MacLeod, R., & Brooks, D. H. (2010) · Springer eBooks · doi:10.1007/978-1-84882-046-3_9 · Cited by 116
Forward Problem of Electrocardiography
Bear, L. R., Cheng, L. K., LeGrice, I. J., Sands, G. B., Lever, N. A., Paterson, D. J., & Smaill, B. H. (2015) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.114.001573 · Cited by 116
Ultrasound Current Source Density Imaging
Olafsson, R., Witte, R. S., Huang, S.-W., & O’Donnell, M. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.919115 · Cited by 100
How Accurate Is Inverse Electrocardiographic Mapping?
Bear, L. R., LeGrice, I. J., Sands, G. B., Lever, N. A., Loiselle, D. S., Paterson, D. J., Cheng, L. K., & Smaill, B. H. (2018) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.117.006108 · Cited by 89
Imaging the electromechanical activity of the heart in vivo
Provost, J., Lee, W.-N., Fujikura, K., & Konofagou, E. E. (2011) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1011688108 · Cited by 84
Cluitmans, M. J. M., Peeters, R., Westra, R. L., & Volders, P. G. A. (2015) · Netherlands Heart Journal · doi:10.1007/s12471-015-0690-9 · Cited by 82
Clinical magnetocardiography: the unshielded bet—past, present, and future
Brisinda, D., Fenici, P., & Fenici, R. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1232882 · Cited by 80
How computer simulations of the human heart can improve anti‐arrhythmia therapy
Trayanova, N. A., & Chang, K. C. (2015) · The Journal of Physiology · doi:10.1113/jp270532 · Cited by 63
Pezzuto, S., Prinzen, F. W., Potse, M., Maffessanti, F., Regoli, F. D., Caputo, M. L., Conte, G., Krause, R., & Auricchio, A. (2020) · EP Europace · doi:10.1093/europace/euaa330 · Cited by 55
Provost, J., Gurev, V., Trayanova, N. A., & Konofagou, E. E. (2010) · Heart Rhythm · doi:10.1016/j.hrthm.2010.12.034 · Cited by 55
Noninvasive epicardial and endocardial mapping of premature ventricular contractions
Wißner, E., Revishvili, A. S., Metzner, A., Tsyganov, A., Kalinin, V., Lemeš, C., Saguner, A. M., Maurer, T., Deiß, S., Sopov, O. V., et al. (2016) · EP Europace · doi:10.1093/europace/euw103 · Cited by 52
Electrocardiographic imaging for cardiac arrhythmias and resynchronization therapy
Pereira, H., Niederer, S., & Rinaldi, C. A. (2020) · EP Europace · doi:10.1093/europace/euaa165 · Cited by 51
Berger, T., Pfeifer, B. E., Hanser, F., Hintringer, F., Fischer, G., Netzer, M., Trieb, T., Stuehlinger, M., Dichtl, W., Baumgärtner, C., et al. (2011) · PLoS ONE · doi:10.1371/journal.pone.0016255 · Cited by 47
Self-citationHan, C., Pogwizd, S. M., Killingsworth, C. R., & He, B. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2011.03.014 · Cited by 48
Rantner, L. J., Vadakkumpadan, F., Spevak, P. J., Crosson, J. E., & Trayanova, N. A. (2013) · The Journal of Physiology · doi:10.1113/jphysiol.2013.255109 · Cited by 45
Investigations of sensitivity and resolution of ECG and MCG in a realistically shaped thorax model
Mäntynen, V., Konttila, T., & Stenroos, M. (2014) · Physics in Medicine and Biology · doi:10.1088/0031-9155/59/23/7141 · Cited by 45
Inverse problems in ion channel modelling
Burger, M. (2011) · Inverse Problems · doi:10.1088/0266-5611/27/8/083001 · Cited by 38
Rudic, B., Chaykovskaya, M. K., Tsyganov, A., Kalinin, V., Tülümen, E., Papavassiliu, T., Dösch, C., Liebe, V., Kuschyk, J., Röger, S., et al. (2016) · Journal of the American Heart Association · doi:10.1161/jaha.116.004095 · Cited by 38
Trayanova, N. A., O’Hara, T., Bayer, J. D., Boyle, P. M., McDowell, K. S., Constantino, J. L., Arevalo, H., Hu, Y., & Vadakkumpadan, F. (2012) · EP Europace · doi:10.1093/europace/eus277 · Cited by 36
Graham, A. J., Orini, M., & Lambiase, P. D. (2017) · Arrhythmia & Electrophysiology Review · doi:10.15420/aer.2017.20.1 · Cited by 36
Konofagou, E. E., & Provost, J. (2012) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2011.11.027 · Cited by 34
Liu, C., Skadsberg, N. D., Ahlberg, S. E., Swingen, C., Iaizzo, P. A., & He, B. (2008) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2007.01066.x · Cited by 33
The inverse problem of bioelectricity: an evaluation
van Oosterom, A. (2012) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-012-0941-5 · Cited by 33
Solving the Inverse Problem of Electrocardiography on the Endocardium Using a Single Layer Source
Kalinin, A. V., Potyagaylo, D., & Kalinin, V. (2019) · Frontiers in Physiology · doi:10.3389/fphys.2019.00058 · Cited by 32
Villongco, C., Krummen, D. E., Stark, P., Omens, J. H., & McCulloch, A. D. (2014) · Progress in Biophysics and Molecular Biology · doi:10.1016/j.pbiomolbio.2014.06.011 · Cited by 31
Computational Cardiology: The Heart of the Matter
Trayanova, N. A. (2012) · ISRN Cardiology · doi:10.5402/2012/269680 · Cited by 30
Temporal Sparse Promoting Three Dimensional Imaging of Cardiac Activation
Yu, L., Zhou, Z., & He, B. (2015) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2015.2429134 · Cited by 25
Prakosa, A., Maxime Sermesant, B., Allain, P., Villain, N., Rinaldi, C. A., Rhode, K., Razavi, R., Delingette, H., & Ayache, N. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2281619 · Cited by 25
Liu, C., Eggen, M. D., Swingen, C., Iaizzo, P. A., & He, B. (2012) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2012.2202914 · Cited by 25
Chávez, C. E., Zemzemi, N., Coudière, Y., Alonso‐Atienza, F., & Álvarez, D. (2015) · Lecture notes in computer science · doi:10.1007/978-3-319-20309-6_45 · Cited by 22
Introduction to Noninvasive Cardiac Mapping
Bear, L. R., Cuculich, P. S., Bernus, O., Efimov, I. R., & Dubois, R. (2015) · Cardiac Electrophysiology Clinics · doi:10.1016/j.ccep.2014.11.015 · Cited by 22
Noninvasive Estimation of Global Activation Sequence Using the Extended Kalman Filter
Liu, C., & He, B. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2066564 · Cited by 20
Janssen, A. M., Potyagaylo, D., Dössel, O., & Oostendorp, T. F. (2017) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-017-1715-x · Cited by 20
Noninvasive Imaging of Cardiac Excitation: Current Status and Future Perspective
van der Graaf, A. W. M., Bhagirath, P., Ramanna, H., van Driel, V. J. H. M., de Hooge, J., de Groot, N. M. S., & Götte, M. (2014) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12140 · Cited by 19
Jiang, M., Zhu, L., Wang, Y., Xia, L., Shou, G., Liu, F., & Crozier, . (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/6/013 · Cited by 18
Yu, L. H., Jin, Q., Zhou, Z., Wu, L., & He, B. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2758369 · Cited by 17
Lai, D., Sun, J., Li, Y., & He, B. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/11/3897 · Cited by 17
Experimental Measures of Ventricular Activation and Synchrony
Sutherland, D. R., Ni, Q., MacLeod, R., Lux, R. L., & Punske, B. B. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01227.x · Cited by 14
Modern Electrophysiology Mapping Techniques
Sy, R. W., Thiagalingam, A., & Stiles, M. K. (2012) · Heart Lung and Circulation · doi:10.1016/j.hlc.2012.04.007 · Cited by 12
Fenici, R., & Brisinda, D. (2006) · International journal of cardiac imaging · doi:10.1007/s10554-006-9076-9 · Cited by 12
Yang, T., Pogwizd, S. M., Walcott, G. P., Yu, L. H., & He, B. (2018) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2018.2866951 · Cited by 11
Seger, M., Hanser, F., Dichtl, W., Stuehlinger, M., Hintringer, F., Trieb, T., Pfeifer, B. E., & Berger, T. (2014) · EP Europace · doi:10.1093/europace/euu045 · Cited by 10
Self-citationWang, Y., Wang, H., Yang, Y., Liu, Z.-Y., Pang, J.-J., Cui, P.-L., & Xiang, M. (2025) · Measurement · doi:10.1016/j.measurement.2024.116353 · Cited by 8
Lai, D., Liu, C., Eggen, M. D., Iaizzo, P. A., & He, B. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/13/027 · Cited by 9
Schulze, W. H. W., Chen, Z., Relan, J., Potyagaylo, D., Krueger, M. W., Karim, R., Sohal, M., Shetty, A., Ma, Y., Ayache, N., et al. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-016-1566-x · Cited by 9
Caulier‐Cisterna, R., Sanromán‐Junquera, M., Muñoz-Romero, S., Blanco–Velasco, M., Goya–Esteban, R., Garcı́a-Alberola, A., & Rojo‐Álvarez, J. L. (2020) · Sensors · doi:10.3390/s20113131 · Cited by 8
Localizing the Accessory Pathway in Ventricular Preexcitation Patients Using a Score Based Algorithm
Stuehlinger, M., Nowak, C., Spuller, K., Etsadashvili, K., Stühlinger, X., Berger, T., Dichtl, W., Gothe, R. M., Fischer, G., Hintringer, F., et al. (2011) · Methods of Information in Medicine · doi:10.3414/me11-01-0046 · Cited by 8
Self-citationvan der Graaf, A. W. M., Bhagirath, P., de Hooge, J., Ramanna, H., van Driel, V. J. H. M., de Groot, N. M. S., & Götte, M. (2015) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-015-0054-9 · Cited by 7
Namdar, M., Chmelevsky, M. P., Hachulla, A.-L., & Shah, D. C. (2016) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.116.003949 · Cited by 7
Method for guiding the ablation catheter to the ablation site: a simulation and experimental study
Fukuoka, Y., Oostendorp, T. F., & Armoundas, A. A. (2009) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-009-0441-4 · Cited by 7
A training whole-heart model for simulating propagation and ECG patterns
Pfeifer, B. E., Seger, M., Hanser, F., Hintermüller, C., Fischer, G., Mühlthaler, H., & TILG, B. (2007) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2007.06.002 · Cited by 7
Self-citationZhou, S., Sapp, J. L., Dawoud, F., & Horáček, B. M. (2019) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2872983 · Cited by 6
Activation recovery interval imaging of premature ventricular contraction
Yang, T., Yu, L. H., Jin, Q., Wu, L., & He, B. (2018) · PLoS ONE · doi:10.1371/journal.pone.0196916 · Cited by 6
Fenici, R., & Brisinda, D. (2007) · Lecture Notes in Computer Science · doi:10.1007/978-3-540-72907-5_24 · Cited by 6
Noninvasive imaging of cardiac electrophysiology.
Berger, T., Hintringer, F., & Fischer, G. (2007) · PubMed Central · Cited by 6
Self-citationGillette, K. K., Winkler, B., Kurath‐Koller, S., Scherr, D., Vigmond, E. J., Bär, M., & Plank, G. (2024) · EP Europace · doi:10.1093/europace/euae223 · Cited by 5
Comparison of minimum-norm estimation and beamforming in electrocardiography with acute ischemia
Konttila, T., Mäntynen, V., & Stenroos, M. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/4/623 · Cited by 5
The electrical resistance of the lungs, intercostal muscles, and kidneys in hypertensive ISIAH rats
Kolomeyets, N. L., Смирнова, С. Л., & Roshchevskaya, I. M. (2016) · BIOPHYSICS · doi:10.1134/s0006350916030076 · Cited by 5
Sohn, K., Lv, W., Lee, K., Galea, A. M., Hirschman, G. B., Hayward, A. M., Cohen, R. J., & Armoundas, A. A. (2013) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2013.2268012 · Cited by 4
Analysis of the criteria of activation-based inverse electrocardiography using convex optimization
Erem, B., van Dam, P. M., & Brooks, D. H. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6090972 · Cited by 4
Spitaler, P., Pfeifer, B. E., Mayr, A., Bachler, R., Bilgeri, V., Adukauskaitė, A., Bauer, A., Stuehlinger, M., Barbieri, F., & Dichtl, W. (2023) · Journal of Clinical Medicine · doi:10.3390/jcm12134510 · Cited by 4
Self-citationA Method to Noninvasively Identify Cardiac Bioelectrical Sources
Sohn, K., Lv, W., Lee, K., Galea, A. M., Hirschman, G. B., Barrett, C. D., Cohen, R. J., & Armoundas, A. A. (2014) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.12380 · Cited by 4
Catheter ablation for ventricular tachycardia: indications and techniques
Kessler, E. J., & Knight, B. P. (2007) · Expert Review of Cardiovascular Therapy · doi:10.1586/14779072.5.5.977 · Cited by 4
Chang, Y., Dong, M., Fan, L., Kang, B., Sun, W., Li, X., Zhang, Y., & Ren, M. (2025) · BMC Cardiovascular Disorders · doi:10.1186/s12872-025-04728-2 · Cited by 3
Three-Dimensional Imaging of Ventricular Activation and Electrograms From Intracavitary Recordings
Liu, C., Iaizzo, P. A., & He, B. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2097598 · Cited by 3
Shou, G., Xia, L., Hl, D., & Qian, M. (2010) · Computing in cardiology · Cited by 3
Noninvasive Cardiac Electrical Activity Mapping Systems: Current Available Options
Ballatore, A., Saglietto, A., Cavallone, E., Ferrari, G. M. D., Anselmino, M., & Dusi, V. (2025) · Reviews in Cardiovascular Medicine · doi:10.31083/rcm44335 · Cited by 2
Non-invasive electrophysiological cardiac mapping from development to practice
Artyukhina, Е. А., Yashkov, М. V., & Revishvili, A. S. (2020) · Patologiya krovoobrashcheniya i kardiokhirurgiya · doi:10.21688/1681-3472-2020-4-33-41 · Cited by 2
Cen, W., Hoppe, R., Sun, A., Ding, H., & Gu, N. (2019) · Journal of International Medical Research · doi:10.1177/0300060518824247 · Cited by 2
Tyšler, M., Švehlíková, J., Punshchykova, O., Kneppo, P., & Maksymenko, V. B. (2015) · IEEE International Conference on Electronics and Nanotechnology (ELNANO) · doi:10.1109/elnano.2015.7146902 · Cited by 1
Cakulev, I. T., Sahadevan, J., & Waldo, A. L. (2015) · Cardiac Electrophysiology Clinics · doi:10.1016/j.ccep.2014.11.005 · Cited by 3
Application of L0-Norm Regularization to Epicardial Potential Reconstruction
Wang, L., Li, X., Chen, Y., & Qin, J. (2015) · Lecture notes in computer science · doi:10.1007/978-3-319-24571-3_59 · Cited by 1
Владимирович, З. С., Chmelevsky, M. P., Potyagaylo, D., Budanova, M. A., Дмитриевич, Р. С., Рыжков, А. В., & Lebedev, D. S. (2018) · Computing in cardiology · doi:10.22489/cinc.2018.347 · Cited by 1
Sui, B., Ji, Y., Wang, Y., Gao, L., & Li, M. (2025) · Cardiovascular Innovations and Applications · doi:10.15212/cvia.2025.0006 · Cited by 1
Sohn, K., & Armoundas, A. A. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1437-x · Cited by 1
The shortest path to cardiac activation.
van Dam, P. M. (2010) · Radboud Repository (Radboud University) · Cited by 1
Modeling of the Body Surface Potential Map for Anisotropic Human Heart ActivationPreprint
Elaff, I. (2025) · Research Square · doi:10.21203/rs.3.rs-6840196/v1 · Cited by 1
Effect of the Human Heart Material Anisotropy on Modeling of the Body Surface Potential MapPreprint
Elaff, I. (2025) · Preprints.org · doi:10.20944/preprints202506.1121.v1 · Cited by 1
Cardiac Electrophysiological Imaging: Solving the Inverse Problem of Electrocardiography
He, B., & Liu, C. (2010) · Springer eBooks · doi:10.1007/978-1-4419-6658-2_18 · Cited by 1
Noninvasive Imaging of Cardiac Electrophysiology (NICE)
Seger, M., Pfeifer, B. E., & Berger, T. (2012) · InTech eBooks · doi:10.5772/30019
Self-citationNovel Imaging Strategies for Cardiac Arrhythmias
Deshmukh, A. J., Narula, J., & Sengupta, P. P. (2012) · doi:10.1002/9781118481585.ch55
Liu, C., Skadsberg, N. D., Ahlberg, S. E., Swingen, C., Iaizzo, P. A., & He, B. (2008) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2008.4650223
Berger, T., Hintringer, F., & Fischer, G. (2009) · Circulation · doi:10.1161/circulationaha.108.819789
Self-citationReview Article Noninvasive Imaging of Cardiac Electrophysiology
Berger, T., Hintringer, F., & Fischer, G. (2007)
Self-citationAblation von Vorhofarrhythmien
Dössel, O. (2011) · Elsevier eBooks · doi:10.1016/b978-3-437-24880-1.10034-0
Seger, M., Netzer, M., Pfeifer, B. E., Hanser, F., Baumgärtner, C., Hintringer, F., Pachinger, O. M., Dichtl, W., Stuehlinger, M., Baumgärtner, C., et al. (2010) · Biomedical Engineering · doi:10.2316/p.2010.723-154
Self-citationThe Inverse Problem of Electrocardiography; has it been solved?
van Oosterom, A. (2012)
Cakulev, I. T., Sahadevan, J., & Waldo, A. L. (2015)
Principles and Applications of Computer Modeling in Patients With Devices
Trayanova, N. A. (2016) · Elsevier eBooks · doi:10.1016/b978-0-323-37804-8.00022-5
Non-invasive imaging of ventricular activation during pacing and arrhythmia: Methods and validation
Yu, L. H., Pogwizd, S. M., & He, B. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7590653
Kolomeyets, N. L., & Roshchevskaya, I. M. (2016) · BIOPHYSICS · doi:10.1134/s0006350916050110
BOUNDARY ELEMENT METHOD IN SPATIAL CHARACTERIZATION OF THE ELECTROCARDIOGRAM
Stenroos, M. (2008) · Aaltodoc (Aalto University)
Noninvasive reconstruction of cardiac electrical activity
Cluitmans, M. J. M. (2016) · doi:10.26481/dis.20160929mc
The use of a simple model in the inverse characterization of cardiac ischemic regions
Chávez, C. E., Alonso‐Atienza, F., & Álvarez, D. (2017) · Journal of Inverse and Ill-Posed Problems · doi:10.1515/jiip-2016-0017
Неинвазивное электрокардиографическое картирование желудочковых нарушений ритма сердца
Хлынин, М. С., Batalov, R. E., Popov, S. V., & Криволапов, С. Н. (2013) · Сибирский журнал клинической и экспериментальной медицины
Functional Mapping of Three-Dimensional Electrical Activation in Ventricles
Liu, C. (2010) · University of Minnesota Digital Conservancy (University of Minnesota)
Packer, D. L., & Padmanabhan, D. (2019) · doi:10.1002/9781119152637.ch96
Metadata
- Authors
- Berger, T., Fischer, G., Pfeifer, B., Modre, R., Hanser, F., Trieb, T., Roithinger, F. X., Stuehlinger, M., Pachinger, O., Tilg, B., & Hintringer, F.
- Affiliations
- Innsbruck Medical University (AT)
- UMIT TIROL - Private University for Health Sciences and Health Technology (AT)
- Corresponding author
- Berger, Thomas
- Venue
- Journal of the American College of Cardiology
- Publisher
- Elsevier BV
- Published
- 2006-11
- Volume
- 48
- Issue
- 10
- Pages
- 2045-2052
- ISSN
- 0735-1097 (ISSN-L)
- OpenAlex
- W2062081311
- Cited by
- 107
- References
- 17
- Citations by year
- 2026: 1 · 2025: 7 · 2024: 2 · 2023: 2 · 2021: 1 · 2020: 6 · 2019: 5 · 2018: 5 · 2017: 4 · 2016: 10 · 2015: 11 · 2014: 6 · 2013: 6 · 2012: 9
- FWCI
- 3.32
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- Locations
- Journal of the American College of Cardiology (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Austrian Science Fund
- Oesterreichische Nationalbank
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac Arrhythmias and Treatments
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac electrophysiology and arrhythmias
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Keywords
- Wolff-Parkinson-White syndrome, preexcitation, accessory pathway localization, electroanatomical mapping, bidomain model, inverse problems, catheter ablation, cardiac arrhythmia mechanisms
- MeSH terms
- Adult, Electrodiagnosis, Electroencephalography, Feasibility Studies, Female, Humans, Magnetic Resonance Imaging, Male, Models, Cardiovascular, Wolff-Parkinson-White Syndrome, Catheter Ablation
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review