Investigations of sensitivity and resolution of ECG and MCG in a realistically shaped thorax model
Open accessThe DOI matches its Crossref record; 57 of the 58 references it cites were found in a registry or an index.
Abstract
Solving the inverse problem of electrocardiography (ECG) and magnetocardiography (MCG) is often referred to as cardiac source imaging. Spatial properties of ECG and MCG as imaging systems are, however, not well known. In this modelling study, we investigate the sensitivity and point-spread function (PSF) of ECG, MCG, and combined ECG+MCG as a function of source position and orientation, globally around the ventricles: signal topographies are modelled using a realistically-shaped volume conductor model, and the inverse problem is solved using a distributed source model and linear source estimation with minimal use of prior information. The results show that the sensitivity depends not only on the modality but also on the location and orientation of the source and that the sensitivity distribution is clearly reflected in the PSF. MCG can better characterize tangential anterior sources (with respect to the heart surface), while ECG excels with normally-oriented and posterior sources. Compared to either modality used alone, the sensitivity of combined ECG+MCG is less dependent on source orientation per source location, leading to better source estimates. Thus, for maximal sensitivity and optimal source estimation, the electric and magnetic measurements should be combined.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Indeed, a large body of literature has provided evidence that MCG enhances the non-invasive diagnostic and prognostic capability in numerous clinical fields, such as but not limited to the earlier detection of (sometimes electrically silent) myocardial ischemia (59, 88, 90, 323, 319), the preventive arrhythmogenic risk assessment in adults (53, 225, 325–322) as well as in fetuses (83, 323, 324),”
From Clinical magnetocardiography: the unshielded bet—past, present, and future (2023) · cites this paper as [324]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Inverse Problem Theory and Methods for Model Parameter Estimation
Tarantola, A. (2005) · doi:10.1137/1.9780898717921 · Cited by 6,585
Interpreting magnetic fields of the brain: minimum norm estimates
Hämäläinen, M. S., & Ilmoniemi, R. J. (1994) · Medical & Biological Engineering & Computing · doi:10.1007/bf02512476 · Cited by 1,995 · cited 1×
“00 © 2014 Institute of Physics and Engineering in Medicine Printed in the UK & the USA Phys. Med. Biol. 59 (2014) 7141–7158 doi:10.”
BackgroundBasic mathematical and electromagnetic concepts of the biomagnetic inverse problem
Sarvas, J. (1987) · Physics in Medicine and Biology · doi:10.1088/0031-9155/32/1/004 · Cited by 1,843
The Resolving Power of Gross Earth Data
Backus, G., & Gilbert, F. (1968) · Geophysical Journal International · doi:10.1111/j.1365-246x.1968.tb00216.x · Cited by 1,523
Amann, A., Costello, B. D. L., Miekisch, W., Schubert, J., Buszewski, B., Pleil, J., Ratcliffe, N., & Risby, T. (2014) · Journal of Breath Research · doi:10.1088/1752-7155/8/3/034001 · Cited by 804
Monte Carlo simulation studies of EEG and MEG localization accuracy
Liu, A. K., Dale, A. M., & Belliveau, J. W. (2002) · Human Brain Mapping · doi:10.1002/hbm.10024 · Cited by 349
Distributed current estimates using cortical orientation constraints
Lin, F., Belliveau, J. W., Dale, A. M., & Hämäläinen, M. S. (2006) · Human Brain Mapping · doi:10.1002/hbm.20155 · Cited by 289
Improving source reconstructions by combining bioelectric and biomagnetic data
Fuchs, M., Wagner, M., Wischmann, H.-A., Köhler, T., Theißen, A., Drenckhahn, R., & Buchner, H. (1998) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/s0013-4694(98)00046-7 · Cited by 239
On the theory of the electrocardiogram
Geselowitz, D. B. (1989) · Proceedings of the IEEE · doi:10.1109/5.29327 · Cited by 224
Bioelectricity and biomagnetism
Gulrajani, R. M. (1998) · Cited by 217
Research works printed but not linked: 3
- Dössel O, 2000, Int. J. Bioelectromagn., 2
- 10.5405/jmbe.1026 · doi:10.5405/jmbe.1026
- Oostendorp T, 2001, Proc. 12th Int. Conf. on Biomagnetism, 813
Cited by
A new wearable multichannel magnetocardiogram system with a SERF atomic magnetometer array
Yang, Y., Xu, M., Liang, A., Yin, Y., Ma, X., Gao, Y., & Ning, X. (2021) · Scientific Reports · doi:10.1038/s41598-021-84971-7 · Cited by 146
“To obtain the cardiac magnetic field distribution on the whole thoracic surface of the subject simultaneously, 20-channel MCG signals at a 4 × 5 region were detected, covering the area of (B, C, D, E) × (2, 3, 4, 5, 6), which is close to the heart.”
MethodMillimetre-scale magnetocardiography of living rats with thoracotomy
Arai, K., Kuwahata, A., Nishitani, D., Fujisaki, I., Matsuki, R., Nishio, Y., Xin, Z., Cao, X., Hatano, Y., Onoda, S., et al. (2022) · Communications Physics · doi:10.1038/s42005-022-00978-0 · Cited by 85
Clinical magnetocardiography: the unshielded bet—past, present, and future
Brisinda, D., Fenici, P., & Fenici, R. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1232882 · Cited by 80
Preprocessing and Denoising Techniques for Electrocardiography and Magnetocardiography: A Review
Jia, Y., Pei, H., Liang, J., Zhou, Y., Yang, Y., Cui, Y., & Xiang, M. (2024) · Bioengineering · doi:10.3390/bioengineering11111109 · Cited by 92
Cardiac Conduction Velocity, Remodeling and Arrhythmogenesis
Han, B., Trew, M. L., & Zgierski-Johnston, C. M. (2021) · Cells · doi:10.3390/cells10112923 · Cited by 69
Metadata
- Authors
- Mäntynen, V., Konttila, T., & Stenroos, M.
- Affiliations
- Helsinki University Hospital (FI)
- Aalto University (FI)
- Corresponding author
- Mäntynen, Ville
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2014-11-03
- Volume
- 59
- Issue
- 23
- Pages
- 7141-7158
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W2070183748
- Cited by
- 45
- References
- 58
- Citations by year
- 2026: 10 · 2025: 9 · 2024: 9 · 2023: 1 · 2022: 2 · 2021: 3 · 2020: 2 · 2019: 2 · 2017: 3 · 2016: 2 · 2015: 2
- FWCI
- 1.60
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- iopscience.iop.org/info/page/text-and-data-mining (tdm)
- creativecommons.org/licenses/by/3.0/ (vor)
- Locations
- Physics in Medicine and Biology (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrical and Bioimpedance Tomography
- Physical Sciences › Mathematics › Mathematical Physics › Numerical methods in inverse problems
- Keywords
- electrocardiogram, magnetocardiography, inverse problems, sensitivity distribution, point spread function, thorax model, volume conductor model, distributed source model, source orientation
- MeSH terms
- Electrocardiography, Heart, Humans, Models, Cardiovascular, Sensitivity and Specificity, Thorax, Magnetocardiography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Investigations of sensitivity and resolution of ECG and MCG in a realistically shaped thorax model. https://app.cochat.ai/papers/10.1088/0031-9155/59/23/7141
- The paper
- Mäntynen, V., Konttila, T., & Stenroos, M. (2014). Investigations of sensitivity and resolution of ECG and MCG in a realistically shaped thorax model. Physics in Medicine and Biology. https://doi.org/10.1088/0031-9155/59/23/7141
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.