Voxel-based morphometry at ultra-high fields. A comparison of 7T and 3T MRI data
Open accessAbstract
Recent technological progress enables MRI recordings at ultra-high fields of 7 T and above leading to brain images of higher resolution and increased signal-to-noise ratio. Despite these benefits, imaging at 7 T exhibits distinct challenges due to B1 field inhomogeneities, causing decreased image quality and problems in data analysis. Although several strategies have been proposed, a systematic investigation of bias-corrected 7 T data for voxel-based morphometry (VBM) is still missing and it is an ongoing matter of debate if VBM at 7 T can be carried out properly. Here, an optimized VBM study was conducted, evaluating the impact of field strength (3T vs. 7 T) and pulse sequence (MPRAGE vs. MP2RAGE) on gray matter volume (GMV) estimates. More specifically, twenty-two participants were measured under the conditions 3T MPRAGE, 7 T MPRAGE and 7 T MP2RAGE. Due to the fact that 7 T MPRAGE data exhibited strong intensity inhomogeneities, an alternative preprocessing pipeline was proposed and applied for that data. VBM analysis revealed higher GMV estimates for 7 T predominantly in superior cortical areas, caudate nucleus, cingulate cortex and the hippocampus. On the other hand, 3T yielded higher estimates especially in inferior cortical areas of the brain, cerebellum, thalamus and putamen compared to 7 T. Besides minor exceptions, these results were observed for 7 T MPRAGE as well for the 7 T MP2RAGE measurements. Results gained in the inferior parts of the brain should be taken with caution, as native GM segmentations displayed misclassifications in these regions for both 7 T sequences. This was supported by the test-retest measurements showing highest variability in these inferior regions of the brain for 7 T and also for the advanced MP2RAGE sequence. Hence, our data support the use of 7 T MRI for VBM analysis in cortical areas, but direct comparison between field strengths and sequences requires careful assessment. Similarly, analysis of the inferior cortical regions, cerebellum and subcortical regions still remains challenging at 7 T even if the advanced MP2RAGE sequence is used.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Recent advances in ultra-high resolution with 7 T allow, with certain restrictions to several brain areas, a minimal resolution below 1 mm3.
“While high resolution structural MRI at field strengths of 3 T and computerized volumetric analyses are the current state-of-the-art techniques in vivo to quantify gray matter microstructure at a minimal resolution around 1 mm 3 (Lenglet et al., 2012) recent advances ultra-high resolution with 7 T allow, with certain restrictions to several brain areas, a minimal resolution below 1 mm 3 (Seiger et”
From Serotonin and neuroplasticity – Links between molecular, functional and structural pathophysiology in depression (2017) · cites this paper as Seiger et al., 2015
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Intraclass correlations: Uses in assessing rater reliability.
Shrout, P. E., & Fleiss, J. L. (1979) · Psychological Bulletin · doi:10.1037/0033-2909.86.2.420 · Cited by 23,433 · see citing works
Tzourio-Mazoyer, N., Landeau, B., Papathanassiou, D., Crivello, F., Etard, O., Delcroix, N., Mazoyer, B., & Joliot, M. (2002) · NeuroImage · doi:10.1006/nimg.2001.0978 · Cited by 16,319 · see citing works
Cortical Surface-Based Analysis
Dale, A. M., Fischl, B., & Sereno, M. I. (1999) · NeuroImage · doi:10.1006/nimg.1998.0395 · Cited by 10,910 · see citing works
Voxel-Based Morphometry—The Methods
Ashburner, J., & Friston, K. J. (2000) · NeuroImage · doi:10.1006/nimg.2000.0582 · Cited by 8,676 · see citing works
A fast diffeomorphic image registration algorithm
Ashburner, J. (2007) · NeuroImage · doi:10.1016/j.neuroimage.2007.07.007 · Cited by 8,122 · see citing works
Forming inferences about some intraclass correlation coefficients.
McGraw, K. O., & Wong, S. P. (1996) · Psychological Methods · doi:10.1037/1082-989x.1.1.30 · Cited by 6,903 · see citing works
Measuring the thickness of the human cerebral cortex from magnetic resonance images
Fischl, B. R., & Dale, A. M. (2000) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.200033797 · Cited by 5,789 · see citing works
QUANTIFYING TEST-RETEST RELIABILITY USING THE INTRACLASS CORRELATION COEFFICIENT AND THE SEM
Weir, J. P. (2005) · The Journal of Strength and Conditioning Research · doi:10.1519/15184.1 · Cited by 4,864 · see citing works
A nonparametric method for automatic correction of intensity nonuniformity in MRI data
Sled, J. G., Zijdenbos, A. P., & Evans, A. C. (1998) · IEEE Transactions on Medical Imaging · doi:10.1109/42.668698 · Cited by 4,731 · see citing works
A Voxel-Based Morphometric Study of Ageing in 465 Normal Adult Human Brains
Good, C. D., Johnsrude, I. S., Ashburner, J., Henson, R. N. A., Friston, K. J., & Frackowiak, R. S. J. (2001) · NeuroImage · doi:10.1006/nimg.2001.0786 · Cited by 4,561 · see citing works
Navigation-related structural change in the hippocampi of taxi drivers
Maguire, E. A., Gadian, D. G., Johnsrude, I. S., Good, C. D., Ashburner, J., Frackowiak, R. S. J., & Frith, C. D. (2000) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.070039597 · Cited by 3,225 · see citing works
MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high field
Marques, J. P., Kober, T., Krüger, G., van der Zwaag, W., de Moortele, P. V., & Rolf Gruetter, A. (2009) · NeuroImage · doi:10.1016/j.neuroimage.2009.10.002 · Cited by 1,463 · see citing works
Three‐dimensional magnetization‐prepared rapid gradient‐echo imaging (3D MP RAGE)
Mugler, J. P., & Brookeman, J. R. (1990) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910150117 · Cited by 1,276 · see citing works
Brain morphometry with multiecho MPRAGE
van der Kouwe, A. J. W., Benner, T., Salat, D. H., & Fischl, B. R. (2008) · NeuroImage · doi:10.1016/j.neuroimage.2007.12.025 · Cited by 597 · see citing works
Wright, I. C., McGuire, P. K., Poline, J.-B., Travère, J.-M., Murray, R. M., Frith, C., Frackowiak, R. S. J., & Friston, K. J. (1995) · NeuroImage · doi:10.1006/nimg.1995.1032 · Cited by 530 · see citing works
Transmit and receive transmission line arrays for 7 Tesla parallel imaging
Adriany, G., de Moortele, P. V., Wiesinger, F., et al. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20321 · Cited by 424 · see citing works
Draganski, B., Ashburner, J., Hutton, C., Kherif, F., Frackowiak, R. S. J., Helms, G., & Weiskopf, N. (2011) · NeuroImage · doi:10.1016/j.neuroimage.2011.01.052 · Cited by 288 · see citing works
de Moortele, P. V., Auerbach, E. J., Olman, C. A., Yacoub, E. S., Uǧurbil, K., & Moeller, S. (2009) · NeuroImage · doi:10.1016/j.neuroimage.2009.02.009 · Cited by 266 · see citing works
Bernstein, M. A., Huston, J., & Ward, H. A. (2006) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20698 · Cited by 255 · see citing works
Intensity non-uniformity correction in MRI: Existing methods and their validation
Belaroussi, B., Milles, J., Carme, S., Zhu, Y., & Benoit-Cattin, H. (2005) · Medical Image Analysis · doi:10.1016/j.media.2005.09.004 · Cited by 250 · see citing works
Measurement of basal forebrain atrophy in Alzheimer's disease using MRI
Teipel, S. J., Flatz, W. H., Heinsen, H., Bokde, A. L. W., Schoenberg, S. O., Stöckel, S., Dietrich, O., Reiser, M. F., Möller, H., & Hampel, H. J. (2005) · Brain · doi:10.1093/brain/awh589 · Cited by 248 · see citing works
Cortical abnormalities in bipolar disorder investigated with MRI and voxel-based morphometry
Nugent, A. C., Milham, M. P., Bain, E. E., Mah, L., Cannon, D. M., Marrett, S., Zarate, C. A., Pine, D. S., Price, J. L., & Drevets, W. C. (2006) · NeuroImage · doi:10.1016/j.neuroimage.2005.09.029 · Cited by 206 · see citing works
Magnetic resonance-based morphometry: a window into structural plasticity of the brain
May, A., & Christian Gaser, B. (2006) · Current Opinion in Neurology · doi:10.1097/01.wco.0000236622.91495.21 · Cited by 171 · see citing works
A computational framework for ultra-high resolution cortical segmentation at 7Tesla
Bazin, P., Weiß, M., Dinse, J., Andreas Schäfer, A., Trampel, R., & Turner, R. (2013) · NeuroImage · doi:10.1016/j.neuroimage.2013.03.077 · Cited by 157 · see citing works
Cortical thickness determination of the human brain using high resolution 3T and 7T MRI data
Lüsebrink, F., Wollrab, A., & Speck, O. (2012) · NeuroImage · doi:10.1016/j.neuroimage.2012.12.016 · Cited by 135 · see citing works
Fujimoto, K., Polimeni, J. R., van der Kouwe, A. J. W., Reuter, M., Kober, T., Benner, T., Fischl, B. R., & Wald, L. L. (2013) · NeuroImage · doi:10.1016/j.neuroimage.2013.12.012 · Cited by 106 · see citing works
High-resolution functional MRI of the human amygdala at 7T
Sladky, R., Baldinger, P., Kranz, G. S., Tröstl, J., Höflich, A. S., Lanzenberger, R. R., Moser, E., & Windischberger, C. (2011) · European Journal of Radiology · doi:10.1016/j.ejrad.2011.09.025 · Cited by 92 · see citing works
Self-citationCaramanos, Z., Fonov, V., Francis, S. J., Narayanan, S., Pike, G. B., Collins, D. L., & Arnold, D. L. (2009) · NeuroImage · doi:10.1016/j.neuroimage.2009.08.008 · Cited by 74 · see citing works
van Tol, M., van der Meer, L., Bruggeman, R., Modinos, G., Knegtering, H., & Aleman, A. (2013) · NeuroImage: Clinical · doi:10.1016/j.nicl.2013.12.008 · Cited by 73 · see citing works
Kraus, C., Ganger, S., Losak, J., Hahn, A., Savli, M., Kranz, G. S., Baldinger, P., Windischberger, C., Kasper, S., & Lanzenberger, R. (2014) · NeuroImage · doi:10.1016/j.neuroimage.2013.08.036 · Cited by 54 · see citing works
Self-citationComparing neural response to painful electrical stimulation with functional MRI at 3 and 7T
Hahn, A., Kranz, G. S., Seidel, E., Sladky, R., Kraus, C., Küblböck, M., Pfabigan, D. M., Hummer, A., Grahl, A., Ganger, S., et al. (2013) · NeuroImage · doi:10.1016/j.neuroimage.2013.06.010 · Cited by 47 · see citing works
Self-citationMarchewka, A., Kherif, F., Krüger, G., Grabowska, A., Frackowiak, R. S. J., Draganski, B., & Initiative, T. A. D. N. (2013) · Human Brain Mapping · doi:10.1002/hbm.22297 · Cited by 41 · see citing works
Test–retest variability of [11C]raclopride‐binding potential in nontreatment‐seeking alcoholics
Yoder, K. K., Albrecht, D., Kareken, D. A., Federici, L. M., Perry, K. M., Patton, E. A., Zheng, Q., Mock, B. H., O’Connor, S., & Herring, C. M. (2010) · Synapse · doi:10.1002/syn.20874 · Cited by 25 · see citing works
Dinse, J., Tardif, C. L., Wähnert, M., et al. (2013) · MPG.PuRe (Max Planck Society) · Cited by 1 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Kraus, C., Castrén, E., Kasper, S., & Lanzenberger, R. (2017) · Neuroscience & Biobehavioral Reviews · doi:10.1016/j.neubiorev.2017.03.007 · Cited by 457 · see citing works
Self-citation7T-guided super-resolution of 3T MRI
Bahrami, K., Shi, F., Rekik, I., Gao, Y., & Shen, D. (2017) · Medical Physics · doi:10.1002/mp.12132 · Cited by 56 · see citing works
- Authors
- Seiger, R., Hahn, A., Hummer, A., Kranz, G. S., Ganger, S., Küblböck, M., Kraus, C., Sladky, R., Kasper, S., Windischberger, C., & Lanzenberger, R.
- Affiliations
- Medical University of Vienna (AT)
- Corresponding author
- Lanzenberger, Rupert
- ORCID
- Venue
- NeuroImage
- Publisher
- Elsevier BV
- Published
- 2015-06
- Volume
- 113
- Pages
- 207-216
- ISSN
- 1053-8119 (ISSN-L)
- PMC
- PMC5341769
- OpenAlex
- W2063399989
- Cited by
- 48
- References
- 34
- Citations by year
- 2026: 6 · 2025: 4 · 2024: 3 · 2022: 1 · 2021: 6 · 2020: 6 · 2019: 5 · 2018: 6 · 2017: 5 · 2016: 3 · 2015: 3
- FWCI
- 4.59
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- NeuroImage (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Funders
- Austrian Science Fund (P23021)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced Neuroimaging Techniques and Applications
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Medical Image Segmentation Techniques
- Keywords
- voxel-based morphometry, 7T MRI, ultra-high field MRI, gray matter volume, B1 inhomogeneity, MP2RAGE, MPRAGE, bias correction, gray matter segmentation, test-retest variability, subcortical structures, cerebellum, hippocampus
- MeSH terms
- Adult, Algorithms, Brain, Brain Mapping, Electromagnetic Fields, Female, Humans, Image Processing, Computer-Assisted, Magnetic Resonance Imaging, Male, Reproducibility of Results, Whole Body Imaging, Young Adult, Signal-To-Noise Ratio
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review