IFCC recommendation on reporting results for blood glucose
The DOI matches its Crossref record; 14 of the 14 references it cites were found in a registry or an index.
Abstract
In human beings, glucose is distributed like water between erythrocytes and plasma. The molality of glucose (amount of glucose per unit water mass) is the same throughout the sample. Different water concentrations in calibrator, plasma, and erythrocyte fluid can explain some differences that are dependent on sample type, methods requiring sample dilution, and direct reading biosensors detecting molality. Different devices for the measurement of glucose detect and report fundamentally different analytical quantities. The differences exceed the maximum allowable error of glucose determinations for diagnosing and monitoring diabetes mellitus, and they complicate the treatment. The goal of the International Federation of Clinical Chemistry, Scientific Division, Working Group on Selective Electrodes (IFCC-SD WGSE) is to reach a global consensus on reporting results. The document recommends harmonizing to the concentration of glucose in plasma (with the unit mmol/l), irrespective of sample type or technology. A constant factor of 1.11 will convert measured concentration in whole blood to the equivalent concentration in plasma.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“An IFCC working group recommended that glucose meters report the plasma glucose concentration, irrespective of the sample type or technology (119,120).”
From Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus (2011) · cites this paper as [119]
Citations
Cites
Report of the Expert Committee on the Diagnosis and Classification of Diabetes Mellitus
The Expert Committee on the Diagnosis and Classification of Diabetes Mellitus* (2002) · Diabetes Care · doi:10.2337/diacare.25.2007.s5 · Cited by 10,338
Report of the Expert Committee on the Diagnosis and Classification of Diabetes Mellitus
The Expert Committee on the Diagnosis and Classification of Diabetes Mellitus (1997) · Diabetes Care · doi:10.2337/diacare.20.7.1183 · Cited by 6,220
Self-Monitoring of Blood Glucose
American Diabetes Association (1994) · Diabetes Care · doi:10.2337/diacare.17.1.81 · Cited by 255
Consensus Statement on Self-Monitor ing of Blood Glucose
(1987) · Diabetes Care · doi:10.2337/diacare.10.1.95 · Cited by 248
Kost, G. J., Vu, H.-T., Lee, J. H., Bourgeois, P., Kiechle, F. L., Martin, C., Miller, S. S., Okorodudu, A. O., Podczasy, J. J., Webster, R., et al. (1998) · Critical Care Medicine · doi:10.1097/00003246-199803000-00036 · Cited by 120
What is Blood Glucose: Can it be Measured?
Burrin, J. M., & Alberti, K. G. M. M. (1990) · Diabetic Medicine · doi:10.1111/j.1464-5491.1990.tb01370.x · Cited by 105
Proposal for standardizing direct-reading biosensors for blood glucose
Fogh-Andersen, N., & D’Orazio, P. (1998) · Clinical Chemistry · doi:10.1093/clinchem/44.3.655 · Cited by 84
Whole Blood Glucose: What are We Actually Measuring?
Wiener, K. (1995) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456329503200101 · Cited by 61
Niskanen, L. K., Uusitupa, M. I., Sarlund, H., Siitonen, O., & Pyörälä, K. (1990) · Diabetes Care · doi:10.2337/diacare.13.1.41 · Cited by 54
Direct reading glucose electrodes detect the molality of glucose in plasma and whole blood
Fogh‐Andersen, N., Wimberley, P. D., Thode, J., & Siggaard‐Andersen, O. (1990) · Clinica Chimica Acta · doi:10.1016/0009-8981(90)90232-h · Cited by 52
Research works printed but not linked: 3
- Siggaard-Andersen, 1987, Physicochemical quantities and units in clinical chemistry with special emphasis on activities and activity coefficients. IUPAC/IFCC recommendation, J. Clin. Chem. Clin. Biochem., 25, 369
- Siggaard-Andersen, 1987, Physicochemical quantities and units in clinical chemistry with special emphasis on activities and activity coefficients. IUPAC/IFCC recommendation, Ann. Biol. Clin., 45, 89
- 1990, Consensus statement-self monitoring of blood glucose, Diabetes Care, 13, 41
Cited by
Items 101–115 of 115
Comparative Study of Blood Glucose Collection Methods at PSMMC
Sobki, S. H., Grooni, K. S. A., Anwar, T., & Qasim, M. (2014) · European Journal of Biotechnology and Bioscience · Cited by 1
Consideraciones acerca del uso de glucómetros durante la prueba de tolerancia oral a la glucosa
Lavín, B. A., & Guerra-Ruiz, A. R. (2024) · Advances in Laboratory Medicine / Avances en Medicina de Laboratorio · doi:10.1515/almed-2024-0108 · Cited by 1
Ignell, C. (2015) · Lund University Publications (Lund University) · Cited by 3
Hêches, X. (2005) · Immuno-analyse & Biologie Spécialisée · doi:10.1016/j.immbio.2005.06.011 · Cited by 1
César, K. R., Lamonica, R., & Rabelo, R. (2003) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/00134384-200309000-00011 · Cited by 1
Anderberg, E. (2010) · Lund University Publications (Lund University) · Cited by 1
Sacks, D. B., Arnold, M. A., Metzger, B. E., & Nathan, D. M. (2012) · Cited by 1
Evaluation of negative energy and mineral balance in periparturient Holstein-Friesian cattle
Abdelhameed, A. A. (2016) · Purdue e-Pubs (Purdue University System) · Cited by 1
Diabetes Care in Community Pharmacy - Focus on Self-monitoring of Blood Glucose
Kjome, R. L. S. (2010) · NVA (Nasjonalt vitenarkiv) · Cited by 1
Genetic and Metabolic Markers for the Development of Diabetes after Gestational Diabetes Mellitus
Ekelund, M. (2010) · Lund University Publications (Lund University)
Improving Decision Making in Intensive Care
Meynaar, I. A. (2012) · RePub (Erasmus University Rotterdam)
Arora, G. P. (2017) · Lund University Publications (Lund University)
Mdlalose, T. I. (2017) · ResearchSpace (University of KwaZulu-Natal)
Considerations about the use of glucometers in glucose tolerance tests
Lavín, B. A., & Guerra-Ruiz, A. R. (2024) · Advances in Laboratory Medicine / Avances en Medicina de Laboratorio · doi:10.1515/almed-2024-0176
Methods for the validation and verification of the diagnostic performance ...
Deniz, A., & Pekmezci, A. (2026) · Medycyna Weterynaryjna · doi:10.21521/mw.7116
Metadata
- Authors
- Working Group on Selective Electrodes Scientific Division
- Corresponding author
- Working Group on Selective Electrodes Scientific Division
- Venue
- Clinica Chimica Acta
- Publisher
- Elsevier BV
- Published
- 2001-05
- Volume
- 307
- Issue
- 1-2
- Pages
- 205-209
- ISSN
- 0009-8981 (ISSN-L)
- OpenAlex
- W305401523
- Semantic Scholar
- 2350ae7dbd90b550f8ebde9f187bdac39bded6ae
- Cited by
- 110
- References
- 14
- Citations by year
- 2025: 1 · 2024: 3 · 2023: 6 · 2021: 1 · 2020: 5 · 2019: 3 · 2018: 1 · 2017: 5 · 2016: 7 · 2015: 4 · 2014: 8 · 2013: 4 · 2012: 6
- FWCI
- 2.59
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- Clinica Chimica Acta (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Keywords
- blood glucose, plasma glucose, whole blood glucose, sample dilution, diabetes mellitus
- MeSH terms
- Blood Glucose, Diabetes Mellitus, Humans, Biosensing Techniques, Guidelines as Topic, Clinical Chemistry Tests
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.