IFCC recommendation on reporting results for blood glucose
The DOI matches its Crossref record; 14 of the 14 references it cites were found in a registry or an index.
Abstract
In human beings, glucose is distributed like water between erythrocytes and plasma. The molality of glucose (amount of glucose per unit water mass) is the same throughout the sample. Different water concentrations in calibrator, plasma, and erythrocyte fluid can explain some differences that are dependent on sample type, methods requiring sample dilution, and direct reading biosensors detecting molality. Different devices for the measurement of glucose detect and report fundamentally different analytical quantities. The differences exceed the maximum allowable error of glucose determinations for diagnosing and monitoring diabetes mellitus, and they complicate the treatment. The goal of the International Federation of Clinical Chemistry, Scientific Division, Working Group on Selective Electrodes (IFCC-SD WGSE) is to reach a global consensus on reporting results. The document recommends harmonizing to the concentration of glucose in plasma (with the unit mmol/l), irrespective of sample type or technology. A constant factor of 1.11 will convert measured concentration in whole blood to the equivalent concentration in plasma.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“An IFCC working group recommended that glucose meters report the plasma glucose concentration, irrespective of the sample type or technology (119,120).”
From Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus (2011) · cites this paper as [119]
Citations
Cites
Report of the Expert Committee on the Diagnosis and Classification of Diabetes Mellitus
The Expert Committee on the Diagnosis and Classification of Diabetes Mellitus* (2002) · Diabetes Care · doi:10.2337/diacare.25.2007.s5 · Cited by 10,338
Report of the Expert Committee on the Diagnosis and Classification of Diabetes Mellitus
The Expert Committee on the Diagnosis and Classification of Diabetes Mellitus (1997) · Diabetes Care · doi:10.2337/diacare.20.7.1183 · Cited by 6,220
Self-Monitoring of Blood Glucose
American Diabetes Association (1994) · Diabetes Care · doi:10.2337/diacare.17.1.81 · Cited by 255
Consensus Statement on Self-Monitor ing of Blood Glucose
(1987) · Diabetes Care · doi:10.2337/diacare.10.1.95 · Cited by 248
Kost, G. J., Vu, H.-T., Lee, J. H., Bourgeois, P., Kiechle, F. L., Martin, C., Miller, S. S., Okorodudu, A. O., Podczasy, J. J., Webster, R., et al. (1998) · Critical Care Medicine · doi:10.1097/00003246-199803000-00036 · Cited by 120
What is Blood Glucose: Can it be Measured?
Burrin, J. M., & Alberti, K. G. M. M. (1990) · Diabetic Medicine · doi:10.1111/j.1464-5491.1990.tb01370.x · Cited by 105
Proposal for standardizing direct-reading biosensors for blood glucose
Fogh-Andersen, N., & D’Orazio, P. (1998) · Clinical Chemistry · doi:10.1093/clinchem/44.3.655 · Cited by 84
Whole Blood Glucose: What are We Actually Measuring?
Wiener, K. (1995) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456329503200101 · Cited by 61
Niskanen, L. K., Uusitupa, M. I., Sarlund, H., Siitonen, O., & Pyörälä, K. (1990) · Diabetes Care · doi:10.2337/diacare.13.1.41 · Cited by 54
Direct reading glucose electrodes detect the molality of glucose in plasma and whole blood
Fogh‐Andersen, N., Wimberley, P. D., Thode, J., & Siggaard‐Andersen, O. (1990) · Clinica Chimica Acta · doi:10.1016/0009-8981(90)90232-h · Cited by 52
Research works printed but not linked: 3
- Siggaard-Andersen, 1987, Physicochemical quantities and units in clinical chemistry with special emphasis on activities and activity coefficients. IUPAC/IFCC recommendation, J. Clin. Chem. Clin. Biochem., 25, 369
- Siggaard-Andersen, 1987, Physicochemical quantities and units in clinical chemistry with special emphasis on activities and activity coefficients. IUPAC/IFCC recommendation, Ann. Biol. Clin., 45, 89
- 1990, Consensus statement-self monitoring of blood glucose, Diabetes Care, 13, 41
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
“An IFCC working group recommended that glucose meters report the plasma glucose concentration, irrespective of the sample type or technology (119,120).”
Sacks, D. B., Bruns, D. E., Goldstein, D. E., Maclaren, N. K., McDonald, J. M., & Parrott, M. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.3.436 · Cited by 1,130
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Diabetes Care · doi:10.2337/dc11-9998 · Cited by 718
“An IFCC working group recommended that glucose meters report the plasma glucose concentration, irrespective of the sample type or technology (119,120).”
Dunkley, A. J., Bodicoat, D. H., Greaves, C. J., Russell, C., Yates, T., Davies, M. J., & Khunti, K. (2014) · Diabetes Care · doi:10.2337/dc13-2195 · Cited by 542
“Capillary blood glucose values were converted to plasma equivalent values (28).”
MethodSacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
“An IFCC working group recommended that glucose meters report the plasma glucose concentration, irrespective of the sample type or technology (119,120).”
Accuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System
Weinstein, R. L., Schwartz, S. L., Brazg, R. L., Bugler, J. R., Peyser, T. A., & McGarraugh, G. V. (2007) · Diabetes Care · doi:10.2337/dc06-1602 · Cited by 228
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 208
“An IFCC working group recommended that glucose meters report the plasma glucose concentration, irrespective of the sample type or technology (119,120).”
Effects of Curcuma longa (turmeric) on postprandial plasma glucose and insulin in healthy subjects
Wickenberg, J., Ingemansson, S. L., & Hlebowicz, J. (2010) · Nutrition Journal · doi:10.1186/1475-2891-9-43 · Cited by 197
Brantsma, A. H., Bakker, S. J. L., Hillege, H. L., de Zeeuw, D., de Jong, P. E., Gansevoort, R. T., & the PREVEND Study Group (2005) · Diabetes Care · doi:10.2337/diacare.28.10.2525 · Cited by 141
Prediction of postpartum diabetes in women with gestational diabetes mellitus
Ekelund, M., Shaat, N., Almgren, P.-E., Groop, L., & Berntorp, K. E. (2009) · Diabetologia · doi:10.1007/s00125-009-1621-3 · Cited by 137
“6 mmol/L) was regarded as diagnostic for GDM, while a glucose concentration below that limit was considered as normal glucose tolerance during pregnancy [14,15].”
Josic, J., Olsson, A. T., Wickeberg, J., Lindstedt, S., & Hlebowicz, J. (2010) · Nutrition Journal · doi:10.1186/1475-2891-9-63 · Cited by 104
Overview on self-monitoring of blood glucose
Montagnana, M., Caputo, M., Giavarina, D., & Lippi, G. (2009) · Clinica Chimica Acta · doi:10.1016/j.cca.2009.01.002 · Cited by 116
“A Working Group of the International Federation of Clinical Chemistry (IFCC) has recently recommended that results provided by PGMs should consistently overlap with those obtained on venous blood with traditional laboratory testing, regardless of the matrix and the technology [36,37].”
MethodImpaired fasting glucose prevalence in two nationwide cohorts of obese children and adolescents
Hagman, E., Reinehr, T., Kowalski, J. H., Ekbom, A., Marcus, C., & Holl, R. W. (2013) · International Journal of Obesity · doi:10.1038/ijo.2013.124 · Cited by 109
Analytical performance of glucometers used for routine glucose self-monitoring of diabetic patients
Solnica, B., Naskalski, J. W., & Sieradzki, J. (2003) · Clinica Chimica Acta · doi:10.1016/s0009-8981(03)00079-2 · Cited by 106
Performance Evaluation of Blood Glucose Monitoring Devices
Chen, E. T., Nichols, J. H., Duh, S.-H., & Hortin, G. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322526969 · Cited by 96
Advances of capillary electrophoresis in clinical and forensic analysis (1999-2000)
Thormann, W., Lurie, I. S., McCord, B. R., Marti, U., Cenni, B., & Malik, N. J. (2001) · Electrophoresis · doi:10.1002/1522-2683(200111)22:19<4216::aid-elps4216>3.0.co;2-w · Cited by 100
The Correlation of Hemoglobin A1c to Blood Glucose
Sikaris, K. A. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300305 · Cited by 92
Age and cystatin C in healthy adults: a collaborative study
Odden, M. C., Tager, I. B., Gansevoort, R. T., Bakker, S. J. L., Katz, R., Fried, L. F., Newman, A. B., Canada, R. B., Harris, T., Sarnak, M. J., et al. (2010) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/gfp474 · Cited by 83
“transformed to whole blood values using an internally validated correction factor [18,19].”
MethodAdmission glycaemia and outcome after acute coronary syndrome
Pétursson, P., Herlitz, J. D., Caidahl, K., Guðbjörnsdóttir, S., Karlsson, T., Perers, E., Sjöland, H., & Hartford, M. (2006) · International Journal of Cardiology · doi:10.1016/j.ijcard.2006.04.043 · Cited by 76
“11 was used to convert measured glucose concentrations in whole blood to the equivalent concentration in plasma [11].”
MethodKristensen, G. B., Christensen, N. G., Thue, G., & Sandberg, S. (2005) · Clinical Chemistry · doi:10.1373/clinchem.2005.049080 · Cited by 70
Population response to information on reversibility of Type 2 diabetes
Steven, S., Lim, E. L., & Taylor, R. (2013) · Diabetic Medicine · doi:10.1111/dme.12116 · Cited by 69
Rahim, N. E., Flood, D., Marcus, M. E., Theilmann, M., Aung, T. N., Agoudavi, K., Aryal, K. K., Bahendeka, S., Bicaba, B. W., Bovet, P., et al. (2023) · The Lancet Global Health · doi:10.1016/s2214-109x(23)00348-0 · Cited by 63
“28,29 For 11 surveys that did not specify the device used to measure capillary glucose, we assumed no correction was required as plasma glucose concentration was the most common form of reporting across surveys (ie, no plasma equivalent was computed).”
Trends in the prevalence of gestational diabetes mellitus in southern Sweden, 2003–2012
Ignell, C., Claesson, R., Anderberg, E., & Berntorp, K. (2014) · Acta Obstetricia et Gynecologica Scandinavica · doi:10.1111/aogs.12340 · Cited by 57
“11 was agreed on (7), resulting in a 2-h threshold value of 10.”
BackgroundGerber, K. L., & Freeman, K. P. (2016) · Veterinary Clinical Pathology · doi:10.1111/vcp.12310 · Cited by 62
“Reporting glucose concentrations as plasma equivalent values has been recommended by the International Federation of Clinical Chemistry (IFCC) in an attempt to standardize and facilitate comparison of glucose results between diagnostic laboratories (measuring serum or plasma) and PBGM (measuring glucose in whole blood…”
MethodLacher, D. A., Berman, L. E., Chen, T., & Porter, K. S. (2013) · Clinica Chimica Acta · doi:10.1016/j.cca.2013.03.032 · Cited by 60
Genetic prediction of postpartum diabetes in women with gestational diabetes mellitus
Ekelund, M., Shaat, N., Almgren, P.-E., Anderberg, E., Landin‐Olsson, M., Lyssenko, V., Groop, L., & Berntorp, K. E. (2012) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2012.04.020 · Cited by 57
Accuracy of AccuChek glucose measurement in intensive care patients *
Meynaar, I. A., van Spreuwel, M., Tangkau, P. L., Dawson, L., Visser, S. S., Rijks, L., & Vlieland, T. V. (2009) · Critical Care Medicine · doi:10.1097/00003246-200910000-00002 · Cited by 42
“11 to estimate serum glucose values (16).”
BackgroundEvaluation of 4 Handheld Blood Glucose Monitors for Use in Seabird Rehabilitation
Lieske, C. L., Ziccardi, M. H., Mazet, J. A. K., Newman, S. H., & Gardner, I. A. (2002) · Journal of Avian Medicine and Surgery · doi:10.1647/1082-6742(2002)016[0277:eohbgm]2.0.co;2 · Cited by 49
Geluk, C. A., Asselbergs, F. W., Hillege, H. L., Bakker, S. J. L., de Jong, P. E., Zijlstra, F., & van Gilst, W. H. (2005) · European Heart Journal · doi:10.1093/eurheartj/ehi253 · Cited by 48
Pregnancy to postpartum transition of serum metabolites in women with gestational diabetes
Chorell, E., Andersson-Hall, U. K., Gustavsson, C., Berntorp, K. E., Puhkala, J., Luoto, R. M., Olsson, T., & Holmäng, A. B. (2017) · Metabolism · doi:10.1016/j.metabol.2016.12.018 · Cited by 45
“Capillary blood was analyzed with a HemoCue Glucose+ Analyzer (HemoCue, 86 momoCue, Sweden), and blood glucose concentrations were converted to equivalent plasma 87 glucose concentrations [17].”
MethodGhys, T., Goedhuys, W., Spincemaille, K., Gorus, F., & Gerlo, E. (2007) · Clinica Chimica Acta · doi:10.1016/j.cca.2007.07.025 · Cited by 45
Blood glucose measurements in the critically ill: more than just a blood draw.
Brunkhorst, F. M., & Wahl, H. G. (2006) · Critical Care · doi:10.1186/cc5110 · Cited by 44
High prevalence of prediabetes in a Swedish cohort of severely obese children
Ek, A. E., Rössner, S. M., Hagman, E., & Marcus, C. (2014) · Pediatric Diabetes · doi:10.1111/pedi.12136 · Cited by 44
“11 according to international standard (24), 42% (n = 57) of the included subjects had glucose measurements converted from blood glucose to plasma glucose.”
BackgroundSeasonal Pattern in the Diagnosis of Gestational Diabetes Mellitus in Southern Sweden
Katsarou, A., Claesson, R., Ignell, C., Shaat, N., & Berntorp, K. (2016) · Journal of Diabetes Research · doi:10.1155/2016/8905474 · Cited by 41
“The HemoCue blood glucose 201+ system (HemoCue AB, Ängelholm, Sweden) is used to perform immediate analysis of capillary glucose concentrations, which are reported as equivalent plasma glucose concentrations [14].”
MethodHbA1c as a predictor of diabetes after gestational diabetes mellitus
Claesson, R., Ignell, C., Shaat, N., & Berntorp, K. (2017) · Primary Care Diabetes · doi:10.1016/j.pcd.2016.09.002 · Cited by 35
Anderberg, E., Källén, K., & Berntorp, K. (2010) · Acta Obstetricia et Gynecologica Scandinavica · doi:10.3109/00016349.2010.526186 · Cited by 36
Straseski, J. A., Lyon, M. E., Clarke, W. A., DuBois, J. A., Phelan, L. A., & Lyon, A. W. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2011.165480 · Cited by 39
Ignell, C., & Berntorp, K. (2011) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.3109/00365513.2011.619703 · Cited by 35
“equivalent plasma glucose concentrations [11,12].”
Wickenberg, J., Lindstedt, S., Berntorp, K., Nilsson, J., & Hlebowicz, J. (2012) · British Journal of Nutrition · doi:10.1017/s0007114511005113 · Cited by 34
Berntorp, K. E., Anderberg, E., Claesson, R., Ignell, C., & Källén, K. (2015) · BMC Pregnancy and Childbirth · doi:10.1186/s12884-015-0722-x · Cited by 34
“11 was agreed on [13], resulting in a 2-h threshold value of 10.”
Davies, M. J., Gray, L. J., Ahrabian, D., Carey, M., Farooqi, A., Gray, A. M., Goldby, S., Hill, S., Jones, K. L., Leal, J., et al. (2017) · Programme Grants for Applied Research · doi:10.3310/pgfar05020 · Cited by 34
“Capillary blood glucose values were converted to plasma equivalent values.(67) If studies did not directly report the mean and standard deviation (SD) for change from baseline to 12 months for the outcomes of interest, they were calculated.”
MethodHow accurately do we measure blood glucose levels in intensive care unit (ICU) patients?
Wahl, H. G. (2009) · Best practice & research. Clinical anaesthesiology · doi:10.1016/j.bpa.2009.09.003 · Cited by 32
Megahed, A., Hiew, M. W. H., Townsend, J. R., Messick, J. B., & Constable, P. D. (2015) · Journal of Veterinary Internal Medicine · doi:10.1111/jvim.13608 · Cited by 31
“At a median Hct of 33%, equation (11) predicts that [gluc]p,molar = 1.”
BackgroundMethodOdden, M. C., Tager, I. B., Gansevoort, R. T., Bakker, S. J. L., Fried, L. F., Newman, A. B., Katz, R., Satterfield, S., Harris, T. B., Sarnak, M. J., et al. (2013) · Annals of Epidemiology · doi:10.1016/j.annepidem.2012.12.004 · Cited by 29
“Plasma glucose values in PREVEND were transformed to whole blood values using an internally validated correction factor [23,24].”
MethodLaboratory Tests in Diagnosis and Management of Diabetes Mellitus. Practical Considerations
Miedema, K. (2003) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2003.193 · Cited by 31
Gijzen, K., Moolenaar, D., Weusten, J. J. A. M., Pluim, H. J., & Demir, A. Y. (2012) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2012-0104 · Cited by 28
Probable Blind Spot in the International Diabetes Federation Definition of Metabolic Syndrome
Chen, H., & Pan, W. (2007) · Obesity · doi:10.1038/oby.2007.633 · Cited by 27
Erroneous inflation of diabetes prevalence: Are there global implications?
Taylor, R. J., Zimmet, P., Naseri, T. K., Hufanga, S., Tukana, I., Magliano, D. J., Lin, S., Linhart, C., & Morrell, S. L. (2016) · Journal of Diabetes · doi:10.1111/1753-0407.12447 · Cited by 26
Role of HbA1c in post-partum screening of women with gestational diabetes mellitus
Claesson, R., Ekelund, M., Ignell, C., & Berntorp, K. E. (2014) · Journal of Clinical & Translational Endocrinology · doi:10.1016/j.jcte.2014.10.003 · Cited by 26
“11 for the conversion between concentration of glucose in blood and the equivalent concentration in plasma is recommended, assuming a normal hematocrit [26].”
BackgroundSteffes, M. W., & Sacks, D. B. (2005) · Clinical Chemistry · doi:10.1373/clinchem.2004.044867 · Cited by 24
What Is Accuracy and How Close Must the Agreement Be?
Nichols, J. H. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.558 · Cited by 22
Quality Control of Self-Monitoring of Blood Glucose: Why and How?
Solnica, B., & Naskalski, J. W. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100204 · Cited by 22
Accuracy Evaluation of a New Real-Time Continuous Glucose Monitoring Algorithm in Hypoglycemia
Mahmoudi, Z., Jensen, M. H., Johansen, M. D., Christensen, T. F., Tarnow, L., Christiansen, J. S., & Hejlesen, O. K. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0043 · Cited by 21
Arora, G. P., Åkerlund, M., Brøns, C., Moen, G.-H., Wasenius, N., Sommer, C., Jenum, A. K., Almgren, P., Thaman, R. G., Orho‐Melander, M., et al. (2019) · Journal of Internal Medicine · doi:10.1111/joim.12903 · Cited by 20
Samtani, M. N. (2010) · Biopharmaceutics & Drug Disposition · doi:10.1002/bdd.700 · Cited by 21
“11 was used to convert measured concentration in whole blood to the equivalent concentration in plasma [17].”
MethodPapadopoulou, A., Lynch, K. F., Anderberg, E., Landin‐Olsson, M., Hansson, I., Agardh, C., Lernmark, Å., & Berntorp, K. E. (2011) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2011.10.037 · Cited by 19
“6 mmol/L) was regarded as diagnostic for GDM, while a glucose concentration below that limit was considered as normal glucose tolerance during pregnancy [7,8].”
ANDERBERG, E., LANDIN‐OLSSON, M., KALÉN, J., FRID, A., URSING, D., & BERNTORP, K. (2011) · Acta Obstetricia et Gynecologica Scandinavica · doi:10.1111/j.1600-0412.2011.01214.x · Cited by 19
“11 was agreed on (15); this results in a somewhat lower cut-off for IGT than that given by the WHO (1.”
BackgroundEvaluation of a glucose meter against analytical quality specifications for hospital use
Dhatt, G. S., Agarwal, M., & Bishawi, B. (2004) · Clinica Chimica Acta · doi:10.1016/j.cccn.2004.01.021 · Cited by 19
Kunze, P. E., Perrault, J. R., Chang, Y., Manire, C. A., Clark, S. A., & Stacy, N. I. (2020) · PLoS ONE · doi:10.1371/journal.pone.0229800 · Cited by 17
“To standardize and simplify the comparison of blood glucose results between methods, the International Federation of Clinical Chemistry and Laboratory Medicine recommends glucose concentrations to be reported as plasma equivalent values [12].”
MethodEvaluation of the VIA® Blood Chemistry Monitor for Glucose in Healthy and Diabetic Volunteers
Ganesh, A., Hipszer, B. R., Loomba, N., Simon, B. F., Torjman, M. C., & Joseph, J. I. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200203 · Cited by 15
The Pituitary Gland and Age-Dependent Regulation of Body Composition
van Beek, A. P., Wolffenbuttel, B. H. R., Runge, E., Trainer, P. J., Jönsson, P., & Kołtowska‐Häggström, M. (2010) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2009-2506 · Cited by 16
“Whole blood concentrations were transformed to plasma values using an internally validated correction factor (30).”
MethodEvaluation of Portable Blood Glucose Meters. Problems and Recommendations
Slingerland, R. J., & Miedema, K. (2003) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2003.186 · Cited by 15
Febiyanto, N., Yamazaki, C., Kameo, S., Sari, D. K., Puspitasari, I. M., Sunjaya, D. K., Herawati, D. M. D., Nugraha, G. I., Fukuda, T., & Koyama, H. (2017) · Biological Trace Element Research · doi:10.1007/s12011-017-1013-3 · Cited by 14
Clinical and Analytical Performance of the Accu-Chek Inform Point-of-Care Glucose Meter
Brouwers, A., Bernard, D. R., & Langlois, M. R. (2005) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/01.poc.0000157098.59514.86 · Cited by 13
Karapinar, T., Tumer, K. C., & Buczinski, S. (2020) · Journal of Veterinary Internal Medicine · doi:10.1111/jvim.15794 · Cited by 11
“In human blood samples, increased Hct can cause decreases in blood glucose measurements and vice versa.(10,11) Changes in Hct seen in calves with different diseases such as diarrhea, and infection might therefore affect agreement of glucometer relative to reference method.”
Kristensen, K. H., Wangel, A., Katsarou, A., Shaat, N., Simmons, D. S., Fadl, H., & Berntorp, K. E. (2020) · Journal of Diabetes Research · doi:10.1155/2020/7937403 · Cited by 11
Postma, M. R., van Beek, A. P., Jönsson, P., van Bunderen, C. C., Drent, M. L., Mattsson, A. F., & Camacho‐Hübner, C. (2019) · Neuroendocrinology · doi:10.1159/000499430 · Cited by 10
Hopf, S., Graf, B. M., & Gruber, M. A. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0051 · Cited by 10
“To generate comparable results to the handheld glucose meters, data from the YSI 2300 analyzer were converted to plasma values according to International Federation of Clinical Chemistry and Laboratory Medicine recommendations.(32) The performance of the YSI 2300 analyzer was tested semidaily, with QC material provide…”
Tietäväinen, J., Mäkelä, S., Huhtala, H., Pörsti, I. H., Strandin, T. M., Vaheri, A. I., & Mustonen, J. T. (2021) · Viruses · doi:10.3390/v13061177 · Cited by 9
Svensson, H. G., Wetterling, L., Andersson-Hall, U. K., Jennische, E., Edén, S., Holmäng, A. B., & Lönn, M. (2018) · Adipocyte · doi:10.1080/21623945.2018.1521230 · Cited by 9
“with a HemoCue Glucose 201+ Analyzer (HemoCue, Ängelholm, Sweden), and blood glucose concentrations were converted to equivalent plasma glucose concentrations.(29) Between 5 and 6 years after GDM diagnosis, the women were contacted by telephone.”
MethodPeña, L. W., Flatland, B., Behrend, E. N., Arzón-Pereira, A., Cole, J. E., & Raz, M. L. (2023) · Journal of Veterinary Diagnostic Investigation · doi:10.1177/10406387231195386 · Cited by 8
“4,10 Whole blood and plasma glucose equivalent concentrations differ because glucose is dissolved in water, and the water content of RBC cytosol (reportedly 71% in humans) and plasma (93”
BackgroundPétursson, P., Herlitz, J. D., Caidahl, K., From-Attebring, M., Sjöland, H., Guðbjörnsdóttir, S., & Hartford, M. (2006) · Scandinavian Cardiovascular Journal · doi:10.1080/14017430600797626 · Cited by 8
“11 was used to convert measured glucose concentration in whole blood to the equivalent concentration in plasma (17).”
MethodStolz, M., Schlawne, C., Hoffmann, J., Hartmann, V., Marini, I., Fritsche, A., Peter, A., Bakchoul, T., & Schick, F. (2020) · NMR in Biomedicine · doi:10.1002/nbm.4358 · Cited by 7
Diabetes mellitus - laboratory diagnostics and patient status monitoring
Friedecky, B., Kratochvíla, J., Springer, D., Prázný, M., Pelikánová, T., Zima, T., & Racek, J. (2019) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2019.007 · Cited by 7
Performance Evaluation of Glucometer Barozen H Based on ISO 15197 Standards
Kim, Y.-B., Seo, J.-Y., Lee, S., & Park, H. (2014) · Laboratory Medicine Online · doi:10.3343/lmo.2015.5.1.6 · Cited by 6
Particularidades de la medición de la glucemia capilar: aspectos técnicos, clínicos y legales
Parramón, M. (2004) · Endocrinología y Nutrición · doi:10.1016/s1575-0922(04)74640-9 · Cited by 6
Seccia, T. M., & Rossi, G. P. (2005) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408360500295600 · Cited by 6
Analytical Performance Evaluation of Glucose Monitoring System Following ISO15197
An, D., Chung, H., Lee, H.-W., Lee, W., Chun, S., & Min, W. (2009) · Annals of Laboratory Medicine · doi:10.3343/kjlm.2009.29.5.423 · Cited by 6
Kinchiku, S., Kotani, K., Kajiya, S., Yodo, K., Maruguchi, Y., & Uenomachi, H. (2012) · Journal of Primary Health Care · doi:10.1071/hc12298 · Cited by 5
“jp J PRIM HEALTH CARE 2012;4(4):294–298.”
Towards normalized birthweight in gestational diabetes mellitus
Nilsson, C., Ursing, D., Strevens, H., & Landin‐Olsson, M. (2015) · Acta Obstetricia Et Gynecologica Scandinavica · doi:10.1111/aogs.12695 · Cited by 5
“0 mmol/L was used (18) but this was changed to plasma glucose in 2004 to align with international standards.”
MethodHagman, E. (2016) · Figshare · Cited by 5
Fabre-Estremera, B., Martínez-Chávez, E., Ocaña, M. M., Carcavilla, A., Sánchez, M. D. L. Á. M., Tejado, I. P., Casado, I. G., Losantos-García, I., Fernández–Calle, P., Soto, A. B., et al. (2024) · Advances in Laboratory Medicine / Avances en Medicina de Laboratorio · doi:10.1515/almed-2024-0017 · Cited by 4
Fabre-Estremera, B., Martínez-Chávez, E., Ocaña, M. M., Carcavilla, A., Sánchez, M. D. L. Á. M., Tejado, I. P., Casado, I. G., Losantos-García, I., Fernández–Calle, P., Soto, A. B., et al. (2023) · Advances in Laboratory Medicine / Avances en Medicina de Laboratorio · doi:10.1515/almed-2023-0089 · Cited by 4
“– Accu-Chek ® Inform-II (Roche Diagnostics, Basel, Switzerland): POCT ACI , connected, enzymatic method (quinoprotein glucose dehydrogenase), venous sample (whole blood, converted to plasma [11]).”
MethodEarly versus late initiation of GH replacement in adult-onset hypopituitarism
Postma, M. R., Burman, P., & van Beek, A. P. (2020) · Endocrine Connections · doi:10.1530/ec-20-0098 · Cited by 4
Haeckel, R., Hänecke, P., Koschinsky, T., Luppa, P. B., Schlebusch, H., & Wahl, H. G. (2004) · Journal of Laboratory Medicine · doi:10.1515/labmed.2004.035 · Cited by 4
Haeckel, R., & Hänecke, P. (2003) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2003.145 · Cited by 4
Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter
Solnica, B., Kuśnierz‐Cabala, B., Słowińska-Solnica, K., Witek, P., Cempa, A., & Malecki, M. T. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500640 · Cited by 4
Diabetes mellitus - laboratory diagnostics and monitoring of patients state
Friedecky, B., Kratochvíla, J., Springer, D., Prázný, M., & Zima, T. (2016) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2016.008 · Cited by 4
Diabetes mellitus - laboratory diagnostics and follow-up of patient condition
Friedecky, B., Zima, T., Kratochvíla, J., & Springer, D. (2012) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2012.021 · Cited by 4
Bahar, B., DeChristopher, P. J., HOLMES, E. W., & Kahn, S. E. (2016) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0000000000000093 · Cited by 3
Aspects of Abnormal Glucose Regulation in Various Manifestations of Coronary Artery Disease
Pétursson, P. (2012) · Gothenburg University Publications Electronic Archive (Gothenburg University) · Cited by 3
Development of a Systematic Quality Control Program for Point-of-Care Glucose Testing
Jo, Y. H., Choi, S., Lee, J. J., Kim, J. G., & Lee, Y.-W. (2023) · Imsang byeong'ri wa jeongdo gwanri/Journal of laboratory medicine and quality assurance · doi:10.15263/jlmqa.2023.45.3.107 · Cited by 2
Prasertsri, P., Padkao, T., Boonla, O., Buddhisa, S., Prakobkaew, N., Sripinyowanich, S., & Phoemsapthawee, J. (2025) · International Journal of Molecular Sciences · doi:10.3390/ijms262412054 · Cited by 2
“Plasma glucose concentrations were determined using an enzymatic hexokinase-based assay following standard clinical laboratory protocols [62].”
Lee, M.-S., Tsao, L., Tseng, Y.-C., Tsai, S.-J., Tsai, M.-L., Huang, T. C., & Chiu, H.-Y. (2011) · Pediatrics & Neonatology · doi:10.1016/j.pedneo.2010.12.008 · Cited by 2
Li, X., Ma, Y., Chen, T., Tang, J., & Ma, X. (2017) · The American Journal of the Medical Sciences · doi:10.1016/j.amjms.2017.05.005 · Cited by 2
Geluk, C. A., Asselbergs, F. W., & Hillege, H. L. (2005) · ACC Current Journal Review · doi:10.1016/j.accreview.2005.09.018 · Cited by 2
Mdlalose, T. I., Christison, K. W., & Vosloo, A. (2017) · Aquaculture · doi:10.1016/j.aquaculture.2016.12.024 · Cited by 2
Metrological Aspects of Glucose Measurements by Biosensors
Wunderli, S., & Andres, H. (2009) · Electroanalysis · doi:10.1002/elan.200904625 · Cited by 2
Nichols, J. H. (2004) · Journal of Laboratory Medicine · doi:10.1515/labmed.2004.036 · Cited by 2
Zandén, L., & Bergh, H. (2014) · Scandinavian Journal of Primary Health Care · doi:10.3109/02813432.2014.909204 · Cited by 1
Metadata
- Authors
- Working Group on Selective Electrodes Scientific Division
- Corresponding author
- Working Group on Selective Electrodes Scientific Division
- Venue
- Clinica Chimica Acta
- Publisher
- Elsevier BV
- Published
- 2001-05
- Volume
- 307
- Issue
- 1-2
- Pages
- 205-209
- ISSN
- 0009-8981 (ISSN-L)
- OpenAlex
- W305401523
- Semantic Scholar
- 2350ae7dbd90b550f8ebde9f187bdac39bded6ae
- Cited by
- 110
- References
- 14
- Citations by year
- 2025: 1 · 2024: 3 · 2023: 6 · 2021: 1 · 2020: 5 · 2019: 3 · 2018: 1 · 2017: 5 · 2016: 7 · 2015: 4 · 2014: 8 · 2013: 4 · 2012: 6
- FWCI
- 2.59
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- Clinica Chimica Acta (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Keywords
- blood glucose, plasma glucose, whole blood glucose, sample dilution, diabetes mellitus
- MeSH terms
- Blood Glucose, Diabetes Mellitus, Humans, Biosensing Techniques, Guidelines as Topic, Clinical Chemistry Tests
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.