STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 15,601–15,700 of 47,868
Accuracy and Reliability of Pulse O2 Saturation Measured by a Wrist-worn Oximeter
Hermand, É., Coll, C., Richalet, J., & Lhuissier, F. (2021) · International Journal of Sports Medicine · doi:10.1055/a-1337-2790 · Cited by 35
Test–retest reliability and validity of body composition methods in adults
de Zepetnek, J. O. T., Lee, J. J., Boateng, T., Plastina, S., Cleary, S. R., Huang, L., Kucab, M., Paterakis, S., Brett, N. R., & Bellissimo, N. (2021) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.12716 · Cited by 37
Agreement between laboratory-based and non-laboratory-based Framingham risk score in Southern Iran
Rezaei, F., Seif, M., Gandomkar, A., Fattahi, M. R., & Hasanzadeh, J. (2021) · Scientific Reports · doi:10.1038/s41598-021-90188-5 · Cited by 35
Kavurt, A. V., Bağrul, D., Gül, A. E. K., Özdemiroğlu, N., Ece, İ., Çetin, İ. İ., Özcan, S., Uyar, E., Emeksız, S., Çelikel, E., et al. (2021) · Pediatric Cardiology · doi:10.1007/s00246-021-02738-3 · Cited by 36
Busk, H., Holm, P. M., Skou, S. T., Seitner, S., Siemsen, T., & Wienecke, T. (2022) · Physiotherapy Theory and Practice · doi:10.1080/09593985.2022.2030830 · Cited by 37
Zhan, J., den Eynde, J. V., Cordrey, K., Long, R., Danford, D. A., Hays, A. G., Barnes, B., & Kutty, S. (2022) · International Journal of Cardiology · doi:10.1016/j.ijcard.2022.06.067 · Cited by 31
High-throughput 3DRA segmentation of brain vasculature and aneurysms using deep learning
Lin, F., Xia, Y., Song, S., Ravikumar, N., & Frangi, A. F. (2023) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2023.107355 · Cited by 36
Eylon, G., Tikotzky, L., & Dinstein, I. (2023) · Sleep Health · doi:10.1016/j.sleh.2023.04.001 · Cited by 34
Yang, P.-K., Filtjens, B., Ginis, P., Goris, M., Nieuwboer, A., Gilat, M., Slaets, P., & Vanrumste, B. (2024) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-024-01320-1 · Cited by 39
Papastefanou, I., Souka, A. P., Pilalis, A., Panagopoulos, P. K., & Kassanos, D. (2011) · Prenatal Diagnosis · doi:10.1002/pd.2869 · Cited by 34
Assessing Hepatomegaly: Automated Volumetric Analysis of the Liver
Linguraru, M., Sandberg, J., Jones, E., Petrick, N., & Summers, R. (2012) · Academic Radiology · doi:10.1016/j.acra.2012.01.015 · Cited by 38
Møller‐Sørensen, H., Graeser, K., Hansen, K. L., Zemtsovski, M., SANDER, E. M., & Nilsson, J. C. (2013) · Acta Anaesthesiologica Scandinavica · doi:10.1111/aas.12227 · Cited by 31
Minsart, A.-F., Sima, F. N., Vandenhoute, K., Jani, J. C., & Pachterbeke, C. V. (2012) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.11155 · Cited by 31
Steiner, M. F. C., Aikman‐Green, S., Prescott, G. J., & Dick, F. D. (2011) · Skin Research and Technology · doi:10.1111/j.1600-0846.2011.00509.x · Cited by 31
Dvir, Z. (2003) · Isokinetics and Exercise Science · doi:10.3233/ies-2003-0109 · Cited by 31
What are the current limits of the Kinect sensor
Bonnechère, B., Jansen, B., Bouzahouene, H., Omelina, L., Cornelis, J. P., Rooze, M. A., & Jan, S. V. S. (2012) · Dépôt institutionnel de l'Université libre de Bruxelles (Université Libre de Bruxelles) · Cited by 44
Mitra, B., O’Reilly, G. M., Collecutt, M. F., Cameron, P. A., Phillips, L. E., & Davis, A. K. (2012) · Emergency Medicine Australasia · doi:10.1111/j.1742-6723.2012.01556.x · Cited by 31
Reliability and validity of a MR‐based volumetric analysis of the intrinsic foot muscles
Cheuy, V. A., Commean, P. K., Hastings, M. K., & Mueller, M. J. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24069 · Cited by 33
Ballesteros-Pomar, M. D., Calleja‐Fernández, A., Díez-Rodríguez, R., Vidal‐Casariego, A., Blanco-Suárez, M. D., & Cano-Rodríguez, I. (2012) · PubMed · doi:10.3305/nh.2012.27.5.5989 · Cited by 36
Cottrell, P. E., Ahmed, S., James, C., Hodson, J. A., McDonnell, P. J., Rauz, S., & Williams, G. P. (2014) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.14-15140 · Cited by 38
Wait, J. M. S., Cody, D., Jones, A. K., Rong, J., Baladandayuthapani, V., & Kappadath, S. C. (2015) · American Journal of Roentgenology · doi:10.2214/ajr.14.13093 · Cited by 31
A Debriefing Session with a Nutritionist Can Improve Dietary Assessment Using Food Diaries
Cantwell, M. M., Millen, A. E., Carroll, R. J., Mittl, B., Hermansen, S., Brinton, L. A., & Potischman, N. A. (2006) · Journal of Nutrition · doi:10.1093/jn/136.2.440 · Cited by 33
Myrvik, M. P., Drendel, A. L., Brandow, A. M., Yan, K., Hoffmann, R. G., & Panepinto, J. A. (2015) · Journal of Pediatric Hematology/Oncology · doi:10.1097/mph.0000000000000306 · Cited by 40
Fujita, Y., Yamamoto, T., Sano, I., Yoshioka, N., & Hinenoya, H. (2004) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000136776.35317.ec · Cited by 31
Effect of Pharmacologic Pupil Dilation on OLCR Optical Biometry Measurements for IOL Predictions
Arriola‐Villalobos, P., Díaz‐Valle, D., Garzón, N., Ruiz‐Medrano, J., Fernández, C., Alejandre‐Alba, N., & Garcia-Sanchez, J. (2013) · European Journal of Ophthalmology · doi:10.5301/ejo.5000332 · Cited by 31
Fuchs, T. A., Stehli, J., Dougoud, S., Sah, B.-R., Bull, S., Clerc, O. F., Possner, M., Buechel, R. R., Gaemperli, O., & Kaufmann, P. A. (2014) · International journal of cardiac imaging · doi:10.1007/s10554-014-0474-0 · Cited by 31
Joneli, J., Lanz, C., & Thormann, W. (2006) · Journal of Chromatography A · doi:10.1016/j.chroma.2006.05.035 · Cited by 31
Assessment of left ventricular size and function in horses using anatomical M-mode echocardiography
Grenacher, P. A., & Schwarzwald, C. C. (2010) · Journal of Veterinary Cardiology · doi:10.1016/j.jvc.2010.01.002 · Cited by 31
Regulating intensity using perceived exertion: effect of exercise duration
Kang, J., Chaloupka, E. C., Biren, G. B., Mastrangelo, M. A., & Hoffman, J. R. (2008) · European Journal of Applied Physiology · doi:10.1007/s00421-008-0922-4 · Cited by 31
Port, R. E., Bernstein, L. J., Barboriak, D. P., Xu, L., Roberts, T. P. L., & van Bruggen, N. (2010) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22399 · Cited by 31
Guoyan Zheng, A., von Recum, J., Nolte, L.-P., Grützner, P. A., Steppacher, S. D., & Franke, J. (2011) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-011-0644-7 · Cited by 31
Cortés, R., Rivera, M., Salvador, A., Bertoméu, V., de Burgos, F. G., Roselló‐Lletí, E., Portolés, M., Payá, R., Martínez‐Dolz, L., & Climent, V. E. (2007) · Heart · doi:10.1136/hrt.2006.096891 · Cited by 31
Chalbot, S., Zetterberg, H., Blennow, K., Fladby, T., Grundke, I., & Iqbal, K. (2010) · Neuroscience Letters · doi:10.1016/j.neulet.2010.05.012 · Cited by 34
Redondo, S., Hristov, M., Gordillo‐Moscoso, A., Ruiz, E., Weber, C., & Tejerina, T. (2008) · Journal of Immunological Methods · doi:10.1016/j.jim.2008.02.011 · Cited by 31
Comparative values of gated blood-pool SPECT and CMR for ejection fraction and volume estimation
Sibille, L., Bouallègue, F. B., Bourdon, A., Micheau, A., Vernhet‐Kovacsik, H., & Mariano‐Goulart, D. (2010) · Nuclear Medicine Communications · doi:10.1097/mnm.0b013e32834155f1 · Cited by 34
Wall, A. J., Capo, R. C., Stewart, B. W., Phan, T. T., Jain, J. C., Hakala, J. A., & Guthrie, G. D. (2013) · Journal of Analytical Atomic Spectrometry · doi:10.1039/c3ja30350k · Cited by 31
Sumikawa, H., Johkoh, T., Nagareda, T., Sekiguchi, J., Matsuo, K., Fujita, Y., Natsag, J., Inoue, A., Mihara, N., Honda, O., et al. (2007) · European Journal of Radiology · doi:10.1016/j.ejrad.2007.03.013 · Cited by 31
Whitehead, K., Langley‐Evans, S. C., Tischler, V., & Swift, J. A. (2013) · Journal of Human Nutrition and Dietetics · doi:10.1111/jhn.12136 · Cited by 34
Hong, Y. T., Kim, S. W., Kim, E. K., & Kim, T. (2010) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2009.10.048 · Cited by 31
Evaluation of Capnography in Nonintubated Emergency Department Patients with Respiratory Distress
Plewa, M. C., Sikora, S., Engoren, M., Tome, D., Thomas, J. O., & Deuster, A. (1995) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1995.tb03106.x · Cited by 31
Comparison between oscillometric- and Doppler-ABI in elderly individuals
Takahashi, I., Furukawa, Ohishi, Takahashi, Matsumoto, & Fujiwara, S. (2013) · Vascular Health and Risk Management · doi:10.2147/vhrm.s39785 · Cited by 34
Shermock, K. M., Bragg, L., Connor, J. T., Fink, J. M., Mazzoli, G., & Kottke‐Marchant, K. (2002) · Pharmacotherapy The Journal of Human Pharmacology and Drug Therapy · doi:10.1592/phco.22.16.1397.33699 · Cited by 32
Roy, A. (2009) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400802535141 · Cited by 31
Zhang, W., Gabos, S., Schopflocher, D., Li, X.-F., Gati, W. P., & Hrudey, S. E. (2009) · Journal of Water and Health · doi:10.2166/wh.2009.009 · Cited by 30
Comparison of Four Methods of Assessing Root Surface Debridement
Chan, Y. K., Needleman, I. G., & Clifford, L. R. (2000) · Journal of Periodontology · doi:10.1902/jop.2000.71.3.385 · Cited by 34
Reproducibility of Doppler ultrasound measurement of resistance index in renal allografts
London, N. J. M., Aldoori, M. I., Lodge, V., Bates, J., Irving, H. C., & GILES, G. R. (1993) · British Journal of Radiology · doi:10.1259/0007-1285-66-786-510 · Cited by 31
Tissue-Informative Mechanism for Wearable Non-invasive Continuous Blood Pressure Monitoring
Woo, S. H., Choi, Y. Y., Kim, D. J., Bien, F. D., & Kim, J. J. (2014) · Scientific Reports · doi:10.1038/srep06618 · Cited by 33
Schellenberg, S., Glaus, T. M., & Reusch, C. E. (2007) · Veterinary Record · doi:10.1136/vr.161.12.418 · Cited by 40
Assessment of Reliability in Isokinetic Testing Among Adolescent Basketball Players
Carvalho, H. M., Coelho-e-Silva, M. J., Ronque, Ê. R. V., Gonçalves, R. S., Philippaerts, R. M., & Malina, R. M. (2011) · Medicina · doi:10.3390/medicina47080063 · Cited by 42
Mafham, M. M., Niculescu‐Duvaz, I., BARRON, J. L., Emberson, J., Dockrell, M. E. C., Landray, M., & Baigent, C. (2007) · Nephron Clinical Practice · doi:10.1159/000102997 · Cited by 31
Jønsson, L. R., Kristensen, M. T., Tibæk, S., Andersen, C., & Juhl, C. B. (2011) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2011.04.020 · Cited by 41
Computerized Adaptive Testing—Ready for Ambulatory Monitoring?
Rose, M., Bjørner, J. B., Fischer, F. H., Anatchkova, M., Gandek, B., Klapp, B. F., & Ware, J. E. (2012) · Psychosomatic Medicine · doi:10.1097/psy.0b013e3182547392 · Cited by 31
Simister, R. J., McLean, M. A., Barker, G. J., & Duncan, J. S. (2009) · Epilepsy Research · doi:10.1016/j.eplepsyres.2008.11.006 · Cited by 33
Elter, K., Şişmanoğlu, A., & Durmuşoğlu, F. (2005) · Gynecological Endocrinology · doi:10.1080/09513590400021128 · Cited by 33
Assessment of Pulmonary Artery Morphology by Optical Coherence Tomography
Li, N., Zhang, S., Hou, J., Jang, I., & Yu, B. (2012) · Heart Lung and Circulation · doi:10.1016/j.hlc.2012.07.014 · Cited by 31
Simple formula for the surface area of the body and a simple model for anthropometry
Reading, B. D., & Freeman, B. J. C. (2005) · Clinical Anatomy · doi:10.1002/ca.20047 · Cited by 32
Eades, S., Read, A. W., Stanley, F. J., Eades, F., McAullay, D. R., & Williamson, A. B. (2008) · Journal of Paediatrics and Child Health · doi:10.1111/j.1440-1754.2008.01306.x · Cited by 32
García, O., Romero, I., Mesa, J. E. G., Moreno, D., Cuétara, E., Rivero, Y., Gutiérrez, A., Pérez, C. L., Álvarez, A., Carnesolta, D., et al. (2010) · Mutation Research/Genetic Toxicology and Environmental Mutagenesis · doi:10.1016/j.mrgentox.2010.11.011 · Cited by 31
Walsh, P. F., Thornton, J. M., Walker, N., McCoy, J. G., Baal, J., Baal, J., Mendoza, N., Banimahd, F., & Asato, J. (2014) · PeerJ · doi:10.7717/peerj.651 · Cited by 37
Falorni, A., Chen, S., Zanchetta, R., Yu, L., Tiberti, C., Bacosi, M. L., Furmaniak, J., Bini, V., Dotta, F., Eisenbarth, G. S., et al. (2011) · Clinical Immunology · doi:10.1016/j.clim.2011.04.012 · Cited by 31
Assessment of Body Composition by Bioelectrical Impedance in Adolescent Patients With Celiac Disease
LORENZO, A. D., Campli, C. D., Andréoli, A., Sasso, G. F., Bonamico, M., & Gasbarrini, A. (1999) · The American Journal of Gastroenterology · doi:10.1111/j.1572-0241.1999.01443.x · Cited by 36
Rozenman, Y., Schwartz, R. S., Shah, H. D., & Parikh, K. H. (2007) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.11.021 · Cited by 33
Hagberg, K., Tranberg, R., Zügner, R., & Danielsson, A. (2010) · Physiotherapy Research International · doi:10.1002/pri.477 · Cited by 31
Porcari, C. E., Amdur, R. L., Koch, E. I., Richard, D. C., Favorite, T., Martis, B., & Liberzon, I. (2009) · Journal of Telemedicine and Telecare · doi:10.1258/jtt.2008.080612 · Cited by 30
D’Assunção, F. L. C., de Albuquerque, D. S., Salazar-Silva, J. R., Santos, V. C. D., & Sousa, J. C. N. (2010) · International Endodontic Journal · doi:10.1111/j.1365-2591.2009.01674.x · Cited by 31
Validity of the International Physical Activity Questionnaire in the Arctic
Dahl‐Petersen, I. K., Hansen, A. W., Bjerregaard, P., Jørgensen, M. E., & Brage, S. (2013) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e31827a6b40 · Cited by 34
O’Halloran, S., & Ilett, K. F. (2008) · Clinical Chemistry · doi:10.1373/clinchem.2008.103952 · Cited by 31
Enhancing runoff modeling with GIS
Coroza, O., Evans, D., & Bishop, I. D. (1997) · Landscape and Urban Planning · doi:10.1016/s0169-2046(97)00015-7 · Cited by 31
Henry, C. J., Lightowler, H. J., Tydeman, E. A., & Skeath, R. (2006) · International Journal of Food Sciences and Nutrition · doi:10.1080/09637480600931626 · Cited by 32
Košťálová, M., Bártková, E., Šajgalíková, K., Dolenská, A., Dušek, L., & Bednařík, J. (2008) · Brain Injury · doi:10.1080/02699050802372190 · Cited by 33
Reproducibility of bladder pressure measurements in critically ill patients
Kimball, E. J., Mone, M. C., Wolfe, T. R., Baraghoshi, G. K., & Alder, S. C. (2007) · Intensive Care Medicine · doi:10.1007/s00134-007-0641-z · Cited by 31
Yuan, L., Wang, T., Liu, F., Cohen, E. D., & Patel, V. V. (2010) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2010.05.017 · Cited by 31
Verschuren, O. W., Zwinkels, M., Obeid, J., Kerkhof, N., Ketelaar, M., & Takken, T. (2013) · Developmental Medicine & Child Neurology · doi:10.1111/dmcn.12214 · Cited by 33
Genton, L., Karsegard, V. L., Zawadynski, S., Kyle, U. G., Pichard, C., Golay, A., & Hans, D. B. (2005) · Clinical Nutrition · doi:10.1016/j.clnu.2005.11.002 · Cited by 34
Japúr, C. C., Penaforte, F. R. O., Chiarello, P. G., Monteiro, J. P., Vieira, M. N. C. M., & Basile-Filho, A. (2009) · Journal of Critical Care · doi:10.1016/j.jcrc.2008.12.007 · Cited by 40
Videler, A. J., Beelen, A., van Schaik, I. N., de Visser, M., & Nollet, F. (2008) · Acta Dermato Venereologica · doi:10.2340/16501977-0143 · Cited by 31
Characterization of the count rate performance of modern gamma cameras
Silosky, M. S., Johnson, V. E., Beasley, C., & Kappadath, S. C. (2013) · Medical Physics · doi:10.1118/1.4792297 · Cited by 31
Thermometry and calorimetry assessment of sweat response during exercise in the heat
Flouris, A. D., & Cheung, S. S. (2009) · European Journal of Applied Physiology · doi:10.1007/s00421-009-1302-4 · Cited by 33
A Chronic Model for Experimental Investigation of Epidural Anesthesia in the Rabbit
Malinovsky, J. M., Bernard, J. M., Baudrimont, M., Dumand, J.-B., & Lepage, J.-Y. (1997) · Regional Anesthesia & Pain Medicine · doi:10.1016/s1098-7339(06)80060-x · Cited by 31
Raundal, P. M., Andersen, P. H., Toft, N., Forkman, B., Munksgaard, L., & Herskin, M. S. (2014) · Veterinary Anaesthesia and Analgesia · doi:10.1111/vaa.12159 · Cited by 34
Lebenberg, J., Buvat, I., Lalande, A., Clarysse, P., Casta, C., Cochet, A., Constantinidès, C., Cousty, J., Cesare, A. D., Jehan‐Besson, S., et al. (2012) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2012.2201737 · Cited by 43
Optoelectronic plethysmography compared to spirometry during maximal exercise
Layton, A. M., Moran, S. L., Garber, C. E., Armstrong, H. F., Basner, R. C., Thomashow, B. M., & Bartels, M. N. (2012) · Respiratory Physiology & Neurobiology · doi:10.1016/j.resp.2012.09.004 · Cited by 39
Stavem, K., Røgeberg, O. J., Olsen, J. A., & Boe, J. (2008) · Addiction · doi:10.1111/j.1360-0443.2008.02278.x · Cited by 42
Renehan, A. G., Jones, J., O’Dwyer, S., & Shalet, S. M. (2003) · Growth Hormone & IGF Research · doi:10.1016/s1096-6374(03)00112-6 · Cited by 33
Lee, H. S., Ko, B. S., Ahn, S. H., Son, B. H., Lee, J. W., Kim, H. J., Yu, J. H., Kim, S., Jung, K. H., Ahn, J., et al. (2014) · Breast Cancer Research and Treatment · doi:10.1007/s10549-014-2920-z · Cited by 31
Björklund, K. L., Sköld, C., Andersson, L., & Dalén, N. (2005) · Knee Surgery Sports Traumatology Arthroscopy · doi:10.1007/s00167-005-0658-1 · Cited by 34
Safari, R., Rowe, P. J., McFadyen, A. K., & Buis, A. W. P. (2013) · The Scientific World JOURNAL · doi:10.1155/2013/486146 · Cited by 31
Joe, E., Lee, J. M., Kim, K. W., Kyung Bun, L. E. E., Kim, S., Baek, J. H., Shin, C., Suh, K., Yi, N. J., Han, J. K., et al. (2012) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.23738 · Cited by 31
Broeren, J., Sunnerhagen, K. S., & Rydmark, M. (2007) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-4-13 · Cited by 33
An Automated Hirschberg Test for Infants
Model, D., & Eizenman, M. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2085000 · Cited by 31
Klarskov, N., & Lose, G. (2007) · British Journal of Urology · doi:10.1111/j.1464-410x.2007.06922.x · Cited by 31
Reliability of radiological and cosmetic measurements in hallux valgus
Saro, C., Johnson, D., de Aragón, J. M., Lindgren, U., & Felländer‐Tsai, L. (2005) · Acta Radiologica · doi:10.1080/02841850500270852 · Cited by 31
Comparison of field methods to estimate fat mass in children
Paineau, D., Chiheb, S., Banu, I., Valensi, P. E., Fontan, J., Gaudelus, J., Chapalain, V., Chumlea, C., Bornet, F. R. J., & Boulier, A. (2008) · Annals of Human Biology · doi:10.1080/03014460801914874 · Cited by 32
A simple method to approximate liver size on cross-sectional images using living liver models
Muggli, D., Müller, M., Karlo, C. A., Fornaro, J., Marinček, B., & Frauenfelder, T. (2009) · Clinical Radiology · doi:10.1016/j.crad.2009.02.013 · Cited by 31
Pablo, L. E., Ferreras, A., Fogagnolo, P., Figus, M., & Pajarín, A. B. (2009) · Eye · doi:10.1038/eye.2009.14 · Cited by 31
Clough, A. R., D'abbs, P., Cairney, S., Gray, D., Maruff, P., Parker, R., & O'reilly, B. (2005) · Australian & New Zealand Journal of Psychiatry · doi:10.1080/j.1440-1614.2005.01634.x · Cited by 32
Validation of the combined olfactory test
Robson, A., Woollons, A. C., RYAN, J., HORROCKS, C., Williams, S., & Dawes, P. J. D. (1996) · Clinical Otolaryngology · doi:10.1111/j.1365-2273.1996.tb01101.x · Cited by 31
Tsuchida, K., García‐García, H. M., Ong, A. T. L., Valgimigli, M., Aoki, J., Rademaker, T., Morel, M.-A. M., van Es, G., Bruining, N., & Serruys, P. W. J. C. (2006) · Catheterization and Cardiovascular Interventions · doi:10.1002/ccd.20581 · Cited by 31
A Novel Analytical Method Using OCT to Describe the Corneoscleral Junction
Tan, B., Graham, A. D., Tsechpenakis, G., & Lin, M. C. (2014) · Optometry and Vision Science · doi:10.1097/opx.0000000000000267 · Cited by 31
Darwent, D., Zhou, X., van den Heuvel, C. J., Sargent, C., & Roach, G. D. (2011) · Chronobiology International · doi:10.3109/07420528.2011.597530 · Cited by 34
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.