Performance Evaluation of Material Decomposition With Rapid-Kilovoltage-Switching Dual-Energy CT and Implications for Assessing Bone Mineral Density
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: The purpose of this article is to quantitatively investigate the accuracy and performance of dual-energy CT (DECT) material density images and to calculate the areal bone mineral density (aBMD) for comparison with dual-energy x-ray absorptiometry (DEXA). MATERIALS AND METHODS: A rapid-kilovoltage-switching DECT scanner was used to create material density images of various two-material phantoms of known concentrations under different experimental conditions. They were subsequently also scanned by single-energy CT and DEXA. The total uncertainty and accuracy of the DECT concentration measurements was quantified by the root-mean-square (RMS) error, and linear regression was performed to evaluate measurement changes under varying scanning conditions. Alterations to accuracy with concentric (anthropomorphic) phantom geometry were explored. The sensitivity of DECT and DEXA to changes in material density was evaluated. Correlations between DEXA and DECT-derived aBMD values were assessed. RESULTS: The RMS error of DECT concentration measurements in air ranged from 9% to 244% depending on the materials. Concentration measurements made off-isocenter or with a different DECT protocol were slightly lower (≈ 5%), whereas measurement in scattering conditions resulted in a reduction of 8-27%. In concentric phantoms, higher-attenuating material in the outer chamber increased measured values of the inner material for all methods. DECT was more sensitive than DEXA to changes in BMD at 2 mg/mL K2HPO4. Measurements of aBMD using DECT and DEXA were highly correlated (R(2) = 0.98). CONCLUSION: DECT material density images were linear in response but prone to poor accuracy and biases. DECT-based aBMD could be used to monitor relative change in bone density.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Cited by
van Hamersvelt, R. W., Schilham, A. M. R., Engelke, K., den Harder, A. M., de Keizer, B., Verhaar, H. J., Leiner, T., de Jong, P. A., & Willemink, M. J. (2017) · European Radiology · doi:10.1007/s00330-017-4801-4 · Cited by 94
Li, X., Li, X., Li, J., Jiao, X., Jia, X., Zhang, X., Fan, G., Yang, J., & Guo, J. (2020) · Clinical Radiology · doi:10.1016/j.crad.2019.11.008 · Cited by 41
Nowak, T., Eberhard, M., Schmidt, B., Frey, D., Distler, O., Saltybaeva, N., Alkadhi, H., & Euler, A. (2021) · Radiology · doi:10.1148/radiol.2020202767 · Cited by 41
Moser, L. J., Pitteloud, J., Mergen, V., Frey, D., Nowak, T., Distler, O., Eberhard, M., & Alkadhi, H. (2024) · American Journal of Roentgenology · doi:10.2214/ajr.23.30347 · Cited by 5
Metadata
- Authors
- Wait, J. M. S., Cody, D., Jones, A. K., Rong, J., Baladandayuthapani, V., & Kappadath, S. C.
- Venue
- American Journal of Roentgenology
- Publisher
- American Roentgen Ray Society
- Published
- 2015-06
- Volume
- 204
- Issue
- 6
- Pages
- 1234-1241
- ISSN
- 0002-9580 (ISSN-L)
- 0361-803X
- 1546-3141
- OpenAlex
- W1939395931
- Cited by
- 31
- References
- 1
- FWCI
- 1.57
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Performance Evaluation of Material Decomposition With Rapid-Kilovoltage-Switching Dual-Energy CT and Implications for Assessing Bone Mineral Density. https://app.cochat.ai/papers/10.2214/ajr.14.13093
- The paper
- Wait, J. M. S., Cody, D., Jones, A. K., Rong, J., Baladandayuthapani, V., & Kappadath, S. C. (2015). Performance Evaluation of Material Decomposition With Rapid-Kilovoltage-Switching Dual-Energy CT and Implications for Assessing Bone Mineral Density. American Journal of Roentgenology. https://doi.org/10.2214/ajr.14.13093
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.