STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 16,201–16,300 of 47,868
Zhu, B., Bronas, U. G., & Fritschi, C. (2018) · Sleep Medicine · doi:10.1016/j.sleep.2018.03.008 · Cited by 34
Rhudy, M. B., & Mahoney, J. M. (2018) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2018.1470692 · Cited by 34
Reproducibility of native T1 mapping for renal tissue characterization at 3T
Dekkers, I. A., Paiman, E. H. M., de Vries, A. P. J., & Lamb, H. J. (2018) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26207 · Cited by 31
Comparison of EQ-5D-5L, VAS, and SF-6D in Thai Patients on Peritoneal Dialysis
Thaweethamcharoen, T., Noparatayaporn, P., Sritippayawan, S., & Aiyasanon, N. (2018) · Value in Health Regional Issues · doi:10.1016/j.vhri.2018.08.005 · Cited by 31
Radiology reporting of low-grade glioma growth underestimates tumor expansion
Gui, C., Lau, J. C., Kosteniuk, S., Lee, D. H., & Megyesi, J. F. (2019) · Acta Neurochirurgica · doi:10.1007/s00701-018-03783-3 · Cited by 30
Validation of the SymptoMScreen with performance-based or clinician-assessed outcomes
Fitzgerald, K. C., Salter, A. R., Tyry, T. M., Fox, R. J., Cutter, G., Mowry, E. M., & Marrie, R. A. (2019) · Multiple Sclerosis and Related Disorders · doi:10.1016/j.msard.2019.01.031 · Cited by 31
Bui, K.-L., Mathur, S., Dechman, G., Maltais, F., Camp, P. G., & Saey, D. (2019) · Physical Therapy · doi:10.1093/ptj/pzz059 · Cited by 32
Assessing bias, precision, and agreement in method comparison studies
Taffé, P. (2019) · Statistical Methods in Medical Research · doi:10.1177/0962280219844535 · Cited by 30
Coventry, L. L., Jacob, A. M., Davies, H. T., Stoneman, L., Keogh, S. J., & Jacob, E. R. (2019) · Journal of Advanced Nursing · doi:10.1111/jan.14078 · Cited by 46
Bush, L. A., Hutchinson, J., Hooson, J., Warthon-Medina, M., Hancock, N., Greathead, K., Knowles, B., Vargas-García, E. J., Gibson, L. E., Margetts, B., et al. (2019) · BMC Nutrition · doi:10.1186/s40795-019-0312-9 · Cited by 38
Concussion-Symptom Rating Correlation Between Pediatric Patients and Their Parents
Patsimas, T., Howell, D. R., Potter, M. N., Provance, A. J., Kirkwood, M. W., & Wilson, J. C. (2020) · Journal of Athletic Training · doi:10.4085/1062-6050-200-19 · Cited by 31
Tzamaloukas, A. H., Rohrscheib, M. R., Ing, T. S., Siamopoulos, K. C., Elisaf, M., & Spalding, C. T. (2004) · The International Journal of Artificial Organs · doi:10.1177/039139880402700904 · Cited by 30
Digital and manual cephalometric analysis
Thurzo, A., Javorka, V., Stanko, P., Lysy, J., Suchancova, B., Lehotská, V., Valkovic, L., & Makovnik, M. (2010) · Bratislavské lekárske listy/Bratislava medical journal · doi:10.1007/bf03657297 · Cited by 30
The Child Self-Refraction Study
He, M., Congdon, N. G., MacKenzie, G. E., Zeng, Y., Silver, J. D., & Ellwein, L. B. (2011) · Ophthalmology · doi:10.1016/j.ophtha.2010.10.003 · Cited by 30
An Evaluation of Three-Dimensional Ultrasonography for The Measurement of Gallbladder Volume
Hashimoto, S., Goto, H., Hirooka, Y., Ito, A., Ishiguro, Y., Kojima, S., Hirai, T., Hayakawa, T., & Naitoh, Y. (1999) · The American Journal of Gastroenterology · doi:10.1016/s0002-9270(99)00671-1 · Cited by 30
Watzinger, N., Lund, G., Higgins, C. B., Wendland, M. F., Weinmann, H., & Saeed, M. (2002) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.10206 · Cited by 31
Bogaard, H. J., Woltjer, H. H., Postmus, P. E., & VRIES, P. M. J. M. D. (1997) · Physiological Measurement · doi:10.1088/0967-3334/18/2/001 · Cited by 30
Metabolic acidosis and breathlessness during exercise and hypercapnia in man.
Lane, R. J. M., & Adams, L. M. (1993) · The Journal of Physiology · doi:10.1113/jphysiol.1993.sp019500 · Cited by 30
GUIDELINES FOR MEASUREMENT VALIDATION IN CLINICAL TRIAL DESIGN
Sanchez, M. M., & Binkowitz, B. S. (1999) · Journal of Biopharmaceutical Statistics · doi:10.1081/bip-100101185 · Cited by 30
Rambaran, C., Jiang, B., Ritter, J. M., Shah, A. M., Kalra, L., & Chowienczyk, P. (2007) · British Journal of Clinical Pharmacology · doi:10.1111/j.1365-2125.2007.03006.x · Cited by 30
Rana, A. K., Wardlaw, J. M., Armitage, P. A., & Bastin, M. E. (2003) · Magnetic Resonance Imaging · doi:10.1016/s0730-725x(03)00087-0 · Cited by 30
Marco, F. D., Sperti, L. R., Milan, B., Stucchi, R., Centanni, S., Brochard, L. J., & Fumagalli, R. (2007) · Intensive Care Medicine · doi:10.1007/s00134-007-0833-6 · Cited by 30
A Standard Procedure for Using TURN180
Simpson, J. M., Worsfold, C., Reilly, E., & Nye, N. (2002) · Physiotherapy · doi:10.1016/s0031-9406(05)60747-x · Cited by 41
Lopez, D. M., Weingarten‐Arams, J., Singer, L. P., & Conway, E. E. (2000) · Critical Care Medicine · doi:10.1097/00003246-200006000-00053 · Cited by 30
Smith, J. (1998) · Journal of Advanced Nursing · doi:10.1046/j.1365-2648.1998.00745.x · Cited by 30
Malyar, N. M., Schlosser, T. W., Barkhausen, J., Gutersohn, A., Buck, T., Bartel, T., & Erbel, R. M. (2007) · Cardiology · doi:10.1159/000105554 · Cited by 30
Kupari, M. M., & Koskinen, P. (1993) · The American Journal of Cardiology · doi:10.1016/0002-9149(93)90989-p · Cited by 30
Evaluation of cardiac biventricular segmentation from multiaxis MRI data: A multicenter study
Lötjönen, J. M. P., Järvinen, V. M., Cheong, B., Wu, E., Kivistö, S., Koikkalainen, J. R., Mattila, J., Kervinen, H., Muthupillai, R., Sheehan, F. H., et al. (2008) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21520 · Cited by 30
In vivo monitoring of microvessels in skin flaps: Introduction of a novel technique
Langer, S., Biberthaler, P., Harris, A. G., Steinau, H., & Messmer, K. (2001) · Microsurgery · doi:10.1002/micr.1058 · Cited by 40
Ishii, M., Kato, H., Kawano, T., Akagi, T., Maeno, Y. V., Sugimura, T., Hashino, K., & Takagishi, T. (1995) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(95)00154-r · Cited by 30
Maternal middle cerebral artery velocimetry in normal pregnancy and postpartum
Serra, V., Kyle, P. M., Chandran, R., & Redman, C. W. G. (1997) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.1997.tb14349.x · Cited by 30
Reproducibility of Cardiopulmonary Exercise Parameters in Patients With Valvular Heart Disease
Lehmann, G., & Kölling, K. (1996) · CHEST Journal · doi:10.1378/chest.110.3.685 · Cited by 30
Carrasco, S., Gonzàlez, R., Jimenez, J. E., Román, R., Medina, V., & Azpiroz, J. (1998) · Journal of Medical Engineering & Technology · doi:10.3109/03091909809032542 · Cited by 31
Automated selection of trabecular bone regions in knee radiographs
Podsiadło, P., Wolski, M., & Stachowiak, G. (2008) · Medical Physics · doi:10.1118/1.2905025 · Cited by 30
Takase, B., Nagata, M., Matsui, T., Kihara, T., Kameyama, A., Hamabe, A., Noya, K., Satomura, K., Ishihara, M., Kurita, A., et al. (2004) · Japanese Heart Journal · doi:10.1536/jhj.45.81 · Cited by 30
Mâaroufi, H., Benbouazza, K., Faïk, A., Bahiri, R., Lazrak, N., Abouqal, R., Amine, B., & Hajjaj-Hassouni, N. (2007) · Spine · doi:10.1097/brs.0b013e318060a63d · Cited by 30
Fluospheres for Cardiovascular Phenotyping Genetically Modified Mice
Richer, C., Domergue, V., Gervais, M., Bruneval, P., & Giudicelli, J. (2000) · Journal of Cardiovascular Pharmacology · doi:10.1097/00005344-200009000-00017 · Cited by 30
Molnár, Z., Umgelter, A., Toth, I., Livingstone, D., Weyland, A., Sakka, S. G., & Meier‐Hellmann, A. (2007) · Intensive Care Medicine · doi:10.1007/s00134-007-0743-7 · Cited by 30
Grebe, O. C., Paetsch, I., Kestler, H. A., Herkommer, B., Schnackenburg, B., Hombach, V., Fleck, E., & Nagel, E. C. (2003) · Journal of Cardiovascular Magnetic Resonance · doi:10.1081/jcmr-120025231 · Cited by 30
Development and psychometric testing of a symptom index for pelvic organ prolapse
Bradshaw, H., Hiller, L., Farkas, A. G., Radley, S. C., & Radley, S. C. (2006) · Journal of Obstetrics and Gynaecology · doi:10.1080/01443610500537989 · Cited by 30
Lohse, L., Uttenthal, Å., Enøe, C., & Nielsen, J. P. (2010) · Acta veterinaria Scandinavica · doi:10.1186/1751-0147-52-29 · Cited by 30
Lehmann, K. G., Gelman, J. A., Weber, M. A., & Lafrades, A. (1998) · The American Journal of Cardiology · doi:10.1016/s0002-9149(98)00080-0 · Cited by 30
Thomas, K. A., Burr, R., Wang, S.-Y., Lentz, M. J., & Shaver, J. (2004) · Biological Research For Nursing · doi:10.1177/1099800403260620 · Cited by 30
Fraenkel, D. J., Purdie, D., Garrett, P., & Venkatesh, B. (2004) · Intensive Care Medicine · doi:10.1007/s00134-003-2102-7 · Cited by 30
Robinson, I. A., McKee, G., & Kissin, M. W. (1995) · Diagnostic Cytopathology · doi:10.1002/dc.2840130315 · Cited by 30
Miura, T., Takeuchi, T., Sato, H., Nishioka, N., Terakado, S., Fujieda, Y., & Ibukiyama, C. (1998) · Japanese Circulation Journal · doi:10.1253/jcj.62.649 · Cited by 30
Relationships of ratings of appetite to food intake in healthy older men and women
Parker, B. A., Ludher, A. K., Tam, K. L., Horowitz, M., & Chapman, I. M. (2004) · Appetite · doi:10.1016/j.appet.2004.05.004 · Cited by 36
MEM and Nott dynamic retinoscopy in patients with disorders of vergence and accommodation
García, Á., & Cacho, P. (2002) · Ophthalmic and Physiological Optics · doi:10.1046/j.1475-1313.2002.00026.x · Cited by 30
A stair-climbing test for ambulatory assessment of children with cerebral palsy
Bar‐Haim, S., Belokopytov, M., Harries, N., & Frank, A. (2004) · Gait & Posture · doi:10.1016/j.gaitpost.2003.09.002 · Cited by 30
Multiparametric ISODATA analysis of embolic stroke and rt-PA intervention in rat
Ding, G., Jiang, Q., Zhang, L., Zhang, Z., Knight, R. A., Soltanian-Zadeh, H., Lü, M., Ewing, J. R., Li, Q., Whitton, P. A., et al. (2004) · Journal of the Neurological Sciences · doi:10.1016/j.jns.2004.05.017 · Cited by 32
Comparison of a New Desktop Spirometer (Diagnosa®) with a Laboratory Spirometer
Maree, D., Videler, E. A., Hallauer, M., Pieper, C. H., & Bolliger, C. T. (2001) · Respiration · doi:10.1159/000050534 · Cited by 30
Regression Models for Method Comparison Data
Dunn, G. (2007) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400701329513 · Cited by 30
Pressure-volume curves with and without muscle paralysis in acute respiratory distress syndrome
Decailliot, F., Demoule, A., Maggiore, S. M., Jonson, B., Duvaldestin, P., & Brochard, L. J. (2006) · Intensive Care Medicine · doi:10.1007/s00134-006-0265-8 · Cited by 30
Complete Blood Cell Count in Psittaciformes by Using High-Throughput Image Cytometry: A Pilot Study
Beaufrère, H., Ammersbach, M., & Tully, T. N. (2013) · Journal of Avian Medicine and Surgery · doi:10.1647/1082-6742-27.3.211 · Cited by 30
Extrapolation to zero-flow pressure in cerebral arteries to estimate intracranial pressure
Bühre, W., Heinzel, F. R., Grund, S., Sonntag, H., & Weyland, A. (2003) · British Journal of Anaesthesia · doi:10.1093/bja/aeg073 · Cited by 30
Papathakis, P. C., Rollins, N. C., Brown, K. H., Bennish, M. L., & Van Loan, M. D. (2005) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/82.3.538 · Cited by 30
Bruynesteyn, K., van der Heijde, D. M. F. M., Boers, M., van der Linden, S., Lassere, M. N., & van der Vleuten, C. P. M. (2004) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2003.10.014 · Cited by 31
Slart, R. H. J. A., Bax, J. J., de Boer, J., Willemsen, A. T. M., Mook, P. H., Oudkerk, M., van der Wall, E. E., van Veldhuisen, D. J., & Jager, P. L. (2005) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-005-1785-0 · Cited by 30
Sandmire, D. A., Vroman, K. G., & Sanders, R. (2000) · PubMed · Cited by 30
Automatic Determination of Aortic Compliance With Cine-Magnetic Resonance Imaging
Lalande, A., Kien, P. K. V., Salve, N., Salem, D. B., LEGRAND, L., WALKER, P. M., Wolf, J., & Brunotte, F. (2002) · Investigative Radiology · doi:10.1097/00004424-200212000-00008 · Cited by 30
Nelesen, R. A., & Dimsdale, J. E. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200210000-00002 · Cited by 30
Chan, J. S., Segara, D., & Nair, P. (2006) · Critical Care and Resuscitation · doi:10.1016/s1441-2772(23)02064-1 · Cited by 30
Deplano, A., Migaleddu, V., Pischedda, A., Garrucciu, G., Gregu, G., Multinu, C., Piredda, M., Tocco, A., Urigo, F., Cossu, P. A., et al. (1999) · Digestive Diseases and Sciences · doi:10.1023/a:1026622721389 · Cited by 30
Tamburrelli, C., Salgarello, T., Caputo, C. G., Giudiceandrea, A., & Scullica, L. (2000) · PubMed · Cited by 30
Mansencal, N., Bouvier, E., Joseph, T., Farcot, J., Pillière, R., Redheuil, A. B., Lacombe, P., Jondeau, G., & Dubourg, O. (2004) · Echocardiography · doi:10.1111/j.0742-2822.2004.03045.x · Cited by 30
Egger, N. G., Carlson, G., & Shaffer, J. L. (1999) · Nutrition · doi:10.1016/s0899-9007(98)00136-1 · Cited by 37
Parr, C. L., Barikmo, I., Torheim, L. E., Ouattara, F., Kaloga, A., & Oshaug, A. (2002) · Public Health Nutrition · doi:10.1079/phn2002357 · Cited by 42
Hall, W. L., Jeanes, Y., Pugh, J., & Lodge, J. K. (2003) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.1263 · Cited by 30
Agreement Between Large And Small Cuffs In Sphygmomanometry: A Quantitative Assessment
Iyriboz, Y., Hearon, C. M., & Edwards, K. (1994) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf02886825 · Cited by 30
Digital Photoplethysmography in the Diagnosis of Suspected Lower Limb DVT: Is It Useful?
Tan, Y. K., & da Silva, A. F. (1999) · European Journal of Vascular and Endovascular Surgery · doi:10.1053/ejvs.1999.0886 · Cited by 30
Arslan, G., Gilja, O. H., Lind, R. A., Florvaag, E., & Berstad, A. (2005) · Scandinavian Journal of Gastroenterology · doi:10.1080/00365520510012163 · Cited by 31
Rodríguez, L., Marquer, B., Mardrus, P., MOLENAT, F., Grand, J. L. L., Reboul, M., & Garrigues, B. (1999) · Intensive Care Medicine · doi:10.1007/s001340050812 · Cited by 30
Lee, D. S., Ahn, J. Y., Kim, S. K., Oh, B., Seo, J. D., Chung, J., & Lee, M. C. (2000) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s002590050025 · Cited by 30
Houzé, P., Borron, S. W., Scherninski, F., Bousquet, B., Gourmel, B., & Baud, F. J. (2004) · Journal of Chromatography B · doi:10.1016/j.jchromb.2004.10.017 · Cited by 30
Review of past and present techniques of measuring corneal topography
Fowler, C. W., & Dave, T. (1994) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.1994.tb00556.x · Cited by 30
The measurement of diastolic blood pressure during pregnancy: Which Korotkoff phase should be used?
López, M. C., Belizán, J. M., Villar, J., & Bergel, E. F. (1994) · American Journal of Obstetrics and Gynecology · doi:10.1016/s0002-9378(94)70230-6 · Cited by 30
Can a Computer Reliably Elicit an Alcohol History?
Bernadt, M., DANIELS, O. J., Blizard, R., & Murray, R. M. (1989) · British Journal of Addiction · doi:10.1111/j.1360-0443.1989.tb00584.x · Cited by 30
Matsukawa, T., Ozaki, M., Hanagata, K., Iwashita, H., Miyaji, T., & Kumazawa, T. (1996) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03013429 · Cited by 30
Rossi, A., Dujardin, K., Bailey, K. R., Seward, J. B., & Enriquez-Sarano, M. L. (1998) · Journal of the American Society of Echocardiography · doi:10.1016/s0894-7317(98)70072-0 · Cited by 30
Limitations of Statistical Measures of Error in Assessing the Accuracy of Continuous Glucose Sensors
Kollman, C., Wilson, D. M., Wysocki, T., Tamborlane, W. V., & Beck, R. W. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.665 · Cited by 34
Monitoring cardiac output using the femoral and radial arterial pressure waveform
de Wilde, R. B. P., Breukers, R.-M. B., Berg, P. C. M. V. D., & Jansen, J. R. C. (2006) · Anaesthesia · doi:10.1111/j.1365-2044.2006.04712.x · Cited by 30
Kuznetsova, T., Staessen, J. A., Stolarz, K., Ryabikov, A. N., Tikhonoff, V., Olszanecka, A., Bianchi, G., Brand, E., Casiglia, E., Dominiczak, A. F., et al. (2004) · Journal of Hypertension · doi:10.1097/00004872-200402000-00012 · Cited by 30
Mason, H. J. (2003) · Occupational Medicine · doi:10.1093/occmed/kqg096 · Cited by 30
Quantitative surface EMG of pericranial muscles. Reproducibility and variability
Jensen, R., Fuglsang‐Frederiksen, A., & Olesen, J. (1993) · Electroencephalography and Clinical Neurophysiology/Evoked Potentials Section · doi:10.1016/0168-5597(93)90078-4 · Cited by 30
Beker, A., Pinchas, A., Erel, J., & Abboud, S. (1996) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1996.tb03277.x · Cited by 30
Evaluation of Breath-by-Breath Measurement of Respiratory Gas Exchange in Pediatric Exercise Testing
Reybrouck, T., Deroost, F., & Hauwaert, L. G. V. D. (1992) · CHEST Journal · doi:10.1378/chest.102.1.147 · Cited by 30
Estimating venous admixture using a physiological simulator
Hardman, J. G., & Bedforth, N. M. (1999) · British Journal of Anaesthesia · doi:10.1093/bja/82.3.346 · Cited by 30
Kuecherer, H. F., Will, M., da Silva, K. G., Borst, M. M., Schwencke, C., Kübier, W., & Strasser, R. H. (1996) · The American Journal of Cardiology · doi:10.1016/s0002-9149(96)00265-2 · Cited by 30
Inter-rater Reliability and Between-days Repeatability of Eight Physical Performance Tests
Horneij, E., Holmström, E., Hemborg, B., Isberg, P., & Ekdahl, C. S. (2002) · Advances in Physiotherapy · doi:10.1080/14038190260501596 · Cited by 31
Pattanapanyasat, K., Lerdwana, S., Noulsri, E., Chaowanachan, T., Wasinrapee, P., Sakulploy, N., Pobkeeree, V., Suksripanich, O., Thanprasertsuk, S., Spira, T. J., et al. (2005) · Clinical and Vaccine Immunology · doi:10.1128/cdli.12.12.1416-1424.2005 · Cited by 30
Delanaye, P., Nellessen, E., Grosch, S., Depas, G., Cavalier, E., Defraigne, J.-O., Chapelle, J.-P., Krzesinski, J.-M., & Lancellotti, P. (2006) · Clinical Transplantation · doi:10.1111/j.1399-0012.2006.00523.x · Cited by 30
Use of the Activated Partial Thromboplastin Time for Heparin Monitoring
Smythe, M. A., Koerber, J. M., Westley, S. J., Nowak, S. N., Begle, R. L., Balasubramaniam, M., & Mattson, J. C. (2001) · American Journal of Clinical Pathology · doi:10.1309/n7ra-d6xn-9eqx-abj9 · Cited by 30
Couris, R. R., Tataronis, G. R., Booth, S. L., Dallal, G. E., Blumberg, J. B., & Dwyer, J. T. (2000) · Journal of the American College of Nutrition · doi:10.1080/07315724.2000.10718081 · Cited by 30
Tuning of pulmonary arterial circulation evidenced by MR phase mapping in healthy volunteers
Laffon, É., Bernard, V., Montaudon, M., Marthan, R., Barat, J. L., & Laurent, F. (2001) · Journal of Applied Physiology · doi:10.1152/jappl.2001.90.2.469 · Cited by 30
Phillips, S., Walker, S., Wyndham, L., & Shaw, J. (1988) · The Medical Journal of Australia · doi:10.5694/j.1326-5377.1988.tb120533.x · Cited by 30
A comparison of alternative assessments of depressive symptom severity: a pilot study
Biggs, M. M., Shores‐Wilson, K., Rush, A. J., Carmody, T. J., Trivedi, M. H., Crismon, M. L., Toprac, M. G., & Mason, M. (2000) · Psychiatry Research · doi:10.1016/s0165-1781(00)00235-3 · Cited by 35
Defining the femoral axis on lateral knee fluoroscopy
Rees, J. L., Price, A. J., Beard, D. J., Robinson, B. J., & Murray, D. W. (2002) · The Knee · doi:10.1016/s0968-0160(01)00132-6 · Cited by 30
Vertical optic disk diameter: discrepancy between planimetric and SLO measurements.
Spencer, A. F., Sadiq, S. A., Pawson, P., & Vernon, S. A. (1995) · PubMed · Cited by 30
Imaging blood vessels of the head and neck.
Sellar, R. J. (1995) · Journal of Neurology Neurosurgery & Psychiatry · doi:10.1136/jnnp.59.3.225 · Cited by 30
Comparison of three methods for measuring respiratory mechanics in ventilated children
Seear, M. D., Wensley, D., & Werner, H. (1991) · Pediatric Pulmonology · doi:10.1002/ppul.1950100412 · Cited by 30
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.