STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 19,901–20,000 of 47,868
The Effect of Insulin Infusion on Capillary Blood Flow in the Diabetic Neuropathic Foot
Flynn, M. D., Boolell, M., Tooke, J. E., & Watkins, P. (1992) · Diabetic Medicine · doi:10.1111/j.1464-5491.1992.tb01858.x · Cited by 23
Graves, M. J., & Dommett, D. (2000) · British Journal of Radiology · doi:10.1259/bjr.73.872.11026856 · Cited by 23
The usefulness of celiteactivated thromboelastography for evaluation of fibrinolysis
Yamakage, M., Tsujiguchi, N., Kohro, S., Tsuchida, H., & Namiki, A. (1998) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03012308 · Cited by 23
Transesophageal echocardiographic measurements of cardiac output in cardiac surgical patients
Izzat, M. B., Regragui, I. A., Wilde, P., Angelini, G. D., & Bryan, A. J. (1994) · The Annals of Thoracic Surgery · doi:10.1016/0003-4975(94)91941-0 · Cited by 23
Desbleds‐Mansard, C., Canet‐Soulas, E., Anwander, A., Chaabane, L., Neyran, B., Serfaty, J., Magnin, I. E., Douek, P., & Orkisz, M. (2004) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10722 · Cited by 24
Jiang, L., Morrissey, R. L., Handschumacher, M. D., de Prada, J. A. V., He, J., Picard, M. H., Weyman, A. E., & Levine, R. A. (1996) · American Heart Journal · doi:10.1016/s0002-8703(96)90536-0 · Cited by 23
Assessment of glomerular filtration rate in small animals by intraperitoneal 99mTc DTPA
Nankivell, B. J., Fawdry, R. M., & Harris, D. C. H. (1992) · Kidney International · doi:10.1038/ki.1992.63 · Cited by 23
Application of Magnetic Resonance Imaging to Evaluation of Femoral Neck Structure in Growing Girls
McKay, H. A., Sievänen, H. T., Petit, M. A., MacKelvie, K. J., Forkheim, K. M., Whittall, K. P., Forster, B. B., & MACDONALD, H. M. (2004) · Journal of Clinical Densitometry · doi:10.1385/jcd:7:2:161 · Cited by 23
Grating and recognition acuities of young amblyopes.
Moseley, M. J., Fielder, A. R., Thompson, J. R., Minshull, C., & Price, D. (1988) · British Journal of Ophthalmology · doi:10.1136/bjo.72.1.50 · Cited by 23
Cowpe, J. G., Ogden, G. R., & Green, M. W. (1993) · Cytopathology · doi:10.1111/j.1365-2303.1993.tb00070.x · Cited by 23
Gerber, T. C., Behrenbeck, T. R., Allison, T. G., Mullan, B. P., Rumberger, J. A., & Gibbons, R. J. (1999) · The American Journal of Cardiology · doi:10.1016/s0002-9149(99)00008-9 · Cited by 23
Couture, P., Denault, A. Y., Carignan, S., Boudreault, D., Babin, D., & Ruel, M. M. (1999) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03012970 · Cited by 23
Shiota, T., Jones, M., Aida, S., Tsujino, H., Yamada, I., Delabays, A., Pandian, N. G., & Sahn, D. J. (1997) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(97)00210-6 · Cited by 23
Ramstad, H., Larsen, S. E., & Aune, T. (2001) · Toxicon · doi:10.1016/s0041-0101(01)00098-8 · Cited by 23
Krebs, W., Klues, H. G., Steinert, S., Sivarajan, M., Job, F. P., Flachskampf, F. A., Franke, A., Reineke, T., & Hanrath, P. (1996) · European Heart Journal · doi:10.1093/oxfordjournals.eurheartj.a015047 · Cited by 23
Wergel‐Kolmert, U., & Wohlfart, B. (1999) · Clinical Physiology · doi:10.1046/j.1365-2281.1999.00160.x · Cited by 23
A digital 3-dimensional method for computing great artery flows: In vitro validation studies
Irvine, T., Li, X.-N., Mori, Y., Wanitkun, S., Li, X., Detmer, P. R., Martin, R. W., Pope, A., Schwartz, G. A., Rusk, R. A., et al. (2000) · Journal of the American Society of Echocardiography · doi:10.1067/mje.2000.106049 · Cited by 23
Pulmonary capillary pressures during the acute respiratory distress syndrome
Nunes, S. L., Ruokonen, E., & Takala, J. A. (2003) · Intensive Care Medicine · doi:10.1007/s00134-003-2036-0 · Cited by 23
Swanstrom, R. I., Bosch, R. J., Katzenstein, D. A., Cheng, H., Jiang, H., Hellmann, N. S., Haubrich, R., Fiscus, S. A., Fletcher, C. V., Acosta, E. P., et al. (2004) · The Journal of Infectious Diseases · doi:10.1086/422692 · Cited by 23
Assessment of respiratory system mechanics by artificial neural networks: an exploratory study
Perchiazzi, G., Högman, M., Rylander, C. C., Giuliani, R., Fiore, T., & Hedenstierna, G. (2001) · Journal of Applied Physiology · doi:10.1152/jappl.2001.90.5.1817 · Cited by 23
Dietz, U., Rupprecht, H., Brennecke, R., Fritsch, H., Woltmann, J., Blankenberg, S., & Meyer, J. (1997) · International journal of cardiac imaging · doi:10.1023/a:1005768523234 · Cited by 23
Elia, M., Jones, M. G., Jennings, G., Poppitt, S. D., Fuller, N. J., Murgatroyd, P. R., & Jebb, S. A. (1995) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1995.269.1.e172 · Cited by 23
Elecsys Testosterone Assay Evaluated
Sänchez‐Carbayo, M., Mauri, M. F., Alfayate, R., Miralles, C., & Soria, F. (1998) · Clinical Chemistry · doi:10.1093/clinchem/44.8.1744 · Cited by 23
Venkatesh, B., Morgan, J., Jones, R. D. M., & Clague, A. E. (1998) · Anaesthesia and Intensive Care · doi:10.1177/0310057x9802600106 · Cited by 23
Renström, A., Gordon, S., Larsson, P. H., Tee, R. D., Taylor, A. J. N., & Malmberg, P. (1997) · Clinical & Experimental Allergy · doi:10.1111/j.1365-2222.1997.tb01177.x · Cited by 23
Sun, J. P., Stewart, W. J., Yang, X. S., Lee, K. S., Sheldon, W. S., & Thomas, J. D. (1995) · Journal of the American Society of Echocardiography · doi:10.1016/s0894-7317(05)80355-4 · Cited by 23
Spencer, D., GREEN, S. E., Evans, J., Pj, P., & Costello, J. (1990) · Clinical & Experimental Allergy · doi:10.1111/j.1365-2222.1990.tb03145.x · Cited by 23
Mario, C. D., Roelandt, J. R. T. C., de Jaegere, P. P. T., Linker, D. T., Oomen, J. A. F., & Serruys, P. W. J. C. (1993) · Catheterization and Cardiovascular Diagnosis · doi:10.1002/ccd.1810280112 · Cited by 23
Anavekar, N. S., Wong, B. F., Foley, T. A., Bishu, K. G., Kolipaka, A., Koo, C. W., Khandaker, M. H., Oh, J. K., & Young, P. M. (2012) · International journal of cardiac imaging · doi:10.1007/s10554-012-0101-x · Cited by 23
Dahl, A. (2009) · University Library of Skövde (University of Skövde) · Cited by 24
Baumstark, A., Jendrike, N., Pleus, S., Haug, C., & Freckmann, G. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0408 · Cited by 22
Goran, M. I., & Khaled, M. A. (1995) · Obesity Research · doi:10.1002/j.1550-8528.1995.tb00187.x · Cited by 22
Naraghi, S., Mutsvangwa, T., Goliath, R., Rangaka, M. X., & Douglas, T. S. (2018) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2018.05.009 · Cited by 22
Automated quantitative evaluation of fetal atrioventricular annular plane systolic excursion
Herling, L., Johnson, J., Ferm‐Widlund, K., Zamprakou, A., Westgren, M., & Acharya, G. (2021) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.23703 · Cited by 22
Temporal and spatial dispersion of human body temperature during deep hypothermia
Opatz, O., Trippel, T., Lochner, A., Werner, A., Stahn, A., Steinach, M., Lenk, J., Kuppe, H., & Gunga, H. C. (2013) · British Journal of Anaesthesia · doi:10.1093/bja/aet217 · Cited by 22
Mitchell, A. E., & Fraser, J. A. (2011) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2010.06.008 · Cited by 22
Exploring Differences in Measurement and Reporting of Classroom Observation Inter-Rater Reliability
Wilhelm, A. G., Rouse, A. G., & Jones, F. (2018) · University of Massachusetts Amherst · doi:10.7275/at67-md25 · Cited by 20
TRANSIT: a GIS toolbox for estimating the duration of ancient sail-powered navigation
Alberti, G. (2018) · Cartography and Geographic Information Science · doi:10.1080/15230406.2017.1403376 · Cited by 22
T2 relaxation mapping MRI of healthy and inflamed gingival tissue
Newbould, R. D., Bishop, C., Janiczek, R. L., Parkinson, C. R., & Hughes, F. J. (2016) · Dentomaxillofacial Radiology · doi:10.1259/dmfr.20160295 · Cited by 22
A Deep Learning Approach for Meibomian Gland Appearance Evaluation
Swiderska, K., Blackie, C. A., Maldonado‐Codina, C., Morgan, P. B., Read, M., & Fergie, M. (2023) · Ophthalmology Science · doi:10.1016/j.xops.2023.100334 · Cited by 24
Milther, C., Winther, L., Stahlhut, M., Curtis, D. J., Aadahl, M., Kristensen, M. T., Sørensen, J. L., & Dall, C. H. (2023) · European Journal of Pediatrics · doi:10.1007/s00431-023-05014-z · Cited by 27
Whitton, C., Collins, C. E., Mullan, B. A., Rollo, M. E., Dhaliwal, S. S., Norman, R., Boushey, C. J., Delp, E. J., Zhu, F. M., McCaffrey, T. A., et al. (2024) · American Journal of Clinical Nutrition · doi:10.1016/j.ajcnut.2024.04.030 · Cited by 30
Dos’Santos, T., Evans, D., & Read, D. B. (2024) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000004785 · Cited by 22
Jiang, J., & Strother, C. M. (2012) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2012.2216542 · Cited by 22
Measurement of children's physical activity using a pedometer with a built-in memory
Trapp, G., Giles‐Corti, B., Bulsara, M. K., Christian, H. E., Timperio, A. F., McCormack, G. R., & Villanueva, K. (2012) · Journal of science and medicine in sport · doi:10.1016/j.jsams.2012.06.011 · Cited by 28
Lifesource XL-18 pedometer for measuring steps under controlled and free-living conditions
Liu, S., Brooks, D., Thomas, S. G., Eysenbach, G., & Nolan, R. P. (2014) · Journal of Sports Sciences · doi:10.1080/02640414.2014.981847 · Cited by 23
Atypical atrial septal defects in children: noninvasive evaluation by cardiac MRI
Beerbaum, P., Parish, V., Bell, A., Gieseke, J., Körperich, H., & Sarikouch, S. (2008) · Pediatric Radiology · doi:10.1007/s00247-008-0977-8 · Cited by 22
DNA fragmentation and sperm head morphometry in cat epididymal spermatozoa
Vernocchi, V., Morselli, M. G., Lange-Consiglio, A., Faustini, M., & Luvoni, G. C. (2014) · Theriogenology · doi:10.1016/j.theriogenology.2014.07.014 · Cited by 22
Horváth, K., Aschermann, Z., Ács, P., Bosnyák, E., Deli, G., Pál, E., Janszky, J., Faludi, B., Késmárki, I., Komoly, S., et al. (2014) · Parkinson s Disease · doi:10.1155/2014/806169 · Cited by 22
Alyahya, K. O., Lee, W. T. K., Al-Mazidi, Z., Morgan, J., & Lanham-New, S. A. (2014) · Archives of Osteoporosis · doi:10.1007/s11657-014-0178-z · Cited by 24
van de Wouw, E., Evenhuis, H. M., & Echteld, M. A. (2013) · Journal of Intellectual & Developmental Disability · doi:10.3109/13668250.2013.816274 · Cited by 26
Li, X. S., Fan, H. X., Fang, H., Song, Y. L., & Zhou, C. W. (2015) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24837 · Cited by 22
Baltes, C., Hansen, M. S., Tsao, J., Kozerke, S., Razavi, R., Pedersen, E. M., & Boesiger, P. (2007) · Radiology · doi:10.1148/radiol.2453061366 · Cited by 22
Kasicka-Jonderko, A. (2011) · World Journal of Gastroenterology · doi:10.3748/wjg.v17.i45.4979 · Cited by 22
Yılmaz, S., Uysal, H. B., Avcil, M., Yılmaz, M., Dağlı, B., Bakış, M., & Ömürlü, İ. K. (2016) · Saudi Medical Journal · doi:10.15537/smj.2016.3.13539 · Cited by 25
Karakochuk, C. D., Whitfield, K. C., Rappaport, A. I., Barr, S. I., Vercauteren, S., McLean, J., Hou, K., Talukder, A., Houghton, L. A., Bailey, K. B., et al. (2016) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2016-0421 · Cited by 22
Alaa, M. H., Abdellatif, M., Tavares-Silva, M., Oliveira‐Pinto, J., Lopes, L., Leite, S., Leite-Moreira, A. F., & Lourenço, A. P. (2016) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00202.2016 · Cited by 22
Precision of recumbent crown-heel length when using an infantometer
Ismail, L. C., Puglia, F. A., Ohuma, E. O., Ash, S., Bishop, D. C., Carew, R. M., Dhaheri, A. S. A., & Chumlea, W. C. (2016) · BMC Pediatrics · doi:10.1186/s12887-016-0725-4 · Cited by 23
Intra‐ and inter‐ examiner repeatability of cycloplegic retinoscopy among young children
McCullough, S. J., Doyle, L. A., & Saunders, K. J. (2016) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12341 · Cited by 22
Ohya, N., Yamada, T., Satoh, Y., & Kawamura, H. (2017) · Journal of Physical Therapy Science · doi:10.1589/jpts.29.754 · Cited by 22
Freckmann, G., Baumstark, A., Jendrike, N. R. M., Rittmeyer, D., Pleus, S., & Haug, C. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0341 · Cited by 23
Megahed, A., Hiew, M. W. H., Townsend, J. R., & Constable, P. D. (2017) · Veterinary Clinical Pathology · doi:10.1111/vcp.12493 · Cited by 23
Ruiz, G. V., Cayuela, J. G., Montes, J. O., Caparrós, M. P., Rojo, B. G., & Albasini, J. L. A. (2017) · Revista Española de Anestesiología y Reanimación · doi:10.1016/j.redar.2017.03.013 · Cited by 22
Validation of a sheet-shaped body vibrometer for screening of obstructive sleep apnea
Kogure, T., Kobayashi, M., Okawa, T., Nakajima, T., & Inoue, Y. (2017) · Drug Discoveries & Therapeutics · doi:10.5582/ddt.2017.01015 · Cited by 22
Improving Rolling Bearing Fault Diagnosis by DS Evidence Theory Based Fusion Model
Yao, X., Li, S., & Hu, J. (2017) · Journal of Sensors · doi:10.1155/2017/6737295 · Cited by 26
de Oliveira, J. K., Montiani‐Ferreira, F., & Williams, D. L. (2018) · Open Veterinary Journal · doi:10.4314/ovj.v8i1.12 · Cited by 22
Freckmann, G., Jendrike, N. R. M., Baumstark, A., Pleus, S., Liebing, C., & Haug, C. (2018) · Diabetes Therapy · doi:10.1007/s13300-018-0392-6 · Cited by 25
The Test-Retest Reliability of New Generation Power Indices of Wingate All-Out Test
Özkaya, Ö., Balcı, G. A., As, H., & Vardarli, E. (2018) · Sports · doi:10.3390/sports6020031 · Cited by 29
A novel approach to skull‐base and orbital osteotomies through virtual planning and navigation
Franz, L., Isola, M., Bagatto, D., Tuniz, F., & Robiony, M. (2018) · The Laryngoscope · doi:10.1002/lary.27479 · Cited by 22
Mandoli, G. E., Cameli, M., Minardi, S., Crudele, F., Lunghetti, S., & Mondillo, S. (2018) · Echocardiography · doi:10.1111/echo.14180 · Cited by 22
Brewster, P. J., Durward, C. M., Hurdle, J. F., Stoddard, G. J., & Guenther, P. M. (2018) · Journal of the Academy of Nutrition and Dietetics · doi:10.1016/j.jand.2018.08.165 · Cited by 23
Differences in muscle contraction onset as determined by ultrasound and electromyography
Tweedell, A. J., Tenan, M. S., & Haynes, C. A. (2018) · Muscle & Nerve · doi:10.1002/mus.26395 · Cited by 24
Hughes, T., Jones, R. K., Starbuck, C., Sergeant, J. C., & Callaghan, M. J. (2019) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2019.01.001 · Cited by 26
Weizman, Y., Tan, A. M., & Fuss, F. K. (2019) · Measurement · doi:10.1016/j.measurement.2019.03.023 · Cited by 22
Chandrakantan, A., Jasiewicz, R., Reinsel, R. A., Khmara, K., Mintzer, J. P., DeCristofaro, J. D., Jacob, Z., & Seidman, P. A. (2019) · Medical Devices Evidence and Research · doi:10.2147/mder.s198707 · Cited by 22
Dart, A. B., McGavock, J., Sharma, A., Chateau, D. G., Schwartz, G. J., & Blydt‐Hansen, T. D. (2019) · Pediatric Nephrology · doi:10.1007/s00467-019-04250-6 · Cited by 24
Howell, D. R., Lugade, V. A., Taksir, M., & Meehan, W. P., III (2019) · The Physician and Sportsmedicine · doi:10.1080/00913847.2019.1632155 · Cited by 26
Ko, J., Wikstrom, E. A., Li, Y., Weber, M., & Brown, C. N. (2019) · Journal of Sport Rehabilitation · doi:10.1123/jsr.2018-0078 · Cited by 22
Moon, Y., Moon, H. S., Kim, D., Kim, J.-M., Lee, J.-K., Shim, W. H., Kim, S., Hwang, G., & Choi, J. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8091419 · Cited by 23
Caballero, Y., Ando, T., Nakae, S., Usui, C., Aoyama, T., Nakanishi, M., Nagayoshi, S., Fujiwara, Y., & Tanaka, S. (2019) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph17010216 · Cited by 23
Automating Spine Curvature Measurement in Volumetric Ultrasound via Adaptive Phase Features
Zhou, G., Li, D., Zhou, P., Jiang, W., & Zheng, Y. (2020) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2019.11.012 · Cited by 22
Barbosa, K. S. S., de Castro, S. S., Leite, C. F., Nacci, F. R., & Accioly, M. F. (2020) · Ciência & Saúde Coletiva · doi:10.1590/1413-81232020253.18992018 · Cited by 33
Angst, C., Eliades, T., & Papageorgiou, S. N. (2020) · European Journal of Orthodontics · doi:10.1093/ejo/cjaa004 · Cited by 22
Alqahtani, B. A., Sparto, P. J., Whitney, S. L., Greenspan, S. L., Perera, S., & Brach, J. S. (2020) · BMC Geriatrics · doi:10.1186/s12877-020-1489-0 · Cited by 23
Intraocular Pressure Reduction After Real-world Cataract Surgery
Leal, I., Chu, C. J., Yang, Y. Y., Manasses, D., Sebastian, R. T., & Sparrow, J. M. (2020) · Journal of Glaucoma · doi:10.1097/ijg.0000000000001527 · Cited by 22
Neu, C., Baumbach, P., Plooij, A. K., Skitek, K., Götze, J., von Loeffelholz, C., Schmidt-Winter, C., & Coldewey, S. M. (2020) · Frontiers in Immunology · doi:10.3389/fimmu.2020.00757 · Cited by 24
Jacobs, K., Rigdon, J., Chan, F. P., Cheng, J. Y., Alley, M. T., Vasanawala, S. S., & Maskatia, S. A. (2020) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-020-00612-4 · Cited by 22
Wu, P., Hsu, P., Chen, T., Huang, J., Chou, C., & Wang, J. (2020) · PM&R · doi:10.1002/pmrj.12432 · Cited by 22
Sager, R., Güsewell, S., Rühli, F. J., Bender, N., & Staub, K. (2020) · PLoS ONE · doi:10.1371/journal.pone.0234552 · Cited by 25
Bianchim, M. S., McNarry, M. A., Larun, L., BARKER, A. R., Williams, C. A., & Mackintosh, K. A. (2020) · Preventive Medicine Reports · doi:10.1016/j.pmedr.2020.101142 · Cited by 24
Kessler, D. A., Mackay, J., McDonald, S., McDonnell, S., Grainger, A. J., Roberts, A. R., Janiczek, R. L., Graves, M. J., Kaggie, J. D., & Gilbert, F. J. (2020) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.27278 · Cited by 22
Chatzistergos, P., Healy, A., Balasubramanian, G. V., Sundar, L., Ramachandran, A., & Chockalingam, N. (2020) · Gait & Posture · doi:10.1016/j.gaitpost.2020.07.011 · Cited by 30
Ding, W., You, T., Gona, P. N., & Milliken, L. A. (2020) · Sports Medicine and Health Science · doi:10.1016/j.smhs.2020.08.001 · Cited by 22
Spatiotemporal Gait Measurement With a Side-View Depth Sensor Using Human Joint Proposals
Hynes, A., Czarnuch, S. M., Kirkland, M. C., & Ploughman, M. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2020.3024925 · Cited by 23
A new assay for quantitative detection of hepatitis A virus
Persson, S., Alm, E., Karlsson, M., Enkirch, T., Norder, H., Eriksson, R., Simonsson, M., & Ellström, P. (2020) · Journal of Virological Methods · doi:10.1016/j.jviromet.2020.114010 · Cited by 24
Christenson, C., Martinez-Vazquez, P., Breidenstein, M. W., Farhang, B., Mathews, J., Melia, U. S. P., Jensen, E. W., & Mathews, D. M. (2020) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-020-00619-3 · Cited by 22
Raiano, L., Tocco, J. D., Massaroni, C., Pino, G. D., Schena, E., & Formica, D. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3043140 · Cited by 24
The Impact of Functional Overreaching on Post-exercise Parasympathetic Reactivation in Runners
Bellenger, C. R., Thomson, R. L., Davison, K., Robertson, E. Y., & Buckley, J. D. (2021) · Frontiers in Physiology · doi:10.3389/fphys.2020.614765 · Cited by 24
Relationship Between the Critical Power Test and a 20-min Functional Threshold Power Test in Cycling
Karsten, B., Petrigna, L., Klose, A., Bianco, A., Townsend, N., & Triska, C. (2021) · Frontiers in Physiology · doi:10.3389/fphys.2020.613151 · Cited by 34
Baral, R. M., Freeman, K. P., & Flatland, B. (2021) · Journal of Feline Medicine and Surgery · doi:10.1177/1098612x20983260 · Cited by 27
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.