STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 35,501–35,600 of 47,868
Lorås, H. W., Sandseter, E. B. H., Storli, L., Kleppe, R., Barnett, L. M., & Sando, O. J. (2024) · Perceptual and Motor Skills · doi:10.1177/00315125241245175 · Cited by 6
Sorkin, N., Zadok, R., Savini, G., Kan‐Tor, Y., Benjamini, Y., Levinger, E., BARDUGO, J., & Abulafia, A. (2024) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2024.04.013 · Cited by 5
Sayee, R. H., Hosamani, M., Krishnaswamy, N., Shanmuganathan, S., Nagasupreeta, S. R., Charan, M. S. S., Sheshagiri, G., Gairola, V., Basagoudanavar, S. H., Sreenivasa, B. P., et al. (2024) · Journal of Virological Methods · doi:10.1016/j.jviromet.2024.114959 · Cited by 5
Russell, J. P., Germain, M. S., Osborn, S. D., Schmitt, T. L., Herrick, K. E. S., & Robeck, T. R. (2024) · Frontiers in Marine Science · doi:10.3389/fmars.2024.1369287 · Cited by 5
Lozano, C. P., Neubig, K., Saha, S., Broyles, S. T., Apolzan, J. W., & Martin, C. K. (2024) · American Journal of Clinical Nutrition · doi:10.1016/j.ajcnut.2024.05.023 · Cited by 5
Marencin, N. C., Edwards, A. A., & Terry, N. P. (2024) · Language Speech and Hearing Services in Schools · doi:10.1044/2024_lshss-23-00134 · Cited by 6
Diktas, H. E., Lozano, C. P., Saha, S., Broyles, S. T., Martin, C. K., & Apolzan, J. W. (2024) · Journal of Nutrition Education and Behavior · doi:10.1016/j.jneb.2024.05.226 · Cited by 7
Validation of the Internet Addiction Test for Adolescents (IAT-A) in the Georgian language
Nadiradze, T., Bakhtadze, S., Kvachadze, I., & Khachapuridze, N. (2024) · European Journal of Paediatric Neurology · doi:10.1016/j.ejpn.2024.06.006 · Cited by 6
Paticcie, T. M. F., José, A., Paiva, L. G., de Oliveira, T. M. D., Pacheco, C. D. R., Silveira, G. W. S., Corso, S. D., Oliveira, C. C., & Malaguti, C. (2024) · Disability and Rehabilitation · doi:10.1080/09638288.2024.2367605 · Cited by 5
The Impact of PEEP on Ventilation Distribution in ARDS
Louis, B., Cour, M., Argaud, L., & Guérin, C. (2024) · Respiratory Care · doi:10.4187/respcare.11569 · Cited by 5
Nieuwenhoff, M. D., Nguyen, H., Niehof, S. P., Huygen, F. J. P. M., Verma, A., Klaassen, E. S., Bechakra, M., Geelhoed, W. J., Jongen, J. L. M., Moll, A. C., et al. (2024) · Muscle & Nerve · doi:10.1002/mus.28213 · Cited by 7
Park, S., Ha, J., & Kim, L. (2024) · Frontiers in Physiology · doi:10.3389/fphys.2024.1325784 · Cited by 6
Moussa, G., Jalil, A., Lippera, M., Alnafisee, N., & Ivanova, T. (2024) · BMJ Open Ophthalmology · doi:10.1136/bmjophth-2023-001531 · Cited by 5
Chowdhury, N. S., Chang, W., Cavaleri, R., Chiang, A. K., & Schabrun, S. M. (2024) · BMC Neuroscience · doi:10.1186/s12868-024-00885-w · Cited by 5
Method for the determination of the luminance of two-photon vision stimuli
Kaczkoś, O., Zielińska, A., Pniewski, J., Wojtkowski, M., & Komar, K. (2024) · Biomedical Optics Express · doi:10.1364/boe.525180 · Cited by 5
Hockett, S. A., Dunbar, S., Williams, C., Sturdivant, R. X., GARNER, B. A., & Rylander, J. H. (2024) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2024.112310 · Cited by 5
Performance of patient-collected dried blood specimens for HIV-1 viral load testing in South Africa
Sahu, M., Schaafsma, T. T., Szpiro, A. A., van Rooyen, H., Asiimwe, S. B., Shahmanesh, M., Krows, M. L., Sithole, N., van Heerden, A. C., & Barnabas, R. V. (2024) · AIDS · doi:10.1097/qad.0000000000004011 · Cited by 5
Hua, A. X., Velasco, C., Muñoz, C., Milotta, G., Fotaki, A., Bosio, F., Granlund, I., Sularz, A., Chiribiri, A., Kunze, K., et al. (2024) · Journal of Cardiovascular Magnetic Resonance · doi:10.1016/j.jocmr.2024.101100 · Cited by 7
The test–retest reliability and validity of food photography and food diary analyses
Curtis, C., Hills, S. P., Arjomandkhah, N., Cooke, C. B., Ranchordas, M. K., & Russell, M. A. (2024) · Nutrition & Dietetics · doi:10.1111/1747-0080.12901 · Cited by 6
Applied Experimental Research on In-Situ Online Monitoring Instrument for Soil Radon in Fault Zone
Zhao, Y., Chen, Z., Wang, C., Hu, X., Liu, Z., & He, H. (2024) · Journal of Environmental Radioactivity · doi:10.1016/j.jenvrad.2024.107546 · Cited by 5
Tolnai, J., Rárosi, F., Tóth, I., Babik, B., Novàk, Z., Peták, F., & Fodor, G. H. (2024) · Scientific Reports · doi:10.1038/s41598-024-75808-0 · Cited by 5
Ferri‐Morales, A., Ando-Lafuente, S., Lírio-Romero, C., Marzetti, E., & Bravo-Esteban, E. (2024) · Sensors · doi:10.3390/s24227343 · Cited by 5
Scientific knowledge role transition prediction from a knowledge hierarchical structure perspective
Yang, J., & Hu, J. (2024) · Journal of Informetrics · doi:10.1016/j.joi.2024.101612 · Cited by 7
Geometry and local wall thickness of abdominal aortic aneurysms using intravascular ultrasound
Fasen, F., Aarle, D. A. C. V., van der Horst, A., van Sambeek, M. R. H. M., & Lopata, R. G. P. (2024) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2024.109514 · Cited by 5
Automated Detection of Change of Direction in Basketball Players Using Xsens Motion Tracking
Pinelli, S., Zinno, R., Jòdar-Portas, A., Prats‐Puig, A., Font‐Lladó, R., & Bragonzoni, L. (2025) · Sensors · doi:10.3390/s25030942 · Cited by 5
Cancellerini, C., Caravelli, A., Esposito, E., Belotti, L. M. B., Soldà, M., Derus, N. R., Merlotti, A., Casadei, F., Mostacci, B., Vignatelli, L., et al. (2025) · Talanta · doi:10.1016/j.talanta.2025.128018 · Cited by 6
Day‐To‐Day Variation in Masseteric Electromyographic Waveforms During the Diurnal Awake State
Nakagawa, Y., Yamaguchi, T., Ishimaru, T., Saito, T., Hattori, Y., Ono, T., Arai, Y., Hasegawa, Y., Shiga, H., Tamaki, K., et al. (2025) · Journal of Oral Rehabilitation · doi:10.1111/joor.14050 · Cited by 5
Zheng, W. L., Zhang, M., Dong, R., Qiu, M., & Wang, W. (2025) · Sensors · doi:10.3390/s25134003 · Cited by 11
The reliability of bilateral cerebral laterality for word generation: Who is left in the middle?
Gerrits, R., & Vingerhoets, G. (2025) · Imaging Neuroscience · doi:10.1162/imag.a.102 · Cited by 5
Quality management and accreditation
Björndahl, L., Mortimer, D., Barratt, C. L. R., Castilla, J. A., Menkveld, R., Kvist, U., Álvarez, J. G., & Haugen, T. B. (2010) · Cambridge University Press eBooks · doi:10.1017/cbo9780511729942.010 · Cited by 5
Latent variable approach to correct errors in radiographic measurements
Tu, Y., Donos, N., Pometti, D., & Nibali, L. (2010) · European Journal Of Oral Sciences · doi:10.1111/j.1600-0722.2010.00781.x · Cited by 5
Cadenas‐Sánchez, C., Arellano, R., Taladriz, S., & López-Contreras, G. (2015) · Journal of Sports Sciences · doi:10.1080/02640414.2015.1046394 · Cited by 5
Álvarez, A. L., & Estrada-Álvarez, J. M. (2009) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas) · doi:10.19052/sv.1072 · Cited by 5
Bhavsar, R. P., Juhl‐Olsen, P., Sloth, E., & Jakobsen, C.-J. (2012) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2012.02655.x · Cited by 5
Designing studies for diagnostic tests
Bakke, P. S. (2008) · The Clinical Respiratory Journal · doi:10.1111/j.1752-699x.2008.00087.x · Cited by 5
Complete blood count at the ED: preanalytic variables for hemoglobin and leukocytes
Carraro, P., Vettore, G., Padoan, A., Piva, E., & Plebani, M. (2015) · The American Journal of Emergency Medicine · doi:10.1016/j.ajem.2015.05.011 · Cited by 6
A glucose meter evaluation co-designed with both health professional and consumer input.
Thompson, H., Chan, H., Logan, F. J., Heenan, H. F., Taylor, L., Murray, C., Florkowski, C. M., FRAMPTON, C. M. A., & Lunt, H. (2013) · PubMed · Cited by 6
The prediction of neck extensor force using surface electromyography
Swaminathan, R., Williams, J. M., Jones, M. D., & Theobald, P. S. (2016) · Journal of Back and Musculoskeletal Rehabilitation · doi:10.3233/bmr-150626 · Cited by 7
Intra- and interobserver repeatability of radiographic measurements in hallux surgery
Schneider, W., Csepan, R., Kasparek, M., Pinggera, O., & Knahr, K. (2002) · Acta Orthopaedica Scandinavica · doi:10.3109/17453670209178034 · Cited by 5
Repeatability of the Universiti Teknologi MARA reading charts
Buari, N. H., Yusuf, N. H., Satali, A. M., & Chen, A. H. (2015) · Bangladesh Journal of Medical Science · doi:10.3329/bjms.v14i3.18964 · Cited by 6
Wittboldt, S., Cider, Å., & Bäck, M. (2015) · European Journal of Cardiovascular Nursing · doi:10.1177/1474515115593840 · Cited by 5
Zang, J., Xu, J., Xiang, C., Zou, S., & He, J. (2015) · Journal of Epidemiological Research · doi:10.5430/jer.v1n1p25 · Cited by 6
Kanda, T., Fujita, M., Iida, O., Masuda, M., Okamoto, S., Ishihara, T., Nanto, K., Shiraki, T., Takahara, M., Sakata, Y., et al. (2015) · Circulation Journal · doi:10.1253/circj.cj-15-0575 · Cited by 5
Joo, G. E., & Sohng, K. (2012) · Journal of Korean Academy of Nursing · doi:10.4040/jkan.2012.42.3.424 · Cited by 5
Antonsen, L., Thayssen, P., Junker, A., Veien, K. T., Hansen, H. S., Hansen, K. N., Hougaard, M., & Jensen, L. O. (2015) · Cardiovascular revascularization medicine · doi:10.1016/j.carrev.2015.08.010 · Cited by 5
Agreement, reliability, and bias in measurement: Commentary on
Beckstead, J. W. (2010) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2010.08.006 · Cited by 5
Kämpe, R., Lind, E., Ståhlberg, F., van Westen, D., Knutsson, L., & Wirestam, R. (2015) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.12288 · Cited by 5
Selvester scoring in patients with strict LBBB using the QUARESS software
Xia, X., Chaudhry, U. A., Wieslander, B., Borgquist, R., Wagner, G. S., Strauss, D. G., Platonov, P. G., Ugander, M., & Couderc, J. (2015) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2015.06.003 · Cited by 5
Midpubic Line to Stage Pelvic Organ Prolapse by MRI Defecography As Compared to POP-Q
Schachar, J. S. (2014) · Obstetrics & Gynecology International Journal · doi:10.15406/ogij.2014.01.00011 · Cited by 6
The S ‐Measurement in the Diagnosis of Canine Hip Dysplasia
Ogden, D. M., Scrivani, P. V., Dykes, N. L., Lust, G., Friedenberg, S. G., & Todhunter, R. J. (2011) · Veterinary Surgery · doi:10.1111/j.1532-950x.2011.00874.x · Cited by 5
Giles, M., Parker, V., Bevan, H., & Wright, I. M. R. (2010) · Journal of Clinical Nursing · doi:10.1111/j.1365-2702.2010.03357.x · Cited by 5
Qu, X., Fang, W., Ye, J., Koh, A. S.-M., Xu, Y., Guan, S., Li, R., & Shen, Y. (2011) · European Journal of Radiology · doi:10.1016/j.ejrad.2011.03.062 · Cited by 5
Kabasakal, L., Türkmen, C., Özmen, Ö., Selçuk, N. A., Önsel, Ç., & Uslu, İ. (2002) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-002-0923-1 · Cited by 5
de Greeff, A., & Shennan, A. H. (2015) · Tropical Doctor · doi:10.1177/0049475515581542 · Cited by 5
Short communication: Evaluation of an automated in-house hematology analyzer for bovine blood
Roland, L., Drillich, M., Fidlschuster, B., Schwendenwein, I., & Iwersen, M. (2014) · Journal of Dairy Science · doi:10.3168/jds.2013-7736 · Cited by 5
Wang, Z., & Pierson, R. N. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/19/021 · Cited by 9
Everett, T. R., Mahendru, A., McEniery, C. M., Lees, C., & Wilkinson, I. B. (2012) · Artery Research · doi:10.1016/j.artres.2012.01.003 · Cited by 5
Balasubramanian, M., Bowd, C., Weinreb, R. N., & Zangwill, L. M. (2010) · Eye · doi:10.1038/eye.2010.124 · Cited by 5
Sipola, P., Niemitukia, L. H., Hyttinen, M. M., & Arokoski, J. P. (2010) · Skeletal Radiology · doi:10.1007/s00256-010-1016-z · Cited by 5
Tsujimura, S., Nakamura, H., Satō, I., Tsuduki, K., Shirahata, T., Yoshida, S., Chubachi, S., Miyazaki, M., Aoki, H., Nakamura, M., et al. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/10/015 · Cited by 5
Repeatability of health economic data in COPD
Nielsen, R., Johannessen, A., Schnelle, H. M., Bakke, P. S., Askildsen, J. E., Omenaas, E. R., & Gulsvik, A. (2008) · Respiratory Medicine · doi:10.1016/j.rmed.2008.06.008 · Cited by 7
Native Acetabular Version: 3D CT Analysis of the Psoas Valley
Osmani, H. T., Henckel, J., Cobb, J. P., & Hart, A. J. (2013) · Hip International · doi:10.5301/hipint.5000007 · Cited by 5
Limits of Agreement Method for Comparison of Pavement Friction Measurement
Izeppi, E. D. L., Flintsch, G. W., & McGhee, K. K. (2012) · Transportation Research Record Journal of the Transportation Research Board · doi:10.3141/2306-22 · Cited by 5
Different Techniques For Body Composition Assessment
Hassan, N. E., El-Masry, S. A., Soliman, N. L., & El-Batran, M. M. (2007) · Journal of Medical Sciences(Faisalabad) · doi:10.3923/jms.2008.15.21 · Cited by 5
Chang, S., Su, P.-F., Lo, A. Y., & Huang, J. (2014) · Taiwan Journal of Ophthalmology · doi:10.1016/j.tjo.2014.08.001 · Cited by 5
Fisher, A. C., McCulloch, D. L., Borchert, M., Garcia-Filion, P. C., Fink, C., Eleuteri, A., & Simpson, D. M. (2015) · Documenta Ophthalmologica · doi:10.1007/s10633-015-9493-y · Cited by 5
Pedometer Accuracy in Elementary School Children while Walking, Skipping, Galloping, and Sliding
Smith, J. D., & Schroeder, C. A. (2010) · Measurement in Physical Education and Exercise Science · doi:10.1080/10913671003715540 · Cited by 8
Takaoka, M., Kubota, Y., Tani, K., Kin, H., Ogura, M., Fujimura, K., Mizuno, T., & Inoue, K. (1993) · Internal Medicine · doi:10.2169/internalmedicine.32.758 · Cited by 5
Effect of electronic compensation on plethysmographic airway resistance measurements
Broughton, S. J., Rafferty, G. F., Milner, A. D., & Greenough, A. (2007) · Pediatric Pulmonology · doi:10.1002/ppul.20661 · Cited by 5
Tseng, K., Tsai, C., Wu, S., Lu, D. V., Hsu, H., Lu, I., & Chu, K. (2009) · Acta anaesthesiologica Taiwanica · doi:10.1016/s1875-4597(09)60037-0 · Cited by 5
Richardson, S., Cooper, A., Alghamdi, G., Alghamdi, M., & Altowaijri, A. M. (2012) · International Journal of Therapy and Rehabilitation · doi:10.12968/ijtr.2012.19.3.163 · Cited by 5
Athanasiou, L. V. (2013) · American Journal of Animal and Veterinary Sciences · doi:10.3844/ajavsp.2013.73.78 · Cited by 5
Rantalainen, T., Silvennoinen, M., Kainulainen, H., & Sievänen, H. T. (2011) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2011.02.010 · Cited by 5
Kasai, T., DePuey, E. G., & Shah, A. A. (2004) · Journal of Nuclear Cardiology · doi:10.1016/j.nuclcard.2003.11.005 · Cited by 5
MR image analysis: Longitudinal cardiac motion influences left ventricular measurements
Berkovic, P., Hemmink, M., Parizel, P. M., Vrints, C. J., & Paelinck, B. P. (2008) · European Journal of Radiology · doi:10.1016/j.ejrad.2008.10.027 · Cited by 5
A new precipitation method with magnetic separation for high-density-lipoprotein cholesterol assay
Iida, S., Osawa, S., & Yonemitsu, H. (1994) · Clinica Chimica Acta · doi:10.1016/0009-8981(94)90284-4 · Cited by 5
Determinants of Ocular Deviation in Esotropic Subjects Under General Anesthesia
Daïen, V., Turpin, C., Lignereux, F., Belghobsi, R., Meur, G. L., Lebranchu, P., & Péchereau, A. (2013) · Journal of Pediatric Ophthalmology & Strabismus · doi:10.3928/01913913-20130226-01 · Cited by 5
A fast and effective method of quantifying myocardial perfusion by magnetic resonance imaging
Aquaro, G. D., Todiere, G., Bella, G. D., Guiducci, L., Pingitore, A., & Lionetti, V. (2013) · International journal of cardiac imaging · doi:10.1007/s10554-013-0220-z · Cited by 5
Schück, O., Teplan, V., Malý, J., Franeková, J., Malínská, H., Štollová, M., Látová, I., Urbanová, J., Skibová, J., & Viklický, O. (2014) · Clinical Nephrology · doi:10.5414/cn108341 · Cited by 5
Reliability of Transabdominal Ultrasound in the Measurement of Prostate Size
Hough, D. M., & List, A. F. (1991) · Australasian Radiology · doi:10.1111/j.1440-1673.1991.tb03047.x · Cited by 5
Ipach, I., Arlt, E.-M., Mittag, F., Kunze, B., Wolf, P., & Kluba, T. (2011) · Hip International · doi:10.5301/hip.2011.8821 · Cited by 5
Colquhoun, D. A., Dunn, L. K., McMurry, T. L., & Thiele, R. H. (2013) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-013-9493-y · Cited by 5
Analysis of the robustness of spectral indices during ventricular fibrillation
Carrión, J. R., Alonso‐Atienza, F., Everss, E., Sánchez-Muñoz, J.-J., Ortiz, M., Garcı́a-Alberola, A., & Rojo‐Álvarez, J. L. (2013) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2013.06.013 · Cited by 5
A Simple, Reliable, and Validated Method for Measuring Brow Position
Verduijn, P. S., Selles, R. W., & Mureau, M. A. M. (2012) · Annals of Plastic Surgery · doi:10.1097/sap.0b013e31826d298e · Cited by 6
Maturational Changes in Skin Color of Japanese Newborn Infants
Watanabe, M., Ohki, Y., Yoshizawa, Y., Inoue, Y., Tokuyama, K., & Morikawa, A. (2007) · Neonatology · doi:10.1159/000100082 · Cited by 5
Low-cost piezoelectric footswitch system for measuring temporal parameters during walking
Balbinot, G. C., Schuch, C. P., Zaro, M. A., & Vaz, M. A. (2014) · International Journal of Engineering & Technology · doi:10.14419/ijet.v3i1.1733 · Cited by 5
Appraising surgeons learning sonography: Measuring measurement variability
Gallagher, P. V., Elliott, S. T., & Charnley, R. (2003) · Journal of Clinical Ultrasound · doi:10.1002/jcu.10185 · Cited by 5
Pallett, E. J., Rentowl, P., & Watson, P. (2012) · Clinical Journal of Pain · doi:10.1097/ajp.0b013e31824781d8 · Cited by 5
Fleischer, J., Charles, M., Tarnow, L., Jensen, K. S., Nygaard, H., Sandbæk, A., & Ejskjær, N. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600110 · Cited by 5
Cui, Y., Johnson, T., Baumbach, A., Reeves, B. C., Rogers, C. A., Angelini, G. D., Marsden, D. A., & Madeddu, P. R. (2012) · PLoS ONE · doi:10.1371/journal.pone.0030389 · Cited by 5
Hodnett, P. A., Glielmi, C. B., Davarpanah, A. H., Scanlon, T. G., Ward, E. V. M., Collins, J. D., Weale, P. J., & Carr, J. (2012) · American Journal of Roentgenology · doi:10.2214/ajr.10.5941 · Cited by 5
Hakacova, N., Bass, G. D., Olson, C. W., Robinson, A. M. C., & Selvester, R. H. S. (2009) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2008.12.008 · Cited by 5
Turgeon, A., Bourdages, M., Levallois, P., Gauvin, D., Gingras, S., Deadman, J. E., Goulet, D. L., & Plante, M. (2004) · Bioelectromagnetics · doi:10.1002/bem.20002 · Cited by 5
Graetz, J. P., Zamunér, A. R., & Moreno, M. A. (2012) · Brazilian Journal of Cardiovascular Surgery · doi:10.5935/1678-9741.20120103 · Cited by 5
Tzamaloukas, A. H., Murata, G. H., Piraino, B., Raj, D. S., VanderJagt, D. J., Bernardini, J. Y., Servilla, K. S., Sun, Y., Glew, R. H., & Oreopoulos, D. G. (2009) · Journal of Renal Nutrition · doi:10.1053/j.jrn.2009.08.004 · Cited by 5
Schaafsma, J. D., Velthuis, B. K., Vincken, K. L., de Kort, G. A. P., Rinkel, G. J. E., & Bartels, L. W. (2014) · European Journal of Radiology · doi:10.1016/j.ejrad.2014.02.001 · Cited by 5
Manual and computerized measurement of coronal vertebral inclination on MRI images: A pilot study
Vrtovec, T., Likar, B., & Pernuš, F. (2013) · Clinical Radiology · doi:10.1016/j.crad.2013.03.008 · Cited by 5
Pakravan, M., Pakbin, M., Mohammad, A. A., Yazdani, S., & Yaseri, M. (2012) · European Journal of Ophthalmology · doi:10.5301/ejo.5000167 · Cited by 5
Kwapiszewski, R., Kwapiszewska, K., Kutter, J. P., & Brzózka, Z. (2014) · Analytica Chimica Acta · doi:10.1016/j.aca.2014.08.042 · Cited by 5
Brennan, E. G., & O'Hare, N. J. (1998) · Physiological Measurement · doi:10.1088/0967-3334/19/3/009 · Cited by 5
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.