STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 39,101–39,200 of 47,868
Positional Differences in Pre-Season Scrimmage Performance of Division I Collegiate Football Players
Early, K. S., Lemoine, N. P., Simoneaux, A. B., Mullenix, S., Marucci, J., MacLellan, M. J., & Johannsen, N. M. (2021) · Sensors · doi:10.3390/s21030769 · Cited by 3
Dahm, V., Nemec, U., Auinger, A. B., Arnoldner, M. A., Kaider, A., Riss, D., Czerny, C., & Arnoldner, C. (2021) · Life · doi:10.3390/life11020156 · Cited by 3
Physiological skin FDG uptake: A quantitative and regional distribution assessment using PET/MRI
Nogami, M., Zeng, F., Inukai, J. I., Watanabe, Y., Nishio, M., Kanda, T., Ueno, Y. R., Sofue, K., Kono, A. K., Hori, M., et al. (2021) · PLoS ONE · doi:10.1371/journal.pone.0249304 · Cited by 3
Anderson, O., Kidd, A. C., Goatman, K. A., Weir, A., Voisey, J. P., Dilys, V., Siebert, J. P., & Blyth, K. G. (2021) · Communications in computer and information science · doi:10.1007/978-3-030-72379-8_7 · Cited by 3
Dong, S.-Y., Yang, Y.-T., Wang, W., Zhu, S., Sun, W., Zeng, M., & Rao, S. (2021) · Acta Radiologica · doi:10.1177/02841851211014194 · Cited by 3
Kim, D. K., Yoon, H., Han, K., Kim, J., Lee, M., Kim, S., Koh, H., Han, S. J., & Shin, H. J. (2021) · Abdominal Radiology · doi:10.1007/s00261-021-03197-3 · Cited by 3
Methodology and applicability of the human contact burn injury model: A systematic review
Springborg, A. H., Wessel, C. R., Andersen, L. P. K., & Werner, M. U. (2021) · PLoS ONE · doi:10.1371/journal.pone.0254790 · Cited by 3
Continuous and automated slug flow nanoextraction for rapid partition coefficient measurement
Payne, E. M., Wells, S. S., & Kennedy, R. T. (2021) · The Analyst · doi:10.1039/d1an01156a · Cited by 3
Lee, J.-S., Kim, S.-R., Hwang, H., & Lee, K. C. (2021) · Imaging Science in Dentistry · doi:10.5624/isd.20210091 · Cited by 3
Casal, L., & Mazzadi, A. N. (2021) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2021.09.006 · Cited by 3
Levator-Urethra Gap: Normative Data in a Nonpregnant Nulliparous Population
Gregory, W. T., Cramer, M. S., Holland, A., & Boniface, E. R. (2021) · Urogynecology · doi:10.1097/spv.0000000000001115 · Cited by 3
Validity and reliability of smartphone-based application for chronic ankle instability
Yoon, T., & Lee, J. (2021) · International Journal of Therapy and Rehabilitation · doi:10.12968/ijtr.2021.0007 · Cited by 3
A Comparative Analysis of Agile Teamwork Quality Models
da Silva, M. M., Freire, A., Perkusich, M., Albuquerque, D. W., Gorgônio, K., de Almeida, H. O., Perkusich, Â., & Guimarães, E. T. (2021) · International Conference on Software, Telecommunications and Computer Networks (SoftCOM) · doi:10.23919/softcom52868.2021.9559062 · Cited by 4
Proposal of a New Specific Test to Assess the Aerobic Performance in Judo
Foresti, Y. F., dos Santos, D. R. A., de Souza, R. A., & Higino, W. P. (2021) · Research Quarterly for Exercise and Sport · doi:10.1080/02701367.2021.1905762 · Cited by 7
Tanino, T., Yufu, K., Shuto, T., Sato, H., Takano, M., Ishii, Y., Kira, S., Saito, S., Kondo, H., Fukui, A., et al. (2021) · Heart and Vessels · doi:10.1007/s00380-021-01992-y · Cited by 3
Roelandt, C. M., Roth, A. M., McWilliams, S. R., & Witt, J. C. (2021) · Journal für Ornithologie · doi:10.1007/s10336-021-01942-5 · Cited by 3
Analytical performance of the new analyzer BA 400 Biosystems with LED technology
Molinario, R., Autilio, C., Daloiso, P. D., Pocino, K., Colacicco, L., Zuppi, C., & Antenucci, M. (2014) · Clinical Laboratory · doi:10.7754/clin.lab.2014.140312 · Cited by 3
Remote assessment of QT interval: A new perspective for implantable cardiac monitors
Carretta, D. M., Troccoli, R., Dentico, A., Vita, M. D., Guerra, A. S., Giacopelli, D., & D’Agostino, C. (2021) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2021.12.005 · Cited by 3
Alomri, R. M., Kennedy, G. A., Wali, S., Alhejaili, F. F., Zelko, M. D., & Robinson, S. R. (2022) · Sleep Medicine · doi:10.1016/j.sleep.2022.01.005 · Cited by 3
Lauria, M., Shaffer, J., Crandall, C. S., & Marinaro, J. L. (2022) · ASAIO Journal · doi:10.1097/mat.0000000000001676 · Cited by 3
Ghavamian, A., Liu, C., Kang, B., Yuan, X., Wang, X., Gao, L., & Zhao, X. (2022) · British Journal of Radiology · doi:10.1259/bjr.20211008 · Cited by 5
What Mathematical Models Are Accurate for Prescribing Aerobic Exercise in Women with Fibromyalgia?
Villafaina, S., Printes, C. B., Parraça, J. A., Brauner, F. D. O., & Tomás-Carús, P. (2022) · Biology · doi:10.3390/biology11050704 · Cited by 6
Validity of a Spanish Version of the Leicester Cough Questionnaire in Children With Cystic Fibrosis
del Corral, T., Percegona, J., López, N., Valiente, A., Garriga, M. J., Seborga, M., & Vilaró, J. (2015) · Archivos de Bronconeumología · doi:10.1016/j.arbr.2015.05.011 · Cited by 5
In vivo accuracy of two intraparenchymal intracranial pressure monitors
Lescot, T., Reina, V., Manach, Y. L., Boroli, F., Chauvet, D., Boch, A., & Puybasset, L. (2012) · Springer eBooks · doi:10.1007/978-3-642-28270-6_45 · Cited by 3
Agreement between Observers: Its Measure on a Quantitative Scale
Marubini, E., Pizzamiglio, S., & Verderio, P. (2005) · The International Journal of Biological Markers · doi:10.1177/172460080502000112 · Cited by 3
Escala de percepção de esforço para criança (EPEC): validação para o português em um teste submáximo
Martins, R., Gonçalves, R. M., Minsky, R. C., Mayer, A. F., & Schivinski, C. I. S. (2020) · Revista Brasileira de Educação Física e Esporte · doi:10.11606/1807-5509202000030513 · Cited by 7
Philippou, E., Demetriou, C. A., Loucaides, G., Solomonidou, N., Critselis, E., Polykarpou, M., Sioulis, S., Hadjisavvas, A., & Kyriacou, K. C. (2022) · Journal of Human Nutrition and Dietetics · doi:10.1111/jhn.13032 · Cited by 4
Généreux, P., Kipperman, R., Disla, J. S. P., Aldaia, L., Koulogiannis, K. P., Marcoff, L., Mediratta, A., Slater, J. P., Zaku, B., Redfors, B., et al. (2022) · Journal of the Society for Cardiovascular Angiography & Interventions · doi:10.1016/j.jscai.2022.100309 · Cited by 3
Kudo, Y., Chu, X., Tanaka, K., & Uchimura, T. (2022) · Journal of Chromatography Open · doi:10.1016/j.jcoa.2022.100050 · Cited by 3
Alsaadi, S. M., Ahmed, R., Alotaibi, K., Alzahrani, M. A., ALOTAIBI, N. E. S., Alahmri, F., & Subbarayalu, A. V. (2022) · Rehabilitation Research and Practice · doi:10.1155/2022/6229847 · Cited by 5
Parry, N. M., Blick, T., Foongkajornkiat, S., Lloyd, T., Thompson, E. W., Hugo, H. J., & Momot, K. I. (2022) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2022.07.004 · Cited by 3
Saito, N., Matsumori, K., Kazama, T., Arakawa, N., & Okamoto, S. (2022) · Cosmetics · doi:10.3390/cosmetics9040086 · Cited by 3
Liu, H., Li, Q., Li, Y., Wang, Y., Huang, Y., Bao, D., Liu, H., & Cui, Y. (2022) · Frontiers in Public Health · doi:10.3389/fpubh.2022.950074 · Cited by 5
Raskovalova, T., Scheffen, L., Jacob, M., Vettier, C., Bulabois, B., Szymanski, G., Chevalier, S., Gonnet, N., Park, S., & Labarère, J. (2022) · PLoS ONE · doi:10.1371/journal.pone.0276095 · Cited by 3
Assessing Agreement When Agreement Is Hard to Assess—The Agatston Score for Coronary Calcification
Andersen, K. P., & Gerke, O. (2022) · Diagnostics · doi:10.3390/diagnostics12122993 · Cited by 6
HYDI-DSI revisited: Constrained non-parametric EAP imaging without q-space re-gridding
Tristán‐Vega, A., Pieciak, T., París, G., Rodríguez‐Galván, J. R., & Aja‐Fernández, S. (2022) · Medical Image Analysis · doi:10.1016/j.media.2022.102728 · Cited by 5
Ogbuagu, N. E., Ayo, J. O., Aluwong, T., & Akor-Dewu, M. B. (2022) · Journal of Thermal Biology · doi:10.1016/j.jtherbio.2022.103445 · Cited by 3
Kurihara, K., Maeda, Y., Sugimura, D., & Hamamoto, T. (2022) · IEEE International Conference on Visual Communications and Image Processing (VCIP) · doi:10.1109/vcip56404.2022.10008871 · Cited by 3
Al-Obaidli, N. M.-N., Alhashimi, N., Lucas, V. S., & ROBERTS, G. J. (2023) · Forensic Science Medicine and Pathology · doi:10.1007/s12024-023-00587-5 · Cited by 3
Regularization Methods for Extracting Doppler Broadening From Hall Thruster Plasma
Rahman, M., & Branam, R. D. (2023) · IEEE Transactions on Plasma Science · doi:10.1109/tps.2023.3263375 · Cited by 3
Brühschwein, A., Schmitz, B., Zöllner, M., Reese, S., & Meyer‐Lindenberg, A. (2023) · Frontiers in Veterinary Science · doi:10.3389/fvets.2023.1019216 · Cited by 3
A Violation of Fitts’ Law is Maintained in Ecologically Valid Settings
Malone, Q., Glazebrook, C. M., & Passmore, S. R. (2023) · Journal of Motor Behavior · doi:10.1080/00222895.2023.2210530 · Cited by 3
Mak, I. L., DeGuire, J. R., Lavery, P., Agellon, S., & Weiler, H. A. (2015) · Journal of Bone and Mineral Metabolism · doi:10.1007/s00774-015-0675-1 · Cited by 3
Filho, A. D. D. R., Filho, R., Fett, C. A., Zavala, A. A. Z., & de Várzea, C. U. (2011) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas) · doi:10.18511/rbcm.v19i2.2183 · Cited by 3
Han, J., Liu, S. M., Guo, W., Bao, Z., Wang, X., Li, L., Liu, Y., Zhuang, D., Li, H., Jia, L., et al. (2015) · PLoS ONE · doi:10.1371/journal.pone.0127696 · Cited by 3
Monitoring of Isoflurane and Desflurane Breakdown: Interfering gases and Infrared Detection
Woehlck, H. J., Dunning, M. B., & Nithipatikom, K. (2000) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1011439530765 · Cited by 3
An Automatic Method for Measurements of Arterial Intima-Media Thickness Using Ultrasonic B-Mode Data
Cinthio, M., Ahlgren, Å. R., Jansson, T. T., Nilsson, T., Lindström, K., & Persson, H. W. (2012) · Acoustical imaging · doi:10.1007/978-94-007-2619-2_12 · Cited by 3
Hein, H. (2006) · KARGER eBooks · doi:10.1159/000093143 · Cited by 3
Zhang, J., Ruch, E., & Bloch, K. E. (2001) · Current Medical Science · doi:10.1007/bf02888032 · Cited by 3
Availability and functionality of sphygmomanometers at health care institutions in Enugu, Nigeria
Chijioke, C. P., Okolo, T., Nwadike, K., Ejim, E., Ekochin, F., Aronu, G., Ghasi, S., & Chijioke, O. (2015) · Nigerian Journal of Clinical Practice · doi:10.4103/1119-3077.156899 · Cited by 3
Liu, J., Piao, J., Sun, R., Tian, Y., & Yang, X. (2011) · PubMed · doi:10.3967/0895-3988.2011.04.006 · Cited by 3
Neto, A. R. T., Botelho, R. G., Sousa, K. E., GOMES, J. L., & LIMA, E. M. M. (2011) · Ars Veterinaria · doi:10.15361/2175-0106.2011v27n1p017-021 · Cited by 3
Fonseca, A. A., García, C. L. A., Collante, M. D. C. B., Patiño, J. P., Santisteban, R. N. R., & Carrascal, Y. T. A. (2014) · IATREIA · doi:10.17533/udea.iatreia.14835 · Cited by 5
Analytical Performance of Specific-Protein Assays on the Abbott Aeroset System
Duly, E. B., Barnes, G., Grimason, S., & Trinick, T. R. (2001) · Clinical Chemistry · doi:10.1093/clinchem/47.9.1709 · Cited by 3
Latex enzyme immunoassay for measuring IgG antibodies to rubella virus.
Duverlie, G., Roussel, C., Driencourt, M., & Orfila, J. (1990) · Journal of Clinical Pathology · doi:10.1136/jcp.43.9.766 · Cited by 3
A capillary whole blood method for measuring the INR
Bamford, E. J., Bowen, R. H., Broad, J. P., Hawken, A., Morgan, J. N., Owen, C. L., Powell, L., Sullivan, B. C., Tollick, H., Wakeman, L. J., et al. (2000) · Clinical & Laboratory Haematology · doi:10.1046/j.1365-2257.2000.00331.x · Cited by 3
Stemp, M., McClements, A., Sykes, P., Chapple, V., & MATSON, P. L. (2013) · Asian Pacific Journal of Reproduction · doi:10.1016/s2305-0500(13)60171-2 · Cited by 3
NOLAN, K. M., Dornelly, S. M., Hughes, D. T., & STRUNIN, L. (1992) · Respiratory Medicine · doi:10.1016/s0954-6111(96)80016-3 · Cited by 3
Accuracy of waste blood measurement in critically ill patients
Bauer, P. R., Kashyap, R., Abraham, R. S., Peikert, T., & Gajic, O. (2011) · Intensive Care Medicine · doi:10.1007/s00134-010-2118-8 · Cited by 3
Hentschel, M., & Brink, I. (2003) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-002-1084-y · Cited by 3
Reliability and validity of an activity monitor (IDEEA®) in a paediatric population
Mackey, A. H., Hewart, P., Walt, S., Moreau, M., & Stott, S. (2006) · Gait & Posture · doi:10.1016/j.gaitpost.2006.11.027 · Cited by 3
Casey, C., Gill, D. S., & Jones, S. P. (2013) · Journal of Orthodontics · doi:10.1179/1465313313y.0000000070 · Cited by 3
Erdıl, T. Y., Dede, F., Tuğlular, S., Şen, F., Jones, T., Farsakoglu, Z., İnanır, S., Ozener, C. I., Akoğlu, E., & Turoğlu, H. T. (2006) · Nuclear Medicine Communications · doi:10.1097/01.mnm.0000195674.75664.ba · Cited by 3
An Overview of Observational Sleep Research with Application to Sleep Stage Transitioning
Caffo, B. C., Swihart, B., Crainiceanu, C. M., Laffan, A. M., & Punjabi, N. M. (2009) · CHANCE · doi:10.1007/s144-009-0002-5 · Cited by 3
Pavlaković, G., Züchner, K., Zapf, A., Bachmann, C. G., Graf, B. M., Crozier, T. A., & Pavlaković, H. (2009) · Muscle & Nerve · doi:10.1002/mus.21270 · Cited by 3
The influence of stimulus parameters on the visual field indices by automated projection perimetry
Dengler‐Harles, M., Wild, J. M., Cole, M. D., & O'Neill, E. C. (1993) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/bf00919030 · Cited by 3
Spencer, L., Alison, J. A., & McKeough, Z. J. (2013) · Physiotherapy · doi:10.1016/j.physio.2013.12.001 · Cited by 4
Sirdar, E., Guimond, J., Coiteux, I., B�lisle, S., Babin, D., Guertin, M., & Denault, A. Y. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2392-4 · Cited by 3
Thermodilution left-sided cardiac output for valve area determination after balloon mitral valvotomy
Gamra, H., Zhang, H. P., Clugston, R. A., Whittaker, P., Allen, J. W., Lau, F., & Ruiz, C. E. (1994) · American Heart Journal · doi:10.1016/0002-8703(94)90592-4 · Cited by 3
Sivarajan, M., Klues, H. G., Krebs, W., Steinert, S., Franke, A., Janssens, U., & Hanrath, P. (1998) · Echocardiography · doi:10.1111/j.1540-8175.1998.tb00662.x · Cited by 3
Bergmann, H., Minear, G., Raith, M., & Schaffarich, P. (2008) · BMC Medical Physics · doi:10.1186/1756-6649-8-6 · Cited by 3
Allier, F., & Monnet, S. (2013) · Kinésithérapie la Revue · doi:10.1016/j.kine.2013.03.013 · Cited by 3
Validity and reliability of body composition assessed by the sulphur hexafluoride dilution method
Demura, S., Yamaji, S., Kobayashi, H., Sato, S., & Nagasawa, Y. (2001) · Journal of Sports Sciences · doi:10.1080/026404102320219365 · Cited by 3
Reliability of Peak Cardiorespiratory Responses During Aquatic Treadmill Exercise
Silvers, W. M., & Dolny, D. G. (2008) · International Journal of Aquatic Research and Education · doi:10.25035/ijare.02.02.06 · Cited by 5
Demura, S., Sato, S., & Kitabayashi, T. (2006) · Journal of Sports Sciences · doi:10.1080/02640410500128148 · Cited by 3
Short-term studies of heart rate variability: comparison of two methods for recording
Migliaro, E. R., Canetti, R. M., Contreras, P., Hakas, M., Eirea, G., & Machado, A. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/6/n01 · Cited by 3
Body composition of Sri Lankan children: Effects of ethnicity
Wickramasinghe, P. (2011) · Sri Lanka Journal of Child Health · doi:10.4038/sljch.v40i3.3506 · Cited by 4
Statistical requirements for reporting nutrition research
Batterham, M. J. (2011) · Nutrition & Dietetics · doi:10.1111/j.1747-0080.2011.01541.x · Cited by 3
Immunoquantitative measurement of soybean aeroallergen emissions at industrial sites.
Cruz, M. J., Villalbí, J. R., Swanson, M. C., Pintó, J. M., Rodrigo, M. J., & Morell, F. (2008) · PubMed · Cited by 3
Kirkegaard-Nielsen, H., Helbo-Hansen, H. S., Severinsen, I. K., Lindholm, P., & Bülow, K. (1996) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.1996.tb04517.x · Cited by 3
QT measurement for TU fused ECG morphology as exhibited during hypoglycaemia
Ireland, R. H., Robinson, R. T. C. E., Heller, S. R., & Harris, N. D. (2002) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.1998.745884 · Cited by 3
Validation de la version française du questionnaire Disability of the Arm, Shoulder and Hand (DASH)
Fayad, F., Lefèvre‐Colau, M., Macé, Y., Fermanian, J., Mayoux-Benhamou, A., Rören, A., Rannou, F., Roby-Brami, A., Gautheron, V., Revel, M., et al. (2008) · Revue du Rhumatisme · doi:10.1016/j.rhum.2007.04.026 · Cited by 3
Wild, J. M., Sims, A. J., Pemberton, J., Irvine, T., Kenny, A., & Murray, A. (2005) · Computers in cardiology · doi:10.1109/cic.2004.1442871 · Cited by 3
Deruelle, F., Grosbois, J., Mucci, P., Bart, F., Lensel, G., & Fabre, C. (2004) · Canadian Journal of Applied Physiology · doi:10.1139/h04-036 · Cited by 4
[Methods of body composition and four compartiments model in obese school children].
Vásquez, F., Díaz-Bautista, E., Lydia Lera, M., Vásquez, L. J. S., Anziani, A., & Burrows, R. A. (2012) · PubMed · doi:10.3305/nh.2012.27.4.5819 · Cited by 7
Comparison Between Isotonic and Hypertonic Saline Solution Inhalation in Patients With Asthma
Cataldo, D. D., Foidart, J.-M., Lau, L. C. K., Bartsch, P., Djukanović, R., & Louis, R. E. (2015) · Cited by 3
Improved functional cardiac MR imaging using the intravascular contrast agent CLARISCAN™
Paetsch, I., Thiele, H., Schnackenburg, B., Bornstedt, A., Müller‐York, A., Schwab, J., Fleck, E., & Nagel, E. C. (2003) · International journal of cardiac imaging · doi:10.1023/a:1025432415983 · Cited by 3
Modified Larsen scoring of digitized radiographs in rheumatoid arthritis.
Young‐Min, S. A., Shakhapur, S., Marshall, N., Griffiths, I. D., Cawston, T. E., & Grainger, A. J. (2003) · PubMed · Cited by 3
de Water, E. V., Oosterlinck, M., Duchateau, L., & Pille, F. J. (2016) · Research in Veterinary Science · doi:10.1016/j.rvsc.2016.03.012 · Cited by 3
Reprodutibilidade dos Parâmetros de Adequação da Hemodiálise Crônica
Manente, M., Thomé, F. S., & Manfro, R. C. (2004) · Brazilian Journal of Nephrology · Cited by 3
Assi, A., Yazbeck, P., Massaad, A., Skalli, W., & Ghanem, I. B. (2013) · Journal of Pediatric Orthopaedics · doi:10.1097/bpo.0000000000000135 · Cited by 3
Point-of-care testing in oral anticoagulation: what is the point?
Katz, B., & Marques, M. B. (2004) · PubMed · Cited by 3
A multi-criteria evaluation platform for segmentation algorithms
Laurent, P., Cresson, T., Vázquez, C., Hagemeister, N., & de Guise, J. A. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7592203 · Cited by 4
Marlatt, K. L., Greenway, F. L., & Ravussin, É. (2016) · European Journal of Clinical Nutrition · doi:10.1038/ejcn.2016.236 · Cited by 3
Mitchell, K. M., PRITCHETT, R. C., Gee, D. L., & Pritchett, K. L. (2017) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000001903 · Cited by 3
Fägerquist, M., Sethi, V., Skytteren, E., & Odén, A. (2017) · Ultrasound · doi:10.1177/1742271x17709854 · Cited by 3
Value Assignment of Vitamin D Metabolites in Vitamin D Standardization Program Serum Samples
Phinney, K. W., Camara, J. E., Tai, S. S.-C., Sander, L. C., Wise, S. A., Grande, L. A. C. D., Thienpont, L. M., Possolo, A. M., Toman, B., Sempos, C. T., et al. (2017) · Journal of AOAC International · doi:10.5740/jaoacint.17-0204 · Cited by 3
Validation of Fourier analysis of videokeratographic data
Sideroudi, H., Labiris, G., Ditzel, F., Tsaragli, E., Georgatzoglou, K., Siganos, C. S., & Kozobolis, V. (2017) · International Ophthalmology · doi:10.1007/s10792-017-0602-7 · Cited by 3
de Antioquia, U., Aristizábal, J. C., Giraldo, A., & de Antioquia, U. (2017) · IATREIA · doi:10.17533/udea.iatreia.v30n3a01 · Cited by 4
Little, R. A., & Spary, D. J. (2017) · Journal of Orthodontics · doi:10.1080/14653125.2017.1371965 · Cited by 4
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.